CS-2
OPERATION MANUAL
CODE NO. 0801
Specification .....................................................243
Index ..................................................................247
Index for Purpose
Order
Exam Order Button Display ................................................................................................................97
Icons displayed with the Exam Order .................................................................................................98
Deleting unnecessary exam orders. ...................................................................................................38
Deleting the Exam Order that is suspended .......................................................................................70
Image
Adjusting the density (Density Slide Bar) ..............................................................................................42
Adjusting the contrast (Contrast Slide Bar) ...........................................................................................43
Resetting the processing condition to the initial status. ......................................................................44
Replacing the image that shall be output with detailed settings. ........................................................50
Exchanging the right and left images (2 on 1 output) ........................................................................56
Setting the trimming frame (output in 1 to 1).....................................................................................111
Changing the Exam Order to the other. ..............................................................................................43
Indication of the [Rotate/Flip] button ...................................................................................................99
Rotate/Flip of the Image......................................................................................................................45
Changing the image to be displayed in the output preview field.
(changing the image for the output with detailed settings).................................................................60
For accurate masking process ..........................................................................................................100
Trouble
Disabling the screen saver................................................................................................................202
Disabling the screen saver (Operator Securiy Mode) .......................................................................190
When the [KONICA MINOLTA] button is turned red. ..........................................................................91
Reaction to the problem when the CS-2 is started up
with a floppy disk inserted in the drive. .............................................................................................202
Others
Restarting the suspended examination...............................................................................................70
Deleting the Exam Order that is suspended. ......................................................................................70
Sorting the "Exam List" .......................................................................................................................93
Updating the "Exam List" ....................................................................................................................94
To return to the list prior to extraction ...............................................................................................124
User Tool
File name of the backup data............................................................................................................183
Initial of the [Sex] selection. ..............................................................................................................173
Foreword
<1>
CS-2 Operation Manual Ver.1.10 2005.2
Foreword
Features
Notes
(1) Unauthorized reproduction of this manual in whole or in part is prohibited.
(2) The content of this manual is subject to change without prior notice.
(3) Should any discrepancies, errors or omissions be discovered, KONICA MINOLTA should be contacted.
(4) Notwithstanding item(3) above, KONICA MINOLTA shall accept no responsibility of any claims made against loss or
loss of profits arising from use of the product.
Trademarks
- "Microsoft" and "Windows" are trademarks or registered trademarks of Microsoft Corporation (USA) valid in the USA
and other countries.
- "Windows XP" is the abbreviation of "Microsoft Windows XP Professional Edition Operation System".
- Other company or product names used in this manual are trademarks or registered trademarks. The indication ,
and TM shall be omitted henceforth.
<2>
CS-2 Operation Manual Ver.1.10 2005.2
About
this Manual
Be sure to read this chapter before using this
manual.
<3>
CS-2 Operation Manual Ver.1.10 2005.2
About this Manual
<4>
CS-2 Operation Manual Ver.1.10 2005.2
About this Manual
Page Layout
2 Touch the rough classification of the body parts that operational procedure
Hint shall be registered in the exam order.
The rough classification of the body parts
is shown.
2 changes its color from blue to green.
Points to be Exam tags displayed beneath the rough
classification changes according to the
noted. selected classification.
Operation Procedure
3 Operation procedure is
4 Insert the cassette that has been exposed into explained in order.
the insertion slot of the cassette reader.
Scanned image will be automatically
associated with the exam order selected in the
Caution Step.3, and the scanned image will be
displayed in the display field of each exam
Settings or other Inserting the cassette into the REGIUS MODEL 170 will issue the exam order.
things to be made REGIUS MODEL 170 can hold another cassette as a stack. An exam order is
issued for this stacked cassette according to the selected exam tag as well. If a Reference
through the User
wrong exam order is issued for the stacked cassette, correct it either by following
the forementioned procedures or remove the stacked cassette once, and insert it Shows reference
again into the slot after selecting the correct exam tag.
Tool. pages or sections.
Contact your CS-2 Operation Manual Ver.1.00 2004.9
< 34 >
See the page for
system more details.
administrator to
change the settings
using the user tool.
Parenthesis in a sentence.
The word in this mark shows that it is
displayed in the touch panel of the CS-2 Ex. "Ready"
operation unit.
<5>
CS-2 Operation Manual Ver.1.10 2005.2
About this Manual
In this manual, the main unit of PC is called "CS-2 Controller Unit" while the touch
panel of the CS-2 is called "CS-2 Operation Unit".
Most of the operation of the CS-2 can be implemented by touching the CS-2
Operation Unit. However, it is recommended to use accessories keyboard and
mouse to operate the User Tool.
<6>
CS-2 Operation Manual Ver.1.10 2005.2
Chap.
Safety
Precautions &
1 When using the REGIUS Console CS-2, the
cautions detailed in this chapter must be strictly
Warnings followed in order to correctly use the device.
<7>
CS-2 Operation Manual Ver.1.10 2005.2
Ch.1 Safety Precautions & Warnings
Prohibited Do not touch Do not disassemble Do not touch Do not expose Mobile Multi-Leads
with wet hand to moisture Prohibited Prohibited
Other Symbols
There are 3 degrees of caution labels, and each is used depending on the level of risk and damage caused by
incorrect use and mishandling.
DANGER : Failure to observe the caution will produce high risk of serious or fatal injury.
WARNING : Failure to observe the caution will produce moderate risk of serious or fatal injury.
CAUTION : Failure to observe the caution will produce risk of moderate or light injury.
High Low
NOTE : If the contents of this page are not legible, order a new manual.
<8>
CS-2 Operation Manual Ver.1.10 2005.2
Ch.1 Safety Precautions & Warnings
Cautions in Use
Following cautions must be observed when handling the CS-2 and followed in order to handle the CS-2 in a correct
manner. Before using the CS-2, thoroughly read this operation manual, and become fully familiar with the operation of
the CS-2.
WARNING
CS-2 is classified as Class B Information Technologies Equipment based on the standard prescribed by VCCI (Voluntary
Control Council of Interference by Information Technology Equipment).
Use of this product may cause electromagnetic disturbance. In such cases, the user may be requested to take neces-
sary actions to remedy the problem.
Disposal
Disposal of the CS-2 must be carried out following the cautions listed below.
Always contact your technical representative when it is necessary to replace the lithium battery that is
incorporated in the CS-2.
Follow the Operation Manual attached to each optional device when disposing of them.
Refer to the Operation Manual attached to each device for the operation of the PC and display that are the
components of the product.
Installing the LCD display on an unstable surface may cause it to fall and result in damage. Install the display
on a firm and safe surface.
CD-RW drive is exclusively for the service engineer's use to carry out the maintenance service.
Users are requested not to use this drive for any purpose.
Devices that the user purchased from the supply source other than Konica Minolta cannot be connected
to the CS-2.
Software that the user purchased from the supply source other than Konica Minolta cannot be installed on
the CS-2.
NOTE : If the contents of this page are not legible, order a new manual.
<9>
CS-2 Operation Manual Ver.1.10 2005.2
Ch.1 Safety Precautions & Warnings
Overall Cautions
WARNING
The unit must not be used if smoke, unusual odors or sounds are detected.
In the event of detection of smoke, unusual odors or sounds, the power should be switched off immediately
and the power plug disconnected from the power outlet. Then contact your Service Representative.
Continued use under such circumstances could result in fire.
Do not insert wire or metal blades.
Do not insert wire, metal blades or other foreign objects into the vent or disk slot; there is a danger of electro-
cution.
CAUTION
Do not introduce water or other foreign objects into the CS-2.
Do not introduce foreign objects such as water or other liquids, pins or clips into the CS-2: fire, electrocution
or damage may result. In the event of accidental introduction of such foreign objects or substances into the
unit, the power should be switched off immediately and the power plug disconnected from the power outlet.
No attempts should be made to disassemble the unit. Then contact your technical representative.
CAUTION
Shutdown or removal of the disk from the CS-2 should be implemented only after confirming that the
access lamp of the CS-2's hard disk is extinguished.
When recycling the power of the CS-2, leave it for minimum 10 seconds after the power is turned
OFF, and then turn the power ON.
When any work that may cause vibration to the product should be performed, before starting the
work, leave the product at least 30 seconds after it is shut down.
Regularly clean the CS-2.
Regular cleaning helps to avoid various malfunctions in advance.
NOTE : If the contents of this page are not legible, order a new manual.
< 10 >
CS-2 Operation Manual Ver.1.10 2005.2
Chap.
Product Outline
2
The product outline of CS-2 is described in this
chapter.
< 11 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.2 Product Outline
Which of these three screens to be displayed as an initial screen shall be set by the
technical representative at the time of installation considering the operation style of
the user. When "Exam Registration Screen" or "Exam List Screen" is selected as an
initial screen, they can be interchanged by the operator. p.73 Switching the
Initial Screen
When the "Image Confirm Screen" is set as an initial screen, it is not possible to
switch the screen to "Exam Registration Screen" or "Exam List Screen".
When the "Image Confirm Screen" is set as an initial screen, it is not possible to
search the patient info.
It is possible to check the registered exam info by opening the "Exam List Screen" from the p.93 Exam List
Screen
"Exam Registration Screen" or to newly input the patient info by opening the "Patient Info ("Exam List"
Input/Search Screen" that is similar as the "Exam Registration Screen" from the "Exam List screen as the
Screen". However, image reading on the CS-2 is only initiated from the Image Confirm screen. initial screen)
< 12 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.2 Product Outline
< 13 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.2 Product Outline
Which type of the exam order registration to be used is set up when the system is
installed by your technical representative considering the operation style of the user.
Once it is set up, it can not be changed by the user.
Auto-Registration
Selecting the appropriate Exam Tag key before the cassette is read on the REGIUS MODEL
170/190, the inserted cassette is automatically read using the preset process condition according
to the order. This will save labor required to register the order in advance.
Exam Tag Key
A set of conditions
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e? @@@@@@@@
specifying the body
@@h?
@@h? @@
@@h? @@
@@
@@h?
@@h? @@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
image processing con-
@@ @@
@@ @@
@@ @@
@@ @@
@@
dition (Density,
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@
Contrast, etc.).
@@ @@
@@
@@ @@
1
@@ @@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@ @@
@@g
@@g
@@g
@@g
@@g
@@g
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
?@@
?@@
?@@
?@@
?@@
?@@
?@@@@@@@@
?@@@@@@@@
Selecting an Exam Tag
Key suitable to the
1. Select the Exam Tag 2. Insert the exposed 3. Automatically register Body Part being exam-
key. cassette. the order, and displays ined allows an image
the image accordingly.
processing suitable to
that image.
Barcode Registration
Reading the barcode on the cassette, the cassette is associated with the exam order that has
been registered.
Barcode reader is required when implementing the barcode registration.
XXXXXXXXXXXxxxxx
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?
@@h? @@@@@@@@
@@h? @@
@@h? @@
@@
@@h? @@
@@h?
@@h? @@
@@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@
@@
@@ @@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@g ?@@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
< 14 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.2 Product Outline
1
2
1. Touch the appropriate 2. Touch the exam order to 3. Insert the cassette in
Exam Tag keys as many set the order of cassette order of cassette
times as the cassettes to reading. reading set in the
be read. Touching the exam order will step.2.
Exam orders will be registered. change the number for reading
order.
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?
@@h? @@@@@@@@
@@h? @@
@@
@@h?
@@h? @@
@@h? @@
@@h? @@
@@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@g
@@g ?@@
@@@@@@@@ ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
< 15 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.2 Product Outline
p.95 p.121
p.104
note 1
p.211 p.213
p.109
Properties Screen
A screen for outputting the
image to printer, viewer/
server, FD.
p.125
System Shutdown
p.23
< 16 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.2 Product Outline
p.121
p.104
note1
p.211 p.213
p.109
Properties Screen
A screen for outputting the
image to printer, viewer/
server, FD.
p.127
p.125
System Shutdown
p.23
< 17 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.2 Product Outline
p.103
p.109
Properties Screen
A screen for outputting the
image to printer, viewer/
server, FD.
p.127
p.125
System Shutdown
p.23
< 18 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.2 Product Outline
p.152
System Shutdown
p.23
p.73 Switching
When the initial screen is "Exam Registration" or "Exam List" screen, [Initial Screen] the Initial
button appears on the "System Menu" screen. Touching this button will switch the Screen
initial screen between "Exam Registration" and "Exam List" screen.
< 19 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.2 Product Outline
< 20 >
CS-2 Operation Manual Ver.1.10 2005.2
Chap.
Operation
3
How to operate the CS-2 is detailed in this
chapter.
< 21 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.3 Operation
Start Up
How to start up the CS-2 is described below.
Before turning on the CS-2, check that the devices (REGIUS MODEL 170/190,
Printers, Viewers, Server, etc.) are correctly connected to the CS-2, and the power
of those devices has been turned on.
Check that a floppy disk is not inserted in the CS-2 Controller Unit. If it is inserted,
push the release button to remove it. With the floppy disk inserted in the CS-2
Controller Unit, CS-2 will not start up.
