Académique Documents
Professionnel Documents
Culture Documents
Ch.3 Settings....................................................................................................53
3.1 Outline of DRYPRO 771 Settings...................................................................................54
3.2 Starting the Service Maintenance Mode.........................................................................55
3.3 Closing the Service Maintenance Mode .........................................................................56
3.4 DICOM SCP Settings .....................................................................................................57
3.5 DICOM SCU Settings .....................................................................................................59
3.6 PRINT CONDITION Settings..........................................................................................62
3.7 Setting the Print Quality (U50, U60, U30).......................................................................65
3.8 DICOM PRIORITY Settings............................................................................................72
3.9 FILM SETUP Settings ....................................................................................................74
3.10 SYSTEM SETUP Settings..............................................................................................75
3.11 Checking the Internal Clock............................................................................................78
3.12 SHEDULE Setting ..........................................................................................................79
Ch.5 Print Commanded from the External Device and Backup ..................89
5.1 Print commanded from the External Device ...................................................................90
5.2 Backup............................................................................................................................90
Specifications....................................................................................................95
Preface This chapter describes information with which the
field service engineer should become familiar
before proceeding with installation.
<1>
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Preface
1. To avoid accidents, only personnel properly trained and authorized by the manufacturer should remove covers
or touch internal parts.
2. The DRYPRO 771 incorporates a laser beam device (Class 3b). Direct exposure of the skin or eyes to the laser
beam may cause serious damage. Always use protective goggles when carrying out repairs or adjustments.
3. Under no circumstances should any attempts be made to carry out operational procedures or adjustments other
than those described in this manual: exposure to harmful radiation may result.
4. To ensure safety while using or working on the DRYPRO 771, cautions shown on warning labels must be
observed at all times.
5. The DRYPRO 771 contains internal parts that carry high voltage that may cause electrocution if touched.
Extreme caution must be exercised at all times.
6. Be careful to avoid parts of the body or clothing becoming trapped or entangled in moving parts such as fans.
7. The DRYPRO 771 weighs approximately 155kg(without stand) / 175kg (with stand).
Ensure that there is sufficient work space when carrying out unpacking or installation work.
8. Electrical circuitry in the DRYPRO 771 may be damaged by static electricity. Due care must be exercised when
handling the main body or any electrical parts removed during repair work.
9. The DRYPRO 771 power supply must be switched off when removing circuit boards, disconnecting or
connecting connectors or cables. Under no circumstances should such procedures be carried out with the
power supply switched on: doing so may result in serious accidents.
10. A wrist band must always be worn when handling circuit boards.
11. The DRYPRO 771 incorporates a lithium battery. Improper replacement of the battery may cause damage: only
a qualified service engineer may carry out this procedure.
12. The DRYPRO 771 is a Class-1 laser device furnished with interlocks. The procedures described in this manual
must be followed when disengaging interlocks.
13. Be certain to comply with any and all laws and government regulations including electric codes and utility
regulations, regarding the installation and operation of this equipment.
<2>
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Preface
DANGER : Failure to observe the caution will produce high risk of serious or fatal injury.
WARNING : Failure to observe the caution will produce moderate risk of serious or fatal injury.
CAUTION : Failure to observe the caution will produce moderate risk of serious or fatal injury or
damage to property.
High Low
<3>
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Preface
<4>
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Preface
11) Caution warning of high-temperature (100:) 12) Heat Process Unit Cover warning
of deodorant filter housing of high-temperature (130:)
<5>
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Preface
4) Class-1 Laser
Product Label
Supply Tray
9) Supply tray light-block label
<6>
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Preface
p.5 Details of
Heat Process Unit Cover Warning
Labels
12) Heat Process Unit Cover warning of high-temperature (130:)
<7>
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Preface
<8>
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
1
Before
This chapter describes information with which
Installation field service engineer should become fully familiar
before proceeding with work.
<9>
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
Installation
Unpacking
Making Settings
Density correction
Checks
Test Printing
< 10 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
DRYPRO 771
< 11 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
Settings
Normally when setting the "DICOM PRINT", it is essential to set "Port No." + "IP Address" + "AE Title".
However, it is possible to allocate 1ch to the DRYPRO 771 disregarding the AE title. (use "S09 DEFAULT
CH" to select and set the channel that is allocated to the expansion port)
Using this expansion port, maximum 8 channels can be connected. Leave blank for the AE title inputbox for
both expansion port and device that is connected to the expansion port.
DRYPRO 771
Basic
Backup DRYPRO 771 DRYPRO 771
< 12 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
Select the basic configuration when it is possible to secure sufficient space for
installing several units of DRYPRO 771.
