Vous êtes sur la page 1sur 7

The British Journal of Radiology, 72 (1999), 186^192 E 1999 The British Institute of Radiology

Report of an image quality and dose audit according to


Directive 97/43/Euratom at Spanish private
radiodiagnostics facilities
1
L GONZALEZ, PhD, 1E VANO, PhD, 2S OLIETE, MD, 3J MANRIQUE, MD,
4
J M HERNAEZ, MD, 4J LAHUERTA, MD and 4J RUIZ, MD
1
Medical Physics Group, Radiology Department, Complutense University, 28040 Madrid, 2Diagnostic
Radiology Service, Hospital Ruber Internacional, 28034 Madrid, 3 Diagnostic Radiology Service, Hospital
12 de Octubre, 28041 Madrid, and 4SANITAS, S.A. de Seguros, Medical Division, Serrano 88, 28006
Madrid, Spain

Abstract. An audit of Spanish private medicine radiodiagnostics facilities has been carried out,
based partly on Spanish legislation relating to European Directives on health protection against
ionizing radiation risks in medical exposure. The study included an appraisal of infrastructure
and equipment, and aspects of quality assurance and radiation protection, by means of data col-
lected through surveys. Of the 51 centres audited, a sample of 24 X-ray rooms was chosen, then an
external evaluation with regard to image quality and patient dose was performed, by an advisory
board of radiologists and medical physicists. The methodology used was similar to that of the group
of European Union experts in European dose evaluation and image quality trials. Chest, abdomen,
lumbar spine and breast examinations were monitored. Doses were measured with thermolumines-
cent dosimeters. A third of the X-ray rooms evaluated reached or exceeded dose reference values,
and in a third of the cases the image quality left considerable room for improvement. Breast and
chest examinations showed themselves to be the hardest to perform, not only as a result of exceed-
ing the reference doses, but also due to failure to meet good image quality standards.

The Council of the European Union (EU) intraoral dental radiology the value of 7 mGy is
Directive 84/466/EURATOM [1], recently proposed, in line with the International Atomic
replaced by 97/43/EURATOM [2], was implemen- Energy Agency (IAEA) recommendations [7]. A
ted in Spanish legislation by Royal Decrees 1132/ centre that performs examinations using other
1990 [3], 1891/1991 [4] and 2071/1995 [5]. The rst than these projections may assess the quality of
of these, which lays down basic measures for pro- its images by means of anatomical criteria devel-
tecting persons undergoing medical examinations oped for the purpose by its own eld radiologists,
or treatment with radiation sources, is to be and may set the rst average dose level found as the
applied to X-ray, radiotherapy and nuclear medi- ``provisional reference value of the centre'' until
cine facilities, whereas the other two aect radio- such time as it has representative reference values
diagnostics only. for the examinations concerned. The dose evalua-
The last two Royal Decrees govern the installa- tions have to be based on 10 estimates for simple
tion and use of X-ray equipment for medical diag- examinations or on ve for complex examinations
nostic purposes with regard to technical (more than four images with possible use of uoro-
specications in the design of an installation and scopy). CT requires dose determination on the sur-
provision of radiation protection resources, face of the patient, in the region scanned, or
amongst other aspects. In addition, they lay down related quantities (computed tomography dose
quality criteria for verifying X-ray images, on a index (CTDI), multislice averaged dose (MSAD),
compulsory annual basis, and their adaptation to etc.), as well as recording the characteristic para-
European quality standards, as the basis of a qual- meters (number of slices, thickness, etc.). The
ity assurance plan for gradual implementation. image rejection rate should also be recorded. An
Under this legislation, the recommended image excess in the mean values of 25% over the reference
quality and reference dose criteria are those of values requires investigation of the causes, with
document EUR 16260 [6] for chest, skull, lumbar correction of the protocols and/or suitable repair
spine, pelvis and urinary tract examinations. For or reconditioning of the equipment in a maximum
period of 2 months, followed by a subsequent
Received 8 April 1998 and in revised form 18 September check on the ecacy of the changes that have taken
1998, accepted 20 October 1998. place.