Three types of screens ("Image Confirm", "Exam Registration" and "Exam List" p.12 Three Initial
Screens
screens) are available as an initial screen for the CS-2.
< 22 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.3 Operation
The procedure detailed below must be followed when switching off the CS-2 power
supply. Do not switch off power from the control unit power switch: doing so may re-
sult in loss of data or damage to the system software.
1
1 Touch the [KONICA MINOLTA] button at the top
left of the initial screen ("Image Confirm" screen
The figure at left shows
or "Exam List" screen).
display when the
The system menu will be displayed. "Image Confirm"
screen has been set as
the initial screen.
2
3 Touch the [YES] button.
CS-2 termination process will be carried out
and, after a short time, power will be
automatically switched off.
To discontinue system termination, touch the
3 [NO] button and then the [Back] button on the
system menu. Termination will be aborted and
the initial screen ("Image Confirm", "Exam
Registration" or "Exam List" screen)
displayed.
Upon termination of CS-2, the reader unit will
automatically be put into power-save (sleep)
mode.
If image data is still being transmitted to external devices such as printers, a mes-
sage will be displayed. System shut-down will be executed upon completion of im-
age transmission. Any images not output upon completion of system shut-down will
be transmitted the next time CS-2 is activated.
< 23 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.3 Operation
Input Panel
[ABC/abc]
Displayed when an input column requiring input of
alphabetical or numerical characters is selected.
Touching the "Caps Lock" key will switch between
upper and lower case characters.
[Keyboard]
Displayed when an input column requiring input of
alphabetical or numerical characters is selected.
Touching the "Caps Lock" key will switch between
upper and lower case characters.
< 24 >
CS-2 Operation Manual Ver.1.10 2005.2
Chap.
Sequence of
4
Examination This chapter describes the sequence of
examination using CS-2.
< 25 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
or or
Operation only with "Image Examination can be started Select the patient info that shall
Confirm" screen without after inputting the patient info. be examined from the list of
opening patient info input It is also possible to select the patient info. If no patient info to
screens. Touching the [Patient exam orders suspended or be examined is found in the list,
Info Disp] allows sent from higher HOST by a new patient info can be
addition/change of patient switching to "Exam List" added to the list.
p.12 info. screen.
Three Initial p.27 p.29 p.31
Screens
Exam order is
registered.
p.36 p.39
REGIUS MODEL 170/190 scans and the image is displayed on the CS-2.
p.14
the image. . . . .
Three Ways of
Exam Order
Registration
Normal output Output per order Normal output of Output after changing
Image Output per order with detailed setting all orders the print format
Output is made per Output is made per Outputs all orders at a Outputs 2 or 4 images
image after check. image after setting time. per film at a time.
or details. or or
Detailed
< 26 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Examination Sequence
Procedures for patient info registration through image confirmation are detailed below.
As described in the chapter 2, there are several choices of initial screen, image format and exam order registration.
Examination sequence for each combination of these choices is described below.
As for the image format, because the operation is the same whether it is single-image or four-image format, a single-
image format is taken as an example in the following explanation.
When the system is configured to receive the patient info from upstream HOST in
the network, the patient info will be automatically displayed when it is received, thus
it is not necessary to manually input them.
@@@@@@@@e?
@@@@@@@@e?
@@h? @@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?
@@@@@@@@
@@h? @@
@@h? @@
@@
@@h?
@@h? @@
@@h? @@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@g ?@@
@@
@@
@@
@@g
@@
?@@
@@
@@
@@
@@
@@g ?@@
@@g ?@@
@@g ?@@
@@g ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@ ?@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@ ?@@@@@@@@
When receiving
the patient info
from upstream
HOST. Patient info is displayed in
the Patient Info field.
1
1 Touch the Patient Info filed.
"Exam Registration" screen is displayed.
< 27 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Following this screen, carry out exam order registration and image
scanning.
< 28 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
When the system is configured to receive the patient info from upstream devices in
the network, the patient info will be automatically displayed when it is received, thus
it is not necessary to manually input them.
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e
@@@@@@@@
@@h?
@@h? @@
@@h? @@
@@
@@h? @@
@@h?
@@h? @@
@@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@g
@@g ?@@
@@@@@@@@ ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@
?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
When receiving
the patient info
from upstream
devices. Patient info is displayed in
the Patient Info field.
To move the cursor from one item to the other in the Patient Info field.
Touching [ENTER] button after inputting first item moves the cursor to the next item.
To skip the item that requires no input, touch [ENTER] button or directly touch the
next item to be input.
@@@@@@@@e?
@@@@@@@@e?
@@h? @@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e
@@@@@@@@
@@h? @@
@@
@@h?
@@h? @@
@@h? @@
@@h? @@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@g ?@@
@@
@@
@@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@
?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
Input the ID, and touch the Cursor moves to the name
[ENTER] button. column.
< 29 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Images and informations that have been already sent to the external devices will not be
automatically resent when they are changed after they are sent.
Following this screen, carry out exam order registration and image
scanning.
< 30 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
When the system is configured to receive the patient info from upstream devices in
the network, the patient info will be automatically displayed when it is received, thus
it is not necessary to manually input them.
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?
@@h? @@@@@@@@
@@h? @@
@@
@@h? @@
@@h? @@
@@h?
@@h? @@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@g ?@@
@@g ?@@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@ ?@@
?@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@ ?@@@@@@@@
When receiving
the patient info
from upstream
devices. Patient info is displayed in
the Patient Info field.
1
2 Input the Patient Info using the input panel or
keyboard.
< 31 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Images and informations that have been already sent to the external devices will not be
automatically resent when they are changed after they are sent.
Following this screen, carry out exam order registration and image
scanning.
< 32 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Items of patient info that can be used as a search key: Patient ID, Name
Patient Info will be reg-
istered in the patient
1 Touch the input window for input of patient database when starting
the exam or switching
ID,name (English) and input the characters to
2 be searched from the input panel (or CS-2 the screen.
keyboard). However, it will not be
registered when "NO" is
selected in the confir-
1 mation dialogue.
3 Touch the scroll button to locate the Patient Info When one Patient Info is
the result of search, a list
when there are more than 10 search results.
of the search result will not
be displayed. It will be di-
rectly input in the "Exam
3 Registration" screen.
5
< 33 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
When the wrong exam order is associated with the image, that image will be dis-
played over the whole area of the CS-2 Operation Panel. Touch the icon for the cor-
rect body part after [OK] and [NG] buttons are displayed.
Inserting the cassette into the REGIUS MODEL 170/190 will issue the exam order.
REGIUS MODEL 170 can hold another cassette as a stack. An exam order will be
issued for this stacked cassette according to the selected exam tag as well. If a
wrong exam order is issued for the stacked cassette, correct it either by following
the forementioned procedures or remove the stacked cassette once, and insert it
again into the slot after selecting the correct exam tag.
< 34 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Output of each exam order with detailed settings p.49 Image Output
(Output with
detailed set-
tings per
Exam Order)
Ordinary output of all exam orders at a time p.51 Image Output
(Normally
output of all
Exam Orders)
< 35 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
4
5 Read the barcode of the cassette using a bar- p.232 Using the
Single Bar
Code Reader
code reader.
A short beep sound can be heard when the p.237 Using the
XXXXXXXXXXXxxxxx
barcode reading is successful. Multi Bar
The barcode of the cassette will be displayed in Code Reader
the field of the selected exam order.
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?
@@h? @@@@@@@@
@@h? @@
@@
@@h?
@@h? @@
@@h? @@
@@h? @@
@@
@@
@@
@@
@@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@
@@
@@g ?@@
@@g ?@@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
6 Repeat the Step.4 and 5 to read in the barcode Exposure -> Cassette
Scanning per each ex-
for each exam order. am order is also possi-
ble.
< 36 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
8 When there are several exam orders, the multi-bar code reader.
Image Confirm
Enabling this function allows the image to be automatically output when
a specific time elapses after scanning the image, assuming that the im-
age has been confirmed. When re-exposure is necessary, touch [NG]
button before the time is reached.
Remaining time (sec) before the image is confirmed is indicated on the
[OK] button.
Changing the setting of density, contrast, rotate/flip and masking
before the time is reached will cancel the image confirm process. It is
required to touch the [OK] button to restart the process.
Count down is cancelled if any error message is shown before the time
is reached. Closing the message box will restart the count down.
If the exam order for the cassette that is inserted following the Step.7 is not dis-
played in the "Image Confirm" screen, an image will not be displayed even though
the cassette is scanned without failure. Touching the scroll button to locate the ex-
am order for the cassette allows the scanned image to appear in the view field of the
corresponding exam order.
< 37 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Output of each exam order with detailed settings p.49 Image Output
(Output with
detailed set-
tings per
Exam Order)
Ordinary output of all exam orders at a time p.51 Image Output
(Normally
output of all
Exam Orders)
3 Touch the exam tag to be registered in the order. Deleting the Exam Order
The selected exam order will be registered, and
For the exam order that is
added to the image display list.
registered by mistake,
Continue touching the exam tag button to leave it without registra-
register the exam tag with the exam order.
tion, and touch [Complete]
button after completing
3 normal examination.
A dialogue to confirm
abortion of the exam order
will be displayed. Touch
[OK] on this dialogue to
abort the exam order.
@@@@@@@@e?
@@@@@@@@e?
@@h? @@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?
@@@@@@@@
@@h? @@
@@
@@h?
@@h? @@
@@h? @@
@@h? @@
@@
@@
@@
@@
@@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@
@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@g
@@g ?@@
@@@@@@@@ ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
When the cassettes are scanned according to the order selected in the Step.3, it is
not necessary to specify the scanning order. In this case, skip the Step.4 and pro-
ceed to the Step.5.
< 39 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Image Confirm
Enabling this function allows the image to be automatically output when
a specific time elapses after scanning the image, assuming that the im-
age has been confirmed. When re-exposure is necessary, touch [NG]
button before the time is reached.
Remaining time (sec) before the image is confirmed is indicated on the
[OK] button.
Changing the setting of density, contrast, rotate/flip and masking
before the time is reached will cancel the image confirm process. It is
required to touch the [OK] button to restart the process.
Count down is cancelled if any error message is shown before the time
is reached. Closing the message box will restart the count down.
< 40 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Output of each exam order with detailed settings p.49 Image Output
(Output with
detailed set-
tings per
Exam Order)
Ordinary output of all exam orders at a time p.51 Image Output
(Normally
output of all
Exam Orders)
< 41 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Processing condition can be practiced both before and after the image output. It is
also possible to alter the processing condition using "Exam History" screen.
< 42 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
@@@@@@@@e?
@@@@@@@@e?
@@h? @@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?
@@@@@@@@
@@h? @@
@@h? @@
@@h? @@
@@h? @@
@@
@@h? @@
@@ @@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@ @@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@g
@@@@@@@@ ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
< 43 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
1
2. Touch the Exam Tag that was original-
ly registered for the Exam Order.
Exam Tag for the Exam Order
resumes the original condition.
@@@@@@@@e?
@@@@@@@@e?
@@h? @@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?
@@@@@@@@
@@h? @@
@@h? @@
@@
@@h? @@
@@h? @@
@@h? @@
@@ @@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@g
@@g ?@@
@@@@@@@@ ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
< 44 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
3
Result of "180 Flip"
@@@@@@@@e?
@@@@@@@@e?
@@h? @@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?
@@@@@@@@
@@h? @@
@@
@@h? @@
@@h?
@@h?
@@h?
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
Rotated and/or flipped image is displayed.
p.98 Icons dis-
@@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
played with
@@ @@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
the Exam
@@
@@ @@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
@@
Order
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@g
@@g ?@@
@@@@@@@@ ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
< 45 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Masking
The purpose of this type of image processing is to black out areas outside the field
of X-ray exposure on the scanned image. This prevents undue light leakage and
eye fatigue during reading of printed or soft copy images.
OFF ON
< 46 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Whether this function should be enabled or not will be set by the service engineer
considering the operation at the user.
Input of the exposure dose of the X-ray can be accepted both before or after the im-
age is output. Also to be input on the "Exam History" screen.
When the x-ray info has not been input yet, the value for the X-ray info shown at the
upper left will be indicated as "******".
****** KV
****** mAs
****** dGym2
< 47 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
1 Display the image to be output using the scroll but- Images may be output
ton. one at a time upon com-
pletion of each image
reading operation, or in
batch mode upon com-
pletion of the reading
1 operation for all orders.
3 Check the image, and click the [OK] button. Tool setting.
1
2 Display the image to be output using the scroll but-
ton.
[Output] button will not be activated when the image is not displayed in the output
preview field. In this case, it is requested to select the image to be output following
the Step.2 ~ 5.
< 49 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
6
7 Touch the [Output] button.
The image will be printed according to the
settings made in the Step.6.
To save the image data on a floppy disk,
"Browse for Folder" screen will be displayed.
