OPERATING MENUAL
1
Tru-Test Operation Manual
Key Function
2
Power Switch - on/off
Key Beep - on/off
Increase and decrease Gain
Auto Calibration
3
Initial Start-Up
Power Supply
The Tru-Test can be operated with an external power supply through AC/DC adaptor or with built-in Lithium-Ion battery pack.
It can also be powered through the universal power supply/ battery charger from any standard AC outlet from 110-230V
Internal Charging
4
If a battery is located in the instrument, the charging process is start automatically when you connect the plug-in power supply unit. The instrument can be used
while the battery is being charged. The charging time is approx.8 hours with turn it off and 10 hours with turn on. This charging time applies to ambient
temperatures from 25-30 degree C. Please take into consideration that the battery pack is not charged to their full capacity at higher temperatures.
The LED display on the plug-in power supply unit indicates the status of the charging process. First plug the power supply to the main power ( 220 VAC or 110
VAC ). The LED display on the Power supply/Charger indicates the status of the charging process. When battery fully charged, the power supply charger LED
indicator will turn red into green.
External Charging
The battery can also charged by plugging the charger adaptor cable directly into the connector below the slotted screw on the battery pack.
Charge procedure is same way as above.
Connecting UT Probes
The prepare of Tru-Test for operation, you have to connected a probe to it. Any brand probe can be use for the Tru-Test, provided the appropriate cable is
available and the operating frequency is within an adequate range.
The Tru-test is available with the probe cable connectors Lemo1.
Probe is connected to the socket at top right on the instrument. Both connector sockets are equally suitable ( connected in parallel) for connecting probes
equipped with only one ultrasonic element (ultrasonic transducer) so that it does not matter which one of the two sockets is used.
When connected Dual (TR) probe (having one transmitter element and one receiver element), or two probes (of which one is transmitting and the other one
receiving), attention should be paid to connecting the transmitter element to the right-hand sockets (transmitter, marked with red circle at the rear of the
instrument case) and the receiver element to the left-hand socket (receiver, marked with black circle). Verify correct polarity by reversing the connection and
comparing signals. When the connection is correct, the signal is stronger.
5
The setting of all function values and the basic setting are same as prior to the instrument being turn off.
Reset
If any functions can not be operated after a warm start, or if you want to reset the instrument to the basic defrost setup, then you should carry out a cold start by
simultaneously pressing the
and Key the instrument will reset to the basic setup and clean all memory file
and key the instrument will reset to the basic setup and still keep all memory file
After press power on key turn on Tru-Test, press again will turn off keypad sound
6
Screen Display
The Tru-Test had a digital screen display of A-scan in normal mode
Echo Amplitude
Surface Distance
Depth
P Measurement Mode
Peak or Flank
Battery Indicator
7
Tru-Test Display Sub-Menu Guide
First press GATE Second press GATE First press SETa Second press SETb First Press PULS
into BASE into a GATE into b GATE into DATE into COLOR into PULSE
RANGE a START b START HORN GROUND TYPE
0250mm 0050mm 0150mm OFF
FINE G a WIDTH b WIDTH BRIGHT WORD POWER
0.1dB 0100mm 0050mm HIGH 400V
MTL VEL a THRESH b THRESH UNIT ECHO * DAMP
5920m/s 88.0% 25.0% mm 200 OHM
D-DELY a LOGIC b LOGIC DATE GRID REF-dB
0000mm POS POS 11-28-06 +00dB
P-ZERO ab GATE b GATE TIME SOLID PRF
0.00us OFF OFF 1528 OFF 100Hz
8
Tru-Test Display Sub-Menu Guide
Into RECV Into TRIG Second Press TRIG Into PK MEM Into SAVE Into RECALL Into DAC
9
Basic Setting
button setting the day and year. After setting all correct date. Press
key twice time back to a SET starting, press twice time, the
date will be memorized
Setting the time
10
In the function group, press key, Enter a SET function.
In the function group, press key twice time, Enter b SET function.
11
OVERVIEW OF THE FUNCTION
When setting all parameter, Tru-Test operated via 3 keys, enable to quickly and direct
set the gain.
