C CB 010 002 ESTIMATED AND MEASURED HEPARIN VOL. DIFFER TOO MUCH
Conditional error, confirm by zero-set heparin flow
CF 31 1 BiCart holder
C CO 072, Conductivity
C CO 074, UF Profiling
C CO 080, Attention
C CO 096, Profiling
C COFF 088 004 BLOOD LEAK DETECTOR TEST ERROR, VALUE TOO HIGH
Unconditional error, blood pump available
C COFF 088 005 BLOOD LEAK DETECTOR TEST ERROR, VALUE OUT OF
LIMIT
Unconditional error, blood pump available
C COFF 088 006 BLOOD LEAK DETECTOR TEST ERROR, UNSTABLE OFF
Unconditional error, blood pump available
C COFB 088 008 BLOOD PUMP TEST ERROR, PUMP NOT RUNNING (FIRST
TEST)
Unconditional error, blood pump not available
C COFB 088 009 BLOOD PUMP TEST ERROR, PUMP NOT RUNNING
(SECOND TEST)
Unconditional error, blood pump not available
C COFB 088 010 BLOOD PUMP TEST ERROR, PUMP NOT STOPPED (CPU P)
Unconditional error, blood pump not available
C COFB 088 011 BLOOD PUMP TEST ERROR, PUMP NOT RUNNING
ROTATION INDICATOR, (FIRST CHECK)
Unconditional error, blood pump not available
C COFB 088 012 BLOOD PUMP TEST ERROR, PUMP NOT RUNNING
ROTATION INDICATOR, (SECOND CHECK)
Unconditional error, blood pump not available
Unconditional error, blood pump not available
C COFB 088 013 BLOOD PUMP TEST ERROR, PUMP NOT RUNNING
TACHO SIGNALS MISSING, (FIRST CHECK)
Unconditional error, blood pump not available
C COFB 088 014 BLOOD PUMP TEST ERROR, PUMP NOT RUNNING
TACHO SIGNALS MISSING, (SECOND CHECK)
Unconditional error, blood pump not available
C COFB 088 015 BLOOD PUMP TEST ERROR, PUMP NOT STOPPED (CPU W)
Unconditional error, blood pump not available
C COFB 088 019 BUZZER TEST ERROR, BUZZER NOT OFF (CPU P)
Unconditional error, blood pump not available
C COFB 088 022 BUZZER TEST ERROR, BUZZER NOT OFF (CPU W)
Unconditional error, blood pump not available
C COFB 088 123: Arterial clamp is set NOT installed but is detected
by Fch as present (Clamp is not closed).
C COFB 088 124: Dialysis machine in Single Needle mode but arterial
clamp is not installed
C COFB 088 125: Assisted Priming Alt 1 contains auto venous drip chamber
adjustment but no arterial clamp is installed.
C COFB 088 126: Assisted Priming Alt 2 contains auto venous drip chamber
adjustment but no arterial clamp is installed.
C COFB 088 127: Assisted Priming Alt 3 contains auto venous drip chamber
adjustment but no arterial clamp is installed.
C COFF 088 030 FLOW SWITCH ERROR, FLOW DETECTED (INVA CLOSED)
Unconditional error, blood pump available
C COFF 088 031 FLOW SWITCH ERROR, NO FLOW (INLET PRESSURE OK)
Unconditional error, blood pump available
C COFF 088 037 HEATER TEST ERROR, HEATER NOT ENERGIZED (CPU C)
Unconditional error, blood pump available
C COFF 088 038 HEATER TEST ERROR, HEATER NOT ENERGIZED (INVA
OPEN)
Unconditional error, blood pump available
C COFF 088 039 HEATER TEST ERROR, HEATER NOT ENERGIZED (CPU P)
Unconditional error, blood pump available
C COFB 088 040 HIGH VENOUS PRESSURE TEST ERROR, TOO HIGH VEN.
PRESS.
