OUTPUT CHARACTERISTICS-ELECTROSURGICAL UNIT Electrosurgical Unit
ELECTRO MAXIMUM RATED MAX. POWER **CREST DUTY
MODE *PER
SURGICAL UNIT P-P VOLTAGE LOAD (ohms) (WATTS) FACTOR CYCLE

BIPOLAR/ Plasma Cut 700 50 300 2.5 N/A


SALINE 98 N/A
Plasma Coag 900 50 250 4.5
Optional N/A
BIPOLAR
Micro/Coag 480 100 250 2.5 N/A
Macro/Cut 98 480 100 250 1.5 N/A
475 KHz 100 250
Plasma Bipolar 295 1.5 N/A
Auto 290 100 250 1.5 N/A
Pure/Cut 1 1800 300 400 1.5 N/A
MONOPOLAR
Blend 1 98 2500 500 200 2.4 75%
CUT 3.4 60
Blend 2 2600 500 200
520KHz Endo/Gastro Cut 1500 200 250 1.8 50%

Soft/Contact 2500 150 150 4.6 25%


MONOPOLAR 97
Desiccate 2300 150 150 4.4 6.5%
COAG
Spray/Fulgurate 6000 500 150 7.2 4.6%
450 KHz
Spray 9500 500 150 10.0

* PER (Power Efficiency Rating): A measure of an electrosurgical generators ability to deliver the
selected power into a wide range of impedance.
** Crest Factor is an indicator of a waveforms' ability to coagulate bleeders without cutting effect.
TM
OUTPUT CHARACTERISTICS - RF VESSEL SEALER & RF SCALPEL (OPTIONAL)

Ligasealer Maximum Maximum Short Maximum


**Crest
Modes Open Circuit Circuit Current Power Setting Advanced, computerised monopolar "ESU" with tissue Impedance
Factor
(4MHz - 0.2 MHz) Voltage Vpp(Vp) (Amperes) (Watts) feedback technology.
Low (Seal) 340 (170) 2.1 95 1.5
Large 11'' color Intuitive Touch Screen Human Interface Energy
Medium (Fusion) 770 (400) 2.1 125 1.5
High (RF) Source with Advanced Computerised Monopolar & Bipolar tissue
580 (295) 4.2 300 1.5
RF (scalpel) impedance feedback ESU.
480 (300) 3.0 300 1.5
The Plug and play feature automatically adjusts the specific
CERTIFIED TO IEC-60601-2-2, IEC 60601-1-2 & IEC 60601-1 parameters as soon as the program is selected.

RF Ten preset programs. Additional storage for upto ninety user set
Bi
Weight and Dimensions Input Power Requirements Cl
am programs.
p
Height : 25 cm Volts : 100-270 VAC
Width : 33.5 cm Amps : 3-6A Upgradable into:-
Length : 47.0 cm Input : 800 VA, 50/60 Hz
Weight : 10 Kg. RF Advanced RF-Scalpel technology - RF-Scalpel - reusable, cost
Sc effective instruments for thermofusion and dissection, with minimal
alp
Low Frequency Leakage (50-60Hz) Audio Volume el
thermal spread.
Patient leakage < 10 uArms The mode indicator tones are
adjustable to a minimum level Rapid cutting and reliable 7mm vessel sealing from a single surgical
of 45 dba at 1 meter. instrument (RF Scalpel) for open and laparoscopic procedure.
Pl
as
m Seal can withstand maximum blood pressure of 300mm of Hg (3x
High Frequency Leakage Operating Mode a
Bi
Less than 10 mA rms., all output int 10s/30s po
lar systolic).
Re
se
cto Wet field bipolar with inbuilt irrigator (optional).
sc
op
e
DIVLABS HEALTHCARE PVT. LTD. Plasma bipolar TURP / Hysteroscopic surgeries in saline (optional).
DIVLABS (INDIA)
AN ISO 9001:2000 & ISO: 13485 COMPANY Highly cost effective: RF BiClamp is reusable and the procedure
D-455, Palam Ext., Sector-7, Dwarka, New Delhi-77 saves time and suture material.
Tel.: +91-11-25088146, 45589807 Fax: +91-11-25088146
E-mail: ceo.divlabs@gmail.com, office1.divlabs@gmail.com Web : www.divlabsindia.com INTUITIVE ENERGY

