Biomedical applications of polymer-composite materials: a review


S. Ramakrishna a,*, J. Mayer b, E. Wintermantel c, Kam W. Leong d
a
Department of Mechanical Engineering, National University of Singapore, 9 Engineering Drive 1, Singapore 117576 Singapore
b
Chair of Biocompatible Materials Science and Engineering, Department of Materials, Swiss Federal Institute of Technology (ETH),
Wagistrasse 23, CH-8952 Schlieren, Switzerland
c
Central Institute of Biomedical Engineering, Technical University of Munich, D-85748, Garching, Germany
d
Department of Biomedical Engineering, Ross 726, School of Medicine, Johns Hopkins University, Baltimore, MD 21205, USA

Received 3 April 2000; received in revised form 26 October 2000; accepted 14 November 2000

Abstract
An overview of various biomedical applications of polymer-composite materials reported in the literature over the last 30 years is
presented in this paper. For the bene®t of the readers, general information regarding structure and function of tissues, types and
purpose of implants/medical devices, and various other materials used, are also brie¯y presented. Di€erent types of polymer com-
posite that are already in use or are investigated for various biomedical applications are presented. Speci®c advantages of using
polymer-composite biomaterials in selected applications are also highlighted. The paper also examines the critical issues and sci-
enti®c challenges that require further research and development of polymer composite materials for their increased acceptance in
the biomedical industry. # 2001 Elsevier Science Ltd. All rights reserved.
Keywords: Biomaterials; Biocomposites; Polymer composites; Implants; Prosthesis; Medical devices; Biomedical engineering; Bioengineering

1. Introduction and thus improve the quality of life of the patients.


According to a report published in 1995 by The Insti-
Biomaterials are materials of natural or man-made tute of Materials, London, the estimated world market
origin that are used to direct, supplement, or replace the for all medical devices, including diagnostic and ther-
functions of living tissues of the human body [21]. Use of apeutic equipment is in the region of $100 billion per
biomaterials dates far back into ancient civilizations. year. Within this industry, the world market for bioma-
Arti®cial eyes, ears, teeth, and noses were found on terials is estimated to be around $12 billion per year,
Egyptian mummies [256]. Chinese and Indians used with an average global growth of between 7 and 12%
waxes, glues, and tissues in reconstructing missing or per annum. Biomaterials are expected to perform in our
defective parts of the body. Over the centuries, advance- body's internal environment, which is very aggressive.
ments in synthetic materials, surgical techniques, and For example the pH of body ¯uids in various tissues
sterilization methods have permitted the use of biomater- varies in the range from 1 to 9. During daily activites
ials in many ways [178]. Medical practice today utilizes a bones are subjected to a stress of approximately 4 MPa
large number of devices and implants. Biomaterials in the whereas the tendons and ligaments experience peak
form of implants (sutures, bone plates, joint replace- stresses in the range 40±80 MPa. The mean load on a
ments, ligaments, vascular grafts, heart valves, intrao- hip joint is up to 3 times body weight (3000 N) and peak
cular lenses, dental implants, etc.) and medical devices load during jumping can be as high as 10 times body
(pacemakers, biosensors, arti®cial hearts, blood tubes, weight. More importantly, these stresses are repetitive
etc.) are widely used to replace and/or restore the function and ¯uctuating depending on the activities such as
of traumatized or degenerated tissues or organs, to assist standing, sitting, jogging, stretching, and climbing [21].
in healing, to improve function, to correct abnormalities, In a year, the stress cycles of ®nger joint motion or hip
joint motion estimated to be as high as 1106 cycles,
and for a typical heart 0.5 107±4107 cycles. This
* Corresponding author. information roughly indicates the acute and instantaneous
E-mail address: engsr@nus.edu.sg (S. Ramakrishna). biological environment in which the biomaterials need to
0266-3538/01/$ - see front matter # 2001 Elsevier Science Ltd. All rights reserved.
PII: S0266-3538(00)00241-4
1190 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

Nomenclature

BIS-GMA bis-phenol A glycidyl methacrylate PET polyethylene terepthalate


C carbon PGA poly(glycolic acid)
CF carbon ®bers PHB polyhydroxybutyrate
GF glass ®bers PHEMA poly(HEMA) or poly(2-hydroxyethyl
HA hydroxyapatite/hydroxylapatite methacrylate)
HDPE high density polyethylene PLA poly(lactic acid)
KF Kevlar ®ber PLDLA poly(l-dl-lactic acid)
LCP liquid crystalline polymer PLLA poly(l-lactic acid)
LDPE low density polyethylene PMA polymethylacrylate
MMA methylmethacrylate PMMA polymethylmethacrylate
PA polyacetal Polyglactin copolymer of PLA and PGA
PBT polybutylene terephthalate PP polypropylene
PC polycarbonate PS polysulfone
PCL polycaprolactone PTFE polytetra¯uroethylene
PE polyethylene PU polyurethane
PEA polyethylacrylate PVC polyvinylchloride
PEEK polyetheretherketone SR silicone rubber
PEG polyethylene glycol THFM tetrahydrofurfuryl methacrylate
PELA block copolymer of lactic acid and UHMWPE ultra high molecular weight poly-
polyethylene glycol ethylene

survive. Needless to say, the biological environment also tion to the mechanical behavior of the host tissues.
depends on the patient's conditions and activities. Therefore, structural compatibility refers to the
In the early days all kinds of natural materials such as mechanical properties of the implant material, such as
wood, glue and rubber, and tissues from living forms, elastic modulus (or E, Young's modulus) and strength,
and manufactured materials such as iron, gold, zinc and implant design (sti€ness, which is a product of elastic
glass were used as biomaterials based on trial and error. modulus, E and second moment of area, I), and optimal
The host responses to these materials were extremely var- load transmission (minimum interfacial strain mis-
ied. Some materials were tolerated by the body whereas match) at the implant/tissue interface. Optimal interac-
others were not. Under certain conditions (characteristiccs tion between biomaterial and host is reached when both
of the host tissues and surgical procedure) some materials the surface and structural compatibilities are met. Fur-
were tolerated by the body, whereas the same materials ther more it should be noted that the success of a bio-
were rejected in another situation. Over the last 30 years material in the body also depends on many other factors
considerable progress has been made in understanding such as surgical technique (degree of trauma improsed
the interactions between the tissues and the materials. It during implantation, sterilization methods, etc), health
has been acknowledged that there are profound di€er- condition and activities of the patient. Table 1 sum-
ences between non-living (avital) and living (vital) marizes various important factors that are considered in
materials. Researchers have coined the words `bioma- selecting a material for a biomedical application.
terial' and `biocompatibility' [253] to indicate the biolo- Clinical experience clearly indicates that not all o€-
gical performance of materials. Materials that are the-shelf materials (commonly used engineering materi-
biocompatible are called biomaterials, and the bio- als) are suitable for biomedical applications. The var-
compatibility is a descriptive term which indicates the ious materials used in biomedical applications may be
ability of a material to perform with an appropriate grouped into (a) metals, (b) ceramics, (c) polymers, and
host response, in a speci®c application [22]. In simple (d) composites made from various combinations of (a),
terms it implies compatibility or harmony of the bio- (b) and (c). Researchers also class®ed materials into
material with the living systems. Wintermantel and several types such as bioinert and bioactive, biostable
Mayer [258] extended this de®nition and distinguished and biodegradable, etc. [90]. As the former classi®cation
between surface and structural compatibility of an is known to engineers, it is further followed in this review.
implant [260]. Surface compatibility meaning the che- Alumina, titania, zirconia, bioglass (or bioactive glasses),
mical, biological, and physical (including surface mor- carbon, and hydroxyapatite (HA) are widely considered
phology) suitability of an implant surface to the host as biocompatible ceramics. Metals and alloys that
tissues. Structural compatibility is the optimal adapta- are successful as biomaterials include: gold, tantalum,
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1191

Table 1
Various factors of importance in material selection for biomedical applications

Factors Description

1st Level material Chemical/biological characteristics Physical characteristics Mechanical/structural characteristics


properties Chemical composition Density Elastic modulus
(bulk and surface) Poisson's ratio
Yield strength
Tensile strength
Compressive strength
2nd Level material Adhesion Surface topology Hardness
properties (texture and roughness) Shear modulus
Shear strength
Flexural modulus
Flexural strength
Speci®c functional Biofunctionality (non-thrombogenic, Form (solid, porous, coating, Sti€ness or rigidity
requirements cell adhesion, etc.) ®lm, ®ber, mesh, powder) Fracture toughness
(based on application) Bioinert (non-toxic, Geometry Fatigue strength
non-irritant, non-allergic, Coeceint of thermal expansion Creep resistance
non-carcenogenic, etc.) Electrical conductivity Friction and wear resistance
Bioactive Color, aessthetics Adhesion strength
Biostability (resistant to corrosion, Refractive index Impact strength
hydrolysis, oxidation, etc.) Opacity or translucency Proof stress
Biogradation Abrasion resistance

Processing and Reproducibility, quality, sterilizability, packaging, secondary processability


fabrication
Characteristics of host: tissue, organ, species, age, sex, race, health condition, activity, systemic response
Medical/surgical procedure, period of application/usage
Cost

stainless steel, Co±Cr, NiTi (shape memory alloy), and forms (solids, ®bers, fabrics, ®lms, and gels), and can be
Ti alloys. A large number of polymers such as poly- fabricated readily into complex shapes and structures.
ethylene (PE), polyurethane (PU), polytetra¯uoroethyl- However, they tend to be too ¯exible and too weak to
ene (PTFE), polyacetal (PA), polymethylmethacrylate meet the mechanical demands of certain applications e.g.
(PMMA), polyethylene terepthalate (PET), silicone as implants in orthopedic surgery. Also they may absorb
rubber (SR), polysulfone (PS), polyetheretherketone liquids and swell, leach undesirable products (e.g. mono-
(PEEK), poly(lactic acid) (PLA), and poly(glycolic acid) mers, ®llers, plasticizers, antioxidants), depending on
(PGA) are also used in various biomedical applications. the application and usage. Moreover, the sterilization
HA/PE, silica/SR, carbon ®ber/ultra high molecular processes (autoclave, ethylene oxide, and 60Co irradia-
weight polyethylene (CF/UHMWPE), carbon ®ber/ tion) may a€ect the polymer properties. Metals are
epoxy (CF/epoxy), and CF/PEEK are few examples of known for high strength, ductility, and resistance to
polymer composite biomaterials. Each type of material wear. Shortcomings of many metals include low bio-
has its own positve aspects that are particularly suitable compatibility, corrosion, too high sti€ness compared to
for speci®c application. This paper is intended mainly to tissues, high density, and release of metal ions which
provide an overview of various polymer composite bio- may cause allergic tissue reactions [221]. Ceramics are
materials, and also to stimulate the research in compo- known for their good biocompatibility, corrosion resis-
site biomaterials as this material group has not been tance, and high compression resistance. Drawbacks of
explored extensively with regards to the biomedical ceramics include, brittleness, low fracture strength, dif-
applications. In this paper, the merits and demerits of ®cult to fabricate, low mechanical reliability, lack of
polymer composite materials are emphasized by con- resilience, and high density. Polymer composite materi-
trasting with the other types of materials. However, it is als provide alternative choice to overcome many short-
not the intention of the authors to advocate that polymer comings of homogenous materials mentioned above.
composite biomaterials are the only candidates suitable The speci®c advantages of polymer composites are
for medical applications. highlighted in the following.
A large number of polymers are widely used in var- Generally, tissues are grouped into hard and soft tis-
ious applications. This is mainly because they are avail- sues. Bone and tooth are examples of hard tissues, and
able in a wide variety of compositions, properties, and skin, blood vessels, cartilage and ligaments are a few
1192 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

examples of soft tissues. As the names suggest, in gen- sti€ness. Thus, bone is insuciently loaded compared to
eral the hard tissues are sti€er (elastic modulus) and the implant, and this phenomenon is called `stress-
stronger (tensile strength) than the soft tissues (Tables 2 shielding' or stress protection. Many investigators
and 3). Considering the structural or mechanical com- [44,168,238], have shown that the degree of stress pro-
patibility with tissues, metals or ceramics are chosen for tection is proportional to the degree of sti€ness mis-
hard tissue applications (Tables 2 and 4), and polymers match. The stress-shielding a€ects the bone remodeling
for the soft tissue applications (Tables 3 and 5). A closer and healing process leading to increased bone porosity
look at Tables 2 and 4 reveals that the elastic moduli of (also known as bone atrophy) [44,103,214,251]. It has
metals and ceramics are at least 10±20 times higher than been recognised that by matching the sti€ness of
those of the hard tissues. One of the major problems in implant with that of the host tissues limits the stress-
orthopedic surgery is the mismatch of sti€ness between shielding e€ect and produces desired tissue remodeling.
the bone and metallic or ceramic implants. In the load In this respect, the use of low-modulus materials such as
sharing between the bone and implant, the amount of polymers appears interesting; however, low strength
stress carried by each of them is directly related to their associated with low modulus usually impairs their
potential use. Since the ®ber reinforced polymers i.e.
polymer composite materials exhibit simultanously low
Table 2
Mechanical properties of hard tissues [22]
elastic modulus and high strength, they are proposed for
several orthopedic applications [85,176]. Additional
Hard tissue Modulus Tensile merit of composite materials is that by controlling the
(GPa) Strength volume fractions and local and global arrangement of
(MPa)
the reinforcement phase, the properties and design of an
Cortical bone (longitudinal direction) 17.7 133 implant can be varied and tailored to suit the mechan-
Cortical bone (transverse direction) 12.8 52 ical and physiological conditions of the host tissues. It
Cancellous bone 0.4 7.4
is, therefore, suggested that composite materials o€er a
Enamel 84.3 10
Dentine 11.0 39.3 greater potential of structural biocompatibility than the
homogenous monolithic materials. They have reasonably
adequate strength [145]. Moreover the human tissues are
Table 3 essentially composite materials with anisotropic proper-
Mechanical properties of soft tissues [22] ties, which depend on the roles and structural arragements
of various components (e.g. collagen, elastin, and hydro-
Soft tissue Modulus Tensile
(MPa) strength xyapatite) of the tissues. For example, the longitudinal
(MPa) mechanical properties of cortical bone are higher than
the transverse direction properties (see Table 2). These
Articular cartilage 10.5 27.5
similarities have led to the development of composite
Fibrocartilage 159.1 10.4
Ligament 303.0 29.5 biomaterials. Other reasons for the development of
Tendon 401.5 46.5 polymer composite biomaterials include: absence of
Skin 0.1±0.2 7.6 corrosion and fatigue failure of metal alloys and release
Arterial tissue (longitudinal direction) 0.1 of metal ions such as Nickel or Chromium which may
Arterial tissue (transverse direction) 1.1
cause loosening of the implant, patient discomfort, and
Intraocular lens 5.6 2.3
allergic skin reactions; and low fracture toughness of

Table 4
Mechanical properties of typical metallic and ceramic biomaterials [22] Table 5
Mechanical properties of typical polymeric biomaterials [22]
Material Modulus Tensile
(GPa) strength Material Modulus Tensile
(MPa) (GPa) strength
(MPa)
Metal alloys
Stainless steel 190 586 Polyethylene (PE) 0.88 35
Co±Cr alloy 210 1085 Polyurethane (PU) 0.02 35
Ti-alloy 116 965 Polytetra¯uoroethylene (PTFE) 0.5 27.5
Amalgam 30 58 Polyacetal (PA) 2.1 67
Polymethylmethacrylate (PMMA) 2.55 59
Ceramics
Polyethylene terepthalate (PET) 2.85 61
Alumina 380 300
Polyetheretherketone (PEEK) 8.3 139
Zirconia 220 820
Silicone rubber (SR) 0.008 7.6
Bioglass 35 42
Polysulfone (PS) 2.65 75
Hydroxyapatite 95 50
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1193

ceramic materials which make them a dicult choice for rial because its properties are directionally dependent
load bearing applications. Composite materials o€er (Table 2). Bone is generally weak in tension and shear,
several other signi®cant advantages over metal alloys particularly along the longitudinal plane. Under excessive
and ceramics in correcting the above mentioned or per- loading or impact bone fractures, and there are many
ceived de®ciencies [88,226,229]. Metals alloys and cera- types of bone fractures depending on the crack size,
mics are radio opaque and in some cases they result in orientation, morphology, and location. Readers are
undesirable artifacts in X-ray radiography [14]. In the recommended to refer to AO (Arbeitsgemeinschaft fur
case of polymer composite materials the radio transpar- Osteosynthesefragen)/ASIF (Association of Surgeons for
ancy can be adjusted by adding contrast medium to the Internal Fixation) documents for detailed classi®cation
polymer. Moreover the polymer composite materials are of bone fractures. Bone fractures are treated (anatomic
fully compatible with the modern diagnostic methods reduction) in di€erent ways and they may be grouped
such as computed tomography (CT) and magnetic into two types namely external ®xation and internal
resonance imaging (MRI) as they are non-magnetic. ®xation. The external ®xation does not require opening
Considering their light weight and superior mechanical the fracture site whereas the internal ®xation requires
porperties, the polymer composites are also used as struc- opening of the fracture site. In the external ®xation
tural components of these imaging devices. Some times, approach the bone fragments are held in alignment
the unreinforced polymers may not have properties su- through various means such as splints, casts, braces, and
cient for intended application. For example, ®ber rein- external ®xator systems. Casting materials or plaster
forced UHMWPE has superior creep and fatigue bandages are used to form splints, casts or braces [20].
resistance than the unreinforced UHMWPE. Higher The casting material essentially is a composite material
creep and fatigue resistance properties are desirable in made of woven cotton fabrics (woven gauze) and Plaster
total knee joint replacement. As shown in Fig. 1, over of Paris matrix (calcium sulphate). Other reinforcements
the years a wide variety of polymer composite materials include fabrics of glass and polyester ®bers. Although the
have been developed for various biomedical applications plaster bandages have many advantages, they also have
[198]. The following sections present details of polymer many disadvantages such as messy application, heavy,
composite biomaterials in terms of hard tissue and soft bulky, low speci®c strength and modulus, low water
tissue applications. In each section, for the bene®t of resistance, low fatigue strength, radiopaque, and long set-
readers, general information regarding sturcture and ting time to become load bearing. Recently, casts made of
function of tissues, purpose and type of implants or glass or polyester ®ber fabrics, and water-activated poly-
devices, and various other materials used are also brie¯y urethanes are gaining popularity. An ideal cast material
presented. Glossary of medical terms used in this paper should be easy to handle, light weight, conformable to
is given in Appendix A. anatomical shape, strong, sti€, water proof, radiolucent,
and easy to remove. More over it should be permeable to
ventilation without which the patient's skin may be scor-
2. Hard tissue applications ched or weakened. To address this speci®c problem,
recently Philips [187] developed a new breathable cast
2.1. Bone fracture repair material using double wall knitted fabrics as reinforcement.
A typical external ®xation system [16] comprises of
Bones of the skeletal system provide the supporting Kirschner wires or pins that are pierced through the bone
structure for the body. Bone is a structural composite and held under high tension by screws to the external
composed of collagen ®bers with hydroxyapatite nano- frame (Fig. 1). The wires can be oriented at di€erent
crystalls precipitated along the collagen ®brils [195]. angles across the bone, and their tension is adjusted to
Bone also contains other constituents such as mucopo- provide necessary ®xation rigidity. To ensure stability,
lysaccharides, blood vessels, and bone cells. The low the external ®xators are designed with high rigidity and
elastic modulus collagen ®bers are aligned in bone along strength. Traditional designs are made of stainless steel,
the main stress directions. The high elastic modulus which is heavy and causes discomfort to the patients as
hydroxyapatite mineral comprises approximately 70% they carry the system for several months. External ®xa-
of the dry bone mass and contributes signi®cantly to the tors constructed from CF/epoxy composite materials
bone sti€ness. Bone can remodel and adapt itself to the are gaining acceptance owing to their lightweight yet
applied mechanical environment, which is generally sucient strength and sti€ness [15]. Moreover, the eva-
known as Wol€'s law (see Appendix A). Density of the luation of the bone union by radiography becomes easy,
living bone is in¯uenced by the stress condition applied as the radiolucency of polymer composites is good and
to the bone. Higher applied stress leads to denser bone. they do not cause artifacts in the radiographs. The exter-
Conversely, if the applied stress is lower than the nor- nal ®xation is also used for bone lengthening purposes.
mal physiological load, the bone mass decreases and In the internal ®xation approach the bone fragments
leads to bone weakening. Bone is an anisotropic mate- are held together by di€erent ways using implants such
1194 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

Fig. 1. Various applications of di€erent polymer composite biomaterials.

