Vous êtes sur la page 1sur 5

Plant Physiol.

(1985) 79, 891-895


0032-0889/85/79/0891 /05/$01.00/0

Measurement of the Water Potential of Stored Potato Tubers'


Received for publication June 14, 1985 and in revised form August 6, 1985

WILLIAM L. BLAND -AND CHAMP B. TANNER*


Department of Soil Science, University of Wisconsin-Madison, Madison, Wisconsin 53706 (C.B.T.)

ABSTRACT found poor agreement with liquid-phase equilibrium and fair


agreement with psychrometers at potentials > -0.6 MPa, al-
A method of measuring the water potential of stored potato tubers though poor thermal equilibration between psychrometer and
(Solanum tuberosum L.) was needed to investigate the relationship of tissue degraded their measurements. The validity of the pressure
bacterial soft rot in tubers to water potential. Pressure chamber meas- chamber measurement was indicated by the agreement with
urements, while useful for tubers with functional stolons, cannot be made psychrometers and also by comparable leaf and tuber water
on stored tubers. Measurements could be made on excised tissue pieces potentials down to -0.7 MPa provided transpiration was low
in a hygrometer chamber and with hygrometers implanted into tubers. (10).
We report here our evaluation of these hygrometric methods using a The pressure chamber cannot be used to measure the water
comparison with the pressure chamber on tubers harvested with stolons potential of stored tubers because these tubers lack a conductive
intact. stolon. This article reports on two methods suitable for stored
In tubers of high water potential, measurements on excised tissue were tubers which were evaluated by comparison to the pressure
as much as 0.5 megapascals lower than the pressure chamber, probably chamber on fresh tubers. The first measures the potential of
due to turgor-driven expansion of the sample when freed from constraints excised tissue in a hygrometer chamber, as is frequently done
imposed by surrounding tissue. Good agreement (0.05 megapascals) with leaves. With the second, specially designed in situ hy-
was found between the implanted hygrometer and the pressure chamber grometers are inserted into the tuber.
at potentials higher than -0.5 megapascals. At lower water potentials,
both hygrometer measurements were higher than the pressure chamber.
Respirational heating of the tissue contributed to the increase in the MATERIALS AND METHODS
excised tissue samples, but not with the implanted hygrometers because Tuber Material. Potato plants (Solanum tuberosum L. cv
of the hygrometer design. The osmotic pressure balaced the pressure Russet Burbank) growing in the field were sheltered from water
chamber measurement of potential at -0.7 megapascals, but was too inputs to obtain a range of water potential. Also, entire plants
small to do so at lower potentials. At most, 25% of this discrepancy can were harvested and spread on tables under continuous light and
be accounted for by dilution by apoplastic water. We believe that the circulating air, with the tubers covered with black plastic. Water
pressure chamber measurement is too low at low water potentials and loss through the foliage lowered the tuber water potential. Tubers
that the error is associated with air bubbles in the xylem. At low potentials also were measured hygrometrically immediately after field har-
air emerged from xylem vessels along w-ith sap, and fewer xylem emitted vest (2-3 weeks after vine-kill) and after storage from 1 to 18
sap as potentials decreased. months in several storage environments.
Pressure Chamber. The pressure chamber and procedure de-
scribed in Gandar and Tanner (10) were used except that a lower
rate of pressurization was found necessary when 4,2 < -(.7 MPa.
Pressurization rates of 2 kPa s' or less gave the same endpoint
with repeated pressurization. Tubers were held for at least 16 h
in a plastic bag or a chamber lined with damp blotters before
An unambiguous measurement of the water potential of stored measurements were made. This delay was to allow for dissipation
potato tuber tissue is needed to investigate further an apparent of water potential gradients near the xylem system (9, 10); we
relationship between water potential and the susceptibility of verified that the delay was adequate to bring the tuber to uniform
stored tubers to attack by soft rot erwinias (15). Methodology potential by repeated pressurizations following removal of tubers
developed should also be of value in studying the water relations from the vine. To separate xylem and phloem exudation (10) for
of other fleshy plant storage tissues, particularly for studies of measurement of the osmotic pressure of xylem sap and to
mechanical damage and postharvest pathology (6). Such storage confirm xylem blockage and air entrainment in xylem sap, we
organs are an important part of the world food supply and picked away the cortex, phloem, and pith at the cut end of the
measurements of their water relations have received inadequate stolon so that the xylem vessels extended beyond other tissue.
attention. Hygrometry. Hygrometric measurements were made with a
Early measurements of potato tuber water potential were made Wescor HR-33T microvoltmeter (Wescor, Logan, UT) operated
by liquid-phase equilibrium (18). Later, soil psychrometers were in the dewpoint mode. The meter was modified with an auxiliary
inserted into tubers attached to plants in the field (5) and the rechargeable battery pack and an integrating digital voltmeter
pressure chamber was used with freshly harvested tubers having attached to the recorder outputs. The mode switch was replaced
intact stolons (1 1). Gandar and Tanner ( 11) compared pressure with a six-position, four-pole (two poles in parallel), gold-plated
chamber measurements with liquid-phase equilibrium and with contact rotary switch (Centralab, Fort Dodge, IA). A sixth mode,
soil psychrometers inserted into tubers attached to plants and 'read reverse', was added so the zero offset, zo, could be read in
' Research supported by the College of Agricultural and Life Sciences, 2 Abbreviations: n, 4{i, 4*,, tuber tissue water potential as measured by
University of Wisconsin-Madison, and by United States Department of the pressure chamber, in situ hygrometry, and excised tissue hygrometry,
Agriculture Hatch formula funds (Project 142-5140). respectively; w, osmotic pressure; zo, zero offset, dry-bulb voltage.
891
Downloaded from on September 28, 2017 - Published by www.plantphysiol.org
Copyright 1985 American Society of Plant Biologists. All rights reserved.
892 BLAND AND TANNER Plant Physiol. Vol. 79, 1985
the reverse from normal polarity. -_ _FLEX::*** FLEXANE 94
The zo arises from two sources: the bias current of the meter .25mm
amplifier (40-50 namps) through the hygrometer junction and EVANOHM LEAD
leads and from temperature differences between the hygrometer COPPER HEAT SINK
reference and measuring junctions. This measured temperature
difference reflects a temperature difference between the reference SUPPORT PIN
junction and the vapor source. Reversing the polarity when
reading zo changes the sign of the thermal component but not
that of the bias current contribution, so by reading zo in both
polarities the thermal portion can be determined. All zo values
reported here represent thermal effects only. I
Excised tissue samples were measured in a 10-sample hygro- 0

