Académique Documents
Professionnel Documents
Culture Documents
Calibration Information
Introduction
This document provides the following information for the 1630 Earth Ground Clamp (hereafter
referred to as the Clamp or UUT):
Safety information
Symbols
Specifications
Maintenance
Performance Tests
Calibration Adjustments
Product Warranty Statement
For complete operating instructions, refer to the 1630 Instruction Sheet.
Contacting Fluke
For warranty service, contact Fluke as follows:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's Web site at www.fluke.com.
To register your product, go to register.fluke.com.
October 2007
2007 Fluke Corporation. All rights reserved. Specifications are subject to change without notification.
All product names are trademarks of their respective companies.
1
1630
Earth Ground Clamp
Safety Information
XW Safety Information
To avoid possible electric shock or personal injury and ensure safe
operation and service of the Clamp, follow these instructions:
Read the operating instructions before use and follow all safety
instructions.
Use the Clamp only as specified in the operating instructions;
otherwise, the Instruments safety features may be impaired.
Adhere to local and national safety codes. Individual protective
equipment must be used to prevent shock and arc blast injury where
hazardous live conductors are exposed.
Before each use, inspect the Clamp. Look for cracks or missing
portions of the Instrument housing or output cable insulation. Also,
look for loose or weakened components. Pay particular attention to the
insulation surrounding the jaws.
Never use the Clamp on a circuit with voltages higher than 600 V CAT II
or 300 V CAT III.
CAT II equipment is designed to protect against transients from
energy-consuming equipment supplied from the fixed installation,
such as TVs, PCs, portable tools, and other household appliances.
CAT III equipment is designed to protect against transients in
equipment in fixed-equipment installations, such as distribution
panels, feeders and short branch circuits, and lighting systems in
large buildings.
Use extreme caution when working around bare conductors or busbars.
Contact with the conductor could result in electric shock.
Use caution when working with voltages above 60 V dc or 30 V ac. Such
voltages pose a shock hazard.
Symbols
Table 1 describes the symbols that appear on the Clamp or in this document.
Table 1. Symbols
Symbol Explanation
Application to or removal from hazardous, live conductors is
, permitted.
W Risk of danger. Important information.
T Double insulated.
Battery or battery compartment. Low battery when shown
B on display.
P Conforms to requirements of European Union.
2
Earth Ground Clamp
Specifications
Specifications
General Specifications
Conductor Size.................................................. 33 mm (1.3 in) approximately
Dimensions (L x W x H) .................................... 276 mm x 100 mm x 47 mm (10.8 in x 3.9 in x 1.9 in)
Weight ................................................................ 750 g (1.65 lb)
Electrical Specifications
Display ............................................................... LCD
Operating Humidity........................................... Less than 85 % RH
Storage Temperature........................................ -20 C to 60 C (-4 F to 140 F)
Storage Humidity .............................................. < 75 % RH
Reference Temperature.................................... 23 C 5 C (73 F 9 F)
Temperature Coefficient................................... 0.1 X (specified accuracy) / C (< 18 C or > 28 C)
Operating Temperature .................................... 0 C to +50 C (+32 F to +122 F)
Protective Type ................................................. IP30 according to IEC 60529/EN 60529
Category Rating ................................................ 300 V CAT III, pollution degree 2 and 600 V CAT II
EMC (Emission)................................................. IEC 61000-4-1, IEC 61326-1 Class B
EMC (Immunity)................................................. IEC 61000-4-2 8 kV (air) Criteria B,
IEC 61000-4-3 V/m perf. Criteria A
Range Selection ................................................ Auto
Overload Indication .......................................... OL
Measurement Frequency.................................. 3.333 kHz
Power Requirement .......................................... 9 V alkaline (type IEC 6 LR 61, NEDA 1604A)
Power Consumption ......................................... Approx. 40 mA (in function)
Maximum Non-Destructive Current................. 100 A continuous, 200 A (< 10 sec) 50/60 Hz
Accuracy of Calibration Plate .......................... 0.1 %
Data Logging Capacity ..................................... 116 records
Data Logging Interval ....................................... 1 to 255 seconds
3
1630
Earth Ground Clamp
Maintenance
XWWarning
To avoid possible electric shock or personal injury, only
qualified personnel should perform repairs or servicing not
covered in this manual.
