RADIBEMS SDN BHD 13-07-2017

Fluke Biomedical Ansur


Test and Inspection Procedure
Copyright 2000 - 2014 Fluke Biomedical

Test Record
TEST PASSED
Test Performed Ansur Components Used
Date: 13-07-2017 Ansur Version 2.9.4
Record: 001061086.pdf Plug-In: ESA620 Version 1.1.4
IEC 62353 Direct
Template:
Leakage - CL2.mtt

Test Setup
Selections
Service Events Performed Standards Performed
IEC 62353 (CL2)

Device under test


Serial Number ZT088A06 Type Nebulizers, Nonheate..
Appliance Code 001061086 Model Pari Inhalierboy 37....
Group BEMS Location Germany
Status BD Address 1 HKL
Manufacturer Pari-Werk GmbH Address 2 HKL

MTI Data
Test Instrument Serial Number Firmware Version
ESA 620 2608005 v2.10

Signatures

Test Record 1 Page 1 of 2


Fluke Biomedical Ansur Test Report RADIBEMS SDN BHD 13-07-2017

Test Result
Test Element Test Type Fail
IEC 62353 Direct Leakage - Class II Auto Sequence

Mains Voltage Mains Voltage

Mains Voltage
Live to Neutral Live to Neutral
Result: Value Unit High Limit Low Limit Standard
Live to Neutral 240.9 V IEC 62353 (CL2)

Direct Equipment Leakage Direct Equipment Leakage

Direct Equipment Leakage


Normal Condition Normal Condition
Result: Value Unit High Limit Low Limit Standard
Normal Condition 0.8 uAAC 100 IEC 62353 (CL2)

Direct Equipment Leakage


Normal Condition, Reversed mains Normal Condition, Reversed mains
Result: Value Unit High Limit Low Limit Standard
Normal Condition,
0.8 uAAC 100 IEC 62353 (CL2)
Reversed mains

Insulation Resistance Insulation Resistance


Configuration:
Test Voltage: 500V

Test Record 1 Page 2 of 2