Vous êtes sur la page 1sur 16

SONOLINE G20

US

Troubleshooting Guide
System
G20 Voltage Test Points

08647831
08647815

© Siemens AG 2004
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
Siemens Medical
Ultrasound
ejt DivisionSolutions
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: US06-103.840.01.02.02 English


Replaces: US06-103.840.01.01.02 Doc. Gen. Date: 01.05
n.a.

Part No.:
sd
2 Revision / Disclaimer
1Revision / Disclaimer

Copyright
Copyright © 2004 by Siemens Corporation. All rights reserved. No part of this publication
may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into
any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of
Siemens Corporation. Siemens Corporation reserves the right to change its products and
services at any time. In addition, this manual is subject to change without notice. Siemens
Corporation welcomes customer input on corrections and suggestions for improvements
to this manual.Although Siemens Corporation has attempted to ensure accuracy through-
out this manual, Siemens Corporation assumes no liability for any errors or omissions, nor
for any damages resulting from the application or use of this information.
Trademarks
ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE,
Adara, Antares, G20, G40, G50, G60S, Omnia and KinetDx are trademarks of Siemens
Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000,
WS3000, and WebPro are trademarks of Siemens Corporation.
Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet
Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are
registered trademarks of Netscape Communications Corporation. Java and the JavaScript
name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered
trademark of Adobe Systems, Inc.
Other products and brand names are trademarks of their respective owners.
Disclaimer
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-
mens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
License Agreement
All computer programs copyright 1990-2004 by Siemens Corporation or its suppliers. Pro-
grams are licensed under the following agreement:
Siemens or its suppliers retain(s) ownership of and title to any computer program supplied
with the equipment and to the trade secrets embodied in such computer programs. Sub-
ject to the Buyer’s acceptance and fulfillment of the obligations in this paragraph, Siemens
grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any
computer program supplied with the Equipment that is necessary to operate the Equip-
ment solely on the medium in which such program is delivered for the purpose of operat-
ing the equipment in accordance with the instructions set forth in the operator’s manuals
supplied with the Equipment and for no other purpose whatsoever. Buyer may not
reverse-assemble, reverse-compile or otherwise reverse-engineer such computer pro-
grams nor may Buyer make a copy of such program or apply any techniques to derive the
trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms

SONOLINE G20 US06-103.840.01.02.02 Page 2 of 16 Siemens


01.05 sd Medical Solutions USA, Inc.
Revision / Disclaimer 3

of this license, the license granted by this paragraph shall terminate. Further, because
unauthorized use of such computer programs will leave Siemens without an adequate
remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to
restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s
suppliers of software is a direct and intended beneficiary of this end-user sublicense and
may enforce it directly against Buyer with respect to software supplied by such supplier,
and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indi-
rect, consequential, incidental or other damages arising out of the sublicense of the com-
puter programs supplied with the equipment.

Siemens US06-103.840.01.02.02 Page 3 of 16 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
4 Table of Contents
1- 0Table of Contents

1 _______ Voltage Test Points ______________________________________________ 5

PCB Voltage Test Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5


A30-1 PBIF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
A30 TXRX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
A31 TRBF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
A32 CFMS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
A33 ECPR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
A34 AVIF2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
A36 RGLT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
A37 MPXA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
A37-1 MPXT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
PCB LED Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
SWRG2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
A30 TXRX PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
A31 TRBF PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
A33 ECPR PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
A34 AVIF2 PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Confirm Voltages at the SWRG2 LED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

SONOLINE G20 US06-103.840.01.02.02 Page 4 of 16 Siemens


01.05 sd Medical Solutions USA, Inc.
Voltage Test Points 5
2-
PCB Voltage Test Points
1Voltage Test Points

The PCB Voltage Test Point section provides the test point locations of each PCB and
their corresponding values.

A30-1 PBIF 0

Fig. 1: Test Points


Pos. TP2 -37 VA
Pos. TP4 +B VA

Siemens US06-103.840.01.02.02 Page 5 of 16 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
6 Voltage Test Points

A30 TXRX 0

Fig. 2: Test Points


Pos. TP1 +5 VA
Pos. TP4 A GND
Pos. TP5 A GND

SONOLINE G20 US06-103.840.01.02.02 Page 6 of 16 Siemens


01.05 sd Medical Solutions USA, Inc.
Voltage Test Points 7

A31 TRBF 0

Fig. 3: Test Points


Pos. TP4 A GND
Pos. TP7 A GND
Pos. TP9 A GND
Pos. TP11 AD +5 V
Pos. TP12 D +3 V

Siemens US06-103.840.01.02.02 Page 7 of 16 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
8 Voltage Test Points

