Vous êtes sur la page 1sur 5

of polonium-210 in the tellurium precipitation. Because po- with the chemical values.

Four measurements by direct


lonium-210 emits alpha particles having energies only 20 keV gamma spectrometry gave a mean of 1.444 f 0.018 (f0.009)%
lower than those of uranium-232, any polonium-210 not re- thorium, also in acceptable agreement. In addition, the mean
moved will add to the uranium-232 peak and make the re- ratio of thorium-228 to thorium-232 for the 20 measurements
covery of tracer appear to be higher than it really is, giving low was 0.996 f 0.024 (f0.005),showing that the measurement
results. For example, the solutions can be centrifuged rather of thorium-228 is no less exact and that the system is in
than filtered if desired, but about 1%of the polonium-tellu- equilibrium so that the gamma spectrometric measurements
rium precipitate will be decanted with the supernate because can be related unequivocally to the thorium-232 concentra-
of a scum at the surface that is not pulled down by centrifu- tion. Although no independent method was available to check
gation. A considerably larger error, and one that is not readily the thorium-230 result in thorium ores, there is no reason to
apparent, is produced in the presence of hydrochloric acid. If believe that its determination will be any less accurate than
5 mL of concentrated hydrochloric acid is used instead of the the other two isotopes.
2 mL of concentrated sulfuric acid specified, the reduction of
the tellurous acid and subsequent agglomeration of the ele- ACKNOWLEDGMENT
mental tellurium takes place so much more rapidly that much The author wishes to express his appreciation for the as-
of the polonium will not have been precipitated and/or in- sistance of his associates, particularly to F. D. Hindman for
cluded in the tellurium carrier by the time the available sur- the thorium and uranium separations, to R. L. Williams for
face of the carrier has been reduced to negligible proportions. the electrodepositions and alpha spectrometry, and to J. S.
The results will be 5 to 10%low. Morton for the gamma spectrometry.
The accuracy of the procedure for determination of the
three main thorium isotopes in thorium ores was verified by LITERATURE CITED
comparing the results obtained for thorium-232 by the present (1)J. N. Rosholt, A. P. Butler, E. L. Garner, and W. R. Shlelds, con. Geol., 60,
alpha spectrometric procedure with those obtained for natural 199 (1965).
(2)J. N.Rosholt, E.N. Harshman, W. R. Shlelds, and E.L. Garner, con. GeoL,
thorium by a fluorometric procedure of previously demon- 59, 570 (1964).
strated accuracy (B), and by direct measurement of the lead- (3)J. Kronfeld, Nucl. Sci. Abstr., 26, 2982 (1969),Abstract No. 31018.
(4)C.W. Sill, "Simultaneous Determlnation of U-238.U-234,Th-230, Ra-226
212 daughter by gamma spectrometry. On 24 measurements and Pb-210 in Uranium Ores, Dusts and Mill Tailings, U. S. Energy Research
of 8 different solid aliquots of a standard monazite ore, the and Development Administration, Health Services Laboratory, Idaho Falls,
fluorometric procedure gave a mean of 1.404 f 0.017 Idaho.
(5) C. W. Sill, K. W. Puphal, and F. D. Hlndman, Anal. Chem., 46, 1725
(f0.004)% thorium, where the first uncertainty is that of an (1974).
individual about the mean and the one in parentheses is that (6)C.W. Sill, Anal. Chem., 46, 1426 (1974).
(7)C.W. Sill and R. L. Williams, Anal. Chem., 41, 1624 (1969).
of the mean itself. Using a value of 2428 f 21 dpm/g of tho- (8)C.W. Slll and C. P. Willis, Anal. Chem., 34,954 (1962).
rium-232 per percent natural thorium, the percentage values (9)C. W. Sill and F. D. Hlndman, Anal. Chem., 46, 113 (1974).
convert to 3409 f 51 ( f 3 1 ) dpm/g of thorium-232. On 20 (10)D. R. Perclval and D. 8. Martin, Anal. Chem., 46, 1742 (1974).
(11)C.W. Silland C. P. Willis, Anal. Chem., 37, 1661 (1965).
measurements on the same 8 aliquots of the sample by the
present alpha spectrometric procedure, a mean of 3421 f 74 RECEIVED
for review October 29,1976. Accepted January 14,
(f16) dpm/g was obtained, in excellent statistical agreement 1977.