www.divlabsindia.com INTUITIVE ENERGY


Intuitive User Interface... ADVANCE FEATURES THE "PLASMA ENERGY"- TECHNOLOGY BRIEF (Optional)

10 Pre set Programs and 90 onsite programs - 100 programmable memory The Divlabs system for bipolar resection impresses by its outstanding cutting

locations. quality and maximum safety.


Different programs adjust output power and modes automatically - manual The special insulations of the bipolar system guarantee unique safety for patient
Audio Patient and user.
override also possible from the control panel.
Plate Alarm
Advanced Dual Microprocessor feedback controlled "Intuitive" technology for Enucleation Button Electrode mimics laser like performance
with visual
pop-up high frequency cutting and coagulation/hemostasis of biological tissue by The Divlabs system possesses not only an HF unit, but also a bipolar
concentrating this energy at electrode tip to achieve different surgical effects resectoscope , especially adjusted for the instrument with specic performance
like cutting, multiple blend levels, dessication, fulguration, Spray and parameters for bipolar resection. In spite of this specication, the unit can also be
vapourisation etc. used in all other surgical elds without any restrictions.
PROGRAM 2 PROGRAM 1
Laparoscopic Audio HF Leak Activation is both by footswitch and handswitch for all the different
Open Surgery
Surgery -
RF Scalpel (Default Program) Alarm with modalities - monopolar, bipolar. General advantages of the bipolar Divlabs system
visual pop-up Intuitive "Pop Up" and audio signals for foll. events: against monopolar resection (Optional)
1) Patient/neutral plate contact quality and connection
No TURP Syndrome
for utmost patient safety.
No Pulmonary edema and hyponatraemia as Saline is not toxic as compared
PROGRAM 3 2) HF Leak - for monitoring and control of Leakage
to glycine
Cancer & current.
Ablative Surgery Less conductive trauma, as the ow of the current between active and
Bipolar controls are both automatic and by footswitch.
neutral electrode is strictly limited to the target area.
Advanced Electro surgical system which provides consistent performance by
Lower risk of capsular lesion.
using latest and reproducible technology used to command the functionality
Less thermal damage to neighbouring tissues.
as the need of surgical application . It comes with 100 programs for different
PROGRAM 4
TURP/TCRE PROGRAM 6 specialities. On site program facility possible. Three different modes of cut, BIPOLAR PLASMA RESECTION
Monopolar CTVS-General
coag and bipolar can be stored individually in each program. True digital
power readout with dossage error control. Self diagnostic mode with error
OPEN SURGERY (DEFAULT PROGRAM) display (25 error messages) Spark sensor technology and RS-232 interface.
1
Multi-feature Monopolar - With uro-cut, gastro-cut and spray coagulation for
PROGRAM 5 PROGRAM 7
Plasma Saline all demanding open and underwater procedures. Reduced 0.51 mm depth Constant low voltage
CTVS - IMA complications of penetration (220320 VRMS)
Bipolar TURP/ Harvesting PPCQM - Patient to Patient plate Contact Quality Monitoring technology - result in shorter of plasma corona plasma corona used
TCRE+COBLATION 400 300 catheterization and cuts tissue and to vaporize tissue
monitors the quality and quantity of contact between the patient and the recovery times. seals vessels in with the active and
Patients can even one stroke return electrode
patient plate. It this contact is not proper the unit deactivates & alarms, go home the resulting in in close proximity.
same day excellent hemostasis.
300 300 thus eliminating any risk of Passive electrode site burns.
PROGRAM 8 TRADITIONAL MONOPOLAR RESECTION
Pump Genertor
/Irrigator-
UPGRADABLE INTO:-
Nuero + Spinal Different dedicated programs and accessories/instruments for Laparoscopic
monopolar, bipolar, vessel sealer and RF ScalpelTM (Program 1 and Program 2)
All the vessel sealing , monopolar, RF ScalpelTM and bipolar accessories of both
open and laparoscopic surgeries are of Divlabs make.
Pop-up for seal PROGRAM 10 PROGRAM 9 RF ScalpelTM Technology ensures Bipolar cut and coagulation at various levels.
''Argon Beam'' LLETZ/LEEP Tissue damage 35 mm depth Constant high voltage
completion Intuitive "Pop Up" and audio signals for foll. events: results in a longer of penetration (1,0003,000 VRMS)
+ ''Liver with ''Smoke
recovery time and causing deep required to remove
Resection'' evacuation '' 1) Re grasp- for incorrect application of Vessel sealing higher incidence coagulation of tissue and traveling
of complications. tissue and through patients body.
instrument. charring leading
2) Seal complete - Signal for the user to safely proceed to increased bleeding.