Select the target folder, and touch [Output]
7 button to save the image data .
Upon touching the [Output] button to send the data to the printer or viewer/server,
it is assumed that the image is confirmed (the image is judged "OK").
To save the image data on a floppy disk, always select the floppy disk drive (Drive
"A") in the "Browse for Folder" screen.
It is also possible to export the data to the external PCs, however it is limited to the
PCs exhibited in "My Network".
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@h?@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@ @@
@@h?
@@h? @@
@@h? @@
@@h? @@
@@h? @@
@@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@
@@
@@ @@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@ @@
@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@g
@@g ?@@
@@@@@@@@ ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@
?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
< 50 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
When all Exam Orders are output with the format set to "2 images to 2 frames" or "4 images to
4 frames", images will be aligned from left to right (
upper left -> upper right -> lower left -> lower right for 4 images) in order of the order registration,
and printed.
Main
Sec
Primary image Secondary image Sec
Two Images Aligned Horizontally (AA) Two Images Aligned Vertically (A)
A
Initial Image
Second Image
4 on 1 Output
Four different images may be output on one film.
< 52 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
When [100% Output] is selected in the output method, the image will be trimmed if
the scanned image size is larger than the image display field.
Trimming will be implemented with the image center aligned to the center of the im- When [Output Whole] is
age display field. Trimming layout can be altered using the "Properties' screen. selected in the output
method, the image may
be reduced if the
Output of 2 images to 2 frames, 4 images to 4 frames are only available for identical scanned size is larger
four Exam Orders. than the output size.
Stamps for [1 on 1 Output] (From 1 Image to 2 Frames on 1 Film) are located in one
place. Stamps for [2 on 1 Output] (From 2 Images to 2 Frames on 1 Film) are locat-
ed in one place for each frame. Stamps for [4 on 1 Output] (From 4 Images to
4Frames on 1 Film) are located in one place for each frame.
When "1 image to 2 frames" output is selected with the [Format] button in the
p.59 2 on 1 Output
"Image Confirm] screen, the output orientation is automatically selected so that the
(1 image to 2
maximum output image size is secured. frames) with
detailed set-
tings
2
3 Touch the "Print Format" button.
Print format select dialogue will be displayed.
3
4 Touch [AA] button to output in "1 image to 2
frames" format.
Close the print format select dialogue, and
4 return to the "Image Confirm" screen.
The display of [Format] button changes from
"A" to "AA".
< 53 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Touch the [CANCEL] button when changes in output method (100%/Whole Image), trim- p.59 2 on 1 Output
(1 image to 2
ming layout or format are required. frames) with
"Properties" screen will be displayed. detailed set-
tings
Touching the [Output] button after the change will output the image.
Touching the [Back] button will output the image and return to the "Image Confirm" screen.
"Properties' screen is
displayed. Use this screen to
change the output method,
trimming layout.
< 54 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
When "2 on 1" output is selected with the [Format] button in the "Image Confirm]
screen, the output orientation is automatically selected so that the maximum output p.61 2 on 1 Output
image size is made possible. (2 image to 2
frames) with
detailed set-
1 Scan 2 images following the normal procedure. tings
6
< 55 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Touch the [CANCEL] button when changes in output method (100%/Whole Image), trim- p.61 2 on 1 Output
(2 image to 2
ming layout or format are required. frames) with
"Properties" screen will be displayed. detailed set-
tings
Touching the [Output] button after the change will output the image.
Touching the [Back] button will output the image and return to the "Image Confirm" screen.
"Properties' screen is
displayed. Use this screen to
change the output method,
trimming layout. p.109 Properties
Screen
Terminating the exam after clicking the [OK] button of the Exam Order that is select-
ed for 2 on 1(2 images to 2 frames on 1 film), without pairing the images, will output
the film with only one image on one side.
< 56 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
When outputting the "4 on 1" format using the [Format] button in the "Image
Confirm" screen, the way images are aligned follows the order of image confirma-
p.63 Output of 4
tion. Use the "Properties" screen to change the alignment and to output. on 1 (4 image
to 4 frames)
with detailed
< 57 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Touch the [CANCEL] button when changes in output method (100%/Whole Image), trim- p.63 Output of 4
on 1 (4 image
ming layout or format are required. to 4 frames)
"Properties" screen will be displayed. with detailed
settings
Touching the [Output] button after the change will output the image.
Touching the [Back] button will output the image and return to the "Image Confirm" screen.
"Properties' screen is
displayed. Use this screen to
change the output method,
trimming layout.
Terminating the examination in the 4 on 1(4 images to 4 frames on 1 film) format with-
out specifying all 4 images will output the film in 4 frame format with blank frame.
< 58 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
2
3 Touch the image area of the body part that shall be
output in 1 on 2 format.
The frame of the image display field changes to
3 green.
< 59 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
6 Select
A
either [AA] (2 frames horizontally aligned
or [ A ] (2 frames vertically aligned) using the
[Print Format] button.
The image displayed in the output preview field
changes according to the selection.
6
7 Select the film size, orientation and output
method.
The image displayed in the output preview field
changes according to the selection.
8
Depending on the set up at the time of installation, ei- : No flip
ther of the icons shown in the right may be printed on
the edge of output film(image data when output to the : Flip
Host), indicating the image is flipped or not.
Changing the image to be displayed in the output preview field. (changing the
image for the output with detailed settings)
Procedures to change the image that you want to output with detailed settings are
described below.
1. Touch the image area of the body part
with which the current image should
be replaced.
The frame of the image changes to
green.
1
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e
@@h? @@@@@@@@
@@
@@h?
@@h? @@
@@h? @@
@@
@@h?
@@h? @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@
@@
@@ @@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@g ?@@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g
@@@@@@@@ ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@
?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
< 60 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
2
3 Select
A
either [AB] (2 frames horizontally aligned
or [ B ] (2 frames vertically aligned) using the
[Print Format] button.
The image displayed in the output preview field
changes to "2 on 1" format.
3
4 Touch the first image for the 2 on 1 print.
The frame of the image display area changes to
green.
4
5
6 Touch the second image for the 2 on 1 print.
The frame of the image display area changes to
green.
6
< 61 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
7
8 Select the film size, orientation and output
method.
The image displayed in the output preview field
changes according to these selections.
The image displayed in the output preview field can be overwritten, deleted or
moved. p.60 Changing the
image to be
displayed in
the output
preview field.
Depending on the set up at the time of installation, ei- : No flip
ther of the icons shown in the right may be printed on
the edge of output film(image data when output to the : Flip
Host), indicating the image is flipped or not.
< 62 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
2
3 Select [ABCD] using the [Print Format] button.
The image displayed in the output preview field
changes to "4 on 1" format.
3
4 Touch the first image for the 4 on 1 print.
The frame of the image display area changes to
green.
4
< 63 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
11
12 Select the film size, orientation and output
method.
The image displayed in the output preview field
changes according to these selections.
12
13
Depending on the set up at the time of installation, ei-
: No flip
ther of the icons shown in the right may be printed on
the edge of output film(image data when output to the
: Flip
Host), indicating the image is flipped or not.
< 64 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Changing the image to be displayed in the output preview field. (changing the
image for the output with detailed settings)
Procedures to change the image that you want to output with detailed settings are
described below. (Same operation as for the "4 images to 4 frames" that is taken as
an example below can be applied to the "2 images to 2 frames")
Overwrite
Allocates another image in the place where an image is already allocated.
(In the example below, an AP Cervical Spine image is allocated in "A" position.
@@@@@@@@e?
@@@@@@@@e?
@@h? @@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@@@@@@@e?@@@@@@@@?e
@@@@@@@@
@@h? @@
@@
@@h?
@@h? @@
@@h? @@
@@
@@h? @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@
?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
Delete
Deletes an image that is already allocated in the output preview field.
(In the example below, an image allocated in "A" position is deleted.)
@@@@@@@@e?
@@@@@@@@e?
@@h? @@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@@@@@@@e?@@@@@@@@?e
@@@@@@@@
@@h? @@
@@h? @@
@@h? @@
@@
@@h?
@@h? @@
@@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@g ?@@
@@g ?@@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@
?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
- to be continued -
< 65 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
@@@@@@@@e?
@@@@@@@@e?
@@h? @@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@@@@@@@e?@@@@@@@@?e
@@@@@@@@
@@h? @@
@@
@@h?
@@h? @@
@@h? @@
@@
@@h? @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@ @@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g
@@g ?@@
?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
< 66 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
1
2 Touch the No Good [NG] button.
A confirmation message will be displayed.
2
3 Touch the [YES] button.
The image for the selected Exam Order will be
erased.
Text strings displayed on the buttons can be edited using the User Tool.
< 67 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Changes on Rejected Reason and whether or not reexamination was done can be
implemented on the Image Reference screen even after completing the examination.
When the Image Confirm is enabled, touch the [NG] button before the remaining
time shown on the [OK] button reaches "0".
With the CS-2 that is set to "Screen Menu" selection or Barcode registration, the
Exam Order for which the image is erased will reset the system to the status before
the image was scanned, and requires reexposure.
Because the barcode of the cassette, which was once scanned and read in is also
erased, touch the exam tag or read the barcode of the cassette again.
With the CS-2 that is set to "Auto-Registration", Exam Order is also erased when the
image is erased.
< 68 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Examination Termination
After completing the scanning for all Exam Orders, touch the [Complete] button.
However, when the initial screen is set to "Exam Registration" or "Exam List" screen, it is possible to suspend the
examination and to start another examination according to the exam order sent from the upstream devices or restart
the examination for suspended Exam Orders.
When Some Images have not been Read ("Image Confirm" screen for initial screen)
The following message is displayed.
[YES] Touching the [YES] button will result in
deletion of orders for which images
have not been read and terminate
examination.
1 3
4
When "Exam List" screen is set as the initial screen
Select the Exam Order marked with "Suspend", and touch the [Perform] button.
< 70 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Browsing the "Exam History" (outputting the Exam Order in the "Exam History")
Procedures to browse the images that have been output, and to re output these images are described below.
6
7 Touch the [Output Device] button for the device
to which the re output is to abe made.
8 Touch the [Scroll] button, and display the image p.133 Viewer/Server
Status Check
Screen
for re output.
p.135 Printer Status
Check Screen
8
< 71 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
10
To save the image data on a floppy disk, always select the floppy disk drive (Drive
"A") in the "Browse for Folder" screen.
When the "Image Confirm" screen is set as an initial screen, it is not possible to
change it to "Exam Registration" or "Exam List" screen.
p.12 Three Initial
Screens
1 Display the initial screen currently set.
[KONICA MINOLTA] button will be displayed at
the lower left of the screen.
2
3 Touch the [Screen Type] button.
Upon touching, the initial screen type changes
one after the other.
displayed,
@@h?
@@h? @@
@@h? @@
@@
@@h?
@@h? @@
@@
@@
@@
@@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@
@@
@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@g
@@g ?@@
@@@@@@@@ ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
?@@@@@@@@
?@@@@@@@@
< 73 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Several filtering items such as "Hospital Dept", "Location (Ward)" and "Exam Room"
can be selected at a time.
Touching the [Clear] button located at the lower left of the "Filter Condition" screen
will cancel all filtering conditions set by "Hospital Dept", "Location (Ward)" and
"Exam Room". While touching the already-selected button (green) will cancel the fil-
tering condition allocated to that button.
< 74 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Image Sharing
Using this function, images processed on the other console can be browsed from your console and can be reprinted.
Procedures to browse the images processed on the other console and to reprint these images are described below.
It may take a little longer before the Exam History screen of the target Console
appears depending on the data size of the image stored in the target Console.
The images that can be browsed using the Image Sharing function are only those
available at the time when the connection to the target Console is made. (step 5)
Changes brought during browsing will not be updated. To update the browsing
images, close the browsing screen first, then make a connection again to the target
console. A message Importing exam data. Please wait .... will be displayed during
update process.
< 75 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
8 9
9 Touch the [Properties] to make a reprint.
"Properties" screen will be displayed.
Output devices set on
the Output Setup
screen are those set on
10 Select the [Device Select] button of the output your own Console.
device (output devices set to your own Console) Actual output is also
from which the image shall be output. performed from your
own Console.
10
11 Touch the [Scroll] button, and display the image for Multiple output devices
can be selected at a
re output.
time.
11
13
< 76 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
1
2 Touch the button in the local column (name of
your own CS-2 is displayed).
2
Frame of the touched button changes to green.
< 77 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Maximum number of characters that can be used for annotation is 128 (full size) or
256 (half size) characters.
Rotation
90 Rotate
Flip
Flip
< 78 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
< 79 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
Once the text string is registered on the preset button, it will be retained even after
turning the power OFF.
< 80 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
< 81 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
< 82 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
When the text strings are selected, they change the color to Black when the background is
black, to White when the background is white, to Blue when the background is transparent.
< 83 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
The text strings of the rotated annotation cannot be changed. Text can be changed
only after restoring the original status where the annotation is not rotated. Font size,
font color and background color can be changed even in the rotated status.