Separation and Determination of Nanogram Amounts of Inorganic


Arsenic and Methylarsenic Compounds

Robert S. Braman,* David L. Johnson,' Craig C. Foreback,2 James M. Ammons, and Joseph L. Bricker
Department of Chemistry, University of South Florida, Tampa, Fla. 33620

Arsenate and arsenite ions, methylarsonlcacid, and dimeth- Considerable interest in the environmental chemistry of
ylarslnic acid in aqueous solutions are reduced to arslne and arsenic stems from the toxicity of its compounds and their use
the corresponding methylarslnes respectlvely at pH 1-2 by as silvicides or pesticides. An extensive annotated bibliogra-
sodlum borohydrlde In a reaction chamber. Entrained by He phy on arsenic in the environment has been prepared (1).
carrier gas the arsines are frozen out In a liquid nltrogen cooled Nearly all of this referenced work used analytical methods for
U-trap. Their separation Is accompllshed by volatlilzatlon upon total arsenic with no differentiation of arsenic by chemical
warming the U-tube trap. Arslnes carried out of the trap by the form except for thin-layer and paper chromatography sepa-
carrler gas are passed through a direct current electrlcai ration methods reported by Sachs, Anastasia, and Wells ( 2 ) .
dlscharge. Arsenic atomic emlsslon lines produced in the Methylarsenic acids are important compounds of arsenic
discharge are detected by a recording, scannlng monochro- produced by biomethylation and could play an important role
mator system. Llmlts of detectionfor arsenic are approximately
in its environmental chemistry. Concentrations of arsenic
found in the environment are generally small, several parts
1 ng for each of the arsines.
per billion in sea water to less than 1ppb in many fresh waters
and parts per million in soil.
Prior to the work of Braman and Foreback (3) little had
Present address, State University of New York, College of Envi- been available for the determination of the arsenic forms in
ronmental Science and Forestry. Syracuse, N.Y. 13210.
* Present address, Chemistry Division, Henry Ford Hospital, De- environmental samples. Johnson and Pilson ( 4 ) developed a
troit, Mich. 48202. spectrophotometric method for differentiation of arsenic(II1)

ANALYTICAL CHEMISTRY, VOL. 49, NO. 4, APRIL 1977 621


Table I. Reduction of Arsenic Compounds
agb at (YO a t Reduction Reduction
Molecular form PKa PH 1 PH 4 PH product bP
As(III), Arsenous acid, HAsOz 9.23 1.00 1.00 4 ASH^ -55
As(V), Arsenic acid, H3As04 2.25 (pK1) 0.948 0.017 1-2 ASH^ -55
Methylarsonic acid, CH3AsO(OH)2 2.60 0.975 0.038 1-2 CH3AsH2 2
Dimethylarsenic acid, (CH3)2AsO(OH) 6.19 1.00 1.00 1-2 (CW2AsH 35.6
Trimethylamine," (CH3)3As ... ... ... 1-4 (CH~)~AS 70
Phenylarsonic acid, C6HsAsO(OH)z 3.59 1.00 0.280 1-2 CsH&Hz 148
Or its oxidized form presumed t o be (CH3)sAsO. cy0 is the fraction of the acid in the undissociated form.

