Dear Customer,

Instruction Manual Thank you for choosing a Hanna product. This SPECIFICATIONS DISPLAY CODE GUIDE
manual will provide you with the necessary infor-
HI 93711 mation for the correct operation of the meter. This indicates that the meter is in
Free & Total Chlorine Please read it carefully before using the meter. If
you need additional technical information, do
a ready state and zeroing can be
performed.
ISM not hesitate to e-mail us at
Sampling in Progress. This prompt
techserv@hannacan.com.
This instrument is in compliance with directives appears each time the meter is
EN 50081-1 and EN 50082-1. performing a measurement.
This indicates that the meter is in
a zeroed state and measurement
PRELIMINARY EXAMINATION can be performed.
A zero reading was not taken. Insert
Remove the instrument from the packing mate- a sample before adding reagent
rial and examine it carefully to make sure that and press ZERO.
no damage has occurred during shipment. If Under range. A blinking "0.00"
there is any damage, notify your Dealer. indicates that the sample absorbs
less light than the zero reference.
Each Ion Specific Meter is supplied complete SPECIFICATIONS Check the procedure and make
with Range Free Cl2 0.00 to 2.50 mg/L sure you use the same cuvet for
9V Battery Total Cl2 0.00 to 3.50 mg/L reference (zero) and measurement.
Two Sample Cuvets and Caps Resolution 0.01 mg/L Over range. A flashing value
One Transport Cap Accuracy 0.03 mg/L 3% of reading higher than the maximum
Note: Conserve all packing material until the Typical EMC 0.01 mg/L concentration readable (see speci-
Deviation fications) indicates that the sample
This Instrument is in instrument has been observed to function
www.hannacan.com Light Source Light Emitting Diode @ 555 nm absorbs too much light, meaning
Compliance with the CE correctly. Any defective item must be
Directives Method Adaptation of the EPA recommended DPD that the concentration is too high.
returned in its original packing. method 330.5. The reaction between the Dilute the sample.
chlorine and the DPD reagent causes a Light over range. The cuvet is not
GENERAL DESCRIPTION pink tint in the sample
inserted correctly and an excess
WARRANTY The HI 93711 meter measures the free and total
Light Detector Silicon Photocell ambient light is reaching the de-
Environment 0 to 50C (32 to 122F); tector. If the cover is properly
HI 93711 is warranted for two years against defects in work-
chlorine (Cl2) content in water and wastewater. max 95% RH non-condensing installed, then contact your dealer
manship and materials when used for its intended The meter uses an exclusive positive-locking Battery Type/Life 1 x 9 volt/40 hours or the nearest Hanna Customer
purpose and maintained according to instructions. Auto-Shut off After 10' of non-use Service Center.
This warranty is limited to repair or replacement free of
system to ensure that the cuvet is in the same
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8") Light under range. The zero sample
charge. place every time it is placed into the measure-
Weight 290 g (10 oz.). is too dark for proper zeroing. If
Damages due to accident, misuse, tampering or lack ment cell.
of prescribed maintenance are not covered. REQUIRED REAGENTS this is not the case, contact your
If service is required, contact the dealer from whom The reagents are in liquid and powder form POWDER: dealer or the nearest Hanna
you purchased the instrument. If under warranty, report and are supplied in bottles and in packets. The Customer Service Center.
Code Unit Description Quantity
the model number, date of purchase, serial number
and the nature of the failure. If the repair is not covered amount of reagent is precisely dosed to ensure HI 93701-0 Free Cl2 DPD 1 packet The "V" indicates that the battery
by the warranty, you will be notified of the charges maximum repeatability. HI 93711-0 Total Cl2 DPD 1 packet voltage is getting low and the
incurred. If the instrument is to be returned to Hanna battery needs to be replaced.
Instruments, first obtain a Returned Goods Authorization Display codes aid the user in routine LIQUID:
Number from the Customer Service department and
V
This indicates that the battery is
operations. Code Unit Description Quantity dead and must be replaced.
then send it with shipment costs prepaid. When shipping
HI 93701A-T Free and Total Cl 2 DPD1 indicator 3 drops
any instrument, make sure it is properly packaged for The meters have an auto-shut off feature that Note: once this indication is dis-
complete protection. HI 93701B-T Free and Total Cl 2 DPD1 buffer 3 drops
will turn the instrument off after 10 minutes of HI 93701C Total Cl2 only DPD3 solution 1 drop played, the meter will
To validate your warranty, fill out and return the enclosed
warranty card within 14 days from the date of purchase. non-use. lockup. Change the battery to
All rights are reserved. Reproduction in whole or in part restart.
is prohibited without the written consent of the copyright
owner, Hanna Instruments inc.

