Vous êtes sur la page 1sur 13

ORIGINAL ARTICLE INTERNATIONAL JOURNAL OF LA BO RATO RY HEMATOLOGY

Can automated blood film analysis replace the manual


differential? An evaluation of the CellaVision DM96
automated image analysis system
C. BRIGGS*, I. LONGAIR*, M. SLAVIK †, K. THWAITE †, R. MILLS*, V. THAVARAJA*, A. FOSTER ‡,
D. ROMANIN ‡, S. J. MACHIN*

*Department of Haematology, SUMMARY


University College London
Hospital, London, UK Automation of differentials is desirable for economic and time-saving

Department of Haematology, The reasons. Over the last 20 years, automated imaging processes have
Doctors Laboratory, London, UK

Sysmex UK Ltd., Sysmex House,
started to be introduced where stained blood films are scanned by a
Garamonde Drive, Wymbush, computer-driven microscope and leucocytes classified; however, early
Milton Keynes, UK methods were slow and had difficulty in classifying abnormal cells.
More recently the CellaVision DM96 (CellaVision AB, Lund, Swe-
Correspondence:
Carol Briggs, Department of Hae- den) has been introduced with added features such as continuous
matology, University College Lon- loading of slides and a faster throughput than previous instruments.
don Hospital, 60 Whitfield Street, The accuracy of CellaVisionTM DM96 has been evaluated by comparing
London W1T 4EU, UK.
results to reference manual differentials. Results from different oper-
Tel.: +44 207 3809882;
E-mail: carolbriggs@hotmail.co.uk ators using the DM96 were compared with their own manual
differential and to a 400-cell reference manual differential. Precision
doi:10.1111/j.1751-553X.2007.01002.x of the instrument was compared to the manual differential. The
Received 26 June 2007; accepted
preclassification accuracy of the DM96 was 89.2%. Precision was
for publication 2 October 2007 similar to that of the 100-cell manual differential. The DM96 was faster
than the manual method, even after reclassification by a laboratory
Keywords scientist of any cells wrongly categorized by the instrument. The DM96
Leucocyte differential, microscopy, accuracy in morphological classification of leucocytes and red blood
automated image analysis
cells; depends upon both blood pathology and experience of the
laboratory scientist using the instrument. For some cell types and
operators, DM96 accuracy was better than the individual’s 100 cell
manual differential.

Staff members represent the major expenditure in


INTRODUCTION
most laboratories so in many places where their num-
Over recent years, there has been increasing demand bers have not increased or have even been reduced,
for haematological tests with clearly defined turn- there is now more work for fewer laboratory scien-
around times along with a need for cost containment. tists. Fortunately, there have been advances in blood
 2007 The Authors
48 Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 48–60
C. BRIGGS ET AL. EVALUATION OF THE CELLAVISION DM96 49

