Vous êtes sur la page 1sur 6

Original ResearchOtology and Neurotology

Otolaryngology
Head and Neck Surgery

Bone Conduction Hearing: Device 2014, Vol. 150(5) 866871


American Academy of
OtolaryngologyHead and Neck
Auditory Capability to Aid in Device Surgery Foundation 2014
Reprints and permission:
Selection sagepub.com/journalsPermissions.nav
DOI: 10.1177/0194599814524530
http://otojournal.org

Mark J. Syms, MD1,2, and


Kelly E. Hernandez, AuD3

Sponsorships or competing interests that may be relevant to content are dis- Received September 20, 2013; revised December 24, 2013; accepted
closed at the end of this article. January 29, 2014.

Abstract

C
linical interest and treatment options for patients with
Objective. To obtain identical laboratory measures of 8 (sur- single-sided deafness (SSD) have increased markedly
gical and nonsurgical) bone conduction devices and relate over the recent past with many more device options
them to clinical function. available. The clinical manifestations of SSD are multiple
Study Design. Each device was measured with a single labora- and well documented.1 Patients with SSD are unable to hear
tory system and characterized with descriptive statistics. on the SSD side and often have to turn their head to engage
in conversation. In addition, patients are unable to locate
Setting. Laboratory. sounds and have problems with background noise. Many
Subjects and Methods. Seven surgical devices (Intenso, BP110, patients are disturbed by the hearing struggles associated
BP100, and Cordelle [Cochlear, Denver, Colorado]; Ponto with SSD and can be effectively rehabilitated with devices
Pro and Ponto Pro Power [Oticon Medical, Somerset, New intended to overcome the hearing difficulties.2-4
Jersey]; and Alpha 2 [Sophono, Inc, Boulder, Colorado]) and When counseling patients regarding the various available
1 nonsurgical dental device (SoundBite; Sonitus Medical, Inc, bone conduction technologies, the mechanisms and surgical
San Mateo, California) constituted the independent variables. differences can be clearly articulated. However, patients
Measured maximum output and gain parameters were the often inquire about the hearing performance of the various
dependent variables. bone conduction options. The hearing performance is depen-
dent on both the specific capabilities of the device as well
Results. Maximum output varied across devices in the pure- as the status of the cochlea that is going to be stimulated.
tone average (PTA; 500-3000 Hz) frequency range (mean, Multiple systems have a simulator that can be used to
109.7 dB re 1 mN; range, 98.8-119.2 dB) and in the above- demonstrate the technology. For an osseointegrated post
PTA (4000-8000 Hz) frequency range (mean, 102.6 dB re 1 system, the simulations often have limited efficacy since
mN; range, 88.99-119.6 dB). Maximum gain varied in the there is intervening soft tissue during the simulation that is
PTA frequency range (mean, 40 dB; range, 29.1-49.1 dB) and not present in the implanted system. For a magnetically
was higher in the frequency range above the PTA (mean, coupled system, the simulation also has limited efficacy since
32.0 dB; range, 20.8-46.0 dB). the implant placed in the bone is absent. For a dental system,
Conclusion. All devices have sufficient maximum output and the simulator (Acoustiq; Sonitus Medical, Inc, San Mateo,
gain for the PTA frequency range for single-sided deafness California) does directly contact the patients teeth and is not
(SSD). The devices differed in maximum output and gain for limited by soft tissue or lack of the presence of implant but
the frequency range above the PTA, a consideration for
accommodating presbycusis and optimizing auditory function 1
Neurotology, Barrow Neurological Institute, Phoenix, Arizona, USA
2
for SSD. The surgical devices have less maximum output and Otology/Neurotology, Arizona Ear Center, Phoenix, Arizona, USA
3
gain in the above-PTA range than in the PTA range. The non- Audiology, Arizona Ear Center, Phoenix, Arizona, USA
surgical dental device had the highest output (up to 30 dB This article was presented at the 2013 AAO-HNSF Annual Meeting & OTO
higher) and gain (up to 26 dB higher) in the above-PTA range. EXPO; September 29October 3, 2013; Vancouver, British Columbia,
Canada.
Corresponding Author:
Keywords Mark J. Syms, MD, Director, Neurotology, Barrow Neurological Institute,
Arizona Ear Center, 2627 North Third St, Suite 201, Phoenix, AZ 85004,
bone conduction, Baha, dental, presbycusis, single-sided USA.
deafness Email: m.syms@arizonaear.com
Syms and Hernandez 867