meter (model SC-10 with a Peltier thermocouple; Decagon De-


vices, Pullman, WA). We polished the inside surfaces of the
unpolished stainless steel sample holders to speed equilibrium
(present Decagon holders are polished). We also washed the
sample holders with 20% NH4OH and cotton swabs, rinsed with
distilled H20, and dried them with lint-free wipes prior to each 0 0 THERMOCOUPLE
use. Cylinders of tissue (13 mm diameter x10 mm long) were
excised at the xylem ring, parallel to the major axis of the tuber lO '--VAPOR PORT
with a sharp, thin-wall, guided sampling tube. The cores were
lightly blotted with lint-free tissue immediately after excision. -r-6
16 mm_
,-TEFLON PLUG
Rinsing the samples with distilled water immediately after exci-
sion and blotting dry had no effect on the potential measured.
Two samples were taken from each tuber, either both from the FIG. 1. Cross-section of hygrometer designed for implantation into
tuber midsection or one each from the apical and basal thirds of potato tubers.
the tuber, and their values averaged. The mean difference be-
tween duplicate samples was 0.04 MPa, believed to be random the tuber with a twist drill followed by a corkborer-like cutter
error. Calibration solutions (usually -0.45 and -0.90 MPa) equal with its cutting edge to the outside. After inserting the hygro-
in volume to the tissue cores were included with each set of meter, the juncture of the periderm and the hygrometer was
samples. The SC-10 was kept under an insulated box in the covered with petroleum jelly, the Evanohm leads were taped to
laboratory (about 22C) when in use except during collection of the tuber surface, and about 15 cm of lead was coiled against the
the 1983 fresh tuber data, when it was mounted inside of a tuber. The exposed end of the hygrometer and adjacent leads
constant temperature air bath (25C). Knobs needed to operate then were covered with 3-mm thick polyethylene foam and
the hygrometer were extended outside of the bath with phenolic wrapped in aluminum foil. Measurements were madedata in a con-
rods. Thermal equilibrium in the hygrometer body was checked stant temperature air bath (25C).The stored tuber
collected with both polyethylene and stainless steel units; the
were
by measuring zo over a calibration solution. In order to ensure
vapor and thermal equilibrium, 4e was recorded 2 h after the fresh and just-harvested tuber data were collected
with the stain-
less steel hygrometers. With fresh tubers, pressure chamber meas-
samples were excised and placed in the hygrometer. At this time
the zo of fresh tuber samples was 0.20 0.08 MV (tissue warmer urements were made both before hygrometers were inserted and
than reference junctions); between 2 and 5 h after excision zo after they were removed to test for shifts in water potential; 14
increased by about 0.16 MV. Samples measured at field harvest changed -0.04 0.05 MPa between the two determinations. In
and during storage had average zo's of 0.10 and 0.03 MV, respec- comparisons {i of and 0, the second pressure chamber value
was used because it was made immediately following {i,
the
tively. No dependence on water potential was observed in the zo
values. Determinationsof 4.,e also were made using Wescor C-52 measurement. Steady values of 4i (0.01
MPa) obtained 8 to 12
h after implanting the hygrometers. Readings at 8 h were within
hygrometers; tissue cores 8 mm in diameter and 4 mm thick
were measured. 0.03 MPa ofthe final value. Adjustment ofthe tuber temperature
Measurements of water potential were also made with hygro- to that of the constant temperature box was the limiting process