cell counting instrumentation and this has resulted in In this study, we have evaluated the accuracy and
a reduction in the need for manual blood film review precision of the DM96 in identifying all classes of
and white blood cell (WBC) differentials (Lewis & leucocytes including abnormal and immature cells. The
Rowan, 1991). The examination of blood films by results from five different laboratory scientists’ manual
microscopy remains one of the major labour-intensive differentials and those from the DM96 have been com-
procedures in the laboratory and the challenge is to pared with a standard reference manual differential
reduce the number of films examined without missing (Clinical and Laboratory Standards Institute H20-A2,
important diagnostic information. Automated blood 2007) and the time taken for these slides to be analysed
cell counters offer a leucocyte count, red cell and was recorded. An evaluation of the red cell morphology
platelet count and five-part (some six-part) leucocyte from the DM96 has also been undertaken.
differential. In addition, some instruments provide a For any automated image system to be introduced
nucleated red blood cell (NRBC) count. Haematology in the haematology laboratory, it will have to demon-
instrument differentials provide only limited informa- strate that it is at least as reliable, or preferably more
tion on cell morphology using abnormal cell flags and reliable, in correctly identifying WBC as the routine
are often unable to reliably classify abnormal and manual method and should also provide significant
immature cells. A blood film is examined for a num- labour savings and so costs to the laboratory.
ber of reasons: to explain an unexpected blood count,
to examine red cell and platelet morphology, to con-
MATERIALS AND METHODS
firm an abnormal automated leucocyte count or to
undertake an extended differential including imma-
Blood samples
ture and abnormal cells.
The examination of blood films is not only time Residual patient blood samples, after all requested hae-
consuming and labour intensive but it also requires matology tests had been completed, were obtained
highly trained staff. The impact of a wrong diagnosis from the routine workload of the Department of
necessitates that experienced staff are present in the Haematology at University College London Hospital.
laboratory 24 h a day. However, manual cell classifi- Samples were collected into K2EDTA as anticoagulant
cation is subjective, with significant inter- and intra- (Becton Dickinson, Franklin Lakes, NJ, USA). After
observer variation (Koepke, Dotson & Shifmann, collection samples were stored at room temperature,
1985) and any count is also subject to significant sta- blood films were prepared and stained using an auto-
tistical variance (Rumke, 1985a). mated slide maker and stainer (SP-100, Sysmex, Kobe,
Automated morphological analysis systems have Japan). For the evaluation 136 samples were selected,
been introduced in the past, the first being described these included 45 normal samples, all parameters
in 1966 (Prewitt & Mendlesohn, 1966). These earlier within our laboratory reference ranges, and the rest
systems were slow, offering a limited degree of auto- consisting of various clinical conditions such as infec-
mation and therefore failed to provide significant tions, chronic and acute leukaemais, lymphomas, hea-
improvements in workflow and counting statistics moglobinopathies, iron deficiency, renal disease, liver
(Bentley, 1990). disease, HIV and chemotherapy. The WBC in these
More recently the CellaVision DM96 (CellaVision samples ranged from 0.51 · 109/l to 73.89 · 109/l. All
AB, Lund, Sweden) has been introduced. The instru- samples included in this study had been analysed on
ment scans the slides at low power to identify poten- the Sysmex XE-2100 to ascertain whether the results
tial WBCs and then takes digital images at high were normal or abnormal and to collect the WBC in
magnification. The images are analysed by an artificial order to calculate the absolute values for the leuco-
neural network based on a database of cells and pre- cytes from the DM96 percentage results. For the preci-
classified according to WBC class. The cells are pre- sion study, 10 samples were selected with varying
sented on a computer screen for conformation or WBC counts, both normal and abnormal samples were
reclassification. The system also allows review of the included. For the comparison of results between oper-
red cell and platelet morphology and estimation of the ators 30 samples were selected, again normal and
platelet count. abnormal samples were included.
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 48–60
50 C. BRIGGS ET AL. EVALUATION OF THE CELLAVISION DM96

user definable computer display (Figure 1). The opera-


Reference method
tor can enlarge single cells for a more detailed view,
The reference method was the manual differential leave cells in the category suggested, accept the pre-
count of 200 cells performed by two experienced labo- classification or move individual cells to a different
ratory scientists (Clinical and Laboratory Standards category, reclassify, by clicking on the cell and using
Institute H20-A2, 2007) using light microscopy, if the drag and drop function. Results from the DM96
there were discrepancies between the two results a are not complete and will not be released until all
third 200-cell differential was performed by a different preclassified cell categories have been reviewed. To
laboratory scientist. conform with the way our laboratory reports WBC
differentials the band cell count was added to the
segmented neutrophil count (band cells are not
Analysis with CellaVision DM96
counted separately) and the variant lymphocyte count
The DM96 consists of a slide scanning unit and a com- and plasma cell count were added to the total lympho-
puter with the CellaVision Blood (CellaVision, AB, cyte count, atypical lymphocytes and plasma cells are
Lund, Sweden) differential Software (version 1.5). The commented on, but not quantitated in the routine
scanning unit consists of a motorized microscope (·10, manual differential. For the evaluation of the red cell
·50 and ·100 objective), a digital CCD camera, an and platelet morphology, the DM96 provides an image
automatic immersion oil unit, slide feeder unit with which corresponds to eight areas of the smear viewed
barcode reader, rack feeder unit and a control unit at ·100 objective. The red cell and platelet images can
which controls motors, sensors, oil applying and illu- be enhanced by clicking on the screen to magnify the
mination. Slides are loaded onto the system and pro- cells. Red blood cell morphology is partially classified
cessed immediately and continuously. The instrument by the DM96; polychromasia, hypochromia, micro-
scans the slides, identifies potential WBCs, takes digital cytosis, macrocytosis, anisocytosis and poikilocytosis
images of them and uses artificial neural network- are reported using a scale 0–3. Abnormalities in red
based software to analyse the cells. Digital images of cell shape, such as fragmented red cells or sickle cells
preclassified cells are presented to the operator on a are not automatically classified by the DM96 but can

Figure 1. Preclassified white blood cells presented on the CellaVision DM96 computer screen.