Different Foot switch Congurations with the cutter after successful sealing.
Pop-up for
improper contact/ RESPONSE OF POLAR MOLECULES
incomplete seal IN AN ELECTRIC FIELD DURING
The Complete Seal Cycle
5
DIVLABS VESSEL LIGATION Dual Microcontroller maintains flat power curve
4

Voltage/Current
ELECTRODE
Alternating 3 Fusion
tion F
Initia ormatio
Electric
2 Seal aveform
Curre
nt (I) n
1 o lt a g e (V) W Wave
form
V Power = Voltage X Current
Field
ELECTRODE
0 1 2 3 4 5 6 7 8 9 10
(RF) 2 MHz
www.divlabsindia.com INTUITIVE ENERGY
Time (in seconds)
The Power house... TRINITY - q RF
TM
OUTPUT CHARACTERISTICS-ELECTROSURGICAL UNIT Intuitive Energy Source with RF Scalpel Technology
ELECTRO MAXIMUM RATED MAX. POWER **CREST DUTY
MODE *PER
SURGICAL UNIT P-P VOLTAGE LOAD (ohms) (WATTS) FACTOR CYCLE

BIPOLAR/ Plasma Cut 700 50 300 2.5 N/A


SALINE 98 N/A
Plasma Coag 900 50 250 4.5
Optional N/A
BIPOLAR
Micro/Coag 480 100 250 2.5 N/A
Macro/Cut 98 480 100 250 1.5 N/A
475 KHz 100 250
Plasma Bipolar 295 1.5 N/A
Auto 290 100 250 1.5 N/A
Pure/Cut 1 1800 300 400 1.5 N/A
MONOPOLAR
Blend 1 98 2500 500 200 2.4 75%
CUT 3.4 60
Blend 2 2600 500 200
520KHz Endo/Gastro Cut 1500 200 250 1.8 50%

Soft/Contact 2500 150 150 4.6 25%


MONOPOLAR 97
Desiccate 2300 150 150 4.4 6.5%
COAG
Spray/Fulgurate 6000 500 150 7.2 4.6%
450 KHz
Spray 9500 500 150 10.0

* PER (Power Efficiency Rating): A measure of an electrosurgical generators ability to deliver the
selected power into a wide range of impedance.
** Crest Factor is an indicator of a waveforms' ability to coagulate bleeders without cutting effect.
TM
OUTPUT CHARACTERISTICS - RF VESSEL SEALER & RF SCALPEL (OPTIONAL)