Connecting with NO
5 or more external
devices
YES
Connecting with NO
8 or less external
devices
YES
2 or more devices
allocated to
NO
expansion port
YES
5 or less devices NO
allocated to
expansion port
YES
3 or less devices NO
not allocated to
expansion port
YES
< 13 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
p.59 Setting
DICOM SCU
p.57 Setting
DICOM SCP CH1 CH2 CH3 CH4
DICOM SCP
S09 DEFAULT CH
DRYPRO 771
DRYPRO 771 OFF
Ex-2a : Expansion Port CS-1 Printlink2 REGIUS IM LiteView Liteview Liteview Liteview Liteview
DICOM SCP
S09 DEFAULT CH DRYPRO 771
DICOM SCP
S09 DEFAULT CH
DRYPRO 771
DRYPRO 771 CH2
< 14 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
Maintenance Space
100mm or more
Normal 630mm Left Door Open 250mm Exhaust Duct, Spacer Attached
(for jam release)
25mm
600mm
Front Cover Open Supply Tray Open Stand Front Door Open
(for deodorant filter change, (for film change)
jam release)
250mm
600mm 580mm
< 15 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
Maintenance Space
800mm or more
DRYPRO
771
1000mm or more
Check of Route along which Power Supply and Network Cables will be Laid
The following items should be checked with the facility supervisor.
The location of the hub to which the DRYPRO 771 will be connected, the route along which the
ethernet cable will be laid and necessary cable lengths.
The location of the power outlet (3P) to which the DRYPRO 771 will be connected, the route along
which power supply cable will be laid and necessary cable lengths.
Check also that the cables will be laid along not, for example corridors, but along routes where safety
can be ensured.
< 16 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
Maintenance PC
PC with 10BaseT/100BaseTX network port.
PC with TCP/IP protocol installed.
PC that operates on Windows95/98/2000/XP.
Supports Internet Explore (ver.5.5 or higher).
Java applet to be installed.
Equipment for Network Connection
Hub (used for direct connection of DRYPRO 771 with maintenance PC).
Ethernet cable.
(For DRYPRO 771 -> hub connection and maintenance PC -> hub connection).
Tools
Cutter
Phillips screwdriver
Long phillips screwdriver
(120mm (47.3inch) or more) to disengage front cover lock.
Flathead screwdriver
Flathead screwdriver (with edge of 4mm (0.16inch) or less) to open front cover lock
disengagement cover.
Radio pliers
13mm box wrench (with a long neck) or socket wrench (with extension cover).
Densitometer
Jigs
DRYPRO 771 interlock killer
< 17 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
2) Front Cover
5) Air Intake
6) Right Cover
7) Front Cover
Unlock Cover
3) Supply Tray
8) Power Breaker
9) Stand
< 18 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
1) Top Cover
3) Air Outlets
7) Operation Switch
8) Left Cover
9) Rear Cover
7) Operation Switch
Activates the DRYPRO 771.
< 19 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
Operation Panel
1) Message Display
Window 6) [timer] Key
2) "operation" Lamp
7) [maintenance] Key
3) "timer" Lamp
4) [menu/select] Key
8) [reset / open] Key
5) [select] Key
9) [enter] Key
12) "0-9" Lamp
10) [exit] Key
13) "A-Z" Lamp
16) Ten-keypad
14) "mode" Key
< 20 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
Under normal use by user, "0-9 Lamp", "A-Z Lamp", [mode] key, [shift] key, "Ten-
Keypad" and [del/BS] key are concealed by service maintenance key cover. Detach this
cover when required in installation, maintenance and service work. However, make sure
that the cover is put back when the user operates the DRYPRO 771.
Slightly push towards the message display window, and lift to remove the cover.
< 21 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
Interior
5) Deodorant Filter
2) Lever-BC
Filter Housing
(Yellow)
3) Cover-C
4) Film Check Cover
< 22 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
Structure
An outline of the structure each of the sections in the DRYPRO 771 is set out below.
Film will be transported from 1) through 6) in a printing process.
5) Heat
6) Cooling Section Processing
Section
1) Supply Section
4) Elevator
2) Descent Transport Transport
Section Section
Film Flow
8) Electrical Section
3) Exposure
Section
1) Supply Section
Picks up film one sheet at a time from the supply tray and feeds it to the descent
transport section.
3) Exposure Section
Executes laser scanning in synchronization with the film feed process and writes
images on the film.
6) Cooling Section
Cools the film that is heated in the heat processing.
The cooling section is comprised of the primary cooling unit and secondary cooling unit.
8) Deodorant Filter
Eliminates odors arising during heat processing. This filter must be periodically
replaced.
9) Electrical Section
This section comprises the power supply unit as well as units for overall control of the
DRYPRO 771, communication with diagnostic devices, image processing and other
control units.
< 23 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.1 Before Installation
< 24 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
2
Unpacking and
Installation This chapter describes the procedure for
unpacking and installing the DRYPRO 771.
< 25 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
Lid
Accessory Box
Deodorant filter unit, Suction cups (spare),
Cutter, One-touch spacers (2 pcs),
Power cable, Clamp (2 pcs), Exhaust duct,
Warranty, Operation manual,
Operation sheet, Attached document for
doctors, Inspection sheet,
Protection parts (red parts) recycle bag,
Accessory list.
DRYPRO 771
< 26 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
2.2 Unpacking
The procedure for unpacking the DRYPRO 771 is described below.
The weight of the DRYPRO 771 is approximately 155kg (without stand) or When disposing of the
175kg(with stand) and the weight of the packing is approximately 189kg (without packing materials,
stand) or 205kg (with stand). follow the regulations
Allow adequate space when transporting or moving the unit and carry out work with and rules of the local
due care. Work should be carried out by at least 2 personnel. government.
2 Remove the lid from the outer box. Do not stand directly in
front of the packing
band when cutting. The
band may rebound and
cause injury when cut.
