Product Information PI

ROTANODE
E7894X

Rotating Anode X-ray Tube Assembly

Rotating anode X-ray tube assembly for the


purpose of general diagnostic X-ray procedures.

Specially processed Rhenium-tungsten faced


molybdenum target of 74 mm diameter.

This tube has foci 1.2 and 0.6, and is available for a
maximum tube voltage 150 kV.

Accommodated with IEC60526 type high-voltage


cable receptacles.

General Data

IEC Classification (IEC60601-1:2005) ......................................... Class ME EQUIPMENT

Electrical:
Circuit:
High Voltage Generator ............................................... Constant potential high-voltage generator
Grounding .......................................................................................................... Center-grounded
Nominal X-ray Tube Voltage (IEC60613:2010):
Radiographic ...................................................................................................................... 150 kV
Nominal Focal Spot Value (IEC60336:2005):
Large Focus ............................................................................................................................. 1.2
Small Focus ............................................................................................................................. 0.6
Nominal Anode Input Power (at 0.1s):
60 Hz 50 Hz
Large Focus ..................................................................................................... 30 kW 29 kW
Small Focus ..................................................................................................... 15 kW 14 kW
Nominal Radiographic Anode Input Power (IEC60613:2010):
60 Hz 50 Hz
Large Focus ..................................................................................................... 30 kW 29 kW
Small Focus ..................................................................................................... 15 kW 14 kW

The information contained herein is presented only as a guide for the application of our products. No responsibility is assumed by Toshiba Electron
Tubes & Devices Co., Ltd. (TETD) for any infringements of patents or other rights of the third parties which may result from its use.
No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
The information contained herein may be changed without prior notice. It is therefore, advisable to contact to TETD before processing with the design of
equipment incorporating this product.
E7894X

Motor Ratings:
Stator: XS-BF
Starting Running
Driven Frequency [Hz] 50/60 50/60
Input Power [W] 1050 270 43
1) 3)
Voltage [V] 200 100 40
2)
Current [A] 6.0 3.0 1.2
4)
Min. Speed Up [s] 0.8 1.5 -
Capacitor [F] 24 24 24
Note: 1) Applied voltage between common and main terminal.
2) Common current.
3) The every applied voltage must be never exceeded 110% of the above specification.
4) The speed-up time is allowed up to 110% of the above specification.

Anode Speed:
60 Hz ............................................................................................................ Minimum 3200 min-1
50 Hz ............................................................................................................ Minimum 2700 min-1
Stator Resistance:
Common-Main Winding ...................................................................................................... 27.5
Common-Auxiliary Winding ................................................................................................ 58.0
Resistance between Housing and Low Voltage Terminals ........................................ Minimum 2 M
Normal Operating Range of the Housing Temperature ..................................................... 16 ~ 75 C
Mode of Operation ........................................................................................................... Intermittent

Mechanical:
Dimensions .................................................................................................. See dimensional outline
Overall Length .................................................................................................................. 437 mm
Maximum Diameter ....................................................................................................... 152.4 mm
Target:
Anode Angle ............................................................................................................ 12.5 degrees
Diameter ............................................................................................................................ 74 mm
Construction .................................................................... Rhenium-Tungsten faced Molybdenum
Permanent Filtration ................................................................. 0.9 mm Al / 75 kV IEC60522:1999
Radiation Protection (In accordance with IEC60601-1-3:2008):
Leakage Technique Factor .................................................................................. 150 kV, 0.4 mA
X-ray Coverage ................................................................................ 430 430 mm at SID 1000 mm
Weight (Approx.) ..................................................................................................................... 16.4 kg
High Voltage Receptacle ..................... To meet the requirements of IEC60526 Corrigendum1:2010
Cooling Method .................................................................................................. Natural or forced air
Tube Housing Model Number ................................................................................................ XH-196

-2-
E7894X

Absolute Maximum and Minimum Ratings


(At any time, these values must not be exceeded.)

Maximum X-ray Tube Voltage (IEC60613:2010):


Radiographic ...................................................................................................................... 150 kV
Between Anode (or Cathode) and Ground ................................................................................ 75 kV
Minimum X-ray Tube Voltage ................................................................................................... 40 kV
Maximum X-ray Tube Current (IEC60613:2010) .................................................... See rating charts
Large Focus ...................................................................................................................... 500 mA
Small Focus ...................................................................................................................... 200 mA
Maximum Filament Current:
Large Focus .......................................................................................................................... 5.8 A
Small Focus .......................................................................................................................... 5.2 A
Filament Voltage:
Large Focus (At maximum filament current 5.8 A) ................................................... 14.9 ~ 20.1 V
Small Focus (At maximum filament current 5.2 A) ................................................... 10.2 ~ 13.8 V
Filament Frequency Limits ................................................................................................ 0 ~ 25 kHz
Continuous Anode Input Power (IEC60613:2010) .................................................... 60 W (85 HU/s)
Thermal Characteristics:
Anode Heat Content ......................................................................................... 100 kJ (140 kHU)
Maximum Anode Heat Dissipation ................................................................... 475 W (667 HU/s)
X-ray Tube Assembly Heat Content ................................................................ 900 kJ (1260 kHU)
Nominal Continuous Input Power (IEC60613:2010):
Without Air-circulator .................................................................................... 170W (238 HU/s)

