Vous êtes sur la page 1sur 6

Anal Bioanal Chem (2009) 395:473478

DOI 10.1007/s00216-009-2980-9

ORIGINAL PAPER

Application of external micro-spectrophotometric detection


to improve sensitivity on microchips
Attila Gspr & Istvn Bcsi & Erika F. Garcia &
Mihly Braun & Frank A. Gomez

Received: 3 May 2009 / Revised: 28 June 2009 / Accepted: 15 July 2009 / Published online: 1 August 2009
# Springer-Verlag 2009

Abstract The goal of this work was to increase the Introduction


sensitivity of a UVVis spectrophotometer by decreasing
the background noise and lengthening the optical path. A Microfluidic lab-on-a-chip technologies and devices repre-
microphotometer has been modified to precisely select very sent a revolution in laboratory experimentation, bringing
small parts of a microfluidic channel pattern of a chip and the benefits of miniaturization, integration, and automation
to measure light absorbance on a magnified area of the to many research-based laboratories and industries. The use
selected part of the channel. The viability of combining a of microfluidic chips has attracted great attention as they
projection microscope and a spectrophotometer for external offer several advantages over conventional analytical
absorbance measurements on disposable PDMS chips was techniques including small sample volume requirements,
studied. Besides the external direct detection above a portability, fast sampling times, ability to multiplex and
microfluidic channel, the optical pathlength was lengthened compatibility with other techniques [1]. The development
by detecting in the region of the perpendicular exit port. of novel and miniaturized detection systems employed in
Increasing the cross-sectional area of the zone of irradiation microfluidic technologies will both simplify and hasten
improved the signal-to-noise ratio and the limits of analysis on chip platform. The most simple, universal, and
detection (LOD). inexpensive detection mode is that based on UVvisible
spectrophotometry. Unfortunately, optical detection can be
Keywords Chip . Poly(dimethylsiloxane) . problematic as microchips have small pathlengths greatly
External UV detection . Micro-spectrophotometer . reducing sensitivity and limits of detection (LOD). Hence,
Flow injection . Microfluidics/microfabrication . UV/Vis few applications of spectrophotometric absorbance meas-
urements in microfluidic chip systems have been reported.
For example, a silicon flow cell was used to detect dyes
whereby the silicon was utilized as mirrored channel [2].
A. Gspr (*) : M. Braun Haswell and coworkers used spectrophotometric detection
Department of Inorganic and Analytical Chemistry, in chip for the determination of phosphate [3], nitrate [4],
University of Debrecen, and nitrite [5] integrating the optical fibers to the channel
Egyetem tr 1,
ends. Numerous groups have used fiber optics in microchip
Debrecen 4032, Hungary
e-mail: gaspara@tigris.unideb.hu capillary electrophoresis [6, 7]. In this approach, alignment
of the fibers with the liquid channel is difficult. Of
I. Bcsi particular concern is light, which circumvents the liquid
Cetox Ltd,
channel and reaches the detector. Others used optical
Egyetem tr 1,
Debrecen 4032, Hungary waveguides that are incorporated into the chip to enhance
the effective usage of the analytical light beam (effective-
E. F. Garcia : F. A. Gomez ness of the interaction of the light with the analyte) [8, 9].
Department of Chemistry and Biochemistry,
Numerous designs have been proposed to increase the
California State University, Los Angeles,
5151 State University Drive, pathlength by passing the light along the length of the flow
Los Angeles, CA 90032-8202, USA channel rather than across it [2, 10, 11].
474 A. Gspr et al.