Performance Tests
XWWarning
To avoid possible electric shock, ensure the Clamp is
completely assembled before performing any test procedures.
Only qualified personnel should perform these tests.
The following performance tests verify the complete operation of the Clamp and check the
accuracy of each function against its specifications. Before performing any of the following tests,
check the battery and replace if necessary.
The recommended calibration interval is 12 months. In the performance tests, the Clamp is
referred to as the unit under test (UUT). If the UUT fails any performance test, contact Fluke
Service for repair. See Contacting Fluke.
4
Earth Ground Clamp
Performance Tests
Required Equipment
Table 2 lists the required equipment to complete the performance tests. If the recommended
models are unavailable, use equipment with equivalent specifications.
Precision Decade
Accuracy, 1 to 1100 : 0.475 % Yokogawa 2793 or equivalent
Resistance Box
1-loop copper wire coil 1 turn, 14-gauge copper wire, 6-in. diameter ---
%
KWH
AC BAL
AP KmVA
fcy01.eps
Figure 1. LCD Test
5
1630
Earth Ground Clamp
Accuracy Tests
Accuracy specifications are valid for one year after calibration when measured at an operating
temperature of 18 C to 28 C. Allow the UUT to stabilize at room temperature before
performing the accuracy tests.
To verify the accuracy of the UUT functions, complete the following steps:
1. Connect a 6-inch diameter loop of wire across the output terminals of the decade resistor
box.
2. Turn the UUT rotary switch from OFF to O and verify the UUT displays CAL 7, 6,
5CAL 1 and a single beep is heard.
3. Set the decade box for the Applied Value in Table 3, steps 1-4, and clamp the UUT
around the wire loop at the output terminals.
4. Compare the displayed reading with Display Reading Limits.
5. Continue accuracy tests for the other functions listed in Table 3. If any display readings
fall outside of the Display Reading Limits in Table 3, the UUT requires calibration
adjustment or repair. See Contacting Fluke.
Table 3. 1630 Accuracy Tests
1 1 0.935 to 1.065
2 9 8.765 to 9.235
3 12 11.52 to 12.48
4 90 88.15 to 91.85
Decade Resistance Box
5 O and 6-in. diameter wire 110 105.7 to 114.3
loop
6 200 193.0 to 207.0
6
Earth Ground Clamp
Performance Tests
7
1630
Earth Ground Clamp
Calibration Adjustment
To prepare the UUT for calibration, remove the back case and complete the following steps:
1. Remove the battery compartment by sliding a small, flathead screwdriver down
between the tabs and pca and lifting the compartment out.
2. Remove the bottom black shield to access the calibration potentiometers underneath
this shield. See Figure 2.
To enter calibration adjustment mode:
1. Press H, F, and D simultaneously while turning the UUT power on.
2. Continue to hold the buttons down until you hear a beep. Harmonic displays on the
LCD when the buttons are released.
The UUT is now in the calibration adjustment mode. Refer to Figure 2 for adjustment locations
and complete the adjustments as instructed in Table 4. If the UUT fails to meet any expected
results, contact Fluke Service for repair. See Contacting Fluke.
Before proceeding to Step 13, reassemble the UUT by replacing the shield, battery compartment, and back
case.
8
Earth Ground Clamp
Calibration Adjustment
Calibrator and 3
turns copper 30 A / Press Honce when
17 ~A
magnet wire, 6-in. 60 Hz reading is stable.[1]
diameter
Note
[1] Disregard reading accuracy during these calibration steps.
9
1630
Earth Ground Clamp
61
6LR
I E C Y
TER
BAT
VR1
Shield removed for demonstration
VR2 purposes only. Do not remove
during adjustment.
VR11
VR10
VR20
fcy02.eps
Figure 2. Disassembly and Adjustment Locations
10
Earth Ground Clamp
Limited Warranty and Limitation of Liability
11/99
To register your product online, visit register.fluke.com.
11
1630
Earth Ground Clamp
12