as wires, pins, screws, plates, and intramedullary nails. greater trochanter, which is often detached during total
The conventional implants are made of stainless steel, hip joint replacement. They are also used to provide
Co±Cr, or Ti alloys. The surgeon based on his experi- additional stability in long oblique or spiral fractures of
ence and the type of fracture judges the bone fracture long bones (femur, humerus, radius, ulna, tibia, and
treatment method. Surgical wires and pins are the sim- ®bula). Most widely used bone screws are two types,
plest implants used to hold the small fragments of bones cortical bone screws (with smaller threads), and cancel-
together. For example wires are used to reattach the lous screws (with larger threads). They are used either to
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1195

directly fasten bone fragments together or to attach a fracture site, primary healing is no longer possible and is
plate to the fractured bone. However proper implant replaced by a more physiological bone healing process,
design and surgical technique must be utilized to ensure which is characterized by the formation of an external
the desired biomechanical outcome of the ®xation and callus bridging the fracture. Thereby, the callus increa-
to avoid additional tissue trauma and devascularization ses the cross-section of the newly formed bone and,
at the fracture site [41]. Fracture healing also would thus, prevents refracture. In early studies, researchers
depend on the patient activities, as they determine the tried using polymers such as PA, PTFE, and polyester
stable or unstable mechanical conditions at the fracture for bone plate applications, and found them to be not
site. It may be noted that all these implants are tem- suitable because of their too low sti€ness. They over-
porarily placed inside the body. After satisfactory heal- looked the fact that the materials proposed for bone
ing of the bone fracture, the implants may be removed plate application must also posses suciently high fati-
based on the discretion of the surgeon. gue strength (comparable to stainless steel), as the
orthopedic devices are subjected to extremely high cyclic
2.1.1. Bone plates loads, and must not lead to large strains at the fracture
Plate and screw ®xation as shown in Fig. 1 is the most site, which may a€ect the bone union. It is now clearly
popular method for rigid internal ®xation of the frac- established that any new material proposed for bone
tured bone. The bone plates are also called osteosynth- plate application must have suciently high fatigue
esis plates. They are made of stainless steel, Cr±Co and strength and appropriate sti€ness. Polymer composite
Ti alloys. The rigid ®xation is designed to provide high materials o€er desired high strength and bone like elas-
axial pressures (also known as dynamic compression) in tic properties [28]. Hence, several investigators proposed
the fragments of the bone, which facilitate primary bone a variety of polymer composite materials for bone plate
healing without the formation of external callus. This applications (Fig. 2) [86,227]. They may be grouped into
method allows the exercise of joints near the fracture non-resorbable, partially resorbable, and fully resorbable
site just after the operation. After a complete bone bone plates [47,133]. The non-resorbable composite plates
healing has been obtained by the plate ®xation, nor- are made of either thermoset polymer composites or ther-
mally it takes from 1 year to 2 years after the operation, moplastic composite materials. CF/epoxy, GF/epoxy are
the plate and screws are removed. However, the rigid few examples of non-resorbable thermoset composites
®xation is not free from complications and reported that [5,29,30,159,223]. Some researchers expressed concern
it results in bone atrophy beneath the plate. There is a over the toxic e€ects of monomers in partially cured
possibility of refracture of bone after the removal of the epoxy composite materials [167,184] and hence research
plates due to bone atrophy [60,95,264]. This is attrib- activity on these materials gradually decreased. As the
uted to the stress shielding e€ect explained earlier. It technology for making good quality thermoplastic com-
may be noted that the modulus of stainless steel (210±230 posites made available, researchers developed CF/PMMA
GPa) is much higher than 10±18 GPa modulus of the [263], CF/PP [43], CF/PS [48,105,107,159], CF/PE [209],
bone. The sti€ness mismatch results in a situation that the CF/nylon, CF/PBT [77], and CF/PEEK [118,135,157,185,
plate transmits the majority of the stress, and the bone 200,249,253] non-resorbable thermoplastic composite
directly beneath the plate experience less stress even after
the fracture has been repaired [233]. The bone under-
neath the plates adapts to the low stress and becomes
less dense and weak. Therefore, the strength of the
healed bone is low. Consequently, there is a possibility
of bone refracture upon removal of the ®xation plate
[44]. The stress shielding e€ect is more pronounced with
the stainless steel plates than the Ti alloy plates. Moyen
et al. [168] and Uhtho€ and Finnegan [238] reported
that the magnitude of bone atrophy under a Ti alloy
plate is signi®cantly lower than that under a stainless
steel plate. It may be noted that the modulus of stainless
steel (230 GPa) is higher than that of the Ti alloy (110
GPa). This example suggests that `less rigid ®xation
plates' diminish the stress-shielding problem and it is
desirable to use plates whose mechanical properties are
close to those of the bone. In other words reduced sti€-
ness mismatch between the implant and the host tissues.
The adaptation of sti€ness also changes the fracture Fig. 2. Bone plates made of (a) CF/epoxy and (b) CF/PEEK compo-
healing mechanisms. Due to the higher strains at the site materials.
1196 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

bone plates. Unlike the thermoset composites, the ther-


moplastic composites are considered free from the
complications associated with unused monomers. More
over, similar to metal alloy plates, thermoplastic com-
posite plates can be bent or contoured (under some
conditions) to the shape of the bone at the time of sur-
gery. At the moment there is insucient data on the
long-term in vivo behavior of non-resorbable thermo-
plastic composite materials. Among various materials
investigated, the CF/PEEK is reportedly biocompatible
[167] and has good resistance to hydrolysis and radia-
tion (a sterilization method) degradation. The other
promising properties include high strength, fatigue
resistance [51,157], and biological inertness with no
mutagenicity or carcinogenicity [44]. The tissue response
to carbon ®bers and composite debris has been described
as minimal. Initially, researchers used short carbon ®ber Fig. 3. CF/PEEK composite screws.
reinforced PEEK composites, as the technologies for
fabricating continuous ®ber reinforced PEEK compo- in vivo conditions with no change in the plate sti€ness
sites were not available at that time. As can be expected with implantation time.
from the composite reinforcement principles, the short As the bone healing progresses, it is desirable that the
®ber composites posses low modulus and strength com- bone is subjected to gradual increase of stress, thus
pared to continuous ®ber reinforced composite materi- reducing the stress-shielding e€ect. In other words, the
als [77]. This means that plates made of short ®ber stress on the plate should decrease with time whereas the
composites must have greater bulk to approximate the stress on the bone should increase. This is possible only if
mechanical sti€ness required for a bone plate. The bulk the plate looses rigidity in in vivo environment. The non-
limitation of short ®ber composites may be increased resorbable polymer composites do not display this
considering their susceptibility to in vivo degradation. desired characteristic. To meet this need, researchers
Hence, there is a need to develop suitable technologies to introduced resorbable polymers for bone plate applica-
fabricate good quality continuous carbon ®ber rein- tions [75]. The polymers such as poly(lactic acid) (PLA)
forced PEEK composites. Mayer [155,156] developed and poly(glycolic acid) (PGA), resorb or degrade upon
knitted CF/PEEK composite bone plates using com- implantation into the body [150,177]. As such these
mingled yarns of carbon and PEEK ®bers. Recently, polymers are either brittle or too weak and ¯exible for
Ramakrishna et al. [200] developed braided CF/PEEK safe clinical use in load bearing applications. Many
composite bone plates using a new technique [276]. bioresorbable polymers found to loose most of their
They initially made micro-braided yarns by combining mechanical properties in few weeks. Tormala et al. [236]
carbon and PEEK ®bers. Micro-braided yarns were and Choueka et al. [42] proposed fully resorbable com-
again braided into ¯at fabrics of desired dimensions. posites by reinforcing resorbable matrices with resorb-
Compression molding above the melting point of PEEK able ®bers (poly(l-lactic acid) (PLLA) ®bers and
matrix resulted in continuous CF/PEEK composites calcium phosphate based glass ®bers). One of the
bone plates. Considering the superior mechanical prop- advantages often sighted for resorbable composite pros-
erties of continuous carbon ®ber reinforced PEEK com- theses is that they need not be removed with a second
posites, it is possible to produce relatively less bulky bone operative procedure, as is recommended with metallic or
plates with out compromising the mechanical require- non-resorbable composite implants. The maximum
ments of the plate. Researchers also developed CF/car- mechanical property of resorbable materials is continues
bon [23] and CF/PEEK [147,185] composite screws (Fig to be a limitation and hence they are limited to only
3), for osteosynthethesis. The squeeze casting method applications where the loads are moderate [215]. In
developed by Peter et al. [185] uses a new net shape ¯ow order to improve mechanical properties, the resorbable
process, which allows fabrication of complex shaped polymers are reinforced with variety of non-resorbable
components with ®ber contents as high as 62% by materials including carbon ®bers [55,170,180,235,272] and
volume. The fatigue properties of the implants made by polyamide ®bers [206,208]. Because of the non-resorbable
this process surpass those of the titanium implants by nature of reinforcements used these composites are called
up to 100%. Combination of polymer composite plates partially resorbable composites. According to Zimmer-
and screws overcomes the corrosion problem faced by man et al. [272], CF/PLA composite possessed superior
the metal plates and screws. The non-resorbable poly- mechanical properties before the implantation. How-
mer composite materials are designed to be stable in ever, they lost mechanical properties too rapidly in
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1197

in vivo environments because of delamination. Further proper blending of mobility, stability, and structural
work is necessary to tailor the composite material such integrity is essential to ful®ll these goals simultaneously.
that the resorption of the plate and the healing rate of The dual function is realized by a linked structure con-
the bone are synchronized [65]. The long-term e€ects of sisting of 33 vertebrae superimposed on one another.
resorbed products, and biostable or slowly eroding The vertebrae are separated by ®brocartilaginous inter-
®bers in the living tissues are not known, and these are vertebral discs (IVD) and are united by articular cap-
the concerns yet to be resolved [27]. sules and ligaments. The IVD is a composite structure
made up of a core, nucleus pulposus, surrounded by
2.1.2. Intramedullary nails multilayered ®bers (90 concentric layers) of the annulus
Intramedullary nails or rods are mainly used to ®x the ®brosis. The orientation of annulus ®bers vary from 62
long bone fractures such as fracture of femoral neck or at the periphery to 45 in the vicinity of the nucleus, thus
intertrochanteric bone fracture. It is inserted into the imparting structurally graded architecture to the disc
intramedullary cavity of the bone and ®xed in position [10]. The disc is covered on the upper and lower surfaces
using screws or friction ®t approach (Fig. 1). From the by a thin layer of cartilaginous endplates, which contain
surgery point of view they can be inserted through a small perforations that allow the exchange of water, nutrients
skin incision without opening the fracture site which is and products of metabolism. The main role of the disc is
not the case with the bone plates. However, the insertion to act as a shock absorber for the spine, to cushion
of nail often requires reaming of the medullary canal, adjacent vertebral segments. A number of spine related
which a€ects intramedullary blood vessels and nutrient disorders is identi®ed over the years. Often reported
arteries. As opposed to the plate system mentioned spine disorders include metastasis of vertebral body and
above, the intramedullary nail ®xation method places the disc, disc herniation, facet degeneration, stenosis, and
neutral-axis of the nail-bone structure at the center of the structural abnormalities such as kyphosis, scoliosis, and
bone itself. This also allows early mobilization and load spondylolistheses. Often one disorder has cascading
bearing of the limb without the plaster support. In the e€ect on the other, and primary causes of many spinal
case of plate ®xation system, the neutral axis of the plate- disorders remain largely speculative. A variety of rea-
bone structure is along the plate, and dynamic forces may sons including birth deformities, aging, tumorous
cause fatigue failure of plate or screws. The nail must be lesions (metastasis), and mechanical loads caused by
of sucient strength to carry the weight of the patient sports and work, lead to spine disorders.
without bending in either ¯exure or torsion, yet not com- In the case the defect is limited to few vertebrae alter-
pletely disrupt the blood supply. In order to achieve these native approaches such as: (a) spinal fusion and (b) disc
objectives intramedullary rods with a number of cross- replacement are used. These methods are used alone or in
sectional areas and end designs have been employed. combination depending on the patient condition and
Stainless steel is one of the widely used materials in prognosis. In broader sense, spinal fusion means surgical
intramedullary nails. Recently, Lin et al, [145] proposed immobilization of joint between two vertebrae. Various
short GF/PEEK composite material for intramedullary methods are employed in spinal fusion. One such
application. The rationale behind this proposal is the approach is the surgical removal of the a€ected (portions
claimed biocompatibility of the composite material and of) vertebrae and restore the defect using synthetic bone
its matching mechanical properties compared to the cor- graft, as the autologous or homologous bone grafts are
tical bone. Kettunen et al. [122] developed unidirectional limited by risk of infection, shortage of donor bone sites
carbon ®ber reinforced liquid crystalline (Vectra A950) (with risk of AIDS and hepatitis in the case of auto-
polymer composite intramedullary rod. The material is logous donors), and postoperative resorption and col-
biologically inert, with ¯exural strength higher than the lapse of the graft. Synthetic bone graft material must
yield strength of stainless steel and elastic modulus close have adequate strength and sti€ness, also capable of
to the bone. Compared to the plate ®xation, the intrame- bonding to the residual vertebrae. Ignatius et al. [109]
dullary nail ®xation is better positioned to resist bending and Claes et al. [50] developed Bioglass/PU composite
since it is located in the center of the bone. However, its material for vertebral body replacement. Similarly
torsional resistance is much less than that of the plate, Marcolongo et al. [151] developed Bioglass/PS compo-
which may be physiologically critical. site material for bone grafting purposes. In vivo studies
indicated that these materials are bioactive and facilitate
2.2. Spine instrumentation direct bone bonding (osseous integration). Another
approach is to use special vertebral prostheses such as
The spine serves two distinct and apparently con¯ict- baskets, cages, and threaded inserts, which are made of
ing roles. First, it must provide a strong, yet mobile, cen- metals or bioceramics [240,259]. They are designed such
tral axis onto which the appendicular skeleton is applied. that tissues grow into the prostheses there by ensuring
Second, it must protect the spinal cord and the roots of rigid anchoring of protheses to the bone. Sometimes
delicate nerves connecting the brain to the periphery. The stainless steel or titanium rods, plates, and screws are
1198 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

used in conjunction with these prostheses to provide prostheses either alone or in combinations. However,
necessary stabilization. Several problems have arisen their performance is not yet been acceptable for long-
with these devices. Due to the poor form ®t of these term applications. To date, there has been no arti®cial
implants, local stress concentrations are considered as a disc that is able to reproduce the unique mechanical and
possible reason for bone resorption and implant loos- transport behavior of a natural disc satisfactorily. This
ening. Additionally the metallic implant systems com- may be as a result of the diculty in ®nding a suitable non-
plicate postoperative assessment with X-rays, computed human experimental model to test devices in vivo. For
tomography (CT), and magnetic resonance imaging total disc replacement, it is important to select materials
(MRI) through re¯ection and artifacts. Inadequate bio- and create designs, which possess the required bio-
mechanical capabilities of bioceramic prostheses may compatibility and endurance, while providing kinematic
lead to the collapse of instrumented spine and injury of and dynamic properties similar to the natural disc.
neurological structures and blood vessels. To over come Structural abnormalities or curvatures (lordosis,
disadvantages of conventional materials, Brantigan et kyphosis, and spondylolistheses) of spine are corrected
al. [32] and Ciappetta et al. [46] developed CF/PEEK using either external or internal ®xations. Splints and casts
and CF/PS composite cages for lumbar interbody fusion. form the external ®xation devices. The internal ®xations
The composite cage has an elastic modulus similar to that require surgery and there are many types of instrumenta-
of the bone, thus eliciting maximum bone growth into the tion (screws, plates, rods, and expanding jacks) available
cage. The composite cages are radiolucent and therefore [33]. In some cases, an adjustable stainless steel rod, also
do not hinder radiographic evaluation of bone fusion. known as a Harrington spinal distraction rod, is used to
Moreover they produce fewer artifacts on CT images than stabilize or straighten the curvature. The rod is attached
other implants constructed of metal alloys. Researchers to the spinous process at two points and by adjusting
also developed CF/PEEK and CF/PS [44,185] composite the rod length between the attachment points, the spine
plates and screws for stabilizing the replacement body and is straightened. Schmitt-Thomas et al. [213] made initial
spine. Flexural and fatigue properties of the CF/PEEK attempts to develop a polymer composite rod using uni-
composites are comparable to those of the stainless directional and braided carbon ®bers and biocompatible
steel, which is normally used for spine plates and screws. epoxy resin. The main motivation for this work is to over
The success rate of spinal fusion is poorly de®ned in the come the problems of metal alloys such as corrosion and
literature and varies in a very wide range between 32% interference with the diagnostic techniques.
and 98%. Biomechanical study also shows that fusion It may be noted that ecient ®xation of spinal defor-
alters the biomechanics of the spine and causes increased mities is dicult. This is attributed to the irregular
stresses to be experienced at the junction between fused shape of the vertebrae, and complex and large forces the
and unfused segments. This promotes further disc degen- prostheses need to withstand. Most of the designs used
eration. This seems to contradict a primary purpose of the in various spine instrumentation, and the criteria that
patient seeking treatment and that is to improve the have evolved are primarily based on general biologic
mobility of his back, in addition to alleviating the pain. and engineering principles. Unfortunately, the speci®c
Such arguments have given rise to intervertebral disc mechanical and physical properties required for ideal
prostheses. spine instrumentation have not yet been de®ned. Until
Problems related to intervertebral discs are treated by controlled clinical investigations provide these guide-
replacing a€ected nucleus with a substitute material or by lines, many materials and designs must be evaluated in
replacement of the total disc (nucleus and annulus) using the laboratory.
an arti®cial disc [17]. Both methods require duplication of
the natural structure, signi®cant durability to last longer 2.3. Joint replacements
than 40 years, and ease and safety during implant place-
ment or removal. Some researchers used metal balls to Joints enable the movement of the body and its parts.
replace the nucleus after discectomy. These nucleus Many joints in the body are synovial types, which per-
substitutes did not restore the natural ¯exibility of the mit free movement. Hence, we are able to do various
disc. Problems included migration and subsidence of the physical activities such as walk, jog, run, jump, turn,
balls into the vertebral bodies as pressure was not bend, bow, stand, and sit in our daily life. Hip, knee,
evenly distributed, and no pressure modulation was shoulder, and elbow are a few common examples of
possible with position change. Concurrent to the devel- synovial joints. They all posses two opposing articular
opment of metals balls, other researchers proposed surfaces, which are protected by a thin layer of articular
injectable silicone elastomers or hydrogels as nucleus sub- cartilage and lubricated by elastic-viscous synovial ¯uid.
stitutes. Several arti®cial disc designs are proposed over The ¯uid is made of water, hyaluronic acid, and high
the years [17]. A variety of materials such as stainless steel, molecular weight mucopolysaccharides. The synovial
Co±Cr alloy, PE, SR, PU, PET/SR [202,203,239], and ¯uid adheres to the cartilage and upon loading can be
PET/hydrogel [8] composites are proposed for disc permeated out onto the surface to reduce friction. The
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1199

coecient of friction in a synovial joint is less than 0.01, tissue reaction and even formation of granuloma. This
better than that of a skate blade on ice. Coordinating is attributed to its low compressive sti€ness and strength,
the ligaments, tendons, and muscles performs the actual and increased wear under high stresses during sliding.
articulation of the joint. Osteoarthritis is one of the PTFE is no longer used in such load bearing applications.
common causes for joint degeneration and sometimes Subsequently acetabular cups made of UHMWPE were
hypertrophic changes in the bone and cartilage of joints developed and found to be successful. The UHMWPE
in middle aged people. This is associated with pro- cups are usually supported with a metal backing. Some
gressive wearing down of opposing joint surfaces with reported data suggest that creep deformation, plastic
consequent distortion of joint position. Joints also distortion, and high wear or erosion of UHMWPE is
become damaged upon exposure to severe mechanical possible. Although the short-term function of
or metabolic injury. Over the years a number of arti®- UHMWPE acetabular cups is satisfactory, their long-
cial joints have been designed to replace or augment term performance has been a concern for many years.
many joints in the body. Unlike those used to treat bone To improve the creep resistance, sti€ness and strength,
fractures, the arti®cial joints are placed permanently in researchers proposed reinforcing UHMWPE with car-
the body. The extensive bone and cartilage removed bon ®bers [209,216,222] or UHMWPE ®bers [61]. Deng
during implantation makes this procedure irreversible. and Shalaby [61] found no appreciable di€erence in wear
Considering the extent of loading, complexity of joint properties of reinforced and unreinforced UHMWPE.
function, and severity of the physiological environment, With opposite results reported in the literature, the e€ect
joint replacement is one of the most demanding of all of carbon ®bers on the wear characteristics of the
the implant applications in the body. The most com- UHMWPE is a controversial subject. In recent years,
monly used arti®cial joints are total hip replacement certain designs use dense alumina or zirconia ball and
(THR) and total knee replacement (TKR) (see Fig. 1). matching acetabular cup made of similar materials mainly
because of potential advantages of ceramic materials in
2.3.1. Total hip replacement terms of high hardness and compressive strength, low
THR is the most common arti®cial joint in human coecient of friction, low wear rate, and good biological
beings [63]. For example, over 150,000 total hip repla- acceptance of wear particles.
cements are performed every year in USA alone. Over Although THRs are used widely, one of the major
the years the design of total hip replacement evolved unsolved problems in this important application has
completely from a simple intuitive design to bio- been the mismatch of sti€ness of the femur bone and the
mechanics based functional design. A typical THR prosthesis. As mentioned above the commercial hip
consists of a cup type acetabular component, and a joint stems are made from metal alloys, which are iso-
femoral component whose head is designed to ®t into tropic and at least ®ve to six times sti€er than the bone.
the acetabular cup, thus enabling joint articulations. It has been acknowledged that the metallic stems due to
The shaft of the femoral component (also called femoral sti€ness mismatch induce unphysiological stresses in the
stem) is tapered such that it can be ®xed into a reamed bone, thereby a€ecting its remodeling process. It is dis-
medullary canal of the femur. Several types of THRs cussed that this leads to bone resoprtion and eventual
are designed by changing the material and geometry of aseptic loosening of the prosthesis (it may be noted that
acetabular cups and femoral stems, and ®xation meth- the aseptic loosening is also linked to wear particles/
ods. Conventional THRs use stainless steel, Co±Cr and debris) [9,24,37,214,244,247,251]. This is particularly a
Ti alloys for the femoral shaft and neck, and Co±Cr problem with young and more active patients. This may
alloy or ceramics such as alumina and zirconia materials cause severe pain and clinical failure necessitating repeat
for the head or ball. Earlier designs of acetabular cups surgery. About 10±15% fail within 5±7 years. Gese et al.
were made of Co±Cr alloy. An e€ort to minimize fric- [74] demonstrated that Ti alloy stems result in a 50%
tion and eliminate metallic wear on particles led reduction in the femur peak stress compared to the Co±
Charnley in the early 1960s to use polymers for the Cr alloy stem. It has been acknowledged that the
acetabular component. He ®rst implanted the stainless implant loosening and eventual failure could be reduced
steel femoral component with a mating acetabular through improvements in the prosthesis design and
component made of PTFE. The PTFE was selected for using a less sti€ material with mechanical properties close
a number of reasons. It has a high thermal stability, it is to the properties of bone (i.e. isoelastic materials). How-
hydrophobic, stable in most types of chemical environ- ever, because of the high strength requirement for hip
ments, and generally considered to be inert in the body. prosthesis design, materials suitable for these implants are
It does not adhere to other materials. It has the lowest very limited. Fortunately, the advanced polymer compo-
coecient of friction of all solids. However, clinical sites can o€er strength comparable to metals, and also
studies involving PTFE acetabular cups in the total hip more ¯exibility than metals. Strength of composite
replacement prostheses showed unacceptably high wear stems can be changed without a€ecting sti€ness and vice
and distortion. The wear debris resulted in extensive versa. More over they also o€er the potential to tailor
1200 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

