Capnography Extension

to the Multi-Measurement Module (MMS)


Philips M3014A Technical Data Sheet

Features Mainstream and sidestream sensors use infrared


absorption spectroscopy to measure CO2
The analyzer portion of the sensor cannot be
The Capnography Extension works together with contaminated with patient secretions.
the M3001A or M3002A Multi-Measurement
A choice between reusable and single-patient use
Module (MMS) and the M8102A IntelliVue MP2 airway adapters for mainstream CO2
Patient Monitor. It can be attached to the MMS and measurement.
MP2 without cables and sends its processed waves, Mainstream CO2 measurements taken at the
numerics and basic alerts to the host monitor. patients airway, resulting in faster response time
and less chance of erroneous artefact data.
It provides the following measurements:
Easy mounting of sidestream CO2 sensor with
- Mainstream CO2 or Sidestream CO2 and, integrated clip.
as options:
A choice of disposable sidestream CO2
- Invasive Pressure accessories for both intubated and non-intubated
- additional Invasive Pressure or Temperature applications.
(one at a time) Attached to the MMS, it provides a lightweight and
- Cardiac Output /Continuous Cardiac Output rugged package that
(PiCCO) - can be mounted to the patient monitor, or
positioned close to the patient to reduce cable clutter. Environmental Specifications
- can be transferred from one monitor to another to ensure
seamless data transfer with the patient. Temperature Range
All patient connections are fully electrically isolated individually for
Operating: 0 to 45C (32 to 113F).
maximum patient safety.
The M3014A can be used with plug-and-play convenience with the Storage: -40 to 70C (-40 to 158F).
following patient monitor families:
- IntelliVue MP20/30/40/50/60/70/80/90 (M8001A, M8002A, Humidity Range
M8003A, M8004A, M8005A, M8007A, M8008A, M8010A) with
software revision C.0 or higher. Sidestream CO2 requires Operating: 95%RH max. @ 40C (104F), non-condensing
software revision D.0 or higher.
Storage: 90%RH max. @ 65C (150F)

Description Altitude Range (at normal ambient pressure)


Operating: -500 to 4,600 m (-1,600' to 15,000')
The M3014A Capnography Extension is designed for use by clinicians
Storage: -500 to 15,300 m (-1,600' to 50,000')
to measure etCO2 and awRR of intubated and non-intubated adult,
pediatric and neonatal patients in Critical and Intermediate Care areas,
in Operating Rooms and General Wards, and for transport inside the Mainstream CO2 Specifications
hospital.
EN ISO 21647:2004 + Cor.1:2005 (except alarm system; alarm system
complies with IEC 60601-2-49:2001)

Specifications
Measurement Range:
Safety Specifications 0 to 150 mmHg (0 to 20.0 kPa)

The M3014A Capnography Extension complies with the Council


Directive 93/42/EEC (together with the Multi-Measurement Module Resolution
and the IntelliVue MPxx patient monitors). Numeric: 1 mmHg (0.1 kPa)
In addition, the product complies with Wave: 0.1 mmHg (0.01 kPa)
- IEC 60601-1:1988 + A1:1991 + A2:1995
EN60601-1:1990 + A1:1993 + A2:1995 Accuracy (after 2 minutes warm-up and adapter zero)
- UL 2601-1
These specifications are valid for standard gas mixtures, balance air,
- CAN/CSA C22.2#601.1-M90
fully hydrated at 35C, flow rate = 2 l/min. Outside these conditions
The applied parts incl. accessories are designed to have the accuracy reaches at a minimum 0.43% of Pamb + 8% of reading.
special protection against electric shocks (particularly
0 to 40 mmHg (0 to 5.3 kPa): 2.0 mmHg (0.29 kPa)
regarding allowable leakage currents), and are defibrillator proof
41 to 70 mmHg: 5% of reading

Physical Specifications 71 to 100 mmHg: 8% of reading

Size (WxDxH): Endtidal CO2 (etCO2) Alarm Limits


190.0 mm 98 mm 40 mm Range:
(7.5" 4" 1.6") etCO2 low: 10 to 90 mmHg (1 to 12.0 kPa)

etCO2 high: 20 to 95 mmHg (2 to 13.0 kPa)


Weight:
450g (0.99 lb)