Hanna Instruments reserves the right to modify the design,


construction and appearance of its products without
advance notice.
Wait for a few seconds, until are requested.
OPERATIONAL GUIDE all the air bubbles have com- INTERFERENCES
pletely dissolved and reinsert the Interference may be caused by:
cuvet into the instrument. Bromine Iodine
MEASUREMENT PROCEDURE ON Fluorine Ozone
Press READ and the display will
Turn the meter on by pressing ON/ OFF show "SIP" during measure-
Oxidized manganese and Chromium
Alkalinity above 250 mg/L or acidity above 150
ACCESSORIES
OFF. ment.
mg/L will not reliably develop the full amount of
Note: For total chlorine mea- REAGENT SETS
When the LCD displays "- - -", color or it may rapidly fade. To resolve this,
surements wait for 2 HI 93701-01 Reagents for 100 free chlorine tests
it is ready. neutralize the sample with diluted HCl or NaOH.
minutes and 30 seconds (powder)
Select free or total Chlorine by In case of water with hardness greater than 500 HI 93701-03 Reagents for 300 free chlorine tests
before pressing the mg/L CaCO3, shake the sample for approximately
pressing FREE/TOTAL. An "I" or a "C" will READ key. (powder)
appear on the right corner to indicate free or 1 minute after adding the reagent. HI 93711-01 Reagents for 100 total chlorine tests
The instrument directly displays concentration
total chlorine, respectively. (powder)
Fill the cuvet with 10 mL of unreacted sample, HI 93711-03 Reagents for 300 total chlorine tests
(powder)
FREE HI 93701-T Reagents for 300 chlorine tests (liquid)
OTHER ACCESSORIES
TOTAL
HI 710009 Blue rubber boot
in mg/L of free or total chlorine on the Liquid HI 710010 Orange rubber boot
Crystal Display. HI 721310 9V battery (10 pcs)
up to the mark, and re- HI 731318 Tissue for wiping cuvets (4 pcs)
Liquid reagents procedure HI 731321 Glass cuvets (4 pcs)
place the cap.
To measure Free Chlorine, add 3 drops of HI
10 mL 93701A-T DPD1 indicator and 3 drops of
TIPS FOR AN ACCURATE HI 731325 Caps for cuvets (4 pcs)
Place the cuvet into the HI 93703-50 Cuvets cleaning solution (230 mL).
holder and ensure that
HI 93701B-T DPD1 buffer to the cuvet. To MEASUREMENT
measure Total Chlorine, also add 1 drop of HI
the notch on the cap is 93701C. Shake gently before adding 10 mL of The instruction listed below should be carefully
positioned securely into
the groove.
unreacted sample. Replace the cap and shake followed during testing to ensure best accuracy. CE DECLARATION
gently again. Do not touch the cuvet walls with hands. OF CONFORMITY
In order to maintain the same conditions
Press ZERO and "SIP" will Reinsert the cuvet into the instrument. during the zeroing and the measuring phases,
appear on the display.
it is necessary to close the cuvet to prevent any
Wait for a few seconds and the display will contamination.
show "-0.0-". Now the meter is zeroed and
Do not let the test sample stand too long after
reagent is added or accuracy will be lost.
ZERO Whenever the cuvet is placed into the mea-
surement cell, it must be completely free of
fingerprints, oil or dirt. Wipe it thoroughly with
HI 731318 or a lint-free cloth prior to insertion.
ready for measurement. x3 x1 It is important that the sample does not contain
Remove the cuvet any debris. This would corrupt the readings.
Powder reagents procedure It is possible to take multiple readings in a
Add the specific test reagent: row, but it is recommended that a zero read-
Free Chlorine: 1 packet of Total Chlorine: 1 ing be taken for each sample and that the
packet of same cuvet is used for zeroing and
DPD Free Chlorine reagentDPD Total Chlorine measurement.
Wait for approximately 30
reagent. seconds and then press It is important to discard the sample immedi-
READ. The display will show ately after the reading is taken because the
"SIP" during measurement. glass might become permanently stained.
Replace the cap and shake gently.
The instrument directly dis- Shaking the cuvet can generate bubbles in the
plays concentration in mg/L of free or total sample, causing higher readings. To obtain
chlorine on the accurate measurements, remove such bubbles Recommendations for Users
by swirling or by gently tapping the vial.
HI 93711
HI 93701

Before using these products, make sure that they are entirely suitable
Liquid Crystal for the environment in which they are used. Operation of these instru-
Display. All the reaction times reported in this manual ments in residential area could cause unacceptable interferences to
are referred to 20C (68F). As a general rule of radio and TV equipments, requiring the operator to take all necessary
Note: Free and total steps to correct interferences. Any variation introduced by the user to
thumb, they should be doubled at 10C (50F) the supplied equipment may degrade the instruments' EMC perfor-
chlorine have mance.
and halved at 30C (86F).
to be measured To avoid damages or burns, do not perform any measurement in mi-
crowave ovens.
separately with fresh unreacted samples
following the above procedure if both values PRINTED IN CANADA IST93711R2 11/99