Ligasealer Maximum Maximum Short Maximum


**Crest
Modes Open Circuit Circuit Current Power Setting Advanced, computerised monopolar "ESU" with tissue Impedance
Factor
(4MHz - 0.2 MHz) Voltage Vpp(Vp) (Amperes) (Watts) feedback technology.
Low (Seal) 340 (170) 2.1 95 1.5
Large 11'' color Intuitive Touch Screen Human Interface Energy
Medium (Fusion) 770 (400) 2.1 125 1.5
High (RF) Source with Advanced Computerised Monopolar & Bipolar tissue
580 (295) 4.2 300 1.5
RF (scalpel) impedance feedback ESU.
480 (300) 3.0 300 1.5
The Plug and play feature automatically adjusts the specific
CERTIFIED TO IEC-60601-2-2, IEC 60601-1-2 & IEC 60601-1 parameters as soon as the program is selected.

RF Ten preset programs. Additional storage for upto ninety user set
Bi
Weight and Dimensions Input Power Requirements Cl
am programs.
p
Height : 25 cm Volts : 100-270 VAC
Width : 33.5 cm Amps : 3-6A Advanced RF-Scalpel technology - RF-Scalpel - reusable, cost
Length : 47.0 cm Input : 800 VA, 50/60 Hz effective instruments for thermofusion and dissection, with minimal
Weight : 10 Kg. RF thermal spread.
Sc
alp
Low Frequency Leakage (50-60Hz) Audio Volume el
Rapid cutting and reliable 7mm vessel sealing from a single surgical
Patient leakage < 10 uArms The mode indicator tones are instrument (RF Scalpel) for open and laparoscopic procedure.
adjustable to a minimum level
of 45 dba at 1 meter. Seal can withstand maximum blood pressure of 300mm of Hg (3x
Pl systolic).
as
m
High Frequency Leakage Operating Mode a
Bi
Less than 10 mA rms., all output int 10s/30s po
lar Wet field bipolar with inbuilt irrigator (optional).
Re
se
cto Plasma bipolar TURP / Hysteroscopic surgeries in saline (optional).
sc
op
e
DIVLABS HEALTHCARE PVT. LTD. Highly cost effective: RF BiClamp is reusable and the procedure
saves time and suture material.
DIVLABS (INDIA)
AN ISO 9001:2000 & ISO: 13485 COMPANY
D-455, Palam Ext., Sector-7, Dwarka, New Delhi-77
Tel.: +91-11-25088146, 45589807 Fax: +91-11-25088146 INTUITIVE ENERGY
E-mail: ceo.divlabs@gmail.com, office1.divlabs@gmail.com Web : www.divlabsindia.com
www.divlabsindia.com INTUITIVE ENERGY
Intuitive User Interface... ADVANCE FEATURES THE "PLASMA ENERGY"- TECHNOLOGY BRIEF (Optional)

10 Pre set Programs and 90 onsite programs - 100 programmable memory The Divlabs system for bipolar resection impresses by its outstanding cutting

locations. quality and maximum safety.


