Académique Documents
Professionnel Documents
Culture Documents
Flexural wave suppression by an elastic metamaterial beam with zero bending stiffness
Journal of Applied Physics 121, 134902 (2017); 10.1063/1.4979686
Electrical properties of planar AlGaN/GaN Schottky diodes: Role of 2DEG and analysis of non-idealities
Journal of Applied Physics 121, 135701 (2017); 10.1063/1.4979530
FIG. 1. Composite motions of a transducer and a cylindrical structure: (a) circularly moving transducer and horizontally moving cylindrical structure; (b) hori-
zontally moving transducer and circularly moving cylindrical structure; (c) helically moving cylindrical structure.
of nine point scatterers is used to compare the performance only exist for one point. Therefore, the limitation of n is
of the HPSM algorithm with that of a DAS algorithm. In defined as follows:
Section IV, six rods and a letter pattern are measured using
an ultrasonic pulse-echo system and the measured data are 0 n < p: (2)
used to evaluate the proposed algorithm. In Section V, a dis-
In helical scanning, a transducer moves along the helix,
cussion of the results is presented and conclusions are
and the ultrasound beam is aimed towards the central axis. A
drawn.
3D helical scan contains a set of helical scans with different
offsets in their n axes or with transducer arrays distributed
II. THEORY along their n axes.
A. Helical coordinate system
B. Solution to wave equation in the helical coordinate
A cylindrical coordinate system constructed based on
system
r; h; z coordinates is shown in Fig. 2. Based on this cylin-
drical coordinate system, a nonorthogonal right-handed The wave equation is
helical coordinate system that is constructed based on
1 @2
r; /; n coordinates is established using the following r2 w 0; (3)
transformation: v2 @t2
8 where v is the ultrasonic wave velocity. In the helical coordi-
< r r;
> nate system, the Laplacian operator can be expressed as
j/j jhj; (1) follows:19
>
:
n z ph; 2 2 ! 2
1 @ @ 1 @ p @
r2 r 2 2 1
where p p=2p > 0, and p is the helical pitch. / and h are r @r @r r @/ 2pr @n2
equal numerically but differ in their physical meanings. In 2
p @
the correlation between n and /, nonunique representations 2 : (4)
2pr2 @/@n
Based on this helical Laplacian operator, the solution to
the helical wave equation is given as
pb
qn : (6)
2p
The relationship among kr , x, and b is
s
x 2
kr b2 : (7)
FIG. 2. Cylindrical and right-handed helical coordinate systems. v
134901-3 Jin et al. J. Appl. Phys. 121, 134901 (2017)
D. Imaging
The acoustic field Wq x; b; r in (10) is transformed
using a 3D inverse Fourier transform. The results at t 0 are
then used for imaging in the following form:
1
1
Ir; /; n F /;n Wq x; b; RWq R; r; v; x; bdx ;
1
(13) FIG. 3. Flowchart for the HPSM.
134901-4 Jin et al. J. Appl. Phys. 121, 134901 (2017)
Parameter Right-handed Left-handed Right-handed Left-handed The geometrical positions of the scatterer and the
transducer are set in the cylindrical coordinate system
p
(Fig. 5). The distance between the transducer and the scat-
b
terer is
8 p
>
> dis dz2 ds2
>
>
p and b are different in the solution to the wave equation >
< dz zs zd
when using the left-handed helical coordinate system. Thus, q (15)
>
> ds rs2 rd2 2rs rd cos b
the formula used to calculate q must also be altered. The >
>
>
:b h h ;
sign table is listed below. s d
According to Table I, the formula for q for an outgoing
diverging wave in the left-handed helical coordinate system where rs ; hs ; zs and rd ; hd ; zd represent the positions of the
is defined as scatterer and the transducer in the cylindrical coordinate sys-
tem, respectively.
pb
qn : (14) The angle a shown in Fig. 5 is calculated using the
2p cosine theorem as
When the transducer moves along the helix during
the helical scan, the physical scan trajectory is confirmed a arccos dis2 rd2 m2 =rd m ; (16)
using the helical coordinate system. However, the formula p
where m rs2 dz2 .
for q that is used in the HPSM algorithm cannot simply
Using the angular spectrum theorem, the back-scattered
be determined based on the physical scan trajectory
signal that is received by the transducers can be formulated
because it is also related to the arrangement of the image
in the cylindrical coordinate system as follows:
(Fig. 4).
When the directions of the n and / axes are drawn on 2 1
1 dis
the right-handed helical scan trajectory [Figs. 4(a) and 4(c)], srd ; hd ; zd ; t Sxei2x v D2 x; aeixt dx;
q can be calculated using (6). In Figs. 4(b) and 4(d), the for- dis
1
mula for q is replaced with that of (14). The opposite cases
(17)
occur on the left-handed helical scan trajectory. The positive
scan direction of / in the physical space is clockwise; there- where Sx is the Fourier transform of the Gaussian-
fore, the complex transfer function in the HPSM algorithm dis
modulated sinusoidal pulse, ei2x v is the phase factor, and
should be revised based on both the physical helical scan Dx; a denotes the directivity function, which is defined as
system and the image arrangement.
sina
Dx; a jinc xa : (18)
2v
FIG. 4. Formulas for q in right-handed helical scan system using (6) for (a)
and (c), and (14) for (b) and (d). FIG. 5. Geometrical positions of the scatterer and the transducer.