implant properties by selecting material ingredients and


spatially controlling ingradient composition and con®g-
uration, which is useful in reducing the development of
high stress regions. This allows one to control engineer-
ing properties such as strength and modulus according
to the performance requirements of the prosthesis. A
prosthesis made of polymer composite with spatially or
locally varying mechanical properties along the bound-
ary of the prosthesis, results in a more uniform and
ecient transfer of stress from the stem to the bone.
This may lead to better bone remodeling and longer
implant service life. Researchers introduced CF/PS [222]
and CF/C [45] composite stems. They reported faster
bone bonding in the case of composite implants com-
pared to the high sti€ness conventional implants. The
quicker bone bonding or bone contact was attributed to
the lower sti€ness of the implant. The composite stems
were found to be stable with no release of soluble com-
pounds, and high static and fatigue strength. Chang et
al. [40] made CF/epoxy stems by laminating 120 layers
of unidirectional plies in a pre-determined orientation
and stacking sequence. Simoes et al. [220] made com-
posite stems using braided hybrid carbon±glass ®ber
preforms and epoxy resin. Some researchers
[4,185,259,261] designed and injection molded CF/
PEEK composite stems (Fig. 4), which possess a
mechanical behavior similar to that of the femur. Ani-
mal studies indicated that CF/PEEK composite elicits
minimal response from muscular tissue. Both the in vivo
and in vitro aging studies con®rmed mechanical stability
of CF/PEEK up to 6 months (it may be noted that this
period is short and further long term testing is needed). Fig. 4. An injection molded CF/PEEK composite stem for total hip
joint replacement.
Finite element analyses and in vitro measurements
[4,268,269] indicated that compared to conventional
metallic stems more favorable stresses and deformations Currently the cemented and ingrown approaches are
could be generated in the femur using composite stems. widely used. As the name suggests in the ®rst approach,
Due to complexities in the geometry of hip prostheses, the implant is secured in the bone by press ®tting and/or
hip loads, and material properties of composites, design using a wide range of pegs, posts, and screws. In the last
of composite implants require greater attention in order method, ®xation is achieved by direct adhesion of stem
to achieve the desired in vivo performance of the to the bone. In the `cemented' approach, the PMMA or
implants. It is in order to mention here that if one tries PMMA variant bone cements are used to ®x the total
to reduce stress shielding by using a less sti€ implant it hip replacement. More details of bone cements are
leads to increased implant deformation and relative described in Section 2.3.4. The quality of cemented pros-
movement (also called micromotion) between the theses ®xation depends on various factors such as cement
implant and bone tissue during loading. The micromotion thickness, voids in cement or blood and tissues in contact
also in¯uences bone remodeling [214,244] and often leads with the cement bed during operation [31,53,119,143,165].
to residual pain. The stress shielding and micro motion Problems cited include thermal damage to the bone due to
are con¯icting phenomena [104,134]. In other words, for cement curing, cytotoxic e€ects of methacrylate mono-
appropriate structural compatibility the implant design mers, migration of cement and other wear particles in the
should reduce stress shielding and micromotion simul- cement±bone interface or the physiological process of
taneously. bone resorption and intramedullary canal widening
In addition to the prosthesis design and material, the [64,103,191,252,266]. The best way to overcome these
®xation method is also important for the success of problems is not use the bone cement. An alternative
THRs [207]. Various methods for ®xing THRs to the approach, known as `cementless approach', promotes
bones can be grouped into four generic types namely ®xation by encouragement of tissue growth into porous
mechanical means, cemented, ingrown, and adhered. surface of the stem. Porous surface coatings have been
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1201

fabricated from various materials including Bioglass, ness, strength and creep resistance properties of reinforced
bioactive glass-ceramics, hydroxyapatite, and bioactive UHMWPE are desirable for the joint replacement
polymers. In other designs the prosthesis surfaces are application.
sintered with metal wire meshes or beads. The surface
bioactiveness and/or porosity facilitate in growth of 2.3.3. Other joint replacements
bone tissues and thus good anchoring of the prosthesis Other joint replacements include ankle, toe, shoulder,
to the bone. The main shortcoming of these cementless elbow, wrist, and ®nger joints. The success rate of these
approaches is that the long time required for achieving joint replacements is limited due to loosening of pros-
rigid ®xation. On the other hand, in the case of cemen- theses and hence they are used less commonly compared
ted implants, the ®rm ®xation is immediate. with THR and TKR. The prostheses failures are attrib-
uted to limited bone stock available for ®xation, minimal
2.3.2. Total knee replacement ligamentous support, and high stresses on the pros-
The knee joint has a more complicated geometry and theses. More details on these joints can be found in
biomechanics of movements than the hip joint. The references [178,179]. Materials such as Co±Cr and Ti
incidence of knee injuries and degeneration is higher alloys, HDPE, and UHMWPE remain to be the candi-
than most other joints. Similar to most other joint date materials for these joint replacements. Some designs
replacements, the knee joint replacement development use CF/UHMWPE instead of UHMWPE to provide
has been an evolutionary process, relying on intuitive higher strength and creep resistance. In certain types
design, empirical data, and laboratory studies. A typical (space ®ller design) of ®nger joint replacements, silicone
TKR mainly consists of femoral and tibial components rubber (SR) is considered. Tearing of SR at the junction of
(Fig. 1). The femoral component articulates on the tibial prosthesis and roughened arthritic bone is a major con-
component. The materials used for femoral components cern. In order to improve the tear strength and ¯exural
are predominantly Co±Cr and Ti alloys [245]. The tibial properties of SR, it is reinforced with PET fabrics.
component is made of UHMWPE polymer supported Goldner and Urbaniak [79] reported that the composite
by a metallic tibial tray. Clinical data indicated that the prosthesis also successful in decreasing pain, improving
UHMWPE undergoes cold deformation, which leads to stability, increasing hand function, and in providing an
sinking of prosthesis. Inoue et al. [111] simulated and adequate range of motion.
compared the performance of metal alloy femoral com-
ponent articulating on a UHMWPE tibial component, 2.3.4. Bone cement
and metal alloy femoral component articulating on a Proper ®xation to the bones is as important as the
®ber reinforced UHMWPE composite tibial compo- design of joint replacement itself. Several di€erent
nent. It is reported that the former material combination methods are adopted for ®xing the arti®cial joints to the
resulted in a high stress concentration in the vicinity of bones. One of the earliest methods, is to press-®t the
tibial stem, whereas the later material combination joint prosthesis into the bone using a grouting material
resulted in minimal stress concentration. This also called bone cement. The most widely used bone cement
explains the reasons for sinking of knee prostheses. Car- is based on PMMA, also called acrylic bone cement
bon ®bers were used to reinforce UHMWPE to reduce its [210]. It is self-polymerizing and contains solid PMMA
cold ¯ow (creep) deformation [219]. The reinforcement powder and liquid MMA monomer. It has minimal
enhances the sti€ness, tensile yield strength, creep resis- adhesive properties, because of which attachment
tance, and fatigue strength of UHMWPE [265]. How- requires undercuts, holes, or furrows in the prosthesis.
ever, the results describing the e€ect of carbon ®bers on Therefore, when the bone cement sets or hardens, it
the wear characteristics of UHMWPE are contradictory. mechanically interlocks with the roughened bone sur-
Early studies reported that wear is reduced because of face and the prosthesis. Cement must endure consider-
carbon ®bers. But the later studies reported that the able stresses in in vivo applications, thus strength
composite wear rates were 2.6±10.3 larger than those of characteristics are important for its clinical success. The
unreinforced UHMWPE. This was attributed to the poor main function of the bone cement is to transfer load
bonding between the carbon ®bers and UHMWPE. The from the prosthesis to the bone or increase the load
addition of carbon ®bers does not improve the resistance carrying capacity of the surgical construct. Researchers
of the material to surface damage. It should be empha- expressed concern over the release of monomers into the
sized that the composite by itself may not be suitable for blood stream. Concerns were also expressed about the
low friction bearing but a combination of a UHMWPE exothermic reaction associated with polymerization
surface and a composite substrate appears to o€er some process, which produces elevated temperatures in the
advantages. Recently, Deng and Shalaby [61] reinforced tissues that may induce locally bone necrosis [64]. The
UHMWPE polymer with UHMWPE ®bers. They repor- polymerization process is also associated with undesir-
ted no di€erence in the wear characteristics of unreinforced able shrinkage of acrylic polymer. Another issue is the
and reinforced UHMWPE. However, the improved sti€- deterioration of cement/implant or cement/bone inter-
1202 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

face with time, leading to problems of mechanical fail- complex internal cavities in bone, it must harden in situ
ure and instability [31]. Fatigue failure has been found and develop mechanical properties sucient to permit
to be a predominant in vivo failure mode of bone functional loading of the implant site, it should maintain
cement [114,131]. Researchers have tried to improve adequate mechanical integrity long enough to provide
bone cement mechanical properties by reinforcing with useful stabilization of the implant, and it should not be
stainless steel and Ti alloy wires, and polymer ®bers such a barrier to bone remodeling.
as UHMWPE [192,231,243,267], Kevlar, carbon [189], It is in order to mention that wear of articulating sur-
and PMMA [76]. Use of such ®ber reinforcement also faces is the major concern of many joint replacements [21].
reduces the peak temperature during polymerization of Particulate debris that is formed becomes incorporated
the cement, and thus reducing the tissue necrosis [231]. into the surrounding tissues, and even though the mate-
The reinforced cement posses higher fracture toughness, rial may be quite inert in the bulk form, the ®ne parti-
fatigue resistance and damage energy absorption cap- cles are much more reactive and thus cause tissue
abilities than the unreinforced cement. In another irritation and in¯ammation. This process if repeated
approach, bone particles or surface-reactive glass pow- excessively, leads to bone resorption, bone loss, implant
ders are mixed with PMMA bone cement in order to loosening, and fracture of bone. Hence, wear rate and
combine immediate mechanical ®xation of PMMA with wear products are of great importance in the design of
chemical bonding of bone particles [137,175,179] or joint replacements. Many e€orts have been made to mea-
surface-active glasses (Bioglass) with the bone [225]. sure the rate of wear debris production in the laboratory.
Formation of this chemical bond makes it possible for In general, the results depend on the geometry of the test,
mechanical stresses to be transferred across the cement/ on the lubricant selected to simulate synovial ¯uid, and to
bone interface in a manner that prevents the fracture of some degree, on the experimenter. There have been great
the interface even when the implant or the bone is loa- diculties encountered in reproducing in vitro experi-
ded to failure. Despite the experimental evidence of mental results. Due to the inherent complexity of con-
superior mechanical performance, reinforced cements ducting a wear test, the exact mechanisms of wear and
have not yet been accepted in current clinical practice, wear rate, and isolated e€ects of wear debris on the
primarily because of limitations such as the addition of body are not clear. It is believed that more than one
®bers increases the apparent viscosity of bone cement mechanism may take place simultaneously. Many stu-
thereby severely decreasing its workability and deliver- dies are being conducted to understand the local and
ability. Furthermore, uniform distribution of ®bers in systemic e€ects of wear particles or debris.
the bone cement is dicult, if not impossible, to obtain.
Gerhart et al. [71] proposed partially resorbable bone 2.4. Bone replacement (synthetic bone graft) materials
cement, which is a composite of tricalcium phosphate
particles and a gelatin matrix. It is intended to provide Synthetic bone grafts are necessary to ®ll bone defects
immediate structural support and subsequent resorption or to replace fractured bones [128]. The bone graft
of resorbable component of the composite cement facil- material must be suciently strong and sti€, and also
itates bone ingrowth and direct bonding by the host capable of bonding to the residual bones. PE is con-
bone. In contrast, the standard PMMA bone cement sidered biocompatible from its satisfactory usage in hip
does not permit direct bonding by the host bone even and knee joint replacements for many years. Sti€ness
though it provides the immediate structural support. and strength of PE are much lower than those of the
PMMA is vulnerable to the accumulation of fatigue bone. For load bearing applications, properties of PE
damage, as repetitive mechanical stresses lead to need to be enhanced. In order to improve the mechan-
loosening at the cement±bone interface. It is in order ical properties some researchers [25,26,58,91,223,255]
to mention that the usefulness of the partially resorb- reinforced PE using HA particles, which are bioactive.
able bone cement may be limited by a tendency for The resulting composite has an elastic modulus of 1±8
particle migration away from the implant site. More- GPa and strain to failure value of over 90±3% as the
over the strength of partially resorbable bone cement is volume fraction of HA increases to 50%. It was repor-
considerably lower than that of the PMMA bone ted that for HA particulate volume fractions above 40%
cement. the composite is brittle. More over the bioactivity of the
Optimum use of bone cement is very important, other- composite is less than optimal because the surface area
wise, cement failure leads to loosening of the implant, of HA available is low and the rate of bone bonding of
which in turn causes pain to the patient. As the implant HA is slow. Further work requires consideration of
loosens, greater loads are experienced by the implant. using more bioactive materials such as Bioglass as rein-
Excessive loosening necessities removal of the implant and forcements in PE [91,92]. A typical composition of Bio-
also some times leads to implant failure. The guiding glass is 45% SiO2, 6% P2O5, 24.5% Cao, 24.5% Na2O
principles for developing new bone cements include, the by weight. The Bioglass reacts with physiological ¯uids
cement can be shaped, molded or injected to conform to and forms tenacious bond to hard or soft tissues
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1203

through cellular activity. To increase the interface gold are mainly used in the restoration of posterior
between HA particles and the bone tissues, some teeth, and not preferred for anterior teeth for cosmetic
researchers developed partially resorbable composites. reasons. Moreover there is concern over the long-term
They reinforced resorbable polymers such as PEG, PBT toxicity of silver-mercury amalgam ®llings. Acrylic
[146], PLLA [96,205,241,242], PHB [25,126], alginate resins and silicate cements have been used for anterior
and gelatin [124] with bioactive particles. Upon implan- teeth. However, they exhibit poor mechanical proper-
tation, as the matrix polymer resorbs, more and more ties, which lead to short service life and clinical failures.
bioactive particles come in contact with the growing tis- Dental composite resins, which are translucent with a
sues, thus achieving good integration of the biomaterial refractive index matching that of the enamel, have vir-
into the bone. The wide range of material combinations tually replaced these materials and are very commonly
o€ers the possibility of making composites with various used to restore posterior teeth as well as anterior teeth.
desired properties such as sti€ness, strength, biodegrada- The dental composite resin comprises of BIS-GMA as
tion, and bioactivity. the matrix polymer and quartz, barium glass, and col-
loidal silica as ®llers. The BIS-GMA is derived from the
2.5. Dental applications reaction of bis (4-hydroxyphenol) and glycidylmetha-
crylate. Low viscosity liquids such as triethylene glycol
All teeth are made of two portions, crown and root, dimethacrylate are used to lower the viscosity and inhi-
which are demarcated by the gingiva (gum). The root is bitors such as BHT (butylated trioxytoluene, or 2,4,6-
placed in a socket called alveolus in the maxillary tri-tert-butylphenol) are used to prevent premature
(upper) or mandibular (lower) bones. Teeth possess a polymerization. Polymerization can be initiated by a
thin (<1 mm) surface layer of highly mineralized (90%) thermochemical initiator such as benzoyl peroxide, or by
dental enamel (the hardest substance found in the a photochemical initiator (benzoin alkyl ether) that gen-
body). The calcium salts of enamel are arranged as ®ne erates free radicals when subjected to ultraviolet light
prisms running perpendicular to the surface. Underlying from a lamp used by the dentist. In other types of com-
and supporting this is dentine, a less mineralized (70%) posites a urethane dimethacryate resin is used rather
tissue that contains ®ne liquid-®lled tubules running than the BIS-GMA. The ®ller particle concentration
through to the pulp chamber. The pulp chamber carries varies from 33 to 78% by weight and size varies from
the nerve and extends up through the root to the center 0.05 to 50 mm. The glass ®llers reduce the shrinkage
of the tooth. In place of enamel, the surface of the root upon polymerization of the resin, and also the coe-
portion of tooth is covered by cementum, a mineralized cient of thermal expansion mismatch between the com-
tissue similar to bone. Teeth are non-homogenous, ani- posite resin and the teeth. They impart high sti€ness
sotropic, and unsymmetrical. Teeth experience a varied and strength, and good wear resistance to the dental
amounts and types (compression, ¯exural, torsion, and composite resins [121]. Strong bonding between the ®l-
their mixed versions) of forces during mastication or lers and resin is achieved using silane-coupling agents
chewing. Masticatory and traumatic forces vary from [132]. Key requirements for a successful restorative
100N to 450N [54,99,112]. material include: suciently low viscosity so as to enable
Dental treatment is one of the most frequent medical it to ®ll the cavity completely; controllable polymeriza-
treatments performed upon human beings. Dental tion; coecient of thermal expansion similar to the den-
treatment ranges from ®lling cavities (also called `dental tine/enamel, otherwise the stresses due to the mismatch
caries') to replacing fractured or decayed teeth. A large is thought to contribute to leakage of saliva and bac-
variety of materials are used in the dental treatments teria at the interface margins; low shrinkage; and good
such as cavity lining, cavity ®lling, luting, endodontic, resistance to creep, wear and water absorption. When
crown and bridge, prosthetic, preventive, orthodontic, the dental composites are used as a posterior restorative
and periodontal treatment of teeth. These materials are material, their radio-opacity is very important. The
also generally described as biomaterials. The choice of detection of caries under a non-radio-opaque composite
material is dependent on its ability to resemble the is virtually impossible, and would allow the caries
physical, mechanical and esthetic properties of natural process to continue undetected for far too long. It is not
tooth structure. Here we only consider the applications clear what the optimum radio-opacity for composite is,
in which composite materials are used, or the potential since excessive radio-opacity can potentially mask out
of using composite materials, is considerably high. caries lying behind the restoration. Nevertheless, the
Dental restorative materials as the name suggests are composite should at least be as radio-opaque as the
used to ®ll the tooth cavities (caries) and some times to enamel. Active research is being pursued to develop
mask discoloration (veneering) or to correct contour dental composite resins with improved performance.
and alignment de®ciencies. Amalgam, gold, alumina, In cases when the severely damaged tooth lacks the
zirconia, acrylic resins and silicate cements are com- structure to adequately retain a ®lling or restoration,
monly used for restoring decayed teeth. Amalgam and often pins are used. In situations where the amount of
1204 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

coronal tooth structure remaining is small (also referred provides a base that has sucient bulk and retention for
as pulpless tooth), a dental post or a cast dowel is used the ®nal restoration, the crown. Cores are usually
to reinforce the remaining tooth structure [100,149], on formed from dental composite resins or amalgam or
which the core and crown are built (Fig. 5a). The post is may be cast in precious or nonprecious alloys in com-
normally inserted in the root canal and ®xed in position bination with the metal post [237]. Traditionally posts
using dental cement. It provides a retentive support to made of stainless steel, Ni±Cr, Au±Pt or Ti alloys are
the core and crown assembly, and also distributes the used based on the assumption that the post should be
forces of mastication to the supporting structures: the rigid. Failures reported include corrosion of posts,
root, periodontal ligament, and surrounding bone. bending or fracture of posts, loss of retention, core
Sometimes pins are used either alone or in combination fracture and root fracture. In recent years this old basic
with the post to provide retention to the core material. tenet has been strongly questioned and it has been sug-
The core replaces the coronal tooth structure that has gested that the modulus mismatch between the post and
been lost because of caries and previous restorations. It the dentine should be reduced so as to minimize the
occurrence of root fractures (root fracture frequency is
2±4%) and failure of restorations. Newer posts made of
zirconia, short glass ®ber reinforced polyester, and uni-
directional carbon ®ber reinforced epoxy composite
posts [113,234] are introduced. These new posts are
adequately rigid, resistant to corrosion and fatigue
[196]. In the frame of an ongoing project at the National
University of Singapore, one of the authors looked at
the function of a dental post in order to fully under-
stand its mechanical requirements. In addition to pro-
viding support to the core, the dental post also helps to
direct occlusal and excursive forces more apically along
the length of the root. A ®nite element study by Caille-
teau et al. [38] indicated that a post-restored model
results in a decreased level of stress along the coronal
facial portion of the root surface which peaked abruptly
near the apical end of the post (labels 1 and 13 in Fig. 5a
indicate coronal and apical ends respectively). These
®ndings contradict the belief that the conventional posts
strengthen the tooth by evenly distributing the external
forces acting on the tooth. An ideal post should have
varying sti€ness along its length. Speci®cally, the cor-
onal end of the post should have higher sti€ness for
better retention and rigidity of the core, and the apical
end of the post should have lower sti€ness matching
that of the dentine so as to over come the root fractures
due to stress concentration. In other words, it is desir-
able to have a post with varying sti€ness. A post with
varying sti€ness but no change in the cross-sectional
geometry along its length is only possible by using
functionally graded composite materials. Ramakrishna
et al. [201] designed and developed a functionally gra-
ded dental post using braided CF/epoxy composite
technology [70,277]. It has a high sti€ness in the coronal
region and this sti€ness gradually reduces to a value
comparable to the sti€ness of dentine at the apical end.
In addition to overcoming the root fracture, the graded
sti€ness post decreases the chances of the post loosening
from the dentine by means of eliminating stress con-
Fig. 5. (a) Post restored tooth, and (b) normal and shear stress dis-
centration in the dentine, and reduction of post/dentine
tributions along the post-dentine interface. S.Steel indicates stainless
steel post and FGM indicates functionally graded polymer composite interfacial shear stresses (Fig. 5b). This clearly suggests
post. Numbers 1 and 13 on the x-axis correspond to the coronal and that innovations in composites design and fabrication
apical ends, respectively. lead to better prostheses with improved performance.
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1205