 2007 The Authors


Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 48–60
C. BRIGGS ET AL. EVALUATION OF THE CELLAVISION DM96 51

be added by the user after examining the image pre- pared with the results for the preclassified and then
sented by the instrument. The default settings for the their reclassified cells from the DM96, additionally all
number of cells exhibiting an abnormality and being results were compared with the reference 400-cell dif-
classified as normal or 1–3 severity can be adjusted by ferential. The operators were also asked to evaluate
the user. An estimate of platelet count can be made by whether the areas presented by the DM96 for red
the DM96 if this feature is selected. The image is then blood cell and platelet morphology were adequate.
subdivided into sub-images and the user is then
required to count the number of platelets in each field
Timing studies
(or sub-image), a calculated estimate of the platelet
concentration is then made based on the counts per The same 30 blood films and five laboratory scientists
field and a platelet estimate factor. This feature was were used for the timing study as for the accuracy
not selected or evaluated in this study. study above. The time taken to perform the total 30
100-cell differentials manually and on the DM96 from
loading the slides onto the instrument to the end of
Accuracy
reclassifying the cells was noted for each individual.
One hundred and thirty-six samples were analysed on
the DM96 and by a 100-cell differential, which is our
Precision
established routine practice, by an experienced labora-
tory scientist. WBC counts ranged from 0.51 · 109/l Ten samples with varying WBC counts, normal sam-
to 73.89 · 109/l. The operator was not provided with ples and samples containing abnormal cells were
any previous or clinical information on the patient or selected. WBC counts ranged from 1.62 · 109/l to
results from the haematology analyser. The DM96 32.87 · 109/l. Each blood film was analysed on the
was set to count 105 cells; this was to allow for the DM96 five times and by the same five laboratory sci-
possible inclusion of artefacts in the WBC count. entists used for the above accuracy study.
However, in samples with low WBC counts 100-cell
counts were not always achieved on the DM96. The
Statistical analysis
accuracy of red blood cell analysis on the DM96 was
also assessed by comparing morphological changes Statistical analysis was performed using Microsoft Excel
with those reported by the microscopic method. (Microsoft Ltd., Reading, UK) and MINITABTM STATISTICAL
SOFTWARE (Minitab Ltd., Coventry, UK) (www.minitab.
com). Multiple comparisons between the different
Accuracy of results when the DM96 is operated by
WBC differential methods and by the different opera-
different laboratory scientists
tors performing them were assessed using Friedman’s
Laboratory scientists were selected with different lev- ANOVA test. Specific differences were shown using
els of experience in blood cell morphology. Two were Mann–Whitney test for paired data with a P-value
very experienced, one moderately experienced and <0.05 considered statistically significant.
two newly qualified members of staff who perform
differentials routinely in the laboratory, but are still
RESULTS
expected to need some help with difficult blood films.
For this study, the operators were not provided with
Accuracy
any previous or clinical information on the patient or
results from the haematology analyser. Thirty blood The DM96 monitor shows leucocytes preclassified
films were first analysed on the DM96 and then by a according to cell type. The operator has to review
100-cell microscopic differential. The samples con- each cell and can move a cell to a different cell class
sisted of 10 from normal patients and the remaining (reclassify the cell) or leave the cell in the category
with abnormal WBCs and red blood cells present, suggested by the instrument. Table 1 shows the
WBC count range 0.51–51.63 · 109/l. The 100-cell percentage agreement of the various cell classes
manual differentials for each individual were com- preclassified correctly by the DM96. The percentage of
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 48–60
52 C. BRIGGS ET AL. EVALUATION OF THE CELLAVISION DM96

