In this section you will learn how to program BX 1 according to your application,
To enter the programming and calibration menu, press the <F> key followed by Info
We press. The display will show [FunC] and [PASSWr] respectively. Service
Program. To access the parameters that can be entered with the service password,
The calibration jumper must be installed (see section 2.2). the password
Only the parameters related to the use can be changed. With service mode
The parameters that can be changed are displayed in the flow field on the next page;
The calibration jumper requirements are indicated by black squares in the upper right
corners.
The Programming and Calibration menu displays the main blocks in the form [X--] and the
bottom
Build from blocks. And typing in the next yada to the previous main blog
accessible. When the main block you want to enter is selected and then press Enter,
The first sub-block within the block is accessed and the display [X0-] appears. And tuna
To reach the desired sub-blog and press Enter to select the sub-block
You can enter the first parameter. Indicating the first parameter of the selected sub-block
The message [XY0] appears. In the same quadrant and in the next yada
You can reach the previous parameter. To change the parameter values, press 0?
Use. To enter a numeric value, use the <T> button to select the digit you want to change,
While at any stage within the Programming and Calibration menu, press <F>
When pressed, a top block is reached from the end point (parameter bottom block).
When the <F> key is pressed while in the main blocks, the message [SAVE] is displayed.
The <F> key is pressed to ignore the changes made. In these three cases
The display will briefly display the message [WAIT] and the BX 1 returns to the weighing
menu.
With the INDFACE program you can change the settings of your device from your computer
and
HRKT DDK.
21 _
In this section only zero gain adjustments can be made. Thus, the entire calibration process
This parameter is only to reset the zero setting. Press <Enter> for
Is entered into the parameter. When the display [ZE.CO] message appears, the scale
The kefesi is emptied and the Enter key is pressed. Indication during zero setting [WAIT]
Message appears. In this process, there should not be a load on the weighing instrument,
It must be ensured that it is stable. The indication [SAVE] appears after the measurement.
With this parameter, the gain setting of the device is renewed. By pressing <Enter>
Displayed [XXXXXX] posts appear. This number is set for gain calibration
Is the required load value. Display weight is a different calibration weight than the displayed
value
, The weight of the calibrated calibration is adjusted with the help of the <Tare> and <Zero>
The desired weight value is entered and the weight is placed on the scale and press <Enter>
The gain setting made is loaded into the memory by pressing the <Enter> key. Calibration
In this step, without lifting the current load of a weighing instrument under load,
Is the gain adjustment made by loading the weight required for calibration. It
When entering this parameter, the device will first determine the current state,
The instrument should not be touched and the calibration weight should not be added.
Press <Enter> key in this parameter, the display will show [P.ZEro]. It,
BX 1 is the stage of determining the current state of the weighing instrument. Temporary
Zero setting
There shall be no additional load on the weighing instrument and the stability of the
weighing instrument
It must be provided. The device that detects the current state switches to gain calibration.
The display will briefly read [LoAd] and then add the calibration
[XXXXXX], which is the load value that should be maintained. Set if necessary
Calibration weight is entered with <Dara> and <Zero> keys, additional load on scale
The [SAVE] message appears on the BX 1 indicator, which completes the calibration setting.
The gain setting made is loaded into the memory by pressing the <Enter> key. Calibration
Caution: Before starting to eCal, the capacity and division from parameter [212]
It provides. Weighing instrument capacity, total load cell capacity, load cell output and
weighing
The tool dead load values are the values we need for this calibration.
Automatic zero instead of entering dead load value if conditions are appropriate
[LC.CAP]
[XXXXXX]
Example: If the weight system contains 4 1000 kg Load Cells, LC.CAP = 4000.
[LC.oUt]
[XXXXXX]
The output value (mV / V) specified in the load cell certificate can be set with the <Tare> and
<Zero>
Enter. If there are more than one load cell in the system, the average load cell output
Mean value for LC1: 2.0010, LC2: 1.9998, LC3: 1.9986 and LC4: 2.0002:
[Zer.adj]
[XXXXXX]
Press <Enter> key to perform zero calibration. If this step is empty and zero
[WAIT] the message came to mind - now that the weighing instrument is empty and stable
If zero calibration is not possible, press (Info) to enter the estimated dead load value.
[Pre-Ld]
[XXXXXX]
Enter the dead load value with the <Dara> and <Zero> keys and press <Enter>
Note: If you need to make a setting after the calibration is completed, either [310]
Set the parameter to zero; Or dead load value [read on empty balance