Attaching Motorized stage to 90i

The motorized stage includes the stage and a mounting ring.

FRONT

Stage Mounting
Holes
Fig 1 Fig 2

Because the motors on the stage block access to the two mounting screws at
the right and left rear of the stage carrier you need to attach the mounting
ring to the stage carrier before you mount the stage.

Fig 3
When mounting the ring you need to make sure that it is centered (Fig 3) to
the carrier to insure that the condenser does not hit the stage when you
kohler the microscope. You will need to power up the stage and position the
stage surface as shown in Fig. 2 to access the two stage mounting holes.
Test mount the stage with the condenser raised to its highest position to
make sure the condenser does not hit the left or right side of the stage
surface.