- Press the key to set up the parameter by moving the arrow to different
function setting
The gain control function is always directly available from left three key.(see right side)
Press select a certain increment for setting the gain. , dB setting choice from
1,2,6,12,20 dB 5 steps setting.
When setting to 20dB, user can press and setting the step gain
12
Adjusting the Display Range ( Function Group BASE )
The function BASE enable you to make the basic adjustment of the screen display range. The display on the
screen must be adjusted for the material to be tested.( function MTLVEL) and for the probe used (function for
P-ZERO).
Note:
In order to accurately adjust the material velocity and the probe delay, please read the section Calibration of
Tru-Test.
The adjustment from 25, 50, 125, 200, 250, 500, 1000, 2000, 4000, 6000 mm. by and
Press the key, the high-light on the left of RANGE will display RANGE. Press and to fine range adjustment.
When press key, the range display will return back to the 25 mm, user can set another display range .
13
MTVEL (Sound Velocity)
Use MTVEL to set sound velocity within the test object. You can use sound velocity between 1000-9999m/s
Coarse adjustment 5920 m/s or 3240 m/s. press key move high-light cursor to MTVEL, then choose by and
Press key for fine adjustment, the high-light on the left of MTVEL will display MTVEL, Fine adjustment 1000-9999m/s in step of 1m/s through
and key.
14
D-DELAY ( Display Starting Point)
This function allow you to shift the complete screen display and consequently also the display zero. If the display should for example start from the surface of
the test object, the value in D-DELAY must be set to 0.
press key move high-light cursor to D-DELY, press and key to increase or decrease the screen delay
Press key move high-light cursor to P-ZERO, press and key to increase or decrease the screen delay0.01 us per step
Or you can press key, When high-light show P-DELY, press and key to double the value
15
Adjusting the Pulse
Use the function POWER to set the pulse voltage. You can choose between two setting ( 400V and 200V). The setting high REF-dB
(400V) is recommended. +00dB
Inspections in which maximum sensitivity is important, e.g. for the detection of small flaws. Choose the setting low (200V) for PRF-MOD
broadband probes or if narrow echoes are required (better resolution). 100Hz
press key move high-light cursor to POWER, press and key to select 400V or 200 V
DAMPING (Probe Matching)
This function serves for matching the probe. the damping of the probes oscillating circuit and consequently change the
height, width and resolution of the echo display(100 or 200)
- Damping 200, this setting matching the echo height but mostly also produces narrow echoes with higher resolution.
REF-dB
This function for inspector follow the code setting or regulation, press and key set the REF dB
16
PRF-MOD ( Pulse Repetition Frequency) TYPE
- The pulse repetition frequency indicates the number of times an initial pulse is triggered per second. You can determine
whether you need the highest possible PRF value, or whether you are satisfied with a low value. POWER
- You 8 step selection (10 25 50 100 250 500 750 1000Hz) 400V
- The larger you work piece, the smaller PRF value are needed in order avoid phantom echoes. In the case of the smaller DAMP
PRF value, however, the A-scan update rate becomes lower, for this reason, high value are required if a work piece 200 ohm
should be scanned fast. The best way to determine the suitable PRF value is by experimenting; start from the highest REF-dB
step and reduce the value until there are no such phantom echoes. +00dB
PRF
- Press and key set the PRF 100Hz
17
Adjusting the receiver
REJECT
The function REJECT allows you to suppress unwanted echo indications, for example structural noise from your test object. REJECT
The % screen height setting indicates the minimum height that the echoes should attain in order for them to be displayed on the 00%
screen at all. The reject setting cannot be higher than the lowest threshold setting ( min 1%) of any gate. MHz
0.5-5
- Press key move high-light cursor to REJECT, press and key increase or decrease the reject % RF-MODE
You should handle this function carefully; as it may of course happen that you suppress echoes from flaws as well. Many test OFF
specifications expressly forbid using the reject function. AGC 80%
The LED R is therefore light with active when REJECT function on OFF
MHz (Frequency Range)
In this function, you can adjust the operating frequency according to the frequency of your probe.
You have a choose between 2 frequency range
- Narrow Band from 0.5- 5 MHz
- Board Band from 4-15 MHz
- Press key move high-light cursor to FREQE, press and key to select narrow or board band
18
RF-MODE ( RF=Radio Frequency)
You can select the rectification mode of the echo pulses according to your application in the function RF-MODE. When RF mode on,
it applies to the display range up to 50 mm ( steel). If user want to increase the range, press key to adjust D-DELAY.
