Académique Documents
Professionnel Documents
Culture Documents
A Practical Guide
to TG-142
Quality Assurance of Medical Accelerators
Jimmy Jones, MS
The Methodist Hospital
Houston, TX
In collaboration with
A Practical Guide to TG-142
A Practical Guide to TG-142
Table of Contents
Table of Contents
Introduction ............................................................................................................................... 6
D-1: Daily Dosimetry, Mechanical, and Safety QA .................................................................. 7
D-1A: Dosimetry Procedures ............................................................................................................... 7
D-1B: Mechanical Procedures ............................................................................................................. 7
D-1C: Safety Procedures .................................................................................................................... 8
D-2: Daily Wedge QA ................................................................................................................ 9
D-2A: Morning Check-Out Run for One Angle .................................................................................... 9
D-3: Daily Imaging QA ............................................................................................................ 10
D-3A: Collision Interlocks .................................................................................................................. 10
D-3B: Positioning/Repositioning & Imaging and Treatment Coordinate Coincidence ........................ 10
W-1: Weekly MLC QA .............................................................................................................. 13
W-1A: MLC Qualitative Test, Travel Speed, and Leaf Position Accuracy ......................................... 13
M-1: Monthly Dosimetry QA ................................................................................................... 16
M-1A: X-Ray and Electron Output, Backup Monitor Chamber, Typical Dose-Rate Output and
Electron Beam Energy Constancy Tests ........................................................................................... 16
M-1B: Photon and Electron Beam Profile Constancy Tests .............................................................. 17
M-2: Monthly Mechanical QA ................................................................................................. 18
M-2A: Light/Radiation Field Coincidence, Jaw Position Indicators, MLC Settings vs. Radiation Field
for Two Patterns (non-IMRT), and Photon Beam Profile Constancy ................................................. 18
M-2B: Distance Check Device for Lasers Compared with Front Pointers and Localizing Lasers ...... 21
M-2C: Gantry/Collimator Angle Indicators ......................................................................................... 23
M-2D: Accessory Trays QA ............................................................................................................... 23
M-2E: Cross-hair Centering and Jaw Positing Indicators .................................................................. 24
M-2F: Treatment Couch Position Indicators QA ................................................................................ 24
M-2G: Wedge Placement Accuracy QA ............................................................................................ 26
M-2H: Latching of Wedges, Blocking Tray QA .................................................................................. 26
M-3: Monthly Imaging QA ....................................................................................................... 27
M-3A: Planar MV Imaging (EPID) AND kV Imaging QA .................................................................... 27
M-3B: Cone-Beam CT (kV and MV) QA ............................................................................................ 29
M-4: Monthly Wedge QA ......................................................................................................... 32
M-4A: Wedge Factor for all Energies ................................................................................................ 32
A-1: Annual Dosimetry QA ..................................................................................................... 33
A-1A: SRS Arc Rotation Mode and Arc Mode QA ............................................................................. 33
A-1B: X-Ray / Electron Output Calibration (TG-51) ........................................................................... 34
A-1C: Spot Check of Field-Size Dependent Output Factors for X-Ray, Output Factors for Electron
Applicators, and Physical Wedge Transmission Factor Constancy ................................................... 34
A-1D: X-Ray Beam Quality and Electron Beam Quality .................................................................... 35
A-1E: X-Ray Monitor Unit Linearity, Electron Monitor Unit Linearity, and X-Ray Output Constancy vs.
Dose-Rate ......................................................................................................................................... 35
A-1F: X-Ray Output Constancy vs. Gantry Angle, Electron Output Constancy vs. Gantry Angle, and
Electron and X-Ray Off-Axis Factor Constancy vs. Gantry Angle ..................................................... 36
A-2: Annual Mechanical QA ................................................................................................... 37
A-2A: Collimator Rotation Isocenter, Gantry Rotation Isocenter, Couch Rotation Isocenter QA ....... 37
Jimmy Jones, MS / Mobius Medical Systems, LP 4
Version 1.0
A Practical Guide to TG-142
Table of Contents
Introduction
This manual was designed to assist the clinical physicist with implementation of the American
Association of Physicists in Medicine’s (AAPM) Task Group 142 (TG-142) recommendations
for linear accelerator and associated components quality assurance (QA). It provides examples of
tests that can be performed to fulfill TG-142’s recommendations.
The manual focuses on tests that can be measured with Electronic Portal Imaging Devices EPID
or film measurements. Ion chamber and other device-oriented procedures are described in less
detail. The recommended tests are in no way the only method of fulfilling TG-142’s
recommendations and will most likely require customization for the individual facility. In some
cases alternate workflows are provided and are distinguishable by their light green font. Text in a
light blue font refers to specific test or quantity indicated in a TG-142 table.
The manual has been developed by Jimmy Jones in collaboration with Mobius Medical Systems,
LP. Mobius Medical Systems is the manufacturer of DoseLab TG-142 and DoseLab Pro,
commercial software designed to perform TG-142 analysis. Screenshots in the document were
captured using DoseLab Pro version 6.40.
In using this manual, you may wish to reference the three tables in the Microsoft Excel file
included with download of the document. The tables are combinations of the six tables in the
original AAPM TG-142 report, organized into Daily, Monthly, and Annual recommended tests.
The tables also indicate whether a given test can be performed using DoseLab Pro, and indicate
the section of this document where the test procedure is described.
Workflow
Follow the manufacturer’s recommendations for specialized devices.