Table 1. Manufacturer, Device Name, Transducer Type, and Coupling Method of 8 Bone Conduction Devices.
Manufacturer Device Transducer Coupling Method

1. Cochlear BP100 Electrodynamic Implanted percutaneous post


2. Cochlear BP110 Electrodynamic Implanted percutaneous post
3. Cochlear Cordelle Electrodynamic Implanted percutaneous post
4. Cochlear Intenso Electrodynamic Implanted percutaneous post
5. Oticon Medical Ponto Pro Electrodynamic Implanted percutaneous post
6. Oticon Medical Ponto Pro Power Electrodynamic Implanted percutaneous post
7. Sonitus Medical SoundBite Piezoelectric Removable dental module
8. Sophono Alpha 2 Electrodynamic Implanted subdermal plate

also may be limited because the front teeth can be used device, constituting the independent variable. The surgical
rather than the molars. devices used electrodynamic transducers but differed by
An additional limitation for simulation of all systems is coupling method, with 6 devices coupled to a percutaneous
the status of the cochlea to be stimulated by the device. If post implanted into the temporal bone, traditionally called a
the patient has a high-frequency hearing loss, the simulation Baha (Intenso, BP110, BP100, and Cordelle [Cochlear,
device needs to be programmed to match the patients hear- Denver, Colorado] and Ponto Pro and Ponto Pro Power
ing loss. Given that programming the device is time- [Oticon Medical Somerset, New Jersey]), and 1 held against a
consuming and consideration of the soft-tissue limitations, magnetic subcutaneous plate screwed into a surgically pre-
the simulation is often performed without programming and pared well in the temporal bone (Alpha 2; Sophono, Inc,
in the audiology booth. Overall, the simulation of the vari- Boulder, Colorado). The nonsurgical device consisted of a
ous devices does demonstrate the phenomenon of bone con- piezoelectric transducer mounted in a patient-removable dental
duction hearing. However, demonstrating the hearing module held against the lateral surfaces of 2 maxillary teeth
experience for a patient is limited. (SoundBite; Sonitus Medical, Inc). Table 1 lists the manufac-
When counseling patients regarding the various bone con- turers, devices, transducer types, and coupling methods.
duction technologies, clinicians should consider both the cur- The study design consisted of a comparison of laboratory
rent status of the cochlea and the natural history of hearing measures from an example of each of these devices obtained
loss in the functional cochlea. A patient who may be able to with a single laboratory measurement system. Standard
be rehabilitated with a device based on the current status of descriptive statistics (mean, range, and SD) were used to
his or her cochlea should be informed of the possible progres- characterize the measures for each device and to quantify
sion of the hearing loss in the only hearing ear. Clinicians and measurement variability.
patients should consider the efficacy of the device over the Devices were characterized using a custom laboratory
patients lifetime. The ability of a device to provide lifelong system (Sonitus Medical, Inc) that enabled testing of both
rehabilitation should be emphasized for patients undergoing implantable and nonsurgical dental devices. The system con-
surgery if additional surgery may be needed. sisted of a bench-top anechoic chamber (Anechoic Test Box
Patients considering a bone conduction option for hearing Type 4232; Bruel & Kjaer, Naerum, Denmark) with a loud-
loss often request a comparison of performance among the speaker driven by a computer with a sound card that delivered
various devices. Clinical studies with head-to-head compari- calibrated acoustic signals (swept frequency from 100-10,000
sons are either compromised because they often involve ear- Hz) to the microphone of the device and a calibrated precision
lier generation technologies or unavailable because it is force gauge (B&K 8001) that measured output of the device
virtually impossible to randomize devices that require sur- (force in dB re 1 mN). The force gauge was assembled with a
gery specific to the device. In the hearing device dispensing vibration mount and bracket, having a total mass that pre-
environment, the objective gains of the devices during simu- sented a sufficiently high mechanical impedance load to the
lation are used for comparison. The purpose of this article is transducer under test for accurate measures of output force
to obtain identical laboratory measures for direct compari- (.20 dB higher than transducer output impedance).
son of 8 bone conduction devices and relate the laboratory Each processor was configured to have the widest fre-
measures to clinical function. quency response, highest gain, and minimum effect from
automatic gain control and other adaptive features such as
Methods noise or acoustic feedback suppression to ensure valid com-
Eight bone conduction devices were investigated represent- parisons across devices. Maximum force output and acousto-
ing those currently cleared by the US Food and Drug mechanical gain were evaluated according to methods
Administration (FDA) in the United States from 4 manufac- defined in the international standard IEC60118-9:1985 using
turers. There were 7 surgical devices and 1 nonsurgical 90-dB sound pressure level (SPL) and 60-dB SPL input
868 OtolaryngologyHead and Neck Surgery 150(5)

