Vous êtes sur la page 1sur 13

GDRP Hints & Tips

Good Density Practice


Brochure

Hints & Tips

How to Achieve the Best Results


Day-to-Day Density Measurement
Tips and Hints
Contents

Density Measurement
Modern digital instruments are easy to use and allow the density of
liquids to be determined with a high degree of accuracy. High-resolution
instruments are however no guarantee for accurate results. This docu-
ment explains what precautions should be taken to avoid errors when
measuring the density of liquids.

Contents

Test and Adjustment 4
Samples 12
Sampling 14
Cleaning 16
Result verification and documentation 20

3
Test and Adjustment
Test and Adjustment

The commonly held opinion that frequent adjustment of the instrument Test (Calibration)
guarantees accurate results is not true. Any adjustment operation results
in changes being made to the internal instrument settings. How often? Tests should be done routinely in relatively short intervals (days, weeks).
If the adjustment is not properly performed, all the measurements Often a test with water is done every day, as it is done quickly and
performed afterwards will be wrong. ensures that the instrument works accurately.
METTLER TOLEDO DM Density Meters offer the possibility to define fixed
Instead of frequent adjustment, it is better to regularly verify the meas- intervals for test sets with an automatic reminder for the operator. Meas-
urement accuracy of the system by measuring a sample of accurately urement Methods can be set up in way that the operator gets warned
known density (e.g. distilled water or a standard) which is called test, again or the instrument is blocked from use if the defined test interval is
calibration or check. Then the measured density is compared to the expired.
known nominal value of the sample.

Which substance? Readily available and stable solvents like deionized water, analytical
grade toluene, or similar can be used.
The most frequently used test substance is deionized water as it is avail-
able in almost every laboratory and in a high and reproducible purity. It
might be necessary to degas it by boiling.
A different test can be defined separately with larger intervals (months, a
year), using certified and traceable standards for quality assurance and
traceability purposes.
METTLER TOLEDO offers combined (density and refractive index) certified
standards in different ranges:
Water (0.99... g/cm3; nD 1.33)
Dodecane (0.75... g/cm3; nD 1.42)
2,4-dichlorotoluene (1.25... g/cm3; nD 1.55)
1-bromonaphthalene (1.48... g/cm3; nD 1.66)

4 5
Test and Adjustment

Which tolerance The following guidelines may help to define reasonable tolerances so that Instrument = DM50 Density Meter with a resolution of 0.00001 g/
should be applied? frequent error messages caused by tolerances which are too strict can be cm3, limit of error of 0.00003 g/cm3 (for the density range of the
avoided. standard) and limit of error for the temperature of 0.02C.
For deionized water (where the uncertainty of the water is usually not (a) Uncertainty of the standard: 0.00003
known) the tolerance should be defined at 2 times the instrument reso- (b) Limit of error instrument: 0.00003
lution plus the operator repeatability (the non-linearity of the cell has (c) Temperature error: 0.00002
been set to zero for deionized water, if it was used for the adjustment 0.02 C (limit of error for the temperature) * 0.00105 g/cm3/C
as well) ( calculated from given densities at different temperatures of the
Never go below that value range, otherwise there is a high risk of standard = 1.24432 1.25477 g/cm3 / 25 15C)
frequently producing false failed tolerance tests, which are only (d) Operator Repeatability: 0.00001 (example, has to be
caused by internal rounding. But keep it in general as narrow as determined)
possible according the instrument resolution and operator repeat- Tolerance = sum of the 4 components = 0.00009 g/cm3
ability.
Example: DM40 Density Meter with a resolution of 0.0001 g/cm3 This is an example and the tolerance has to be calculated specifi-
Operator repeatability (as example) = 0.00005 g/cm3 (standard cally for each combination of standard and instrument. The toler-
deviation when the operator measures the same sample 3 times in ance for a certified standard may become larger than the 2 to 5
a row. If an operator works properly, he should not get a S.D. more times instrument resolution as it is the case for a normal test with
than that of the instruments rounding capability). local deionized water.
Tolerance = 2 x instrument resolution + operator repeatability =
0.0002 g/cm3 + 0.00005 g/cm3 What to do If the value obtained deviates from the expected (true) value more than
round up to a tolerance of 0.0003 g/cm3. if the test fails? the defined tolerance, proceed as follows:
When using certified organic standards which usually have a relatively 1. Check if the correct substance has been used, e.g. pure fresh deion-
high temperature coefficient (density change with temperature change), ized water
please also allow for the specified temperature error of the instrument. 2. Clean the cell thoroughly and completely dry it in the end (see chapter
So there are four components which normally have to be summed up cleaning)
to form the tolerance, in order to avoid establishing tolerances which 3. Measure the density air value and verify if it is correct, i.e. if the instru-
are too strict: ment is completely clean and dry
Uncertainty of the standard, limit of error instrument, temperature error 4. Repeat the Test
and repeatability. 5. If the test continues to fail with varying results from each test, then
Example: certified standard 2,4-dichlorotoluene with the following the cleaning should be continued with more care, possibly using
stated values: more powerful cleaning solvents and longer cleaning cycles, until the
test plus the air measurements in between show repeatable behavior
Temperature Density [g/cm3] (within a few hours the oscillation value of DM Density Meters varies
15 C 1.25477 0.00003 not more than 1). Only when the tests fail with repeatable results
20 C 1.24954 0.00003 should a new adjustment be performed.
25 C 1.24432 0.00003