for analysis and gives the pKa data and (YO, fraction of arsenic acids
in the undissociated form present at analysis pH conditions.
Arsenous acid or arsenic(II1) ions, dimethylarsinic acid, and oxi-
dized forms of trimethylarsine are partially reduced above pH 4, even
though, considering the reduction potential for sodium borohydride
(-1.24 V), one might predict that inorganic arsenic(V) ions and
methylarsonic acid would also be reduced. A pH 3.5-4.0 condition was
1 ElDiSC'
selected for determination of arsenic(II1) ions in the analysis proce-
dure because arsenic(V) was not reduced in this pH range and because
reduction of arsenic(II1) was more rapid than a t higher pH values.
The reduction of arsenic(V) noted in earlier work (7) was found to
be due to a partial reduction observed when strongly acidic solutions
of arsenic(V) were injected into sodium borohydride solutions or vice
versa. Therefore, it is necessary to buffer samples to the appropriate
pH prior to addition of NaBH4 when arsenic(II1) is to be deter-
mined.
Reduction of arsenic(V) by sodium cyanoborohydride in a pre-
reduction step prior to analysis proved less convenient than using
several treatments with 2% aqueous NaBH4 a t pH 1-1.5. The cya-
noborohydride ion decomposes and produces hydrogen cyanide which
obscures the methylarsine and the dimethylarsine peaks, thus pre-
Figure 1. Sample reaction chamber, U-trap, and detector arrange- venting their analysis in the same sample and at the same time as total
ment inorganic arsenic. Sodium cyanoborohydride was also unsuitable as
a prereducing agent for seawater analyses.
Although sodium borohydride rapidly hydrolyzes at pH 1-1.5, it
apparently has sufficient life to effect reduction of arsenic compounds.
and arsenic(V) anions. Peoples, Lakso, and Lais (5) reported Completeness of arsenic(V) reduction was studied. Results indicated
that over 95% of arsenic(V) was reduced after 4 to 5 additions of 2 mL
a method to some extent capable of distinguishing between each of 2% NaBH4 in both seawater and distilled water. Reduction
inorganic arsenic and organic arsenic acids in aqueous samples of all other arsenic compounds was over 99% complete by the analysis
containing more than 2 pg of arsenic. procedure.
Gas chromatography of methyliodide-treated dimeth- Oxalic acid was found to be the best acid for pH control a t pH 1-1.5.
ylarsinic acid has been reported (6)but this technique has not The use of sulfuric acid to produce pH 1or lower was tried but found
to cause approximately a 2-5% disproportionation in analysis of di-
been applied to the other arsenic compounds. methylarsinic acid. This type of effect was not observed with the
Braman, Justin, and Foreback (7)have reported a method procedures reported here. The analysis process by itself did not cause
for inorganic arsenic based upon continuous detection of vo- the methylation of inorganic arsenic either in standard mixes or in
latilized arsine. The low limits of detection served as a basis natural water samples to which inorganic arsenic has been added.
of the method reported here. Methylarsenic compounds were not demethylated.
T a l m i and Bostick (8) have recently reported based upon Zinc and hydrochloric acid were also found to reduce methylarsonic
acid and dimethylarsinic acid but were not as rapid or as convenient
gas chromatographic separation of sodium borohydride re-
to use as sodium borohydride.
duced arsenic compounds with detection by a microwave Apparatus. The apparatus arrangement used and characteristics
discharge emission type detector. Talmi and Norvell(9) have of the detector and the scanning monochromator system have been
reported a method for inorganic As and Sb based upon gas reported previously (12,13).
chromatographic separation after preparation of triphenyl Arsines are generated in a reaction chamber, trapped out in the
derivatives. E d m o n d s and Francesconi (10) have recently U-trap, and carried through the detector by helium carrier gas in the
apparatus shown in Figure 1.All glassware was constructed locally.
reported a separation similar to the one reported here; t h e y Quartz tubing, 6-mm 0.d. was used to construct the detector chamber.
used a flame type atomic absorption spectrometer as t h e de- Polytetrafluoroethylene (PTFE) tubing, 0.25 inch 0.d. was used to
tection unit. connect parts of the glassware system. The tubing connectors fit
This paper reports i n detail on the analytical method of tightly over 6-mm 0.d. glass tubing, yet could be easily disconnected
B r a m a n and Foreback (3) including improvements since its to facilitate disassembly of the apparatus. The detector was attached
first use in their environmental studies. The method has been to the monochromator by means of a linen-filled Bakelite mounting
adapted to air analysis i n environmental studies ( 3 , l l ) . board.
The reaction chamber was constructed from a ground glass joint
EXPERIMENTAL size 34/35 and could be used with samples up to approximately 70-mL
volume. A side arm injection port was constructed with a 0.25-inch
Reduction Chemistry of Arsenic Compounds. Development of Swagelok nut fitting sealed to the 6-mm 0.d. tube by epoxy cement.
this method proceeded from our finding that the several arsenic acid Medium or coarse frits were used. Heavy hooks were used to secure
compounds may be reduced to arsine or corresponding methylarsines the top and bottom parts of the chamber by means of rubber bands.
in aqueous solution by sodium borohydride. Further, the reduction Ball joints and O-ring joints have also been successfully used in similar
reaction was found to be pH dependent, and related to the pK, of the reaction chamber designs.
arsenic acids, a fact which aids the identification of arsenic compounds The U-trap, 8 inches long each arm, was half-filled with 60-80 mesh
found in samples. Table I shows the reduction conditions now used glass beads.