186 The British Journal of Radiology, February 1999


Image and patient dose QC in Spanish private X-ray centres

University of Madrid. An Advisory Committee,


comprising experienced radiodiagnostics specia-
lists, was formed to evaluate the image quality.
The audit was carried out in two stages.
(1) Evaluation of infrastructure and basic qual-
ity aspects of the services. The starting point was a
clinical audit survey model prepared by SANITAS
on infrastructure and equipment, which was sup-
plemented with quality assurance and radiation
protection aspects regulated in the legislation.
The survey consisted of two parts, one to be com-
pleted by an external auditor, who gathered the
immediately accessible information, and the other
to be sent by post and completed by the person in
charge of the centre, which enabled documents and
technical details to be obtained.
Both parts were used to create a database
on such aspects as accessibility to the centre for
handicapped patients, comfort, existence of ade-
quate technical means, percentage of examinations
supplied with the corresponding written reports,
renovation and maintenance of the equipment,
number of radiologists in relation to hours avail-
able for patient appointment, presence and train-
ing of sta and level of computerization of the
centre etc., as well as the degree of compliance with
Figure 1. Averaged values of entrance surface dose guidelines laid down by the directors of the centre,
obtained from the individual dose values of each centre characteristic doses imparted in examinations,
(represented by a bar), for chest and breast examina- when available etc. The survey covered 51 centres.
tions. (2) From the centres which supplied complete
information and agreed to collaborate with the
The application of this legislation in public hos- programme (about 75% of the total), a selection
pitals was immediate, although its actual imple- was made of 24 X-ray rooms from 12 centres, as a
mentation has been gradual. In private X-ray sample for an external appraisal on image quality
facilities, however, the process has been slower and patient dose, supplementary to the one in the
due to, among other reasons, a rather slow distri- previous paragraph. Thus, the bias that inevitably
bution of information and awareness of the appears as a result of either self-evaluation or dis-
requirements, and to possible economic implica- crepancies attributable to dierent methods of
tions as a result of expenses derived from changes measuring were corrected. Rooms were selected
or repairs of faulty equipment. In addition, there is by the Advisory Committee taking into account
still no surveillance mechanism to check the said the type of examinations and workload to assure
items by the Health Authority. samples of patients as required by the standard in
The above legislation has been used by a private the examinations evaluated.
insurance company, ``SANITAS, S. A. de This stage was planned with a methodology similar
Seguros'', to audit the infrastructure and general to that used by the group of experts of the European
quality of its aliated radiodiagnostics centres. Commission (EC) in its second European dose
Furthermore, the company has carried out an evaluation and image quality trial [9], assessing chest
evaluation of patient dose and image quality in a posteroanterior (PA), abdomen anteroposterior
signicant sample of centres and X-ray rooms. (AP), lumbar spine AP and breast craniocaudal
Taking into account the data available on private (CC) projections. Doses were monitored with lithium
radiology in Madrid [8], this action potentially uoride (LiF) thermoluminescent dose meters
aects over 200 000 patients a year. This study (TLDs), type TLD-100, from Harshaw TLD/
describes the experimental approach and the Bicron/NE-Technology (BICRON-NE, Solon, OH,
results obtained. USA), individually calibrated, in a sample of ve
patients per room.The images obtained were initially
evaluated by the radiologists ofeach centre according
Method to the image quality criteria of document EUR 16260
The audit was carried out in conjunction with [6]. Then, these images, their quality reports and the
the Medical Physics Group of the Complutense dosemeters used were collected for evaluation by the

The British Journal of Radiology, February 1999 187


L Gonzalez, E Vano, S Oliete et al

Table 1. Summary of ndings in relation to image quality and the radiation doses measured

Examination Total Image quality Measured dose


number
of rooms Defective Fair Good Very good High Reference Low
checked

Chest 5 1 2 1 1 2 2 1
Lumbar spine 8 1 4 3 2 3 3
Abdomen 3 1 2 3
Breast 8 1 3 2 2 3 1 4

