Académique Documents
Professionnel Documents
Culture Documents
CENTERVIEW
8000 / 8000e
Version 3.3 - September 16, 2015
TABLE OF CONTENTS
1. CONTACT ................................................................................. 5
5. INSTALLATION ..........................................................................29
5.1 DEVICE ASSEMBLY ....................................................................................... 31
5.1.1 TRANSPORT LOCK ......................................................................................................... 32
5.1.2 ALIGN THE MEASURING HEAD ....................................................................................... 33
5.1.3 FIXING THE DEVICE STAND ............................................................................................ 34
5.1.4 SPECIALTY FOR CENTERVIEW 8000 C............................................................................. 35
5.2 WIRING ...................................................................................................... 37
5.2.1 GROUNDING ................................................................................................................ 38
5.2.1.A GROUNDING SCREW AT THE DEVICE STAND .............................................................. 41
5.2.1.B GROUNDING AT CENTERVIEW 8000 C ....................................................................... 41
5.2.2 VOLTAGE SUPPLY .......................................................................................................... 42
5.2.3 SIGNAL LINES ................................................................................................................ 44
5.3 HIGH FREQUENCY GENERATOR (QGE) ......................................................... 46
5.3.1 QGE - DIMENSIONS ...................................................................................................... 47
5.3.2 QGE - CONNECTIONS ................................................................................................. 48
5.3.2.A QGE - CONNECTIONS (OLD VERSION) ...................................................................... 49
6. INITIATION ...............................................................................56
6.1 ZERO ADJUSTMENT ECCENTRICITY ............................................................... 56
6.1.1 REQUIREMENTS ............................................................................................................. 57
6.1.2 INSTALL THE CALIBRATION CONSTRUCTION ................................................................. 58
6.1.3 PROCEEDING ................................................................................................................ 60
6.2 MEASURING OPERATION ............................................................................. 65
6.2.1 SMALL DIAMETER AT CENTERVIEW 8010 / 8010E ........................................................... 66
6.3 DEVICE CONNECTIONS (SIOB-XX) ................................................................ 67
6.3.1 ANALOG OUTPUT / DIGITAL INPUT / RELAY CONTACT .................................................. 68
6.4 CONNECTION CABLES TO SIKORA DISPLAY DEVICES ..................................... 69
6.4.1 CONNECTION TO ECOCONTROL 6000 / 1000 ........................................................... 69
6.4.2 CONNECTION TO ECOCONTROL 2000 ....................................................................... 69
6.4.3 CONNECTION TO ECOCONTROL 600 ......................................................................... 70
6.4.4 CONNECTION TO REMOTE 2000 / DISPLAY 2000 ......................................................... 70
6.5 CONFIGURATION VIA DIAGNOSTIC SOFTWARE ............................................ 71
6.5.1 RS232 CONNECTION CABLE FOR SETUP/DIAGNOSTIC ................................................. 72
6.6 SC400 (ASCII) TELEGRAM ............................................................................. 73
6.6.1 TELEGRAM STRUCTURE .................................................................................................. 73
6.6.2 VALUE SETTINGS ............................................................................................................ 74
6.6.3 CHECKSUM BYTE BCC ................................................................................................... 74
6.6.4 TIME OUT ...................................................................................................................... 74
6.6.5 RECEIVING TELEGRAM SLAVE (MEASURING DEVICE) ....................................................... 75
6.6.6 SENDING TELEGRAM SLAVE (MEASURING DEVICE) ......................................................... 76
6.6.7 REMOTE INDICATION (MEASURING DEVICE) - EXTERNAL PC .......................................... 80
6.7 'OPTION | PROFIBUS DP .............................................................................. 81
6.7.1 INTERFACE .................................................................................................................... 81
6.7.2 TELEGRAM STRUCTURE .................................................................................................. 82
6.7.3 DATA MODULES ............................................................................................................ 82
6.7.3.A MODULES 1 TO 3 - IN DATA (FROM DEVICE) ............................................................... 83
6.7.3.B MODULES 1 TO 3 - OUT DATA (TO DEVICE) ................................................................ 85
6.7.3.C MODULE 4 - IN DATA (FROM DEVICE) ......................................................................... 86
6.7.3.D MODULE 4 - OUT DATA (TO DEVICE) .......................................................................... 88
7. OPERATION .............................................................................91
7.1 SWITCH DEVICE ON .................................................................................... 91
7.2 SWITCH DEVICE OFF ................................................................................... 91
7.3 MOTOR STOP FUNCTION ............................................................................ 92
1. CONTACT
SIKORA AG
Bruchweide 2
28307 Bremen
-Germany-
Phone : +49 (0) 421 / 48 900 - 0
Email : sales@sikora.net
Internet : www.sikora.net
An actual overview for all sales and service offices can be found on our website under Sales &
Service. Contact your country specific office for service advices. Most countries offer locally SIKORA
service engineers. If you could not found an own sales and service office for your country or have no
direct contact person, refer to our service central office in Germany.
Email : service@sikora.net
In case of error or problems using your SIKORA device, we recommend reading the corresponding
chapters of this instruction manual first. If this provides insufficient, contact your responsible service
for assistance.
Make sure you have the corresponding device number when talking to your SIKORA contact
person. This will assure a quick processing of your request!
Neither the document nor parts of it may be reproduced in any form or processed, duplicated or
broadcasted by using electronic systems without the prior permission of the SIKORA AG.
All corresponding documents / appendices have been produced with the appropriate care. The
instruction manual content based on the ISO/IEC 82079-1 also as the directive 2006/42/EC of the
European Parliament and of the Council of 17 May 2006 on machinery and amending directive
95/16/EC (recast) annex I chapter 1.7.4.1 general principles for the drafting of instructions. The
use of this guideline reference to content requirements of an instruction manual for all SIKORA
devices only and is not automatically binding to fulfill this directive. To comprehend the fulfilled
directives during production of the accordant SIKORA device, consult the corresponding declaration
of conformity.
The SIKORA AG meets the demands of instruction responsibility about installation, operation,
maintenance, possible safety risks and foreseeable misuse with the appropriate care. Possible wrong
information respect to the documentation will not be impossible and exist without own knowledge.
If you find such wrong information contact your respective contact person to correct / actualize this
state of knowledge.
There is no responsibility at all for direct, indirect, random or other damages as well as subsequent
damages, resulting from incorrect or abusive use of this document and the associated products. The
same applies for a correct or incorrect use of the product with the assistance of this document by
persons that are not destined resp. trained for this product.
The SIKORA AG is not liable for damages or problems resulting from the use of options which are no
original SIKORA products.
The listed information here makes the use with the corresponding device or the corresponding event easier.
For further information, read the chapter / documentation named at this reference.