6 7
Test and Adjustment

Adjustment

Which temperature Usually density meters are adjusted with air and water.
and substances? The temperature for the adjustment should be the same as the usual
measurement temperature, e.g. if samples are measured at 40 C the ad-
justment should be done at 40 C as well. (Take care when using water
at higher temperatures! See chapter samples with dissolved gases for
the proper sample preparation)
If possible, never switch off instruments with built-in Peltier thermostats.
They are designed to ensure and maintain absolutely stable (e.g. 2
to 4 mK for a DM40 Density Meter) and correct temperature and are not
made for quick changes. If the instrument is switched off, wait for at least
one hour after switching on again before performing an adjustment. Every error or offset caused by a wrong adjustment is a systematic
error it will add on top of the error specification of the instrument!
How to proceed? Only adjust the instrument if the test (calibration) fails but in a repeatable
way (see chapter Test).
Prior to the adjustment the measuring cell must be completely clean and Follow the instrument instruction to perform the adjustment.
perfectly dry. To ensure dryness, the cell should be rinsed with a volatile After performing an adjustment:
solvent (e.g. acetone) and then dried by passing dry air through it for at Always question the adjustment which you have performed do these
least three minutes before performing an adjustment. values make sense?
Example: In the graph below 2 different adjustments are shown. Verify the validity by measuring air and deionized water afterwards.
Blue dots and line the cell was clean and absolutely dry at the begin- If you are not able to reproduce the nominal values which have been
ning for the air part of the adjustment. The water used as second stand- assigned in the adjustment, something went wrong, either in the ad-
ard is pure and not contaminated. justment or in the following measurements.
Pink dots and dashed line the cell is not dried completely; there re- When deviations are observed, repeat the adjustment (or the measure-
mains a small film of moisture (condensed water) along the cell walls, ment of air and water) after correcting the problems (usually a more
not completely removed by the dry air. The following water point then thorough cleaning and complete drying is needed)
still measures correctly, as the water is not contaminated. The instrument
uses a Zero point (for air) which is wrong (too high) due to the insuf-
ficiently dry cell in the beginning. Due to this wrong Zero point it also
calculates an incorrect slope which is too high and samples measured
after the adjustment report inaccurate densities:
Samples with densities lower than water give too low results.
Samples with a density around water, and water, give correct results.
Samples with densities lower than water give too high results.

8 9
Test and Adjustment

Adjustment a nalysis METTLER TOLEDO DM Density Meters offer the possibility to activate an Define your Test procedures (interval, substances, tolerances) and
and automatic adjustment analysis. This brings a warning if the deviation in the oscilla- your Adjustments (substances, temperatures):
a djustments tion values (air and/or water) turned out too high and the adjustment can
be rejected. In addition, the adjustment history can be viewed anytime,
also as a graph.

The METTLER TOLEDO SC1 and SC30 automation units help to avoid er-
roneous adjustments:
Adjustments are very easy to perform by just clicking the Adjust short-
cut on the home screen.
Prior to each adjustment, the measuring cell can be thoroughly
cleaned and completely dried (if selected).
A Cell Test can be activated (verifying the density of air).
Standard is pushed into measuring cell at an accurately defined and
reasonably low speed to ensure complete wetting of the cell glass
walls.