Poly(dimethylsiloxane) (PDMS) is a favorite material to prepared in water. All solutions (methanol, water) were
construct microfluidic chips from due to its low cost, ease degassed and filtered through a 0.45-m syringe filter.
of fabrication and robustness [12]. There are a number of
studies that detail novel detection schemes for PDMS-based Microphotometer
microchips but they either require one, the compounds of
interest to be labeled (in the case of fluorescent detection) The analytes injected into the chip were detected by a
or two, integration of detection components (for example, modified microphotometer (Schnellphotometer-II, VEB
conductivity and amperometry) into the chips [13]. The use Carl Zeiss, Jena; Fig. 1) equipped with a fiber optic
of label-free detection modes is growing and will increase miniaturized spectrophotometer. For many years, micro-
in usage in lab-on-a-chip devices. Similarly, incorporation photometers (microdensitometers) have been used for
of sensor components into chips is quite developed [14]. spectrographs to evaluate photographic plates (photoem-
On the other hand, it would be beneficial if neither labeling ulsions placed on glass plates). This device directs a steady,
nor chip modification were needed to detect molecules in narrow beam of light through the photographic plate to a
the ultraviolet and visible range using a PDMS-based light sensitive detector. As the plate was stepped along the
microchip. One advantage of PDMS is that it does not dark and transparent part of the plate, the detector measured
absorb visible light and only slightly above 250 nm. and recorded the amount of light.
Herein, we describe the use of a microphotometer, The microphotometer consists of an xy translational
originally designed to measure photographic density on a stage to move and accurately control the position of the
photographic plate, to measure light absorbance on a magni- microfluidic chip where the absorbance measurement is
fied area of a microchannel. The instrument is a combination intended to be carried out. The chip can be fixed onto a
of a projection microscope and a spectrophotometer. The movable stand and positioned manually by means of
applicability of this instrument for external absorbance adjustable screws with 0.0025 mm accuracy. The part of
measurements on disposable PDMS chips was studied. the chip to be investigated should be positioned above and
Besides external direct detection above a microfluidic channel, below the lighting and projective objectives, respectively.
the optical pathlength was lengthened by detecting in the After focusing, approximately a 1-mm2 area of the chip can
perpendicular exit port. be projected onto the screen to obtain an enlarged picture of
the spot (Fig. 1, right side). The screen provides a rapid
visual survey of the section of the channel under examina-
Experimental tion and in fine detail. Approximately 100 magnification
of the section of the chip under study can be obtained on
Chemicals and reagents the screen.
On the spot of the chip projected onto the screen, a
Stock solutions of Brilliant Blue FCF food dye (FD&C rectangle opening with variable sides can be specified by
blue#1 or E133 from Szilas Ltd, Hungary) and 0.1 mM adjusting slits in both horizontal and vertical directions
phosphate complexed in molybdenum blue compound (the (Fig. 1, right side). Through the specified rectangle, the
phosphate was mixed with a reagent containing ammonium light of the projected picture can pass to the detector, thus
molybdate, potassium antimonyl tartarate and sulfuric acid the absorbance measurement can be limited only to the
followed by ascorbic acid according to Ref. [15]) were specified area of the chip (inside a given length of the

Fig. 1 The construction of the


microphotometer (Schnellpho-
tometer, Carl Zeiss, Jena) used
projection screen
for spectrophotometric detection
adjustable slit
on chip. The optical fiber of a
miniaturized spectrophotometer magnified
(Avaspec-2048-2, Avantes, The picture of the chip
Netherlands) was positioned
behind the projection screen.
The magnified picture about the
projection screen is shown in
the right side focus setting
projecting objective
chip on movable stand
lighting objective
Application of external micro-spectrophotometric detection 475