Advisory Committee, and returned to the centre the rooms where every one of the images supplied
next day. The decision to monitor ve instead of 10 met practically all the criteria and the radiologists
exposures, as in the case of European scale trials, on the panel of experts considered them of a very
was taken to audit centres with a small workload, good quality overall. Since the average dose values
thereby avoiding the need to store images for several are not based on 10 measurements, the dose classi-
days. In this way, the usual process of providing the cation ``High'' is conferred on X-ray rooms above
images to the patients (immediate in private practice the reference value and in the band that is indica-
in Spain) did not suer any signicant delay. tive of anomalies, according to Spanish guidelines
Furthermore, we were able to assure the availability (25% above European reference levels), ``Reference
of images 24 h subsequentto their evaluation, bearing value'' is applied to cases with that value or in a
in mind the limited capacity to analyse large numbers band of 10% around it, and ``Low'' doses are those
of images by the board of experts. clearly below the reference value.

Results Discussion
Evaluation of the survey questionnaires showed Since the results are obtained from a small num-
that the facilities have a satisfactory level of equip- ber of measurements, the data should be
ment and material resources. It was also important approached with caution. In addition, the partici-
to verify that all the sta taking the X-ray images pation of some centres with a small number of
were duly accredited, with academic qualications patients (members of SANITAS), adds a further
at medical or technical levels, plus accreditation on degree of uncertainty, as some images come from
radiation protection from the Nuclear Safety individuals whose weight and measurements do
Council (the Spanish regulatory body). At the pre- not lie completely within the standard of the
sent time, no periodical retraining programmes are EUR 16260 document [6]. Bearing in mind these
followed, since this is not yet a compulsory provisos, a third of the X-ray rooms investigated
requirement. produce images of a quality that is merely fair or
The second stage evaluation enabled results to below the threshold of acceptability, while almost
be obtained with regard to the level of compliance a third generate them with doses that are clearly
with image quality and patient dose guidelines. above the reference value, according to Table 1.
Figure 1 shows examples of the average radiation Thus, both groups would benet considerably
doses used to produce the images, for chest and from an optimization programme.
breast examinations. Figures 2a and b allow us to These ndings underline the urgent need to
compare the evaluations of the board of experts implement a quality assurance programme. In
with those of the eld radiologists regarding how any case, since image quality can be appraised by
each image quality criterion is fullled in chest inspection and the centres may be assumed to have
and breast examinations. made an eort to send their best images, the situa-
The ndings with regard to the doses and quality tion of the X-ray rooms with poor image quality is
of the images are summarised in Table 1. The clas- qualitatively more acceptable, compared with
sication ``Defective'' refers to rooms which pro- those that generate doses above the reference
vided images regarded by the expert board as not value, as doses are not ``seen'' and discrepancies
suitable for diagnosis; ``Fair'' is assigned to rooms from the reference value cannot be corrected. In
contributing images that permit diagnosis, even fact, the written evaluations made by the radiolo-
though they fail to meet signicant quality criteria; gists of each centre with their own images, in accor-
``Good'' refers to rooms with images which, dance with the image quality criteria of the EUR
besides permitting diagnosis, meet the most rele- 16260 document [6], show that, barring special
vant quality criteria in all the images supplied, situations, the specialist identies some unsatisfac-
and the classication ``Very good'' is reserved for tory image details in his images, even in the cases in

188 The British Journal of Radiology, February 1999


Image and patient dose QC in Spanish private X-ray centres

(a) (b)