100-cells on 16 (11.8%) of samples, this was usually


Table 1. Percentage of cells from 286 blood films
correctly preclassified by the CellaVision DM96 when the total WBC count was <1.5 · 109/l but also
occurred if there was more than the usual amount of
Preclassifying artefact on the slide. All manual differentials counted
Cell class agreement (%) 100 cells. Pre- and postclassification results for the red
blood cells are presented in Table 2. Preclassification
Neutrophil (Neut) 99.5
Lymphocyte (Lymph) 94.9 agreement with the microscopic morphology for aniso-
Monocyte (Mono) 87.6 cytosis and macrocytosis is low; this is because of a
Eosinophil (Eos) 79.9 high number of false positives reported by the DM96.
Basophil (Baso) 54.1 Classification of the red cells by the DM96 failed for
Metamyelocyte 32.6 one sample and for seven samples (5%) the operator
Myelocyte 37.7
reported the cells presented by the DM96 as inade-
Promyelocyte 77.6
Blast 76.6 quate for morphological comment. All blood films
Nucleated red blood cell 89.6 were adequate when viewed under the microscope.
Neut, Lymph and Mono 97.3
Neut, Lymphs, Mono, Eos and Baso 87.2
All cell classes 89.2 Accuracy of results when the DM96 is operated by five
Abnormal cells called normal 0.9 different laboratory scientists
Normal cells misclassified as 9.1
other normal cells The correlation coefficients of determination (r2) from
Normal cells called abnormal cells 1.8 30 blood films analysed on the DM96 by five labora-
tory scientists are presented in Table 3. Each individ-
ual’s manual differential result as well as the results
correctly classified cells was highest for the mature for the preclassified and reclassified cells from the
cells and lowest for the immature granulocytes. NRBC DM96 are compared with the reference 400-cell
preclassification was good. Reclassification of the cells differential.
was performed by the same person who performed Correlation to the reference method is similar for
the 100-cell microscopic differential counts. an operator whether the differential is performed
Assessment of the accuracy of results was carried manually or by the DM96 with reclassification of cells.
out by determining the correlation coefficient (r2) Operators 1 and 2 were experienced morphologists,
between one experienced laboratory scientist’s manual operator 3 had moderate experience and 4 and 5 were
differential and the pre- and reclassified results from newly qualified. For some cell types and for some
the DM96 for all cell classes (Figure 2). Because of the operators, the accuracy of the DM96 compared with
low number and subjectivity of classification of met- the reference method was better than that of the indi-
amyelocytes, myelocytes and promyelocytes the results vidual’s 100-cell manual differential. Correlation with
for these cells are also demonstrated as a total imma- the reference method for the lymphocyte count from
ture granulocyte count. The correlation coefficients the reclassified results from the DM96 is very poor for
were higher after the cells had been reclassified. R2 operator 4. One sample contained 90% blasts which
values were greater than 0.9 for neutrophils (including the DM96 preclassified as lymphocytes, operator 4
band cells), lymphocytes, total immature granulocytes, failed to reclassify these cells while all other operators
blasts, NRBC and metamyelocytes. Correlation for the correctly reclassified the cells. If the results from this
reclassified monocytes was 0.81 and eosinophils slide are removed, correlation for lymphocytes by
0.67.There is one significant outlier on the preclassi- operator 4 improves to a r2 of 0.61 Correlation for the
fied vs. manual differential monocyte graph, the DM96 eosinophils on the DM96 are not as good as the man-
preclassified myeloblasts in this samples as monocytes. ual method even after reclassification.
By removing this data point, correlation is then Statistically the preclassification results for lympho-
improved to 0.77. Because of the low number of cells cytes, immature granulocytes, myelocytes and to a les-
counted, correlation for basophils, myelocytes and ser extent basophils had significantly higher values
promyelocytes was poorer. The DM96 failed to count than those by the reference method, reclassified results
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 48–60
C. BRIGGS ET AL. EVALUATION OF THE CELLAVISION DM96 53

Man v DM96 reclassified neutrophils Man v DM96 neutrophils


60 60
y = 1.0079x + 0.2756

DM 96 neutrophils x 109/l
50 y = 0.9923x + 0.3026 50 R 2 = 0.9507
neutrophils x 109/l
R 2 = 0.9859
DM96 reclassified

40 40

30 30

20 20

10 10

0 0
0 10 20 30 40 50 60 0 10 20 30 40 50 60
Manual neutrophils x 109/l Manual neutrophils x 109/l

Man v DM96 reclassified lymphocytes 40 Man v DM96 lymphocytes


35
35 y = 1.0597x + 0.8266

DM96 lymphocytes x 109/L


y = 0.9706x - 0.1226
30 R 2 = 0.7267
R 2 = 0.9591
lymphocytes x 109/l

30
DM96 reclassified

25
25
20
20
15 15
10 10
5 5
0 0
0 5 10 15 20 25 30 35 0 5 10 15 20 25 30 35 40
Manual lymphocytes x 109/l Manual lymphocytes x 109/l

Man v DM96 reclassified monocytes 45 Man v DM96 monocytes


14
40
12
DM96 monocytes x 109/l

y = 0.8529x - 0.012 35
monocytes x 109/l
DM96 reclassified

10 R 2 = 0.805 30
y = 0.8393x + 0.363
8 25
R 2 = 0.1164
6 20
15
4
10
2
5
0 0
0 2 4 6 8 10 12 14 0 2 4 6 8 10 12 14
Manual monocytes x 109/l Manual monocytes x 109/l

Figure 2. Correlation of cell classification on the CellaVision DM96 with the 100-cell manual microscopic method.
Man vs. DM96: manual microscopic method vs. the preclassified results from the DM96; Man vs. DM96 reclassified:
manual microscopic method vs. the results from the DM96 after reclassification of cells; Imm Gran, immature gran-
ulocytes, the sum of metamyelocytes, myelocytes and promyelocytes.

or each operator’s manual differential. The DM96 has a For reclassified data, far fewer statistically signifi-
tendency to wrongly classify other cells into these cell cant differences were identified between the specific
classes. users’ manual and DM96 WBC differentials when
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 48–60
54 C. BRIGGS ET AL. EVALUATION OF THE CELLAVISION DM96