meters placed in close-fitting holes drilled axially in tubers. Two in equilibration.
hygrometers were designed for implantation into tubers, one Once the output was steady it remained so for at least 30 h.
with a polyethylene body and the other with a stainless steel To test whether the cut surface was still exchanging vapor with
body. The stainless steel unit (Fig. 1) was designed to lock the the hygrometer, a hygrometer was removed and replaced with
hygrometer and the tissue together thermally, minimizing error another after the first had been implanted 22 h. The second
caused by temperature gradients. To minimize heat flow along hygrometer was read 4 h after implanting and indicated the same
the leadwires, 12 cm of 0.25-mm (30 gauge) Evanohm3 wire with water potential (within 0.02 MPa) as the first hygrometer.
12 Am thick polyurethane varnish was used for leads, crimped The osmotic pressure of the tubers used in the comparison
into holes drilled in the heat sinks. The wires were extended by also was determined. Duplicate tissue cores (8 mm diameter, 12
crimping them to stranded copper wire. The hygrometer meas- mm length) were cut from the xylem ring and put in a 3-ml
uring junction was welded from 25 Mm chromel and constantan polyethylene syringe. To prevent the tissue from plugging the
wire and crimped into notches at the ends of the heat sinks. A syringe orifice, a disc of stainless steel screen (IO00-m
pores) had
been placed at the needle-end of the syringe. The plunger was
0.076 mm Evanohm/constantan thermocouple glued into a
small cavity in the brass core was used to measure the hygrometer replaced, the needle inserted in a rubber stopper, and the syringe
temperature. Snug-fitting holes for the hygrometers were cut in was frozen in dry ice and kept in a freezer until measured.
Measurement of r was made by thawing the tissue in the sy-
1
ringe at room temperature for h, then expressing the The exudate
3 Evanohm wire is available from Pelican Wire Co., Naples, FL. onto a disc of filterpaper in a Wescor C-52 hygrometer.
Relative to copper, it has an electromotive force of only 0.2 ,V/'C, but plicate samples were measured with different hy-
du-
lower thermal conductivity and higher tensile strength. grometers; duplicates differed, on average, by 0.0 16 MPa. In-
Downloaded from on September 28, 2017 - Published by www.plantphysiol.org
Copyright 1985 American Society of Plant Biologists. All rights reserved.
WATER POTENTIAL OF STORED POTATO TUBERS 893
creases in r due to enzymic action following thawing (3) were NV (MPb)
not observed, whether the cell sap was separated quickly after
rapid thawing or allowed to remain in contact with structural
tissue components. We tested for possible enzymic changes in r
during storage of samples in the freezer for as long as 16 weeks
(7). Decreases in r averaging 0.013 MPa week-' occurred; ac-
cordingly, samples reported here were typically stored no more
than 4 d.
Frequently, the ir of the cell sap freed when tissue samples
were excised was measured to compare with that of frozen tissue.
Sap was collected by touching a filter paper disc to a freshly cut
tissue surface, which was then placed in a Wescor C-52 hygro-
meter. To reduce evaporation, these operations were done in a
box lined with damp blotters. The osmotic pressure measured
on frozen and thawed tissue was the same as that of the sap freed
when tissue was cut. For 51 samples of both fresh and stored
tubers with water potentials from -0.58 to -1.53 MPa, r (frozen/
thawed) averaged only 0.005 MPa higher (drier) than ir (cut cell
sap); this was not significant at P < 0.40.