- Press key move high-light cursor toRF-Mode, press and key to select RF ON or OFF
( When choose RF on, The screen display will automatic change into 25 mm full Screen )
19
AGC 80% ( Automatically increasing the gain raise the echo amplitude to 80% high )
- Press key move high-light cursor to REJECT, press , the echo cross the gate will automatic increase to 80% height.
See below
20
Setting The Gates (Function Group aGAT and bGAT)
Display of Gates
All functions for setting the (dual) gate are arranged in the function group aGAT and bGAT.
Select the function group aGAT or bGAT
21
instrument is used. i.e. when function parameters are changed. Possible alarms occurring during instrument
operation (setting of function) are to be ignored. To make the assignment easier, the gate are displayed in
different colors. You can not vary the colors of gates because they are fixed as following
Gate A yellow
Gate B - Green
aSTART/bSTART (Starting point of the gates)
You can fix the starting point of the gates A or B within the adjustment range of 0-5999 mm.
- - Press one time or two time select the gate and choice aSTART or bSTART
- Use the and key to set gate right position right value
22
- - Select the function aTHRSH or bTHRSH
23
Setting TRIG
24
Setting CAL
- MEAS = PEAK and Flank ( select depth measurement use Peak or Flank, Usually we use flaw detect by Peak Mode, thickness for
Flank Mode )
- S-REF-1( Setting the reference distance or measurement)
- S-REF-2( Setting the reference distance or measurement)
- a START( Gate A start point)
- b START(Gate A start point)
Setting DAC
25
Function Group PK.m
26
ENVOLOP
Press ENVOLOP=on can hold the all the echoes cross the gate
Put probe on 20 mm thick block to get Open envelope function, all the echo pass the gate will
back echo cross the gate hold and display on the screen.
SHADOW
Press SHADOW=on. The initial and background echo will change color into solid display
Put probe on 20 mm thick block to get Press SHADOW set it on, the echo color will change When move the probe. The show still hold on the
back echo cross the gate into shadow display type display for reference
27
Time B-Scan
1. Select the probes which already calibrated the probe delay. If its a new probe, please follow the probe delay measurement procedure calibrate the probe
delay
2. Press BASE key, enter probe delay ( measurement by Peak method )
3. Press CAL set MEAS=Peak
4. Press PULS, set contact probe
5. Press BASE, press set the screen display range to 50 mm
6. Put the probe on the step block make sure the gate can pass the min. and max. range.
7. Adjust the gain to get 75% amplitude from thick step as following
8. check the min step 20 mm back echo, see Sa=20.0 mm
9. Move the probe on the 20 mm step. Press PK MEM key, press T-SCAN right arrow key set T- scan on and start time-Bcan
Move the probe from 20 mm to 10/5 mm step. After completed of the T-scan,
press enter key and memorize the data.
10. Press COPY recall key, recall the Time B-scan data.
11. This Time B-scan map show the step block from 20-10-5mm.
28
5 MHz Contact Probe Delay Measurement
Select one 5 MHz contact probe connected to the cable and plug it to the Tru-test
Get 1 4 step thickness block ( 20, 10, 5 , 2.5mm )
Press Tru-Test ON/OFF switch, turn on the Tru-test.
Press and key to select probe type single contact probe symbol.
29
Use Flank
30
Note:
Use Peak or Flank have different measurement value display. The probe delay is different. This will case inspection tolerance. Most of time,
PEAK use for flaw detection and FLANK use for thickness measurement.
Select one 4 MHz, 8x9 mm, 70 degree probe connected to the cable and plug it to the Tru-test.
Press Tru-test ON/OFF switch, turn on Tru-Test.
Put probe on V1 steel test block, target the probe to the 100 mm curve
31
6. Put probe on V1 and target to 100mm curve, get echo
7. Set gain until echo rise 80% full screen
8. Press key. Move the gate across the echo and adjust the gate width, set gate thresh to 80%
9. Press key, enter TRIG set angle =70 degree, s-VALE=12.0 mm.
10. Press key back BASE setting, increasing P-ZERO until the sound path= 100mm
11. check Sa=100.0. Now the P-DELY for this probe is 3.89 us
32
4 MHz Dual Crystal Contact Probe Delay Measurement
Select one 4 MHz,12mm, 0 degree dual probe connected to the cable and plug it to the Tru-test.