3 4
4 Remove the lock piece (4) from the outer box.
5 Remove the outer box.
< 27 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
the box
containing
accessories
(2)
9 Using the restrainer as a lever, raise the main
(1) body at the front and remove the shock
Be extremely careful not
(2) absorbers (2) from under the left and right front
to jam fingers when
side.
removing the shock
< 28 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
< 29 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
< 30 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
Starting up the
DRYPRO 771 with
protective parts still in
place may result in
damage.
< 31 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
2.4.2 Removing protective parts from the Rear Panel of Exposure Unit
How to remove the protective parts from the rear panel of exposure unit is set out below.
120mm minimum
< 32 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
Hook
Underside of
exposure unit
2 Remove the two screws (at both ends) of the
tree securing the protective parts for fixing the
rear side of the exposure unit, and pull out these When removing the
protective parts. protective parts for
fixing the exposure unit
at its back, be careful
not to damage the
internal parts of the
Next, remove the protective parts from the front
panel of the exposure unit. DRYPRO 771.
2.4.3 Removing protective parts from the Front Panel of the exposure unit
How to remove the protective parts from the front of the exposure unit following the removal from the
rear panel of the exposure unit is set out below.
< 33 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
BC
< 34 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
2.4.5 Removing protective parts from the Rear Panel of the exposure unit
How to remove the protective parts from the supply section following the removal of protective films
from exposure section is set out below.
6 Take out the screw (1) securing the supply unit Hook
Underside of
Supply unit
When attaching the protective parts, first remove the supply tray, then attach the
protective parts for fixing the supply section.
< 35 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
The procedure for attaching the supply unit cover is detailed below.
1) Bring the upper edge of the supply unit cover into contact with the main
body.Ensure that the groove at the top left of the supply unit cover fits under
the bar that secures the main body and front cover.
2) Bring the lower edge of the supply unit cover into contact with the main body.
3) Ensure that the two hooks located on the back of the supply unit cover latch
into the corresponding slots on the main body.
4) Secure the supply unit cover to the main body using the six screws provided.
< 36 >
1 2
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
3
< 36 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
The heat-developer
Hold both sides of the heat process unit cover drum shaft cover is
when removing, as shown in the illustration. attached to the heat
process unit cover at
the rear of the main
body. This axis cover is
only latched onto the
unit cover: be careful
not to forget to replace
the axis cover when
< 37 >
setting the unit cover
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
back in place.
Ch.2 Unpacking and Installation
Cleaner Unwoven Cloth 10 Check that the surface of the unwoven cleaner surface of the drum will
adversely affect output
cloth is not wavy.
images.
If it is found wavy, remove the waviness according
to the procedure described in "2.4.8 Adjusting the
Cleaner Unwoven Cloth".
< 38 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
a
1 Remove the five screws "a"(truss screws,
M4x12) securing the rear cover.
a
a
a
b
a
2 Loosen the two screws "b" (truss screw, M4x12)
Rear Cover
on the bottom of the rear cover.
b
< 39 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
a
2 Attach the rear cover.
a
a
a
3 Tighten the five screws "a" (truss screw, M4x12)
to secure the rear cover.
a
b
Rear Cover
b
4 Tighten the two screws "b" on the bottom of the
rear cover.
< 40 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
< 41 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
< 42 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
< 43 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
< 44 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
2 Pull the tray out to its furthest extent and open the
light blocking cover.
11x14
14x14
17x17
3 Remove the side stopper screws (M4x8, two in
front and rear) from the original position.
Do not remove
these two.
4 3
5 6 6 Attach the two screws (M4x8) removed in the Make sure all screws
once removed are
step-4 into the original position. installed.
Destination Original Pos. The screw holes at the original position are Failure to install all may
blocked, thus preventing the external light to leak in. cause fogging of the film.
< 45 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
1. Operation
Operating sequence of the battery, one is recommended by the manufacturer, the other not is
detailed below.
If the serial cable is accidentally pulled off during operation of the DRYPRO 771, it starts
shutdown sequence in the same manner as a power failure occurs.
If the power is abruptly shut down while the DRYPRO 771 is in operation, data such as
film data may not be kept correct.
The above mentioned, connection test-proven battery for UPS is available in each
country. However, please note that the product code may differ depending on the
electrical standard)
Check on the homepage of APC in each country, and select a correct type
compatible with those listed above.
3. Installation
Follow the instruction detailed in the user manual attached to the purchased device.
The UPS battery can be set within the stand if the DRYPRO is delivered with a
stand.
When installing the UPS battery for the DRYPRO that is not delivered with a stand,
install the battery on either side of the unit so that it will not hinder the operation.
< 46 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
< 47 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
< 48 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
2.6 Start Up
After completing the removal and installation of specified parts, start up the DRYPRO 771.
ON
OFF
ON
OFF
< 49 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
Displayed
alternately.
1 Press the [reset/open] key.
Shutter closing... .
Please wait... The message indicating that the supply tray is
being unlocked is displayed.
3 Open the light-blocking cover on the supply tray. The light-blocking cover
cannot be opened
unless the supply tray
4 Take the new film package out of the box and has been pulled out
load it into the supply tray so that the bar-code is fully.
positioned at the front-left corner.