microfluidic channel) resulting in good utilization of light 8 developer (Microchem, Newton, MA). The PDMS chip
for the spectrophotometric detection of the analyte. The was fabricated by cast molding of a 10:1 mixture of PDMS
horizontal and vertical apertures of the slits are 05 and 0 oligomer and cross-linking agent (Sylgard 184, Dow
20 mm, respectively. Corning, Midland, MI). The PDMS mixture was degassed
In our experiments, a tungsten lamp was used as the light and baked at 80 C for 30 min. The PDMS replicas were
source. The optical fiber of a miniaturized spectrophotom- peeled off from the mold.
eter (Avaspec-2048-2, Avantes, The Netherlands) was Holes of 0.3 mm diameter for the liquid connections to C
positioned behind the projection screen. The light reached (carrier) and S (sample) ports, and hole of 0.5 mm diameter
the optical fiber after passing through the slit of the adjusted for the perpendicular detection to the E (exit) port (Fig. 2.)
area and collected by lens. The spectrophotometric meas- were punched (Schmidt Technology GmbH (St. Georgen,
urements were controlled and recorded by Avasoft Germany) through the PDMS chip. The inner diameter of
(Avantes, Eerbeek, The Netherlands). Since the detection the E port can be decreased to 50400 m if a short piece
was performed externally, any point along the chip could be of fused-silica capillary is tightly inserted into the port.
detected. The obtained absorption data were exported to a Then, the chip was irreversibly sealed onto a standard glass
Microsoft Excel program for further analysis. The standard slide of 1 mm thickness (VWR, USA) using air plasma
PDMS chip with glass slide could be used for the oxidation (PDC-32G, Harrick Plasma, Ithaca, NY, USA).
spectrophotometric detection in the visible range. Figure 3 shows the pieces of fused-silica capillaries of
50 m and 400 m ID (360 m and 600 m OD,
Chip fabrication respectively) inserted into the port of PDMS chip.
In order to remove the liquid coming from the exit port
The PDMS chips were prepared by using a mold created by and to keep clear and bubble-free the top of the E port for
soft photolithography [10]. The pattern consisting of the spectrophotometric detection, the E port is covered
standard T-type and cross T-type channel of 50 m wide (sealed by air plasma oxidation) with a thin (about 1 mm)
was designed using AutoCAD software (San Rafael, CA) layer of PDMS including a 0.2-mm-wide, 50-m-deep, and
and printed as a high resolution (20,000 dpi) photomask 5-mm-long channel. One end of this channel is positioned
(CAD/Art Services, Inc., Bandon, OR). Negative type to the E port, and a 0.3-mm diameter for the liquid
photoresist (SU-8 2025, Microchem, Newton, MA) was connections to the W (waste) port (Fig. 2.) was punched
spin-coated onto a 3 silicon wafer at 2,600 rpm for 30 s to through the PDMS layer at the other end of the channel.
a thickness of 30 m. The photoresist-coated wafer was
baked for 15 min at 95 C. The pattern on the mask was Injection of sample plugs
transferred to the wafer through UV exposure for 2 min.
The exposed wafer was baked at 95 C for 5 min and Sample plugs were injected into the chip by pressure or
unexposed areas were removed by rinsing with SU- high voltage. Upon detection of the sample plugs in the
Fig. 2 Schematic representation Detection site
of the standard T-type channel
chip made from PDMS from the a, Detection site through the exit port
side (a) and from above (b). The through the channel
top of the E exit port is covered
with a thin layer of PDMS
including a channel to lead the
Fluid channel W
C S
waste liquid out from the chip.
The arrows indicate the E
possible detection sites of PDMS layers
the external detection with
the microphotometer
Glass/quartz slide

b, S
W
C

E
476 A. Gspr et al.

the same even if the cross-sectional area of the irradiation is


changed. This is advantageous for detection on chips,
because it means that we get the same absorbance value
even if we measure a very small area of the sample plug
(e.g., 104 mm2) in the microfluidic channel (its width is
typically less than 100 m) as compared to the conven-
tional measurement through a cuvette of six orders of
magnitude larger irradiated area (a 1-cm-side cuvette).
Using the microphotometer and selecting a section of the
microfluidic channel on the magnifying projection screen,
we can precisely select very small parts of the channel
pattern of the chip for measuring absorbance. On the
projected picture of the chip, an area of rectangular
dimension inside the microfluidic channel (in the analyte
plug) can be specified by adjusting the slits horizontally
and/or vertically (Fig. 1, right side). Thus, a rectangle with
sides of 250 m and 2200 m can be specified for the
absorbance measurements. Using the microphotometer, the
absorbance measurement is limited only to the small area of
200 m the chip inside the microfluidic channel; therefore, very
good utilization of light for the analyte detection can be
achieved. In Fig. 4, the absorbance signals collected for
Fig. 3 Microscopic pictures around the pieces of fused-silica 60 s were compared using slit areas of 20, 40, 200, and
capillaries of 50 (a) and 400 m (b) ID (360 and 600 m OD, 400 m2 (104 times larger on the projected, magnified
respectively) inserted into the port of the PDMS chip screen). Although the obtained absorbance values are not
altered (as expected due to the LambertBeer law),
microfluidic channel, a subnanoliter volume of dye sample considerable differences can be found in the signal noise.
(length in channel less than 200 m) was used for the Increasing the cross-sectional area of the irradiation
detection. Electrokinetic injection was used in the cross T- decreases the noise.
type channel. Here, 400 V was applied to produce constant The LOD of a measurement depends on the signal-to-
flow in the sample channel and when the junction of the noise (standard deviation) (LOD 3s=S, where and S
chip filled with the sample, 800 V was applied to the ends
of the separation channel. For the electrokinetic injections
two miniaturized high-voltage (0.22 kV) power supplies 50 mAU
were used (Microply-01, Cetox Ltd., Hungary). The dye
sample migrated toward the cathode (electrolyte, 25 mM
phosphate, 50 mM SDS; pH, 3).
For detection of the sample plugs through the E port, a,
only a single-channel peristaltic pump (Ismatec IPC,
Switzerland) was used. The sample (0.12 L) was injected
into the peristaltic pump tubing (ID, 0.3 mm), previously
filled with water, and pumped into the S port while the C b,
port was closed.
c,