Different programs adjust output power and modes automatically - manual The special insulations of the bipolar system guarantee unique safety for patient
Audio Patient and user.
override also possible from the control panel.
Plate Alarm
Advanced Dual Microprocessor feedback controlled "Intuitive" technology for Enucleation Button Electrode mimics laser like performance
with visual
pop-up high frequency cutting and coagulation/hemostasis of biological tissue by The Divlabs system possesses not only an HF unit, but also a bipolar
concentrating this energy at electrode tip to achieve different surgical effects resectoscope , especially adjusted for the instrument with specic performance
like cutting, multiple blend levels, dessication, fulguration, Spray and parameters for bipolar resection. In spite of this specication, the unit can also be
vapourisation etc. used in all other surgical elds without any restrictions.
PROGRAM 2 PROGRAM 1
Laparoscopic Audio HF Leak Activation is both by footswitch and handswitch for all the different
Open Surgery
Surgery -
RF Scalpel (Default Program) Alarm with modalities - monopolar, bipolar. General advantages of the bipolar Divlabs system
visual pop-up Intuitive "Pop Up" and audio signals for foll. events: against monopolar resection (Optional)
1) Patient/neutral plate contact quality and connection
No TURP Syndrome
for utmost patient safety.
No Pulmonary edema and hyponatraemia as Saline is not toxic as compared
PROGRAM 3 2) HF Leak - for monitoring and control of Leakage
to glycine
Cancer & current.
Ablative Surgery Less conductive trauma, as the ow of the current between active and
Bipolar controls are both automatic and by footswitch.
neutral electrode is strictly limited to the target area.
Advanced Electro surgical system which provides consistent performance by
Lower risk of capsular lesion.
using latest and reproducible technology used to command the functionality
Less thermal damage to neighbouring tissues.
as the need of surgical application . It comes with 100 programs for different
PROGRAM 4
TURP/TCRE PROGRAM 6 specialities. On site program facility possible. Three different modes of cut, BIPOLAR PLASMA RESECTION
Monopolar CTVS-General
coag and bipolar can be stored individually in each program. True digital
power readout with dossage error control. Self diagnostic mode with error
OPEN SURGERY (DEFAULT PROGRAM) display (25 error messages) Spark sensor technology and RS-232 interface.
1
Multi-feature Monopolar - With uro-cut, gastro-cut and spray coagulation for
PROGRAM 5 PROGRAM 7
Plasma Saline all demanding open and underwater procedures. Reduced 0.51 mm depth Constant low voltage
CTVS - IMA complications of penetration (220320 VRMS)
Bipolar TURP/ Harvesting PPCQM - Patient to Patient plate Contact Quality Monitoring technology - result in shorter of plasma corona plasma corona used
TCRE+COBLATION 400 300 catheterization and cuts tissue and to vaporize tissue
monitors the quality and quantity of contact between the patient and the recovery times. seals vessels in with the active and
Patients can even one stroke return electrode
patient plate. It this contact is not proper the unit deactivates & alarms, go home the resulting in in close proximity.
same day excellent hemostasis.
300 300 thus eliminating any risk of Passive electrode site burns.
PROGRAM 8 Different dedicated programs and accessories/instruments for Laparoscopic TRADITIONAL MONOPOLAR RESECTION
Pump Genertor TM
/Irrigator- monopolar, bipolar, vessel sealer and RF Scalpel (Program 1 and Program 2)
Nuero + Spinal All the vessel sealing , monopolar, RF ScalpelTM and bipolar accessories of both
open and laparoscopic surgeries are of Divlabs make.
RF ScalpelTM Technology ensures Bipolar cut and coagulation at various levels.
Intuitive "Pop Up" and audio signals for foll. events:
Pop-up for seal PROGRAM 10 PROGRAM 9 1) Re grasp- for incorrect application of Vessel sealing
''Argon Beam'' LLETZ/LEEP Tissue damage 35 mm depth Constant high voltage
completion instrument. results in a longer of penetration (1,0003,000 VRMS)
+ ''Liver with ''Smoke
recovery time and causing deep required to remove
Resection'' evacuation '' 2) Seal complete - Signal for the user to safely proceed higher incidence coagulation of tissue and traveling
of complications. tissue and through patients body.
with the cutter after successful sealing. charring leading
to increased bleeding.
Compatible with Da Vinci Robtic System CUSA and Argon Beam Coagulator.
Different Foot switch Congurations
Pop-up for
improper contact/ RESPONSE OF POLAR MOLECULES
incomplete seal IN AN ELECTRIC FIELD DURING
The Complete Seal Cycle
5
DIVLABS VESSEL LIGATION Dual Microcontroller maintains flat power curve
4