Press Tru-test ON/OFF switch, turn on Tru-Test.
1. Press CAL key, select the MEAS= PEAK
2. Press BASE key, set range 25 mm
3. Set MTL VEL=5920 m/s
4. Press PULS, selecting probe type= dual probe.
5. Put probe on 20 mm step block
6. Set gain until echo rise 80% full screen
7. Press TRIG set angle = 0 degree, s-VALE=0.0 mm.
8. Press GATE. Move the gate across the echo
9. Press BASE back BASE setting
10. Increasing P-ZERO Until Sa=20.0mm
11. check Sa=20.0. Now the P-ZERO for this probe is 3.39 us
33
Its very important to calibrate the probes delay and record it. If you did not calibrate probe delay and entry the parameter. The inspection result will generate big
tolerance. Each probe has different probe delay.
34
13. Move the probe to 10 mm step, move gate across back echo, the Sa=10.0 and Da=10.0
14. Move the probe to 10 mm step, move gate across back echo, the Sa=5.0 and Da5.0
35
Dual Probe Thinckness measurement
Select one 4 MHz,12mm, 0 degree dual crystal probe connected to the cable and plug it to the Tru-test
Press Tru-Test ON/OFF switch, turn on the Tru-Test.
1 Press the CAL key, select the MEAS= PEAK
2 BASE key, set range 25 mm
3 Set MTL VEL=5920 m/s
4 Press PULS to Dual probe
5 Put probe on 20 mm step block
6 Set gain until echo rise 80% full screen
7 Press TRIG set angle = 0 degree, s-VALE=0.0 mm, MTL THK =20.0 mm
8 Press GATE. Move the gate across the echo and adjust the width and thresh to 50%
9 Press BASE back BASE setting
10 Move the gate across the echo
11 Press P-ZERO tight arrow key increasing P-ZERO until Sa=20.0
12 check Sa=20.0. Now the P-DELY for this probe is 3.39 us
13 Move the probe to 10 mm step, move gate across back echo, the Sa=10.0 and Da=10.0
14 Move the probe to 10 mm step, move gate across back echo, the Sa=5.0 and Da=5.0
36
Note: When use Flank, the probe delay calibrate is different.
Select one 5 MHz contact probe connected to the cable and plug it to the Tru-Test.
Get 1 4 step thickness block ( 20, 10, 5 , 2.5mm )
Press Tru-test ON/OFF switch, turn on Tru-Test.
3. Press key move high-light cursor to MTLVEL set the velocity = 5920 m/s
4. Press key move high-light cursor to P-DELY set the probe delay=3.72 us, already calibration data before.
37
5. Press CAL set the MEAS=Peak
38
If use FLANK, The probe delay calibrate same as before. Just take care set the aGATE and bGATE down cross echo wave first rise line. This will give you
more precision measurement.
Dual contact probe echo-echo do the same way.
39
Tru-Test Probe Zero Auto Calibration
1. Select the probe type ( contact or angle beam ), frequency, range and velocity
Contact Probe
Use V2 block 20 mm thick. Due to the probe near field resolution, please use second echo calibrate. The second echo should be 40 mm.
Or, to use V1 block 100 mm
Angle beam probe
Target the probe to V1 R100 mm curve
2. Move gate cross and cover the max. response echo
3. Enter CAL mode, Press Set key, now the Means/Peak will show double arrow cursor. This mean, the instrument already enter calibration procedure.