Device description General overview of the measuring device, existing measuring principles,
technical data and available interfaces. At the end of the chapter is a
listing of the included and optional scope of delivery.
Installation This chapter describes all general instructions to mount the measuring
device. It consists of the device assembling and wiring also as optional
components if available, to install them to/into the measuring device.
Initiation This chapter describes all functions and instructions for the configuration
and initiation of the measuring device. It includes besides general
information of the measuring device the different customer interfaces and
possible data telegrams for communication with the measuring device.
Operation This chapter describes all instructions and proceedings for the general
operation of the measuring device.
Maintenance An overview of all instructions and operations for maintenance, care and
servicing the measuring device.
3. SAFETY ADVICES
This is essential for the opening at the bottom as well as for the
measuring openings.
Obey the shown information absolutely.
Non-compliance can lead to device damages or to slight, moderate injuries!
The compliance has to be provided by the customer. The SIKORA AG is not liable for non-compliance / disregard.
DECLARATION OF CONFORMITY
We confirm that the product
This declaration is valid for all products which are produced in accordance with the technical
documentation which is part of this declaration.
For verification of conformity with regard to safety requirements and electromagnetic compatibility
the following harmonized standards are applied:
EN 61010-1 Safety requirements for electrical equipment for measurement,
control and laboratory use.
Part 1: General requirements.
The following declaration of conformity is a content report of the original document only, which is part of the product delivery / available at the SIKORA AG
(depending on order). It has no serial number, date or signing.
Before using the measuring device / system, read the instruction manual.
A basic instruction for the personnel will be realized during an optionally initial operation (depending
on customer order), according to the locally necessity. A specific or safety training of operating
personal is not necessary.
The use of the device or its components as hand tool or striking tool.
The use of the device as throwing object.
The use of the device frame as seating or footprint.
Further foreseeable misuses for the measuring device and its components are not known on side of
the SIKORA AG.
If possible foreseeable misuses appear, the SIKORA AG has to be informed immediately so they can
be documented to guarantee the intended use of the measuring device ant its components. The
SIKORA AG will assume no liability for not known foreseeable misuses.
In the case of foreseeable misuse which is not noted in this documentation so far or not sufficient
description of operating processes / safety advices, contact the respective contact person (sales or
service), so we can add your information to future documentation updates.
Unknown and therefore not documented residual risks are not excluded.
The SIKORA AG endeavors to complete the explanations for residual risk and precaution. In the case
of a missing documentation, contact the respective contact person.
4. DEVICE DESCRIPTION
The CENTERVIEW 8000 / 8000e measuring head is a precision device for measuring the eccentricity,
diameter and ovality during production. The device is able to determine the location of the round,
stranded or solid conductor in the insulation material at any line speed. Thus the operator can
readjust the die head during production in order to reach maximum concentricity.
The measuring system is equipped with two sensor-guided motors which are used to adjust the
measuring head horizontally and vertically so the conductor is always located in the middle of the
measuring head opening.
The CENTERVIEW 8000 / 8000e has the ability to measure and display the eccentricity of isolation
wall thickness at 8 points. Also it could collect single measuring values with high precision to show
them into point form inside a coordinate system at the display unit. This will lead to an allocation of
commonness, showing the operator the deviation range of the eccentricity values. In case of rotating
eccentricity values, the operator could initiate adapted procedures with this information.
The CENTERVIEW 8000 / 8000e has three serial interfaces for transferring measuring and operational
data to and from the device. The interfaces can be used to connect the CENTERVIEW 8000 / 8000e to
a SIKORA display unit or an external line computer, as Profibus interface (Option) and as a Setup and
Diagnostic Interface.
Also, there are four analogue outputs (optional board configurable via setup software) for
transmitting the measuring data, one digital input which is used for the MOTOR STOP function and a
fault contact.
Data are measured without contacting the surface of the measuring object by applying an
inductive-optical measuring principle. The inductive part is used to locate the conductor position
inside the measuring head by observing the current flow in the conductor. The optical system is used
to measure the position and the diameter of the insulated cable.
Optical measurement is done by sending out light from a laser diode which shines on the measuring
object thus casting a shadow on the receiver, a CCD line. The image information is then evaluated by
integrated signal processors. Measuring data acquisition remains precise even with vibrating
measuring objects due to the extremely short exposure time.
Checking the measuring values of just one particular cable spot in the laboratory will not be sufficient
at all:
The CENTERVIEW 8000 / 8000e displays a value of eccentricity which is averaged from the values
measured over a certain length of cable (compare the corresponding settings and their
explanation in the configuration section Customer Setup of the Diagnostic Program) and not a
value at a certain selected point. As a result, this displayed value cannot be equivalent to the one
measured at one single spot. Additionally, the eccentricity of the cable produced does not
necessarily occur homogeneously and may vary over the length of the cable. Optical
measurement of just one single sample cannot consider this behaviour.
Merely cutting the measuring object may already lead to its deformation, thus leading to
differing values. Any cutting-off process, however careful it is done, means layer deformation,
especially when forces applied for separation affect the cut and if a temperature rise at cutting
can cause strain. Cutting also causes a compensation of the forces formed by material shrinkage,
especially in longitudinal direction of the insulation, with corresponding effect on the thickness
of each single layer.
Placing one single sample differently in the microscope may already lead to different results.
This means:
Values from a single cut do not compare with values displayed by our measuring devices!
There should be identical temperature of the tested cable at each measuring process in order not
to neglect the temperature coefficient of the insulation.
If the conductor is removed too, the cable sample will be partially relieved concerning the radial
forces. This causes considerable changes in all measuring values including outer diamter and
conductor diameter.
Thus, to obtain at least approximately comparable values from samples, several precautions must be
taken.
More than just one cut is required; preferably 10 cuts should be prepared. To do so, take from the
cable produced a section whose length is proportional to the averaging time configured (see
Diagnostic Program) for the CENTERVIEW 8000 / 8000e.
Mark the beginning and the end as well as the cable position and the direction of production
accordingly! Note down the values displayed at the position of mark 2, these represent the average
values for the section concerned.
Taking the sample must be done carefully, using appropriate tools and avoiding deformation of the
cable.
Divide the sample section into 10 segments of equal length. In order to measure each sample
identically concerning position and place, each segment must be marked so its position during
production can be seen, like by
When measuring the samples, their alignment in the profile gauge must be identical for each sample
and the measurement must be taken at identical position and side.
The following example is to illustrate how to proceed, showing the measurement of four values per
sample.
Compare this calculated value with the one indicated by the CENTERVIEW 8000 / 8000e.
