Académique Documents
Professionnel Documents
Culture Documents
FACULTY OF ENGINEERING
LAB SHEET
*Note: On-the-spot evaluation may be carried out during or at the end of the experiment.
Students are advised to read through this lab sheet before doing experiment. Your
performance, teamwork effort, and learning attitude will count towards the marks.
Page 1 / 20
Simultaneous Capacitance Voltage (CV) Measurement
1. Introduction
1.4 Objectives:
2. CV Measurement System
Page 2 / 20
Simultaneous Capacitance Voltage (CV) Measurement
System Controller
o Installed with General Purpose Interface Bus (GPIB) conforming to
IEEE488.2 for communication between Software and instruments
Page 3 / 20
Simultaneous Capacitance Voltage (CV) Measurement
Figure 2.2
Please ensure that the addresses of the hardware and software are the
same settings.
Two Curves
oQuasistatic CV (low frequency)
oHigh Frequency CV (100kHz or 1MHz)
Figure 3.0 shows a typical simultaneous CV measurement for a p-type
semiconductor of a MOS-CAP structure
Page 4 / 20
Simultaneous Capacitance Voltage (CV) Measurement
Accumulation region
In the accumulation region, the majority carriers will accumulate near the
semiconductor surface.
Capacitance = Oxide Capacitance, Cox
Page 5 / 20
Simultaneous Capacitance Voltage (CV) Measurement
Inversion region
During Inversion, minority carriers will be generated and dominant near
the semiconductor surface. Semiconductor Surface Charge is inverted.
Minority carriers do not respond to high frequency stimulus and
generation are relatively slow.
Capacitance = Oxide Capacitance and Maximum Depletion Layer
Capacitance in Series.
(For High Frequency Only)
Low Frequency Capacitance = Oxide Capacitance
3.2 Characteristic of CV
Page 6 / 20
Simultaneous Capacitance Voltage (CV) Measurement
Page 7 / 20
Simultaneous Capacitance Voltage (CV) Measurement
Threshold Voltage
The Threshold Voltage, VTH is obtained when the surface potential, is S
twice that of the bulk potential, (Please refer to Figure 3.2.1 and 3.2.2).
B
Once the value of CFB is known, the value of VFB is interpolated from the
closest Vgs value. Based on doping, the calculation of C FB uses N at 90%
Wmax, or user-applied NA (bulk doping for p-type, acceptors) or N D (bulk
doping for n-type, donors) The Flatband capacitance is first calculated as
follows:
Page 8 / 20
Simultaneous Capacitance Voltage (CV) Measurement
Page 9 / 20
Simultaneous Capacitance Voltage (CV) Measurement
4. CV Measurement Procedure
4.2 Connection
Please check the system connection with the diagram as in Figure 2.2.
Once the connection is established, power on the CV measurement
system by pushing the power ON / OFF switch of the respective
instruments / equipments:
Turn the vacuum switch to OFF position to disable the vacuum. (Switch
is positioned on the right of the chassis probe enclosure) Place the
assigned experiment wafer on top of the chuck and turn ON the vacuum
Page 10 / 20
Simultaneous Capacitance Voltage (CV) Measurement
switch. Ensure that the wafer is positioned properly on the chuck. Check
the wafer map sheet and select a appropriate site for your CV
measurement from the listed wafer map sheet. Position the tip of the
probe that is connected to the OUTPUT of Keithley model 5951 (adjust
using the X, Y and Z-axis fine adjust) on the DOT wafer that CV
measurement is to be performed. Ensure that the tips are just contacting
the DOT wafer and not scratching it. Similarly, position the tip of the probe
that is connected to the INPUT of Keithley model 5951 (adjust using X, Y
and Z-axis fine adjust) on the top surface of chuck. Once the probes are
properly positioned to the DOT wafer, probe up again for both left and
right probes. (That is, the tips of the probe should not make contact with
the DOT wafer or chuck) Close the probe station enclosure.
Page 11 / 20
Simultaneous Capacitance Voltage (CV) Measurement
Page 12 / 20
Simultaneous Capacitance Voltage (CV) Measurement
From the Setup Editor (See Figure 4.4.2), click on the OUT
instrument icon (That is connected to the Gate,G Terminal), the
Source Setup for KI82-WIN will appear as in Figure 4.4.3.
Depending on the wafer provided for the experiment, select an
appropriate site for the CV measurement that is listed on the wafer
map sheet (As done in procedure 4.3).
From the wafer map sheet, check the Start and Stop Voltage of the
wafer site selected and key the respective values to the Start V
and Stop V accordingly.
Select 20mV as the Step V
Select Mode to Single Stair
Select 100kHz for High Frequency Stimulus
Page 13 / 20
Simultaneous Capacitance Voltage (CV) Measurement
Page 14 / 20
Simultaneous Capacitance Voltage (CV) Measurement
Once the appropriate and necessary setup had been completed from
procedures 4.1 to 4.4, CV measurement can be made.
Once the ZERO CANCEL had been completed, adjust the probes to
contact the wafer and the chuck according to procedure 4.3
Once the probes are in contact with the wafer, close the probe
shield enclosure
Press on the SINGLE button of the Measure Dialogue to begin the
CV Measurement
Once the CV measurement is completed, the results will be
reflected in the Data Sheet View
Page 15 / 20
Simultaneous Capacitance Voltage (CV) Measurement
CH (min) = ____________ pF
Page 16 / 20
Simultaneous Capacitance Voltage (CV) Measurement
Cox = _________________ nF
(Hint: Cox = Maximum Capacitance at High Frequency)
Calculate the oxide thickness (Tox) from the given equation EQ (3.2.1)
Tox = __________________ nm
(Note: Please check the Area, A from the Wafer Map Sheet)
W(90%) = __________________ nm
= _______________________ nm
CFB = _____________________ nF
Extract the Flatband Voltage, VFB at CFB from the CV graph or Data View
Spreadsheet
VFB = _____________________ mV
Page 17 / 20
Simultaneous Capacitance Voltage (CV) Measurement
B = ______________________ mV
VTH = _____________________ V
Note: Nbulk to take at N90%W
5. Questions
Page 18 / 20
Simultaneous Capacitance Voltage (CV) Measurement
Marking Scheme
Page 19 / 20
Simultaneous Capacitance Voltage (CV) Measurement
Page 20 / 20