In the extreme case, the damaged or condemned tooth dimensions [273]. Commonly used materials in the
is extracted and replaced with a dental implant. Dental manufacture of brackets are stainless steel, polycrystal-
implants are an arti®cial tooth roots that permanently line alumina, and single crystal alumina. Brackets made
replace missing teeth, and they are an alternative to from metal alloys show high strength and sti€ness but
bridges or false teeth. The dental implant may be su€er from poor aesthetics. The ceramic brackets have
designed to enter the jawbone or to ®t on to the bone improved aesthetics, however, ceramic brackets are
surface. The types of dental implants available are bulkier than the metal alloy brackets. Furthermore,
numerous. For example, certain root forms have threads, ceramic is abrasive to tooth enamel, and this has,
which facilitate to secure the root form into the jaw bone, therefore, limited the use of ceramic brackets to upper
whereas in some other designs, the surface is coated with teeth. Some patients are hypersensitive to metals (Ni,
porous bioactive materials, which allow bone growth and Cr, and Co). It has been reported that these patients'
osseointegration. They are made of a wide range of mate- immune system responds with vigorous foreign body
rials [36,274] such as metals (Co±Cr±Mo alloys, Ti alloys, allergic reactions causing dermatitis. Use of metallic
stainless steel, platinum, silver,), ceramics (zirconia, restorations or braces is not recommended for metal
alumina, glass, and carbon), polymers (UHMPE, sensitive patients. There is a need to develop suitable
PMMA, PTFE, PS, and PET), and composites (SiC/ polymer composite orthodontic brackets. For any
carbon and CF/carbon) [1,35,148,166]. Compared to material combination to succeed in orthodontics, it is
ceramic and metal alloys, the outstanding properties of also important to consider the friction and abrasive
composites are high or sucient strength combined with wear characteristics of arch wires and brackets.
low modulus. Such composite materials may o€er pro-
tection against the alveolar bone resorption. Moreover
fatigue properties of composites are far superior to the 3. Soft tissue applications
metal alloys and ceramics. The dental implants need to
be designed to withstand extremely large and varying Many di€erent types of implants are used in the surgery
forces applied during mastication. to correct soft tissue deformities or defects which can be
A bridge is a partial denture (false teeth) used to congenital, developmental, or acquired defects, the last
replace one or more tooth completely. In an extreme case category usually being secondary to trauma or tumor
removable dentures are used to overcome the loss of all excision. Depending on the intended application, the soft
the teeth. A large percentage of adults over the age of 50 tissue implants perform various functions: ®ll the space
years have full or upper or lower dentures. The root form from some defect; enclose, store, isolate, or transport
mentioned previously is also used to anchor dentures and something in the body; and mechanical support or serve
bridges to the jawbone. The high cost and time consuming as a sca€old for tissue growth.
preparation of current gold bridges has led to the
development of relatively inexpensive and easy to use 3.1. Bulk space ®llers
CF/PMMA [19], KF/PMMA [93], UHMWPE/PMMA
[56] and GF/PMMA [164] composite bridges and den- Bulk space ®llers are used to restore cosmetic defects,
tures [67]. atrophy, or hypoplasty to an aesthetically satisfactory
Orthodontic arch wires (approximately 0.5 mm dia- condition [158]. They are mostly used in the head and
meter) are used to correct the alignment of teeth. This is neck [39]. The materials used in these applications include
facilitated by bonding orthodontic brackets on to the SR, PE and PTFE [59,84]. The space-®llers are also inves-
teeth. An arch wire is placed through the brackets and tigated for the replacement of articular cartilage in the case
retained in position using a ligature, a small plastic piece. of its deterioration by osteoarthritis. Articular cartilage, 1±
By changing the tension in the arch wire the alignment 2 mm thick, covers the opposing bony surfaces of typical
of the teeth is adjusted. The bracket acts as a focal point synovial joints. The cartilage provides a means of
for the delivery of forces to the tooth generated from absorbing force and provides low-friction bearing sur-
wire. It is important for a bracket to have high strength faces for joints. The cartilage replacement material must
and sti€ness to prevent distortion during tooth move- be hydrophilic with controlled water content, must have
ment. This technique is also used to splint the trauma- sucient strength, and should be very smooth. Poly-
tized teeth. Traditionally, the arch wires were made of mers such as SR and PTFE [178] are proposed to ®ll the
stainless steel and Ni±Ti (beta titanium) alloys. Jancar defects in the articular surfaces or to replace meniscus
and Dibenedetto [115], Jancar et al. [116] and Imai et al. or ®brous tissues following the condylar shave or high
[110] proposed GF/PC, GF/Nylon, GF/PP, and GF/ condylectomy in the treatment of painful arthritis and
PMMA composite materials for arch wires. The stated to restore normal joint function. Messner and Gillquist
advantages of using composite arch wires include aes- [163] reported that composites comprising PET or
thetics, easy forming in the clinic, and the possibility PTFE fabrics and PU are more suitable for this
of varying sti€ness without changing component purpose, as they were found to reduce the cartilage
1206 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

degeneration following the meniscectomy. At the same artery pressure, gases, collecting blood samples for
time Pongor et al. [190] clinically used woven carbon ®ber analysis). PU and SR are widely used materials for
fabrics and their composites for the treatment of carti- making catheters because of their ¯exibility and ease of
lage defects. No in¯ammatory change or deterioration fabrication into a variety of sizes and lengths in order to
in joint damage was reported, indicating the usefulness accommodate the wide range of vessels to be cannu-
of the prostheses. Further improvements in the compo- lated. SR is reinforced with silica particles to improve its
site materials in terms of retaining the shape of the tear strength and to decrease wettability. Andreopoulos,
implant could further improve the joint biomechanics. et al. [11] reported that with increasing the volume
fraction of silica particles up to 35%, the tensile
3.2. Encapsulants and carriers strength and elongation at break increased, whereas the
elastic modulus only changed marginally. Since the
3.2.1. Wound dressing catheter interfaces with blood, it is important that its
Burn victims are often treated with skin dressings. In design and material properties ensure blood compat-
order to conform to irregular surfaces, the skin dressing ibility, nonthrombogenicity, and inhibit infection. An
must be elastic and ¯exible. There are two opposite ideal vascular catheter also must be ¯exible enough to
requirements for skin dressing to meet: it should prevent allow vein and patient movement without becoming
loss of ¯uids, electrolytes and other biomolecules from extravascular and damaging both the vessel and the sur-
the wound and obstruct bacterial entry, but it should rounding structures. Catheters that are initially supple
also be permeable enough to allow the passage of dis- may become brittle over time, resulting in vascular wall
charge through pores or cuts. In addition it should be damage. Newer designs consist of polymers (PU, LDPE,
able to adhere to the wound surface, and be easy to peel and PVC) reinforced with braided Nitinol (Ni±Ti alloy)
from the skin without disturbing new tissue growth. ribbons with the purpose of making a catheter having
Woven fabrics or porous layers of resorbable polymers an exceptionally thin wall, controlled sti€ness, high
such as collagen, chitin, and PLLA are used in many skin resistance to kinking, and complete recovery in vivo
dressings. In hybrid skin dressings, synthetic polymers from kinking situations.
and cultured cells are combined to form vital/avital
composites. They are designed to initiate, accelerate and 3.3. Functional load-carrying and supporting implants
control the natural skin repair process. Until now there
is no synthetic material that can meet all the require- 3.3.1. Tendons and ligaments
ments of a skin substitute exactly. Arti®cial tendons and ligaments are the best examples
of load-bearing soft tissue implants. A tendon is a strong
3.2.2. Ureter prosthesis ®brous band of tissue that extends from a muscle to the
Ureter prostheses made of PVC, PE, nylon, PTFE, periosteum of the bone. A ligament is a connective tissue
and SR were used without much long-term success. band that links bones in the vicinity of every synovial
They were not very successful because of the diculty joint. Tendons and ligaments hold the bones of a joint
of joining a ¯uid-tight prosthesis to the living system. In thus facilitating their stability and movement. They also
addition, constant danger of microbial infection and transmit force between muscle and bone. They are
blockage of passage by calci®cation deposits from urine essentially composite materials comprising undulated
have proven to be dicult to overcome. Polyester ®ber collagen ®ber bundles aligned along the length and
reinforced glycol methacrylate gel prostheses with a immersed in a ground substance, which is a complex
fabric backing was reported to be successful [92,130]. made of elastine and mucopolysaccharide hydrogel
The fabric backing facilitated easy attachment of a [193]. The unique mechanical feature of tendons/liga-
prosthesis ®rmly on to the mucous membrane without ments is their non-linear J-shaped convex stress±strain
irritation, and the hydrophilic nature of the gel helped curve as opposed to the concave stress±strain relation-
to maintain a clear inner space. A similar solution was ships of common engineering materials. For example,
proposed for the replacement of portions of intestinal the static tensile curve of ligaments characteristically
wall. There is a need to develop new materials with exhibit a `toe' region (low modulus) at low strain, a lin-
improved surface properties of minimal microbial ear region at intermediate strain, and eventually a fail-
adhesion, low friction, and control of cell and protein ure at high strain. Tissue structural parameters such as
adsorption. ®ber composition and structure, hydration, ®ber±matrix
interaction, and ®ber±®ber interaction determine its the
3.2.3. Catheters mechanical behavior. Tendons have little regenerative
Catheters (tubes) are increasingly used to access capacity and require very long times to regenerate fully.
remote regions of the human body to administer ¯uids The use of biomaterials in tendon/ligament repair is
(e.g. nutrients, isotonic saline, glucose, medications, one of the most demanding applications of prostheses in
blood and blood products) as well as to obtain data (e.g. soft tissues. A ligament or tendon prosthesis should: (a)
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1207

possess the same ¯exibility as the natural tissue in order currently using prostheses of natural tissues (homografts,
to bend around articulations and assure the transmis- allografts, and xenografts). Many consider that a com-
sion of the force to the muscle always in the mode of a bination of autegenous tissue and synthetic materials is an
traction (Seedhom, [218] reported that estimated forces ideal choice for tendon/ligament prostheses. These materi-
in the anterior cruciate ligament of the knee joint are als reportedly possess the desired biomechanical properties
196 N for level walking, 72 N for ascending stairs, 93 N such as low coecient of friction, and improved com-
for descending stairs, 67 N for ascending a ramp, and pliance, strength, creep, and fatigue resistance.
445 N for descending a ramp), (b) reproduce similar
mechanical properties including J-shaped stress-strain 3.3.2. Vascular grafts
behavior, large extensibility without permanent defor- Arterial blood vessels are complex, multilayered
mation, and damping properties, and (c) assure time structures comprising collagen and elastin ®bers, smooth
invariance of the mechanical properties. Biomaterials muscle, ground substance and endothelium. The blood
are used in a number of ways in tendon healing. They vessel is anisotropic because of the orientation of inher-
may be used to replace the tendon, they may be used to ent ®brous components. Like other soft tissues, the
hold a damaged tendon in proper alignment, or they blood vessel also behaves in a non-Hookean way when
may be used to form a new sheath. In the last approach, subjected to physiological loads, and displays J-shaped
a two-stage surgical procedure is followed. In the ®rst stress±strain behavior. Vascular grafts are used to
operation, the tendon is replaced by a gliding implant replace segments of the natural cardiovascular system
that facilitates the formation of a new tendon sheath. In (mainly successful in the case of blood vessels with
the second operation, a tendon graft replaces the gliding lumen diameter of over 5 mm) that are diseased or
implant inside the newly formed sheath. blocked (atherosclerosis, deposits on the inner surface
Synthetic biomaterials used thus far include UHMWPE, of the vessels restricting the ¯ow of blood and increas-
PP, PET, PTFE, PU, Kevlar 49, carbon, and recon- ing blood pressure). A typical example is to replace a
stituted collagen ®bers in the multi®lament form or brai- section of aorta where an aneurysm has occurred.
ded form [13,18,62,66,117,123,152,161,183,186,228,248]. Another example is the arteries in the legs of diabetic
Permanent ®xation of the implant assumed to be pro- patients that have a tendency to be blocked. Grafts,
vided by tissue ingrowth into the spaces between the essentially tubular structures, are inserted to bypass the
®laments. The clinical experience with synthetic pros- blockages and restore circulation. Most widely used vas-
theses has so far been disappointing. The problems with cular grafts are woven or knitted fabric tubes of PET
synthetic prostheses include diculty of anchorage to material or extruded porous wall tubes of PTFE and PU
the bone, and abrasion and wear of the prostheses, materials. The most important property of a graft is its
which deteriorate in strength in the long term and lead porosity. Certain porosity is desirable as it promotes tissue
to mechanical failure (such as fatigue). Further, the growth and acceptance of the graft by the host tissues.
particulate matter generated by abrasion against rough However, excessive porosity leads to leakage of blood.
bony surfaces may cause synovitis, as well as in¯amma- Most synthetic grafts are preclotted prior to transplan-
tion of the lymph nodes should the size of the particu- tation to minimize blood leakage. In another approach,
late matter produced allow its migration to the nodes vascular grafts are impregnated with collagen or gelatin
[218]. To reduce particle migration and improve hand- to seal the pores and also to improve the dimensional
ling properties, prostheses are coated with polymers stability of grafts. These are known as composite grafts.
such as SR, poly(2-hydroxyethyl methacrylate) The seal degrades in approximately 2±12 weeks after the
(PHEMA), and PLA. Pradas and Calleja [193] reported implantation. In addition to porosity, good handling
that by combining ¯exible polymer such as PMA or PEA and suturing characteristics, satisfactory healing (rapid
with crimped Kevlar-49 ®bers, the stress-strain behavior tissue growth), mechanical and chemical stability (good
of natural ligaments can be reproduced to a certain tensile strength and resistance to deterioration) are
extent. Iannace et al. [108] and Ambrosio et al. [6,7] major requirements of vascular grafts. Since vascular
developed a ligament prosthesis by reinforcing a hydro- grafts are subjected to static pressure and repeated stress
gel matrix (PHEMA) with helically wound rigid PET of pulsation in application, they should have good dila-
®bers, and demonstrated that both static and dynamic tion and creep resistance. The fabric tubes are crimped
mechanical behavior of natural ligaments can be repro- to make them bulky, resilient, and soft. Moreover,
duced. This has been achieved by controlling the struc- crimping facilitates extensibility, and bending of fabric
tural arrangement of reinforcing ®bers and the tubes without kinks and stress concentrations, which are
properties of the components. It may be noted that PET very important in blood transporting vascular grafts.
is sensitive to hydrolytic, stress induced degradation. PET (Dacron) vascular grafts (woven or knitted fab-
Surgeons are still looking for suitable synthetic materials ric tubes) are mainly successful in the replacement of
that adequately reproduce the mechanical behavior of large diameter blood vessels (12±38 mm diameter). A
natural tissue for long-term application, while they are major issue for a vascular grafts is the reaction between
1208 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

the surface of the material and blood that can cause 3.4. Others
destruction of blood cells and thromboembolism. Bio-
compatibility of PET ®bers and fabrics is generally con- Hernia is an irregular protrusion of tissue, organ, or a
sidered to be acceptable. Protein and platelet absorption portion of an organ through an abnormal break in the
of PET is minimal, however it is thrombogenic. PET surrounding cavity's muscular or connective tissue wall. A
vascular grafts are seeded with endothelial cells to reduce number of materials such as nylon, PP, PTFE, PET, car-
the thrombogenic character and to improve patency. bon, stainless steel, and tantalum in the form of fabrics or
These grafts are essentially composites of PET fabrics meshes are used to repair hernias [246]. The fabrics or
and cells (see Section 5.2 for further details). meshes facilitate tissue ingrowth thus providing stability
Expanded PTFE (e-PTFE or Gore Tex) is widely used to the prosthesis. Recently, Werkmeister et al. [250]
for medium diameter (6±12 mm) vascular grafts. A developed PET fabrics coated with collagen and PU
modi®ed extrusion process produces the porous e-PTFE. materials suitable for repairing hernia and abdominal
The porous non-woven microstructure of e-PTFE pro- wall (abdominal wall lines the abdominal cavity that
vides vascular grafts with a mechanical behavior contains liver, gallbladder, spleen, stomach, pancreas,
matching to that of the host blood vessels compared to intestine, and kidney) defects. The composite is designed
the vascular grafts made of non-porous (solid) materi- to display adequate mechanical properties as well as
als. Moreover, the inner (luminal) surface of e-PTFE facilitate tissue ingrowth. The composite material is
graft facilitates formation of neointima (newly formed reportedly superior to uncoated fabrics in terms of bio-
endothelial tissue lining) that avoid the complications compatibility. Other suitable applications being currently
such as formation of thrombi (blood clot) and emboli investigated include tracheal prostheses (combined with
(dislodged blood clot). However, the exact mechanisms stainless steel mesh or SR), prosthetic sphincters for
of neointima formation are not clear. gastrointestinal tracts, and urethral prostheses.
It is widely accepted now that a major requirement for Prostheses are also used for restoring the conductive
optimal healing and patency of a vascular graft is match- hearing loss from otosclerosis (a hereditary defect which
ing of its mechanical properties to those of the anasto- involves a change in the bones of the middle ear). Otology
mosed natural tissues. Lack of compliance matching with prostheses made of polymers namely PMMA, PTFE, PE,
the host artery is detrimental to the acceptance of syn- and SR, and CF/PTFE composites have been tried to
thetic vascular grafts, when used in the reconstruction replace defective ossicles (three tiny bones of middle ear,
of arteries. A compliance mismatch results in a malleus, incus, and stapes) (it may be noted that the clini-
mechanical incongruity, and in a blood ¯ow of high cally established prostheses are made from titanium, gold,
shear stress and turbulence, with local stagnation. These stainless steel, hydroxyapatite, alumina und glasscera-
factors may lead to local thrombosis, and may damage mics). Migration of prostheses is the main problem repor-
the arterial wall. Hence, there is a greater need to match ted and it is essential to apply suitable surgical method.
compliance of both the vascular graft and the attached Researchers [202,230] are also developing PE/PU ¯exible
blood vessel. The conventional vascular prostheses are composite materials as tympanic membrane replace-
predominantly rigid structures, lacking anisotropy and ments. Tympanic membrane transmits sound vibrations
non-linear compliance. Gershon et al. [72,73] and Klein to the inner ear through three auditory ossicles.
et al. [125] developed composite grafts comprised of
polyurethane (Lycra trade name) ®bers in a matrix of
polyurethane (Pellathane trade name) and PELA (block 4. Other biomedical applications
copolymer of lactic acid and polyethylene glycol) mix-
ture. The non-linear stress strain behavior and com- 4.1. Prosthetic limbs
pliance of the composite graft are varied by controlling
the ®ber orientation [197]. The composite graft is ani- Initial arti®cial legs are designed primarily to restore
sotropic, and isocompliant with the natural artery. The walking of the amputees. They were made of wood or
matrix material is designed to resorb in in vivo condi- metallic materials. These materials are limited by their
tion. At the time of implantation the impervious graft weight, and poor durability due to corrosion and moist-
prevents any loss of blood. The resorption of matrix ure induced swelling. As a result the user is often restric-
material during healing process will result in pores. The ted to slow and non-strenuous activities. Strenuous
ingrowth of granulation tissue into pores provides a activities, such as playing ball games and running are
stable anchorage for the development of a viable cel- not possible due to the weight of these devices and their
lular lining. The optimum pore size of the outer and poor elastic response during stance. The lightweight,
inner layers of the graft can be designed to meet the corrosion resistance, fatigue resistance, aesthetics, and
exact needs of ingrowth and anchorage. The composite ease of fabrication of polymer composite materials made
grafts are in the clinical research phase and yet to be them ideal choice for modern limbs systems [204]. Several
used clinically. designs of arti®cial limbs with di€erent commercial names
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1209

are available. Thermoset polymer composites reinforced fabrication also reduces the number of patient visits and
with glass, carbon, or Kevlar ®bers are widely used in improves service to the physically disabled people. The
these systems [52]. A typical arti®cial leg system consists direct sockets appeared in the market in recent years,
of three parts namely socket, shaft, and foot (Fig. 6). are made using a combination of knitted or braided
The most highly customized and important part of the carbon or glass ®ber fabrics and water-curable (water-
prosthesis is the socket, which has to be fabricated activated) resins. As expected the braided fabric rein-
individually to the satisfaction of each amputee. Sockets forced sockets are sti€ and strong, whereas the knitted
can be divided into two categories, namely, direct and fabric reinforced sockets are ¯exible and more con-
indirect sockets. A widely used indirect socket is fabri- formable to the patient's stump [102].
cated by wrapping several layers of knitted or woven The shaft or stem is often made of ®lament wound or
fabrics [224] on a customized plastic mold, vacuuming laminated woven/braided fabric carbon ®ber reinforced
the fabrics enclosed in a plastic bag, and impregnating epoxy composites. It provides structural support and
the vacuumed fabrics with polyester resin. The socket is force trasmittance to mimic the skeleton [69]. In some
formed after the resin is cured under the vacuum pres- designs, the foot unit consists of heel and forefoot com-
suring condition. It is reported that the performance of ponents, which are made of laminated CF/epoxy compo-
an indirect socket depends mainly on the quality of the sites and are designed to serve as ¯at spring-like leaves so
mold. Moreover, the fabrication process is time-con- that the foot provides strong cushioning and energy stor-
suming and greatly in¯uenced by the prosthesist skills. ing e€ect [232]. They are designed to store energy during
A direct socket, as the name suggests, is fabricated stance and release energy as body weight progresses for-
directly on the stump of a patient, without using any ward, thus helping to propel the body and to achieve
kind of mold. Compared with indirect sockets, the ben- smooth ambulation. This gives the user a higher degree of
e®t of direct socket fabrication is that it can reduce the mobility with a more natural feel compared with conven-
amount of skill dependency in the creation of a socket tional wood prosthetic feet [78]. Delamination of plies
and lead to reduction of ®tting errors between the is a major concern and need to be addressed for longer
stump and the socket. In addition, the direct socket life of the foot. Polymer composites are also used in knee
braces.