622 ANALYTICAL CHEMISTRY, VOL. 49, NO. 4, APRIL 1977


Table 11. Reproducibility of Volatilization Timesa T a b l e 111. Calibration C u r v e s f o r Arsenic Compounds
Standards in H2O As(II1) (sodium arsenite)
Methylarsine 9.2 s f 1.2 s.d. N =5
y (cm2) = 0.172 f 0.010 ng (as As) - 0.25 f 0.16
n=5
Dimethylarsine 23.2 s f 2.0 s.d. N =7
As(V) (sodium arsenate)
Lake water analyses y (cm2) = 0.164 f 0.005 ng (as As) - 0.23 f 0.11
Methylarsine 13.0 s f 0.8 s.d. N =3 n=7
Dimethylarsine 24.1 s f 0.9 s.d. N =3 Sodium methylarsonate
Trimethylamine 34.2 s f 1.3 s.d. N =3 y (cm2) = 0.172 f 0.0017 ng (as As) - 0.26 f 0.051
0 Appearance of peak maximum after starting t o heat t h e U- n=6
trap. Dimethylarsinic acid
y (cm2) = 0.165 i 0.0033 ng (as As) - 0.40 f 0.11
n = 5
The addition of a carbon dioxide absorber with a bypass, a 2- to
3-inch long column packed with small sodium hydroxide beads, was
a major improvement removing carbon dioxide interference with
arsine detection. These beads used were not the usual larger pellets trogen level. Retention time reproducibility is not critical to the
but were approximately 2-mm 0.d. obtained from Fisher Scientific analysis.
Co., NaOH No. S612. The carbon dioxide absorber occasionally needs Reagents. Powdered sodium borohydride and sodium methylar-
replacement and must be protected from water vapor. The bypass sonate were obtained from Ventron Corp., Beverly, Mass. Dimeth-
directs methylamines around the sodium hydroxide trap so as to avoid ylarsinic acid was obtained from Research Organic/Inorganic Chemi-
small losses of the methylamines. cal Corp., Sun Valley, Calif. Oxalic acid and other reagent grade
Characteristics of the Separation Method. Differences in boiling chemicals used were obtained from Matheson, Coleman and Bell
points of the arsines suggested separation by a gas chromato- Corp., East Rutherford, N.J. Sodium arsenate and sodium arsenite
graphic-like procedure. were Baker Analyzed Reagents grade. All non-arsenic reagents must
Water vapor, arsines, carbon dioxide, and any other vapors from be analyzed for the presence of traces of the various arsenic com-
the sample or generated in the reduction process are frozen out in the pounds. Methylarsenic acid compounds were analyzed for purity by
U-trap. After removal of the liquid nitrogen flask and upon warming, acidimetric titration. Airco laboratory grade helium has been used
all trapped compounds are evolved from the U-trap generally in order with no purification.
of their boiling points. Reasonably good separation of the arsines is Sodium borohydride used for reductions should be low in carbonate
obtained. The carrier gas is then passed through the detector cell ion content. Commercial aqueous solutions of sodium borohydride
where an arsenic atomic emission line intensity is observed and re- have been found to contain much larger amounts of carbonate ion
corded to provide a higher degree of selectivity. than the powders. Although the carbon dioxide trap prevents inter-
The separation is carried out under non-isothermic conditions. The ference in arsine detection, large amounts of carbon dioxide can in-
liquid nitrogen is removed from the U-tube which is at first allowed terfere in methylarsine detection. Prepared solutions of 2% sodium
to warm up without heating. Any low boiling gases volatilize out of borohydride in distilled water were kept in a closed plastic dispenser
the trap in 10-30 s. Arsine volatilizes out just prior to carbon dioxide bottle to avoid absorption of carbon dioxide from air.
which is always present, at least in small amounts, as an impurity in Standards. Stock solutions of from 500 to 1000 ppm of the several
sodium borohydride. After the carbon dioxide has been vaporized out arsenic compounds were prepared in distilled water. These were
of the U-tube, the wire heater is turned on with the heating rate reasonably stable for several days to approximately one week. Serial
controlled by a laboratory autotransformer. dilutions, usually 1 : l O O of the stock solutions were made. Diluted
A slow heating rate is optimum; U-tubes wound with approximately solutions had to be prepared fresh each time a calibration sample was