Figure 2. Graphs showing responses for each image quality criterion, according to the opinions of the eld radiolo-
gists and the board of experts, for (a) chest PA and (b) breast CC projections. The following image criteria are taken
from reference [6], although, for chest examinations, they are presented as in reference [9] for an easy comparison
with results of the 1991 EC trial. Chest: 1, performed at full inspiration and with suspended respiration; 2, symmetrical
reproduction of the thorax; 3, medial border of the scapulae to be outside of the lung elds; 4, reproduction of the
whole rib cage above the diaphragm; 5, visually sharp reproduction of the vascular pattern in the whole lung, particu-
larly the peripheral vessels; 6.a, visually sharp reproduction of the trachea and proximal bronchi; 6.b, visually sharp
reproduction of the borders of the heart; 6.c, visually sharp reproduction of the aorta; 7.a, visually sharp reproduction
of the diaphragm; 7.b, visually sharp reproduction of the lateral costophrenic angles; 8, visualization of the retrocar-
diac lung and the mediastinum. Breast: 1, visually sharp reproduction of pectoralis muscle at image margin; 2, visually
sharp reproduction of retroglandular fat tissue; 3, visually sharp reproduction of medial breast tissue; 4, visually
sharp reproduction of lateral glandular tissue; 5, no skinfolds seen; 6, symmetrical images of left and right breast;
7, visualization of skin outline with bright light; 8, reproduction of vascular structures seen through most dense par-
enchyma; 9, visually sharp reproduction of vessels and brous strands and pectoralis muscle margin (absence of
movement); 10, visually sharp reproduction of skin structure (rosettes from pores) along the pectoralis muscle.

which the overall image quality is considered values are relatively high. On the other hand, for
appropriate. In this way, this audit has fostered chest and breast examinations the reference dose
the optimization of the images, with the result that values are often exceeded, even above the thresh-
only three X-ray rooms have produced images that old, indicating major anomalies according to
are clearly decient. On the other hand, an anom- Spanish guidelines. In mammography we should
aly in the image chain that may give rise to high investigate the reasons. These include the need for
patient doses is not usually detected by the specia- good radiographic technique and adequate com-
list. pression, which is not always successfully
The individual doses for the same study exhibit achieved. In addition, the reference value in force
variations of one order of magnitude, and even in Spain (7 mGy for a breast 4.5 cm thick, using
more, depending on the equipment used, the antiscatter grid), which is the one originally
technique and other determining factors. Average adopted in the EC documents [10] (currently
doses range by a factor of 2 (between 4 and 8 mGy) updated as Report EUR-16260 [6]), is low in com-
in the abdomen, by factors of about 3 in the chest parison with the values adopted for other examina-
and the breast (Figure 1) and between 2.2 and tions. This is shown by the subsequent adoption by
17 mGy in the lumbar spine. For abdomen and the EC of a reference value of 10 mGy for breast
lumbar spine AP projections, the average dose 5 cm thick using an antiscatter grid. In the case of
values obtained from the dierent centres are chest examinations, it is not unusual to nd focal^
below the reference dose value (10 mGy in both skin distances and operating voltages below the
examinations). This suggests that these reference values recommended in the EC guideline,