4. Press enter key, now the cursor will move to S-REF1.
5. Use up and down key set 100 mm ( if use V1 block )
6. Press enter key again. Now Tru-Test will auto-calibrate and the sound path will display 100.0
7. Press BASE key, Now the P-delay showed 2.77 us.
8. Press Freeze key, press enter key again, Save this probe calibrate P-delay.
Example
Contact Probe Auto-Calibrate Procedure
1. Get 4 MHz, 12 mm, 0 degree contact probe and V2 block . Press BASE set range into 50 mm
2. Press ENTER key move cursor to MTLVEL. Press Up Key, set MTLVEL 5920 m/s.
- Set angle 0 degree ( if now showed angel=70, press and key select the angle=0.0 degree )
40
5. Plug probe cable to Tru-Test.
6. Put couplant on V2, put probe on the block to get response echo.
8. Press Gate, move Gate A across the second echo and cover it
9. Press key, enter CAL mode, press SET key, the MEAS/Peak showed double arrow.
10. Press enter key, the cursor move to S-REF-1. use and key adjust the thickness into 40 mm
11. press enter key again. Now the probe sound path will show 40.0 mm.
14. Press FREEZE key, Press ENTER key enter SAVE mode. Press key Save the setting
15. Next time when use the same probe, press COPY key, enter recall mod, press ENTER key move cursor to LOAD, Press for
loading the previous calibration setting.
3. Press ENTER key move cursor to MTLVEL. Press SET key, press Up Key, set MTLVEL 3240 m/s.
41
4. Press enter to PULS mode
Set Probe Type-angle beam
5. Press enter TRIG mode
Set angle 60 degree ( if now showed angel=70, press SET key showed double arrow. Press and select the angle=60.0 degree )
6. Plug probe cable to Tru-test
7. Put couplant on V1, put probe on the block target R100 curve to get response echo.
10. Press enter CAL mode, press SET key, the MEAS/Peak showed double arrow.
11. Press enter key, the cursor move to S-REF-1. use and up/down key adjust the thickness into 100 mm
12. press enter key again. Now the probe sound path will show 100.0 mm.
15. Press FREEZE key, Press ENTER key enter SAVE mode. Press key Save the setting
16. Next time when use the same probe, press COPY key, enter recall mod, press ENTER key move cursor to LOAD, Press load the
previous calibrate setting.
42
DAC Curve Setting
3. move high-light cursor to DAC 3C and press up key turn on the DAC 3C on
4. Put probe on the test block and target to the first hole. Adjust the gain amplitude into 80%. Press key set point 1.
5. Move the probe to second hole. Adjust the gate , press set point 2
6. Move the probe to third hole. Adjust the gate , press set point 3
9. Move cursor to 3point DAC, press up key set on, the DAC curve will display
+6dB method DAC Curve
43
+6dB DAC CURVE SETTING
( only use if sound path over 100 mm and probe sensitivity not enough )
12. move high-light cursor to DAC 3C and press up key turn on the DAC 3C on
13. Put probe on the test block and target to the first hole. Adjust the gain amplitude into 80%. Press key set point 1
14. Move the probe to second hole. Adjust the gate , press set point 2
15. Move the probe to third hole. Adjust the gate , press set point 3, , then press the key, the echo will double. Press freeze set point 3
44
again.
16. Press DAC key, the setting will increasing 6dB. Press set point 4
17. Move the probe to no. 4 hole, adjust the gate, press set point 5
18. Move the probe to no. 5 hole, adjust the gate, press set point 6
45
QUIT DAC CURVE, PRESS DAC KEY, PRESS ENTER MOVE CURSOR TO RESET, PRESS DAC AGAIN TO QUIT THE DAC
SETTING
46
WELDING SECTION SIMULATE DISPLAY
( THIS SECTION ONLY AVAILABLE WHEN YOU SET ANGLE BEAM PROBE CONDITION )
PRESS OR ADJUST ECHO AMPLITUDE.
MOVE a GATE CROSS THE DEFECT ECHO PRESS GATE ENTER a GATE SETTING.
SET a GATE POSITION, WIDTH AND LEVEL
PRESSTRIC KEYMOVE CURSOR TO
SETMTL THK MTLPRESS OR OR ENTER
SPECIMEN THICKNESS
PRESSTRIG KEYSELECT AND SET WELD
SET WELD TYPE, WIDTH, ANGLE , PARAMETERPRESS OR KEY SET IT
DISTANCE AND W DISPL
--STEP 1
47
2. EXCUTE PROBE INDEX POINT CALIBRATION, RECORD DATA
3. EXCUTE PROBE ANGLE CALIBRATION
4. AFTER COMPLETED THE CALIBRATION, PRESS TRIG KEY TWO TIME. ENTRY WELD SETTING DISPLAY
48
49
STEP 2
PUT PROBE ON THE INSPECTION SPECIMEN, FIND DEFECT RESPONSE ECHO.