48 inches of B & S gauge 24 Chrome1 A wire, approximately 1.65 R per needed; their arsenic content decreased perceptibly in 15-30 min.
foot, were heated by an impressed voltage of 12-20 V from the auto- Microliter range syringes (1-10) were used to sample the diluted
transformer. The optimum heater voltage for each U-tube is deter- standard solutions.
mined by experiment with standards. Generally, the more rapid the Instrument Parameters. The 228.81 nm or 234.98 nm atomic
heating rate, the more rapidly do the methylamines volatilize out of emission lines for arsenic are the most sensitive (7) but the former was
the U-tube, but the separation can become poorer a t high heating used in most work. Each has approximately the same sensitivity if a
rates. The response of the arsines was found to be uninfluenced by 1P28 photomultiplier (PM) tube is used. Slit widths used were from
heating rate with the exception of dimethylarsine. At high heating 100 to 200 pm. Photomultiplier voltages used ranged from 900 to 1050
rates, a 2040% decrease in signal was noted for dimethylarsine at- V. The slit width and P M tube voltage may be adjusted to whatever
tributable to thermal decomposition. After removal of the arsines, voltage range is desired to meet sensitivity. At these settings and with
water finally starts to volatilize out of the trap, quenches the dis- a 40-mA discharge current, output currents of the photomultiplier
charge, and ends the analysis. module were in the 1 X 10-8-1 X A range. A chart speed of 4
The U-trap volatilization separation could be considered to be a inchedmin was used.
programmed temperature gas chromatographic effect on a solid phase Procedure for Inorganic Arsenic(II1). Samples containing from
with a rapid heating rate. The heating rate is greater than 100 "C per 0 to 100 ng of arsenic in up to 70 mL of solution are neutralized, if
minute since the U-trap is heated from -195 to approximately 60 "C necessary, and placed in a clean bubbler. To this is added 1-3 mL of
in 1-2 min. 5% potassium biphthalate. The sample should now be approximately
The half-packed U-trap is an interesting feature. A fully-packed pH 3.5-4. Carrier gas is then allowed to pass through the system at
U-trap did not function well nor did an empty open U-tube. Water 300-350 mL/min for 1-2 min to flush out air. The U-trap is then
vapor condenses and freezes in the up-stream end of the U-tube while connected and cooled by the liquid nitrogen trap. Two milliliters of
the lower boiling arsines and other gases apparently freeze out further 2% sodium borohydride is then added and the system is allowed to
down stream. During the volatilization and analysis step, the lighter stand with the carrier gas passing through it for 5 min. The discharge
gases pass through or out of the packed section before water. Water is turned on 1 min prior to analysis and allowed to warm up. The liquid
vapor apparently is kept from volatilizing out of the U-tube at the nitrogen flask is removed and arsine passes through the carbon
same times as the arsines by the still-cold glass beads. Nevertheless, dioxide trap into the detector. A small fraction of dimethylarsine and
the gas velocity through a fully packed column may be sufficient to trimethylamine may also be detected. The system is then prepared
distribute water vapor almost throughout the entire column. Water for the next analysis. The U-trap is disconnected from the system and
vapor then emerges in time to quench the dimethylarsine and tri- heated while passing carrier gas through it to remove water prior to
methylarsine peaks. reuse.
The effect of carrier gas flow rate on separation of peaks was found Procedure for Arsenic(V). A sample just analyzed for arsenic(II1)
to be important. A flow rate of 300 mL per min was optimum for according to the previous procedure may be further analyzed for ar-
analysis. Considerably lower values gave poorer separation of the alkyl senic(V). The sample is buffered to pH 1-1.5 by adding 5 mL of a
arsines. Table I1 gives typical retention time reproducibility data saturated solution (approximately 10% wt/vol) of oxalic acid in water.
under optimum operating conditions and with the same liquid ni- After assembling the trapping system, four additions of 2 mL of 2%