RESULTS AND DISCUSSION


FIG. 3. Comparisons of the water potential of freshly harvested potato
Comparisons among the pressure chamber, excised tissue, and tubers as determined by the pressure chamber (4,6) and in situ hygrometers
implanted hygrometer measurements of tuber water potential (4's). The sy, is the standard error of the estimate.
are presented in Figures 2 to 4. Osmotic pressure as a function
of A1 is presented in Figure 5. ejz (MPG)
The pressure chamber values presented in this paper have not
been corrected for the osmotic pressure of the xylem sap, as is
thought necessary for pressure chamber values to yield total
water potential (1). We found the osmotic pressure of the xylem
sap by collecting xylem exudate on filter paper discs and meas-
uring these in Wescor C-52's psychrometrically. The mean of
several tubers was 0.10 MPa, so correcting the chamber meas-
urements would shift all data points in Figures 2, 3, and 5 to the
left, worsening the agreement.
Excised Tissue Measurements. For pressure chamber meas-
urements ranging from 0 to -0.3 MPa, the excised tissue water
potentials were -0.25 to -0.35 MPa, independent of 0, (Fig. 2).
The best-fit straight line describing l, as a function of 46^ for
pressure chamber values less than -0.3 MPa (Fig. 2) had a slope
%(MPO)