RAISE THE GAIN UNTIL RESPONSE ECHO AMPLITUDE BETWEEN 60-80% FULL
SCREEN , MOVE GATE A ACROSS THE DEFECT ECHO, THEN MOVE THE CURSOR
TO W DISP. PRESS ENTER KEY
AFTER PRESS ENTER KEY THE WELDING MAP WILL DISPLAY ON THE TOP OF THE SCREEN
FOR EXAMPLE:
50
INSTRUMENT: TRU-SONIC
PROBE: ABC-0460 4 MHz, 8X9 mm, 60 DEGREE
CABLE: LEMO1-LEMO 00
CALIBRATION BLOCK: V1, STEEL
TEST PROCEDURE
1. FIRST PUT PROBE ON THE V1 BLCOK AND CHECK PROBE INDEX POINT = 13 mm
2. PUT THE PROBE ON THE V1 BLOCK, TARKET TO THE V1 r=100 mm CURVE. CALIBRATE THE PROBE ZERO=2.53 Us
3. PRESS BASE SET THE INSTRUMENT RANGE=100 mm
4. SET VELOCITY = 3240 m/s
5. SET P-ZERO = 2.53 us
6. PRESS PULS, SET TYPE= ANGLE BEAM
7. SET FREQU=0.5-5 M, SET REJECT=0%
8. PRESS TRIG 1 TIME, SET ANGLE=60 DEGREE, X-VALUE=13 mm, MTL THK=25 mm
PRESS TRIG 2 TIME, SET WELD TYPE=V, W WIDTH=30 mm, W ANGLE= 30 DEGREE,
9. PUT PROBE ON THE INSPECT WELDING BLOCK. MOVE THE BLOCK AND GET THE DEFECT SIGNAL.
IF THE DEFECT COLOR DISPLAY GREEN, THIS IS FIRST HALF V SOUND PASS DEFECT. THE LARGE D WILL SHOW THE DEFECT DEPTH FROM
15-25 mm, ABOVE 25 mm SOUND PATH, THE ECHO WILL CHANGE INTO YELLOW. THE DEPTH DISPLAY
51
HALF V INSPECTION FULL V INSPECTION
THE GREEN ECHO AND YELLOW ECHO WILL LET USER EASY TO UNDERSTAND WHICH ECHO DISPLAY ECHO IS HALF V AND FULL V SOUND
PATH.
52
10. PRESS GATE, MOVE A GATE ACROSS TO THE DEFECT ECHO, THE W DISPTANCE SHOW THE ON EQUIPMENT SCREEN.ITS THE SURFACE
DISTANCE.
11. PRESS TRIG TWO TIME, MOVE THE CURSOR TO W DIST, THEN SET THE CORRECT W DIST.
12. PRESS W DISP RIGHT ARROW KEY, MOVE HIGH LIGHT CUSOR TO THE W DISP.
PRESS ENTER KEY, THE WELDING SECTION DEFECT SIMULATE MAP WILL SHOW ON THE TOP.
53
13. PRESS ENTER KEY, DISPLAY WILL SHOW SAVE MENU , STORGE THE A-SCAN DISPLAY AND WELD-ZONE DISPLAY WITH PROGRAM NO. AND
FILE NAME.
14. PRESS COPY KEY TWO TIME TO RETURN BACK THE TEST AGAIN.
15. IF MOVE THE PROBE AND USE FULL V PASS INSPECTION, THE ECHO COLOR OVER 25 mm WILL CHANGE INTO YELLOW COLOR. SEE BELOW
54
THE DEPTH SHOW 27.2 mm.
THE DEFECT POSTION CACULATED AS
DEPTH DISPLAY 27.2- MATERIAL THICKNESS 25 = 2.2 mm.
FLAW POSTION IS 25 mm-2.2 mm = 22.8 mm ABOVE THE SURFACE.
ENTER THE W DIST = 47 mm, MOVE HIGH LIGHT CUSOR TO W DISPL. PRESS ENTER
55