Information about X rays irradiator:

- Features of the Emitter


Tube type: Varian NDI 226
Max tube voltage: 225 KV
Max tube current: 17.5 mA
Max tube watt: 4 KW
Max filament voltage: 5.5 V
Max filament current: 4.2 V
Beam size: 7.5 mm
Irradiation angle: 40º
Weight: 11 Kg

- Chamber design
The chamber is completely shielded by lead to reduce the leakage of beam. Inside the
chamber there is a cooper filter with different thickness to adjust the quality of beam.
In addition, a rotatory table which can hold two irradiation canisters can rotate at a
speed of 4 rpm when exposure. With the moving slide, the rotatory table is able to
adjust the distance from the radiation source and radiation canisters.

- Chamber maximum capacity


Dimensions of chamber: 500 mm*600 mm*544 mm (H*W*D)

- Dose Rate
Central Dose rate at 30 cm from the radiation source: 2.8 Gy/min (with water). This
value is measured through 0.3 mm cooper filter under the voltage of 225 KV.

- Minimum/Maximum Dose Ratio


Sorry we could not catch your question. Do you mean the minimum dose and
maximum dose of the irradiation tube, respectively?
- Electrical requirements
Standard powder supply: AC 220 V ±10%, 50/60 Hz, 7.5 K VA
Or: AC 110V/380V, 50/60Hz, 7.5 K VA

- Equipment Dimensions. Weight.


Dimensions of irradiator: 1836 mm*1460 mm*875 mm (H*W*D)
Weight of irradiator: 1500 Kg

- External Water Cooling Module? Dimensions.Weight.


Water cooling system; Dimensions: 850 mm*560 mm*800 mm(H*W*D)
Weight: 100 Kg (with water)

- Parts and Labor warranty, Tube warranty


We provide one year warranty for the whole machine (both parts and tube) under the
normal operation.

- Safety interlocks
1. Irradiator cannot work under such conditions: the water cooling system is closed or
broken and the lead door is open.
2. The high-voltage generator will close the voltageoutput when there is any
malfunction with the irradiator.