Vous êtes sur la page 1sur 4

Aurora 1030C

Total Organic Carbon Analyzer


Aurora 1030C High Temperature

The Aurora 1030C employs high temperature


(680 C) catalytic combustion to oxidize and
convert organic compounds present in aqueous
samples to CO2 for measurement by a solid state
non-dispersive infrared (SS-NDIR) detector.

The high temperature combustion technique is


most effective for analysis of samples containing
high molecular weight, difficult-to-oxidize
organics (e.g; humic acid) at levels > 500 ppb C.

The Aurora 1030C is equipped with a patented*


two chamber combustion reactor that overcomes
problems associated with TOC analyzers that use
a single chamber reactor. The first chamber of the
1030C reactor contains a bed of quartz to protect
the platinum catalyst in the second chamber from
deposition of noncombustible constituents, and
ensures consistent oxidation conditions for stable
blanks. This reactor design extends catalyst life
and reduces instrument maintenance costs.

*Patent: US 7,306,770 B2

Test Methods Supported by the Aurora 1030C TOC Analyzer


The high temperature combustion technique has been approved and adopted in numerous regulatory
compliance methods, standards, and norms for water quality testing.

Method Sample / Application


USEPA 415.1 Drinking Water
Standard Method 5310B Wastewater, Ground & Surface Waters
USEPA 9060A Wastewater, Ground & Surface Waters
USEPA-DBPR Disinfection Byproduct Rule
ISO 8245 Drinking Water, Wastewater
EN 1484 Surface & Ground Waters, Potable Water
ASTM D7573-09 Water, Wastewater, Seawater
Combustion TOC Analyzer
The Aurora 1030C can be equipped
with a number of instrument options Aurora 1030D Dual Oxidation
and automation accessories to improve Mode TOC Analyzer
sample throughput and productivity.

1030D Dual Oxidation Mode


Analyzer option performs both high
temperature combustion and heated
persulfate wet oxidation techniques
on the same instrument (with manual
adjustment).

An 88-position rotary autosampler


fits directly underneath the 1030C
to conserve bench space.

ATOC software provides fully automated


data collection, analysis, reporting, and
storage in a LAN/LIMS environment.

Multi-stream sampling module for at-


line monitoring of up to four process
streams.

1030S Solids Module operates in High Temperature Wet Oxidation


conjunction with a 1030C analyzer to Combustion Tube Reaction Chamber
combust solid materials for analysis of
TC or TOC. Organic C + 02 680 C CO2 S 0 2 + 2H 0 98 C 2SO 2 + 2H+ + 2OH
2 8 2 4
Pt 2
Organic C + 2S04 + OH

C02+ H20

Applications
OI Analytical has been an innovator in
TOC instrumentation since 1972. Hundreds
of laboratories and industrial facilities rely
on our TOC analyzers for their water quality
monitoring applications.
Drinking Water
Municipal Wastewater
Industrial Wastewater
Ground Water / Surface Water
Sewage Effluent
Aurora 1030C Specifications
Operating Principle High temperature (680 C) catalytic combustion
Measurement Technique Non-dispersive infrared (NDIR) detection
Measurement Range 100 ppb C 30,000 ppm C
(multiple calibration ranges or dilution required)
Instrument Detection Limit (IDL) 50 ppb C
Operator Interface Color LCD touchscreen display with Windows
CE-based software
Operating Modes Standalone (Windows CE), PC-controlled, or LAN/
LIMS network connectivity
Basic Software Single instrument operation with data transfer to PC
Optional ATOC Software Network LAN/LIMS operation for data collection,
management and custom report generation
Autosampler 88 position rotary autosampler designed to fit
directly underneath Aurora 1030C analyzer
Sample Injection Manual syringe, sipper tube, autosampler, or multi-
stream at-line sampling module
Certification CE, EMC: EN61326 / Safety: IEC 61010-11 2001
Reagents Required Hydrochloric acid, rinsewater
Sample Injection Volume 10L 0.8 mL
Method TC 680 C Pt catalyst, 900 C non-catalyst packing
Method TIC Acidification and sparging
Method TOC NPOC, combustion of TIC-free sample, TOCTIC
Furnace Temperature Adjustable, 680 950 C in 1 C increments
Repeatability 3.0%
Accuracy 2% FS or 2% relative, whichever is greater
Sample pathway Color coded Teflon tubing
Sample handling Syringe with isolation loop to prevent contamination
Gas Supply Zero-grade air, or O2 (99.998%)
Power Supply Variable voltage, 100-240VAC, 50/60 Hz, 950W
Dimensions Aurora 1030 42.5 cm H x 49.5 cm W x 41.9 cm D
(16.75 in. H x 19.5 in. W x 16.5 in. D)
Dimensions Aurora 1030 + 1088 Autosampler 26.75 in. H x 19.5 in. W x 23 in. D
Weight Aurora 1030 + 1088 Autosampler 15.4 kg (34 lbs.), 34.5 kg (76 lbs.) 1030W + 1088

151 Graham Road Aurora 1030W Wet Oxidation


PO Box 9010 TOC Analyzer
College Station, Texas
77842-9010 The Aurora 1030W
(979) 690-1711 analyzes TOC by the
info@srainstruments.com
www.srainstruments.com

(800) 653-1711 USA/Canada heated persulfate wet


FAX (979) 690-0440 Aurora 1030W
oxidation technique. Total Organic Carbon Analyzer
www.oico.com
SOLUTION E-mail: oimail@oico.com
Via alla Castellana, 3
20063 Cernusco s/N MI
Tel. 02 9214 3258
Fax 02 9247 0901
Publication 39910413
Windows is a registered trademark of Microsoft Corp.