ANALYTICAL CHEMISTRY, VOL. 49, NO. 4, APRIL 1977 623


Table IV.Reproducibility of Water Analyses
Golf Course Pond (10-mL samples)
Arsenic compound Total found, as As
+
As(II1 V) 64 f 4.2 (s.d.) ng, 6.5% rel, N = 4
Methylarsenic acid 74 f 4.4 (s.d.) ng, 5.9% rel, N = 4
Dimethylarsenic acid trace
Trimethylarsine not detected
Lake Carroll (25-mL samples)
+
As(I1 V) 50 f 5.5 (s.d.) ng, 11%rel, N = 3
Methylarsonic acid 4.39 f 0.23 (sad.)ng, 5.2%rel, N = 3
Dimethylarsenic acid 6.31 f 0.38 (s.d.) ng, 5% rel, N = 4
Trimethylarsine compounds 5.31 f 0.44 (sod.)ng, 8.4%rel, N = 3
DMAA Standard in distilled water, replicates, (25-mL Samples)
20.0 f 0.49 ng (s.d.), 2.5% rel, N = 5

sodium borohydride each are made at 30-8intervals. Carrier gas is Silver and copper in interferences were removable by
passed through the system for 6 min. Arsine in the U-trap is measured passing sample solutions through a cation exchange column
as in the prior procedure. The balance of dimethylarsine and tri- in the ammonium ion form. The Amberlite IR124 (Rohm and
methylamine and all of the methylamine produced will also be de- Haas) exchange resin had no arsenic blank background.
tected.
Procedure for Total Inorganic Arsenic and Alkylarsenic Samples which contain large amounts of surface active
Compounds. It is most convenient to analyze samples for total in- materials may foam excessively. T o prevent this, 1-3 mL of
organic arsenic compounds and alkylarsenic compounds in a single a 0.1% (wt/vol) solution of Anti-Foam B (Technicon product)
procedure. Samples are neutralized, if necessary, and are buffered by may be added to the sample.
addition of 5 mL of oxalic acid reagent. Four 2-mL additions of 2% Antimony does not interfere in the arsenic analysis in a ratio
sodium borohydride are made 30 s apart. After collectionof the arsines
for 6-8 min, the liquid nitrogen trap is removed. Arsine is passed
of up to at least 1001, despite the fact that it is reduced to
through the carbon dioxide trap. The methylarsines bypassed around stibine under the same conditions. If comparatively large
it while heating the U-trap. samples of arsenic, over 1 pg, are run through the detector,
small amounts may deposit in the chamber. This arsenic is
removed by the stibine vapor and can lead to a positive arsenic
signal. Deposition of arsenic in the discharge can also decrease
RESULTS AND DISCUSSION the signal response by blocking the light path. Removal of
Response a n d Limits of Detection. Response curves for deposited arsenic was done by injection of air or ammonia
the four arsenic compounds studied were found to be linear vapors into the discharge while it was operating. Samples
from the limit of detection near 1 ng to the upper limit of containing up to a t least 100 ng of arsenic did not deposit ar-
sample sizes studied, 100 ng (as As). Table I11 gives typical senic in the detector chamber a t an appreciable rate.
calibration curves for four arsenic compounds. The four ar- Identification of Arsenic Compounds. Comparison of
senic compounds give approximately the same calibration retention times for detected compounds to that of methylar-
curve based upon response as arsenic. senic standards is useful for identification. The disappearance
The uncertainty in detecting small amounts of the arsenic of arsenic peaks (by analysis of samples a few angstroms off
compounds in the 1-10 ng range as measured by the noise level of the arsenic emission line) indicates the presence of arsenic
of the discharge response was found to be: 0.2 ng for arse- compounds. The pH effect on reduction of arsenic compounds
nic(II1) and (V), 0.2 ng for methylarsonic acid, and 1 ng for may also be used. Analysis at pH 4-5 will eliminate signals of
dimethylarsinic acid and trimethylamine. Dimethylarsinic inorganic arsenic(V) ions, methylarsonic acid and most of the
acid has a poorer limit of detection, likely because it is evolved dimethylarsinic acid present. Thus, disappearance of these
from the cold trap more slowly than the other arsines and peaks by pH effect is also positive evidence of the presence of
background noise is integrated over a larger time period. arsenic acids. GC-mass spectroscopy has also been used to
Concentration limits of detection calculated using 50-mL positively identify the methyl arsines.
samples range from 0.004 to 0.02 ppb (as arsenic). Applications. The analytical procedures have been used
Interferences. Any volatile compounds which are scrubbed for the analysis of a variety of samples in environmental
out of the reaction chamber and which are evolved out of the chemistry studies. Lakes and rivers in the Tampa Bay, Florida
U-trap prior to water can be an interference if present in area, seawater, human urine, and various types of biological
sufficient amounts to increase the background emission, or materials (3, 14, 15) have been analyzed.
if they quench arsenic emission. I t is possible by wavelength Analysis of air particulate and of filtered air for the arsines
scanning to determine if detected peaks are arsenic line (11, 16) has been successfully carried out. Air particulate on
emission or background or band emission. Fortunately, few glass wool is analyzed after tearing apart the filter pad in a
compounds encountered in environmental samples have the small amount of 0.05 N NaOH. Inorganic arsenic appears to
requisite volatility. None have been observed in any appli- be completely trapped by the Gelman Instrument Co. type
cation of the method. A glass fiber filters used. No apparent volatility of AS203 was
A number of metal ions were tested for interference in the observed in the air analysis work reported (11)as none was
method for inorganic arsenic. The following ions were not an ever observed on the silvered bead column used under each
interference a t 20 ppm: A13+, Cd2+, C r ~ 0 7 ~ Fe3+,
-, Mn2+, filter pad to pick up alkyl arsines.
Ni2+,Pb2+,Zn2+,NO3-, Br-, and HP0d2-. The ions Ag+ and Reproducibility of the procedures for the replicate analyses
Cu2+ at 20 ppm partially inhibit arsine evolution, but had no of water samples is given in Table IV. The standard error of
effect at 2 ppm. Seawater and a number of fresh waters had the technique which is approximately f10% relative is likely
no effect on evolution of the arsines. satisfactory for most trace analysis applications.