4.2. Medical instrumentation

High technology machines such as CT and MRI


scanners are gaining wider usage for medical diagnostic
purposes. These machines have larger bodies ®tted with
moving tables for the patients. The moving table needs
to strong and sti€, at the same time lightweight, radi-
olucent and non-magnetic to obtain clear sliced images
of the patient. As expected the moving tables are made
of carbon ®ber reinforced polymer composites [129].
These materials are also used in making surgical clamps,
head rests frames, X-ray ®lm cassettes and CT scan
couches.

5. Critical issues

From the previous sections it is apparent that a wide


variety of polymer composite materials were investigated
or developed for possible biomedical applications. For
the purpose of clarity, the various man-made polymer
composite materials are classi®ed into several sub-
groups as shown in Fig. 7. A composite material made
of avital (non-living) matrix and reinforcement phases,
is called `avital/avital composite'. Alternatively, a com-
posite material comprising of vital (living) and avital (non-
living) materials is called `vital/avital composite'. These
Fig. 6. Photograph of a typical prosthetic leg showing socket, shaft, composites are further discussed in Section 5.2. The avital/
and foot. avital composites are analogous to polymer composites
1210 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

Fig. 7. Classi®cation of man-made polymer composite biomaterials.

known to engineers. The avital/avital composites are and arti®cial skin. The need and usefulness of non-
further divided into non-resorbable, partially resorbable resorbable and resorbable composites are highlighted in
and fully resorbable composite biomaterials. The non- the previous sections. The speci®c issues common to
resorbable composites are designed not to degrade in various avital/avital composite materials are discussed
the in vivo (inside the body) environment. They are further in the following sections.
particularly promising for long-term implants such as
total joint replacements, bone cement, spine rods, fusion 5.1. Avital/avital composites
cages, discs, plates, dental posts, and hernia patches.
They are also proposed for short term applications such 5.1.1. E€ects of in-vivo environment, and new failure
as bone plates, rods, screws, ligaments, and catheters. On criteria
the other hand the resorbable composites are intended to As mentioned earlier, the non-resorbable composites
loose their mechanical integrity in in vivo conditions. are intended not to degrade in in vivo conditions. How-
They are particularly promising as short-term or tran- ever, some researchers pointed out that the in vivo condi-
sient implants namely bone plates, screws, pins, rods, tions might introduce profound changes in the physical,
ligaments, tendons, bone replacement, vascular grafts, chemical, and mechanical properties of composite bio-
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1211

materials. Hence, knowledge of the e€ects of the in vivo ®ber pull-out and ¯exural tests to investigate the e€ect
environment on the composite properties is very impor- of silane treatment on the interfacial bond strength of
tant [249]. McKenna et al [159] investigated the stability GF/PC and GF/PP composites. They reported best
of GF/epoxy and CF/PS composites in simulated in vivo results for composites with untreated ®bers compared to
conditions (i.e. in vitro testing in saline solution). They the composites reinforced with silane treated ®bers. The
reported only a small change in sti€ness and strength of silane treatment reportedly led to the problems of hydro-
GF/epoxy composite whereas a signi®cant reduction in lytic instability under extreme conditions of stress and
the properties of CF/PS composite material. This dif- moisture. The best results in the case of untreated ®ber
ference was attributed to the variations in the ®ber/ composite were attributed to the annealing treatment
matrix interfacial bond strengths of both the composite given to the composite, which resulted in a strongly bon-
materials. Latour and Black [140] investigated the e€ect ded, highly water resistant interface through nucleation of
of simulated in vivo environments such as saline and highly ordered polycarbonate at the ®ber/matrix interface.
exudate (it is acellular biologic ¯uid similar to inter- The above studies clearly indicate that the quality of ®ber
stitial ¯uid) on the ®ber/matrix interfacial bond strength and matrix interface is of principal importance in deter-
of CF/PC, CF/PS, KF/PC, and KF/PS composites, which mining the response of polymer composite materials to in
are candidates for orthopedic applications. They adopted vivo environments. The e€ect of in vivo exposure upon
a single ®ber pull-out test to measure the interfacial bond the ®ber/matrix interface, and the subsequent e€ect
strength. The bond strength of each material combination upon the implant's mechanical properties must be con-
was signi®cantly degraded by exposure to either saline or sidered in the design and selection of polymer composites
exudate. The water and/or salt ions were found to be to ensure satisfactory long term durability/performance
responsible for the deterioration of interfacial bond in vivo. The review of present knowledge on the polymer
strength. Later, Latour and Black [141] also conducted composite biomaterials leads to the recognition that
fatigue studies on the CF/PS and KF/PS composites in there is lack of accumulated experience and knowledge
simulated in vivo environments. They found that the about the long-term stability of these materials in physi-
®ber/matrix interface failed at approximately 105 load ological environment. The studies reported in the litera-
cycles at a maximum applied load level of only 15% of ture only illustrate the e€ect of di€usion of environment
its ultimate dry bond strength without indication of an on the mechanical properties of composite materials. It is
endurance limit being reached. They expressed serious to be noted that the in vivo conditions depending on the
concern about the durability of polymer composites in purpose and the site of implantation include di€erent tis-
load bearing orthopedic applications. In another study, sue ¯uids and dynamic mechanical loads. Hence, knowl-
Brown et al. [34] investigated the e€ect of exposure to sal- edge of combined e€ects of di€usion of environment and
ine solution (0.9% NaCl) on the ¯exural and fracture mechanical stresses (static and dynamic) on the long-term
toughness properties of short carbon ®ber reinforced PS, behavior of composite materials is important. More
PBT, and PEEK composites. CF/PS and CF/PBT com- importantly, for the same implant, the results obtained
posites showed signi®cant degradation of mechanical from one composite material system cannot be extra-
properties following exposure to saline solution. However, polated to another system, even though there may be one
no such reduction in mechanical properties was reported common phase in both the systems. Similarly, one com-
for the CF/PEEK composites. This was attributed to good posite system evaluated and found suitable for one bio-
bond between the carbon ®bers and PEEK matrix [254]. medical application cannot be used in another application
Suwanprateeb et al. [223] conducted in vitro tests on without systematic studies and design. This also calls for
HDPE and HA/HDPE in a simulated body environ- thorough experimental evaluation of durability of dif-
ment, Ringer's solution. They reported that unreinforced ferent polymer composite biomaterials in in vivo condi-
HDPE properties were una€ected by the solution, tions. This knowledge is very important in making
whereas the composite creep resistance and sti€ness proper judgements for their clinical use.
decreased. The e€ect increased with increasing volume The readers are reminded that the above discussion is
fraction of HA and time of immersion. The decrease in limited to non-resorbable composites, which are designed
properties was attributed to penetration of solution into to remain stable in vivo environment. In contrast, the
the material through the interface. Various methods resorbable composites are designed to be in¯uenced by
have been developed to improve the interface of HA the in vivo environment. The components of resorbable
with a polymer matrix. Silane coupling agents [58], zin- composites are selected such that the water absorption
conyl salts, polyacids and isocyanates [146] were used to (hydration) and/or enzymatic degradation leads to con-
form direct chemical linkage between the HA particles trolled degradation of mechanical integrity of the compo-
and the polymer matrix. By optimizing the surface treat- site material. This involves simple intentional delamination
ment, a further improvement of in vivo behavior of to loss of total mass of the composite material. Current
composites can be expected. However, Jancar and Dibe- constitutive models and failure criteria used for engi-
nedetto [115] found opposite results. They used single neering polymer composites may not be applicable to
1212 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

the resorbable composites, as they are developed 5.1.3. Wear debris, and leached or resorbed products
assuming no change in the material geometry and total Wear of materials is particularly important for articu-
mass. Hence, there is a need to develop new constitutive lating joint applications. Research reports published on
models as well as failure criteria to understand or simulate the wear characteristics of polymer composites from the
the in vivo behavior of resorbable composite materials. viewpoint of biomedical applications are very few. The
With regard to resorbable composite materials, the goal e€ect of reinforcement on the wear characteristics of
that remains to be achieved is how to tailor the composite polymers is a controversial subject, and further sys-
material such that it would loose its mechanical properties tematic investigations are necessary to clearly under-
at approximately the same rate as required by the inten- stand the in vivo wear mechanisms of polymer
ded application (related to tissue healing). Furthermore, composite materials. Also the long-term systemic e€ects
after loss of the mechanical functionality the implant of polymer composite wear debris are still unclear, and
should disappear as fast as possible. Otherwise, the long hence, accumulation of clinical data and its careful
residual time of the implant may lead to formation of a analysis is needed [171].
thick ®brous capsule, which subsequently results in In the case of thermoset polymer composites, there
undesirable calci®cation. An important aspect of bior- are concerns about possible harmful e€ects of residual
esorbable biomaterials is that not only the original monomers, catalysts, and additives that may leach into
material but also the degradation products have to be the tissues. Further e€orts are necessary to develop
non-toxic and removed from the body without side newer thermoset polymers, which are biocompatible.
e€ects. Moreover they need to have adequate initial In the case of resorbable polymers, concerns are
strength and sti€ness at the time of implantation. Cur- expressed over the long-term e€ects of resorbed pro-
rently, this is an area of intensive research. ducts. E€orts are needed to design these materials such
that they are removed from the body without side e€ects.
5.1.2. Improved test methods and new design criteria
Among biomedical researchers, there has been a con- 5.1.4. Improved manufacturing methods, and e€ect of
siderable variability in the method of testing or evalu- sterilization
ating implants. It is very important to standardize the The success of polymer composite biomaterials also
test methodology so as to obtain a meaningful compar- relies greatly on the quality of the implant, which is
ison of various results and also to reproduce results with determined by the reproducibility of the fabrication pro-
con®dence. The problem has been compounded with the cess, sterilization treatment, material storage and hand-
introduction of polymer composite biomaterials, which ling. Many of the polymer composite biomaterials
are anisotropic and inhomogenous. Testing methods that investigated so far were produced in biomedical research
have been used to evaluate implants made of homogenous laboratories with limited success. This is because of the
isotropic materials may not work for testing composite trial-and-error approach followed in making the com-
material implants. This aspect has been illustrated by posites without proper understanding and implementa-
Heiner et al. [89] with regards to the testing of metallic and tion of ®ner aspects of polymer composite fabrication
polymer composite femoral stems. Further improvements processes. More over, the composite fabrication meth-
and standardization of evaluation methods could con- ods used for engineering applications have been used
tribute to the design of better implants. directly for producing implants. It is important to
A major ¯aw in the majority of the literature dealing acknowledge that the requirements for both the appli-
with implants made of polymer composite biomaterials is cations are di€erent, and the composite fabrication
the lack of proper understanding of composite behavior methods need to be tailored to suit the biomedical
and theories. Many researchers used directly the implant applications. For example, for a hip joint replacement
geometry/design originally meant for isotropic materials application, the composite material surface should be
in producing the polymer composite implants. As the completely covered with a continuous matrix layer in
composite materials are distinctly di€erent from the order to prevent a potential release of ®ber particle
homogenous materials in terms of anisotropy, fracture debris during implantation. More over the fabrication
behavior, and environmental sensitivity, the polymer method need to be optimized such that it enables
composite implants must be designed using criteria sepa- desired local and global arrangement of reinforcement
rate from those, which have been used for isotropic mate- phase so as to make the composite implant structurally
rial-based implants. This may even lead to design of compatible with the host tissues. Review of existing lit-
superior performance implants. Innovations such as spa- erature suggests that the various ¯exibilities of compo-
tially varying ®ber volume fraction and/or ®ber orientation sites in terms of material combinations, ®ber/matrix
are leading to new types of functionally graded composite interface control, ®ber volume fractions, and ®ber and
materials. New design criteria need to be developed to matrix distributions are yet to be fully exploited in fab-
harness the potential of this new class of materials and ricating functionally superior implants. Thus far, poly-
to design implants with improved performance. mer composite biomaterials are mainly reinforced with
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1213

particulates, short ®bers and unidirectional ®bers, and 5.2. Vital/avital composites
very few works reported on woven fabric composites. The
many advantages o€ered by textile composite materials Current trend in biomaterials development is to grow
have not been exploited in the biomedical ®eld. E€orts tissues in the laboratory using cells (patient's cells, auto
should be made to harness the potential of textile compo- or xenologous cells, human stem cells or genetically
site materials in designing implants with improved per- engineered cells) of the target tissue (i.e. tissue to
formance. It is also important to consider the cost of replaced or augmented) and porous sca€olds. The com-
composite implant. E€orts must be made to develop bination of polymers (avital or non-living) in the form of
suitable manufacturing methods for composite implants foams or fabrics (woven, braids, knits, and non-wovens)
so as to compete with the current commercial implants. and cells (vital or living) results in special type of com-
Like any other material, polymer composite bioma- posite materials, namely vital/avital composites [57]. If
terials are also sterilized prior to implantation. It is known the patient's own cells can be used, the vital/avital
that the polymer properties are sensitive to the sterilization composites are readily biocompatible and well accepted
procedure used [181]. For example, the gamma steriliza- by the host tissues. Many consider the vital/avital com-
tion reportedly causes long-term embrittlement of posites are ideal for implant applications. The vital/avital
UHMWPE (used in hip joint cups) due to radiation- composites are in their infant stage of development,
induced oxidation. Hence, some e€ects of sterilization however, it is an area of intensive research worldwide
on the mechanical propitioes of polymer composites can and called by di€erent names including `tissue engineer-
also be expected. McKenna et al. [160] investigated the ing' and `cellular engineering'. Researchers are develop-
e€ect of autoclave sterilization on a number of candi- ing vital/avital composites for a number of applications
date composite materials. They reported that CF/PP including vascular grafts [162], tendon/ligament pros-
composites did not undergo signi®cant degradation theses [13,18,21,275], arti®cial skin [137], dural substitutes
even at long autoclave times. The CF/PS composites [188], hernia patches, arti®cial bladder wall, and regener-
degraded at even the shortest autoclave times. This ated cartilage. A wide variety of non-resorbable polymers
study highlights the need for evaluating the degradation such as PET, PU, and PTFE, and resorbable polymers
resistance of composite materials under sterilization such as PGA, PLA, and their blends are used as porous
conditions. A suitable sterilization procedure for com- sca€olds. In order to introduce time dependent poros-
posite of interest needs to be established through careful ity, some researchers [68,270,271] used bicomponent
experiments. sca€olds containing both resorbable and non-resorb-
able polymers. To facilitate the attachment of cells to
5.1.5. Surface coatings the avital sca€olds, they have been coated with di€erent
As mentioned earlier, the success of an implant also systems including pyrolitic carbon [3,212], collagen,
depends on its surface chemistry, which determines the albumin, gelatin, and antibiotic drug-releasing gels.
interactions at the implant material±tissue interface. To It may be noted that the sca€old surfaces are functio-
elicit desirable material±host tissue interactions, the nalized for a variety of other reasons. Di€erent kinds
polymer composite implants may need to be coated with of cells are seeded onto the porous sca€olds depending
suitable coatings. Another important reason for a suitable on the intended application (target tissue) of the
surface coating is the wear of the implant surface being in composite material. The cell attachment to the avital
contact with the host tissues. For example, the hardness of sca€olds, and the di€erentiation and maturation of the
bone leads to very heavy abrasion by fretting or direct ingrown or in situ newly formed tissue depend on a
wearing as soon as the interfacial strains between the number of variables including pore size and geometry,
implant and hard tissue occur. Thus, there is a need for porosity, pore distribution, nature (two dimensional
developing suitable coating methods for polymer com- or three dimensional), inter-connectivity of pores,
posite implants. For example, Ha [81] and Ha et al. sca€old thickness, surface topography and biochemical
[80,82] developed a method of coating CF/PEEK com- functionalization, types of cells, external stimuli
posite hip stem surface with bioactive coating. They ®rst (mechanical, electrical or chemical), etc. Speci®c
vacuum plasma sprayed the composite surface with tita- details are outside the scope of the present paper. Inter-
nium. Subsequently, the surface is treated with NaOH ested readers may consult the references cited appro-
and immersed in simulated body ¯uid (SBF), containing priately [83,97,98,120,136,138,139,142,144,154,172±174,
ions in concentrations similar to those of human blood 182,199,262,270,271]. The majority of the information
plasma. Formation of biocompatible and bioactive cal- reported in the literature on the vital/avital composites
cium phosphate layer similar to hydroxyapatite on the is chemistry, biochemistry, and biology related. Little is
composite surface was reported. To date very limited known about the mechanical characteristics of this new
knowledge is available with regards to surface coating class of polymer composite materials. The vital/avital
of polymer composite implants, and this warrants fur- composite materials require relooking into the traditional
ther research and development. composite principles and theories originally developed
1214 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

for mostly linear and small deformation composite mate- Aneurysm Abnormal dilatation of bulging
rials. Further work illustrating the principles of deforma- of a segment of a blood vessel
tion behavior of these materials would be very useful to Ankylosis Fixation of a joint; in dentistry,
innovatively design new implants, and also would be use- the rigid ®xation of the tooth to
ful to understand the behavior of natural tissues itself. the aveolar bone and ossi®cation
Ultimately, this knowledge may give insights to unravel of the periodontal
the mysteries of many natural tissues. membrane
Anterior Direction referring to the front
side of the body
6. Conclusions Arthritis In¯ammation of joints
Arthrodesis Fusion or ®xation of a joint
With increased understanding of function and inter- Arthroplasty Surgical repair of a joint
action of implants with the human body, it is clear now Articular The cartilage at the ends of bones
that for greater success, the implants should be surface cartilage in joints which serve as the
compatible as well as structurally compatible with the articulating, bearing
host tissues. In this respect, the polymer composite bio- surface.
materials are particularly attractive because of their tai- Arti®cial organ A medical device that replaces, in
lorable manufacturing processes, and properties part or in whole, the function of
comparable to those of the host tissues. Innovations in one of the organs of the body.
the composite material design and fabrication processes Atrophy Wasting away of tissues or organs
are raising the possibility of realizing implants with Autograft A transplanted tissue or organ
improved performance. However, for successful appli- transferred from one part of a
cation, surgeons must be convinced with the long term body to another part of the same
durability and reliability of polymer composite bioma- body
terials. Monolithic materials have long been used and Biocompatibility Acceptance of an implant by
there is considerable experimental and clinical data surrounding tissues and by the
supporting their continued usage. Such data with body as a whole. The implant
respect to polymer composite biomaterials is relatively should be compatible with tissues
small. This requires further research e€orts to elucidate in terms of mechanical, chemical,
the long-term durability of composite biomaterials in surface, and pharmacological
the human body conditions. properties. Simply it is the ability
of the implant material to
perform with an appropriate
host response in a speci®c
application.
Bioglass Surface-active glass compositions
that have been shown to bond
to tissue
Biomaterial The term usually applied to living
Appendix A or processed tissues or to
materials used to reproduce the
function of living tissue in
Apical Near the apex or extremity of a conjunction with them. Simply
conical structure, such as the tip it is a material intended to
of the root of a tooth interact with biological
Acetabulum The socket potion of the hip joint systems.
Allograft Transplanted tissue or organ Bone cement A biomaterial used to secure a
between unrelated individuals of ®rm ®xation of joint prostheses,
the same species. Also called such as hip and knee joints. It is
`homograft' primarily made of polymethyl
Alveolar bone The bone structure that supports methacrylate powder and monomer
and surrounds the roots of teeth methyl methacrylate liquid
Amalgam An alloy of two or more metals, Callus The hard substance that is formed
one of which is mercury around a bone fracture during
Anastomosis Interconnection between two healing. It is usually replaced
blood vessels with compact bone.
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1215