< 50 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.2 Unpacking and Installation
< 52 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
3
Settings
This chapter describes the procedure for making
DRYPRO 771 settings.
< 53 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
The setting of the DRYPRO 771 can also be implemented using the maintenance
tool. Refer to the Service Manual for details.
< 54 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
U90 PASSWORD /.
4 Press the [enter] key.
The message "INPUT PASSWORD" will be
displayed in the message display window.
Once the service maintenance mode has been started up, it will remain activated
either until it is exited or until the DRYPRO 771 power supply is shut down.
< 55 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
The service
maintenance mode
1 Select "U90 PASSWORD" in the user may be closed by
maintenance mode. switching off the
DRYPRO 771 power
The message "U90 PASSWORD" will be displayed
U90 PASSWORD /. in the message display window. supply.
< 56 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
S09 SCP DEFAULT CH NO Setting of default channel for use by the DRYPRO 771 only.
Select "NO" when connection is made using the basic
configuration. Selection may be made from "NO", "CH1,"
"CH2," "CH3" and "CH4."
S0A HOST NAME blank Setting of name for DRYPRO 771 on the network.
A maximum of 16 alphanumerical characters may be input.
< 57 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
When [exit] key is pressed in the "DICOM SCP Setting", a confirmation screen will be
displayed. Select "YES" or "NO" for the register of the changed data using [select]
key [[][\].
When "YES" selected : Changes to the setting (data) are registered, and the
display returns to the maintenance menu screen.
< 58 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
Service Manual
3.5 DICOM SCU Settings S10 DICOM SCU /. Web Maintenance
The procedure for setting DICOM SCU (DICOM information for diagnostic devices connected to Tool>
the DRYPRO 771) using the service maintenance mode, is described below. SCU>
The following DICOM SCU items are available for setting. DICOMSCU
S11 CHANNEL USE : This setting determines whether or not to use the channel of the selected
diagnostic device.
"ON" or "OFF" may be selected. default setting : OFF
S12 SCU NAME : Sets the name of the selected diagnostic device.
A maximum of 12 alphanumerical characters may be input.
The name set here is used in the DRYPRO 771 only and will be displayed as
the diagnostic device name.
S13 N-EVENT PORT : Sets the N-EVENT PORT number.default setting : 00000
S14 IP ADDRESS : Sets the IP address of the selected diagnostic device.default setting : 0.0.0.0
S15 AE TITLE : Sets the AE title of the selected diagnostic device.
A maximum of 16 alphanumerical characters may be input.
default setting : blank
S16 HOST NAME : Sets the name of the selected diagnostic device to be used on the network.A
maximum of 16 alphanumerical characters may be input.
default setting : blank
S17 BORDER SIZE : Sets the output film border size."REGIUS" (no margin) or "CT/MRI" (with
margin) may be selected.default setting : CT/MRI
"REGIUS" : Border with almost no margin used, for example, with
REGIUS CR.
"CT/MRI" : Sets a margin suitable for CT or MR multi-format images.
S18 FILM SIZE/TYPE : This setting determines whether or not a notification of film size and type check
is given.
"NO CHECK" or "CHECK" may be selected. default setting : CHECK
S19 EMPTY NOTICE : This setting determines whether or not notification is sent to the diagnostic
device when the supply tray is empty.
"ON" or "OFF" may be selected. default setting : ON
S1A DICOM ERR LV : Sets the method of return of DICOM errors from the DRYPRO 771 to diagnostic
devices.
"STANDARD," "FAILURE" or "N-ACTION" may be selected.
default setting : STANDARD
"STANDARD" : When the diagnostic device checks the DRYPRO 771 status
(N-GET), the DRYPRO 771 returns to the diagnostic device a
"WARNING" signal to indicate, for example, that the supply
tray is empty and a "FAILURE" signal for serious errors.
"FAILURE" : A "FAILURE" signal is returned for all errors.
"N-ACTION" : In cases where the diagnostic device does not check the
DRYPRO 771 status (N-GET), the DICOM connection is
automatically broken (association aborted) by a print
command (N-ACTION).
< 59 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
S1B FRAME LAYOUT : Sets the spacing between frames in multi-format output.
"EVEN"(even layout) "CENTER" (centered layout), "H EVEN" (even layout, Hitachi
format) or "H CENTER" (centered layout, Hitachi format) may be selected.
default setting : EVEN
"EVEN" : Film is printed with the top/bottom/left/right margins unchanged
and the spaces between frames as wide as possible. (Even
layout)
"CENTER" : Film is printed with the spaces between frames uniform and the
top/bottom/left/right margins broadened. (Centered layout)
"H EVEN" : Selected to produce even layout when a Hitachi diagnostic
device is connected.
"H CENTER" : Selected to produce centered layout when a Hitachi diagnostic
device is connected.
S1C DMAX/DMIN : This setting determines whether maximum and minimum densities will be printed
according to the DRYPRO 771 settings or the diagnostic device settings.
"DICOM" (diagnostic device) or "P-CON" (DRYPRO 771) may be selected.
default setting : DICOM
S1D P-VALUE RCV : Sets the tone conversion method of the image.