Unconditional error, blood pump not available
C COFB 088 041 HIGH VENOUS PRESSURE TEST ERROR, TOO BIG
DIFFERENCE
Unconditional error, blood pump not available
C COFB 088 050 SAFETY RELAY TEST ERROR, REFERENCE CURRENT TOO
HIGH
Unconditional error, blood pump not available
C COFB 088 069 VENOUS CLAMP TEST ERROR, CLAMP NOT OPEN AT
START
Unconditional error, blood pump not available
C COFB 088 070 VENOUS CLAMP TEST ERROR, CLAMP NOT OPEN (CPU P)
Unconditional error, blood pump not available
C COFB 088 071 VENOUS CLAMP TEST ERROR, CLAMP NOT OPEN (CPU W)
Unconditional error, blood pump not available
C COFB 088 072 VENOUS CLAMP TEST ERROR, CLAMP NOT CLOSED (CPU
P)
Unconditional error, blood pump not available
C COFB 088 073 VENOUS CLAMP TEST ERROR, CLAMP NOT CLOSED (CPU
W)
Unconditional error, blood pump not available
C COFF 088 099 DIVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 102 BYVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
General description of conditions for occurring:
- BYVA should be open but the measured valve current indicates that it is
closed.
C COFF 088 103 BYVA C OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 104 BYVA P OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 105 TAVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 106 EVVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 110 DRVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 111 REVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 112 ZEVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 113 RIVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 114 INVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 115: The negative pressure of -350 mmHg is not accomplished.
Probable cause is huge air leakage somewhere between INVA and ZEVA.
Other causes could be a malfunctioning HPG sensor or a malfunctioning
suction pump.
C COFF 088 116: The closed in pressure have increased above the negative pressure
of
-250 mmHg during the stabilization period of 4 seconds.
robable cause is huge air leakage somewhere between INVA and ZEVA.
C COFF 088 117: The closed in pressure have decreased more then 10 mmHg
within 10 seconds. Probable cause is minor airleakage in
ZEVA, BYVA, RIVA or degassing chamber.
C COFF 088 118: The closed in pressure have increased more then 10 mmHg
within 10 seconds. Probable cause is minor airleakage INVA,
AIVA, CBVA or fluid path is leaking.
C COFF 088 119: The closed in pressure have increased above the negative pressure
of
-250 mmHg during the stabilization period of 4 seconds.
Probable cause is huge air leakage somewhere between INVA and ZEVA.
C COFF 088 120: The closed in pressure have decreased more then 10 mmHg within
10 seconds.
Probable cause is minor airleakage in ZEVA, BYVA, RIVA or degassing
chamber.
C COFF 088 121: The closed in pressure have increased more then 10 mmHg within
10 seconds.
Probable cause is minor airleakage INVA, AIVA, CHVA or fluid path is
leaking.
C COFF 088 122: Control venous pressure sensor and protective venous pressure
sensor
differs more then 10 mmHg.Probable cause is malfunctioning sensors
or not correct calibrated sensors.
The following error codes are silent error codes. If a test corresponding to
each of the error codes fails, the fluid monitor functional check will be restarted
and the error code C FRFF XXX XXX will be stored in the error code buffer. If
the test fails a second time, the error code C COFF XXX XXX will appear, together
with buzzer alarm and spanner key light.
The following error codes are silent error codes. If a test corresponding to
each of the error codes fails, the fluid monitor functional check will be restarted
and the error code C FRFF XXX XXX will be stored in the error code buffer. If
the test fails a second time, the error code C COFF XXX XXX will appear, together
with buzzer alarm and spanner key light.
P SDT 004 001 CPU C DOES NOT SEND SHUT DOWN ACKNOWLEDGE
P SDT 004 002 CPU W DOES NOT SEND SHUT DOWN ACKNOWLEDGE
P SDT 047 001: This error is generated if the conditions for performing the high
venous-pressure test are still unfulfilled after
10 minutes into treatment.
The table below is a brief sequence diagram of the start in each CPU.
The number in brackets refer to number shown on LED display at start-up.
Init. of chip select logic. Init. of chip select logic. Cold or warm start?(1)
If warmstart jump to A
Init. I/O ports and timers. Init. of I/O ports and timers. Internal memory and register
verification.(2)
Cold or warm start? (1) Cold or warm start? (1) Program verification.(3)
If warmstart jump to A If warmstart jump to A
OS start(6) OS start(6)