mN; range, 98.8-119.2 dB) and in the above-PTA frequency


range (mean, 102.6 dB re 1 mN; range, 88.99-119.6 dB).
Maximum acousto-mechanical gain was also found to be
comparable to the manufacturers published data (generally
within 62 dB). As with maximum output force, gain differs
across devices and frequency region. Maximum gain varied
in the PTA frequency range (mean, 40.0 dB; range, 29.1-
49.1 dB) and in the above-PTA frequency range (mean,
32.0 dB; range, 20.8-46.0 dB). Figure 4 shows the maxi-
mum gain for each device (in dB) for the standard clinical
frequency range (PTA) and the high-frequency range (above
PTA).
Figure 5 illustrates the presbycusis consideration for all
devices. This figure shows the maximum output of each
device displayed in an audiogram format (dB hearing level
Figure 1. Photo of the custom laboratory measurement system [HL]) with a frequency range indicated for the conventional
measuring a Baha device. clinical description (PTA range) and the frequency range
indicated for changes in thresholds for 2 typical presbycusis
cases represented by the average hearing thresholds for
signals, respectively. Implantable bone-anchored devices women and men aged 70 years.5 In addition, the range of
were tested with a Baha abutment modified with an M2/10- the hearing loss by frequency for these 2 groups (10th-90th
32 brass fitting to interface to the force gauge. Figure 1 percentiles) is displayed.6
shows an image of the force gauge assembly with the
Cochlear Intenso attached. The nonsurgical dental device was Discussion
tested using a steel fitting attached to the force gauge that The measures reported enable direct comparison of output
allowed coupling via a spring wire to replicate its attachment and gain among devices in a clinically relevant format and
to the teeth. Data for the dental device were compensated for are very useful since they can help in device selection. The
the effect of the tooth interface to the skull to provide a mea- maximum output information is particularly useful for
sure of the skull stimulation force for direct comparison to device selection since these maximum output values suggest
the surgically implanted devices. a fitting range for each device. The maximum gain informa-
tion is also useful, but less so for purposes of device selec-
Results tion because the actual use gain is always much less than
Figure 2 shows the maximum output (force in dB re 1 mN) maximum and adjusted for each patient individually.
for the measurements (solid line) and for the published man- All of these devices are FDA approved for patients with
ufacturers results (filled circles) for each device. The mea- a normal-functioning cochlea with either conductive hearing
sures for the subdermal surgical device were comparable to loss or SSD. In these cases, the reported maximum output is
the manufacturers published results, although only the man- sufficient, but the maximum gain for all devices will far
ufacturers values are shown for this device because a soft- exceed the patients gain requirements. The maximum gain
tissue component was not developed for the measurement can be adjusted for all these devices down to the required 0-
system. The nonsurgical dental device (SoundBite) response dB to 10-dB target value for activating a normal cochlea.7
is compensated for the effect of the tooth interface to the For the devices that are FDA approved for patients in whom
skull and represents the actual skull stimulation force. The the cochlea is not normal (mixed hearing loss), the maxi-
measures for all devices were reliable (test-retest SD of mum gain measure will be useful in device selection
\0.5 dB) and valid (generally within 62 dB of the manu- depending on the individual bone conduction sensitivity.
facturers published result). This indicates that the measures Two very common clinical entities also should be consid-
are comparable across devices. ered during the process of device selection. First, all of
Figure 3 shows the maximum output for all devices dis- these devices are intended to be a long-term solution for
played in a format that uses a standard (ANSI S3.6:2004) patients, and this is especially true of the patients using a
clinical audiometric frequency range (250-8000 Hz) with device requiring a surgical intervention. Changes in hearing
values at octave frequencies (250-8000 Hz) and select inter- not associated with the primary pathology, presbycusis in
octave frequencies (750-6000 Hz). The maximum output particular, should be considered for all cases. Second,
differed across devices and by frequency region. These patient selection is often described in terms of air conduc-
results can be described using the clinically relevant pure- tion hearing loss over a very restricted frequency range, the
tone average (PTA) of the values at 500, 1000, 2000, and PTA, when in fact bone conduction sensitivity is the only
3000 Hz and the average of the values above the PTA factor to consider for all bone conduction devices, by defini-
(4000, 6000, and 8000 Hz). Maximum output varied across tion. Each of these 2 considerations will be described in a
devices in the PTA frequency range (mean, 109.7 dB re 1 clinical context.
Syms and Hernandez 869