10 11
Samples
Samples

Viscous samples Be careful when measuring viscous samples and make sure that the be introduced into the measuring cell under positive pressure and not, as
samples do not contain air bubbles when you introduce them into the is usually the case with many automated sampling pumps, sucked into
measuring cell. It often helps to warm the sample in a closed vessel the cell as this bears the risk of degassing due to reduced pressure. The
and to allow it to stand for a few minutes before the measurement. The METTLER TOLEDO SC1 and SC30 automation units are suitable for such
viscosity of the sample decreases on warming and air bubbles can more samples.
easily escape.
The shear forces that arise when viscous samples are measured in the Samples with Some samples require degassing prior to the density measurement.
measuring cell can lead to inaccurate results. The value of the density d issolved gases To degas CO2 of samples such as soft drinks, stir them for several min-
displayed by the meter is then usually too high. All METTLER TOLEDO utes until bubbling stops and/or use a paper filter.
DM Density Meters are able to automatically correct the sample viscosity To remove dissolved air different techniques can be applied. Pure organic
measurement error. If your density meter has such viscosity correction, solvents can be put in an ultra-sonic bath for a few minutes. Other sam-
you should always switch it on if you want to obtain the best possible ples can be boiled for several minutes to remove the dissolved air.
accuracy for samples with a viscosity greater than 25 mPas. Example: when Water is measured at temperatures above approx.
METTLER TOLEDO SC1 and SC30 automation units are able to handle 3540 C; often problems often occur (due to dissolved air). To avoid
samples with a viscosity of up to 30000 mPas (which equals ap- this, boil the water for at least 10 minutes, keep it boiling until use and
proximately liquid honey). For samples which have a higher viscosity or introduce it to the measuring cell while still hot so that the cell has to cool
which are even solid at room temperature (e.g. paraffine) the heated ver- the sample by at least 1 C.
sions SC1H and SC30H can be used.
Non-homogeneous If solutions or suspensions are allowed to stand, part of the solid mate-
Aggressive samples Make sure that all parts which come in contact with the sample are samples/suspensions rial may settle or a concentration gradient may form. Stir the sample well
resistant! before taking a sample. Make sure that no air bubbles are introduced
If concentrated acids or bases (e.g. H2SO4, HCl, NaOH) are measured, during stirring. If you have to measure suspensions, it is often not pos-
rinse the cell rapidly after the measurement with lots of water to avoid sible to completely homogenize the sample (e.g. ketchup or cream). In
excessive heating stress for the cell. such cases, the measurement should be repeated several times and the
For measuring aggressive or toxic samples, the METTLER TOLEDO SC1 mean value of the individual measurements calculated in order to obtain
and SC30 automation units offer very high occupational safety as the reliable values. (Example for a METTLER TOLEDO DM Density Meter: use
handling of these materials is limited to a minimum and no evaporation measurement reliability medium or low, or even fixed measurement dura-
occurs. For strongly oxidizing acids (e.g. conc. HNO3) SC1 and SC30 tion. Use the multiple measurement function as well)
can be equipped with a special Hastelloy needle (C22) to avoid corro-
sion.

Volatile samples Air bubbles may form during the measurement as a result of the sample
degassing. The solubility of a gas in a liquid usually decreases with in-
creasing temperature. For this reason, samples that show a tendency to
degas (e.g. winter gasoline which contains dissolved butane) should be
cooled to a low temperature below that of the measurement (e.g. stored
in a fridge) before introduction into the measuring cell and stored in a
closed vial filled to the top. Furthermore, such samples should always

12 13
Sampling
Sampling

With a syringe Manual sample handling with a syringe is always operator-dependent, Avoid bubbles After the filling, check if the cell is bubble-free. Air bubbles (or a thin air
which means that it is error-prone, difficult to reproduce, and time-con- film) are a serious problem. Even very small quantities of air cause big
suming. Make sure that your operator is properly trained and injects the errors when doing density determinations:
sample in a reproducible manner.
Use 3-component syringes with a rubber O-ring as they allow a Diameter of the air bubble Max. Measuring error caused
much better speed control than cheap 2-component syringes. [mm] [g/cm3]
2 0.000838
Do NOT fill the cell just barely! Possible contamination would remain in 1 0.000052
the cell and yield a false measurement: 0.5 0.000003

The problem is that most small bubbles or air cushions (due to bad wet-
ting of the cell) are not or hardly visible by eye. Also in dark samples
Make sure that the sample comes out of the cell at least an extra 10 cm, bubbles are hardly visible.
so that contamination can be pushed out and only new sample is left in METTLER TOLEDO DM Density Meters have a built-in Bubble Check
the cell. to detect bubbles. The most secure way for reliable results is a Multiple
Measurement though. (see chapter error detection)

Fill the measuring cell at a slow speed and with a laminar flow
(510 cm per second) to ensure complete wetting of the cell walls
(no trapped bubbles along the walls).
Make sure that no air is entrapped in the syringe. The plunger has to be
pressed slowly and continuously without stopping.

