CDS Boule Con-Diff Multi-Parameter Assayed Hematology Control

The picture can't be display ed.

Control LOT
Tri-Level 21705-5K 09/28/17 Open vial stability 14 days
Low 21705-51 09/28/17 CA
Normal 21705-52 09/29/17
High 21705-53 09/29/17

MEDONIC
Abnormal low lot: 21705-51 Normal lot: 21705-52 Abnormal high lot: 21705-53
CA620/530 Mean Range Mean Range Mean Range
12 2.17 0.12 4.02 0.18 4.60 0.22
RBC 10 /L
MCV fL 62.1 5.5 79.3 5.5 85.8 5.5

RDW % 14.2 4.0 13.8 4.0 13.2 4.0

Hct % 13.5 2.5 31.9 3.0 39.5 3.5

Plt * 109/L 87 15 264 30 555 60

MPV fL 10.4 1.5 10.2 1.5 10.5 1.5

WBC 109/L 3.4 0.3 8.8 0.6 20.8 1.8

Hgb g/dl 5.1 0.3 11.5 0.4 14.2 0.5

MCH pg 23.5 1.5 28.6 1.5 30.9 1.5

MCHC g/dl 37.8 3.5 36.1 3.5 35.9 3.5

Lymph# 109/L 1.5 0.4 3.9 0.6 12.9 1.5

Mid# 109/L 0.2 0.3 0.6 0.5 1.3 1.0

Gran# 109/L 1.7 0.8 4.3 1.0 6.6 2.5

Lymph % 43.2 7.0 44.1 5.0 62.2 5.0

Mid % 5.7 5.0 6.7 5.0 6.1 5.0

Gran % 51.1 7.0 49.2 5.0 31.7 5.0

Mseries
M-series Mean Range Mean Range Mean Range
12 2.16 0.12 4.03 0.18 4.66 0.22
RBC 10 /L
MCV fL 63.6 5.5 81.6 5.5 89.1 5.5

RDW % 14.3 4.0 13.7 4.0 12.9 4.0

Hct % 13.7 2.5 32.9 3.0 41.5 3.5


9 78 15 241 30 509 60
Plt * 10 /L
MPV fL 10.6 1.5 10.2 1.5 10.3 1.5

WBC 109/L 3.4 0.3 8.5 0.6 21.0 1.8

Hgb g/dl 5.0 0.3 11.6 0.4 14.5 0.5

MCH pg 23.1 1.5 28.8 1.5 31.1 1.5

MCHC g/dl 36.5 3.5 35.3 3.5 34.9 3.5

Lymph# 109/L 1.4 0.4 3.7 0.6 13.0 1.5


9 0.2 0.3 0.5 0.5 1.2 1.0
Mid# 10 /L
9 1.8 0.8 4.3 1.0 6.8 2.5
Gran# 10 /L
Lymph % 42.0 7.0 43.3 5.0 61.7 5.0

Mid % 5.4 5.0 6.3 5.0 5.7 5.0

Gran % 52.6 7.0 50.4 5.0 32.6 5.0

*"FD" Flags may be ignored on control results providing the flagged parameter
recovers within the expected range.

Produced for Boule Medical AB


by Clinical Diagnostic Solutions
th
1800 NW 65 Avenue
Plantation, FL 33313

PN 201078E R11.04.16
For CA620 with barcode reader
1. Go to Menu 7.4.1

2. Scan in the calibrator/control in sequence top to bottom

This bar-code sheet is also available at :


www.medonic.se

202290B R12.19.14
For Medonic M-series with barcode reader
1. In the Main Menu go to Q/C Menu Enter Con/Cal Control/Calibrator Assay Value Input

2. Scan in each blood calibrator/control column by column, sequence 1 to 9