134901-5 Jin et al. J. Appl. Phys. 121, 134901 (2017)
FIG. 6. Simulation images. (a) Angle-R image and (b) n-angle image formed using the raw data; (c) angle-R image and (d) n-angle image reconstructed using
the DAS algorithm; (e) angle-R image and (f)n-angle image reconstructed using the HPSM algorithm.
n 2(0 mm, 2 mm) are weakened because of the zero padding and n-angle images, respectively. Although some pseudo-
used when n < 0 for the DAS computations. morphic images are present in the angleR image, the
The images that were reconstructed using the HPSM amplitudes of these pseudomorphic images are lower than
algorithm are shown in Figs. 9(e) and 9(f). The scatterers those of the scatterers and could easily be eliminated using
are focused as points and pencil-shaped rods in the angle-R a threshold filter. To quantitatively compare the resolution
FIG. 7. Experimental set-up. FIG. 8. Dimensions and distributions of the rod pairs.
134901-7 Jin et al. J. Appl. Phys. 121, 134901 (2017)
FIG. 9. Experimental images of the rods. (a) Angle-R image and (b) n-angle image using raw data; (c) angle-R image and (d) n-angle image reconstructed
using the DAS algorithm; (e) angle-R image and (f) n-angle image reconstructed using the HPSM algorithm.
of imaging methods, i.e., the DAS and HPSM, we recorded caliper are listed in the column labelled as Actual values
the maximum amplitudes along the Angle axis, as shown in in cylindrical coordinate system in Table III. The rod
Fig. 10(a). Then, the resolution is calculated based on full dimensions that were acquired from the ultrasound images
width half maximum (FWHM) and shown in Fig. 10(b). are listed in the column of Measured in helical coordinate
From the raw data, we can see that the resolution is increas- system. For ease of comparison, the data based on
ing with the wave propagation distance. After applying the helical coordinates were transformed into cylindrical
DAS, the resolution is improved to 68 pixels. The resolu- coordinate data. The equation used for the transformation
tion is further improved when the raw data are recon- from helical coordinates to cylindrical coordinates is
structed using HPSM and the resolution is 23 pixels, defined as
which is much better than that of the DAS method. Since 8
the angle pitch is 0.5 , the ultimate resolution of HPSM is < r r;
>
about 1 1.5 . jhj j/j; (22)
>
:
The rod lengths were also quantitatively measured z n p/=360 z0 :
based on the high-resolution images, and the results are as
presented in Table III. The geometries and positional Here, the helical pitch p is 108 mm and the starting off-
dimensions of the rods that were measured using a vernier set z0 is 39 mm in this experiment. Based on the results, the
134901-8 Jin et al. J. Appl. Phys. 121, 134901 (2017)
Measured in helical coordinate system Measured in cylindrical coordinate system Actual values in cylindrical coordinate system
Rod ra /a na ra ha za ra ha za
a
The units for r, n, and z are mm, and the units for / and h are degrees ( ).
134901-9 Jin et al. J. Appl. Phys. 121, 134901 (2017)
FIG. 12. Experimental images of letter pattern UT. (a) Angle-R image and (b) n-angle image using raw data; (c) angle-R image and (d) n-angle image recon-
structed using the DAS algorithm; (e) angle-R image and (f) n-angle image reconstructed using the HPSM algorithm.
diffraction. The proposed algorithm also demonstrates higher algorithm, including interior helical scanning methods and
resolution than the DAS algorithm. the detection of multilayer structures.
In most cases, because the test samples are generally
solids that are covered by a coupling layer for practical ACKNOWLEDGMENTS
ultrasonic NDT, the detection object becomes a multi-layer This work was supported by the National Natural
structure. The refraction that occurs at the boundary Science Foundation of China under Project Nos. 51675480
between the different materials makes the use of the time and 61401392.
domain SAFT more difficult. Fortunately, many PSM-type
algorithms have been shown to maintain their performance
APPENDIX: EXPLODING REFLECTOR MODEL (ERM)
levels for multilayer materials without any increase in the
computational burden. Because the basic principle of the As shown in Fig. 13(a), the ultrasound wave propagates
HPSM algorithm is similar to that of PSM, it therefore back and forth in the pulse-echo ultrasonic system. Because the
holds the potential to be applied to multilayer structures. round trip satisfies the reciprocity theorem, the exploding reflec-
Another point that must be researched in future is the scan tor model (ERM) is introduced to simplify the pulse-echo sys-
direction. In this paper, we mainly focused on exterior heli- tem and represent this system as a single upward wave scenario,
cal scanning. Interior helical scanning is also widely used which assumes that all the scatterers in the medium explode
in ultrasonic NDT and also in ultrasonic medicine. Our simultaneously and thus become upward-emitting acoustic sour-
future work will address the application of the HPSM ces, as shown in Fig. 13(b). The resulting ultrasonic waves are
134901-10 Jin et al. J. Appl. Phys. 121, 134901 (2017)
4
G. Matrone et al., The delay multiply and sum beamforming algorithm in
ultrasound B-mode medical imaging, IEEE Trans. Med. Imag. 34(4),
940949 (2014).