Automatic filling Automatic filling systems ensure that the cell is filled with the correct
speed and in a reproducible manner, independent of operators and
samples.
With the METTLER TOLEDO SC1 and SC30 automation units its even
possible to fill very highly viscous or sticky samples into measuring cell
(such as highly viscous oils, viscosity improvers, creams, molasses or
liquid honey).
The sampling speed can be easily adjusted for different sample viscosi-
ties in the defined products or methods.

14 15
Cleaning
Cleaning

Inappropriate cleaning is the most common source of erroneous Rinse Before the beginning of rinsing, remove all sample from cell and tubes.
results! Make sure that the measuring cell does not contain any For each kind of sample 2 appropriate rinsing solvents have to be
residue from previously measured samples or rinsing solutions.
d efined.
Purpose of the 1st rinse: Must be able to completely and quickly dis-
Deposits of previously measured products are not always visible. For ex- solve all the sample, so that no contamination is left in the cell. This
ample, if products containing oil or fat are measured, the measuring cell solvent is often not very volatile.
may be coated with a very thin film of oil. To prevent this, the cell should Purpose of the 2nd rinse: Must completely dissolve the 1st solvent
be cleaned with suitable rinsing solutions (as described later) and dried - above, and it must evaporate quickly without leaving any residue to en-
preferably after each measurement. able quick drying afterwards.
If all measured samples are of a similar kind and able to dissolve the Please make sure that this solvent is free of water, otherwise it does not
residues in the measuring cell (e.g. when the density meter is used to evaporate quickly enough. The use of acetone is preferred. If acetone is
measure different juices), it is also possible to do a large over-sampling not allowed, use ethanol (must be puriss > 99.9%, NOT 96% and NOT
with the new sample to ensures a complete removal of the old one denatured).
(analytical rinse).
Restrictions: Some general recommendations for the rinsing solutions:
Use at least a sampling pump (e.g. METTLER TOLEDO FillPal), a Sample Solvent 1 Solvent 2
proper over-sampling is difficult to achieve with a syringe Water based Water Acetone or Ethanol (puriss)
Immerse the sampling tube of the pump in the sample, then remove Acids Lots of water Acetone or Ethanol (puriss)
it so that air is sucked in the tube (~23 cm air in the tube) and im- Fats and oils Deconex* Acetone or Ethanol (puriss)
merse it again in the sample. Repeat this procedure approx. 5 times (0.3 to 0.5% in water)
before the cell is filled for the measurement. This ensures that the old Petrochemicals Toluene or Petrol Ether Hexane or similar if temp.
sample is properly flushed out of the cell. is > 30 C
Verify the procedure to make sure that the required repeatability and At room temp. use low-
boiling petrol ether mixture
limit of error are maintained. Measure the most critical sample first (for
or acetone.
instance, the one with the highest sugar content) followed by deionized
Conc. Sugar Water (use enough water Acetone (puriss)
water then repeat this step a couple of times. solutions / syrup before rinse with acetone
If for instance sugar containing products are measured, make sure that risk of polymerization)
the cell remains filled with either sample or with water between meas-
* Deconex dissolves well in water, acetone, and ethanol! Available
urements to avoid the sample drying out and crystallizes sugar on the from METTLER TOLEDO
cell walls.
Completely clean and dry (as described below) the measuring cell at
least once after each working day.

16 17
Cleaning

Dry Even very small quantities of rinsing liquids (or residues of previously Define the cleaning procedures for your samples:
measured samples) in the cell may cause substantial measuring errors.
Example: If the density meter cell was rinsed with ethanol and not com-
pletely dried, and then a measurement of water is performed, the error
due to remaining rinsing liquid is as follows:

Remaining Ethanol in Cell [L] Measuring error caused [g/cm3]


10 0.001500
1 0.000150
0.1 0.000015
0.01 0.000001

Dry the cell completely with dry air by using a drying pump.
Usually, ambient air is aspired through a bed of silica gel drying agent.
Please make sure to replace or re-activate the silica gel well before it is
spent (usually indicated by a color change). Using inactive silica gel car-
ries the risk that moisture from the ambient air is condensed out inside
the cold measuring cell of your instrument, thus contaminating each
sample with a layer of condensed water which clings to the cell walls.
Such condensation will hardly be visible by eye. This risk increases the
more the cell temperature lies below room temperature.
To check if the cell has been dried completely, i.e. an appropriate drying
time has been chosen, measure the air density and compare it to the
known value at that temperature.