Select from "OFF" (as in the past) and "ON" (tone conversion using "P" value) .
default setting : OFF
Relation among Status of the DRYPRO 771, "DICOM ERR LV" setting and communication with external
devices.
Ready for print warming up
(including the case where "S19 EMPTY NOTICE" is set to "OFF" and film empty.
Status of DRYPRO 771
DICOM ERR LV DICOM Comm. Message Queue Reg.
displayed on external device
STANDARD NORMAL A/V READY
FAILURE NORMAL A/V READY
N-ACTION NORMAL A/V READY
In Error Status
note 2 : Queue registration is available when output under "WARNING" is set to be allowed on the external diagnostic device.
note 3 : Forcefully interrupt the DICOM connection using the print command (N-ACTION).
< 60 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
< 61 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
S21 - S2B
3.6 PRINT CONDITION Settings S20 PRINT COND. /. Service Manual
The procedure for setting printing parameters for each of the diagnostic devices connected is Web Maintenance
described below. Tool>
Following items for PRINT CONDITION needs to be set. SCU>
STAMP
S21 STAMP SELECT : This setting determines whether or not to print the date as a stamp.
TIME/DATE "ON" or "OFF" may be selected. default setting : ON S2C - S2G
S22 STAMP SELECT : This settings determines whether or not to print the printed sheet count Service Manual
COPY NO. as a stamp."ON" or "OFF" may be selected. default setting : ON Web Maintenance
S23 STAMP SELECT : This settings determines whether or not to print the patient ID as a stamp. Tool>
PATIENT ID "ON" or "OFF" may be selected. default setting : ON SCU>
S24 STAMP SELECT : This settings determines whether or not to print the ID and page ID as a
DEFAULT SETUP
SCU ID/PAGE ID stamp."ON" or "OFF" may be selected. default setting : ON
S25 STAMP SELECT : This settings determines whether or not to print the message print
STAMP MESSAGE count as a stamp."ON" or "OFF" may be selected. default setting : ON
S26 STAMP SELECT : Sets the print format of the patient ID and the patient name.
PATIENT DATA : "ASCII" (alphabetical input) or "2BYTE" (full-size character input) may be
selected.default setting : ASCII
The S26 setting is valid
S27 STAMP MESSAGE : Sets the message to be printed as a stamp. A maximum of 32 alphanumerical
only when S23 is set to
characters may be input.
ON.
default setting : blank
S28 STAMP DIRECT : Sets the stamp position.
"UPPER," "LOWER," "RIGHT" or "LEFT" may be selected.
default setting : LOWER
S29 YEAR FORMAT : Sets the year format.
"YY" (year expressed in 2 digits) or "YYYY" (year expressed in 4 digits) may be
selected. default setting : YY
S2A MONTH FORMAT : Sets the month format.
"NUMERIC" (month expressed as a numerical value) or "TEXT" (month
expressed in abbreviated alphabetical character form) may be selected.
default setting : TEXT
S2B DATE FORMAT : Sets the date format.
"TIME+DATE" (expressed in the order time, date), "DATE+TIME" (expressed in the
order date, time), "TIME" (time only) or "DATE" (date only) may be
selected.default setting : TIME+DATE
S2C REQ. IMGSIZE : Sets the size of one frame.
"EFFECTIVE" (size specified) or "INVALID" (size not specified) may be
selected.default setting : EFFECTIVE
S2D EXPANSION CHAR : This setting determines whether or not to expand the patient ID and name font
to a 2-line display.
This setting is valid only when "CT/MRI" has been selected under the DICOM SCU
S17 BORDER SIZE setting.
"ON" or "OFF" may be selected.default setting : OFF
S2E EDGE ENHANCE : This setting determines whether or not to show characters in the image in an
easily readable format.
Selection of "ON" will result in characters being shown slightly brighter.
"ON" or "OFF" may be selected.default setting : OFF
< 62 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
S2F REQ. BEHAVIOR : Select the action when the image size sent from the connected device is larger
than the installed film size.
Select from "DECIMATE" (proportional display) and "CROP" (length display).
Selecting "DECIMATE" prints a reduced image by fitting the image size to the
film. The reduction ratio will be printed on the output film.
Selecting "CROP" does not reduce the image, but trim the excess of the image
for the print to fit to the film. The trimmed amount in length will be printed on the
output film.
default setting : DECIMATE
S2G FLIP : This setting determines whether display shall be highlighted.
"ON" or "OFF" may be selected.
default setting : OFF
RIGHT
Image Ori.
Image Ori.
LOWER
UPPER LEFT
LOWER
UPPER RIGHT
LEFT
RIGHT
LOWER
UPPER
LOWER LEFT
UPPER RIGHT
LEFT
RIGHT
Image Ori.
LOWER
UPPER
LEFT
< 63 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
and settings.
After registering the settings, the screen returns to
the maintenance menu screen.
When [exit] key is pressed in the "PRINT CONDITION Setting" of the service
maintenance mode, a confirmation screen will be displayed. Select "YES" or "NO" for
the register of the changed data using [select] key [[][\].
When "YES" selected : Changes to the setting (data) are registered, and the
display returns to the maintenance menu screen.