The British Journal of Radiology, February 1999 189


L Gonzalez, E Vano, S Oliete et al

Table 2. Percentages of criteria not met and doubtful may be seen that the image quality criteria assume
per type of examination healthy patients (as many details would not be seen
Examination type % of criteria not met and doubtful in certain types of disease) and standard anatomy
(e.g. certain symmetry or structure observation
Committee of Field specialists requirements are not achieved in post-surgical
experts
patients, those with deforming pathology, the
Breast 34 15 elderly or uncooperative subjects). Thus, images
Chest 18 11 that do not meet many quality criteria may be con-
Lumbar spine 16 7 sidered to be suitable in practice, in view of the fact
Abdomen 17 8 that they permit identication of the condition that
was the reason for ordering the examination. In
such cases, the application of the quality criteria
of document EUR 16260 [6], which should reason-
circumstances also observed in the instances of ably be met in the image, has been a valuable help
high doses in lumbar spine examinations. and will, also in those cases, assist the optimization
At centres where various rooms were audited, process. In short, acceptance and transmission of
practically any combination of image quality and radiological images is not independent of the
dose levels could be found. On many occasions patient's characteristics and there is a certain sub-
both parameters exhibited the same tendency that jectivity in the judgement. It is possible to identify
could be expected from the rate of renewal of the certain ``key'' criteria, however, for example in the
equipment (i.e. reasonably low doses and relatively chest examination [14], the absence of which makes
high image quality in modern equipment, and high the image unacceptable for diagnostic purposes in
doses and comparatively inferior quality images in most cases. This suggests the possibility of advan-
cases where old equipment is used). However, cing along the same lines for other types of exam-
these problems are often due to ancillary equip- ination.
Figure 2 allows us to see what criteria degrade
ment, such as intensifying screens or cassettes.
the radiographic image rst, and are at the same
This observation supports the supposition that
time more dicult to meet without the aid of a
dramatic improvements can be achieved with mod-
good technique in chest and breast examina-
est investments, or even at no cost at all [11^13], as
tionsstudies in which the criteria are met with
in cases of radiographic technique changes. No
apparent diculty. Table 2 groups the percentages
correlation is found between image quality and
of image criteria that are not met or are doubtful
dose, as was already veried in the European study
and helps to evaluate their restrictiveness or the
[9], but the facilities that supplied decient quality
intrinsic diculty of an examination when it
images produced dose averages above the reference comes to producing very high quality images.
value. Mammography is, by far, the examination that
The sensitivity of intensifying screen^lm sets is requires most optimization. This is apparently
not an item of information supplied systematically due to insucient compression in some cases,
by the centres, since this feature is probably not and to a relatively common incomplete reproduc-
known by the persons in charge. It may be inferred tion of the pectoralis muscle, even in images of
from the information available, however, that high quality with regard to almost all of the other
lm^screen combinations of nominal speed class criteria. Correct patient positioning is therefore
200 are still used in certain cases. Thus, the unsuit- shown to be one of the key criteria for producing
ability of the equipment, its age (especially the age- valid images for diagnosis. Reproduction of the
ing of the intensifying screens), and details skin structure (pore rosettes) appears to be a criter-
concerning the techniques for obtaining the image ion that is better identiable in images obtained
(especially tube potential, kVp) and inappropriate with relatively modern equipment, and certain
geometric parameters (focal^skin distance, colli- intensifying screen^lm combinations. The radi-
mation) among others, could result in the higher ologists on the board of experts underlined the
dose values. importance of adequate training of the personnel
In quantitative terms, out of 24 facilities responsible for obtaining the mammograms and
audited, 10 would benet very substantially from even that it is preferable for them to be female to
an immediate optimization programme, improv- facilitate the relationship with the patient and
ing their image quality and/or reducing the doses. improve the whole process. The PA chest examina-
The rest exhibit varying degrees of acceptability, so tion continues to be a major challenge for the spe-
the implementation of a quality assurance plan cialist, as frequent failures to comply with
would clearly improve the overall situation. signicant criteria are observed, even at specia-
In the nal classication given to the images, lized centres. In lumbar spine examinations, cor-
either as suitable or not suitable for diagnosis, it rect X-ray beam collimation, sometimes not