A32 CFMS 0

Fig. 4: Test Points


Pos. TP8 +3VD
Pos. TP9 +12D
Pos. TP10 -12D
Pos. TP11 +5D
Pos. TP14 DGND
Pos. TP18 DGND

SONOLINE G20 US06-103.840.01.02.02 Page 8 of 16 Siemens


01.05 sd Medical Solutions USA, Inc.
Voltage Test Points 9

A33 ECPR 0

Fig. 5: Test Points


Pos. TP1 +3 VD
Pos. TP5 D GND
Pos. TP6 D GND

Siemens US06-103.840.01.02.02 Page 9 of 16 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
10 Voltage Test Points

A34 AVIF2 0

Fig. 6: Test Points


Pos. TP6 D GND
Pos. TP14 +5 VD
Pos. TP15 +3 VD

SONOLINE G20 US06-103.840.01.02.02 Page 10 of 16 Siemens


01.05 sd Medical Solutions USA, Inc.
Voltage Test Points 11

A36 RGLT 0

Fig. 7: Test Points


Pos. TP2 HV +155 V
Pos. TP3 AGND
Pos. TP4 AGND
Pos. TP5 +BV
Pos. TP6 A +12 V
Pos. TP7 A -12 V
Pos. TP8 A +5 V
Pos. TP9 A -37 V
Pos. TP10 D +3 V

Siemens US06-103.840.01.02.02 Page 11 of 16 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
12 Voltage Test Points

A37 MPXA 0

Fig. 8: No Test Points

A37-1 MPXT 0

Fig. 9: No Test Points

SONOLINE G20 US06-103.840.01.02.02 Page 12 of 16 Siemens


01.05 sd Medical Solutions USA, Inc.
Voltage Test Points 13

PCB LED Indicators 0

Amber-colored LEDs are mounted on each PCB and the SWRG2. These LEDs indicate
component function or voltage settings, depending on where they are located.

SWRG2 0

LED State System State


Normally on The MAINS circuit breaker is on.

A30 TXRX PCB 0

LED State PCB Location System State


Normally on DS1 +5V (A+5V)
Normally on DS2 +12V (A+12V)
Normally on DS3 -12V (A-12V)

A31 TRBF PCB 0

LED State PCB Location System State


Normally on DS2 Received Echo start signal (VALID0)
Freeze = off
Normally on DS3 If it is off, a failure has been detected for HVMUX
on A30-1 PBIF PCB.
Normally on DS11 +12V (A+12V)
Normally on DS12 +5V (AD+5V)
Normally on DS13 +3.3V (D+3V)

Siemens US06-103.840.01.02.02 Page 13 of 16 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
14 Voltage Test Points

A33 ECPR PCB 0

LED State PCB Location System State


B mode on DS1 Valid B-mode signal
M mode off
Freeze off
Normally on DS2 uP-bus transaction
Bus error off
Normally on DS3 DMA transfer action
Normally dim DS7 Tx trigger timing signal
Freeze on
Normally on DS51 D+12V
Normally on DS52 D+5V
Normally on DS53 D-12V
Normally on DS54 D+3V

A34 AVIF2 PCB 0

LED State PCB Location System State


Normally on DS20 +3.3V (D+3V)
Normally on DS21 +5V (D+5V)
Normally on DS22 -12V (D-12V)
Normally on DS23 +12V (D+12V)

SONOLINE G20 US06-103.840.01.02.02 Page 14 of 16 Siemens


01.05 sd Medical Solutions USA, Inc.
Voltage Test Points 15

Confirm Voltages at the SWRG2 LED 0

1. Check the SWRG2 LED (located in the front of the system):


The LED on the SWRG2 is illuminated under the following conditions: power is on and
the system is in standby.
2. Check the LEDs on the I/O panel.
3. Check the LED on the control panel.
4. Check the SWRG2 check points:
- Remove the side cover.
- Remove the front shield cover.
- Check DC voltage at the pin locations listed in Fig. 7.

Fig. 10: SWRG Voltages


Pos. 1 +12V
Pos. 2 -12V
Pos. 3 +5V
Pos. 4 STB +5V
Pos. 5 NC
Pos. 6 +3.3V

Siemens US06-103.840.01.02.02 Page 15 of 16 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
16 Voltage Test Points

SONOLINE G20 US06-103.840.01.02.02 Page 16 of 16 Siemens


01.05 sd Medical Solutions USA, Inc.