< 84 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
3 Hold down the left mouse button and drag the font
size slider.
Dragging the slider increases or decreases the font
3 size.
< 85 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
When the text strings are selected, they change the color to Black when the background is
black, to White when the background is white, to Blue when the background is transparent.
< 86 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
< 87 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.4 Sequence of Examination
< 88 >
CS-2 Operation Manual Ver.1.10 2005.2
Chap.
Description of
5
Screens
This chapter describes each of the CS-2 screens.
< 89 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.5 Description of Screens
Depending on the
setting, there may be
cases where a new
patient ID is
automatically created
Patient Info Display when the screen is
Items of Patient Info are displayed. displayed.
To input data into an item, touch the input window at
the right of that item.
The determination of
which Patient Info is to
be used should be
made depending on the
facility when the system
is installed.
Touching the [ENTER] button will move the cursor to next item.
< 90 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.5 Description of Screens
Patient Info that is input on the "Exam Registration" screen will be registered in the
Patient Database when the examination starts or the screen is switched to the other.
However, when touching the [List] button to open the "Exam List" screen, a dialogue
shown below will appear, and selecting "NO" on this dialogue will not register the
Patient Info.
< 91 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.5 Description of Screens
Exam List Screen (when "Exam List" screen is set as the initial screen)
The screen displays a list of Patient Info input in the "Exam Registration" screen.
Selects the Patient Info that is reserved for which the examination is implemented. p.70 Restarting
Patient data already reserved may be deleted. the suspend-
ed examina-
(Suspended Patient Info cannot be deleted) tion
< 92 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.5 Description of Screens
Downward Arrow : Sorts the listing in descending order by the title item as a key.
Upward Arrow : Sorts the listing in ascending order by the title item as a key.
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@@@@@@@e?@@@@@@@@?e @@@@@@@@e?@@@@@@@@e?@@@@@@@@
@@h? @@@@@@@@
@@
@@@@@@@@e?
@@h? @@@@@@@@e? @@@@@@@@
@@
@@h?
@@h? @@ @@h?
@@h? @@
@@h? @@ @@h? @@
@@h? @@
@@ @@h? @@
@@
@@h? @@ @@h? @@
@@ @@ @@ @@
@@ @@ @@ @@
@@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@
@@ @@ @@
@@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@
@@ @@ @@ @@
@@
@@ @@
@@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@
@@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@
@@
@@ @@ @@ @@
@@ @@
@@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@
@@g ?@@ @@ @@
@@g ?@@ @@ @@
@@g ?@@ @@ @@
@@g ?@@ @@
@@ @@
@@g ?@@ @@ @@
@@g ?@@ @@ @@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@ @@ @@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@ @@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@g
@@@@@@@@ ?@@@@@@@@ ?@@
@@@@@@@@ ?@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
Page Number
Indicates the page number of the registered Exam List
pages.
[Filter] Button
p.123 Filter Key
Used to set parameters for extraction of data from the Screen
"Exam List".
The Filter Key screen is displayed.
[Delete] Button
Deletes the currently selected Patient Info from the
"Exam List".
[OK] Button
Returns to the "Exam Registration" screen.
Patient Info selected in the "Exam List" screen will be
displayed on the "Exam Registration" screen.
< 93 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.5 Description of Screens
< 94 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.5 Description of Screens
"Exam Regist" or
"Exam List" screen as
an initial screen
Exam Regist Image Confirm
Screen Screen
@@ @@
@@ @@
@@ @@
@@
@@
@@ @@
@@ @@
@@ @@
@@ @@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@g
@@@@@@@@ ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@
?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
Skull
Chest
Neck
Arm
Pelvis Abdomen
Others
Legs
Exam Tag
Exposing subject (body part) is classified as Exam Tag
according to the icon selected in the rough
classification.
The main Exam Tag contains the individual image
Exam Tags (Density, Contrast, etc.) that are best suited
for each body part.
Selecting one item (Exam Tag) for a specific
examination here will register a new order in the Exam
Order List.
< 96 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.5 Description of Screens
Image confirmed
Image has been checked and confirmed.
< 97 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.5 Description of Screens
Left of 2 Right of 2
Rotate/Flip
Indicates the rotation angle and upright or flipped image.
Exam Tag
< 98 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.5 Description of Screens
[Scroll] button
When there are more Exam Orders than what is
displayed, touch this button to scroll the list.
[Rotate/Flip] button
Rotates or flips the selected image (framed in green).
"Image Rotate/Flip" dialogue will be displayed. p.103 image
"Rotate/Flip"
Dialogue
Masking
The purpose of this type of image processing is to black out areas outside the field
of X-ray exposure on the scanned image. This prevents undue light leakage and
eye fatigue during reading of printed or soft copy images.
OFF ON
< 99 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.5 Description of Screens
[Complete] button
Terminates the examination, and returns to the initial
screen.
When there are examination not completed or not
confirmed ([OK] button not touched), a dialogue to
confirm termination will be displayed.
Please note the limitation for the stitched images (composite image).
Available : "Original", "Flip", "180 Rotate", "180 Rotate+Flip"
Not Available : "90 Rotate", "90 Rotate+Flip", "270 Rotate", "270 Rotate+Flip"
Detailed Data
Annotation Setup
Setting Screen
Screen
(Overlay)
[Back] Button
Sets the color of the background of the text used for
the annotation.
Select from White, Black and Transparent.
When the text of the annotation is selected, it changes the color to Black when the background
is black, to White when the background is white, to Blue when the background is transparent.
The text strings of the rotated annotation cannot be changed. Text can be changed
only after restoring the original status where the annotation is not rotated.
[Delete] Button
p.87 Deleting one
Deletes one annotation that has been selected. selected
annotation
[CANCEL] Button
Cancels the annotation settings.
Returns to the Detailed Data Setting screen.
[OK] Button
Saves the annotation settings.
Returns to the Detailed Data Setting screen.
Changing the format using this dialogue will automatically select the film size and orien-
tation so that the objective images are printed with their maximum possible sizes.
Print Format
Image Confirm
Select
Screen
Dialogue
[2-Image] button
[AB] Outputs two images in two horizontally or vertically.
[4-Image] button
[ABCD] Outputs four images, two on the top in line and other two on the bottom in line.
Touching the button will change the format, and closes the "Format Select" dialogue.
Use the "Properties" screen to output the "1 image to 2 frames", "2 images to 2 frames" p.109 Properties
or "4 images to 4 frames" with different film size, output method (100%/whole). Screen
Changes in "Format Select" setting are sustained even after termination of the
examination or system restart.
Whether this function should be enabled or not will be set by the service engineer
considering the operation at the user.
KVP
Input the voltage applied to the X-ray bulb.
Exposure
Input the exposure dose.
Ten key
Input the value for KVP, Exposure, Image Area Dose.
[BS] Key
Deletes one character that has been input.
[Cancel] Button
Close the dialogue without saving the input data.
[OK] Button
Close the dialogue after saving the input data.
Properties Screen
A screen to manually set the image size, format and output method, and to output the image.
[Scroll] button
Touching this button will scroll the list by
4 images to locate the image for the other
Exam Orders.
[Device] button
Selects the device to which the image is output.
Touching the [Device] button will display the detailed
setting dialogue of each device.
Printer
Performs the detailed settings for the output to the printer.
When the image stitching (composite) is selected and the magnification is set to
100%, the first image for tile printing will be displayed in the output preview field.
[Size] button
Selects the film size that will be printed on a
printer.
Displays the film sizes that are available on the
printer.
Selecting the size will change the button
color from light blue to dark blue.
[Orientation] button
Selects the orientation of the film printed on a
printer.
Upon selection, the button changes the color
from light blue to dark blue.
Upon selection, the orientation of the image
displayed on the Output Preview field changes.
[Output] button
Prints the image according to the setting made on the
"Properties" screen.
Touching this button will display the print queue
icon on the image display field, and starts printing.
[Output] button is disabled when the image to be printed is not displayed in the
Output Preview field.
Viewer/Server
Outputs the image to the viewer/server.
[Output] button
Outputs the image to the viewer/server.
Touching this button will display the viewer/server
icon on the image display field, and starts output.
Floppy Disk
Outputs the irreversible compression data of the image to the floppy disk drive of the
CS-2.
Annotation and name (patient info) that have been set will not be reflected on the data
exported to the floppy disk.
[Magnification] button
25% : Decrease the image size to 25%.
50% : Decrease the image size to 50%.
75% : Decrease the image size to 75%.
100% : Sets the print size to 1:1(100%).
Selecting the button will change its color
from light blue to dark blue.
[Quality] button
Selects the quality of the image that shall be
output to the floppy disk.
Low : Sets the quality to low level.
Std : Sets the quality to standard resolution.
High : Sets the quality to high level.
Selecting the button will change its color
from light blue to dark blue.
[Output] button
Outputs the image data to the floppy disk.
Touching this button will open the "Browse for
Folder" dialogue.
To save the image data on a floppy disk, always select the floppy disk drive (Drive
"A") in the "Browse for Folder" screen.
When outputting to a floppy disk, always select "31/2 Floppy disk" as the save target in
the "Browse for folders" dialogue.
The image size may exceed the capacity of floppy disk, resulting in save failure.
When a message "Failed to create JPEG image. Check the target disk capacity" is dis-
played, change the magnification (100% 75% 50% 25%) and/or image quali-
ty (HQ Std LQ).
Downward Arrow : Sorts the listing in descending order by the title item as a key.
Upward Arrow : Sorts the listing in ascending order by the title item as a key.
@@@@@@@@e?
@@@@@@@@e?
@@h? @@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@@@@@@@e?@@@@@@@@?e
@@@@@@@@
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@
@@@@@@@@e?
@@@@@@@@
@@h? @@
@@
@@h?
@@h? @@
@@
@@h?
@@h? @@ @@h?
@@h? @@
@@h? @@
@@ @@h? @@
@@
@@h? @@ @@h? @@
@@ @@ @@ @@
@@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@
@@ @@
@@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@
@@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@
@@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@
@@
@@ @@ @@ @@
@@ @@
@@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@
@@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@
@@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
Page Number
Indicates the page number of the registered Exam List
pages.
[Delete] button
Deletes the Patient Info currently selected in the "Exam
List".
When several items are input, search is made for the Patient Info that matches all inputs.
When displaying the new input screen for inputting a new Patient Info, it can be set so that
the new Patient ID is automatically issued.
For the system that searches from the upstream devices, the [Search] button
while it is searching from the upstream devices changes to "Suspend".
Touching the "Suspend" button will stop searching from the upstream
devices.
[CANCEL] button
Cancels the input for the Patient Info.
Returns to the "Exam List" screen.
[OK] button
Adds the input Patient Info to the "Exam List".
Returns to the "Exam List" screen.
Input Patient Info will be added to the "Exam List".
Patient Info
Exam Regist or Search Result
Input/Search
Screen Screen
Screen
Downward Arrow : Sorts the listing in descending order by the title item as a key.
Upward Arrow : Sorts the listing in ascending order by the title item as a key.
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e? @@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@h? @@@@@@@@
@@
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@h? @@
@@h?
@@h? @@ @@h? @@
@@h? @@ @@h? @@
@@h? @@ @@h?
@@h? @@
@@h? @@ @@h? @@
@@ @@
@@
@@ @@ @@
@@ @@
@@ @@
@@ @@ @@
@@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@
@@ @@ @@ @@
@@ @@ @@ @@
@@
@@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@
@@ @@ @@ @@
@@ @@ @@
@@
@@ @@ @@ @@
@@
@@ @@
@@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@
@@ @@ @@ @@
@@ @@ @@ @@
@@
@@ @@ @@ @@
@@ @@
@@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@
@@
@@ @@ @@ @@
@@ @@
@@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@
@@ @@ @@ @@
@@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@
@@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@
@@
@@ @@ @@ @@
@@ @@
@@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@
@@ @@ @@ @@
@@ @@
@@ @@ @@
@@ @@
@@ @@ @@ @@
@@ @@ @@ @@
@@
@@ @@ @@ @@
@@ @@ @@ @@
@@
@@ @@ @@ @@
@@ @@
@@ @@ @@
Scroll button
Scrolls the displayed page of the Patient Info.
[CANCEL] Button
Cancels the selection of the Patient Info, and closes
the "Search Result" screen.
[OK] button
Adds the selected Patient Info to the "Exam Regist"
screen, and closes the "Search Result" screen.
Selected Patient Info will be displayed in the "Exam
Regist" screen.
Modification of the Patient Info in the "Patient Info Modify" screen will be reflected al-
so in the Patient Info registered in the Patient Info database.
[CANCEL] button
Cancels the input for the Patient Info.
Returns to the "Exam List" screen.
[OK] button
Modification of the Patient Info is effected.
Returns to the "Exam List" screen.
[Location(Ward)] Button
Used to select "Location(Ward)" parameters.
Scroll button
Where there are 40 or more parameter selection but-
tons, the scroll button may be used.
[Clear] button
Filter Keys are cleared.