Voltage/Current
ELECTRODE
Alternating 3 Fusion
tion F
Initia ormatio
Electric
2 Seal aveform
Curre
nt (I) n
1 o lt a g e (V) W Wave
form
V Power = Voltage X Current
Field
ELECTRODE
0 1 2 3 4 5 6 7 8 9 10
(RF) 2 MHz
www.divlabsindia.com INTUITIVE ENERGY
Time (in seconds)
CREATION - Plasma Bipolar (PB)
EP T O F TECHNICAL SPECIFICATIONS :- Trinity (220-240V)
T rinity-THE CONC PROTECTION-Effectice heamostasis
DESTRUCTION- EFFORTLESS CUT AND ABLATION
Weight and Dimensions Input Power Supply
Weight : 10 Kg. Volts : 180-270 VAC
Length : 390mm Current Fuse : 6Amps

Trinity Electrosurgical Unit Height : 130mm


Width : 340mm
Power consumption-750 VA Max.
(Inbuilt line filter)
Output Controls
Current Leakage All digital and push switch controls
With programable micro controller assisted quick response technology in all modes
High Frequency Leakage:
<150 mA rms., all output Audio Volume
Low Frequency Leakage (50-60Hz): All isolated tone indications to differenciate
the mode of application and adjustable
< 150 Arms in 4 modes
PPCQM System
Measuring Frequency-140 Khz. Cooling Convection
PPCQM pad-0.135 ohms
Certified to IEC-60601-2-2 and IS-7583
Non-PPCQM pad- not< 10ohms
Protection Class-1. UnitType-CF.
Output - All floating Output.
OUTPUT CHARACTERISTICS CE Certified.

OUTPUT MAXIMUM RATED MAX. CREST PER* DUTY


MODE FREQ. P-P LOAD POWER FACTOR* CYCLE
(kHz) VOLTAGE (ohms) (WATTS)

Micro/Coag 500 380 100 99 2.5 98 N/A


BIPOLAR Macro/Cut 500 480 100 99 1.5 N/A
Plasma 700 T.D T.D 150 T.D N/A
Pure/Cut 1 500 1800 300 400 1.5 98 N/A
MONOPOLAR Blend/Cut 2 500 2600 500 200 2.4 75%
CUT Endo/Gastro Cut 500 1500 200 250 1.8 50%
Soft/Desiccate 450 2300 150 150 4.4 97 25%
Advanced Electro surgical system which provides consistent performance by using latest and reproducible " CONSOLE - WITH MONOPOLAR
Forced/Fulgurate 450 6000 500 150 7.2 6.5%
COAG
technology used to command the functionality as the need of surgical application . It comes with MOBILE CART " Spray 450 9500 500 150 8.0 4.6%
ten programs for different specialities. On site program facility possible. Three different modes of cut,
* Crest Factor is an indicator of a waveforms' ability to coagulate bleeders without cutting effect.
coag and bipolar can be stored individually in each program. True digital power readout with dossage * PER (Power Efficiency Rating): A measure of an electrosurgical generators ability to deliver the selected power
error control. Self diagnostic mode with error display (25 error messages) Spark sensor technology into a wide range of impedance.

and RS-232 interface.