2
Adjustments: Adjustments:

1 mmHg (0.1 kPa) steps 1 mmHg (0.1 kPa) steps

Inspired Minimum CO2 (imCO2) Alarm Limit Inspired Minimum CO2 (imCO2) Alarm Limit
Range: Range:

imCO2 high: 2 to 20 mmHg (0.3 to 3.0 kPa) imCO2 high: 2 to 20 mmHg (0.3 to 3.0 kPa)

Adjustments: 1 mmHg (0.1 kPa) steps Adjustments: 1 mmHg (0.1 kPa) steps

Limit Alarm Delay Limit Alarm Delay


< 14 seconds < 14 seconds

Response Time Total System Response Time


Less than 60 ms (with adult or infant reusable or disposable adapter) 3 seconds

Sidestream CO2 Specifications Sample Flow Rate


5010 ml/minute
Operating Temperature:
0 to 40C (32 to 100F) Airway Respiration Rate (awRR):

Measurement Range: Range


0 to 150 mmHg (0 to 20.0 kPa) 2 to 150 rpm

Resolution Accuracy
Numeric: 1 mmHg (0.1 kPa) 1 rpm

Wave: 0.1 mmHg (0.01 kPa)


Alarm Limits
Accuracy (after 2 minutes warm-up and adapter zero) Low Limit Range:

These specifications are valid for gas mixtures of CO2, balance N2, dry Adult/Pedi: 0 to 95 rpm
gas at 760mmHg within specified operating temperature range.
Neo: 0 to 145 rpm
0 to 40 mmHg (0 to 5.3 kPa): 2.0 mmHg (0.29 kPa)
High Limit Range:
41 to 70 mmHg: 5% of reading
Adult/Pedi: 10 to 100 rpm
71 to 100 mmHg: 8% of reading
Neo: 30 to 150 rpm
101 to 150 mmHg: 10% of reading
Adjustments:

Limit < 20 rpm: 1 rpm steps


Endtidal CO2 (etCO2) Alarm Limits
Limit >20 rpm: 5 rpm steps
Range:

etCO2 low: 10 to 90 mmHg (1 to 12.0 kPa)

etCO2 high: 20 to 95 mmHg (2 to 13.0 kPa)

3
Limit Alarm Delay Transducer Load Impedance
Low Limit Alarm, settings from 0 to 20 rpm: Maximum 200 to 2000 (resistive)
4 seconds after value goes below the limit

Low Limit Alarm, settings above 20 rpm: Transducer Output Impedance


< 14 seconds after value goes below the limit
3000 (resistive)
High Limit Alarm: < 14 seconds after value goes over limit
Measurement Range
Apnea Alarm
40 to 360 mmHg
An Apnea alarm is issued if no breath is detected within the set Apnea
Delay Time. Frequency Response
Range:
dc to 12.5 Hz or 40 Hz
Apnea Delay Time: 10 to 40 seconds
Adjustment: Limit Alarms
5 seconds steps
40 to 360 mmHg (5.0 to 48 kPa)
Apnea Alarm Delay:
Set Apnea Delay Time + 4 seconds
Alarm Delay

PRESS Specifications 12 seconds

Complies with IEC 60601-2-34/EN60601-2-34


Pulse Rate Measurement Range

Input Sensitivity 25 to 350 bpm

Accuracy: 1% Full Range


5 V/V/mmHg (37.5 V/V/kPa)
Resolution: 1 bpm
Sensitivity Adjustment Range
Pulse Rate Limit Alarms
10%
25 to 300 bpm
Zero Adjustment
Range: 200 mmHg ( 26 kPa) TEMP Specifications
Accuracy: 1 mmHg ( 0.1 kPa)
Measurement Range
Drift: Less than 0.1 mmHg/C (0.013 kPa/C)
1 to 45C (30 to 113F)

Gain Accuracy (excluding transducers) Resolution: 0.1C (0.2F)

Accuracy: 1% Full Scale (FS) Accuracy (excluding probe): 0.1C ( 0.2F)

Drift: Less than 0.05%/C


Average Time Constant
Non-linearity and Hysteresis: Error of less than, or equal to 0.4% FS
(@ CAL 200 mmHg) Less than 10 seconds

Overall Accuracy Displayed Test Temperature

4 mmHg ( 0.5 kPa) or 4% of reading, whichever is greater 40C 0.1C (104F 0.2F)