d,
Results and discussion
0 20 40 60
One of the most well-known relationships in analytical Time (s)
chemistry is the LambertBeer law. This law states that
absorption is proportional to the light pathlength and the Fig. 4 The absorbance signals using different slit areas on the
concentration of the absorbing species. Since the obtained projected picture a 2, b 4, c 20, and d 40 mm2 (the projected areas on
chip were 104 times smaller; sample: blue dye (FD&C blue#1), =
absorbance does not depend on the cross-sectional area of 640 nm) The standard deviations were calculated to 0.015 (a), 0.0049
the irradiated sample (cuvette), the absorbance should be (b), 0.0015 (c), and 0.0014 (d)
Application of external micro-spectrophotometric detection 477

Table 1 Limit of detection values of phosphate (in molybdenum blue


0,60 complex) using on-channel and in-exit-port detections with fiber optic
spectrophotometry for chips, and with conventional spectrophotome-
try including 1-cm-length cuvette (=640 nm)
Absorbance

0,40 LOD (M) Volume of sample used


for the detection (nL)

On-channel detectiona
0,20
20 m2 50 0.6
40 m2 38 1.2
0,00 200 m2 5.6 6.0
0 20 40
Time (s)
60 80
400 m2 5.3 12
In-exit-port detectionb 0.78 120
Fig. 5 Absorption signals of sample plugs injected by voltage into the Conventional spectrometryc 0.16 1,000,000
chip (sample: blue dye (FD&C blue#1), sample volume, 0.15 nL, =
640 nm) a
Cross-sectional areas of the irradiation on the chip were ranged
between 20400 m2
b
are the standard deviation of the response and slope of the The height of the exit port was 2.5 mm
c
calibration curve, respectively). Increasing the cross- Supporting that at least 1 cm height of sample should be loaded into
sectional area of the irradiation from 20 to 400 m2 the cuvette
decreases the standard deviations from 0.015 to 0.0017
(resulting in similar improvement in LOD values), respec- be increased due to higher Joule-heat formation during
tively. Further increasing the area did not improve the electrophoretic applications, attenuated electroosmosis, less
standard deviation. These data explain that a larger achievable resolution for separations, etc.). Transverse
irradiation area leads to a large number of parallel detection detection through a given length of the channel provides
in the sensor at the same time and the standard deviation of for more sensitive detection, but is not amenable for
the large number of parallel detections should be smaller. detection of small plugs of sample (which are typical in
The use of electrokinetic injection allows for the microfluidic applications). Furthermore, transverse detec-
injection of subnanoliter volumes (plugs of about 100 m tion frequently requires complicated adjustment of the
length in the channel of 50 m width and 30 m height) optical fibers. The most simple solution to increasing the
into the channel of a cross T-type chip and with high height of the liquid layer is when the detection is performed
reproducibility (Fig. 5). The use of the microphotometer in the exit port perpendicular to the end of the microfluidic
allows for the absorbance measurement within the spot of channel. Using a microphotometer, a 50-m-wide hole of
the flowing subnanoliter plug. the exit port can be specified for the detection (Fig. 3).
The height of the flowing liquid layer in the channels of Since microscopic bubbles are often trapped at the bottom
microchips is extremely small (3050 m) and should not of the capillary thereby complicating spectrophotometric

100 M

0.75 a, 0.79
b,

75 M

0.59
0.5
Absorbance

Absorbance

50 M

0.39

0.25
25 M
0.19
c,
10 M
0 5 M
0 25 50 75 100
Concentration (M) -0.01

Fig. 6 Calibration graphs obtained by conventional spectrophotomet- Fig. 7 Signals of standard solutions of phosphate in concentration
ric measurement with 1-cm-length cuvette (a), in-exit-port of a chip range of 5100 M (sample: phosphate (in molybdenum blue
(b) and on-channel of a chip (c) detection complex), sample volume, 0.2 L =640 nm)
478 A. Gspr et al.