Fully automatic With the METTLER TOLEDO SC1 and SC30 automation units, the measur-
cleaning ing cell is automatically cleaned and dried. The two rinsing liquids for
cleaning (e.g. water and acetone) are mixed with lots of air and pumped
through system at high speed. This results in a pulsating flow which pro-
vides very efficient near-mechanical cleaning and also reduces solvent
consumption.
As the inside and outside of the SC1/SC30 sampling nozzle is thoroughly
cleaned and dried after each measurement, sample carryover is not pos-
sible.
With an automatic Cell Test the cleanness of the cell can be verified.

18 19
Result verification
Result verification and documentation

and documentation
Automatic result Often the result has to be converted using a table. Looking up in or in- A 2 fold measurement with reliability medium (extrapolated) takes about
conversion terpolating from a table is error-prone and time-consuming. Automatic the same time as 1 measurement at maximum measurement reliability
conversion using built-in tables (e.g. alcohol, Brix, temperature compen- but delivers far better data quality!
sation according to API) prevents reading or calculation errors and saves Also in case if a bubble would not be detected by the Bubble Check
time. A digital density meter of the latest generation allows the use of it would be securely detected with the multiple measurement. And for
built-in conversion tables to show the result directly in the desired unit. not homogeneous samples (density gradients or such) a mean value is
METTLER TOLEDO DM Density Meters have to following built-in result much more representative than just a single measurement.
units / concentration tables:
Specific gravity, light and heavy Baum, Twaddell Result limits METTLER TOLEDO DM Density Meters offer a product approach. The
Sugar: Plato, Brix (Emmerich, NBS 113), HFCS 42/55, Invert Sugar, same measurement method might be used for several samples, but
KMW, Oechsle, Babo some sample specific parameters (like viscosity correction, filling speed
Alcohol: OIML, AOAC, Proof Degree, HM C&E, Gay Lussac etc.) can be defined for each product, also the lower and upper result
Petro: API degrees and gravity tables for crude oils, refined products limits can be defined. Measured products are then automatically verified
and lubricants with the limits and the color of the result(s) indicates if specifications are
Up to 30 user-defined concentration tables (can be entered as tables or met: black = passed, red = failed.
formulas)

Error detection It is often very difficult, or even impossible, to visually detect bubbles or
thin air cushions. Therefore digital density meters often have automatic
bubble detection, for instance the Bubble Check of METTLER TOLEDO
DM Density Meters. But there are also other possible error sources such
as remaining rinsing liquids etc.
If you want to be sure about your results, do a 2 (or 3) fold measure-
ment and take the mean value.The data quality drastically increases.

METTLER TOLEDO DM Density Meters in combination with an automation Proper Documentation Hand-written results bear the risk of transcription errors. Depending on
unit offer a fully automatic multiple measurement with a definable maxi- the environment, there is no guarantee that this kind of documentation
mum standard deviation. fulfills regulatory requirements.
Regulatory requirements are easy fulfilled when using a printer or PC soft-
ware. GLP-compliant printouts contain all of the necessary information
for complete documentation. METTLER TOLEDO DM Density Meters can
be used with the USB-P25 dot matrix printer for GLP-compliant printouts.
Printouts must be manually managed and filed, which is a potential
source of errors and incompleteness. Using PC software for data collec-
tion and management has the advantage of greater data reliability and
easier future accessibility.

20 21
Result verification and documentation

The software LabX in connection with DM Density Meters guarantees to Notes:


have all data stored centrally and securely in a database. It offers ERP/
LIMS connectivity, customized reports, secure traceability and 21 CFR
11 support. The software can be connected remotely via Ethernet saving
valuable bench space and enabling the operator to perform all operations
from the instrument terminal.

22 23
Fully automatic adjustments / test
Automatic filling, cleaning & drying. Also for difficult and viscous
samples
Automatic error detection Bubble Check / Multiple Measurement
Automatic result verification product database with stored limits
Automatic storage of results in database, LIMS export etc. with
software LabX

www.mt.com/GDRP
For more information
Mettler-Toledo AG, Analytical
CH-8603 Schwerzenbach, Switzerland
Phone +41-44-806 77 11
Fax +41-44-806 73 50

Subject to technical changes


07/2012 Mettler-Toledo AG
Printed in Switzerland
MCG MarCom Greifensee

Vous aimerez peut-être aussi