4
Limit Alarms ITBV
Range: 1 to 45C (30 to 113F) Range: 50 to 6000 ml

Accuracy: + 10% or 30 ml
Difference TEMP Specifications Repeatability: + 5% or 20 ml

Measurement Range
Continuous Cardiac Output (CCO) Specifications
46C ( 115F)
Range: 0.1 to 25.0 l/min
Accuracy (excluding probe): 0.1C ( 0.2F)
Standard Deviation: 10% or 0.3 l/min

Display Update: 2 seconds nominal


Cardiac Output Specifications (Right Heart)
C.O. Range 0.1 to 20.0 l/min
Cardiac Output Alarm Specifications

Instrument Specification (measured Tblood Alarms


electronically)
Alarm Limit Range: 17 to 43C (62.6F to 109.4F)
C.O. instrument accuracy: + 3% or 0.1 l/min
Adjustment:
C.O. repeatability accuracy: + 2% or 0.1 l/min
Steps of 0.1C (T > 35 C)
Steps of 0.5C (T < 35 C)
System Specification
Alarm Criterion: 8 seconds after the value exceeds the set limit value
C.O. instrument accuracy:
Alarm: within 2 seconds after this criterion is met
+ 5%.or 0.2 l/min (10 l/min)
+ 8% ( 10 l/min)
CCO Alarms
C.O. repeatability accuracy: + 3% or 0.1 l/min
Alarm Limit Range: 0.1 to 25.0 l/min

Adjustment:
Cardiac Output Specifications (Transpulmonary)
Steps of 0.1 l/min (CCO < 10.0 l/min)
C.O. Range: 0.1 to 25.0 l/min
Steps of 0.5 l/min (CCO > 10.0 l/min)

Alarm Criterion: 8 seconds after the value exceeds the set limit value
Instrument Specification (measured
electronically) Alarm: within 2 seconds after this criterion is met.

C.O. instrument accuracy: + 4% or 0.15 l/min

C.O. repeatability accuracy: + 2% or 0.1 l/min

System Specification
C.O. instrument accuracy: + 5% or 0.2 l/min

C.O. repeatability accuracy: + 3% or 0.1 l/min

EVLW (not available in the USA)


Range: 10 to 5000 ml

Standard Deviation: 10% or 1 ml/kg

5
Ordering Information Accessories
M3014A Capnography Extension Only the following accessories may be used with the M3014A. The use
of other accessories may degrade the performance or even cause
The M3014A is available as: damage to the M3014A. Accessories are not included with the
M3014A and may be ordered from the Philips Supplies organization.
Option Description

#A01 Capnography Extension only Mainstream CO2 Accessories

#A02 Capnography Extension with Mainstream CO2 Sensor

#A03 Capnography Extension with Sidestream CO2 Sensor


CO2 Sensors and Accessories Part No.
#C05 Capnography Extension with Press, Press/Temp and
Reusable
C.O.
Mainstream CO2 Sensor M2501A
#C07 Capnography Extension with Press, Press/Temp

#C10 Capnography Extension with Press, Press/Temp, C.O. Reusable adult/pediatric airway adapter M2513A
and PiCCO (1 unit)

Reusable infant airway adapter (1 unit) M2516A


The following accessories can be ordered as a product option:
Disposable (EU/EFTA only. Not available in USA)

Single-patient use adult airway adapter (10 M2533A


Option Description
units)
#K01 Reusable adult/pediatric airway adapter for Mainstream
Single-patient use infant airway adapter (10 M2536A
CO2 (M2513A)
units)
#K02 Reusable infant airway adapter for Mainstream CO2
(M2516A)

#K03 Single-patient use adult airway adapter for Mainstream


CO2 (M2533A)

#K04 Single-patient use infant airway adapter for Mainstream


CO2 (M2536A)

#K41 Single-patient use adult CO2 Oral-Nasal Cannula for


Sidestream CO2 (M2756A)

#K42 Single-patient use pediatric CO2 Oral-Nasal Cannula


for Sidestream CO2 (M2757A)

#K43 Airway Adapter Set H, ET > 4.0 mm (M2772A).

#K44 Airway Adapter Set H, ET 4.0 mm (M2773A)

6
Sidestream CO2 Accessories TEMP Accessories

CO2 Sensors and Accessories Part No. Temperature Probes Part No.