All categories ([Hospital department], [Location(Ward)] or
[Exam. Room] ) are cleared.
[CANCEL] button
Filter Key settings/changes are cancelled.
Display returns to the "Exam List" screen.
[OK] button
Sets Filter Keys.
Display returns to the "Exam List" screen.
The "Exam List" screen is modified to match the set Filter
Keys.
System Menu
Initial Screen
Screen
[Back] Button
Terminates the system menu screen.
Initial screen will be displayed.
[Screen Format] button will be displayed when the "Exam Regist" or "Exam
List" screen is set as the initial screen. Touching this button will switch
between the "Exam Regist" and "Exam List" screens.
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@
@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?
@@@@@@@@
@@h?
@@h? @@
@@h? @@
@@
@@h?
@@h? @@
@@h? @@
@@
@@ @@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@ @@
@@
@@ @@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@g ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@ ?@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@?e@@@@@@@@e?@@@@@@@@ ?@@@@@@@@
JM
JM connected to the CS-2 is displayed in icon format.
Black (normal) : Ready for operation.
Red (error) : Error has been occurred.
Touching this icon will display the status of JM and
the cassette registrations.
p.131 JM Status
Check
Screen
JM
This is an application software that controls the network between REGIUS
MODEL170 and CS-2.
System
Displays the status of the hard disk incor-
porated into the CS-2 System.
When disk capacity is
Guarded Images : Displays the number of
full, images on the
images guarded on the
"Examination History"
"Examination History"
screen are automatically
screen.
deleted from the oldest.
Disk Space (Used) : Used disk capacity is
A guard should be ap-
shown in giga-bytes
plied to selected exami-
(GB).
nations in the
Disk Space (Free) : Remaining disk capacity
"Examination History"
is shown in giga-bytes
screen to prevent eras-
(GB).
ing those images.
p.139 Examination
History
screen
[Back] Button
Terminates the "Device Status" screen and returns to
the original screen ("System Menu", "Image Confirm" or
"Image Browse" screen) p.125 System Menu
Screen
[OK] Button
Terminates the REGIUS MODEL 170/190 status check
screen.
[Primary] Button
Sets the JM for standard use.
[Secondary] Button
Sets the standard JM to backup JM.
This button is disabled when the standard JM is
operating without failure.
Scroll button
Scrolls the barcode list of the cassettes that are
registered on the CS-2.
[Stop] Button
Deletes read status data selected from the read status
data display.
Usually, the [Stop] but-
Read status data may be deleted only from a
ton is not used.
registered CS-2.
The button is used in
cases where the
[OK] button message "This image is
Terminates the JM status check screen. reserved for reading
and cannot be
browsed. The image
may be browsed after
completion of reading
or after disabling the
read operation" is dis-
played and movement
to another screen is not
possible.
< 132 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.5 Description of Screens
Output Reservation
Exam Info reserved for output after
completing the examination is listed.
Selecting the exam info will change its
color to light blue.
[Primary] Button
Sets the viewer/server to the one for standard use.
[Secondary] Button
Changes the output destination from standard
viewer/server to backup viewer/server.
[Scroll] button
Scrolls the displayed page for Exam Order list that is
reserved for examination.
[Delete] button
Deletes the output queue to the viewer/server.
This button is enabled (red) when the reserved Exam
Order is selected in the "Output Reservation" list.
[Output] button
Turns ON/OFF the output to the viewer/server.
Output Reservation
Exam Info reserved for output after
completing the examination is listed.
Selecting the exam info will change its color
to light blue.
[Primary] Button
Sets the printer to the one for standard use.
[Secondary] Button
Changes the output destination from standard printer to
backup printer.
[Scroll] button
Scrolls the displayed page for Exam Order list that is
reserved for examination.
[Delete] button
Deletes the output queue to the printer
This button is enabled (red) when the reserved Exam
Order is selected in the "Output Reservation" list.
[Output] button
Turns ON/OFF the output to the printer.
[Scroll] button
Scrolls the displayed page for Unsent Exam list.
[Primary] Button
Sets the selected External Patient DB/External Exam
DB for standard use.
[Secondary] Button
Sets the standard External Patient DB/External Exam
DB to backup External Patient DB/External Exam DB.
[Delete] button
Deletes the output queue to the External Patient
DB/External Exam DB. This button is enabled (red)
when the reserved Exam Order is selected in the
"Output Reservation" list.
[Output] button
Turns ON/OFF the output to the External Patient
DB/External Exam DB. This button will not be displayed
Touching the [OK] but-
in "External Patient DB/External Exam DB Status ton while the [Output]
Check (Input)" screen. button is kept "OFF" will
: Output "ON" (starts output) not send the info report-
ing that the Exam is
: Output "OFF" (suspends output) completed to the
External Patient DB/
External Exam DB, and
[OK] button the completed Exam
Closes the "External Patient DB/External Exam DB data may remain in the
Status Check" screen. list.
Therefore, do not turn
this button OFF unless
the External Patient
DB/External Exam DB
is in error.
Guard
Guard the image of
selected Patient Info to
protect it from deletion
from the hard disk.
Page Number
Indicates the page number of the registered Exam List
pages.
[Guard] Button
Used to guard the image of selected Patient Info and
protect it from deletion from the hard disk.
Images from Patient Info that are guarded are indicated by
display of " * " in the "guard" item on the "Examination
History" screen.
Touching the guarded Patient Info again removes the
guard.
Image Sharing function is an option for the CS-1. On the CS-1 that does not incor- p.75 Image
porate this function, Console Select screen will not be displayed. Sharing
Local Button
Touch this button to return to your Console that is
currently operated. Name of the Console will be
displayed on the button. Touching this button will select
your Console and changes the frame color to green.
Network Button
Selects the target Console that shall be browsed.
Consoles selectable for browsing are displayed on the
buttons. Touching this button will select the target
Console and changes the frame color to green.
[CANCEL] Button
Cancels the selection of the target Console.
Returns to the Exam History screen.
[OK] Button
Displays the Exam History screen of the selected
Console.
In cases where the Patient Info is insufficient or wrong, touch the p.121 Patient Info
[Modify Patient] button to modify the Patient Info. Modify
Screen
Modification of the Patient Info in the "Image Browse" screen will be reflected upon
the Patient Info registered in the Patient Info database.
Skull
Chest
Neck
Arm
Pelvis Abdomen
Others
Legs
Exam Tag
Processing conditions suitable to the body part are
classified as Exam Tag according to the icon selected
in the rough classification. The main Exam Tag
contains the individual image Exam Tags (Density,
Contrast, etc.) that are best suited for each body part.
Selecting one item (Exam Tag) for a specific
examination here will register a new order in the Exam
Order List.
Order Display
Exam Order that is registered with the Exam Tag can
be viewed on the button.
Select the Exam Order for the scanning before
examination. After scanning, select the [OK] or [NG] of
the image depending on the quality to output the image
(print or save). Touching any of these Exam Orders will
change its frame to green.
[Rotate/Flip] button
Rotates or flips the selected image (framed in green).
"Image Rotate/Flip" dialogue will be displayed.
[Masking] button
Selects to apply masking to the currently selected
image (framed in green).
Touching the button will switch between "ON" and
"OFF".
[ON] Applies "Masking" to the image.
Masking frame is displayed in yellow
around the X-ray exposed area.
Masking
The purpose of this type of image processing is to black out areas outside the field
of X-ray exposure on the image read. This prevents undue light leakage and eye
fatigue during reading of printed or soft copy images.
OFF ON
[Properties] button
Outputs the image to printer , viewer/server or floppy
disk. For the output to the printer, set details (image
size, orientation, format, output method) for each Exam
Order.
"Properties" screen will be displayed.
User Tool
6
How to operate the User Tool is detailed in this
chapter.
Before editing new settings using the User Tool, make sure to back up the current
data using the "Back Up" function of the User Tool. Doing so allows the user to
quickly restore the original setting even when any trouble happens to the CS-2 or
p.181 Backing Up
reader. the User Tool
Settings
System screen
Edits various settings of
general operation of the CS-2.
p.154
p.163
p.172
p.178
Back Up screen
Backs up the settings made by
the User Tool.
p.181
Restore screen
Restores the settings made by
the User Tool.
p.184
to System Menu
p.19
p.187 Operator
[Operator], [Log], [Statistics], [NG Reason Edit] buttons will be displayed in addition
Security
to those illustrated in the above using the Operator Security Mode or Statistics Mode
Analysis (Option) setting, and the screen corresponding to each button will be dis-
played by touching it. p.209 Optional
Functions
"System" Screen
A screen used to edit the general setting of the CS-2.
User Tool
System Screen
screen
Administrator Password - Sets the password that should be input to start up the User Tool.
Clicking the [Alter] button will display the "Admin Password" dialogue.
p.159 Operation of the Admin
Password Dialogue
Date/Time - Sets the internal clock of the CS-2. Clicking the [Alter] button will
display the "Date/Time Properties" dialogue.
p.160 Date/Time Properties
Dialogue
Facility
Name - Input the facility name. Maximum 60 characters can be input. Use the
keyboard to input.
Address - Input the facility address. Maximum 60 characters can be input. Use
the keyboard to input.
Image Confirm Manual Sets whether to use the Image Confirm. Select between "Manual"
and "Auto". When "Use" is selected, set the time (sec) before the
image is automatically confirmed.
Image Confirm
Selecting "Auto" in this function allows the image to be automatically
output when a specific time elapses after scanning the image, assum-
ing that the image has been confirmed. When re-exposure is neces-
sary, touch [NG] button before the time is reached.
Remaining time (sec) before the image is confirmed is indicated on the
[OK] button.
Changing the setting of density, contrast, rotate/flip and masking
before the time is reached will cancel the image confirm process. It is
required to touch the [OK] button to restart the process.
Count down is cancelled if any error message is shown before the time
is reached. Closing the message box will restart the count down.
Output Preview
1 Frame Yes Sets whether to display the "Output Preview" screen when the image
is confirmed on the "Image Confirm" screen. (in the case of 1on 1 or
same 2 on 1 format)
2 or 4 Frames Yes Sets whether to display the "Output Preview" screen when the image
is confirmed on the "Image Confirm" screen. (in the case of 2 on 1 or
4 on 1 format)
Auto-Output
Printer Yes Sets whether to automatically send the image data to a printer when
the image is confirmed on the "Image Confirm" screen. Selecting
"No" will not allow the image to be sent to a printer even when the
[OK] button is touched. In this case, it is required to manually output
the image on the "Properties" or "Exam History" screen.
Viewer/Server 1 Yes Sets whether to automatically send the image data to a Viewer/
Server 1 when the image is confirmed on the "Image Confirm"
screen. Selecting "No" requires to manually output the image on the
"Properties" or "Exam History" screen.
Viewer/Server 2 Yes Sets whether to automatically send the image data to a Viewer/
Server 2 when the image is confirmed on the "Image Confirm"
screen. Selecting "No" requires to manually output the image on the
"Properties" or "Exam History" screen.
Stamp Format OFF Sets whether to add the Patient Info (name) at the edge of the output
film. The style of this Patient Info should be selected from the
following 5 preset styles beforehand at the time of installation.
(However, the data output to a floppy disk will not be accompanied by
the stamp regardless of the setting (ON/OFF) ) Selecting "OFF" will
not print the Patient Info. 5 default styles are as follows. (Styles may
be different when they have been changed according to the request
from the system administrator at the facility)
Style-1 Style-2
Style-3 Style-4
Style-5
Stamp Position Upper Left When other than "OFF" is selected in the "Stamp Format", specify the
print position of the name from "Upper Left", "Upper right", "Lower
Left", "Lower Right".
Upp-L Name Name
Upp-R
Name Name
Low-L Low-R
Border Transparent Sets the background of the output film. Select from "Transparent" and "Black".
Black : Outputs the image with black background.
Transparent : Outputs the image with transparent background.
Black Transparent
[CANCEL] button
Closes the "System" screen without saving the altered
settings.
A message confirming to exit without saving the
settings will be displayed. Touching the [YES] will
return to the "User Tool" screen, while [NO] returns
to the "System" screen.
[OK] button
Closes the "System" screen after saving the altered
settings.
A message confirming to exit after saving the
settings will be displayed. Touching the [YES] will
return to the "User Tool" screen, while [NO] returns
to the "System" screen.
[CANCEL] button
Closes the "System" screen without saving the new
password.
Returns to the "System" screen.
[OK] button
Closes the "System" screen after saving the new
password.
A message requesting a confirmation to save the
new password will be displayed. Touching the [YES]
button will return to the "System" screen.
Date/Time
System screen
Properties
[Date/Time] Tab
Sets the date and time.
Date
Sets the date of the CS-2.
Time
Sets the time of the CS-2.
Time Zone
Select the time zone in the list box at the top of the
screen.
2 Click the [s], [t] button or input the digits. When the
time, min or sec is altered, clock hands automati-
2 cally move according to the time setting.
[Copy] icon
Copies the Exam Tag selected in the Exam Tag list.