Man v DM96 reclassified eosinophils Man v DM96 eosinophils


3.0 3.0
y = 0.8044x + 0.1154

DM96 eosinophils x 109/l


eosinophils x 109/l 2.5 y = 0.997x + 0.0415 2.5 R 2= 0.4973
DM96 reclassified

R 2 = 0.672
2.0 2.0

1.5 1.5

1.0 1.0

0.5 0.5

0.0 0.0
0.0 0.5 1.0 1.5 2.0 2.5 3.0 0.0 0.5 1.0 1.5 2.0 2.5 3.0
Manual eosinophils x 109/l Manual eosinophils x 109/l

Man v DM96 reclassified basophils Man v DM96 basophils


1.2 4.0
y = 0.3346x + 0.0379 3.5
1.0 y = 0.0379x + 0.1292
R 2 = 0.0534 DM96 basophils x 109/l
DM96 reclassified

3.0
basophils x 109/l

R 2 = 0.00
0.8
2.5
0.6 2.0

0.4 1.5
1.0
0.2
0.5
0.0 0.0
0.0 0.2 0.4 0.6 0.8 1.0 1.2 0.0 0.2 0.4 0.6 0.8
Manual basophils x 109/l Manual basophils x 109/l

Man v DM96 reclassified Imm Gran Man v DM96 Imm Gran


12.0 y = 1.5927x + 0.1427
y = 1.2512x + 0.0363 14
R 2 = 0.9129
10.0 R 2= 0.9514
DM96 Imm Gran x 109/l

12
DM96 reclassified
Imm Gran x 109/l

8.0 10
8
6.0
6
4.0
4
2.0
2
0.0 0
0.0 2.0 4.0 6.0 8.0 10.0 12.0 0 2 4 6 8 10 12 14
Manual Imm Gran x 109/l Manual Imm Gran x 109/l

Figure 2. Continued.

compared with the reference method. The greatest counted. For lymphocytes, the median range of
number of statistically significant differences occurred reclassified cells on the DM96 is 1.20–1.42 · 109/L,
again with lymphocytes and basophils, for basophils for the manual differentials 1.07–1.52 · 109/l and for
this may be influenced by the low number of cells the reference method the count is 1.42 · 109/l. The
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 48–60
C. BRIGGS ET AL. EVALUATION OF THE CELLAVISION DM96 55

Man v DM96 reclassified blasts Man v DM96 blast


80 25
y = 0.2912x + 0.2572
70
R 2 = 0.6362
y = 0.9937x + 0.001 20

DM96 blasts x 109/l


60
DM96 reclassified

R 2 = 0.9953
blasts x 109/l

50
15
40
30 10

20
5
10
0 0
0 10 20 30 40 50 60 70 80 0 10 20 30 40 50 60 70 80
Manual blasts x 109/l Manual blasts x 109/l

Man v DM96 reclassified NRBC Man v DM96 NRBC


300 300

250 y = 1.0895x - 2.8732 250 y = 1.0142x - 2.9026

DM96 NRBC/100WBC
R 2= 0.973 R 2 = 0.9636
DM96 reclassified
NR BC/100WBC

200 200

150 150

100 100

50 50

0 0
0 50 100 150 200 250 300 0 50 100 150 200 250 300
Manual NRBC/100WBC
Manual NRBC/100WBC

Man v DM96 reclassified metamyelocyts Man v DM96 metamyelocytes


8.0 8.0
y = 0.9602x + 0.1206
7.0 y = 1.1202x + 0.005 7.0
DM metamyelocytess x 109/l

R 2 = 0.9331 R 2 = 0.7809
metam yelocytes x 109/l

6.0 6.0
DM96 reclassified

5.0 5.0

4.0 4.0

3.0 3.0

2.0 2.0

1.0 1.0

0.0 0.0
0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0
Manual metamyelocytes x 109/l Manual metamyelocytes x 109/l

Figure 2. Continued.

two highly experienced operators showed better All operators found the quality of the WBC images
statistical agreement between the manual and DM96 presented to be excellent and the use of the instru-
differentials. ment to be simple.
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 48–60
56 C. BRIGGS ET AL. EVALUATION OF THE CELLAVISION DM96

Man v DM96 reclassified myelocytes Man v DM96 myelocytes


DM96 reclasified myelocytes 3.5 7.0

3.0 6.0

DM96 myelocytes x 109/l


y = 0.7989x + 0.0884
2.5 R 2 = 0.3709 5.0 y = 1.3028x + 0.1861
R 2 = 0.311
2.0 4.0
x 109/l

1.5 3.0

1.0 2.0

0.5 1.0

0.0 0.0
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 0.0 0.5 1.0 1.5 2.0 2.5 3.0
Manual myelocytes x 109/l Manual myelocytes x 109/l