190 The British Journal of Radiology, February 1999


Image and patient dose QC in Spanish private X-ray centres

performed, allows the reproduction of spinous and ``culture of quality'' would be gradually imple-
transverse apophyses through enhancing the con- mented and the situation should become consi-
trast, apart from the dose saving oered by this derably improved without a substantial cost
practice. In a simple abdominal examination, increase.
incorrect positioning often impedes reproduction Of the examinations analysed, those referring to
of the kidney area. breast and chest seem to be the most dicult to
The data collected on the average radiological perform not only because of the prospect of
voltage used in each type of examination and cen- exceeding the reference dose values, but also due
tre provides global mean tube potential values to failure to comply with the image quality criteria
below those recommended in the European docu- proposed in document EUR 16260 [6]. On the basis
ment EUR 16260 [6]. The causes should very prob- of higher statistics it may be seen that the low per-
ably be sought in the comparative improvement of centage of spine and abdomen images found in this
the contrast in the image through reducing the study with doses above the reference values sug-
tube potential, although the board of expert radi- gests that the values for these examination types
ologists and other specialists consulted suggest could be reduced. The upper threshold of tube
that the voltage intervals applicable, reected as potential recommended in these examinations
an example of good ``technique'' in the lumbar could be too high in certain patients, impairing
spine and abdomen examinations put forward an the image quality due to lack of contrast.
upper limit that may prove excessive in specic Finally, although the quality criteria of the
cases. images should normally be applied to healthy
Lastly, we should mention that this study partly patients, weighing around 70 kg according to the
anticipates the need for the clinical audit specied recommendations of document EUR 16260 [6],
in article 6.4 of Directive 97/43 EURATOM [2], this is not always possible, especially at centres
and is the rst action of this kind carried out in where patients over a certain age or with certain
Europe in the domain of private medicine, conditions are examined. However, the application
according to the information available. Its preli- of the image quality criteria, based on knowledge
minary results allow us to assure an improvement of the patient's characteristics, has assisted the
in the quality of the whole medical process in optimization process, despite the fact that the
aspects of image quality and patient radiation images, although permitting a correct diagnosis,
doses. do not meet all the image quality criteria in such
cases.
Conclusions
A QC check at 51 private radiology centres Acknowledgments
has been carried out, sponsored for the rst The authors thank Professor E Guibelalde and
time in Spain by a private insurance company J M Fernandez for their help in data processing,
and based on self-evaluation of the centres and an and throughout the study.
external audit procedure taking the objective
form of experimental dose measurements
and evaluations of the quality of the radio-
graphic images. In the case of the 24 facilities eval- References
uated externally, a third exceed the patient dose 1. European Union. Laying down the basic measures
reference levels recommended in the document for the radiation protection of persons undergoing
EUR 16260 [6] and a signicant improvement medical examination or treatment, Council
in the quality of the radiological process, in Directive 84/466 Euratom. Ocial J Eur Commun
terms of dose and image quality, would be possible 1984;No L 265:1^3.
2. European Union. On health protection of individuals
in 10. against the dangers of ionizing radiation in relation to
Barring exceptions, the image quality is consid- medical exposure, and repealing Directive 84/466
ered acceptable for diagnosis, but the use of the Euratom, Council Directive 97/43 Euratom. Ocial J
criteria of document EUR 16260 [6] leaves room Eur Commun 1997;No L 180: 22^7.
for improvement, even in cases where the image 3. Bolet| n Ocial del Estado. Establishing fundamental
measures for radiation protection of the persons
quality is presumed to be appropriate. In this undergoing medical examination or treatment,
sense, periodic retraining of sta should be Royal Decree 1132/1990, from the Health and
a requirement. On the other hand, the doses Consumer Aairs Department. State Ocial
for the same study exhibit large variations. For Bulletin 1990;27261 (in Spanish).
both reasons, a systematic quality assurance 4. Bolet| n Ocial del Estado. On installation and utili-
zation of X-ray devices for medical diagnostic pur-
programme is advisable in all the centres, poses, Royal Decree 1891/1991, from the Industry
including measurement of the patient dose and and Energy Department. State Ocial Bulletin
evaluation of the image quality. In this way, the 1992;138^48 (in Spanish).