Pelvis Abdomen
Others
Legs
[CANCEL] button
Closes the "Exam Tag" screen without saving the
altered settings.
A message requesting a confirmation to exit without
saving the new settings will be displayed.
Selecting the [YES] button will return to the User
Tool screen.
Selecting the [NO] button will return to the "Exam
Tag" screen.
[OK] button
Closes the "Exam Tag" screen after saving the altered
settings.
A message requesting a confirmation to exit after
saving the new settings.
Touching the [YES] button will return to the User
Tool screen.
Touching the [NO] button will return to the "Exam
Tag" screen.
2
3 Click the [Scroll] button (for Exam Tag), and dis-
play the page from which the new Exam Tag
key shall be selected.
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@@@@@@@e?@@@@@@@@?e
@@h? @@@@@@@@
@@h? @@
@@
@@h?
@@h? @@
@@h? @@
@@
@@h? @@
@@
@@
@@
@@
@@
@@
@@
@@ @@
@@
@@
@@
@@
@@
@@g
@@
?@@
@@
@@g ?@@
@@g ?@@
@@g ?@@
@@g
@@g ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@ ?@@
?@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
The Exam Tag in the original list will not be deleted even when the [Exam Tag] key
is deleted.
Ex) Copy the "Leg/Foot General", and create "Leg/Foot General-1". Then change the
name of "Leg/Foot General-1" to "Tip of Foot". Using the "Leg/Foot General" for overall leg
examination and the "Tip of Foot" for tip of foot examination, the more they are used for
separate body part, the more the image processing condition suitable to that body part will
be automatically created with the self-learning function.
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@@@@@@@e?@@@@@@@@?e
@@h? @@@@@@@@
@@h? @@
@@
@@h?
@@h? @@
@@h? @@
@@h? @@
@@
@@
@@
@@
@@
@@
@@
@@
@@ @@
@@
@@
@@
@@
@@
@@g
@@
?@@
@@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
3 Change the name of Exam Tag when neces- p.169 Altering the
"Exam Tag"
name
sary.
"Rotate/Flip" can be altered or newly set for the Exam Tags that are originally pro-
vided by KONICA MINOLTA or copied.
2
3 Change the name of Exam Tag when neces- p.169 Altering the
"Exam Tag"
name
sary.
"Exam Tag" name can be altered for the Exam Tags that are originally provided by
KONICA MINOLTA or copied.
1
2 Input the new name using the keyboard.
Name of the "Exam Tag", identical to the one already registered will be rejected.
Exam Tags originally provided by KONICA MINOLTA cannot be deleted. Only those
copied and edited can be deleted.
@@@@@@@@e?
@@@@@@@@e?
@@h? @@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@@@@@@@e?@@@@@@@@?e
@@@@@@@@
@@h? @@
@@
@@h?
@@h? @@
@@h? @@
@@h? @@
@@
@@
@@
@@
@@
@@
@@
@@
@@ @@
@@
@@
@@
@@
@@
@@g
@@
?@@
@@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
Deleting the "Exam Tag" that has been registered for the "Exam Tag" key will also
delete the already registered "Exam Tag" key.
The Exam Tag keys for the composite image are the special ones, do not use them
as the exam condition in the normal examination.
Do not delete or edit the following Exam Tag keys for test/servicing work purposes.
- "S Value", "DR Phantom", "test1", "test2".
When there are more Patient Info's than a certain number, it may become impossi-
p.177 Filtering the
ble to display all Patient Info's in the list. If such happens, extract the Patient Info's Patient Info
to be displayed in the list using the filter function.
[CANCEL] button
Closes the "Patient Info List" screen without saving the
altered settings.
A message confirming to exit without saving the
settings will be displayed. Touching the [YES] will
return to the "User Tool" screen, while [NO] returns
to the "Patient Info List" screen.
[OK] button
Closes the "Patient Info List" screen after saving the
altered settings.
A message confirming to exit after saving the
settings will be displayed. Touching the [YES] will
return to the "User Tool" screen, while [NO] returns
to the "Patient Info List" screen.
Inputting a new Patient Info in the "Exam Registration" screen and starting an exam-
ination can also add the Patient Info to the list.
1
2 Input the new Patient Info using the keyboard. Initial of the [Sex] se-
lection.
M : Male
F : Female
O : Others
Modifying the Patient Info searched from the "Exam Registration" screen and start-
ing the examination will also modify the Patient Info.
Modification of the Patient Info stored in the Patient Info database will not be reflect-
ed upon the Patient Info of the Exam History.
There are two ways available as follows to delete several Patient Info's at a time.
2
< 175 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.6 User Tool
To display all Patient Info's, click the [OK] button without inputting any parameter.
[Location(Ward)] Button
Used to select "Location(Ward)" parameters.
Scroll button
Scrolls the Filter Key List.
[CANCEL] button
Closes the "Filter Key List" without saving the altered
settings.
A message confirming to exit without saving the
settings will be displayed. Touching the [YES] will
return to the "User Tool" screen, while [NO] returns
to the "Filter Condition List" screen.
[OK] button
Closes the "Filter Key List" after saving the altered
settings.
A message confirming to exit after saving the
settings will be displayed. Touching the [YES] will
return to the "User Tool" screen, while [NO] returns
to the "Filter Condition List" screen.
3
@@@@@@@@e?
@@@@@@@@e?
@@h? @@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@@@@@@@e?@@@@@@@@?e
@@@@@@@@
@@h? @@
@@h? @@
@@
@@h? @@
@@h?
@@h? @@
@@
@@
@@
@@
@@
@@
@@ @@
@@
@@
@@ @@
@@
@@
@@
@@
@@
@@g ?@@
@@g ?@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g
@@@@@@@@ ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@
?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
?@@@@@@@@
3
@@@@@@@@e?
@@@@@@@@e?@@@@@@@@e?@@@@@@@@?e@@@@@@@@
@@h?@@@@@@@@e?@@@@@@@@?e@@@@@@@@ @@
@@h?
@@h? @@
@@h? @@
@@h? @@
@@
@@h? @@
@@
@@
@@
@@
@@ @@
@@
@@ @@
@@ @@
@@
@@
@@
@@
@@
@@g
@@g ?@@
@@g ?@@
?@@
@@g ?@@
@@g
@@g ?@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@ ?@@
?@@@@@@@@
@@@@@@@@ ?@@@@@@@@?e@@@@@@@@ ?@@@@@@@@
User Tool
Backup Screen
screen
[Go] button
Starts backup sequence.
Selecting any of the backup item enables this button to
be selected.
"Browse for Folder" screen will be displayed.
[Back] button
Closes the "Backup" screen.
A message requesting a confirmation to exit the
screen. Selecting the [YES] button will return to the
"User Tool" screen. Selecting the [NO] will return to
the "Backup" screen.
When the data backed up for the last time remains in the floppy disk that will be in-
serted, it will be overwritten. Care should be taken.
To save the data on a floppy disk, always select the floppy disk drive (Drive "A") in
the "Browse for Folder" screen.
It is also possible to export the data to the external PCs, however it is limited to the
PCs exhibited in "My Network".
If the data is too large and cannot be saved in a floppy disk, a message reporting
such will be displayed. Remove the floppy disk, and insert the next floppy disk. Take
notes of the order by which the data is saved in floppy disks.
User Tool
Restore Screen
screen
Backup Folder
Displays the directory of the folder that is
specified with the [Browse] button.
[Browse] button
Specify the folder in which the file to be
restored resides.
Upon specification of the folder, the
display in the "Backup Folder" changes.
[Go] button
Starts backup sequence.
Selecting any of the backup item enables this button to
be selected.
"Browse for Folder" screen will be displayed.
[Back] button
Closes the "Backup" screen.
A message requesting a confirmation of exiting the
screen. Selecting the [YES] button will return to the
"User Tool" screen. Selecting the [NO] will return to
the "Restore" screen.
2
2 Click the [Browse] button. the same order as it is
inserted for backup.
"Browse for Folder" button will be displayed.
4
5 Click the button for the item to be restored.
Selected button for the item changes its color
to green.
When two or more floppy disks are required for backup files, insert second, third, ...
floppy disks following the displayed message to restore all files.
8
9 Click the [Back] button.
A screen requesting a confirmation will be
displayed,
10
11 Display the screen for each item that has been
restored, and check that the settings are correctly
restored.
11
Super User
Technologist
Administrator
This user holds the highest authority to access the system.
A system administrator of the CS-2 is entitled to this authority, and capable of doing any
operation.
Super User
This user holds the authority next to the "Administrator", and capable of editing the User
Tool except for "Operator Management" and "Operation Log Management".
Technologist
This user holds no authority to implement any setting on the CS-2, but "Study" only.
Log In
When the operator security mode is in effect, it is necessary to input the user name and password when starting up
the CS-2 and restoring from the screen saver.
p.22 Start Up
Log In at start up
Log Off
While the operator security mode is in effect, implement log off when you are leaving the CS-2 or other user comes to
utilize your station or operate that CS-2.
There are two methods to implement log off.
In the case that the application is logged off in between start and completion of
study, and another user implements log in, a message requesting a confirmation
that the previous user has logged off before the study is completed appears.
Screen Saver
It is required to input the user name and password to cancel the screen saver.
User Tool
According to the user privilege with which the user logged in, functions of the User Tool will
be limited as follows.
When the login is made under the Technologist privilege.
[User Tool] button of the System menu will not be displayed.
On the CS-2 to which the optional Statistic Analysis is licensed, [Statistics Analysis] p.219 Statistics
Analysis
and [NG Reason Edit] buttons will be additionally displayed on the "User Tool"
screen.
Operator Management
Allows to control and manage the operators.
"Operator Management" screen will be displayed only when the login is made under
the Administrator privilege.
Operator List
Displays the name list of registered opera-
tors.
Touch the button either for registering a new
operator or for editing or deleting the current-
ly registered operator.
The frame of the selected button will
change to green.
[Trash] icon
Deletes the selected operator name.
[Scroll] button
Scrolls the operator list page.
[Cancel] button
Cancels the operator name editing, and exits the
operator management screen.
Clicking thus button will display a confirmation
dialogue.
When moving the item upward, it will replace the item just
below the green line which moves together with the cursor.
Frame changes to
green when clicked.
"Operator Log" screen will be displayed only when the login is made under the
Administrator privilege.
Even there are logs on the CS-2 when the screen is switched to the "Operator Log Management"
screen, these logs will not be displayed in the Log list. This Operator log will be displayed only
when the [List Apply] button is touched.
Log Size
Displays the file size of the Operator log.
[Export] button
Exports the Operator log.
[Trash] button
Deletes the selected operator log.
[Return] button
Closes the "Operator log Management" screen.
Return to the "User Tool" screen.
Even there are logs on the CS-2 when the screen is switched to the "Operator Log Management"
screen, these logs will not be displayed in the Log list. This Operator log will be displayed only
when the [List Apply] button is touched.
To save the data on a floppy disk, always select the floppy disk drive (Drive "A") in
the "Browse for Folder" screen.
It is also possible to export the data to the external PCs, however it is limited to the
PCs exhibited in "My Network".
If the data is too large and cannot be saved in a floppy disk, a message reporting such
will be displayed. Remove the floppy disk, and insert the next floppy disk. Take
notes of the order by which the data is saved in floppy disks.
Troubleshooting
8 Remedies for typical troubles and error messages
displayed on the CS-2 are described in this
chapter.
Trouble with JM
When a backup JM is available, switching the standard JM to the backup JM when an error occurs on the standard JM
will help minimize the downtime of the system. Switching the JM to the other can be managed by one of the CS-2s. In
this case, it is not necessary to restart the CS-2, but it is necessary to restart the REGIUS MODEL 170/190.
Following is the procedure how to switch from the standard JM on which an error has occurred to the backup JM. An
error occurred on the CS-2 disables the REGIUS MODEL 170/190. from performing an examination. Therefore,
switching to the other JM is essential when such trouble happens..
Switch over between standard and backup JMs should be implemented only under the
condition where all REGIUS 170/190s and consoles connected to the system have been
turned on.
When the switch over to the backup JM is completed, a following message will be
displayed on the other CS-2s connected to the same network.
Touching the [YES] to close the message window.
It may take approx. 1 min to switch over to the backup JM.
Examination by the REGIUS MODEL 170/190 is enabled
After switching the standard JM to the backup JM, report to your technical represen-
tative without delay.
If the screen does not change, keep pressing the power button till the power for the CS-
2 Control Unit is turned off.
If the phenomena that the CS-2 often rejects inputs becomes eminent, contact your
technical representative.
Reaction to the problem when the CS-2 is started up with a floppy disk inserted in the drive.
Attempting to start up the CS-2 with a floppy disk inserted in the floppy disk drive
(A:) will display a message "Remove disk or other media. Press any key to restart".
Remove the floppy disk, and press any key to start the CS-2.
Error List
Error messages displayed on the CS-2 and remedies to be practiced are described below.