* Optional product diameter: CENTERVIEW 8010 / 8010e MICRO for micro-coax cables
1
Automatically adjustment (Factory setting: 1000 mm | 39.4 inch)
Customer device
Analog output (Option)
like line control, SPS, line
recorder
Overview
external connections Digital input
Customer device
Customer device
Relay contact
like external signalling
device, line control
HINT Check the parts after unpacking carefully for damage and completeness
of the components.
1 Measuring head from device set CENTERVIEW 8000 / 8000e with stand
1 High frequency generator
1 Toroidal transformer
1 Silver wire (for zero adjustment eccentricity)
1 Cleanser
1 Device CD with instruction manual and relevant documentations also as DIAGNOSTIC software
Product 90-5601 Universal field bus module (Profibus, Profinet IO, CANopen, DeviceNet,
EtherNet/IP)
Product 90-0164 Stand for high frequency generator and toroidal transformer
5. INSTALLATION
The measuring device can be installed directly after the extruder crosshead or after the cooling
section. See the installation examples in the Appendix at Subchapter System Survey. Plan the exact
place of installation together with our technicians.
To achieve best results concerning accuracy of extruder die adjustment and saving of insulation
material, it is best to place the measuring head at a point in the line, where the cable has just passed
the extruder crosshead and its core is coated by the insulation material.
When installing the measuring head at a place where the cable has left the cooling trough, care has
to be taken to keep the test object dry, as drops of water may cause measuring errors resp. error
messages.
Keep in mind that the high frequency generator must be installed near the toroidal transformer at a
certain minimum distance.. Also, there are some minimum distances to be kept when installing the
measuring head. These are shown in the figures at chapter System survey (page 119).
NOTICE The HF voltage induced by the toroidal transformer is used for the zero
eccentricity alignment using a bare wire and must not be reduced or
applied to ground!
Regard this advice for intended use of the SIKORA device and its components. Disregard can lead to incorrect use
up to physical device damage.
The Protected Area shown in these illustrations marks a section of the line, where there must not be:
Any contacting measuring device.
Non-insulated line equipment (e.g. guiding rolls).
Using a conductor preheater may cause a substantial current flowing through the cable conductor.
Thus, the resulting voltage may result in varying diameter measuring values. In this case the
conductor must be additionally grounded after the pre-heater, before the extruder, by mounting a
grounded metal guiding roll. According to the type of preheater used, this may not be sufficient. In
this case, contact our service office to consider installing a ferrite tunnel between preheater and
extruder, thus making the line more high-resistive. Grounding by means of a guiding roll or brush
after this tunnel (towards the extruder) is mandatory as otherwise grounding will be too
high-resistive.
At the installation site, temperature should be kept in a range from +5 ... +50 C | +41 ... +122F, air
humidity for the CENTERVIEW 8000 / 8000e is max. 95% (without condensation).
For mounting the stand to the floor there are four slotted holes at the base. If the floor is uneven, we
recommend you first fix a level steel plate on the floor onto which the stand can be mounted.
Installation of a wiper is recommended in those cases, where lumps exceeding the CENTERVIEW
8000 / 8000e measuring opening in size may occur.
On installing the measuring head you have to decide how to place it in regard to the direction of
view. For the device itself a certain direction of production is not mandatory. In most cases, this
depends on the situation of the connectors locality. You have to decide and configure from which
side (red or green) of the measuring head you want to look at the cable when watching the
eccentricity display.
Background: The eccentricity display illustrates a virtual cable cross-section inside the masuring head.
Depending on the direction of view, the eccentricity has to be configured. This also refers to the
angle of eccentricity being counted in clockwise direction always.
The two sides with the measuring openings are marked with colored dots (red and green) for
identification.
You can find a transport screw for protection at the side of the measuring device. It is connected to a
cable, which is fixed to the voltage connector. This cable is just an optical aid to really remove the
transport screw before first initiation (voltage supply) of the motors. The cable has no electrical
meaning.
Transport screw at the device side Cable with sticker as optical aid
The CENTERVIEW 8000 / 8000e is placed correctly, when the test object passes concentrically
through the whole length of the measuring head opening. Check this by reading out the [ P AS S
A N G L E ] at DIAGNOSTIC software. The horizontal and vertical value each must not exceed a value of
0.3. If necessary adjust the measuring head and/or the guiding rolls if any installed.
OK
Measuring head seen from side... ...and seen from the top
NOT
OK
Test object not Sagging test object, not centered on Test object not running straight, not
centered whole length centered on whole length
The line height has been pre-adjusted at the SIKORA AG factory; the motors are in middle position.
Within the range specified, the CENTERVIEW 8000 / 8000e automatically adapts itself to the line
height.
Regard this advice for intended use of the SIKORA device and its components. Disregard can lead to incorrect use
up to physical device damage.
Movements of the test object during production do not influence measurement accuracy.
If it is not, the measuring device position has to be set first, before you start
to arrange the cable and start the device assembly.
The configuration can be realized via the DIAGNOSTIC software at the [ D I AG N O S I S ] >> [ S E R I AL
I N P U T /O U T P U T ] part. A message appears, that the function can be used offline only.
is clicked again.
With the position marker in the middle, you can continue with the actually device assembly.
Completely de-energize the measuring device. Lead the cable through the measuring opening and
visually control its concentricity. Correspondingly adjust the measuring device until the cable visually
lays horizontal and vertical in the middle.
Now reconnect the voltage supply for the measuring device. The device automatically starts to
adjust. By reading the angle at the DIAGNOSTIC software at [ P AS S A N G L E ] , you can check the
adjustment.
5.2 WIRING
Example TN network
Execute the wiring in the corresponding order of the following sub chapters. It guarantees an
appropriate wiring procedure.
5.2.1 GROUNDING
Regard to connect not only the measuring device grounding (PE) correctly, but also provide for the
right voltage potential among conductor L and N.
Especially when using a transformer / converter you have to mind to ground the Zero conductor
respectively the midpoint (e.g. rotating current) additional to PE to avoid sparkover voltage.
NOTICE Sparkover voltage can release incorrect data transfer or damage the
internal serial interfaces! Regard for a competent grounding and cable
laying.
Regard this advice for intended use of the SIKORA device and its components. Disregard can lead to incorrect use
up to physical device damage.
In general, the SIKORA AG basically recommend the installation of a solid ground connection close to
the line and leading straight from the pay-off to the take-up in order to provide fundamental
grounding. Therefore, use copper wire with at least 10 mm.
Take great care to ground your installation properly and sufficiently. This is of vital importance for
the inductive part of the measuring system. If the grounding is insufficient, the inductive part will not
be able to provide a correct measuring result. In order to locate the conductor position inside the
measuring head, the inductive part observes the current flow in the conductor. This current flow
because of the voltage induced by the toroidal transformer, and of course only if the electric circuit
being represented by the conductor, the pay-off and the cooling trough or the take-up is closed.
