This document describes the configuration of the Varian ProStar
363 Fluorescence Detector driver. It also explains how to build a
control method for this driver, how to start and view a running
acquisition and what actions are available during an analysis.
Select from the list the name of the RS422 Bus to which the
detector is connected.
Hardware Window
NOTE: Once the system has been created and correctly configured, it must be
associated with one or more projects in order to start acquisitions. For more
details about the creation and configuration of a system, refer to the Galaxie
Configuration Manager Users Guide.
NOTE: For more details about the programming of each method section, turn to the
Galaxie Chromatography Data System Users Guide.
Parameters Control
Enter the Signal Offset value (from 0 to 1000 mV) that will be
added to the detector output signal to create a positive baseline
offset.
The Method Name, the Run Name, the Run ID, and the
Acquisition length must be filled before starting the analysis.
All these parameters are automatically imported from the
acquisition part if defined in the method.
NOTE: For more details about how to start an acquisition, refer to the Galaxie
Chromatography Data System Users Guide.
GLP Window
Lamp hours: displays the number of hours the detector lamp has
been on.
Lamp On/Off count: displays the number of time the lamp has
been turned on.
Lamp change date: displays the date when the lamp was
changed.