< 64 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
External Diagnostic Device
DR_HIGH_C
USER 1 USER 1 USER 1 USER 1
User LUT USER 2 USER 2 USER 2 USER 2 Specifies the LUT No. registered on
x4 USER 3 USER 3 USER 3 USER 3 DRYPRO 771 for image processing
(able to edit) USER 4 USER 4 USER 4 USER 4
P-CON DICOM
(2) Edit the User LUT (3) Apply LUT in LUT (5) Setting the DICOM PRIORITY
curve point by point. Library to each CH S31 DEFAULT LUT .
of DRYPRO 771 DICOM
U52 LUT1 020-300 .
POINT01 020 X0.01D U60 LUT LIBRARY /.
Uses the standard LUT when the LUT is not
CH 1 MRI-1
specified by the external diagnostic device.
Image Processing
Use web maintenance tool to copy the
User LUT from one CH to other.
(3) Applying the LUT in the LUT Library to each CH of the DRYPRO 771.
p.71
Apply the LUT from the LUT Library which is provided by the manufacturer or the User LUT
to the channel 1~7 through which the corresponding external device is affected.
(same operation can be implemented using "web maintenance toolSCULUT")
(4) Selecting the standard LUT which will be used by each channel.
Set the standard LUT that will be used when the LUT is not specified by the external device p.73
or when "S31DEFAULT LUT" is set "P-CON".
(same operation can be implemented using "web maintenance toolSCUDEFAULT
SETUP)
(5) Setting the DICOM PRIORITY
Select between the LUTs used for image processing, one for standard LUT (P-CON) set for p.75
DRYPRO 771 and the other specified for each image processing by the external device
(same operation can be implemented using "web maintenance toolSCUPRIORITY
SETUP)
< 65 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
To apply the LUT in the library directly to the LUT of the DRYPRO 771, procedures in "(1)
Copying the Library LUT to User LUT and (2) Editing the User LUT curve point by point" are not
necessary. Proceed to "(3) Applying the LUT in the LUT Library to each CH of the DRYPRO
771".
< 66 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
< 67 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
(3) Applying the LUT in the LUT Library to each CH of the DRYPRO 771.
Applies the LUT in the library, which is prepared in advance or the User LUT to any of channel 1~7
through which the external device is affected by the LUT.
(same operation can be implemented using "web maintenance toolSCULUT")
< 68 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
Differences among "KC", "LOW", "HIGH", "LINEAR" of each LUT are illustrated as
below.
< 69 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
(4) Selecting the standard LUT which will be used by each channel.
Set the standard LUT that will be used when the LUT is not specified by the external device or
when "S31DEFAULT LUT" is set "P-CON".
At the same time, "U30 PRINT COND." of the user maintenance mode is set.
Set the print condition for each diagnostic device to handle the case where there is no print
condition specified by the diagnostic device.
Print conditions to be set are listed below.
U31 LUT NO. User LUT USER1 Selection of User LUT with which the print condition is
set.Select 4 User Luts from 7 User Luts. Select from
"USER1", "USER2", "USER3", "USER4".
U32 DENSITY Density 0 Image density setting. Settings may be made within a
range of -7 to +7. Incrementing the value produces lighter
density, decrementing the value produces darker density.
U33 CONTRAST Contrast 0 Image contrast setting. Settings may be made within a
range of -7 to +7. Incrementing the value produces higher
contrast, decrementing the value produces lower contrast.
U34 SMOOTHING Smoothing Type 2 Image smoothing type setting. Settings may be made
within a range of 0 to 7. In the range of "0" through "6", the
higher the value, the smoother the image. However, the
image becomes blurred when the set value is too large.
Setting at "7" gives the same smoothness as "0" while
prints the image at 100% without magnification.
Recommended setting is "2" or "3".
U35 MAX DENSITY Max. Density 300
Setting of the maximum density for printing.
The higher the value, the higher the maximum density, the
lower the value, the lower the maximum density.
U36 MIN DENSITY Min. Density 20 Setting of the minimum density for printing.
The higher the value, the higher the minimum density, the
lower the value, the lower the minimum density.
U37 ORIENTATION Film Orientation PORTRAIT Selection may made between "PORTRAIT" and
"LANDSCAPE".
U38 IMAGE POLARITY Image Polarity POSI Selection may be made between "POSI"(positive) and
"NEGA"(negative).
U39 TRIM ON/OFF Trim ON/OFF CH1 : OFF Selection may be made between the image printed with
CH2 : ON trim or not..
U3A BORDER Border Color BLACK Selection of border colour. Select from black (BLACK)
or white (CLEAR).
U3B ILLUMINATION Light Table 2000 Light table brilliance setting. Normally it is sent from the
Illuminance modality. Setting should be made in the range of "0"
through "38000".
< 70 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
When [exit] key is pressed in the "Print Parameter Settings", a confirmation screen
will be displayed. Select "YES" or "NO" for the register of the changed data using
[select] key [[][\].
When "YES" selected : Changes to the setting (data) are registered, and the
display returns to the maintenance menu screen.
< 71 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
< 72 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
.
.
6 Input the necessary characters or digits in each
. of the setting windows.
.
When [exit] key is pressed in the "PRINT CONDITION Setting" of the service
maintenance mode, a confirmation screen will be displayed. Select "YES" or "NO" for
the register of the changed data using [select] key [[][\].