624 ANALYTICAL CHEMISTRY, VOL. 49, NO. 4, APRIL 1977


LITERATURE CITED (1 1)D. L. Johnson and R. S.Braman, Chemosphere, 6, 333 (1975).
(12) R. S. Braman and A. Dynako, Anal. Chem., 40, 95 (1968).
(1) "Arsenic in the Environment-An Annotated Bibliography", Oak Ridge (13) R. S.Braman, Anal. Chem., 43, 1462 (1971).
National Laboratory, Oak Ridge, Tenn., ORN-2-EIS-73-16 (1973). (14) D. L. Johnson and R. S.Braman, Deep-sea Res., 22,503 (1975).
(2) R. M. Sachs, F. 8. Anastasia, and W. A. Wells, Proc. Northeast. Weed (15) C. C. Foreback, Ph.D. Thesis, University of South Florida, Tampa, Fla.,
Control Conf., 24, 316 (1970). 1973.
(3) R. S. Braman and C. C. Foreback, Science, 182, 1247 (1973). (16) R. S. Braman, "Arsenic in the Environment" in "Arsenical Pesticides",
(4) D. L. Johnson and M. E. Q. Pilson, Anal. Chim. Acta, 58, 289 (1972). E. A. Woolson, Ed., American Chemical Society, Washington, D.C.,
(5) S. A. Peoples, J. Lakso, and T. Lais, Proc. West. Pharmacol. Soc.,14, 178 1975.
(1971).
(6) C. J. Sonderquist, D. G. Crosby, and J. B. Bowers, Anal. Chem., 46, 155
(1974).
(7) R. S. Braman. L. L. Justen, and C. C. Foreback, Anal. Chem., 44, 2195 RECEIVEDfor review October 1, 1976. Accepted December
(1972). 23,1976. The support of this research by the National Science
(8) Y. Talmi hnd D. T. Bostick, Anal. Chem., 47, 2145 (1975).
(9) Y. Talmi and V . Norvell, Anal. Chem., 47, 1510 (1975). Foundation, RANN program, under grants No. GI-43753, and
(10) J. S.Edmonds and K. A. Francesconi,Anal. Chem., 48, 2019 (1976). AEN 74-14598 A01 is gratefully acknowledged.