Cancellous The reticular or spongy tissue of Endosseous In the bone, referring to dental
bone bone where spicules or trabeculae implants ®xed to the jaw bone
form the interconnecting Endosteal Related to the membrane lining
latticework that is surrounded by the inside of the bone cavities
connective tissue or bone Extracorporeal Outside the body
morrow Femur The thigh bone, the bone of the
Catheter An instrument (tube) for gaining upper leg
access to and draining or sampling Fixation devices Implants used during bone-
¯uids in the body fracture repair to immobilize the
Celestin tube A nylon reinforced latex tube used fracture
to bypass esophgeal tumors Fracture plate Plate used to ®x broken bones by
Cochlear A type of surgically implanted open (surgical) reduction. It is
implant hearing aid used to treat ®xed to the bone by using screws.
sesorineural hearing loss Gingiva The gum tissue; the dense ®brous
Collagen The supporting protein from tissue overlying the alveolar bone
which the ®bers of connective in the mouth and surrounding the
tissues are formed necks of teeth
Compression Bone plate designed to give Graft A transplant
plate compression on the fracture site Ground The amorphous polysaccharide
of a broken bone for fast substance material in which cells and ®bers
healing. are embedded
Condylar Arti®cial knee joints Hard tissue The general term for calci®ed
prostheses structures in the body, such as bone
Congenital A physical defect existing since birth Heparin A substance (mucopolysaccharide
Cortical bone The compact hard bone with acid) found in various body
osteons tissues; that prevents the clotting
Crown The part of tooth that is exposed of blood
above the gum line or covered Herniated disk A herniated disk is the rupture of
with enamel. Largely made of the central portion, or nucleus, of
hydroxyapatite mineral. the disk through the disk wall and
CT Computed tomography or into spinal canal. It is also called
computed axial tomography slipped disk.
(CAT), an X-ray technique for Heterograft A graft from one species to
producing cross-sectional image of another. Also called xenograft
the body. Hyaline cartilage Cartilage with a frosted glassy
Dacron Polyethylene terephthalate appearance
polyester that is made into ®bers, Hydrogel Highly hydrated (over 30% by
a product of Dupont Co, USA. If weight) polymer gel. Acrylamide
the same polymer is made into a and poly-HEMA (hydroxyethy-
®lm, it is called Mylar. methacrylate) are two common
Dental caries Tooth decay caused by acid- hydrogels.
forming micro-organisms Hydroxyapatite Mineral component of bone and
Dental Another name for dental ®llings (HA) teeth. It is a type of calcium
restoration phosphate, with composition
Dentine The main substance of the tooth, Ca10(PO4)6(OH)2.
with properties and composition Ilizarov A technique used most often in
similar to bone. technique reconstructive settings to
Dermatitis In¯ammation of skin lengthen limbs, transport bone
Dura mater The dense, tough connective tissue segments, and correct angular
over the surface of the brain deformities
Elastin The elastic ®brous mucoprotein in Implant Any medical device made from one
connective tissue or more materials that is
Enamel A hard, white substance that intentionally placed within the
covers the dentine of the crown body, either totally or partially
of a tooth; enamel is the hardest buried beneath an epithelial
substance in the body surface.
1216 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

Intervertebral A ¯at, circular platelike structure Ossicles The small bones of the middle ear
disc of cartilage that serves as a which transmit sound from ear
cushion, or shock absorber, drum to the body
between the vertebrae Osteoarthritis A degenerative joint disease,
Intima Inner lining of a blood vessel characterized by softening of the
Intramedullary An orthopedic rod or nail inserted articular ends of bones and
rod or nail into the intramedullary marrow thickening of joints, sometimes
cavity of the bone to promote resulting in partial ankylosis
healing of long bone Osteopenia Loss of bone mass due to failure
fractures of osteoid synthesis
Intraosseous An implant inserted into the bone Osteoporosis The abnormal reduction of the
implant density and increase in porosity
In vivo condition Inside the living body of bone due to demineralization,
In vitro condition Simulated in vivo condition in the commonly seen in the
laboratory elderly
Kirschener wire Metal surgical wires Osteotomy Cutting of bone to correct a
Kyphosis Abnormally increased convexity deformity
in the curvature of the lumbar Percutaneous Transcutaneous, of having to do
spine with passing through the
Ligament A sheet or band of ®brous epidermis or skin
connective tissue that join bone Periodontal Periodontium; the connective
to bone, o€ering support to the ligament tissue (ligament) joining the tooth
joint to the alveolar bone
Long bones Bones that are longer than they Polysaccharides Major constitutents of the ground
are wide and with distinctive substance; carbohydrates
shaped ends, such as femur containing saccharide groups
Lordosis Abnormally increased concavity Posterior Direction referring to the back
in the curvature of the lumbar side of the body
spine Proplast A composite material made of
LTI carbon Low-temperature istropic carbon ®brous PTFE and carbon. It is
Lumen The space within a tubular usually porous and has low
structure modulus and low strength.
Mandibular bone Lower jaw of the mouth Prosthesis A device that replaces a limb,
Maxillary bone Upper jaw of the mouth organ or tissues of the body
Medullary cavity The marrow cavity inside the long Proximal Nearest the trunk or point of
bones origin; opposed to distal
Metastasis Transfer of disease producing Pyrolitic carbon Isotropic carbon coated onto a
cancer cells or bacteria from an substrate in a ¯uidized bed
original site of disease to another Resorption Dissolution or removal of a
part of the body with substance
development of a similar lesion in Rheumatoid Chronic and progressive
the new location arthritis in¯ammation of the connective
Myocardium The muscular tissue of the heart tissue of joints, leading to
Necrosis Death of tissues deformation and disability
NMR Nuclear magnetic resonance Scoliosis An abnormal lateral (sideward)
Nonunion A bone fracture that does not join curvature of a portion of the
Occlusion Becoming close together; in spine
dentistry, bringing the teeth Silastic Medical grade silicone rubber,
together as during biting and Dow Corning Corporation
chewing Silica The ceramic SiO2
Orthopedics The medical ®eld concerned with Spondylosis Any of various degenerative
the skeletal system diseases of the spine
Orthotics The science and engineering of Spondylolisthesis Forward bending of the body at
making and ®tting orthopedic one of the lower vertebrae
appliances used externally to the Stapes One of the ossicles of the middle
body. ear
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1217

Stenosis A narrowing or constriction of the References


diameter of a bodily passage or
[1] Adams D, Williams DF, Hill J. Carbon ®ber-reinforced carbon
ori®ce.
as a potential implant material. J Biomedical Materials
Stress-shield Prolonged reduction of stress on a Research 1978;12:35±42.
e€ect bone may result in porotic bone [3] Aebischer P, Goddard MB. Tissue reaction to fabrics coated
(osteoporosis), which may weaken with turbostratic carbon: subcutaneous versus vascular
it. This process can be reversed if implants. Biomaterials 1988;9(1):80±5.
the natural state of stress can be [4] Akay M, Aslan N. Numerical and experimental stress analysis
of a polymeric composite hip joint prostheses. J Biomedical
restored to its original Materials Research 1996;31:167±82.
state [5] Ali MS, Hastings GW, Rae T, Rushton N, Ross ERS, Wynn-
Subcutaneous Beneath the skin Jones CH. Carbon ®ber composite bone plates. J Bone and
Subperiosteal Underneath the periosteum Joint Surgery 1990;72-B(4):586±91.
Synovial ¯uid The clear viscous ¯uid that [6] Ambrosio L, Carotenuto, Nicolais L. Composite materials. In:
Black J, Hastings G, editors. Handbook of Biomaterial
lubricates the surfaces of joints Properties, Editors. London, UK: Chapman and Hall, 1998. p.
and tendons, secreted by the 214±69.
synovial membrane [7] Ambrosio L, De Santis R, Iannace S, Netti PA, Nicolais L.
Tendon A band or cord of ®brous tissue Viscoelastic behavior of composite ligament prostheses. J Bio-
connecting muscle to medical Materials Research 1998;42(1):6±12.
[8] Ambrosio L, Netti PA, Iannace S, Huang SJ, Nicolais L. Com-
bone posite hydrogels for intervertebral disc prostheses. J Materials
THR Total hip replacement Science: Materials in Medicine 1996;7:251±4.
Thromboembolism An obstruction in the vascular [9] Amstutz HC, et al. Mechanism and clinical signi®cance of wear
(blood circulating) system caused debris-induced osteolysis. Clinical Orthopedics and Related
by a dislodged thrombus Research 1992;276:7±18.
[10] Andersson GBJ. Intervertebral disk. In: Wright V, Radin EL,
Thrombosis Formation of a thrombus, blood editors. Mechanics of human joints: physiology, pathophysiol-
clot ogy, and treatment. New York: Marcel Dekker Inc, 1993. p.
Thromus A ®brinous blood clot attached at 293±311.
the site of thromsosis [11] Andreopoulos AG, Evangelatou M, Tarantili PA. Properties of
maxillofacial silicone elastomers reinforced with silica powder. J
TKR Total knee replacement
Biomaterials Applications 1998;13:66±73.
Transplantation Transfer of a tissue or organ from [13] Arnoczky SP, Torzilli PA, et al. Biologic ®xation of ligament
one body to another, or from one prostheses and augmentations. An evaluation of bone in growth
location in a body to in the dog. Am J Sports Med 1988;16(2):106±12.
another [14] Baidya KP, Ramakrishna, S, Rahman M, Ritchie A.
Trachea A cylinder-shaped tube lined with Quantitative radiographic analysis of ®ber reinforced polymer
composites. J Biomaterials Applications 2001;15(3):279±89.
rings of cartilage that is 115 mm [15] Baidya KP, Ramakrishna S, Rahman M, Ritchie A. Advanced
long, from the larynx to the textile composite ring for ilizarov external ®xator system. J
bronchial tubes; the windpipe Engineering in Medicine, IME UK 2001;215(H1):11±23.
Ureter The tube that conducts urine from [16] Baidya KP, Ramakrishna S, Ritchie A, Rahman M. Design and
the kidney to the bladder evaluation of composite external fracture ®xation device. Proc.
ICCM-12, Paris, France, p. 496.
Urethra The canal leading from the bladder [17] Bao QB, McCullen GM, Higham PA, Dumbleton JH, Yuan
to the outside for discharging urine HA. The arti®cial disc: theory, design and materials. Biomater-
Vascular Blood vessels ials 1996;17:1157±67.
Vitallium A Co-Cr alloy, Howmedica Inc. [18] Berry JL, Berg WS, et al. Evaluation of Dacron-covered and
plain bovine xenografts as replacements for the anterior cruciate
Vitreaous carbon A term generally applied to
ligament. Clin Orthop 1988;236:270±8.
isotropic carbon with very small [19] Bjork N, Ekstrand K, Ruyter IE. Implant-®xed, dental bridges
crystallites from carbon/graphite ®ber reinforced poly(methyl methacry-
Wol€'s law The principle relating the internal late). Biomaterials 1986;7:73±5.
structure and architecture of bone [20] Black J. Orthopaedic Biomaterials in Research and Practice.
to external mechanical stimuli. New York: Churchill Livingstone, 1988 (p. 197±210).
[21] Black J. Biological Performance of Materials: Fundamentals of
Remodeling of bone takes place Biocompatibility. New York: Marcel Dekker, 1992.
in response to mechanical [22] Black J, Hastings GW. Handbook of Biomaterials Properties.
stimulation so that the new London, UK: Chapman and Hall, 1998.
structure becomes suitably adapted [23] Blazewicz S, Chlopek J, Litak A, Wajler C, Staszkow E.
to the load. Experimental study of mechanical properties of composite car-
bon screws. Biomaterials 1997;18:437±9.
Xenograft A transplanted tissue or organ [24] Blumenthal NC, et al. A new technique for quantitation of
transferred from an individual metal particulates and metal reaction products in tissues near
of another species implants. J Applied Biomaterials 1994;5:191±3.
1218 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

[25] Bon®eld W. Composites for bone replacement. J Biomedical [46] Ciappetta P, Boriani S, Fava GP. A carbon ®ber reinforced
Engineering 1988;10:522±6. polymer cage for vertebral body replacement: a technical note.
[26] Bon®eld W, Grynpas MD, Tully AE, Bowman J, Abram J. Neurosurgery 1997;41(5):1203±6.
Hydroxyapatite reinforced polyethylene Ð a mechanically [47] Claes LE. Mechanical characterization of biodegradable
compatible implant material for bone replacement. Biomaterials implants. Clinical Materials 1992;10:41±6.
1981;2:185±6. [48] Claes L, Hutter W, Weiss R. Mechanical properties of carbon
[27] Bostman O, Hirvensalo E, Makinen J, Rokkanen P. Foreign- ®ber reinforced polysulfone plates for internal ®xation. In:
body reactions to fracture ®xation implants of biodegradable Christel P, Meunier A, Lee AJC, editors. Biological and Bio-
synthetic polymers. J Bone and Joint Surgery 1990;72-B: mechanical Performance of Biomaterials. Amsterdam, The
592±6. Netherlands: Elsevier Science Publishers, 1997. p. 81±6.
[28] Bradley JS. Cautionary note on the strength of materials for [50] Claes L, Schultheiss M, Wolf S, Wilke HJ, Arand M, Kinzl L. A
internal ®xation devices. Biomaterials 1981;2:124. new radiolucent system for vertebral body replacement: its sta-
[29] Bradley JS, Hastings GW. Carbon ®ber reinforced plastics for bility in comparison to other systems. J Biomedical Materials
orthopedic implants Mechanical properties of biomaterials. Research, Applied Biomaterials 1999;48(1):82±9.
Hastings GW, Williams DF, editors. 379±386 [51] Cogswell FN. Thermoplastic Aromatic Polymer Composites.
[30] Bradley JS, Hastings GW, Johnson-Nurse C. Carbon ®ber UK: Butterworth Heinemann, 1992.
reinforced epoxy as a high strength, low modulus material for [52] Coombes AGA, Greenwood CD, Shorter JJ. Plastic materials
internal ®xation plates. Biomaterials 1980;1:38±40. for external prostheses and orthoses. In: Wise DL, Trantolo DJ,
[31] Bragdon CR, et al. Defects in the cement mantle: a fatal ¯aw in Altobelli DE, Yaszemski MJ, Gresser JD, editors. Human Bio-
cemented femoral stems for THR. San Francisco: Society for materials Applications. (NJ): Humana PressTotowa, 1996. p.
Biomaterials, 1995. 215±55.
[32] Brantigan JW, Ste€ee AD, Geiger JM. A carbon ®ber implant [53] Culleton TP, et al. Fatigue failure in the cement mantle of an
to aid interbody lumbar fusion Mechanical testing. Spine arti®cial hip joint. Clinical Materials 1993;12:95±102.
1991;16(6S):S277±0S282. [54] Darendeliler S, Darendeliler H, Kinoglu T. Analysis of a central
[33] Bridwell KH, DeWald RL, Lubicky JP, Spencer DL, Hammer- maxillary incisor by using a three-dimensional ®nite element
berg KW, Benson DR, Neuwirth MG. The Textbook of method. J Oral Rehabilitation 1992;19:371±83.
Spinal Surgery. Philadelphia, (PA): J.B. Lippincott Company, [55] Dauner M, Caramaro L, Missirlis Y, Panagiotopoulos E.
1991. Resorbable continuous-®ber reinforced polymers for osteo-
[34] Brown SA, Hastings RS, Mason JJ, Moet A. Characterization synthesis. J Materials Science: Materials in Medicine 1998;
of short-®ber reinforced thermoplastics for fracture ®xation 9:173±9.
devices. Biomaterials 1990;11:541±7. [56] Davy KWM, Parker S, Braden M, Ward IM, Ladizesky H.
[35] Bruckmann H, Huttinger KJ. Carbon, a promising material in Reinforcement of polymers of 2,2 bis-4(2-hydroxy-3-methacry-
endoprosthetics. Part 1: the carbon materials and their mechan- loyloxy propoxy) phenyl propane by ultra-high modulus poly-
ical properties. Biomaterials 1980;1:67±72. ethylene ®bers. Biomaterials 1998;13(1):17±19.
[36] Bruckmann H, Keuscher G, Huttinger KJ. Carbon, a promising [57] de Haan J, Peijs T, Mayer J, Wintermantel E. Modeling of the
material in endoprosthetics. Part 2: tribological properties. Bio- deformation behavior of ¯exible knitted composites. In: Proc.
materials 1980;1:73±81. ECCM-7, London: Woodhead Publishing Limited, 1996, p.
[37] Buchert PK, et al. Excessive Metal Relase due to Loosening and 421±7.
Fretting of Sintered Particles on Porous-Coated Hip Prostheses. [58] Deb S, Wang M, Tanner KE, Bon®eld W. Hydroxyapatite-
J Bone and Joint Surgery 1986;68-A:606±9. polyethylene composites: e€ect of grafting and surface treatment
[38] Cailleteau JG, Rieger MR, Ed Akin J. A comparison of intra- of hydroxyapatite. J Materials Science: Materials in Medicine
canal stresses in a post-restored tooth utilizing the ®nite element 1996;7:191±3.
method. J Endodontics 1992;18(11):540±4. [59] Dehl RE. On the characterization of porosity in PTFE-carbon
[39] Chalian VA, Phillips RW. Materials in maxillofacial prosthtics. composite implant materials by mercury porosimetry. J Biome-
J Biomedical Materials Research Symposium 1974;5(2):349±63. dical Materials Research 1982;16:715±9.
[40] Chang FK, Perez JL, Davidson JA. Sti€ness and strength tai- [60] Deluca PA, Lindsey RW, Ruwe PA. Refracture of bones of the
loring of a hip prostheses made of advanced composite materi- forearm after the removal of compression plates. J Bone and
als. J Biomedical Materials Research 1990;24:873±99. Joint Surgery 1988;70-A(9):1372±6.
[41] Chao EYS, Aro HT. Biomechanics of fracture ®xation. In: Mow [61] Deng M, Shalaby SW. Properties of self-reinforced ultra-high-
VC, Hayes WC, editors. Basic Orthopaedic Biomechanics. Phi- molecular weight polyethylene composites. Biomaterials
ladelphia: Lippincott-Raven, 1997. p. 317±50. 1997;18:645±55.
[42] Choueka J, Charvet JL, Alexander H, Oh YH, Joseph G, Blu- [62] Durselen L, Claes L, Ignatius A, Rubenacker S. Comparative
menthal NC, LaCourse WC. E€ect of annealing temperature on animal study of three ligament prostheses for the replacement of
the degradation of reinforcing ®bers for absorbable implants. J the anterior cruciate and medial collateral ligament. Biomater-
Biomedical Materials Research 1995;29:1309±15. ials 1996;17:977±82.
[43] Christel PS, Leray JL, Sedel L, Morel E. (1980). Mechanical [63] Esslinger JO, Rutkowski EJ. Studies on the skeletal attachment
evaluation and tissue compatibility of materials for composite of experimental hip prostheses in the pygmy goat and the
bone plates. In: Mechanical Properties of Biomaterials. Hastings dog. J Biomedical Materials Research Symposium 1973;4:
GW, Williams DF, editors. p. 367±377. 187±93.
[44] Christel P, Claes L, Brown SA. Carbon reinforced composites in [64] Feith R. Side e€ects of acrylic cement, implanted to bone. Acta
orthopedic surgery. In: Szycher M, editor. High Performance Orthop Scand 1975; (Suppl.):161.
Biomaterials: A Comprehensive Guide to Medical and Pharma- [65] Flahi€ CM, Blackwell AS, Hollis JM, Feldman DS. Analysis of
ceutical Applications. Lancaster, (USA): Technomic, 1991. p. 499± a biodegradable composite for bone healing. J Biomedical
518. Materials Research 1996;32:419±24.
[45] Christel P, Meunier A, Leclercq S, Bouquet Ph, Buttazzoni B. [66] Forster IW, Ralis ZA, Jenkins DH. Filamentous carbon ®ber
Development of carbon-carbon hip prostheses. J Biomedical induction of new tendon: tissue reactions and environmental
Materials Research: Applied Biomaterials 1987;21(A2):191±218. conditions. In: Winter GD, Leray JL, de Groot K, editors.
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1219