When "YES" selected : Changes to the setting (data) are registered, and the
display returns to the maintenance menu screen.
< 73 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
When [exit] key is pressed in the "FILM SETUP" of the service maintenance mode, a
confirmation screen will be displayed. Select "YES" or "NO" for the register of the
changed data using [select] key [[][\].
When "YES" selected : Changes to the setting (data) are registered, and the
display returns to the maintenance menu screen.
S71 PATCH CONTROL ON Select whether to implement the density control using the density
patch.
S72 LOG WRITE MODE NORMAL Select the record level with which the logs of system operation is
recorded.
The log contains the record of initial 20 minutes from the start.
Select from "NORMAL" , "DETAILS".
Leave at "NORMAL" unless otherwise instructed.
S73 TIME ZONE 64 Select the time zone for the local where the DRYPRO 771 is
installed.
S74 LANGUAGE ENGLISH Select the language used for messages in user screen and screens
for user maintenance menu."ENGLISH" , Select from "GERMAN" ,
"FRENCH" , "SPANISH" , "ITALIAN" , "PORTUGUESE" ,
"SCANDINAVIA".
Service maintenance mode screens are always displayed in English
disregard of the language chosen here.
The current version of the DRYPRO 771 can accept English and
German.
S75 TOTAL FILM CNT - Displays the total print count since the DRYPRO 771 is installed.
Do not touch at installation.
S76 OPE. HOURS - Displays the total running time since the DRYPRO 771 is installed.
Do not touch at installation.
< 75 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
When [exit] key is pressed in the "SYSTEM SETUP", a confirmation screen will be
displayed. Select "YES" or "NO" for the register of the changed data using [select]
key [[][\].
When "YES" selected : Changes to the setting (data) are registered, and the
display returns to the maintenance menu screen.
< 76 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.3 Settings
Time Code Time Zone Summer Time Code Time Zone Summer
Time Time
< 77 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.4 Checking the Image
Check the iternal clock of the DRYPRO 771, and correct if necessary.
Set "YEAR", "MONTH", "DAY", "HOUR" and "MINUTES".
Data Saving...
Please wait...
.
5 Press the [enter] key after completing the
setting.
Settings will be registered and display returned to
the maintenance menu screen.
UA0 TIME SET /.
When [exit] key is pressed in the "TIME SET" of the service maintenance mode, a
confirmation screen will be displayed. Select "YES" or "NO" for the register of the
changed data using [select] key [[][\].
When "YES" selected : Changes to the setting (data) are registered, and the
display returns to the maintenance menu screen.
SG0 SCHEDULE
MNT.A RESET
/.
4 Select the check schedule to be initialized using
[select] key.
Check schedule item will switch.
RESET OK? .
YES
5 Press the [enter] key.
A confirmation screen will be displayed. Use
[select] key to select "YES" or "NO" for
initialization of the schedule.
Selecting "YES" will initialize the schedule, and the
SG0 SCHEDULE /.
screen returns to the maintenance menu screen.
MNT.A RESET
Selecting "NO" will return to the maintenance
menu screen without initializing the schedule.
< 79 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.4 Checking the Image
< 80 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
4
Checking the
Image Quality This chapter describes the procedure for checking
the image quality..
< 81 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.4 Checking the Image Quality
When a densitometer is available at the facility, use it. If it is not available, use the
one you bought with.
(00)
(01)
(02)
2 Using the density pattern of Service
(03)
(04)
Maintenance Mode "S50", print out a density
(05)
(06) correction sheet.
(07)
(08)
(09)
(10)
S50 DENSITOMETR
INPUT DENSITY
/.
4 Select "INPUT DENSITY" of the Service
Maintenance Mode "S50", and press [enter] key.
Input field for "POINT01" will be displayed.
S51 OUT DENSITY /. 5 Using the ten key, input the density readings
POINT01[000]x0.01D measured in step 3.
Press [] key of the [menu/select] key after
inputting the value in "POINT01" will switch to the
input field of "POINT02".
.
.
.
.
6 Upon completion of inputs for all
"POINT01"~"POINT10", press [enter] key.
The settings will be registered, and returns to the
maintenance menu screen.
When [exit] key is pressed, a confirmation screen
S50 DENSITOMETR /.
will be displayed, and prompts to select YES or NO
INPUT DENSITY with the [select] key for registration of the settings.
< 82 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.4 Checking the Image Quality
< 83 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.4 Checking the Image Quality
< 84 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.4 Checking the Image Quality
< 85 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.4 Checking the Image Quality
< 86 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.4 Checking the Image Quality
If any abnormality is found on the printed film, refer to the "Troubleshooting" of the Service
Manual for remedy.
< 87 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.4 Checking the Image
Density Linearity
Criterion Correction 1 Difference Correction 2 Difference
0 20 20 0 20 0
350
1 39 40 1 39 0
2 57 60 3 56 -1
300
3 76 80 4 75 -1
4 95 100 5 94 -1
250
5 113 121 8 111 -2
6 132 143 11 129 -3 Criterion
200
7 151 160 9 150 -1
Correction 1
8 169 177 8 171 2
150 Correction 2
9 188 195 7 189 1
10 207 212 5 207 0
100
11 225 230 5 226 1
12 244 247 3 246 2 50
13 263 265 2 265 2
14 281 282 1 282 1 0
15 300 300 0 300 0
1 3 5 7 9 11 13 15
( x0.01D)
The above is an example where the film type is "DRB" or "B", and because
the first correction failed to bring the difference within +/- D=0.05, SMPTE
pattern is again printed and correction is made again.