Sidestream CO2 Sensor M2741A Reusable

CO2 Nasal Cannula, Adult M2744A General purpose probe 21075A

Small flexible vinyl probe (Infant/Pediatric) 21076A


CO2 Nasal Cannula, Pediatric M2745A
Attachable surface probe 21078A
CO2 Nasal Cannula, Infant M2746A
Disposable
CO2 / O2 Nasal Cannula, Adult M2750A
General purpose probe M1837A
CO2 / O2 Nasal Cannula, Pediatric M2751A
Skin probe 21091A
CO2 / O2 Nasal Cannula, Infant 989803144471 Esophageal/Stethoscope Probe 21093A
CO2 Oral-Nasal Cannula, Adult M2756A (12 French)

Esophageal/Stethoscope Probe 21094A


CO2 Oral-Nasal Cannula, Pediatric M2757A
(18 French)
CO2 / O2 Oral-Nasal Cannula, Adult M2760A Esophageal/Stethoscope Probe 21095A
(24 French)
CO2 / O2 Oral-Nasal Cannula, Pediatric M2761A
Foley Catheter Probe M2255A
Airway Adapter Set, ET > 4.0 mm M2768A
(12 French)
Airway Adapter Set, ET=< 4.0 mm 989803144531
Foley Catheter Probe 21096A
Airway Adapter Set H, ET > 4.0 mm M2772A (16 French)

Airway Adapter Set H, ET 4.0 mm M2773A Foley Catheter Probe 21097A


(18 French)
Straight Sample Line M2776A
Adapter cable 1.5 m/4.9 21082B
Straight Sample Line H M2777A
Adapter cable 3.0 m/9.8 21082A

7
PRESS Accessories

Pressure Transducers and Accessories Part No. Accessories - PiCCO Method

Reusable C. O. Interface Cable

Reusable pressure transducer CPJ840J6 2.4 + 2.4 m cable M1643A


5 V/V/mmHg sensitivity C.O. Temperature Probes
Sterile disposable pressure domes for CPJ84022 Inline temperature probe for warm and cold M1646A
CPJ840J6 (pack of 50) injectate, Latex free, compatible with housings
Transducer holder for CPJ840J6 CPJ84046 from PULSION
(pack of 4) Pulsion disposables
IV pole mount for CPJ840J6 CPJ84447 Arterial Catheter 4F, 16 cm, or other compatible PV 2014 L16
Disposable (EU/EFTA only. Not available in USA) catheter as validated by Pulsion

Single channel disposable sensor kit (20) M1567A PiCCO Pressure See PRESS Accessories
Transducer Kits
Dual channel disposable sensor kit (20) M1568A compatible with
Transducer holder for M1567/8A M2271A M1646A

IV pole mount for M1567/8A M2272C

Adapter cable for disposable sensor kit, 3.0 M1634A


m/9.8, for M1567/8A Additional Accessories for both methods

Set of ice buckets 14455A


C.O. / CCO Accessories
Remote handswitch 15244A
See Pressure accessories for PULSION continuous cardiac output
accessories.

Accessories - Right Heart Method

C. O. Interface Cable

2.7 +0.2 m cable M1642A

2.4 + 2.4 m cable M1643A

C.O. Temperature Probes

Inline temperature probe for warm and cold M1646A


injectate, Latex free, compatible with housings from
PULSION

2.4 m standard inline injectate temp. probe for cold 23001A


injectate, compatible with Edwards/Baxter housings

0.5m standard inline injectate temp. probe for cold 23001B


injectate, compatible with Edwards/Baxter housings

Ice bath temp. probe 23002A

8
9
Philips Healthcare is part of
Royal Philips Electronics

How to reach us
www.philips.com/healthcare
healthcare@philips.com
fax: +31 40 27 64 887

Asia
+852 2821 5888

Europe, Middle East, Africa


+49 7031 463 2254

Latin America
+55 11 2125 0744

North America
+1 425 487 7000
800 285 5585 (toll free, US only)

Philips Healthcare
Global Information Center
P.O Box 1286
5602 BG Eindhoven
The Netherlands

2008 Koninklijke Philips Electronics N.V.


All rights are reserved.
Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or
obligation and will not be liable for any consequences resulting from the use of this publication.
Printed in The Netherlands.
4522 962 42321 DEC 2008