Man v DM96 reclassified promyelocytes Man v DM96 promyelocytes


2.5 3.0

DM96 promyelocytes x 109/l


2.0 2.5
promyelocytes x 109/l
DM96 reclasified

y = 0.4561x + 0.0273
2.0 R 2 = 0.248
1.5 y = 0.6811x + 0.0392
R 2 = 0.4175 1.5
1.0
1.0

0.5
0.5

0.0 0.0
0.0 0.5 1.0 1.5 2.0 2.5 0.0 0.5 1.0 1.5 2.0 2.5 3.0
Manual promyelocytes x 109/l Manual promyelocytes x 109/l

Figure 2. Continued.

All but one area presented by the DM96 for red of samples with low WBC counts and highly abnor-
blood cell and platelet morphology was judged to be mal cells.
adequate.

Precision
Timing studies
There was no significant difference identified for preci-
Results for the timing study are presented in Table 4. sion on the DM96 preclassified, reclassified or the man-
The total time for the analysis of the 30 blood films ual differential for most cell classes except for basophils,
and reclassification of cells is similar for all operators blasts and NRBC. These differences in precision were
on the DM96, despite the variability of experience in found to be small and the differences between median
the operators (average time 1 h 20 min, range 1 h values ranged only from 0.07 to 0.2 · 109/l.
5 min–1 h 40 min). The time to perform the manual
differential is very different between operators (aver-
DISCUSSION
age time 2 h 54 min, range 1 h 20 min–4 h 10 min),
with the inexperienced taking much longer than the Advances in technology mean that for most blood
experienced, this is probably because of the inclusion samples automated haematology analysers can
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 48–60
C. BRIGGS ET AL. EVALUATION OF THE CELLAVISION DM96 57

on all haematology in-patients to twice per week


Table 2. Percentage agreement for red cell morphology
on the CellaVision DM96 only, despite having daily full blood counts, our blood
film review rate is less than 10% (approximately 200
Red cell Preclassification Reclassification blood films per day). Final WBC differential results
abnormality agreement (%) agreement (%) from the DM96 (after reclassification by the operator)
were compared with the 100-cell manual differential
Polychromasia 76 76
Hypochromia 87 84 and for most cell types, including blast cells and
Microcytosis 85 84 NRBCs, correlation was very good. Correlation for the
Macrocytosis 42 85 eosinophils is not as good as the manual method, this
Anisocytosis 51 74 may be a staining issue which could be modified, the
Poikilocytosis 59 74 eosinophils appear pinker on the DM96 screen than
Preclassification is the number of correct suggestions by under the microscope.
the DM96 and Reclassification is agreement with the For basophils and immature myeloid cells auto-
manual method after the operator has altered the results mated classification was different from the reference
originally presented by the instrument. Two hundred method and this may be because of the small number
and eighty-six blood films were evaluated.
of cells being counted (Rumke, 1985a) or, for the
immature granulocytes, the misclassification of cells
provide precise and accurate WBC differential counts. into the earlier or later maturity class, e.g. myelocyte
However, when abnormal cells are present a manual instead of promyelocyte. This is demonstrated by the
differential or examination of a stained blood film is excellent correlation (r2 = 0.95) between the reference
still required. Manual examination of blood films method and the DM96 when the total immature cells
requires highly trained personnel and is time con- were compared. The subjective classification of imma-
suming. Because of these difficulties in manual cell ture granulocytes into an earlier or later maturity class
counting automated image analysis systems have is also a human failing.
been developed, but earlier systems were slow and The classification of red blood cells was less suc-
failed to identify abnormal cells (Tatsumi & Pierre, cessful with the instrument reporting macrocytosis
2002). and anisocytosis on many samples which had normal
We have evaluated a system capable of automated red blood cells. However, in 95% of samples the qual-
analysis of peripheral blood films and present data ity of the red cell and platelet images were sufficient
that shows that the system can generate WBC differ- for the operator to clearly identify morphological fea-
ential results comparable with a manual differential tures and reclassify the results. In this evaluation, the
performed by a laboratory scientist. 89.2% of cells default red cell precharacterization settings were used
were classified correctly with only 0.9% of abnormal which may have contributed the percentage disagree-
cells misclassified as normal. ment; settings can be readjusted by the user to
The DM96 can be loaded with blood films and the achieve a level of classification equivalent to that by
operator can leave the instrument to perform other manual microscopy.
duties, then on return be presented with the WBCs The overall accuracy of the manual differential by
on the computer screen. These can be reviewed, five different operators does not exceed that of the
quickly reclassified if necessary and the results DM96. This is demonstrated by the comparison of
released into the laboratory information system. In each individual’s manual differential as well as their
busy laboratories, such as ours, which processes 2000– pre- and reclassified differentials from the DM96 to
2500 samples a day, it would be necessary to have the 400-cell reference differential. Correlation to the
two DM96 instruments for the optimum processing of reference method is similar for an operator whether
blood films. Using local decision rules for blood film the differential is performed manually or by the
review, similar to those published by the International DM96 with reclassification of cells. For some cell types
Consensus Group for Hematology Review (Barnes and for some operators, the accuracy of the DM96
et al., 2005), but adapted for our patient population, compared with the reference method was better than
and restricting the number of times a film is looked at that of the individual’s 100-cell manual differential.
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 48–60
58 C. BRIGGS ET AL. EVALUATION OF THE CELLAVISION DM96