Density Linearity
Criterion Correction 1 Difference Correction 2 Difference
0 10 12 2 1 1
350
1 29 32 3 2 3
2 49 50 1 8 1
300
3 68 70 2 9 1
4 87 91 4 8 1
250
5 107 108 1 109 2
6 126 126 0 128 2 Criterion
200
7 145 147 2 146 1
Correction 1
8 165 166 1 167 2
150 Correction 2
9 184 187 3 187 3
10 203 209 6 205 2
100
11 223 229 6 224 1
12 242 246 4 243 1
50
13 261 265 4 263 2
14 281 282 1 281 0 0
15 300 300 0 300 0
1 3 5 7 9 11 13 15
( x0.01D)
The above is an example where the film type is "C", and because the first
correction failed to bring the difference within +/- D=0.05, SMPTE pattern
is again printed and correction is made again.
If an abnormality occurs with the printed image in the test print, take necessary
means to correct it according to the service manual.
Pay attention not to leave the printed flat pattern that is printed for the installation
adjustment, visible to the staff of the facility. They should be brought back.
< 88 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
5
Print Commanded
from the External
Device and Backup This chapter describes the procedure for print
commanded from the external device and backup.
< 89 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.5 Print Commanded from the External Device and Backup
< 90 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.5 Print Commanded from the External Device and Backup
5.2 Backup
Procedures to back up the DRYPRO 771 software program and the set up files to protect from the crush of
the compact flash memory are detailed below.
Back up the CF memory data on the Hard Disk of the DRYPRO 771.
Back up the initial set up files created at the time of installation on the maintenance PC using the Web
maintenance tool.
< 91 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.5 Print Commanded from the External Device and Backup
5.2.2 Back Up of the Initial Set Up Data Created at the Time of Installation
How to back up entire set up data and information including the data identical to the unit using the Web
maintenance tool is set out below.
System Requirements
To use the Web maintenance tool, the maintenance PC should satisfy the following
requirements.
PC with 10BaseT/100BaseTX network port.
PC with TCP/IP protocol installed.
PC that operates on Windows95/98/2000/XP.
Supports Internet Explore (ver.5.5 or higher).
Java applet to be installed.
When the system of your PC is doubtful, download
"j2re_1_3_1_win_i_exe" from the SUN's homepage.
< 92 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.5 Print Commanded from the External Device and Backup
7 Click [LOGIN].
Top screen of the Web maintenance tool will be
displayed.
9 Click [OK].
File download dialogue will be displayed.
< 93 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Ch.4 Checking the Image
< 94 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Specifications
This chapter details the specifications of the
DRYPRO 771.
< 95 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Specification
External 630 (W) X 600 (D) X 1125 (H) mm (24.8 (W) X 23.7 (D) X 44.3 (H) inch)
Dimensions
Weight 175 kg
External 790 (W) X 780 (D) X 1410 (H) mm (30.8 (W) X 31.0 (D) X 55.2 (H) inch)
Dimensions
(Packed)
Weight 205 kg
(Packed)
< 96 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
Specification
Image Matrix
When Connected to REGIUS When Connected to Units Other than REGIUS
14 X 17 inch 4,503 x 5,419 pixel 4,351 x 5,114 pixel
14 X 14 inch 4,500 x 4,452 pixel 4,351 x 4,236 pixel
11 X 14 inch 3,498 x 4,503 pixel 3,396 x 4,236 pixel
Number of 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 25, 30, 35, 36, 42, 48, 54, 56, 60, 63, 64
Recorded Frames
Positive/Negative Available.
Trim Available.
Border Black or Clear.
Stamp Recorded on 1 line or 2 line of film (May be printed at top/bottom/left/right).
Density Furnished in main body.
Correction
Start timer Day/time settings possible.
Applicable IEC60601-1-2 : 2001, IEC60601-1, IEC60825
Standards
Remarks DRYPRO 771 is manufactured in factories certified in compliance with ISO9001 : 2000, EN
ISO13485 : 2000, 93/42/EEC quality control standards as well as with medical device directive
FDA Pre market Notification 510(k) and GMP.
Accessories Deodorant filter unit, Suction cups (spare), cutter, One-touch Spacers (2 pcs), Power cable,
Clamp (2 pcs), Exhaust duct, Warranty, Operation manual, Operation sheet, Attached
document for doctors, Inspection sheet, Protective parts(red) collection bag,
Accessory list
Options Deodorant filter
(available at cost)
< 97 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
0582-55031 A
Konica Minolta Medical Imaging KONICA CANADA INC. KONICA EUROPE GmbH
U.S.A., Inc. 1329 Meyerside Drive, Mississauga, Ontario L5T 1C9, Friedrich-Bergius-Str.6,
411 Netwark Pompton Turnpike, Wayne, NJ 07470, U.S.A CANADA D-85662 Hohenbrunn, GERMANY
TEL.973-633-1500 TEL.905-670-7722 TEL.8102-8040
0310TE