Evaluation of Biomaterials. New York: John Wiley & Sons Ltd, [88] Hastings GW. Biomedical applications of CFRPs. In: Fitzer E,
1980. p. 367±71. editor. Carbon Fibre and their Composites, Springer-Verlag,
[67] Freilich MA, Meiers JC, Duncan JP, Jon Goldberg A. Fiber- Berlin, p. 261±271.
Reinforced Composites in Clinical Dentistry. IL: Qintessence [89] Heiner AD, Brown SA, Davy DT. E€ects of mechanical testing
Publishing, 2000. device variables on polymer composite femoral stem strains.
[68] Galletti PM, Aebischer Petal. Experience with fully bioresorb- Biomaterials 1996;17(23):2211±7.
able aortic grafts in the dog. Surgery 1988;103(2):231±41. [90] Hench LL. Bioceramics: from concept to clinic. J American
[69] Ganesh VK, Ramakrishna S. Prothesis: another concept. The Ceramic Society 1991;74:1487±510.
European periodical for technical textile users. Quarterly [91] Hench LL. Inorganic biomaterials Materials Chemistry: An
Magazine 1998;29(3):56±60. Emerging Discipline. In: Interrante LV, Caspar LA, Ellis AB,
[70] Ganesh VK, Ramakrishna S, Leck HJ. Fiber reinforced com- editors. Advances in Chemistry Series 245. Washington, (DC):
posite based functionally gradient materials. J Advanced Com- American Chemical Society, 1995. p. 523±47.
posite Letters 1998;7:111±5. [92] Hench LL, Ethridge EC. Biomaterials An Interfacial Approach.
[71] Gerhart TN, Miller RL, Kleshinski SJ, Hayes WC. In vitro New York: Academic Press, 1982.
characterization and biomechanical optimization of a biode- [93] Henderson JD, Mullarky RH, Ryan DE. Tissue biocompat-
gradable particulate composite bone cement. J Biomedical ibility of kevlar aramid ®bers and polymethylmethacrylate,
Materials Research 1988;22:1071±82. composites in rabbits. J Biomedical Materials Research
[72] Gershon B, Cohn D, Marom G. Utilization of composite lami- 1987;21:59±64.
nate theory in the design of synthetic soft tissues for biomedical [95] Hidaka S, Gustilo RB. Refracture of bones of the forearm after
prostheses. Biomaterials 1990;11:548±52. plate removal. J Bone and Joint Surgery 1984;64-A(8):1241±3.
[73] Gershon B, Cohn D, Marom G. Compliance and ultimate [96] Higashi S, Yamamuro T, Nakamura T, Ikada Y, Hyon SH,
strength of composite arterial prostheses. Biomaterials Jamshidi K. Polymer-hydroxyapatite composites for biodegrad-
1992;13:38±43. able bone ®llers. Biomaterials 1986;7:183±7.
[74] Gese H, et al. Relativbewegungen und stress-shielding bei [97] Hirt SW, Aoki M, Demertzis S, Siclari F, Haverich A, Borst HG.
zementfreien HuÈftendoprothesen Ð eine Analyse mit der Meth- Comparative in vivo study on the healing qualities of four di€erent
ode der Finiten Elemente. In: Die zementlose HuÈftprothese. presealed vascular prostheses. J Vasc Surg 1993;17(3):538±45.
GraÈfel®ng, Demeter Verlag: p. 75±80. [98] Hirt SW, Dosis D, Siclari F, Rohde R, Haverich A. Collagen-
[75] Giardino R, Glannini S, Fini M, Giavaresi G, Martini L, presealed or uncoated aortic bifurcation Dacron prostheses: a 5-
Orienti L, Experimental in vivo model to evaluate resorbable year clinical follow-up study. Thorac Cardiovasc Surg
implants into bone. In: Kim HL, Chung LP editors. Biodegrad- 1991;39(6):365±70.
able Implants in Fracture Fixation. 1993. p. 143±51. [99] Ho MH, Lee SY, Chen HH, Lee MC. Three-dimensional ®nite
[76] Gilbert JL, Ney DS, Lautenschlager EP. Self-reinforced com- element analysis of the e€ects of posts on stress distribution in
posite poly(methyl methacrylate): static and fatigue properties. dentin. The J of Prosthhetic Dentistry 1994;72:367±72.
Biomaterials 1995;16(14):1043±55. [100] Hoag EP, Dwyer TG. A comparative evaluation of three post
[77] Gillett N, Brown SA, Dumbleton JH, Pool RP. The use of short and core techniques. The Journal of Prosthetic Dentistry
carbon ®ber reinforced thermoplastic plates for fracture ®xa- 1982;47:177±81.
tion. Biomaterials 1986;6:113±21. [102] Huang ZM, Ramakrishna S. Development of knitted fabric
[78] Goh JCH, Tan PH, Toh SL, Tay TE. Gait analysis study of an reinforced composite material for prosthetic application.
energy-storing prosthetic foot-a preliminary study. Gait and Advanced Composites Letters 1999;8(6):289±94.
Posture 1994;2(2):95±101. [103] Huiskes R. Some fundamental aspects of human-joint replace-
[79] Goldner JL, Urbaniak JR. The clinical experience with silicone- ment. Acta Orthop Scand 1980; (Suppl.), 185.
Dacron metacarpophalangeal and interphalangeal joint pros- [104] Huiskes R, et al. The relationship between stress shielding and
theses. J Biomedical Materials Research Symposium bone resorption around total hip stems and the e€ects of ¯exible
1973;4:137±63. materials. Clinical Orthopaedics and Related Research 1992;
[80] Ha SW, Eckert KL, Wintermantel E, Gruner H, Guecheva M, 274:124±34.
Vonmont H. NaOH treatment of vacuum-plasma-sprayed tita- [105] Hunt MS. Development of carbon ®ber/polysulfone orthopae-
nium on carbon ®ber-reinforced poly(etheretherketone). J dic implants. Materials and Design 1987;8(2):113±9.
Materials Science: Materials in Medicine 1997;8:881±6. [107] Huttner W, Keuscher G, Nietert M. Carbon ®ber reinforced
[81] Ha S-W. Bioactive calcium phosphate coatings on poly (ether- polysulfone thermoplastic composites. In: Ducheyne P, Van der
etherketone) and its composites, PhD. thesis, ETH, Zurich, Perre G, Aubert AE, editors. Biomaterials and Biomechanics.
Switzerland, 1997. Amsterdam, The Netherlands: Elsevier Science Publishers, 1984.
[82] Ha S-W, Mayer J, Koch B, Wintermantel E. Plasma-sprayed p. 167±72.
hydroxylapatite coating on carbon ®ber reinforced thermo- [108] Iannace S, Sabatini G, Ambrosio L, Nicolais L. Mechanical
plastic composite materials. J Materials Science: Materials in behavior of composite arti®cial tendons and ligaments. Bioma-
Medicine 1994;5:481±4. terials 1995;16(9):675±80.
[83] Hake U, Gabbert H, Iversen S, Jakob H, Schmiedt W, Oelert H. [109] Ignatius A, Unterricker K, Wenger K, Richter M, Claes L. A
Evaluation of the healing of precoated vascular dacron pros- new composite made of polyurethane and glass ceramic in a
theses. Langenbecks Arch Chir 1991;376(6):323±9. loaded implant model: a biomechanical and histological
[84] Halstead A, Jones CW, Rawlings RD. A study of the reaction of analysis. J Materials Science: Materials in Medicine 1997;8:753±
human tissue to proplast. J Biomedical Materials Research 6.
1979;13:121±34. [110] Imai T, Watari F, Yamagata S, Kobayashi M, Nagayama K,
[85] Harris B. The mechanical behavior of composite materials. The Nakamura S. E€ects of water immersion on mechanical
Mechanical Properties of Biological Materials. Cambridge, UK: properties of new esthetic orthodontic wire. Americal Journal of
Cambridge University Press, 1980 (p. 37±74). Orthodontics and Dentofacial Orthopedics 1999:533±538.
[86] Hastings GW. Is there an ideal biomaterial for use as an implant [111] Inoue N, Hirasawa Y, Hirai T, Katayama T. Composite mate-
for fracture ®xation? In Hastings GW editor. Biodegradable rials in bio-medical engineering. Materiaux and Techniques
Implants in Fracture Fixation, 1993. p. 19±34. 1994;4-5:23±6.
1220 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

[112] Issidor F, Brendum K. Intermittent loading of teeth with [131] Krause W, Mathis RS. Fatigue properties of acrylic bone
tapered individually cast or prefabricated parallel-sided posts. cements: Review of the literature. J Biomedical Materials
Int J Prosthodontics 1992;5:257±61. Research: Applied Biomaterials 1988;22(A1):37±53.
[113] Issidor F, Odman P, Brondum K. Intermitttent loading of teeth [132] Krause WR, Park SH, Straup RA. Mechanical properties of
restored using prefabricated carbon ®ber posts. Int J Prostho- BIS-GMA resin short glass ®ber composites. J Biomedical
dontics 1996;9(2):131±6. Materials Research 1989;23:1195±211.
[114] James SP, Jasty M, Davies J, Piehler H, Harris WH. A fracto- [133] Krebs J, Friedrich K, Heuwinkel R, Rieger P, Kosack P. Fiber-
graphic investigation of PMMA bone cement focusing on the reinforced thermoplastic composite osteosynthesis implants with
relationship between porosity reduction and increased fatigue variable mechanical properties for better treatment of bone
life. J Biomedicl Materials Research 1992;26:651±62. fractures. Proc. ICCM-11, 1997. p. 509±517.
[115] Jancar J, Dibenedetto AT. Fiber reinforced thermoplastic [134] Kuiper JH, Huiskes R. Mathematical optimization of elastic
composites for dentistry, Part I hydrolytic stability of the properties: application to cementless hip stem design. J Bio-
interface. J Materials Science: Materials in Medicine mechanical Engineering 1997;119:166±74.
1993;4:555±61. [135] Kwarteng KB, Stark C. Carbon ®ber reinforced PEEK (APC-2/
[116] Jancar J, Dibenedetto AT, Goldberg AJ. Thermoplastic ®ber- AS4) composites for orthopedics implants. SAMPE Quarterly
reinforced composites for dentistry Part II E€ect of moisture on 1990;22(1):10±14.
¯exural properties of unidirectional composites. J Materials [136] Lacroix H, Boel K, Nevelsteen A, Suy R. Early in¯ammatory
Science: Materials in Medicine 1993;4:562±8. response to gelatin- and collagen-sealed Dacron prostheses.
[117] Jenkins DHR. Experimental and clinical application of carbon AnnVascSurg 1995;9(2):152±4.
®ber as an implant in orthopedics. J Bone and Joint Surgery [137] Lakes R. Composite biomaterials. In: Bronzino JD, editor. The
1977;59-B(4):501. Biomedical Engineering Handbook. Bota Racon, Florida, USA:
[118] Jockisch KA, Brown SA, Bauer TW, Merritt K. Biological CRC Press, 1995. p. 598±610.
response to chopped-carbon-®ber-reinforced peek. J Biomedical [138] Langer R, Vacanti J. Tissue engineering. Science 1993;260:920±6.
Materials Research 1992;26:133±46. [139] Langer R, Vacanti J. Arti®cial organs. Scienti®c American 1995;
[119] Kahl S, et al. Ein¯uss von Zementierungsfehlern auf die Knochen- September 100±3.
zement-Beanspruchung. Biomedizinische Technik 1993;38:298± [140] Latour RA, Black J. Development of FRP composite structural
302. biomaterials: Ultimate strength of the ®ber/matrix interfacial
[120] Kajiwara H. A new cardiac wall substitute made of Dacron bond in in vivo simulated environments. J Biomedical Materials
fabric prosthesis sealed with autologous tissue fragments: com- Research 1992;26:593±606.
parative studies of neointima formation and bleeding resistant [141] Latour Jr. RA, Black J. Development of FRP composite struc-
property against ®brinolysis phenomenon with those of a pre- tural biomaterials: fatigue strength of the ®ber/matrix interfacial
clotted sealing graft. Artif Organs 1995;19(1):64±71. bond in simulated in vivo environments. J Biomedical Materials
[121] Kennedy KC, Chen T, Kusy RP. Behavior of photopolymerized Research 1993;27:1281±91.
silicate-glass-®ber-reinforced dimethacrylate composites sub- [142] Laurencin CT, Norman ME, Elgendy HM, el-Almin SF, All-
jected to hydrothermal ageing. J Materials Science: Materials in cock HR, Pucher SR, Ambrosio AA. Use of polyphosphazenes for
Medicine 1998;9:243±8. skeletal tissue regeneration. J Biomed Mater Res 1993;27:963±73.
[122] Kettunen J, Makela A, Miettinen H, Nevalainen T, Heikkila M, [143] Lee IY, et al. E€ects of variation of prosthesis size on cement
Tormala P, Rokkanen P. Fixation of femoral shaft osteotomy stress at the tip of a femoral implant. J Biomedical Materials
with an intramedullary composite rod: An experimental study Research 1994;28:1055±60.
on dogs with a two-year follow-up. J Biomaterials Science [144] Leong KW, Ramakrishna S. Sca€olds engineering. Annual
Polymer Edition 1999;10(1):33±45. Reviews of Biomedical Engineering (in press).
[123] King MW, Poddevin N, Guidoin R, Marois Y, et al. Designing [145] Lin TW, Corvelli AA, Frondoza CG, Roberts JC, Hungerford
textile structures to repair and replace knee ligaments. Canadian DS. Glass peek composite promotes proliferation and osteo-
Textile Journal 1996;113(3):53±7. calcin production of human osteoblastic cells. J Biomedical
[124] Klein CPAT, van der Lubbe HBM, de Groot K. A plastic Materials Research 1997;36(2):37±144.
composite of alginate with calcium phosphate granules [146] Liu Q, de Wijn JR, van Blitterwijk CA. Composite biomaterials
as implant material: an in vivo study. Biomaterials 1987;8:308± with chemical bonding between hydroxyapatite ®ller particles
10. and PEG/PBT copolymer matrix. J Biomedical Materials
[125] Klein N, Carciente ML, Cohn D, Marom G, Uretzky G, Peleg Research 1998;40(3):490±7.
H. Filament-wound composite soft tissue prostheses: controlling [147] Loher U, Tognini R, Mayer J, Koch B, Wintermantel E.
compliance and strength by water absorption and degradation. J Development of a cortical bone screw made with endless carbon
Materials Science: Materials in Medicine 1993;4:285±91. ®ber reinforced Polyetheretherketone (CF-PEEK) by extrusion.
[126] Knowles JC, Hastings GW. In vitro and in vivo investigation of A new method. In: 7th international conference on polymers in
a range of phosphate glass-reinforced polyhydroxybutyrate- medicine and surgery, Amsterdam, The Netherlands, The Insti-
based degradable composites. J Materials Science: Materials in tute of Materials, 1993. pp 88±97.
Medicine 1993;4:102±6. [148] Louis JP, Dabadie M. Fibrous carbon implants for the main-
[128] Knowles JC, Hastings GW, Ohta H, Niwa S, Boeree N. Devel- tenance of bone volume after tooth avulsion: ®rst clinical
opment of a degradable composite for orthopedic use: in vivo results. Biomaterials 1990;11:525±8.
biomechanical and histological evaluation of two bioactive [149] Lovdahl PE, Nicholls JI. Pin-retained amalgam cores Vs cast-
degradable composites based on the polyhydroxybutyrate poly- gold dowel-cores. J Prosthetic Dentistry 1977;38(5):507±14.
mer. Biomaterials 1992;13(8):491±6. [150] Majola A, Vainionpaa S, Vihtonen K, Mero M, Vasenius J,
[129] Ko FK. Fiber reinforced composites in medical applications. In: Tormala P, Rokkanen P. Absorption, biocompatibility, and
Lectures on textile structural composites. Taipei, Taiwan: China ®xation properties of polylactic acid in bone tissue: an experi-
Textile Institute 1999. p. 3B3-3B6. mental study in rats. Clinical Orthopedics 1991;268:260.
[130] Kocvara S, Kliment CH, Kubat J, Stol M, Ott Z, Dvorak J. [151] Marcolongo M, Ducheyne P, Garino J, Schepers E. Bioactive
Gel-fabric prostheses of the ureter. J Biomedical Materials glass ®ber/polymeric composites bond to bone tissue. J Biome-
Research 1967;1:325±36. dical Materials Research 1998;39(1):161±70.
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1221

[152] Marois Y, Roy R, et al. Histopathological and immunological [172] Noishiki Y, Yamane Y, Tomizawa Y. Sealing of highly porous
investigations of synthetic ®bres and structures used in three fabric vascular prostheses by adipose connective tissue frag-
prosthetic anterior cruciate ligaments: in vivo study in the rat. ments instead of preclotting with fresh blood. J Invest Surg
Biomaterials 1993;14(4):255±62. 1993;6(3):231±40.
[154] Matsumoto A, Noishiki Y, Ichikawa Y, Soma T, Kondo J, [173] Noishiki Y, Yamane Y, Tomizawa Y, Okoshi T, Satoh S,
Kosuge T. Sealing of a fabric vascular prosthesis with auto- Takahashi K, Yamamoto K, Ichikawa Y, Imoto K, Tobe M.
logous adipose tissue: a preliminary report of its clinical appli- Rapid neointima formation with elastic laminae similar to the
cation. Arti®cial Organs 1995;19(1):51±6. natural arterial wall on an adipose tissue fragmented vascular
[155] Mayer J. Gestricke aus Kohlensto€asern fur biokompatible prosthesis. Asaio J 1994;40(3):M267±0M272.
Verbundwerksto€e, dargestellt an einer homoelastischen [174] Noishiki Y, Yamane Y, Tomizawa Y, Okoshi T, Satoh S,
Osteosyntheseplatte. PhD thesis, ETH Zurich, Switzerland, Wildevuur CR, Suzuki K. Rapid endothelialization of vascular
1994. prostheses by seeding autologous venous tissue fragments. J
[156] Mayer J, Wintermantel E. Thermoforming process for knitted Thorac Cardiovasc Surg 1992;104(3):770±8.
fabric reinforced thermoplastics. New manufacturing techniques [175] Okazaki M, Ohmae H. Mechanical and biological properties of
for load bearing anisotropic implants. In: Bhattacharyya D, apatite composite resins. Biomaterials 1988;9:345±9.
editor. Composite Sheet Forming. Amsterdam: Elsevier Science, [176] Orthopedic Composites (1991). Stuart M Lee, editor. Interna-
1997. p. 403±40. tional Encyclopedia of Composites. VCH Publishers, New
[157] Mayer J, Rueux K, Koch B, Wintermantel E, Schulten T, York, 1991. Vol. 4, p. 74±87.
Hatebur A. The double die technique (DDT): biomaterials [177] Paganetto G, Mazzullo S, De Lolliis A, Buscaroli S, Rocca M,
processing for adaptable high fatigue resistance thermoplastic Fini M, Giardino R. Poly-l-lactid acid: biointeraction and
carbon ®ber osteosynthesis plates. J Biomedical Engineering, processing variable relationship. Biomaterials 1991;5(2):179.
Application, Basis, Communication 1993;5:778±83. [178] Park JB. Biomaterials Science and Engineering. New York:
[158] McGrath MH. Biomaterials in plastic surgery. In: Geogiade GS, Plenum Press, 1984.
Georgiade NG, Riefkohl R, Barwick WJ, editors. Textbook of [179] Park JB, Lakes RS. Biomaterials An Introduction. New York:
Plastic, Maxillofacial, and Reconstructive Surgery. Baltimore, Plenum Press, 1992 (p. 169-83).
USA: Williams and Wilkins, 1992. p. 93±101. [180] Parsons JR, Alexander A, Weiss AB. Absorbable polymer-®la-
[159] McKenna GB, Bradley GW, Dunn HK, Statton WO. Mechan- mentous carbon composites: a new concept in orthopedic bio-
ical properties of some ®ber reinforced polymer composites after materials. In: Szycher M, editor. Biocompatible Polymers,
implantation as fracture ®xation plates. Biomaterials Metals, and Composites. Lancaster: Technomic Publishing Co.
1980;1:189±92. Inc, 1983. p. 873±905.
[160] McKenna GB, Bradley GW, Dunn HK, Statton WO. Degra- [181] Pascaud RS, Evans WT, McCullagh PJJ, FitzPatrick DP.
dation resistance of some candidate composite biomaterials. J In¯uence of gamma-irradiation sterilization and temperature on
Biomedical Materials Research 1979;13:783±98. the fracture toughness of ultra-high-molecular-weight poly-
[161] Mendes DG, Iusim M, Angel D, Rotem A, Mordehovich D, ethylene. Biomaterials 1997;18:727±35.
Ro€man M, Lieberson S, Boss J. Ligament and tendon sub- [182] Patel M, Arnell RE, Sauvage LR, Wu HD, Shi Q, Wechezak AR,
stitution with composite carbon ®ber strands. J Biomedical Mungin D, Walker M. Experimental evaluation of ten clinically
Materials Research 1986;20:699±708. used arterial prostheses. Ann Vasc Surg 1992;6(3):244±51.
[162] Menger MD, Hammersen Fetal. In vivo assessment of neo- [183] Pedro Kato Y, Christiansen DL, Hahn RA, Shieh SJ, Goldstein
vascularization and incorporation of prosthetic vascular bio- JD, Silver FH. Mechanical properties of collagen ®bers: a com-
grafts. Thorac Cardiovasc Surg 1992;40(1):19±25. parison of reconstituted and rat tail tendon ®bers. Biomaterials
[163] Messner K, Gillquist J. Prosthetic replacement of the rabbit 1989;10:38±42.
medical menicus. J Biomedical Materials Research 1993;27: [184] Peluso G, Ambrosio L, Cinquegrani M, Nicolis L, Saiello S,
1165±1173. Tajana G. Rat peritoneal immune response to carbon ®ber
[164] Miettinen VM, Vallittu PK. Release of residual methyl metha- reinforced epoxy composite implants. Biomaterials 1991;12:231±
crylate into water from glass ®ber-poly (methyl methacrylate) 5.
composite used in dentures. Biomaterials 1997;18(2): 181±5. [185] Peter T, Tognini R, Mayer J, Wintermantel E. Homoelastic,
[165] MjoÈberg B. Fixation and loosening of hip prostheses. Acta anisotropic osteosynthesis system by net-shape processing of
Orthopedica Scandinavia 1991;62:500±8. endless carbon ®ber reinforced polyetheretherketone (PEEK).
[166] More N, Baquey C, Barthe X, Rouais F, Rivel J, Trinquecoste In: Goh JCH, Nather A editors. Proc. of 9th Conference on
M, Marchand A. Biocompatibility of carbon-carbon materials: Biomedical Engineering, Singapore, National University of Sin-
in vivo study of their erosion using 14carbon labeled samples. gapore, 1997. p. 317±9.
Biomaterials 1988;9:328±34. [186] Peterson CJ, Donachy JH, Kalenak A. A segmented poly-
[167] Morrison C, Macnair R, MacDonald C, Wykman A, Goldie I, urethane composite prosthetic anterior cruciate ligament in vivo
Grant MH. In vitro biocompatibility testing of polymers for study. J Biomedical Materials Research 1985;19:589±94.
orthopedic implants using cultured ®broblasts and osteoblasts. [187] Philips D. Characterization and development of 3D-knitted
Biomaterials 1995;16(13):987±92. composites. PhD thesis, Katholieke University Leuven, Bel-
[168] Moyen BJ-L, Lahey PJ, Weinberg EH, Harris WH. E€ects on gium, 1999.
intact femora of dogs of the application and removal of metal [188] Pietrucha K. New collagen implant as dural substitute. Bioma-
plates. J Bone and Joint Surgery 1978;60A(7):940±7. terials 1991;12(3):320±3.
[170] Nazre A, Lin S. Theoretical strength comparison of bioabsorb- [189] Pilliar RM, Blackwell R, Macnab I, Cameron HU. Carbon
able (PLLA) plates and conventional stainless steel and titanium ®ber-reinforced bone cement in orthopedic surgery. J Biomedi-
plates used in internal fracture ®xation. In: Harvey Jr. JP, cal Materials Research 1976;10:893±906.
Games F. editors. Clinical and Laboratory Performance of [190] Pongor P, Betts J, Muckle DS, Bentley G. Woven carbon sur-
Bone Plates. 1994. p. 53±64. face replacement in the knee: independent clinical review. Bio-
[171] Neugebauer R, Helbing G, Wolter D, Mohr W, Gistinger G. materials 1992;13(15):1070±6.
The body reaction to carbon ®ber particles implanted into the [191] Poss R, et al. Femoral expansion in the presence of total hip
medullary space of rabbits. Biomaterials 1981;2:182±4. arthroplasty. J Arthroplasty 1987;2:259±64.
1222 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