Table 3. Correlation coefficients (r2 values) for comparison of five different operators’ differentials to the 400-cell reference
differential

Ref vs. Ref vs. Ref vs. Ref vs. Ref vs. Ref vs. Ref vs. Ref v Ref v
Operator man reclass preclass Man reclass pre-class man Re-class Pre-class

Neutrophils Basophils Metamyelocytes


1 0.994 0.993 0.995 0.295 0.001 0.056 0.274 0.330 0.822
2 0.993 0.991 0.974 0.060 0.012 0.031 0.537 0.701 0.723
3 0.968 0.986 0.988 )0.053 )0.049 )0.152 0.845 0.238 0.758
4 0.843 0.994 0.989 0.023 0.009 0.063 0.789 0.928 0.914
5 0.990 0.990 0.990 0.089 )0.197 )0.284 0.863 0.728 0.956
Lymphocytes Blasts Myelocytes
1 0.897 0.752 0.218 0.998 0.989 0.804 0.804 0.670 0.707
2 0.788 0.841 0.240 0.999 0.998 0.999 0.763 0.712 0.963
3 0.752 0.764 0.324 0.996 0.998 0.982 0.264 0.927 0.831
4 0.442 0.077* 0.179 0.965 0.975 0.952 0.688 0.333 0.329
5 0.686 0.782 0.193 0.998 0.999 0.986 0.712 0.503 0.712
Monocytes Nucleated red blood cell Promyelocytes
1 0.624 0.663 0.677 0.804 0.978 0.973 0.309 0.000 0.690
2 0.674 0.768 0.707 0.995 0.991 0.973 0.948 0.702 0.936
3 0.752 0.540 0.823 0.861 0.960 0.956 0.227 0.423 0.643
4 0.848 0.822 0.805 0.921 0.953 0.915 0.018 0.600 0.706
5 0.521 0.805 0.724 0.992 0.962 0.956 0.551 0.540 0.600
Eosinophils Immature granulocytes
1 0.802 0.461 0.323 0.831 0.987 0.910
2 0.451 0.528 0.155 0.909 0.971 0.917
3 0.624 0.554 0.301 0.750 0.748 0.950
4 0.338 0.395 0.465 0.777 0.670 0.851
5 0.694 0.394 0.112 0.898 0.887 0.956

Ref vs. man, the operators’ manual differential compared with the reference differential; Ref vs. reclass, the differential
from the CellaVision DM96 after re-classification by the operator; Ref vs. preclass, the differential from the CellaVision
DM96 before reclassification by the operator.
*Operator did not reclassify blast cells in one sample that had been wrongly classified by the DM96 as lymphocytes, if
this one sample is excluded r2 = 0.61.

One sample contained a large number of blast cells The wrongly identified cells in the reclassified results
which the DM96 classified as lymphocytes, all opera- from this operator were clearly seen by recalling the
tors, apart from one, reclassified the cells correctly. differential from the DM96 database. An experienced
laboratory scientist and the individual reanalysed the
Table 4. Comparison of time taken to complete the 30 blood film on a cell-by-cell basis to re-educate the
differentials on the CellaVision DM96 including member of staff who had made the misclassification.
reclassification of cells with time taken to perform the Unlike a manual blood film it is exactly the same 100
same differentials manually cells seen the first time that are being reanalysed. The
ability to determine how an individual has classified
Time for analysis Time for manual
Operator on DM96 differential analysis cells makes training and competency testing much
simpler and more effective.
1 1 h 5 min 1 h 45 min CellaVision can provide specific software for profi-
2 1 h 10 min 1 h 40 min ciency testing and education in manual blood cell dif-
3 1 h 30 min 3 h 45 min ferentials. CellaVision Diff IQ has not been
4 1 h 40 min 4 h 10 min
evaluated in this study but should allow laboratories
5 1 h 14 min 3 h 10 min
to ensure that staff are trained to report results consis-
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 48–60
C. BRIGGS ET AL. EVALUATION OF THE CELLAVISION DM96 59