Band Broadening Studies Using Parameters for an Exponentially


Modified Gaussian

R. E. Pads1 and L. B. Rogers*


Department of Chemistry, University of Georgia, Athens, Ga. 30602

The effects on chromatographic peak shape of dead volume and, r, a nonsymmetrical contribution due to the exponential
and flow rate have been examined using the standard devlatlon decay. These two width terms are additive to give the second
of the Gaussian component of the peak and the exponential moment or peak variance.
decay constant. For a nonretalned solute, addltlon of dead
Mz = u2 + r2 (2)
volume led to an increase in the standard devlation that was
Independent of flow rate while the decay constant was ln- Sternberg (7)has published a comprehensive review of
versely proportional to the flow. Smaller changes were ob- extra-column broadening and discussed the contributions to
served for a retalned species. the peak variance of input profile, connecting tubing, and
detector time Constants. He distinguished contributions from
mixing chambers, diffusion chambers, and tubing-diameter
A gas chromatographic column is normally assumed to act expansions. Several workers ( I , 2, 8, 9) have published on
as a Gaussian operator, broadening the 6 input into a Gaussian extra-column factors, especially on the effects on column ef-
distribution as it passes through the column. However, pure ficiency of dead volume, such as connecting tubing, fittings,
Gaussian peaks are not found experimentally. This is because and detector volume. Perhaps the most extensive experi-
noncolumn factors such as dead volume, detector time-con- mental study on the effects of dead volume on column effi-
stants, and injection profile convolute the Gaussian distri- ciency has been that by Maynard and Grushka (IO).In that
bution. Schmauch ( I ) , as well as Johnson and Stross ( 2 ) ,have work they showed that pre-column dead volume degraded
shown that detector dead-volume will exponentially modify column efficiency more than post-column dead volume. They
a chromatographic peak. McWilliam and Bolton (3) have also also showed that dead-volume effects were much larger for
shown that time constants of detector-amplifier systems will nonretained species, and that expansions in tubing diameter
exponentially convolute a Gaussian input profile. For these seriously affected column performance even if the expansion
reasons, an exponentially modified Gaussian has been widely occurred after the column.
used as a model for chromatographic peak shapes. The purpose of this study was to briefly examine the effect
Several workers (4-6) have applied an exponentially of pre-column dead volume on the values of c and r for a
modified Gaussian as a model in the least-square fitting and chromatographic peak. Post-column effects were not exam-
deconvolution of chromatographic peaks. An exponentially ined because they were reported to be small (10).
modified Gaussian is generated by the following integral: EXPERIMENTAL
N Reagents. Helium (Selox, Inc.) was used as the carrier gas after it
f(t) = --
r u v z had been purified by passage over a molecular sieve. Methane and
n-pentane (spectrophotometric grade, Aldrich Corp., East Ruther-
ford, N.J.) were used as solutes. Chromosorb W, 100/120 mesh, and
SE-30 silicone stationary phase were obtained from Alltech Associates
(Arlington Heights, Ill.).
where N is the peak amplitude, t~ is the center of gravity of Apparatus. A large chromatographic oven described earlier (11)
the Gaussian component, u is the standard deviation of the was used in this study. The temperature in the oven was controlled
Gaussian, r is the time constant of the exponential decay and by a Thermatrol proportional controller (Hallikainen Instrument Co.,
t' is a dummy variable of integration. The width of an expo- Richmond, Calif.). The oven temperature was measured using a
nentially modified Gaussian has two components: u, a sym- platinum resistance thermometer (Omega Engineering Inc., Stamford,
metrical component due to the original Gaussian distribution Conn.) in conjunction with a digital multimeter (Keithley Instru-
ments, Cleveland, Ohio).
Carrier gas was fed to the sampling valve through 1.5 m of capillary
tubing so that back-diffusion of the sample into the gas supply would
Present address, Amoco Research Center, P.O. Box 400, Naper- be negligible. Carrier gas flow was controlled by a Millaflow pressure
ville, Ill. 60540. regulator (Veriflow Corp., Richmond, Calif.) and a manual flow

ANALYTICAL CHEMISTRY, VOL. 49, NO. 4, APRIL 1977 625

Vous aimerez peut-être aussi