Depending on the installation of your CENTERVIEW 8000 / 8000e, the measuring current flows as
shown in the drawings below:
1)
Always install the toroidal transformer close to the CENTERVIEW 8000 / 8000e, in this example at
the beginning of the cooling trough. Regard minimum distance as shown in System Survey
(Appendix).
1)
Always install the toroidal transformer close to the CENTERVIEW 8000 / 8000e, in this example at
the beginning of the cooling trough. Regard minimum distance as shown in System Survey
(Appendix).
In this case, the SIKORA AG strongly recommend the installation of a ground connection with at least
10mm leading straight from the pay-off to the take-up in order to provide fundamental grounding.
Grounding at the take-up may be problematic, as the conductor is already completely insulated here
thus making ground connection unreliable and insufficient. Avoid this situation by grounding all
guiding rolls at the take-up.
Regard this advice for intended use of the SIKORA device and its components. Disregard can lead to incorrect use
up to physical device damage.
CENTERVIEW 8000
Pin 1
L = 200 V AC -5% to 230 V AC +10%
alternative (automatic switch)
100 V AC -5% to 115 V AC +10%
50/60 Hz, 200 VA
CENTERVIEW 8000 C
Pin 1
L =200 V AC -5% to 230 V AC +10%
alternative (automatic switch)
100 V AC -5% to 115 V AC +10%
50/60 Hz, 200 VA
Pin 2 Pin 3
N 0 V AC not connected
QGE
Pin 1
L = 100 to 240 V AC 10%, 50/60 Hz, 140 VA
Pin 2
allready factory-connected
Pin 3
N 0 V AC
In the case of the optional voltage supply the CENTERVIEW 8000 / 8000e will internal switch
automatically to accordant voltage.
When designing the device, special attention has been laid on preventing power-line interferences.
Additionally, when installing the device the SIKORA AG recommend that you use only those power
lines which correspond to the EMC (electromagnetic compatibility).
If in doubt, you better provide a separate power line from your main distributes directly to the device
and do not connect any other loads to it.
Regard this advice for intended use of the SIKORA device and its components. Disregard can lead to incorrect use
up to physical device damage.
Install the wiring as required for your application. Note that the measuring head connection cables
must be sufficiently long for the measuring head to be moved vertically by its motors. The range of
this movement may be up to 200 mm | 8 inch!
Refer also to chapter Device connections (SIOB-xx) (page 67) for pin assignment and where to find
the respective connectors. Note that these are located on the main PCB mounted in the device
housing stand, so you have to open the side panel in order to do the connecting.
NOTICE Before you open the housing, always cut off the device
voltage supply!
Regard this advice for intended use of the SIKORA device and its components. Disregard can lead to incorrect use
up to physical device damage.
Pass your connection cables through the cable clamps. Do not forget to close the panel again.
Regard this advice for intended use of the SIKORA device and its components. Disregard can lead to incorrect use
up to physical device damage.
NOTICE Before connecting the mains, make sure the ground connection exists!
Regard this advice for intended use of the SIKORA device and its components. Disregard can lead to incorrect use
up to physical device damage.
HINT This connection is necessary to use the device. Without this connection
the generator makes a very low output (for safety) only.
The control cable should be preferably be equipped with a tight screen wire mesh (and additional
with a screen sheet at ideal case).
HINT The connection cable to the high frequency generator (ST1B) is not part
of delivery and has to be provide by the customer.
5.3.4 LED
Power-ON-LED: f this LED shines, then the switching power supply is working
The power supply should then not be touched, because some
components conduct voltages.
Output-voltage-LED: The bigger the output amplitude, the brighter the LEDs shines.
At maximum output voltage the LED shines the brightest.
Current-limiter-LED: This LED shines if the persistent current is limited by the generator.
Short-circuit-LED: If this LED shines, then the final stage had shortly been deactivated
because of a short-term current peak resp. short circuit.
Input-current-LED: If this LED shines, the internal current sensor at the toroidal
transformer receive a sufficing signal.
Under these hindered conditions, we recommend to rinse the measuring head with cleaned
compressed-air. To refit the measuring gauge, an appropriate construction set is available optionally.
Use from the SIKORA AG offered / installed compressed-air rinsing only, it is explicit manufactured
for the measuring head. Other rinsing systems can interfere with the measuring or clean the device
insufficient.
Using unfiltered compressed-air could lead to a stronger fouling of the optical device components. To
prevent from oil residues inside compressed-air, you should install an external oil extractor. Also
humid compressed-air could lead to corrosion at the electrical contacts.
1 Filter 3 m
1 Pressure reducer
1 Elbow with adapter
1 Connection tube 8 mm (0.315 inch)
1 Connection tube 4 mm (0.157 inch)
1 Connection part for measuring head
Regard this advice for intended use of the SIKORA device and its components. Disregard can lead to incorrect use
up to physical device damage.
6. INITIATION
After installing and connecting your CENTERVIEW 8000 / 8000e completely, the device has to be
adjusted. Usually this is done by an experienced SIKORE technician. In general, you will not have to
adjust your CENTERVIEW 8000 / 8000e again.
However, repeating the device adjustment may become necessary in certain cases:
Adjustment of the CENTERVIEW 8000 / 8000e means zero-point adjustment, i.e. defining the point of
zero eccentricity. Therefore, use the accordant DIAGNOSTIC software. Also regard the information in
chapter System survey (page 119).
6.1.1 REQUIREMENTS
To perform your adjustment measurement, you must observe certain conditions. Most important of
all, you need a reliable calibration normal which is used to simulate a Zero Eccentricity-object. This
normal must be of very high quality: a round conductor, completely straight, not bent or damaged in
any way, it must not have any scratches or kinks. These requirements can be met by making a
calibration normal from a specific piece of silver wire (part of delivery) which is stretched by a certain
amount directly in your line thus eliminating any bends or uneven parts.
Observe that there must not be any contacting devices in the section being passed by the calibration
normal. If there are any guiding rolls, these must be insulated. This is most important, as the HF
voltage induced is used for aligning the point of zero eccentricity and must not be deducted in any
way.
NOTICE Absolutely never just strip the cable being in the line!
Regard this advice for intended use of the SIKORA device and its components. Disregard can lead to incorrect use
up to physical device damage.
Adjust the guide rolls height with the line. This can be checked via the
DIAGNOSTIC software with the function [P AS S AN G LE] .
The rolls inclusive holder are height adjustable with the two attachment screws.
Maybe you have to shorten the lower rubber seal of the stand, if the application has
to be set lower.
If the guide rolls are also used during running production for the attenuation of vibrating product,
the adjustment is done. Otherwise lower the application for the production and adjust it again for
the next calibration.