Barcode XXXXXXXXXXXXXX for this cassette is illegal. Stop using the cassette with which the problem
Register other cassette. Resource error [X,X]. Restart the device. occurred, and contact your technical
representative.
Trouble with JM occurred. Switch to the backup JM if available, and contact
Call FSE. your technical representative.
When a backup JM is available, switch to the backup JM.
Failed to read ID[xxxx] Name[yyyy] [zzzz] Exam Time[YYYY:MM:DD] Follow the instruction in the REGIUS MODEL
Memory Error 170/190 Operation Manual.
Hardware Error
High Voltage ON Timeout
Laser ON Timeout
Mech Ready Timeout
Standby Timeout
Retransmission Count Overflow
CRC Error Count Overflow
Net Initialize Timeout
Net Ready Timeout
Put Shading Timeout
Status Check Timeout
Barcode Seek Timeout
Net Command Timeout
Transfer End Timeout
Retransmission Count Overflow
CRC Error Count Overflow
H sync timeout
Polygon Timeout
DMA Timeout
SCOM Timeout
Gain Overflow
Gain Underflow
Offset Overflow
Offset Underflow
QRV Overflow
QRV Underflow
Memory Error
Hardware Error
Job Error (Invalid Job)
Job Error (Invalid Mode)
Invalid Consoled
Job Error (delete)
Trans. Unit HP1 Sensor NG
Trans. Unit HP1 Sensor NG
Trans. Unit HP2 Sensor NG
Trans. Unit HP2 Sensor NG
Trans. Unit HP2 Sensor NG
Trans. Unit HP2 Sensor NG
Trans. Unit HP2 Sensor NG
Justification R/L Sensor NG(Reading)
Justification R/L Sensor NG(Absorption)
Justification R/L Sensor NG(Evacuation)
Justification R/L Sensor NG(Discharge)
Justification HP Sensor NG
Justification HP Sensor NG(Initializing)
Justification HP Sensor NG(Initializing)
Elevator Stand Sensor NG(Load) Follow the instruction in the REGIUS MODEL
Elevator Stand Sensor NG(Absorption) 170/190 Operation Manual.
Elevator Stand Sensor NG(Evacuation)
Elevator Stand Sensor NG(Discharge, before elevating)
Lock HP Sensor NG
Elevator Stand HP Sensor NG
Elevator Stand Sensor NG(Discharge, after elevating)
Insertion Slot Shutter Open Sensor NG
Insertion Slot Shutter Close Sensor NG
Discharge Slot Cover Close Sensor NG
Sub-scan HP Sensor NG
Back Plate Absorption Sensor NG
Back Plate Warp Sensor NG
V-SYNC Sensor NG
Back Plate Absorption Sensor NG
Back Plate Absorption Sensor NG
Back Plate Absorption Sensor NG
Back Plate Absorption Sensor NG
Back Plate Absorption Sensor NG
Back Plate Absorption Sensor NG
Insertion Sensor-M NG
Insertion M-Sensor-L NG
Delete Lamp Current Sensor-1 NG
Delete Lamp Current Sensor-2 NG
Front Door Open Sensor NG
Sub-Scan Motor NG
Transport Motor NG
Elevator Motor NG
Justification Motor NG
Insertion Slot Shutter Open/Close Timeout
Lock Timeout
Lock Initializing
Insertion Slot Shutter Initializing
Elevator Stand Initializing
Justification Initializing
Transport Initializing
Sub-Scan Initializing
Intake Ope-1 Start NG
Intake Ope-2 Start NG
Insertion Discharge Ope Start NG
Discharge Ope Start NG
Insertion Shutter Open NG
Insertion Shutter Close NG
Lock Start NG
Cassette Fall Sensor Time NG
Delete Lamp ON Command not reported
Delete Lamp OFF Command not reported
GetStatus Receive Timeout Initialize
GetStatus Receive Timeout Ready Follow the instruction in the REGIUS MODEL
GetStatus Receive Timeout Barcode Detected 170/190 Operation Manual.
GetStatus Receive Timeout High Voltage ON
GetStatus Receive Timeout Laser ON
GetStatus Receive Timeout Barcode Cancelled
GetStatus Receive Timeout Processing
GetStatus Receive Timeout Suspend
GetStatus Receive Timeout Front Door Open
GetStatus Receive Timeout Front Door Close
GetStatus Receive Timeout Error
GetStatus Receive Timeout Standby
GetStatus Receive Timeout Transport in Delete Mode
GetStatus Receive Timeout Power OFF after Ready
Registration OK Receive Timeout
GetStatus Receive Error Laser ON
GoBackward Receive Timeout
Laser ON without sending High Voltage (No.1)
Laser ON without sending High Voltage (No.12)
Illegal Command Received
Cassette Detected at Start up (Elevator Stand)
Cassette Detected at Start up (Justification)
Cassette Detected at Start up (Discharge Slot Open)
Cassette Detected at Start up (Back Plate Absorb)
Delete Lamp Dead(2 tubes)
SCB Comm Error
Cassette Warp Detected
Cassette Separation NG
Cassette Fall Detected
Front Door Open Detected
Default Shading Correction Coefficient Used
Default QRV Table Used
Default Shading Correction Coefficient/Default QRV Table Used
Polygon PLL Error Detected
Laser Error Detected
Barcode Detect Disabled
Detected Barcode Error
2 Cassette Inserted at a Time
Job Unregistered for Barcode Regist
2 Cassette Continuously Inserted
Cassette Reading in the Device Failed
Discharge Stacker Full
Discharge Slot Open
Delete Lamp Dead (1 tube)
Implementing the stitching process will abort examination conditions (image pro-
cessing, print format, overlay) that have been set before the stitching process.
Examination conditions for the composite image will be applied to the images once
treated in stitching process.
Stitching process cannot be applied to the images on the Consoles referred to from outside.
Order List
Orders before stitching and after stitching are listed in the
form of buttons.
Touching the order button will select itself and change its
frame to green.
One order will be added to the list for the image that has
been composited by stitching.
Maximum 9 orders will be displayed in the list. For more
than 10 orders, scroll the list using the scroll button.
[Add/Reset] Button
While the order is being selected, touch the [Add]
button (blue, right arrow) to register the selected order
(green frame) for composite image (stitching). Touch
the [Reset] button (purple, left arrow) to cancel the
registration of the selected order (green frame) from
registration for composite image (stitching).
When the order in the order list, created by stitching is being selected, [Add] button
(blue, right arrow) will not appear.
[Exit] Button
Returns to the Exam Check screen after terminating
the stitching process.
In the case where the screen is switched from the Exam Check screen to
the Composite Image Select screen, and joint borders of the composite
image that has been output are corrected using the [Rebuild] button, resulted
composite image will not be registered in the output queue even returning to
the Exam Check screen after completing modification. This happens
because the image is recognized as already printed. To output such order,
first exit or suspend the examination, then perform output using the Image
Reference screen invoked from the Exam History screen.
Stitching process is not completed only by registering the multiple orders for
the display of composite image. It will be stitched and become effective when
the [Build] button is touched and the Composite Image Adjust screen is
displayed.
[Rotate] Button
Touching this button will rotate the image. Lower image
in the Whole Stitched Images will be rotated. There is
a limit for rotation, and rotation along the arrow
direction beyond this limit is not possible.
: Rotates counterclockwise with the center positioned at the upper right corner of the lower image.
: Rotates clockwise with the center positioned at the upper right corner of the lower image.
: Rotates counterclockwise with the center positioned at the upper left corner of the lower image.
: Rotates clockwise with the center positioned at the upper left corner of the lower image.
[Move Button] ( )
Moves the image up/down or right/left.
Lower image in the Whole Stitched Images moves.
There is a limit (40mm) for move, and move along the
arrow direction beyond this limit is not possible.
[Reset] Button
Resets the image positioning, and restores the original
image position as the automatic stitching is completed.
[Order] Button
Changes overlapping order of the images.
Selecting the Normal simulates the overlapping of the
plates set in the long body part cassette. Selecting the
Reverse allows overlapping of the images in the
opposite way as the plates are overlapped.
Relation between the Plate Positions overlapped in the Long Body Part Cassette
and the [Order] button.
Mid Upper
Normal Upper
Image
Image Lower
Image
Image Lower
Image
[Scale] Button
Select whether to attach a scale to the long body part
image or not .
Zoom Slider
Changes the magnification of composite image by
dragging the slider to the right and left.
Dragging to the left decreases the zoom rate (smaller).
Dragging to the right increases the zoom rate (larger).
[CANCEL] Button
Cancels the composite image adjustment.
Returns to the Composite Image Select screen.
[OK] Button
Saves the setting made by composite image
adjustment and exits the screen.
Returns to the Composite Image Select screen.
[OK] Button
Output the image according to the setting made by
Output ON/OFF button.
[CANCEL] Button
Cancels the image output.
< 215 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.9 Optional Functions
Tiling at Border : Divides the image at the borders of the plates. In this way of
output, stitch borders tend to be less represented in comparison
to Tiling by Film Size.
(With this process, the image is always positioned in the center)
Tiling by Film Size : Divides the image to fit to the output film size. In this way of
output, film sheets to be output may be less than the sheets for
Tiling at Border.
On the CS-2 that is set to Tiling at Border, [Trim Adjust] button becomes invalid
when the divided image is smaller than the film size. [Trim Adjust] button becomes
valid when the divided image is larger than the film size.
On the CS-2 that is set to Tiling at Border, a message Unable to tile due to
insufficient film size. Switches to Tiling by Film Size will be displayed when
the longer side of the film (h2 in the diagram) is shorter than the distance
between the borders (h1 in the diagram), and the output is made by using
Tiling by Film Size.
Composit Image Film for Print
h2
h1
h1
h1
Stitching process is not completed only by registering the multiple orders for the
display of composite image. It will be stitched and become effective when the
[Build] button is touched and the Composite Image Adjust screen is displayed.
Selecting the order for the image that has been stitched enables the [Rebuild] but-
ton. Touching the [Rebuild] button displays the Composite Image Adjust screen,
enabling readjustment of the stitch borders of the images.
Detailed data (image processing, print format, overlay) set to each image becomes
invalid. Set the detailed data again to the orders after they are stitched.
Please note the limitation for the stitched images (composite image).
Only Original, Flip, 180 Rotate or 180 Rotate+Flip is selectable.
Statistics Analysis
This screen will be displayed by clicking the [Backup] button in the User Tool screen, followed by clicking the
[Statistics Analysis] button of the Statistics Analysis.
Statistics Analysis tool will be started up.
Because the database created for the Statistic Analysis tool is stored in both main
and sub JMs, the procedures should be carried out with both main and sub JMs
being active.
Statistic Menu
Select the statistic type that shall be displayed in the list.
[Reject/Repeat] : Indicates the ratio between the
rejected images and total exami-
nations.
[S-value] : Indicates the settings (S-value, G-
value) per body part and expo-
sure method at the examination.
[Console] : Indicates the examination count
per console.
[Plate] : Indicates the examination count
per plate.
[Reader] : Indicates the examination count
per reader.
Clicking the [List Apply] button after selection will display
the statistic data of the selected item in the list.
< 219 >
CS-2 Operation Manual Ver.1.10 2005.2
Ch.9 Optional Functions
Period
Select the period for which the statistic data shall be
displayed in the list. Selecting Not specify in the pull-
down menu for Year in the left column (column for
from) will select all through the past.
Clicking the [List Apply] button after selecting the Period will
display the statistic data for the selected period in the list.
[Export...] Button
Outputs the displayed statistic data in CSV file format.
The information exported by touching the [Export...] button is those listed in the
Statistic Data List. When the selection condition has been changed using a different
"Period" or "Technologist pull-down menu", it will not be reflected upon the export
result unless the Statistic Data List is updated using the [List Apply] button.
p.00 Displaying
1 Display the statistic data that you want to output. the Statistic
Data
To save the data on a floppy disk, always select the floppy disk drive (Drive "A") in
the "Browse for Folder" screen.
It is also possible to export the data to the external PCs, however it is limited to the
PCs exhibited in "My Network".
If the data is too large and cannot be saved in a floppy disk, a message reporting such
will be displayed. Remove the floppy disk, and insert the next floppy disk. Take
notes of the order by which the data is saved in floppy disks.
The statistic data that is exported will be saved under the file name (statistic data type + Date
& Time) as follows. (Example below is the file that is saved at 9:27:47, 2005. 1. 25)
"Reject/Repeat" data : RejectInfo20050125092747.csv
"S-value" data : S-valueInfo20050125092747.csv
"Console" data : ConsoleInfo20050125092747.csv
"Plate" data : PlateInfo20050125092747.csv
"Reader" data : ReaderInfo20050125092747.csv
Reject Reasons
Edit the reason for image abortion and whether the reexamination is done or not.
Available : Tick to select the items that shall be displayed on the buttons in the
Reject Reason screen as reject reason . Maximum 10 items can be
registered.