The British Journal of Radiology, February 1999 191


L Gonzalez, E Vano, S Oliete et al

5. Bolet| n Ocial del Estado. Establishing quality cri- 11. Ortiz P, Maccia C, Padovani R, Vano E, Alm
teria in radiodiagnostics, Royal Decree 2071/1995, Carlsson G, Schibilla H. Results of the IAEA^CEC
from the Health and Consumer Aairs co-ordinated research programme on radiation
Department. State Ocial Bulletin 1996;1958^62 doses in diagnostic radiology and methods for reduc-
(in Spanish). tion. Radiat Prot Dosim 1995;57:95^100.
6. Carmichael JHE, Maccia C, Moores BM, Oestmann 12. Vano E, Gonzalez L, Guibelalde E, Fernandez JM,
JW, Schibilla H, Teunen D, et al, editors, European Calzado A, Ruiz MJ. Some results from a diagnostic
guidelines on quality criteria for diagnostic radiogra- radiology optimization programme in the Madrid
c images, EUR 16260. Brussels: EC, 1996. area. Radiat Prot Dosim 1995;57:289^92.
7. International Atomic Energy Agency. International 13. Van o E, Oliete S, Gonzalez L, Guibelalde E,
basic safety standards for protection against ionizing Velasco A, Fernandez JM. Image quality and dose
radiation and for the safety of radiation sources, in lumbar spine examinations: results of a 5 year
Safety Series No. 115. Vienna: IAEA, 1996. quality control programme following the
8. Vano E, Gonzalez L, Calzado A, Moran P, Delgado European quality criteria trial. Br J Radiol
V. Some indicative parameters on diagnostic radiol- 1995;68:1332^5.
ogy in Spain: rst dose estimations. Br J Radiol 14. Vano E, Guibelalde E, Morillo A, Alvarez-Pedrosa
1989;62:20^6. CS, Fernandez JM. Evaluation of the European
9. Commission of the European Communities. The image quality criteria for chest examinations. Br J
1991 CEC trial on quality criteria for diagnostic Radiol 1995;68:1349^55.
radiographic images: detailed results and ndings,
EUR 16635-EN. Luxembourg: EC, 1996.
10. European Commission. Guidelines on patient expo-
sure. In: Quality criteria for diagnostic radiographic
images. Working Document (2nd edn), CEC,
DG XII, ref. XII/173/90. Brussels: EC, 1990.

Book review
Catalogue of Diagnostic X-ray Spectra and Other ripple (up to 30%) is modelled for tungsten target
Data. IPEM Report No. 78 (CD-ROM), 1997 spectra. Also held on the CD-ROM are half value
(IPEM, York), 20.00 layer data and linear attenuation data for 32 com-
ISBN 0904181 88X mon elements and compounds.
Report SRS 30, detailing spectral data for diagnos- Although the Birch and Marshall algorithm was
tic X-rays, has been one of the most useful and used to calculate the spectra, some of the support-
widely referenced reports produced by the HPA. ing data have changed. The XCOM attenuation
Report 78 from the IPEM, which updates and coecients are used instead of the well known
extends these data, comes to us on CD-ROM. A McMaster data. Furthermore, the SRS 30 compo-
basic installation takes no time at all, requiring sitions for air and tissue have been replaced by
only 5 MB of hard disk space. Once installed, you ICRU-44 data for air and adipose tissue.
have access to Report 78 itself, some software for Perhaps the most useful part of the package is
processing spectra and of course, lots of data. The the spectrum processor. This program lters a
report gives a straightforward account of the var- spectrum with a combination of lter materials,
ious parameters aecting spectrum shape, com- displays the result as a graph and saves the results
pares the new results against SRS 30 and to le, if desired. Some users may nd the range of
measured data and gives an example BASIC code lter materials a little limited and therefore a fea-
for manipulating the spectrum les. ture allowing the calculation of mass attenuation
The photon spectra themselves are saved as les coecients for user dened materials would be a
on the CD-ROM. Three target materials are con- useful addition.
sidered: tungsten, molybdenum and rhodium. The Overall, I think this is an extremely valuable
energy interval for each spectrum is 0.5 keV and package with a wide range of applications, from
the constant potential spectra have been calculated education through to use in routine work and
in 1 kV steps over the useful diagnostic energy research. This CD-ROM oers excellent value for
range. A range of target angles is also modelled, money and is recommended without hesitation.
at 1 intervals. Finally, the eect of waveform N MARSHALL

192 The British Journal of Radiology, February 1999

Vous aimerez peut-être aussi