140 140

120 120

Force (dB re 1 N)

Force (dB re 1 N)
100 100

80 80

60 60

40 40
Cochlear Cochlear
20 BP100 20 BP110
0 0
250 500 1k 2k 4k 8k 250 500 1k 2k 4k 8k
Frequency (Hz) Frequency (Hz)
140 140

120 120
Force (dB re 1 N)

Force (dB re 1 N)
100 100

80 80

60 60

40
Cochlear 40 Cochlear
20 Cordelle 20 Intenso
0 0
250 500 1k 2k 4k 8k 250 500 1k 2k 4k 8k
Frequency (Hz) Frequency (Hz)
140 140

120 120
Force (dB re 1 N)

Force (dB re 1 N)

100 100

80 80

60 60

40 40
Oticon Oticon
20 Ponto Pro 20 Ponto Pro Power
0 0
250 500 1k 2k 4k 8k 250 500 1k 2k 4k 8k
Frequency (Hz) Frequency (Hz)
140 140

120 120
Force (dB re 1 N)

Force (dB re 1 N)

100 100

80 80

60 60

40 40
Sonitus Sophono
20 SoundBite 20 Alpha 2
0 0
250 500 1k 2k 4k 8k 250 500 1k 2k 4k 8k
Frequency (Hz) Frequency (Hz)
Figure 2. Maximum output force (dB re 1 mN) for 90-dB sound pressure level input for measurements (solid line) and from published
manufacturer results (filled circles) for each device.