FIG. 4. Comparisons of the water potential of potato tubers as deter-


mined by excised tissue hygrometry (4e,) and in situ hygrometers (4,). For
{i < -0.3 MPa, the standard error of the estimate of 4' is 0.06 MPa.
of 0.7 (not equal to 1.0 at P < 0.001).
Figure 2 also shows the straight line fit (slope = 0.42) through
21 comparisons (individual points not shown) of 4,, and excised
tissue samples determined with the C-52 hygrometer, data were
evenly distributed along the line. In samples with high turgor, C-
52 measurements were as much as 0.5 MPa below pressure
chamber measurements. Small zo's were recorded in the C-52.
The C-52's design (thermocouple wires only about 1 mm long
[4], and the measuring junction recessed in the metal hygro-
meter body) reduces the zo response to thermal gradients between
the vapor source and the reference junction. However, this does
FIG. 2. Comparisons ofthe water potential of freshly harvested potato not reduce associated errors.
tubers as determined by the pressure chamber (4) and excised tissue We believe the unduly low 4,e at high potentials (Figs. 2, 4)
hygrometry (he). Points and regression line are shown for data collected was caused by turgor-driven expansion of the samples following
with the SC-j0 hygrometer, only the regression line for the C-52 data is their excision. Excision removed constraints imposed on the
shown. The s,, is the standard error of the estimate. sample by surrounding tissue in the intact tuber, allowing the
Downloaded from on September 28, 2017 - Published by www.plantphysiol.org
Copyright 1985 American Society of Plant Biologists. All rights reserved.
894 BLAND AND TANNER Plant Physiol. Vol. 79, 1985
pression. Accordingly, errors from respiratory heating are re-
duced with implanted hygrometers as compared to measure-
ments on excised tissue. Additionally, the implanted hygrometer
is surrounded by tissue into which wound-stimulated respiratory
heat could dissipate.
The possible degradation of membranes following wounding
ofpotato tuber tissue (16) could impact #e and fj. This apparently
did not occur to a great extent during our measurements since
the cells maintained appreciable turgor at high water potentials.
This and other physiological reactions to wounding should have
0.5- had more effect on the excised tissue measurement than on the
in situ because of the large bulk of intact tissue surrounding the
implanted hygrometer. It is known that when excised potato
tuber discs are reimplanted into intact tubers, some wound
responses of the discs are attenuated (21).
In comparisons with the pressure chamber, the excised tissue
measurement (Fig. 2) was more scattered than the implanted
hygrometer measurement (Fig. 3). This was probably due to
variable effects of turgor-driven expansion, respiration, and
-9.0 -0.5
wound responses.
Implanted Hygrometer Measurements. The implanted hy-
Wc (MPo) grometer measurements agreed well with the pressure chamber
FIG. 5. Osmotic pressure (w) of freshly harvested potato tuber tissue at high potentials (Fig. 3). At low potentials, 4#i was higher than
measured by hygrometry of sap from frozen and thawed tissue as a A; the second-order term of the regression in Figure 3 is signifi-
function of the tuber water potential measured by the pressure chamber cant at P < 0.01. This departure from the pressure chamber
(14). The line indicates where 4 = measurement at low potentials also occurred with the excised
tissue measurement (Fig. 2) and when osmotic pressure was
expansion. This expansion occurred at the expense of turgor, determined (Fig. 5). Respiratory heating of the tissue explains
lowering e,. In a plot such as Figure 2 this would tend to push only a small amount of the discrepancy between Ac and 4JJ and
the data points downward from the 1:1 line where turgor exists is negligible with Oi. The negative turgor calculable from the data
(4c > -0.7 MPa; Fig. 5). The smaller C-52 samples yielded lower in Figure 5 seems unlikely in potato tuber tissue at the level of
41e values than did the SC-10 samples at all potentials where there hydration of these tubers.
was turgor, as expected for smaller samples on release of con- Dilution of the symplasm solutes by apoplastic water in the
straints. samples may have contributed to apparent negative turgor (2,
Turgor-driven strain does not affect in situ measurements 23). In potato tuber tissue, however, there does not appear to be
because of the large bulk of tissue surrounding the implanted enough apoplastic water to provide the necesary dilution. To
hygrometer. The small cylindrical cavity cut for the hygrometer cause a decrease in r of 0.1 MPa at 0, = -1.0 MPa (Fig. 5), the
would minimize strain. Also, radial geometry assists water flow apoplastic water must be 10% of the total tissue water. The