5
R. H. Stolt, Migration by Fourier transform, Geophysics 43(1), 2348
(1978).
6
J. Gazdag, Wave equation migration with the phase - shift method,
Geophysics 43, 13421351 (2012).
7
T. Olofsson, Phase shift migration for imaging layered objects and
objects immersed in water, IEEE Trans. Ultrason. Ferroelectr. Freq.
Control 57(11), 25222530 (2010).
8
K. Qin, C. Yang, and S. Feng, Generalized frequency-domain synthetic
aperture focusing technique for ultrasonic imaging of irregularly layered
objects, IEEE Trans. Ultrason. Ferroelectr. Freq. Control 61(1), 133146
(2014).
9
M. H. Skjelvareid, T. Olofsson, Y. Birkelund, and Y. Larsen, Synthetic
aperture focusing of ultrasonic data from multilayered media using an
omega-k algorithm, IEEE Trans. Ultrason. Ferroelectr. Frequency
Control 58(5), 10371048 (2011).
10
L. Huang and Y. Quan, Ultrasound pulse-echo imaging using the split-
step Fourier propagator, Med. Imag. Int. Soc. Opt. Photonics 6513,
651305 (2007).
11
H. Li, L. H. Le, M. D. Sacchi, and E. H. M. Lou, Ultrasound imaging of
long bone fractures and healing with the split-step Fourier imaging meth-
od, Ultrasound Med. Biol. 39(8), 14821490 (2013).
12
S. A. Sanaee and M. C. Bakker, Ultrasound imaging for quantitative mea-
surement of immersed plastic waste particles, in Proceedings of the 18th
World Conference on Non-Destructive Testing, Durban, SouthAfrica,
2012.
13
A. J. Hunter, B. W. Nkwater, and P. D. Wilcox, The wavenumber algo-
rithm for full-matrix imaging using an ultrasonic array, IEEE Trans.
Ultrason. Ferroelectr. Freq. Control 55(11), 24502462 (2008).
14
M. H. Skjelvareid, Y. Birkelund, and Y. Larsen, Synthetic aperture focus-
ing of outwardly directed cylindrical ultrasound scans, IEEE Trans.
FIG. 13. Ultrasonic wave propagation model. (a) Pulse-echo system, and (b) Ultrason. Ferroelectr. Freq. Control 59(11), 24602469 (2012).
exploding reflector model (ERM). 15
M. H. Skjelvareid, Y. Birkelund, and Y. Larsen, Internal pipeline inspec-
tion using virtual source synthetic aperture ultrasound imaging, NDT E
then received by a transducer array. In the ERM, the acoustic Int. 54(3), 151158 (2013).
16
S. Wu, M. H. Skjelvareid, K. Yang, and J. Chen, Synthetic aperture imag-
velocity is half of the actual acoustic velocity, and the time
ing for multilayer cylindrical object using an exterior rotating transducer,
delay of the received ultrasound wave will be therefore be Rev. Sci. Instrum. 86, 083703 (2015).
17
equivalent to that of the pulse-echo system. According to the Standard Practice for Ultrasonic Examination of Metal Pipe and Tubing,
ERM, the scattered wave is focused at t 0, where all waves ASTM E213, 2004.
18
H. Andresen, S. I. Nikolov, and J. A. Jensen, Synthetic aperture focusing
are focused on their sources and are not diffused. As a result,
for a single-element transducer undergoing helical motion, IEEE Trans.
the lateral image resolution is enhanced. Ultrason. Ferroelectr. Freq. Control 58(5), 935943 (2011).
19
R. A. Waldron, A helical coordinate system and its applications in elec-
1
S. R. Doctor, T. E. Hall, and L. D. Reid, SAFTThe evolution of a sig- tromagnetic theory, Q. J. Mech. Appl. Math. 11, 438461 (1958).
20
nal processing technology for ultrasonic testing, NDT Int. 19(3), 163167 J. F. Clrbout, Imaging the earths interior: Blackwell scientific pub-
(1986). lications, Hematol. Oncol. 11(3), 155156 (1985).
2 21
P. T. Gough and D. W. Hawkins, Unified framework for modern syn- E. G. Williams, Fourier Acoustics: Sound Radiation and Nearfield
thetic aperture imaging algorithms, Int. J. Imag. Syst. Technol. 8(4), Acoustical Holography (Academic Press, New York, 1999).
22
343358 (1997). D. Garcia, L. L. Tarnec, S. Muth, E. Montagnon, J. Poree, and G. Cloutier,
3
J. Johnson, Parameter study of synthetic-aperture focusing in ultra- Stolts f-k migration for plane wave ultrasound imaging, IEEE Trans.
sonics, Rev. Prog. Quant. Nondestr. Eval. 1, 735752 (1982). Ultrason. Ferroelectr. Freq. Control 60(9), 18531867 (2013).