Logging for NG reason : Input the character string that shall be displayed on the button in the
Reject Reason screen. The text strings input here as the reject rea-
son will be recorded in the log as the reject reason. Maximum 64 char-
acters can be input.
[CANCEL] button
Close the window without saving the settings.
A confirmation message will be displayed.
1 2
3 2 Input the text string to be displayed as a button
in the Logging for NG Reason column next to
the Available check box.
The text string that shall be displayed as the button
in the Reject Reason screen will be registered.
LED
Beeper
Reading
Window
Trigger
LED
Displays the single-bar code reader status.
Reading Window
The laser beam is output from this window.
Beeper
A tone is generated when the single-bar code reader power is switched on.
Trigger
When using a handheld bar code reader, pulling the trigger will initiate the laser beam
from the reading window.
3 high beeps (Pu, Po, Pi) When the power is turned on or reset.
LED Meaning
Ergonomic Recommendations
Caution: In order to avoid or minimize the potential risk of ergonomic injury follow the rec-
ommendations below. Consult with your local Health & Safety Manager to ensure that you
are adhering to your companys safety programs to prevent employee injury.
Reduce or eliminate repetitive motion
Maintain a natural position
Reduce or eliminate excessive force
Keep objects that are used frequently within easy reach
Perform tasks at correct heights
Reduce or eliminate vibration
Reduce or eliminate direct pressure
Provide adjustable workstations
Provide adequate clearance
Provide a suitable working environment
Improve work procedures.
Regulatory Information
Radio Frequency Interference Requirements
This device has been tested and found to comply with the limits for a Class B digital
device pursuant to Part 15 of the Federal Communications Commissions Rules and
Regulation. These limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction manual, may cause harmful interfer-
ence to radio communications. Operation of this equipment in a residential area is likely to
cause harmful interference in which case the user will be required to correct the interfer-
ence at his own expense.
However, there is no guarantee that interference will not occur in a particular installation. If
the equipment does cause harmful interference to radio or television reception, which can
be determined by turning the equipment off and on, the user is encouraged to try to cor-
rect the interference by one or more of the following measures:
Re-orient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that which the receiver
is connected.
Consult the dealer or an experienced radio/TV technician for help.
This device complies with FCC Part 15. Operation is subject to the following two condi-
tions: (1) this device may not cause harmful interference and (2) this device must accept
any interference received, including interference that may cause undesired operation.
Applicable Directives
Electromagnetic Compatibility Directive 89/336/EEC
Low Voltage Directive 73/23/EEC
Applicable Standards
EN 55022:1998, Limits and Methods of Measurement of Radio Disturbance
Characteristics of Information Technology Equipment
EN 55024:1998; Information Technology equipment -Immunity characteristics - Limits
and methods of measurement
IEC 1000-4-2:1995; Electromagnetic compatibility(EMC); Part 4:Testing and measure-
ment techniques;Section 4.2:Electrostatic discharge immunity test
IEC 1000-4-3:1997; Electromagnetic Compatibility(EMC); Part 4:Testing and mea-
surement techniques;Section 3. Radiated, radio frequency, electromagnetic field
immunity test.
IEC 1000-4-4:1995; Electromagnetic compatibility(EMC); Part 4: Testing and mea-
surement techniques;Section 4:Testing electrical fast transient,/Burst immunity.
IEC1000-4-5:1995; Electromagnetic compatibility(EMC), Part 4: Testing and measure-
ment techniques;Section 5: Surge Immunity
IEC 1000-4-6:1996; Electromagnetic compatibility(EMC), Part 4:Testing and measure-
ment techniques;Section 6: Immunity to conducted disturbances, induced by radio fre-
quency fields.
IEC 1000-4-11:1994; Electromagnetic compatibility(EMC), Part 4: Testing and mea-
surement techniques;Section 11: Voltage Dips, Short Interruptions, and Voltage
Variations.
EN 60 950 + A1+A2+A3+A4+A11 - Safety of Information Technology Equipment
Including Electrical Business Equipment
EN 60 825-1 (EN 60 825) - Safety of Devices Containing Lasers
Laser Devices
Symbol products using lasers comply with US 21CFR1040.10, and IEC825-1:1993,
EN60825-1:1994+A11:1996. The laser classification is marked on one of the labels on the
product.
Caution: Use of controls, adjustments or performance of procedures other than those
specified herein may result in hazardous laser light exposure.
Class 2 laser scanners use a low power, visible light diode. As with any very bright light
source, such as the sun, the user should avoid staring directly into the light beam.
Momentary exposure to a Class 2 laser is not known to be harmful.
ENGLISH HEBREW
CLASS 2 LASER LIGHT
DO NOT STARE INTO BEAM
CLASS 2 LASER PRODUCT
DANISH / DANSK
KLASSE 2 LASERLYF ITALIAN / ITALIANO
SE IKKE IND I STREN CLASSE 2 LUCE LASER
KLASSE 2 LASERPRODUKT NON FISSARE IL RAGGIOPRODOTTO
AL LASER DI CLASSE 2
DUTCH / NEDERLANDS
KLASSE 2 LASERLICHT NORWEGIAN / NORSK
NIET IN STRAAL STAREN KLASSE 2 LASERLYS IKKE STIRR INN I LYSSTRLEN
KLASSE-2 LASERPRODUKT LASERPRODUKT, KLASSE 2
GERMAN / DEUTCH
KLASSE 2 LASERSTRAHLEN
NICHT DIREKT IN DEN LASERSTRAHL SCHAUEN
LASERPRODUKT DER KLASSE 2
SWEDISH / SVENSKA
KLASS 2 LASERLJUS STIRRA INTE MOT STREN
LASERPRODUKT KLASS 2
2 Direct the reading window of the single barcode The barcode may be
reader to the barcode label on the cassette and
read either from the top
depress the trigger.
or the bottom of the
Depressing the trigger will produce laser output to cassette.
read the barcode.
XXXXXXXXXXXxxxxx
XXXXXXXXXXXxxxxx
When using the barcode reader with a hands-free mode, the laser outputs continuously.
Be sure to avoid that the laser is radiated into eyes directly.
When reading the barcode, ensure that the laser scanning line crosses all the bars and When the laser beam
spaces. hits the barcode sur-
face at right angle, it
may fail to read proper-
ly. If failed, slightly
incline the barcode
reader, and try to read
012345 012345
again.
LED
Reading Window
Beeper
LED
Displays the multi-barcode reader status.
In the normal operational status, the green LED illuminates. The red LED illuminates to indi-
cate that the barcode has been correctly read.
Reading Window
The laser beam is output from this window.
Beeper
A tone is generated when the multi-barcode reader power is switched on.
LED Meaning
Notices
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to part 15 of the FCC rules. These limits are designed
to provide reasonable protection against harmful interference in a residential
installation. This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may
cause harmful interference to radio communications. However, there is no guar-
antee that interference will not occur in a particular installation. If this equipment
does cause harmful interference to radio or television reception, which can be
determined by turning the equipment off and on, the user is encouraged to try to
correct the interference by one or more of the following measures:
This device complies with part 15 of the FCC Rules. Operation is subject to the
following two conditions: (1) This device may not cause harmful interference,
and (2) this device must accept any interference received, including interference
that may cause undesired operation.
Notice
This Class B digital apparatus complies with Canadian ICES-003.
Remarque
Cet appareil numrique de la classe B est conforme la norme NMB-003 du
Canada.
Caution
Use of controls or adjustments or performance of procedures other than those
specified herein may result in hazardous laser light exposure. Under no circum-
stances should the customer attempt to service the laser scanner. Never
attempt to look at the laser beam, even if the scanner appears to be nonfunc-
tional. Never open the scanner in an attempt to look into the device. Doing so
could result in hazardous laser light exposure. The use of optical
instruments with the laser equipment will increase eye hazard.
Atencin
La modificacin de los procedimientos, o la utilizacin de controles o ajustes
distintos de los especificados aqu, pueden provocar una luz de lser peligrosa.
Bajo ninguna circunstancia el usuario deber realizar el mantenimiento del lser
del escner. Ni intentar mirar al haz del lser incluso cuando este no est oper-
ativo. Tampoco deber abrir el escner para examinar el aparato. El hacerlo
puede conllevar una exposicin peligrosa a la luz de lser.
El uso de instrumentos pticos con el equipo lser puede incrementar el riesgo
para la vista.
Attention
L'emploi de commandes, rglages ou procds autres que ceux dcrits ici peut
entraner de graves irradiations. Le client ne doit en aucun cas essayer d'en-
tretenir lui-mme le scanner ou le laser. Ne regardez jamais directement le
rayon laser, mme si vous croyez que le scanner est inactif. N'ouvrez jamais le
scanner pour regarder dans l'appareil. Ce faisant, vous vous exposez une ray-
onnement laser q st hazardous. L'emploi d'appareils optiques avec cet
quipement laser augmente le risque d'endommagement de la vision.
Achtung
Die Verwendung anderer als der hier beschriebenen Steuerungen,
Einstellungen oder Verfahren kann eine gefhrliche Laserstrahlung hervorrufen.
Der Kunde sollte unter keinen Umstnden versuchen, den Laser-Scanner selbst
zu warten. Sehen Sie niemals in den Laserstrahl, selbst wenn Sie glauben, da
der Scanner nicht aktiv ist. ffnen Sie niemals den Scanner, um in das Gert
hineinzusehen. Wenn Sie dies tun, knnen Sie sich einer gefhrlichen
Laserstrahlung aussetzen. Der Einsatz optischer Gerte mit dieser
Laserausrstung erhht das Risiko einer Sehschdigung.
Attenzione
L'utilizzo di sistemi di controllo, di regolazioni o di procedimenti diversi da quelli
descritti nel presente Manuale pu provocare delle esposizioni a raggi laser ris-
chiose. Il cliente non deve assolutamente tentare di riparare egli stesso lo scan-
ner laser. Non guardate mai il raggio laser, anche se credete che lo
scanner non sia attivo. Non aprite mai lo scanner per guardare dentro
l'apparecchio. Facendolo potete esporVi ad una esposizione laser rischiosa.
L'uso di apparecchi ottici, equipaggiati con raggi laser, aumenta il rischio di
danni alla vista.
The multi barcode reader is a CDRH Class II a laser device. The following
cautions must be observed during usage. Do not look into the reading win-
dow on the front of the device as the laser may cause damage to the
eyes.
Do not direct the laser at the eyes of others as this may cause damage to
the eyes.
barcode
Mouse Ball 2 Reverse the mouse and turn the mouse ball
cover counterclockwise to take out the mouse
ball.
Roller
5 Return the mouse ball in the mouse.
If the mouse ball or small roller is cleaned with a
dampened cloth or swab in the step 3 or 4, be
sure to dry the mouse ball sufficiently before
returning it in the mouse.
<Optical Type>
To ensure the function of the optical-type mouse, the built-in optical sensor needs to
correctly sense the moving direction and distance. Use the mouse in the low-dust
environment to protect the dust from sticking on the sensor. Periodically clean the
mouse with a dry cloth.
Maintenance by User
Frequency
Weekly Monthly Remarks
Maintenance Task
1) Checking the operation of
the cooling fan.
Specification
CS-2
Image Output
Host Normal 1ch, Backup 1ch
Printer Normal 1ch, Backup 1ch
DICOM Support Storage(CR/DX, SCU), Basic Grayscale Print Management SCU, Modality Worklist
Management SCU, Modality Performed Procedure Step SCU
Operating Condition
15 ~ 30:/40 ~ 80RH% (no condensation)
Transportation/
Storage Condition 40 ~ 60:/5 ~ 100RH% (no condensation)
Accessories
Control Unit power cable, Operation unit power cable, Control Unit/Operation Unit
connecting cable (several), Keyboard, Mouse x 1, Operation manual
Options
barcode reader for cassette registration (single/multi), Software License for additional
output to CS-2 HOST. Software licence for connection with CS-2 electrical medical
record system.
Option
Operating Temperature 0 ~ 50 :
Desk Stand
Interface RS232C
Resolution 0.13mm
For the optional devices other than those listed in the above, please refer to each
Operation Manual.
A M
Annotation.................................................................78 [Mag. Method] button ..............................................111
Auto-Registration ......................................................14 Masking ....................................................................46
C P
Complete ..................................................................27 Primary image...........................................................52
Condition keys ..........................................................15 Print Format ..............................................................52
Contrast Slide Bar.....................................................43 [Properties] button ..................................................102
CS-2 Controller Unit....................................................6
Q
CS-2 Operation Unit ...................................................6
[Quality] button........................................................113
D
R
DEL ...........................................................................90
[Rotate/Flip] button ...................................................99
Density Slide Bar ......................................................42
S
E
Single-Image Format ................................................13
Exam order display ...................................................97
Screen Menu Selection.............................................15
Exam Tag Key ..........................................................14
[Size] button ............................................................110
F Suspend....................................................................69
Four-Image Format ...................................................13 Single Barcode Reader...........................................228
U
I
Unexposed field ........................................................98
Image Confirm ..........................................................37
0502TE