Presbycusis, age-related progressive bilateral sensori- FDA approved for presbycusis. However, many cases will
neural hearing loss, affects men more than women. It is have a presbycusis consideration. Often times, patients
measurable epidemiologically for frequencies .2000 Hz with a longstanding SSD present for evaluation and rehabi-
beginning at 50 years of age and progresses systematically litation with the onset of presbycusis in the only hearing
with age.8 None of the devices considered in this study is ear.9
870 OtolaryngologyHead and Neck Surgery 150(5)

20

40

dB HL
60

80

100
250 500 1k 2k 4k 8k
Frequency (Hz)
Figure 3. Maximum output force for each device (dB re 1 mN) Figure 5. Maximum output of each device (see Figure 3 for
for 90-dB sound pressure level input. Each manufacturer is repre- device identification) displayed in an audiogram format. The gray
sented by a color (Cochlear: blue, Oticon Medical: red, Sonitus circles represent the typical presbycusis audiogram of a 70-year-old
Medical: orange, and Sophono: green) and each device by a sepa- woman, and black circles represent the same for a 70-year-old
rate symbol. man. The correlating bars represent the mean change in threshold
in the 10th to 90th percentiles for women and men age 70 years.8

Figure 5 illustrates that both of the typical presbycusis


PTA Frequency Range cases will require gain and maximum output increases in
>PTA Frequency Range
the frequency range well above the PTA. The Sophono
device will perhaps not have enough output to maximally
rehabilitate the typical 70-year-old male or female presbycu-
sis patient. The Sophono, Oticon Ponto Pro, or Oticon
Ponto Pro Power will not have enough output to maximally
40 rehabilitate a typical 70-year-old man. As demonstrated in
Figure 4, none of the implantable surgical devices evalu-
Maximum Gain (dB)

ated in this study is able to provide enough high-frequency


gain to maximally rehabilitate the more severe cases of
presbycusis. The SoundBite device is able to provide
enough high-frequency gain and output except at 8000 Hz
20 in the range of hearing loss that occurs in the 90th percen-
tile range and worse in 70-year-old men.
The maximum output levels are similar across the each
manufacturers family of devices. The difference in maxi-
mum output among the Cochlear BP100, Cochlear BP110,
and Cochlear Cordelle devices is not great. Similarly, the
0 difference in maximum output between the Oticon Ponto
Pro and Oticon Ponto Pro Power is not great either. This
o
10

lle

ite
er
00

2
ha
Pr
ns

w
P1

de

finding suggests that factors other than the maximum output


P1

dB
Po

lp
te

o
rB

or
rB

un
nt

A
n

such as the potential increased size of the processor, and


rC

rI

o
Po

o
ea

So
ea

Pr

on
ea
ea
hl
hl

on

thus less favorable aesthetic satisfaction, should be dis-


s
hl

ph
oc
oc

hl

tu
nt
oc

tic

So
oc

ni
Po
C
C

cussed with a patient since the potential clinical advantage


C

So
C

on

may not be significant.


tic
O

Single-sided deafness is a common application for these


Figure 4. Maximum acousto-mechanical gain for each device in dB
(difference between force in dB re 1 mN and 60dB sound pressure devices,5 a condition where the device activates a normal
level [SPL] input sound pressure) averaged over the standard clini- cochlea. It is often reported using the air conduction hearing
cal pure-tone average (PTA) frequency range (solid bars, 500-3000 sensitivity of the better hearing ear in the PTA frequency
Hz) and the higher frequency range above the PTA (hashed bars, range. Devices that produce signals in the PTA frequency
4000-8000 Hz). range provide a minimum signal for understanding speech
Syms and Hernandez 871

