Safety
To guarantee a safe measurement you must connect the incidents may induce high transient overvoltages, signifi-
protective earth terminal of the enclosure with an approp- cantly above the allowed values specified by ZES ZIMMER,
riate earthing point. Here you have to make sure that the and therefore a safety risk for man and equipment exists.
earthing conductor has a sufficient cross section, depen- For further data concerning safety and application, refer to
ding on the possible short circuit current of the applica- chapter 7 of the sensor manual ZES Sensors and Accessories.
tion. The existing protecting earth connection of the power
supply is not dimensioned for this. The appropriate safety
instructions are strictly to be observed!
We recommend to use high voltage fuses which are used for
voltage transformer protection, and surge arresters. The
high voltage dividers are not suitable to measure on the
grid, as lightning strokes, switching operations and similar
Measurement principle
3-phase I1
L1
system U12 U31 I2 load
(e.g. medium L2 (e.g. motor)
voltage
U23 I3
frequency L3
U10 U20 U30
inverter)
High voltage fuses with high breaking capacity
with nominal current 1 A, designed for voltage
transformer protection. Installation nearby HV line.
Alternatively:
HSTx-3
Connection to power meter
without star-delta conversion U1* U2* U3*
out1
U10/scale U1 U2 U3
out2
U20/scale
out3
U30/scale
On the high voltage side the HST inputs in1, in2, in3 are 2. Instruments without star-delta conversion are connected
connected with lines L1, L2, L3. All voltage measurements as shown in the lower part of the drawing. The line voltages
have the same reference potential: earth. with reference potential earth can be tapped directly at the
BNC jacks. Here all common mode voltages are captured by
The earthed HST can also measure isolated (non-earthed) the measurement.
sources, as these are always bound to earth by their earth
capacities. Examples for the occurrence of common mode voltages are
the DC voltages at complex rectifier circuits and zero-phase
The low voltages at the measuring outputs out1, out2, out3 systems of fundamental and/or harmonics.
can be connected with the precision power meter LMG or By including the common mode voltages in the measure-
other instruments by two different ways: ment the trms values of voltage and of apparent power are
1. Instruments with internal star-delta conversion are con- altered, and the collective sums of trms voltages and appa-
nected as shown in the upper part of the drawing. This has rent power are increased. In both cases the active power is
the advantage that existing common mode voltages in the measured correctly.
3-phase system are not measured because of the difference
measurement.
Technical Data
Series HST3 HST6 HST9 HST12
no. of
1 2 3 1 2 3 1 2 3 1 2 3
channels
ordering type HST3 HST3 HST3 HST6 HST6 HST6 HST9 HST9 HST9 HST12 HST12 HST12
-1 -2 -3 -1 -2 -3 -1 -2 -3 -1 -2 -3
maximum sine
trms voltage for 3.5kV 7kV 10.5kV 14kV
full scale
maximum trms
4.2kV 8.4kV 12.6kV 16.8kV
input voltage
maximum peak volt-
5kV 10kV 15kV 20kV
age for full scale
maximum
4.2kV 8.4kV 12.6kV 16.8kV
trms voltage*
Overvoltage capability maximum periodic
of high voltage input 5kV 10kV 15kV 20kV
peak voltage*
against earthed case
maximum transient
*) voltages in accordance to 3.8kV 6.8kV 8.8kV 10.2kV
overvoltage*
EN 61010:2010, valid for max.
altitude 2000m over sea level non-repetitive
maximum peak 8.8kV 16.8kV 23.8kV 30.2kV
voltage*
one fixed high voltage lead (length 2m) for each channel,
measurement input open leads to be finished by customer,
earth jack as common reference point
Other safety class class I; device must be earthed additionally to PE of mains supply cord!
Subject to technical changes, especially to improve the product, at any time without prior notification. Copyright ZES ZIMMER Electronic Systems GmbH