-3-
E7894X

Environmental Limits

Operating Limits:
Temperature ................................................................................................................ 10 ~ 40
Humidity ........................................................................................................................ 30 ~ 85 %
(No Condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature .............................................................................................................. -20 ~ 70
Humidity ........................................................................................................................ 20 ~ 90 %
(No Condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa

-4-
E7894X

Maximum Rating Charts


(Absolute Maximum Rating Charts)

Conditions: Tube Voltage


Constant potential high-voltage generator
Stator Power Frequency 60Hz

Nominal Focal Spot Value: 1.2 Nominal Focal Spot Value: 0.6
70kV 60kV 50kV 40kV 80kV 70kV 60kV 50kV 40kV
500 200
90kV
80kV
100kV
400 90kV
150 110kV
TUBE CURRENT [mA]

TUBE CURRENT [mA]


100kV
110kV 125kV
300 125kV
150kV
150kV 100

200

50
100

0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]
Refer to IEC60613:2010

Conditions: Tube Voltage


Constant potential high-voltage generator
Stator Power Frequency 50Hz

Nominal Focal Spot Value: 1.2 Nominal Focal Spot Value: 0.6
60kV 50kV 40kV 70kV 60kV 50kV 40kV
500 200
70kV 80kV

80kV 90kV
400
90kV 150 100kV
TUBE CURRENT [mA]
TUBE CURRENT [mA]

100kV 110kV

300 110kV 125kV


125kV
100 150kV
150kV
200

50
100

0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]
Refer to IEC60613:2010

-5-
E7894X

Emission & Filament Characteristics


Constant potential high-voltage generator

Nominal Focal Spot Value: 1.2


500 20
60kV

80kV
400 100kV 18

FILAMENT VOLTAGE [V]


TUBE CURRENT [mA]

125150kV

300 16

50kV
200 40kV 14

Ef
100 12

0 10
4.5 4.7 4.9 5.1 5.3 5.5 5.7 5.8

FILAMENT CURRENT [A]

Note1) For Reference Only


Note2) Refer to IEC60613:2010

Nominal Focal Spot Value: 0.6


200 14

125150kV
FILAMENT VOLTAGE [V]

150 12
TUBE CURRENT [mA]

100kV
Ef
100 10

80kV
60kV
50 50kV 8
40kV

0 6
4.0 4.2 4.4 4.6 4.8 5.0 5.2

FILAMENT CURRENT [A]

Note1) For Reference Only


Note2) Refer to IEC60613:2010

-6-
E7894X

Thermal Characteristics

X-ray Tube Assembly Heating / Cooling Curve

900
COOLING 170W
800

700
HEAT STORAGE [kJ]

600

500

400

300

200

100
HEATING
0
0 60 120 180 240 300

TIME [min]

Anode Heating / Cooling Curve

100
475W
90
COOLING
80
HEAT STORAGE [kJ]

70
200W
60

50

40

30
60W
20

10
HEATING
0
0 1 2 3 4 5 6 7 8
TIME [min]
The heating curves are showing examples of average
input power to the anode in operation.

-7-
E7894X

Dimensional Outline
Unit mm

142

CENTRAL RAY AND REFERENCE AXIS


ANODE
4-M6 8 DEEP

139
(10)

(138) .5
12
90

4373
(123)

92

(132)

45
(10)

152.4 S
L

CATHODE

TERMINAL CONNECTIONS
122

9 8 7 6 5 1 2 3 4 E
53

2-M63.5, P=1.27
81

NC NC NC NC
C1 M A

ET
TEMPERATURE RELAY
(NORMALLY CLOSED)
Note) Do not connect terminal No.1 and No.5 or 6
in series circuit.
70

CENTER OF FOCAL SPOT

EXPLANATION OF SYMBOLS -1.5mm1.5mm


CATHODE -1.5mm1.5mm
TERMINAL
C : COMMON
L : LARGE FOCUS
S : SMALL FOCUS

TERMINAL CONNECTIONS : CENTRAL X-RAY
ANODE & CATHODE TERMINAL
C1 : COMMON
M : MAIN WINDING OF THE STATOR : IEC60526 TYPE
A : AUX. WINDING OF THE STATOR
NC : NON-CONNECTION
ET : EARTH TERMINAL

-8-
E7894X

Detail of inside of end cap


Unit: mm

141
A 131
A

45
45
4.9

45

45
4M5
CROSS SECTION A-A
4M5

45
6.4

45

D
45

45
B 131 B
141

CROSS SECTION B-B

141
R1
0.5
13

R1 141
15
0.5

Detail drawing C Detail drawing D

-9-
Marketing Engineering Group, Sales Department
1385,Shimoishigami,Otawara-shi,Tochigi 324-8550,Japan
Tel: +81-287-26-6666 Fax: +81-287-26-6060
http://www.toshiba-tetd.co.jp/

The head office of Toshiba Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Environmental Management System ISO14001.
Toshiba Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Quality Management Systems ISO9001 and ISO13485.
Product scope is referred to the following URL. http://www.toshiba-tetd.co.jp/tetd/eng/company/quality.htm.
PE-E7894X 2017-02

Vous aimerez peut-être aussi