and dissipation of potential gradients near the cavity. Finally, apoplastic water present can be estimated in two ways: first by
the tuber tissue volume provides a large capacitance for water, the amount of water in the wall and the amount of wall in the
minimizing the impact of the coring on the tissue. tissue and secondly by the fraction of cell volume that is com-
During the measurements reported here, the excised tissue prised by wall. We measured the water retained by isolated potato
samples were isothermal with or warmer than the hygrometer tuber cell walls (14) at various water potentials with a soil pressure
body, so thermal errors were in the opposite direction to expan- plate apparatus. The weight fractions of water (dry wall weight
sion errors. Tissue warming, as indicated by the zo in the SC-10, base) at various potentials were: 0.90 at -1.2 MPa, 1.08 at -0.8
ordered fresh > harvest > stored as expected for respiratory MPa, 1.25 at -0.4 MPa, and 1.60 at -0.2 MPa. Since these are
heating (20). The size of respirational errors in y6 can be esti- overestimates due to water retained in menisci between the wall
mated from zo in the SC-10. The zo underestimates the temper- particles, we can use 1.00 as the upper limit for water weight
ature gradient between the reference and the tissue because the fraction and 0.40 as a lower limit (that of never-ried wood;
measuring junction is 0.43 of the distance between them and [22]). Weight fractions can be converted to volumetric fractions
because of divergent heat flow geometry. Correcting for the given the specific gravity of dry cell wall, about 1.55 (8, 22).
distance of the measuring junction from the tissue and recogniz- Measurements of the dry wall weight fraction in fiesh tissue (0.01
ing that the measured zo was removed electronically before to 0.02; [13, 14, 17]) and the water content of the tubers (0.78-
making the dewpoint measurement, a minimum error of 0.04 0.80 of fresh weight) lead to apoplastic water fractions between
MPa obtains for the fresh tubers. This value is in good agreement 0.005 and 0.025. The volume fraction of wall in tissue can be
with the displacement of the fresh tuber data from the 1:1 line estimated from wall thickness and cell size observed in cut tissue.
in Figure 4; the lower respiration rate of stored tubers did not Data of Reeve et al. (19) give wall volume fractions between 0.02
cause a similar displacement. This fresh tuber respirational error to 0.05 for stained tissue and we measured values of 0.02 to 0.04
is also present in Figure 2. As reviewed by Laties (16), respiration with unstained, fresh tissue. Data reported by Hayden et al. (12)
increased 3- to 5-fold immediately after cutting the tissue and at yield a value close to 0.05. Since a tuber is about 0.86 water by
least another 2-fold within 8 h. The long-term increase in respi- volume, apoplastic water fractions of 0.009 to 0.035 obtain.
ration was inversely related to slice thickness so the error in the Dilution by apoplastic water can account for 35% of the discrep-
C-52 measurements should have been greater than with the SC- ancy at most, and 20 to 25% is more likely.
10 samples. Contrary to pressure chamber measurements, Oi and -r meas-
In the in situ measurement, the hygrometer body makes good urements gave reasonable turgors for fresh and stored tubers:
thermal contact with the tissue, whereas the excised samples about 0.1 MPa at Ai = -0.7 MPa decreasing to 0.0 at -1.0 MPa.
make poor contact with the holder walls to avoid tissue com- Correction for the dilution error in xwould increase these turgor
Downloaded from on September 28, 2017 - Published by www.plantphysiol.org
Copyright 1985 American Society of Plant Biologists. All rights reserved.
WATER POTENTLAL OF STORED POTATO TUBERS 895
estimates slightly. This protracted decrease in turgor at low 2. BOYER JS, JR POTTER 1973 Chloroplast response to low leaf water potentials.
potential may be due to variation in water potential components I. Role of turgor. Plant Physiol 51: 989-992
3. BROWN PW, CB TANNER 1983 Alfalfa osmotic potential: A comparison of the
or bulk modulus of elasticity within the tissue, as evidenced by water-release curve and frozen-tissue methods. Agron J 75: 91-93
the development of sponginess at the basal end of tubers during 4. CAMPBELL GS 1979 Improved thermocouple psychrometer for measurement
water loss before the entire tuber is spongy. of soil water potential in a temperature gradient. J Phys E Sci Instrum 12:
From the above osmotic pressure arguments we believe the 739-743
5. CAMPBELL MD 1972 The lower limit of soil water potential for potato growth.
difference between the pressure chamber and the hygrometer PhD thesis. Washington State University, Pullman