6.1.3 PROCEEDING
The SIKORA AG recommend you proceed as described below:
>> [O K] .
Watch the motor status message box at the bottom of the page to change to Motor OFF.
Put the wire to be used as calibration normal into the line. The wire must not drag or sag, it must
be straight as during production and in exactly the same position. The bare conductor must be
free; it must not touch any metal parts like metal guiding rolls for instance.
Install the calibration normal / silver wire as follow (marked as blue line at the picture):
A. Fix the wire to the grounding socket at the stand without the crank handle (wind it
several times for a better hold).
B. Pull the wire above the guide roll and through the attachment.
D. Pull the wire through the attachment and then above the guide roll on the other side.
Connect the grounding cable (yellow) and the signal cable (red). Regard the side of the toroidal
transformer.
The yellow grounding cable is always connected at the opposide side of the toroidal.
The red signal cable is connected on the toroidal side and leads to an intended ground
connection at the toroidal (yellow connector).
Cable connections with toroidal right Cable connections with toroidal left
Start the motors again and wait till they automatically center the measuring head.
In the diagnostics program selection bar, click
[ S E T U P ] >> [ E X T E R N AL S E T T I N G S ] >> [S E ND M O TO R S T AT U S ] >> [M O TO R R UN ] >> [OK]
>> [O K] .
Watch the motor status message box to change to Motor ON.
Check the wire passage angle through the measuring head after calibration normal insertion:
The [ P AS S A N G L E [ ] ] will be shown separated in horizontal and vertical direction inside the
DIAGNOSTIC software. It should be a low value with a maximum of 0.3. Therefore, adjust the
cable guides if necessary.
Watch the DIAGNOSTIC software, eccentricity display. Consider the numerical values indicated
there (numerical values ECCENTRICITY and ANGLE) whether these still meet your requirements
YES,
AD J U S T M E N S T E P .
NOTICE Do not touch the light source and sensors with your fingers (fouling the
measuring openings!) or with metal objects (scratches!).
Regard this advice for intended use of the SIKORA device and its components. Disregard can lead to incorrect use
up to physical device damage.
To protect the measuring device keep the covering always closed. Open it for maintenance only.
Note that the cable has to pass through the line far enough before any eccentricity measuring values
are available. This is because the inductive part of the measuring system will not be able to provide a
correct measuring result before the electric circuit being represented by the conductor, the pay-off
and the cooling trough or the take-up is closed. Keep this in mind after line start-up and when cutting
the cable.
HINT Via the standard DIAGNOSTIC interface you can only display the
CENTERVIEW video but not configure this function.
For diameter smaller 0.7 mm | 0.03 inch the setting [ S M AL L D I AM E T E R ] has to be set to [Y E S ] at
the DIAGNOSTIC software.
Start the DIAGNOSTIC software and set the following under [ S E T U P ] >> [ C U S T O M E R S E T U P ] at tab
[ G E N E R AL ] :
For diameter above 0.7 mm | 0.03 inch this option should not be activated.
At the CENTERVIEW 8000 C the connections ST11, ST12 and ST13 are not at the electronic but as
separate connections at the rear side of the measuring device.
REFERENCE For a detailed description to use the DIAGNOSTIC software, read the
DIAGNOSTIC software user manual for the corresponding measuring
device.
Regard to use always the actual version of the software DIAGNOSTIC. It can be found on the Device
CD, delivered with the measuring device or obtain by the SIKORA AG service.
Using an older version, a correct function of the software DIAGNOSTIC with the measuring device can
not be guaranteed. The software is downward compatible; older measuring device can be used with
the actual version also!
The DIAGNOSTIC software is a stand-alone application without need for installation or further
configuration. Copy the software to the desired computer which you want to connect to the
measuring device.
HINT The RS232 interface is for connection and controlling the measuring
device via the SIKORA DIAGNOSTIC software.
To connect (1:1 standard connection) you can use any usable D-Sub
connection system available on market or at the SIKORA AG.
The appropriate device acts as MASTER and sends request telegrams to the SIKORA measuring
device. The SIKORA measuring device only sends a telegram to the MASTER as reaction answer for
the request.
Theoretically the data transfer rate is limited by the Baud rate only.
Checksum =
IDB1 xor IDB2 ... xor DB1 ...xor DBn
xor ETX with setting sent by MASTER
resp.
IDB1 xor IDB2 xor IDB3 xor ETX with request sent by MASTER
IDB1 xor IDB2 ... xor NAK xor ETX with reaction or setting or if ID is
missing (sent by SLAVE)
(yes) (no)
B. Maximum permissible delay time between receiving the last character of a request telegram and
sending the first character of the according reply: 500 ms
If this criterion is fulfilled, the sender reacts by sending a request telegram again.
Setting:
ID DB Content:
-ASCII- -ASCII- Set Value
000 DB1...DB5 Nominal value diameter
001 DB1...DB4 +Tolerance diameter
002 DB1...DB4 -Tolerance diameter
019 DB1...DB4 +Tolerance diameter (short time)
020 DB1...DB4 -Tolerance diameter (short time)
035 DB1 MOTOR STOP
036 DB1..DB6 Tolerance standard deviation eccentricity (ECC)
Request:
ID Meaning:
-ASCII- Identification for requested parameter
001 Status and actual value diameter
003 Status, sign and deviation diameter
004 Nominal value diameter
005 +Tolerance diameter
006 -Tolerance diameter
008 Status and actual value diameter axis - Blue
012 Status and actual value diameter axis - Magenta
016 Status and actual value ovality
023 +Tolerance diameter (short time)
024 -Tolerance diameter (short time)
033 Status and minimum diameter
034 Status and maximum diameter
