Is there any problem running a modal test with 2 KHz excitation but only analyzing up to 500 Hz ?
Let's discuss this.
Now this is an interesting question. There are several issues to There is really no right or wrong answer here but I have some
be discussed relative to this. The more important question is strong feelings regarding the adequacy of this measurement as
maybe why one would want to run a test in this fashion in the shown. Without some very specific details, I really don’t want
first place and then discuss some of the issues that might have to make this measurement as requested. Looking at the input
an effect on the overall measurement and then possibly some power spectrum, cross power spectrum, frequency response
alternate things to consider. function and coherence, there is definitely excitation and
response to 2 KHz. There appears to be considerably higher
So let’s consider a measurement as shown in Figure 1. As the response levels in the higher frequency range as well as many
question was posed, the measurement would be acquired over a more modes of the system. This measurement looks acceptable
2KHz range but the only range up to 500 Hz is to be analyzed. overall but is it really the best possible measurement over the
500 Hz frequency range of interest?
It seems much wiser to limit the excitation used with a low pass
filter and not ever excite the higher frequency modes of the
system. This would then possibly allow the use of more
FREQUENCY
RESPONSE
FUNCTION
1 KHz 2 KHz
Figure 3 – Measured Response to 1 KHz
AVERAGED INPUT AVERAGED CROSS
POWER SPECTRUM POWER SPECTRUM
So what if I am forced to run a test with a 2 KHz excitation but
only analyze to 500 Hz. It might be best to run a test with
2KHz excitation and a second test with 500 Hz excitation. Both
measurements should provide equivalent information if all the
issues identified above have been properly addressed. And if I
am forced to excite the structure to 2 KHz, then I would run
COHERENCE
FUNCTION
both tests and analyze both sets of data to see if there are any
significant differences. Of course, this still would imply that the
ALLOWS MORE SENSITIVE instrumentation would have to be suitable for both frequency
LOW FREQUENCY ACCELEROMETER ranges and therefore may not be optimum for the lower
TO BE USED TO MEASURE SYSTEM frequency range.
I hope that this little discussion has shed some light on the
effects of acquiring data well beyond the actual frequency range
FREQUENCY
RESPONSE of interest. It can be done, if required, but there may be some
FUNCTION issues related to selecting transducers that appropriately
measure the actual frequency band of interest accurately. You
1 KHz 2 KHz
need to evaluate this carefully. If you have any more questions
Figure 2 – Frequency Excitation to 500 Hz
on modal analysis, just ask me.
_________________________________________________________________________________________________________________________
Is there any problem running a modal test with 2KHz excitation but only analyze up to 500 Hz? Copyright 2006
SEM Experimental Techniques – Jun 2006 Page 2 Pete Avitabile