detection, capillaries of 250400 m ID are typically used. analytical light beam, detection, both inside the 50400-m-
The height of the detection cell (E exit port) was 2.5 mm wide hole of the exit port and in the flowing subnanoliter
(optical path length in the chip), comparable to classical volume of sample plug, are needed. In the proposed
cuvette-based spectrophotometric techniques, but the vol- experimental setup, the interaction of the light with the
ume of the cell is only 120 nL (for 250 m ID capillary), analyte is better than in the case of fiber optics in a
which is in the range of the volume of solution found in spectrometer arranged to the channel without the micropho-
microchannels in a chip. This exit port detection is not tometer. On the other hand, the use of adjustable slits on the
suitable for detection of small sample plugs (e.g. in case of projection screen of the microphotometer maximizes the
electrophoretic separations in chip), but it can be applied in irradiated area (e.g., the whole area of a subnanoliter sample
flow injection analysis (FIA) using sample plugs of plug), vis-a-vis, all the light which reaches the detector
submicroliter volumes. The dual use of microfluidics passes through the liquid in the channel and not outside the
integrating FIA and separation systems has recently been channel. The sensitivity of the spectrophotometric measure-
proposed [16]. Figure 6 shows the measurement of standard ments can be improved by detecting in the exit port having a
solutions of phosphate (in molybdenum blue complex) in small inner diameter.
the concentration range of 5100 M. The 0.2 L volumes
of sample solutions were transferred by use of a peristaltic Acknowledgments The authors gratefully acknowledge financial
support for this research by grants from the National Scientific
pump (ID of tubing was 250 m) and injected into the chip
Research Fund, Hungary K75286), the National Office for Research
via the S sample port. As can be seen, even low-micromolar and Technology (Baross-2-2007-0028) and the National Science
concentrations of a phosphate sample can be reproducibly Foundation (DMR-0351848, CHE-0515363, and OISE-0754138).
detected. LOD values of phosphate using on-channel and
in-exit-port detections with fiber optic spectrophotometry
for chips, and with the conventional spectrophotometry was References
summarized in Table 1. While the attainable sensitivity less
than five times worse using in-exit-port detection, the 1. Tabeling P (2005) In introduction to microfluidics. Oxford
University Press
needed sample volume almost 10,000 times smaller than in 2. Verpoorte E, Manz A, Ludi H, Bruno AE, Maystre F, Krattinger
conventional spectrophotometry. The LOD values are far B, Widmer HM, Vanderschoot BH, Derooij NF (1992) Sens
from the desired level when on-channel detection is Actuators B 6:6674
applied. 3. Doku GN, Haswell SJ (1999) Anal Chim Acta 382:113
4. Greenway GN, Haswell SJ, Petsul PH (1999) Anal Chim Acta
In Fig. 7, the calibration graphs obtained by on-channel 387:110
and exit port detection are compared. The latter one is more 5. Petsul PH, Greenway GM, Haswell SJ (2001) Anal Chim Acta
than 50 times more sensitive due to the longer optical path 428:155164
length. In this figure, the calibration graph obtained by 6. Manz A, Verpoorte E, Effenhauser CS, Burggraf N, Raymond DE,
Widmer HM (1994) Fresenius' J Anal Chem 348:567571
conventional spectrophotometric measurement with cuvette 7. Liang Z, Chiem N, Ocvirk G, Tang T, Fluri K, Harrison DJ (1996)
(1 cm side) is also compared. Anal Chem 68:10401046
8. Duggan MP, McCreedy T, Aylott JW (2003) Analyst 128:1336
1340
9. Ruano JM, Benoit V, Aitchison JS, Cooper JM (2000) Anal Chem
Conclusions 72:10931097
10. Daykin RNC, Haswell SJ (1995) Anal Chim Acta 313:155159
We have described the use of a microphotometer coupled to 11. Daridon A, Sequira M, Pennarun-Thomas G, Dirac H, Krog JP,
a spectrometer for detection of analytes in a microchip Gravesen P, Lichtenberg J, Diamon G, Veerporte E, de Rooij NF
(2001) Sens Actuators B 76:235244
format. Although the size of the microphotometer does not 12. Duffy DC, McDonald JC, Schueller OJA, Whitesides GM (1998)
fit well to the microfluidic scale (which is a general Anal Chem 70:49744984
problem for numerous other detection techniques, including 13. Dittrich PS, Tachikawa K, Manz A (2006) Anal Chem 78:3887
LIF or MS detections, used for microfluidics), the applied 3906
14. Marle L, Greenway GM (2005) Trends Anal Chem 24:795802
microphotometer makes universal detection mode possible 15. Galhardo CX, Masini JC (2000) Anal Chim Acta 417:191200
(not the case for LIF). Improved LODs were achieved in 16. Crevillen AG, Pumera M, Gonzalez MC, Escarpa A (2009) Lab
microchannels. To enhance the effective usage of the Chip 9:346353

Vous aimerez peut-être aussi