035 MOTOR STOP
036 Tolerance standard deviation eccentricity (ECC)
ID Meaning:
-ASCII- Identification for requested parameter
209 Status and absolute value eccentricity
210 Status and angle eccentricity
211 Status and standard deviation ECC axis - Blue
212 Status and standard deviation ECC axis - Magenta
213 Status and standard deviation ECC absolute value
300 Actual values
301 Default values
320 Actual values
321 Default values
330 Actual values
331 Default values
Reaction to default:
Default ok
Reaction to request:
ID DB Content:
Request -ASCII- Actual value
-ASCII-
001 DB1 Status diameter
DB2...DB6 Diameter
003 DB1 Status diameter
DB2 Sign diameter
DB3...DB7 Diameter
004 DB1...DB5 Nominal value diameter
005 DB1...DB4 +Tolerance diameter
006 DB1...DB4 -Tolerance diameter
008 DB1 Status diameter axis - Blue
DB2...DB6 Diameter axis - Blue
012 DB1 Status diameter axis - Magenta
DB2...DB6 Diameter axis - Magenta
016 DB1 Status ovality
DB2...DB6 Ovality
023 DB1...DB4 +Tolerance diameter (short time)
024 DB1...DB4 -Tolerance diameter (short time)
033 DB1 Status minimum diameter
DB2DB6 Minimum diameter
034 DB1 Status maximum diameter
DB2DB6 Maximum diameter
035 DB1 MOTOR STOP
0 : run
1 : stop
036 DB1...DB6 Tolerance standard deviation eccentricity
209 DB1 Status
DB2...DB5 Absolute value eccentricity
210 DB1 Status
DB2...DB5 Angle eccentricity
211 DB1 Status
DB2...DB6 Standard deviation eccentricity axis - Blue
212 DB1 Status
DB2...DB6 Standard deviation eccentricity axis - Magenta
213 DB1 Status
DB2...DB6 Standard deviation eccentricity absolute value
300 DB1 Status mean value
DB2...DB6 Mean value
DB7 Sign deviation
DB8...DB12 Deviation
DB13...DB17 Value - Blue >> diameter axis - Blue
DB18...DB22 Value - Magenta >> diameter. axis - Magenta
DB23...DB27 Ovality
DB28DB37 0 as dummy
ID DB Content:
Request -ASCII- Actual value
-ASCII-
301 DB1...DB5 Nominal value
DB6...DB9 +Tolerance
DB10...DB13 -Tolerance
DB14...DB17 Lump tolerance
DB18...DB21 Neck tolerance
DB22DB25 ? as dummy
320 DB1 Status mean value
DB2DB6 Mean value
DB7 Sign deviation
DB8DB12 Deviation
DB13...DB17 Value - Blue diameter >> axis - Blue
DB18DB22 Value - Magenta diameter >> axis - Magenta
DB23DB27 Ovality
DB28 Status eccentricity
DB29DB33 Absolute value eccentricity
DB34...DB38 Angle eccentricity
321 DB1...DB5 Nominal value
DB6...DB9 +Tolerance
DB10...DB13 -Tolerance
DB14...DB17 Lump tolerance
DB18...DB21 Neck tolerance
DB22 MOTOR STOP
330 DB1 Status mean value
DB2DB6 Mean value
DB7 Sign deviation
DB8DB12 Deviation
DB13...DB17 Value - Blue diameter >> axis - Blue
DB18DB22 Value - Magenta diameter >> axis - Magenta
DB23DB27 Ovality
DB28 Status eccentricity
DB29DB33 Absolute value eccentricity
DB34...DB38 Angle eccentricity
DB39 Status standard deviation ECC axis - Blue
DB40...DB46 Standard deviation ECC axis - Blue
DB47 Status standard deviation ECC axis - Magenta
DB48...DB54 Standard deviation axis - Magenta
DB55 Status standard deviation ECC absolute value
DB56...DB62 Standard deviation ECC absolute value
ID DB Content:
Request -ASCII- Actual value
-ASCII-
331 DB1...DB5 Nominal value
DB6...DB9 +Tolerance
DB10...DB13 -Tolerance
DB14...DB17 Lump tolerance
DB18...DB21 Neck tolerance
DB22 MOTOR STOP
DB23...DB29 Tolerance standard deviation eccentricity
DB Meaning:
-ASCII- Setting for requested parameters
DB1DB5 Mean value, 5 Digit ASCII if value o.k. Otherwise BLANK or HELP
PP Position decimal point
S Status mean value
DBB1DBB5 Mean value BOT-camera
SB Status BOT-camera
DBT1DBT5 Mean value TOP-camera
ST Status TOP-camera
DBMI1DBMI5 Minimum value
DBMA1DBMA5 Maximum value
The required projecting files for the SIKORA measuring device are supplied on the delivered product
CD. While using the projecting files, you can select from the amount of data modules for data
transmission listed in this instruction manual.
6.7.1 INTERFACE
The device interface supplies the voltage needed for the Profibus load resistances. The load
resistances themselves are not integrated.
In order to keep a reliable system the SIKORA AG recommend installing a terminating element, like
for instance the Siemens "Load Resistance for Profibus" Order No. 6ES7972-0DA00-AA0.
If an Actor or Sensor is used as terminating element for the bus system, removing this element will
not remain without consequences.
Module 1 Module 2
IN data: DB0DB16 IN data: DB0DB24
OUT data: DB0DB5 OUT data: DB0DB8
cv2000_1.gsd cv2000_3.gsd cv2000_2.gsd cv2000_3.gsd
Module 3
IN data: DB0DB25
OUT data: DB0DB9
cv2000_3.gsd
Module 4
IN data: DB0DB8
OUT data: DB0DB7
cv2000_3.gsd
1) If no configuration with LUMP option is available, this value will be set to zero.
1) Nominal value DM could be used for output DEVI analog or to calculate absolute threshold for lump option.
2) FFT/SRL settings in DB5 and DB8 are ignored if ECOCONTROL (Master) is connected.
1) If no configuration with LUMP option is available, this value will be set to zero.
DB Meaning Value
Content: Remark
Actual Value
DB6...DB7 Setting of SRL- (speed*10) line speed [m/min] or [ft/min]
Parameters 2) 1-100 rel. prop. velo. [%]
0 - 65535 cable length [m] or [ft]
(atten*10) attenuation condition [dB/100m] or [dB/100ft]
attenuation freq. cond. [MHz]
DB6 control: frequency 0 = 260.4 Hz
range 1 = 130.2 Hz
2 = 52.1 Hz
3 = 26.0 Hz
4 = 13.0 Hz
5 = 5.2 Hz
6 = 2.6 Hz
7 = 1.3 Hz
8 = 0.5 Hz
DB7 control: FFT-length 0 = 512
1 = 1024
2 = 2048
3 = 4096
4 = 8192
(atten*10) attenuation dielectrium [dB/100m] or [dB/100ft]
atten. freq. dielectrium [MHz]
DB6 cable construction 0 = Coax: Solid Outer Cond.
1 = Coax: Braided Outer Cond.
2 = Cat.UTP: Small Pairage
Cable
3: Cat.UTP: High Pairage
Cable
4: Cat.STP
DB7 RL Prediction source 0 = O.D. Insul. Core (Layer 2)
1 = O.D. Cond.
2 = O.D. Insul. Layer 1
1 - 255 FFT/SRL average
0 - 65535 nom.diam. insul. core [selected unit]
(layer 2)
0 - 65535 nom.diam. core [selected unit]
1) Nominal value DM could be used for output DEVI analog or to calculate absolute threshold for lump option.
2) FFT/SRL settings in DB4...DB7 are ignored if ECOCONTROL (Master) is connected.