even though sound quality may not be optimal. Primary Sonitus. Kelly E. Hernandez is a consultant and a member of the
complaints of patients with SSD are reductions in spatial Surgeons Advisory Board for Sonitus and an investigator for a proto-
hearing ability and reductions in speech understanding in col sponsored by Sonitus. The protocol, which started in November
noise.1 The frequency range above the PTA optimizes audi- 2013, is called Evaluation for Benefit for Treatment of Single Sided
tory function for spatial hearing and understanding speech Deafness Between Two Bone Conduction Prosthetic Devices:
Osseointegrated Implant versus Maxilla Anchored Removable Oral
in noise as well as improves sound quality.10 In fact, the fre-
Appliance.
quency range for optimizing function for all SSD cases
should be 1500 Hz and higher because frequencies lower Sponsorships: Sonitus performed bench-top testing.
than 1500 Hz are perceived normally with no device in SSD Funding source: Internal funding.
cases.11
Acoustic feedback has been an issue with many of these References
devices. Many variables determine if a patient will experi- 1. Valente M, Oeding K. Single Sided Deafness. New York, NY:
ence acoustic feedback.2 When counseling patients, the Thieme; 2011.
potential for acoustic feedback should be discussed. The 2. Gurgel RK, Shelton C. The soundbite hearing system: patient-
current investigation does not evaluate the level of gain for assessed safety and benefit study. Laryngoscope. 2013;123:
each device in the context of the variables that can lead to 2807-2812.
acoustic feedback. Although a device may have a demon- 3. Rasmussen J, Olsen S, Nielsen LH. Evaluation of long-term
strable level of gain in an anechoic chamber, this maximum patient satisfaction and experience with the Baha bone con-
level may not be able to be obtained in a patient due to duction implant. Int J Audiol. 2012;51:194-199.
acoustic feedback limitations. 4. Kompis M, Pfiffner F, Krebs M, Caversaccio MD. Factors
influencing the decision for Baha in unilateral deafness: the
Conclusion
Bern benefit in single-sided deafness questionnaire. Adv
All but one of the devices can be measured accurately with Otorhinolaryngol. 2011;71:103-111.
a single custom laboratory system. All devices have suffi- 5. Pai I, Kelleher C, Nunn T, et al. Outcome of bone-anchored
cient maximum output and gain for the PTA frequency hearing aids for single-sided deafness: a prospective study.
range (500-3000 Hz). The devices differ in maximum Acta Otolaryngol. 2012;132:751-755.
output and gain for the frequency range above the PTA 6. Hoffman HJ, Dobie RA, Ko CW, Themann CL, Murphy WJ.
(4000-8000 Hz), a consideration for accommodating presby- Hearing threshold levels at age 70 years (65-74 years) in the
cusis and for optimizing the auditory function for SSD. unscreened older adult population of the United States, 1959-
The surgical devices have less output and gain in the fre- 1962 and 1999-2006. Ear Hear. 2012;33:437-440.
quency range above the PTA than in the PTA frequency 7. Bosman AJ, Snik FM, Mylanus EA, Cremers WR. Fitting
range. The nonsurgical dental device has the highest output range of the BAHA Intenso. Int J Audiol. 2009;48:346-352.
(up to 30 dB higher) and highest gain (up to 26 dB higher) 8. Hoffman HJ, Dobie RA, Ko CW, Themann CL, Murphy WJ.
in the frequency range above the PTA. Americans hear as well or better today compared with 40
Acknowledgment years ago: hearing threshold levels in the unscreened adult
population of the United States, 1959-1962 and 1999-2004.
We thank Timothy Proulx, Vice President, Research &
Development, Sonitus Medical, Inc, for his help in conducting the Ear Hear. 2010;31:725-734.
laboratory measures. 9. Faber HT, de Wolf MJ, Cremers CW, Snik AF, Hol MK.
Benefit of Baha in the elderly with single-sided deafness. Eur
Arch Otorhinolaryngol. 2013;270:1285-1291.
Author Contributions
10. Moore BC, Popelka GR. Preliminary comparison of bone-
Mark J. Syms, analysis and writing; Kelly E. Hernandez, analy- anchored hearing instruments and a dental device as treatments
sis and writing. for unilateral hearing loss. Int J Audiol. 2013;52:678-686.
Disclosures 11. Pfiffner F, Kompis M, Flynn M, Asnes K, Arnold A, Stieger
C. Benefits of low-frequency attenuation of Baha in single-
Competing interests: Mark J. Syms is a consultant and shareholder
sided sensorineural deafness. Ear Hear. 2011;32:40-45.
for Cochlear, a consultant and a member of the Surgeons Advisory
Board for Sonitus, and an investigator for a protocol sponsored by

Vous aimerez peut-être aussi