1. General
Automatic calibration is carried out using a CCS2 and a
BSS2 in combination with the INT-Software CALSOFT 01.
Based on ISO 11171 CALSUS 01, a fluid for secondary
calibration, is needing during this process.
Comments:
The drifting of calibration values over time is rather
insignificant.
If drastic changes of calibration values (more than 100%)
occur, malfunction is suspected. Especially if very low
values are determined. In such cases it is to assume that
partial contamination of the sensor or contaminations of
the windows are the reason.
3
2. Initial operation
4
Notice:
Be aware of the fact that this process has to be performed
quickly in order to ensure that particles are spread out even
within the sample. This is important for accurate and constant
measurements.
5
Now open the valve of the BSS 2, flush with [2]
(approximate 3 – 4), stop the flushing with [5] and start the
count by hitting [3] on the CCS 2.
Shut the valve of the BSS 2 again when finished with the
measurements.
6
3.3. Calculating ideal numbers of particles
XN = (XZ - XA) x 10
XZ = 6475 particles / 1 ml
XA = 358 particles / 1 ml
XN = (6475 – 358) x 10
= 61170 particles / 10 ml
7
4. Calibration
The purpose is to align the actual and the ideal numbers
of particles.
Activate the input box by hitting [F2], delete the old value
with [DEL] and after entering the ideal value confirm by
hitting [F2] again.
After entering all ideal values close this menu by hitting [5].
8
4.2. Preparation of the calibration solution
Agitate the calibration solution moderately for app. 10
minutes in all directions.
green: 1,5 %
yellow: 3%
red: > 3%
9
Redo the calibration until ideal and actual values of the
measuring channels: > 4 µm, > 4,6 µm, > 6 µm, > 6,4
µm, > 10 µm, > 14 µm do not differ more than 1.5%.
(b) Print-out:
Print your values by hitting [F1].
10
Notice: Unstable values are an indication for insufficient
preparation and/or bad handling of the solution for calibration.
Careful preparation of samples will result in successful
calibration!
After you are done the calibration please flush the BSS
2 and the CCS 2 with cleanded mineral oil ISO VG 22.
11
North America — HQ Europe/Africa/Middle East For more information, please e-
70 Wood Ave., South, 2nd Friedensstraße 41
Floor D-68804 Altlussheim, mail us at
Iselin, NJ 08830 Germany filtrationinfo@eaton.com
Toll Free: (800) 656-3344 Voice: +49-6205-2094-0
(North America Only)
Voice: (732) 767-4200 Auf der Heide 2 Visit us online eaton.com/filtration
53947 Nettersheim, Germany
Voice: +49-2486-809-0
for a complete list of Eaton´s
filtration products.
China An den Nahewiesen 24
No. 3, Lane 280, Linhong 55450 Langenlonsheim, ©2012 Eaton Corporation. All Rights Reserved.
Road Germany All trademarks and registered trademarks are the property of their
Changning District, 200335 Voice: +49 6704 204-0 respective owners. Litho USA.
Shanghai, P.R. China All information and recommendations appearing in this brochure
Voice: +86-21-5200-0099 concerning the use of products described herein are based on tests
believed to be reliable. However, it is the user’s responsibility to
determine the suitability for his own use of such products. Since the
Brazil actual use by others is beyond our control, no guarantee, expressed
Av. Julia Gaioli, 474 – or implied, is made by Eaton as to the effects of such use or the
results to be obtained. Eaton assumes no liability arising out of the
Singapore Bonsucesso use by others of such products. Nor is the information herein to be
4 Loyang Lane #04-01/02 07251-500 – Guarulhos, construed as absolutely complete, since additional information may
be necessary or desirable when particular or exceptional conditions
Singapore 508914 Brazil or circumstances exist or because of applicable laws or government
Voice: +65-6825-1668 Voice: +55 (11) 2465-8822 regulations.