[192] Pourdeyhimi B, Wagner HD. Elastic and ultimate properties of [213] Schmitt-Thomas KhG, Yang ZG, Hiermer T. Performace char-
acrylic bone cement reinforced with ultra-high-molecular-weight acterization of polymeric composite implant rod subjected to
polyethylene ®bers. J Biomedical Materials Research torsion. In: Proceedings of ICCM-11, Gold Coast, Australia,
1989;23:63±80. 1997. p. V277±V286
[193] Pradas MN, Calleja RD. Reproduction in a polymer composite [214] Schneider E, et al. A comparative study of the initial stability of
of some mechanical features of tendons and ligaments. In: cementless hip prostheses. Clin Orthop 1989;248:200±9.
Szycher M, editor. High Performance Biomaterials: A [215] Scholz C, Huimacher D, Bahr W, Cales L. The development
Comprehensive Guide to Medical and Pharmaceutical and testing of three biodegradable screw plate-systems for max-
Applications. Lancaster, USA: Technomic Publishing, 1991. p. illofacial surgery. In: Kim HL, Chung LP. editors. Biodegrad-
519±23. able Implants in Fracture Fixation; 1993. p. 67±73.
[195] Pro®o AE. Biomedical Engineering. New York: John Wiley & [216] Sclippa E, Piekarski K. Carbon ®ber reinforced polyethylene for
Sons, 1993. possible orthopedic uses. J Biomedical Materials Research
[196] Purton DG, Payne JA. Comparison of carbon ®ber and stainless 1973;7:59 ±570.
steel root canal posts. Quintessence International 1996;27(2):93±7. [218] Seedham BB. Ligament reconstruction with reference to the ante-
[197] Ramakrishna S. Developoment of ¯exible knitted fabric rein- rior cruciate ligament of the knee. Mechanics of Human Joints. In:
forced composites for soft tissue replacements. The Japanese Wright V, Radin EL, editors. Physiology, Pathophysiology, and
Journal of Arti®cial Organs 1997;26(4):S±61. Treatment. New York, USA: Marcel Dekker, 1993. p. 163±201.
[198] Ramakrishna S. Biomedical applications of polymer composite [219] Silverton C, Rosenberg AO, Barden RM, Sheinkop MB,
materials. In: Hong CS, Kim CS, Kim CS, editors. Proc. of the Galante JO. The prosthesis-bone interface adjacent to tibial
second Asian-Australasian Conference on Composite Materials components inserted without cement. The J Bone and Joint
(ACCM-2000), Korea. Advanced Institute of Science and Surgery 1996;78-A(3):340±7.
Technology: Korea, 2000. p. A431±0A436. [220] Simoes JA, Marques AT, Jeronimidis G. Design of a controlled-
[199] Ramakrishna, S. (2000B). Textile sca€olds in tissue engineering. sti€ness composite proximal femoral prosthesis. Composites
Smart ®bers, fabrics, and clothing: fundamentals and Science and Technology 1999;60:559±67.
applications. Editor, X. Tao, Woodhead Publishing Limited, [221] Speidel MO, Uggowitzer PJ. Biocompatible nickel-free stainless
UK. steel to avoid nickel allergy. In: Speidel MO, Uggowitzer, edi-
[200] Ramakrishna S, Fujihara K, Yoshida E, Hamada H. Design tors. Materials in Medicine. Switzerland: vdfHochschulverlag
and development of braided carbon ®ber reinforced polymer AG an der ETH zurich, 1998. p. 191±208.
composites for dental posts and bone plates. In: 29th FRP [222] St.John KR. Applications of Advanced Composites in Ortho-
symposium, Japan Society for Materials Science, Japan, 2000. p. pedic Implants. In: Szycher M., editor. Biocompatible Polymers,
11±4. Metals, and Composites. Lancaster: Technomic Publishing,
[201] Ramakrishna S, Ganesh VK, Teoh SH, Loh PL, Chew CL. 1983. p. 861±71.
Fiber reinforced composite product with graded sti€ness. Sin- [223] Suwanprateeb J, Tanner KE, Turner S, Bon®eld W. In¯uence of
gapore Patent Application No. 9800874-1, 1998. Ringer's solution on creep resistance of hydroxyapatite rein-
[202] Ramakrishna S, Ramaswamy S, Teoh SH, Tan CT. Develop- forced polyethylene composites. J Materials Science: Materials
ment of a knitted fabric reinforced elastomeric Composite in Medicine 1997;8:469±72.
intervertebral disc prosthesis. Proc ICCM-11, Conard Jupiters- [224] Tallent MA, Cordova CW, Cordova DS, Donnelly DS. Ther-
Golad oast, Australia 1997;1:458±66. moplastic ®bers for composite reinforcement. In: Lee SM, edi-
[203] Ramakrishna S, Ramaswamy S, Teoh SH, Hastings GW, Tan tor. International Encyclopedia of Composites, Volume 2. New
CT. Application of textiles and textile composites concepts York: VCH Publishers, 1990. p. 466±80.
for biomaterials development. In: International conference [225] Tamura J, Kawanabe K, Yamamuro T, Nakamura T, Kokubo
on New Textiles for Composites, Aachen, Germany, 1996. p. 27/ T, Yoshihara S, Shibuya T. Bioactive bone cement: The e€ect of
1±27/27. amounts of glass powder and histologic changes with time. J
[204] Robin GC. Below-knee drop-foot braces: stresses during use Biomedical Materials Research 1995;29:551±9.
and evaluation of design. In: Ghista DN editor Biomechanics of [226] Tayton K, Bradley J. How sti€ should semi-rigid ®xation of the
Medical Devices, Marcel Dekker Inc. 1981. p. 535±567. human tibia be? J Bone and Joint Surgery 1983;65-B:312±5.
[205] Rodriguez-Lorenzo LM, Salinas AJ, Vallet-Regi M, San [227] Tayton K, Johnson-Nurse C, Mckibbin B, Bradleym J, Hastings
Roman J. Composite biomaterials based on ceramic polymers. GW. The use of semi-rigid carbon ®ber reinforced plastic plates
I. Reinforced systems based on Al2O3/PMMA/PLLA. J Biome- for ®xation of human fractures. J Bone and Joint Surgery
dical Materials Research 1996;30:515±22. 1982;64B(1):105±11.
[206] Roman JS, Garcia PG. Partially biodegradable polyacrylic- [228] Tayton K, Phillips G, Ralis Z. Long-term e€ects of carbon
polyester composites for internal bone fracture ®xation. Bioma- ®ber on soft tissues. J Bone and Joint Surgery 1982;64-B(1):112±
terials 1991;12:236±41. 4.
[207] Rubin P. Hip joint replacement and morphometric aspects of [229] Tayton KJJ. The use of carbon ®ber in human implants: the
the bone-implant system. Ecole Polytechnique Federale De state of the art. J Med Engng Tech 1983;7:271±2.
Lausanne, Lausanne, 1992. p. 189. [230] Teoh SH, Tang ZG, Ramakrishna S. Development of thin
[208] Rubin LR. Biomaterials in maxillo-facial surgery. In: Szycher elastomeric composite membranes for biomedical applications. J
M, editor. Biocompatible Polymers, Metals, and Composites. Materials Science: Materials in Medicine 1999;10:343±52.
Lancaster, USA: Technomic Publishing, 1983. p. 941±51. [231] Timmie Topoleski LD, Ducheyne P, Cuckler JM. The fracture
[209] Rushton N, Rae T. The intra-articular response to particulate car- toughness of titanium-®ber-reinforced bone cement. J Biomedi-
bon ®ber reinforced high density polyethylene and its constituents: cal Materials Research 1992;26:1599±617.
an experimental study in mice. Biomaterials 1984;5:352±6. [232] Toh SL, Goh JCH, Tan PH, Tay TE. Fatigue testing of energy
[210] Saha S, Pal S. Mechanical properties of bone cement: a review. J storing prosthetic feet. Prosthetics and Orthotics International
Biomedical Materials Research 1984;18:435±62. 1993;17:180±8.
[212] Sbarbati R, Giannessi D, et al. Pyrolytic carbon coating enhan- [233] Tonino AJ, Davidson CL, Klopper PJ, Linclau LA. Protection
ces Te¯on and Dacron fabric compatibility with endothelial cell from stress in bone and its e€ect. J Bone and Joint Surgery
growth. International J Arti®cial Organs 491-;14(8). 1976;58-B(1):107±13.
S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224 1223

[234] Torbjorner A, Karlsson A, Syverud M, Hensten-Pattersen A. [254] Williams DF, McNamara A, Turner RM. Potential of Poly-
Carbon ®ber reinforced root canal posts, mechanical properties etheretherketone (PEEK) and carbon-®ber-reinforced PEEK in
and cytotoxic properties. Eur J Oral Sci 1996;104:605±11. medical applications. J Materials Science Letters 1987;6:188±
[235] Tormala P, Vainionpaa S, Kilpikari J, Rokkanen P. The e€ects 90.
of ®ber reinforcement and gold plating on the ¯exural and ten- [255] Williams D. Concise Encyclopedia of Medical and Dental
sile strength of PGA/PLA copolymer materials in vitro. Bioma- Materials. Oxford, UK: Pergamon Press, 1990.
terials 1987;8:42±5. [256] Williams DF, Cunningham J. Materials in Clinical Dentistry.
[236] Tormala P, Vasenius J, Vainionpaa S, Laiho J, Pohjonen T, Oxford, UK: Oxford University Press, 1979.
Rokkanen P. Ultra-high-strength absorbable self-reinforced [258] Wintermantel E, Mayer J. Anisotropic biomaterials strategies and
polyglycolide (SR-PGA) composite rods for internal ®xation of developments for bone implants. In: Wise DL, Trantolo DJ,
bone fractures: In vitro and in vivo study. J Biomedical Materi- Altobelli DE, Yaszemiski JD, Gresser JD, Schwartz ER, editors.
als Research 1991;25:1±22. Encyclopedic Handbook of Biomaterials and Bioengineering, Part
[237] Trabert KC, Cooney JP. The endodontically treated tooth, B-1. Marcel Dekker: New York, 1995. p. 3±42.
restorative concepts and techniques. Dental Clinics of North [259] Wintermantel E, Bruinink A, Eckert K, Ruex, K Petitmermet
America 1984;28(4):923±51. M, Mayer J. Tissue engineering supported with structured
[238] Uhtho€ HK, Finnegan M. The e€ects of metal plates on post- biocompatible materials: goals and achievements. In: Speidel
traumatic remodeling and bone mass. J Bone and Joint Surgery MO, Uggowitzer, editors. Materials in Medicine. Switzerland:
1983;65B(1):66±71. vdf Hochschulverlag AG an der ETH zurich, 1998. p. 1±136.
[239] Urbaniak JR, Bright DS, Hopkins JE. Replacement of inter- [260] Wintermantel E, Ha SW. Biokompatible Werksto€e und Bau-
vertebral discs in chimpanzees by silicone-Dacron implants: a weisen: Implantate fur Medizin und Umwelt. Berlin, Germany:
preliminary report. J Biomedical Materials Research Sympo- Springer-Verlag, 1998.
sium 1973;4:165±86. [261] Wintermantel E, Koch B, Mayer J, Snobl B, Scheidegger A.
[240] Valdevit ADC, Inoue N, MacWilliams BA, Anderson LL. CAD, CAE and CAM for a new anisotropic carbon ®ber rein-
Methods for mechanical testing of spinal constructs. Spine forced hip endoprosthesis stem, technology, test procedures and
1996;10(2):231±48. qualifying preclinical results. J Biomedical Engineering, Appli-
[241] Verheyen CCPM, de Wijin JR, van Blitterwijk CA, de Groot K. cation, Basis, Coomunication 1994;6:18±21.
Evaluation of hydroxyapatite/poly(l-lactide) composites: [262] Wintermantel E, Mayer J, Blum J, Eckert KL, Luscher P,
mechanical behavior. J Biomedical Materials Research Mathey M. Tissue engineering sca€olds using superstructures.
1992;26:1277±96. Biomaterials 1996;17:83±91.
[242] Verheyen CCPM, de Wijn JR, van Blitterwijk CA, de Groot K, [263] Woo SLY, Akeson WH, Levenetz B, Coutts RD, Matthews JV,
Rozing PM. Hydroxyapatite/poly(l-lactide) composites: an ani- Amiel D. Potential application of graphite ®ber and methyl
mal study on push-out strengths and interface histology. J Bio- methacrylate resin composites as internal ®xation plates. J Bio-
medical Materials Research 1993;27:433±44. medical Materials Research 1974;8:321±38.
[243] Wagner HD, Cohn D. Use of high-performance polyethylene [264] Woo SL-Y, Akeson WH, Coutts RD, Rutherford L, Doty D,
®bers as a reinforcing phase in poly(methylmethacrylate) bone Jemmott GF, Amiel D. A comparison of cortical bone atrophy
cement. Biomaterials 1989;10:139±41. secondary to ®xation with plates with large di€erences in bend-
[244] Walker PS, et al. Strains and micromotions of press-®t femoral ing sti€ness. J Bone and Joint Surgery 1976;58A(2):190±5.
stem prostheses. J Biomechanics 1987;20:693±702. [265] Wright TM, Fukubayashi T, Burstein AH. The e€ect of carbon
[245] Walker PS, Blunn GW. Biomechanical principles of total knee ®ber reinforcement on contact area, contact pressure, and time-
replacement design. In: Mow VC, Hayes WC, editors. Basic dependent deformation in polyethylene tibial components. J
Orthopaedic Biomechanics. Philadelphia, USA: Lippincott- Biomedical Materials Research 1981;15:719±30.
Raven Publishers, 1997. p. 461±93. [266] Wroblewski BM. Wear and loosening of the socket in the
[246] Ward R, Minns RJ. Woven carbon ®ber mesh patch versus Charnley low-friction arthroplasty. Orthop Clin North Am
Dacron mesh in the repair of experimental defects in the lumbar 1988;19:627±30.
fascia of rabbits. Biomaterials 1989;20:425±8. [267] Yang JM, Huang PY, Yang MC, Lo SK. E€ect of MMA-g-
[247] Weinans H. Mechanically induced bone adaptations around hip UHMWPE grafted ®ber on mechanical properties of acrylic bone
implants. PhD thesis. University of Nijmegen, Netherlands, 1991. cement. J Biomedical Materials Research 1997;38(4):361±9.
[248] Wening JV, Marquardt H, Katzer A, Jungbluth KH. Cytotoxi- [268] Yildiz H, Chang FK, Goodman S. Composite hip prostheses
city and mutagenicity of Kelvar1: an in vitro evaluation. Bio- design. II. Simulation. J Biomedical Materials Research
materials 1995;16(4):337±40. 1998A;39(1):102±19.
[249] Wenz LM, Merritt K, Brown SA, Moet A, Ste€ee AD. In [269] Yildiz H, Ha SK, Chang FK. Composite hip prostheses design.
vitro biocompatibility of polyetheretherketone and polysulfone I. Analysis. J Biomedical Materials Research 1993;39(1):92±
composites. J Biomedical Material Research 1990;24:207±15. 101.
[250] Werkmeister JA, Edwards GA, Casagranda F, White JF, Ram- [270] Yu TJ, Chu CC. Bicomponent vascular grafts consisting of
shaw JAM. Evaluation of a collagen-based biosynthetic material synthetic absorbable ®bers. I. in vitro study. J Biomedical
for the repair of abdominal wall defects. J Biomedical Materials Materials Research 1993;27(10):1329±39.
Research 1998;39(3):429±36. [271] Yu TJ, Ho DM, Chu CC. Bicomponent vascular grafts consist-
[251] Whiteside E. The e€ect of stem ®t on bone hypertrophy and ing of synthetic absorbable ®bers: Part II: in vivo healing
pain relief in cementless total hip arthroplasty. Clinical Ortho- response. J Invest Surg 1994;7(3):195±211.
pedics 1989;247:138±47. [272] Zimmerman M, Parsons JR, Alexander H. The design and ana-
[252] Willert HG, Ludwig J, Semlitsch M. Reaction of bone to lysis of a laminated partially degradable composite bone plate
methacrylate after hip arthroplasty. J Bone and Joint Surgery for fracture ®xation. J Biomedical Materials Research: Applied
1974;56A:1368±82. Biomaterials 1987;21(A3):345±61.
[253] Williams DF. Consensus and de®nitions in biomaterials. In: [273] Zufall SW, Kennedy KC, Kusy RP. Frictional characteristics of
de Putter C, de Lange K, de Groot K, Lee AJC, editors. composite orthodontic arch wires against stainless steel and
Advances in Biomaterials. Elsevier Science: Amsterdam, 1988. ceramic brackets in the passive and active con®gurations. J
p. 11±16. Materials Science: Materials in Medicine 1998;9:611±20.
1224 S. Ramakrishna et al. / Composites Science and Technology 61 (2001) 1189±1224

[274] Ratnev, BD, Ho€man, AS, Schoen, FJ, Lemons, JE. An intro- [276] Fujihara, K, Huang, ZM, Ramakrishna, S, Satkunanathan, S,
duction to materials in medicine, Academic Press, New York, Hamada, H. Development of braided carbon/PEEK Compo-
USA, 1996;308±18 site bone plates. Advanced Composites Letters 2001;10(1):13±
[275] Cao, Y, Vacanti, JP, Ma, PX, Ibarra, C, Paige, KT, Upton, J, 20.
Langer, R, Vacanti, CA. Tissue engineering of tendon. In: [277] Loh, PL, Ravi, K, Ganesh, UK, Ramakrishna, S, Chew, CL.
Proceedings of Materials Research Society Symposium, Moisture absorption of carbon ®ber reinforced posts. J. Dental
1995;(394):83±9 Research 2000;79(5):1317.

Vous aimerez peut-être aussi