ACCESSORIES
Advanced Performance Features Advanced Safety Features Enhanced Operating Convenience
STANDARD:-
Monopolar-Computer Controlled intelligent 1 PPCQM- Patient to Patient plate 1 Separate Bipolar generator controls and
cut with Feedback Response System contact Quality monitoring footswitch with auto selectivity reduces
the dependency on the operating room D-004 : Monopolar Handle with
technology - monitors the quality &
1 Three Cut Modes - (Auto Cut-Optional) staff during the procedure. cable
quantity of contact between the patient
(a) Pure cut suitable for under water
& the patient plate. If this contact is not 2 Two independent handswitching
cutting (TURP, TUVP & TCRE) with D-005 : Monopolar twin button
proper the unit deactivates & alarms, monopolar outputs and one
URO-CUT, which eliminates lag Handle with cable
thus eliminating any risk of Passive handswitching bipolar output combined LAP-CLIPON-
time in initiating cut especially suited (Hand switching pencil)
for urological procedure. electrode site burns. (Optional) with Laproscopic Clipons, eleminates MONOPOLAR
(b) Three Blend - modes (program) 2 It deactivates the unit & alarms in the need of footswitches during D-006 : Patient Plate with cable
(c) Low Cut- Endocut (Gastrocut) - case of any discontinuity in the patient endoscopic procedures.
for G.I. Surgery plate cord. Exclusive PEM (Patient D-006 : Bipolar forceps with cable
3 Ergonomically designed trolley to
2 Electrode monitoring)
Three Coag Modes - (Auto Coag-Optional) house diathermy, accessories, argon
(a) Desiccation/Soft - for low voltage 3 Isolated RF output directs current D-009: Monopolar Electrode 4 Nos
to return back to the generator to
unit with cylinders and smoke evacuator.
soft coagulation suitable for laproscopic
& delicate tissue work. reduce the risk of Alternate site 4 Comprehensive clinical training network D-007 : Double pedal footswitch
(b) Fulgurate/Forced - non contact burns & to ensure patient safety in through workshops, seminars, videos & for Monopolar
Coagulation for minimal smoke generation case of any discontinuity in patient literatures.
(c) Spray for non-contact large tissue plate cord. This current has no UPGRADATIONS D-010 : Single pedal foot switch
area coagulation with minimum depth reference to ground so it does not for Bipolar
cause notice able interference with 1 'Vessel Sealing System' both for open and
of necrosis.
the ECG monitor, Cardiac pace Laparscopic Surgeries - 'TRINITY-LS.'
maker etc. 2 Upgradation into Dual Generator System OPTIONAL
Bipolar-SLS (Computer controlled 4 Independent activation of hand for C.T.V.S. Surgeries with specialised
Self Limiting System-Unmodulated-Auto Optional) switching, foot switching & bipolar IMA-program (Internal Mammary Artery D-011 : Tripple pedal footswitch
1 Micro - for delicate tissue work by output to protect against Harvesting)-and vessel sealing System for
inadvertent output from an idle Endo Vessel Harvesting 'TRINITY-CT' D-021 : Mobile Cart
controlled desiccation without charring,
adhesion or blanching of adjacent tissue. accessory lying on or near the 3 Easly upgradable into Argon-Beamer
patient or surgical staff.
D-008 : Bayonet Bipolar forceps with
(opthalmic & nuero-surgery). with DIV-ARGON Module cable.
2 Macro- for new generation of macro 5 Class '2' leakage cancelling, 4 Compatible with DIV-VAC smoke evacuator
bipolar cutting instruments (Tri-polar) a must for lengthy procedure like
5 Upgradable with DIV-BCM Bipolar
for LAVH procedure etc. When used with Cardiac surgery (CABG) etc.
current monitor DIVLABS (India) - An ISO-13485 Company Marketed by :
BIPOL probes provides excellent 6 Meets all general requirements
tamponade capabilities. of Safety for Electromagnetic 6 Compatible with DIV-AES Active electrode
safety system
DIVLABS HEALTHCARE PVT. LTD.
compatibility (EMC) as per D-455, Palam Ext., Sector-7, Dwarka, New Delhi - 110077 (India)
3 Plasma Bipolar (PB) - Advanced IEC-60601-1-2 including 36.201 7 Compatible with DIV-RFTA Radio frequency
Tel.: +91-11-25088146, 45589807, Fax: +91-11-25084483
computerised wet field Bipolar works ( CISPR 11 Group 1) - Emissions, thermal ablation
8 Used with CUSA CEM Module E-mail: ceo.divlabs@gmail.com, officie1.divlabs@gmail.com
under blood, saline etc. 36.202-Immunity and 36.202.1-General.
Web: www.divlabsindia.com, www.ligasealer.in www.rfscalpel.com
9 Under Saline Biplolar TURP-TRINITY-NaCl Specifications subject to change without notice.