measurements at potentials < -0.7 MPa mainly was due to error 6. COOK RJ, RI PAPENDICK 1978 Role of water potential in microbial growth
in the pressure chamber measurement. Errors due to rapid and development of plant disease, with special reference to postharvest
pressurization and to lack of water potential equilibrium within pathology. HortScience 13: 11-16
7. FENNEMA 0 1975 Activity of enzymes in partially frozen aqueous systems. In
the tuber were guarded against, as described earlier. We believe R Duckworth, ed, Water Relations of Foods, Proc. Int. Symp. in Glasgow,
that the low pressure chamber measurements at low potentials 1974. Academic Press, London
were in some way associated with air entering the xylem of 8. FREY-WYSUNG A 1976 The Plant Cell Wall. Gebruder Borntraeger, Berlin
stressed tubers. When the xylem at the end of the stolon was 9. GANDAR PW 1975 Growth and water relations in potatoes (Solanum tubero-
sum L.). PhD thesis. University of Wisconsin-Madison
separated from the surrounding tissue to permit observation, the 10. GANDAR PW, CB TANNER 1976 Potato leaf and tuber water potential meas-
number of xylem exuding sap decreased as the potential de- urements with a pressure chamber. Am Potato J 53: 1-14
creased below about 0.7 MPa. We believe, as observed by Tyree 11. GANDAR PW, CB TANNER 1975 Comparison of methods for measuring leaf
et al. (24), that as the potential decreased, the water in an and tuber water potentials in potatoes. Am Potato J 52: 387-397
12. HAYDEN RI, CA MoYsE, FW CALDER, DP CRAWFORD, DS FENsoM 1969
increased number of xylem elements fractured under tension Electrical impedance studies on potato and alfalfa tissue. J Exp Bot 20: 177-
and air diffused into the xylem. This air embolism resting against 200
perforation plates would block return of sap. In making duplicate 13. HOFF JE, MD CASTRo 1969 Chemical composition of potato cell walls. J Agric
determinations of the chamber end point, the pressure never was Food Chem 17: 1328-1331
decreased rapidly nor more than about 0.1 MPa; accordingly, 14. JARVIS MC, MA HALL, DR THREFALL, J FRIEND 1981 The polysaccharide
structure of potato cell walls: Chemical fractionation. Planta 152: 93-100
the embolism would have been created prior to the chamber 15. KELMAN A, JW BAUGHN, EA MAHER 1978 The relationship of bacterial soft
measurement. We also observed air bubbles entrained in the sap rot susceptibility to water status of potato tuber. Phytopathol News 12: 178
and the number of vessels with entrained air increased as the Abstr 236
potentials decreased. 16. LATIES GG 1978 The development and control of respiratory pathways in
slices of plant storage organs. In G Kahl, ed, Biochemistry of Wounded Plant
Good measurements of tuber water potential appear to be Tissues. W DeGruyter, Berlin
provided by the in situ hygrometers over the range of water 17. MCGUIRE R 1983 Reduced severity of Erwinia soft rot in potato tubers with
potentials encountered under field or storage conditions. Meas- increased calcium content. PhD thesis. University of Wisconsin-Madison
18. MEYER BS, AM WALLACE 1941 A comparison of two methods of determining
urement with excised tissue from stored tubers of water potential the diffusion pressure deficit of potato tuber tissuei Am J Bot 28: 838-843
< -0.4 MPa are acceptable provided the SC-10 hygrometer, with 19. REEVE RM, H TIMM, ML WEAVER 1973 Cell wall thickness during growth of
its large sample holders, is used. Pressure chamber measurements domestic and foreign potato cultivars. Am Potato J 50: 204-211
are satisfactory with fresh tubers at potentials > -0.7 MPa, but 20. SCHIPPERS PA 1977 The rate of respiration of potato tubers during storage. 1.
error obtains at lower potentials. Dilution by apoplastic water Review of the literature. Potato Res 20: 173-188
21. SMrrH BG, PH RUBERY 1981 The effects ofinfection by Phytophihora infestans
affects measurements of wr little, since only about 0.025 of the on the control of phenylpropanoid metabolism in wounded potato tissue.
water in potato tubers is apoplastic. Sap collected from cut Planta 151: 535-540
surfaces yields the same value of X as obtained by killing the 22. STAMM AJ 1971 Review of nine methods for determining the fiber saturation
tissue by freezing. points of wood and wood products. Wood Sci 4: 114-128
23. TYREE MT 1976 Negative turgor pressure in plants: Fact or fallacy? Can J Bot
LITERATURE CITED 54: 2738-2746
24. TYREE MT, MA DIXON, EL TYREE, R JOHNSON 1984 Ultrasonic acoustic
1. BOYER JS 1969 Water status measurements in plants. Annu Rev Plant Physiol emissions from the sapwood of cedar and hemlock. Plant Physiol 75: 988-
20: 351-364 992

Downloaded from on September 28, 2017 - Published by www.plantphysiol.org


Copyright 1985 American Society of Plant Biologists. All rights reserved.

Vous aimerez peut-être aussi