7. OPERATION
The CENTERVIEW 8000 / 8000e can:
Either work as a stand-alone device with its data being sent out via the analog interface (Option),
in this case the analog output channels must be configured using the setup and diagnostic
software unless you prefer to use the default values set at our factory.
Or can be controlled via one of the three serial interfaces by an external computer, an
ECOCONTROL (SIKORA display unit) or by a Profibus system. If necessary, the interfaces have to
be configured using the Setup and Diagnosis Software.
HINT Using the analog output does not exclude the use of any
serial interface and vice versa.
Its relay contact informs you whether the CENTERVIEW 8000 / 8000e is ready or not.
Use the CENTERVIEW 8000 / 8000e digital input to activate the MOTOR STOP function.
While the MOTOR STOP function is active, the generator will be controlled down during line
stand-still, to prevent from danger by touching the energized conductor.
At restart and non-active MOTOR STOP, the generator will be controlled up again.
During line stand-still the wire or conductor could possibly hang down in the measuring head
opening. In this case, its motors will move down the measuring head in order to center the wire
again. When the line is running again, the wire will get pulled tight again and will possibly drag along
the edge of the measuring head opening.
As soon as the line is stopped, the measuring head motors will be stopped as well in order to prevent
the head from following the wire.
When the line runs again, the motors will be enabled again.
OK OK OK
ON ON ON
No Receiver Signal No Receiver Signal No Receiver Signal
OFF Receiver not ready OFF Receiver not ready OFF Receiver not ready /
Internal inteface
SIOB-SPU not ok
(check Hardware
Revision)
OK OK OK
OFF ON OFF
Inductive Error No Receiver Signal Inductive Error
The graphical display can be changed via function key to the Diameter display (page 96), the Wall
thickness display (page 97) and a Cloud display (page 98).
REFERENCE The display language can be set at the configuration page (page 99).
All picture examples at this instruction manual are made with the english
language pack, which is default at the CENTERVIEW 8000e display.
Through definite position and symbolic of every theme it can be assign to
any language.
This graphical display shows the diameter as cross section (means a graphical cross section of the
product).
The outer diameter and the eccentricity are available as display windows in mm or inch.
Therefore, touch the corresponding value (black font) and enter the
desired value via a numeric pad.
This graphical display is just available with a set core diameter (Nom(Core)), because the
CENTERVIEW 8000 / 8000e can determine the outer diameter only.
It shows the wall thickness as cross section (means a graphical cross section of the product).
The outer diameter and the average wall thickness are available as display windows in mm or inch.
Therefore, touch the corresponding value (black font) and enter the
desired value via a numeric pad.
This graphical display shows the eccentricity of the product as a scatter plod (cloud) display. This
makes it possible to have a detailed overview about the measuring values and its occurrence,
without disregarding the single measuring value.
The outer diameter and the average wall thickness are available as display windows in mm or inch.
Therefore, touch the corresponding value (black font) and enter the
desired value via a numeric pad.
Master Mode ON While activated, the core diameter as well as the nominal values and
tolerances at the display windows can be manual set.
OFF Core diameter and nominal values / tolerances can not be manual set.
The CENTERVIEW gets its designations via profibus or ECOCONTROL
and works as REMOTE.
Wall Thickness ON Needs the designation of the core diameter (Nom(Core)) for right
calculation. It has to be larger than zero.
If an ECOCONTROL display device is connected to the CENTERVIEW 8000e, the monitor automatically
switches to SL AV E mode.
In that case, all configurations from the ECOCONTROL are imported and via the monitor are no
configurations changeable.
C L O U D D I S P L AY
If an ECOCONTROL display device is connected to the CENTERVIEW 8000e, the cloud display (scatter
plot display) can be activated either at the ECOCONTROL or at the monitor (not both at same time).
8. MAINTENANCE
The SIKORA AG recommend taking care that the openings for the light are kept clean and free of any
kind of dirt. To be sure, check the openings regularly, for instance after production completion.
If this message occurs occasionally, you should check and clean the measuring openings at the next
possible opportunity. Do not wait until the message keeps occurring permanently!
If the message keeps occurring after cleaning, we recommend you repeat the cleaning procedure
while simultaneously checking the result by watching the Raw Video at the DIAGNOSTIC software
(Function D I AG N O S I S / V I D E O / R AW V I D E O ).
For cleaning, use only the cleansing tissues provided for this measuring head. Do not use anything
else like a piece of cloth, a sponge or similar!
If you run out of cleansing tissues, you can order more at our service office. Only apply those brands
of tissues approved by us. Do not use just any kind of cleansing tissues, as some of these may leave
residue.
Remove the measuring object from the CENTERVIEW 8000 / 8000e before opening the device.
Remove the measuring head cover:
Open the latches at both sides of the measuring head and carefully pull off the cover. Do not
grasp into the openings.
Replace the cover. Regard to position the red point on the protective hood above the red point
on stand. Do not grasp into the openings to prevent from finger crushes.
Close the latches.
NOTICE Always keep the measuring head frame close, otherwise the device
could gets dirty and reacts on extraneous lights!
Regard this advice for intended use of the SIKORA device and its components. Disregard can lead to incorrect use
up to physical device damage.
REFERENCE The calibration set ist for checking only. It can not be used for the
adjustment of the measuring device. Therefore you need a
corresponding calibration construction and a special wire.
Proceed as follow:
Open the cover from your CENTERVIEW 8000 / 8000e measuring head.
Take the desired pin gauge as well as the pin attachment from the calibration equipment.
Lead the pin attachment inside the CENTERVIEW 8000 / 8000e measuring head and start to
check the measuring accuracy via the DIAGNOSTIC software.
Therefore, you have to connect the PC with the DIAGNOSTIC software once to the diagnostic
interface at the device front side and once to the diagnostic interface at the device rear side.
Both times the check-up has to be done. While looking frontal to the device from the side with
the connected diagnostic interface, the left optical system will be checked.
After finish the measuring accuracy check, close the cover of the CENTERVIEW 8000 / 8000e
measuirng head again.
To prevent the measuring device from permanent adjustment, if it detects the pin gauge at both
optical systems but is not insert accurately centered, activate the MOTOR STOP function (page 92)
first.
9. APPENDIX
10. GLOSSARY M
MEASURING AXES
A Measuring axes denotation:
X-axis = Blue axis = Vertical
AC / DC Y-axis = Magenta axis = Horizontal
= Alternating current (voltage)
= Direct current (voltage)
O
C OSC
= Outer semi conductor
CCD
= Charge-coupled device
Light sensitive, electronic component. P
PE
D = Protective earth
Electric conductor for safety purpose (protection against
DM electric strike).
= Diameter
DSP
Q
= Digital signal processor QGE
Used for continuous processing of digital signals. = Quartz crystal generator.