tent with laboratory procedures and helps to guaran- with only a small increase in acquisition time. Statisti-
tee the quality of the differential results. A digital test cally, the more cells counted the more precise the
case is set up by the examiner who also establishes results will be. The DM96 was reliable throughout the
the correct cell classification. A member of staff per- 3-month evaluation period with no breakdowns.
forms a WBC differential and red cell morphology Other advantages of the DM96, outside this evalua-
comments using the DM96 and the results are auto- tion, are the possibility of remote viewing of blood
matically compared with those of the examiner and cells. Software can be installed on any PC which
other participants. There is automatic extraction of allows for verification of cell types from a different
cells possibly misidentified encouraging post-test com- location. A small satellite laboratory without morphol-
parison and discussion. The automatic storage of all ogy expertise can send images to the central labora-
digital images from previous differentials is also useful tory for classification and diagnosis.
for allowing cell comparisons to previous encounters So can automated blood film analysis replace the
on a particular patient. manual differential?
The timing study clearly demonstrates that the We have demonstrated that the differential from
DM96 differential is faster than the manual differen- the DM96 is as good as that by a laboratory scientist;
tial this is consistent with earlier findings where on however, the laboratory scientist operating the DM96
average the manual differential took 1.3 min longer must be skilled in blood cell morphology. The instru-
to perform than when the DM96 was used (Kratz ment gives results of comparable precision with that
et al., 2006).The time saved when using the DM96 is of different individuals performing manual differen-
greatest for the less experienced laboratory scientists. tials on the same sample. The speed of results from
With increased instrument familiarity with the instru- the DM96 is impressive; quicker than all the labora-
ment there may be potential for even more time tory scientists involved in the study, including those
saved. with considerable morphological experience. The
The precision of the DM96 when analysing the DM96 is reliable and certainly has a place in the hae-
same slide five times is similar to that of the manual matology laboratory where it should improve work-
differential performed on the same slide by five differ- flow, make more efficient use of experienced
ent operators. This is unsurprising as the DM96 was laboratory scientists’ time and make training and
set to count 100-cells for equivalence with the man- monitoring of staff in blood cell morphology skills eas-
ual count. The error of these methods is associated ier and more efficient.
with the number of cells counted (Rumke, 1985b) as University College Hospital London was in receipt
well as cell distribution and interobserver variability. of an unrestricted educational grant at the time of this
The DM96 can be adjusted to count more than 100- study from Sysmex Europe GMBH.
cells or more than one slide per sample can be used

REFERENCES Clinical and Laboratory Standards Insti- Performance evaluation of the CellaVi-
tute. (2007) Reference Leukocyte sion DM96 system. American Journal of
Barnes P.W., McFadden S.L., Machin S.J. (WBC) Differential Count (Proportional) Pathology 124, 770–781.
& Simson E. (2005) The international and Evaluation of Instrument Methods; Lewis S.M. & Rowan M. (1991) Assess-
consensus group for hematology Approved Standard, 2nd edn. Vol 27, ment of the need for blood film exami-
review: suggested criteria for action fol- No 4. nation with blood counts by aperture-
lowing automated CBC and WBC differ- Koepke J.A., Dotson M.A. & Shifmann M.A. impedance systems. In: Standard Hae-
ential analysis. Laboratory Hematology (1985) A critical evaluation of the man- matology Practice (ed B. Roberts), pp.
11, 83–90. ual/visual differential leukocyte counting 34–42. Blackwell Scientific Publications,
Bentley S.A. (1990) Automated differen- method. Blood Cells 11, 173–186. Osney Mead, Oxford, UK.
tial white cell counts: a critical apprai- Kratz A., Bengtsson H., Casey J.E., Keefe Prewitt J.M. & Mendlesohn M.L. (1966)
sal. Bailliere’s Clinical Haematology 3, J.M., Beatrice G.H., Grzbek D.Y., Lew- The analysis of cell images. Annals of
851–869. androwski K.B. & Van Cott E.M. (2006)

 2007 The Authors


Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 48–60
60 C. BRIGGS ET AL. EVALUATION OF THE CELLAVISION DM96

the New York Academy of Sciences Rumke C.L. (1985b) Imprecision of ratio- and future of blood cell morphology
128, 1035–1053. derived differential leucocyte counts. identification. Clinics in Laboratory
Rumke C.L. (1985a) Statistical reflections Blood Cells 11, 311–314, 315. Medicine 22, 299–315.
on finding atypical cells. Blood Cells 11, Tatsumi N. & Pierre R.V. (2002) Auto-
141–144. mated image processing. Past, present

 2007 The Authors


Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2009, 31, 48–60

Vous aimerez peut-être aussi