Special high frequency generator for the SIKORA devices
E CENTERVIEW 2000/8000 and SICCON 600.
ECC
= Eccentricity
U
UNIT OF LENGTH
EMC At the manual all corresponding lengths are stated as
= Electromagnetic compatibility metrical and american unit of length.
Describes the freedom from interferences for electrical 1 meter (m) = 3.28 feet (ft)
or electronic devices with their environment. 1 millimeter (mm) = 0.039 inch (inch)
1 micrometer (m) = 0.039 Mil (mil)
G
W
GND
= Ground WT
Reference potential for signal and operating voltage. = Wall thickness
I
INS
= Insulation
IR
= Infrared
ISC
= Inner semi conductor
L
LED
= Light-emitted diode.
A diode emitting light by flowing current (as infrared
radiation or ultraviolet radiation, depending on type).
11. INDEX
EXAMPLE 2 - CALIBRATION AFTER COOLING TROUGH ............. 124
F
A FIXING THE DEVICE STAND ........................................................ 34
AC / DC .................................................................................. 128 FORESEEABLE MISUSE ................................................................ 14
ALIGN THE MEASURING HEAD ................................................... 33 FRAME CLEANING ................................................................... 101
ANALOG OUTPUT / DIGITAL INPUT / RELAY CONTACT .............. 68
APPENDIX ................................................................................ 106
G
GND ....................................................................................... 128
B GROUNDING ............................................................................ 38
BEHAVIOR IN CASE OF EMERGENCY ......................................... 14 GROUNDING AT CENTERVIEW 8000 C ...................................... 41
GROUNDING SCREW AT THE DEVICE STAND ............................. 41
C
CCD ....................................................................................... 128
H
CENTERVIEW 8000E TFT MONITOR .....................................95, 103 HIGH FREQUENCY GENERATOR (CA-250) ............................... 116
CENTERVIEW 8010 .................................................................. 107 HIGH FREQUENCY GENERATOR (QGE) ..............................46, 115
CENTERVIEW 8010 C ............................................................... 109
CENTERVIEW 8010E ................................................................ 108
I
CENTERVIEW 8010E C ............................................................. 110 INITIATION ................................................................................ 56
CENTERVIEW 8025 .................................................................. 111 INS .......................................................................................... 128
CENTERVIEW 8025 C ............................................................... 113 INSTALL THE CALIBRATION CONSTRUCTION ............................. 58
CENTERVIEW 8025E ................................................................ 112 INSTALLATION ........................................................................... 29
CENTERVIEW 8025E C ............................................................. 114 INSTRUCTION MANUAL REFERENCES ........................................... 6
CHARACTERISTIC FOR CENTERVIEW 8010 / 8010E ................... 105 INTENDED USE .......................................................................... 13
CHARACTERISTIC FOR CENTERVIEW 8025 / 8025E ................... 105 INTERFACE ................................................................................ 81
CHECKING THE MEASURING ACCURACY ................................ 104 INTERFACES OVERVIEW ............................................................. 27
CHECKSUM BYTE BCC ............................................................... 74 IR ............................................................................................ 128
CLOUD DISPLAY ........................................................................ 98 ISC .......................................................................................... 128
CONFIGURATION PAGE ............................................................ 99
CONFIGURATION VIA DIAGNOSTIC SOFTWARE ........................ 71 L
CONNECTION CABLES TO SIKORA DISPLAY DEVICES ................. 69
LED ....................................................................................52, 128
CONNECTION GENERATOR - CENTERVIEW ............................... 50
CONNECTION TO ECOCONTROL 2000 ................................... 69 M
CONNECTION TO ECOCONTROL 600 ..................................... 70
MAINTENANCE ........................................................................ 101
CONNECTION TO ECOCONTROL 6000 / 1000 ........................ 69
MEASUREMENT OPENING CLEANING ...................................... 101
CONNECTION TO REMOTE 2000 / DISPLAY 2000 ..................... 70
MEASURING AXES .................................................................... 128
CONSIDERATION OF THE INSTRUCTION MANUAL ..................... 15
MEASURING OPERATION ........................................................... 65
CONTACT ................................................................................... 5
MEASURING PRINCIPLE .............................................................. 19
P Z
PE ........................................................................................... 128 ZERO ADJUSTMENT ECCENTRICITY ............................................ 56
PROCEEDING ............................................................................ 60
Q
QGE ....................................................................................... 128
QGE - CONNECTIONS ............................................................. 48
QGE - CONNECTIONS (OLD VERSION) ..................................... 49
QGE - DIMENSIONS .................................................................. 47
R
RECEIVING TELEGRAM SLAVE (MEASURING DEVICE) ................... 75
REMOTE INDICATION (MEASURING DEVICE) - EXTERNAL PC ....... 80
REQUEST FOR PERSONNEL ........................................................ 13
REQUIREMENTS ......................................................................... 57
RESIDUAL RISK AND PRECAUTIONS ............................................ 15
REVISION OVERVIEW ............................................................... 126
RS232 CONNECTION CABLE FOR SETUP/DIAGNOSTIC ............. 72
S
SAFETY ADVICES ........................................................................ 10
SAFETY LABELING AT DEVICE ..................................................... 15
SC400 (ASCII) TELEGRAM ........................................................... 73
SCOPE OF DELIVERY .................................................................. 28
SENDING TELEGRAM SLAVE (MEASURING DEVICE) ..................... 76
SETTING POSSIBILITIES VIA S1 TO S3 .......................................... 53
SIGNAL LINES ............................................................................ 44
SMALL DIAMETER AT CENTERVIEW 8010 / 8010E ........................ 66
SPECIALTY FOR CENTERVIEW 8000 C ......................................... 35
STRUCTURE AND CONTENT OF THE MAIN CHAPTERS .................. 9
SWITCH DEVICE OFF ................................................................. 91
SWITCH DEVICE ON .................................................................. 91
SYSTEM SURVEY ....................................................................... 119
T
TECHNICAL DATA ...................................................................... 25
TECHNICAL DRAWINGS ........................................................... 106
TELEGRAM STRUCTURE .........................................................73, 82
TIME OUT.................................................................................. 74
TOROIDAL TRANSFORMER ....................................................... 117
TOROIDAL TRANSFORMER AT STAND ....................................... 118
TRANSPORT LOCK ..................................................................... 32
U
UNIT OF LENGTH .................................................................... 128
USE OF AN ECOCONTROL ..................................................... 100
USED SIGNAL WORDS AND SYMBOLS .......................................... 7
V
VALUE SETTINGS ....................................................................... 74
VOLTAGE SUPPLY ...................................................................... 42
W
WALL THICKNESS DISPLAY .......................................................... 97
WIRING ..................................................................................... 37
WT .......................................................................................... 128