Workflow
1. Place the Tel-align phantom at 100 cm SSD,
and align the phantom surface etchings with
the crosshairs.
3. Verify that the laser localization falls within the recommended tolerance distances for
your machine from the etchings on the phantom (side marks for axial and coronal lasers
shown at left; cross-hairs for ceiling and sagittal lasers shown at right).
4. Tape marks on the wall can also serve as quick visual checks for the axial and coronal
lasers.
Workflow
1. While the machine is warming up, open the vault door to ensure the beam turns off (door
interlock)
2. Verify that the machine cannot beam back on until the door is shut (door closing safety)
3. Once the machine is warmed up, verify that the radiation monitor (radiation area monitor)
and beam-on indicator lights are working (beam-on indicator).
4. Verify that you can see the treatment table on the closed circuit monitors and that you can
hear sound from inside the vault using the A/V system (audiovisual monitor)
Workflow
Run a dynamic wedge (45 ° recommended) while the machine is warming up to verify no
interlocks.
Workflow
Manually (and gently) depress each cover on the imaging detector, source housing, and EPID
to verify the machine interlocks.
Workflow
1. Acquire a computed tomography (CT) scan of the phantom with a slice thickness less
than the specified TG-142 positioning/repositioning tolerance for your system. The
central tungsten sphere should be delineated in the treatment planning software (TPS).
The generated digital reconstructed radiographs (DRRs), reference cone beam CT
(CBCT), and structures will facilitate alignment on the imaging system and result in a
more efficient workflow.
2. Place the WL-QA cube on the treatment couch so that the small
crosshairs on the phantom are aligned with the room lasers.
4. Acquire a CBCT image, and position the tungsten sphere at isocenter using the vendor’s
CBCT software for your linear accelerator (linac).
Note: Use a consistent technique for the CBCT image on a daily basis and a common
clinical technique for QA purposes. Consistency helps to identify degradation of the
system due to the imaging components, the mechanical isocenter (or surrogate used to set
up the cube), or the software. Minimize offset errors due to partial volume averaging by
using a minimum slice width of the CBCT scans less than the specified TG-142 tolerance
for your system.
5. Record the couch shift required and compare to the known offsets (provided by the
phantom manufacturer). This will verify that your lasers are within tolerance if they were
used to set up the phantom and that the scaling on the IGRT software is correct.
Note: The difference in the measured offset and known offset should be within the
imaging and treatment coordinate coincidence tolerance for TG-142.
6. Allow the CBCT software to reposition the treatment couch at the determined offset.
7. Enter the room and visually ensure that the lasers match the large crosshairs within the
TG-142 CBCT positioning/repositioning tolerance. Do not move the WL-QA cube.
9. Turn on the digital graticule in the treatment software, and verify that the center of the
tungsten sphere is less than the planar kV (shown below right) and MV (shown below
left) imaging and treatment coordinate coincidence TG-142 tolerance away from the
digital graticule.
Note: The imaging isocenter (red) is displayed with a grid overlay (green) to assist in
drawing vectors in the X and Y directions from the imaging isocenter to the center of the
cube.
Alternatively, the repositioning of the WL-QA cube can be accomplished using orthogonal
kV images rather than a CBCT. Technically, the positioning/repositioning procedure is listed
for both CBCT and planar kV and MV imaging. To fulfill both, it would be necessary to
perform the alternative repositioning of the WL-QA cube in addition to Steps 1-8 so that
positioning/reposition of the IGRT system is accomplished with both a CBCT alignment and
a planar image alignment. If this is the intent of TG-142, then at Step 4 the WL-QA cube
would have to be realigned at the intentional offset first.
Picket fence image (EPID or film-based). DoseLab will only analyze the picket fence
version with separate narrow irradiated strips, not the version with abutting irradiated
strips.
3. Select Machine QA | MLC Strip test from the menu (or select the icon), and the
automated routine will run. Results appear in a new window.
4. Review the summary of the analysis highlighted in green and confirm that all tests pass.
(For a full description of the analysis and user-defined variables and tolerances, please
see the “DoseLab Pro User’s Manual.”)
5. Export the results to PDF, XLS, or to the DoseLab Pro database as desired.
1. Review the “DoseLab Pro User’s Manual” for information on logging the DynaLogs and
obtaining access to them, if necessary.
3. Browse to the folder where the DynaLogs are stored, and click OK. FractionCHECK
automatically analyzes all DynaLogs in the folder. (Refer to the “DoseLab Pro User’s
Manual” for a complete description of the features, settings, and analysis for
FractionCHECK.”)
4. Review the summary of the analysis highlighted in green and confirm that all tests pass.
5. Export the results to PDF, XLS, or to the DoseLab Pro database as desired.
Electrometer
Workflow
1. Measure X-ray and electron output under reference conditions established at the time of
commissioning for all energies, and verify that the backup monitor chamber is within
tolerance of the primary monitor chamber.
2. Repeat the X-ray and electron output measurements at different dose rates to verify dose-
rate output constancy. (This is necessary because the RF driver and gun current are tuned
independently for each energy.)
3. Measure electron beam energy constancy by comparing full profiles (water tank) or
output ratios at different depths (solid water). Verify that the acquired values are within a
one percent tolerance from baseline values.
A water tank (requires more time for data acquisition, although the recommended tool for
annual profile constancy tests)
Sun Nuclear FS-QA phantom, EPID image of the phantom, and DoseLab Pro software
General Workflow
Follow the manufacturer’s recommendations for data acquisition and verify that the obtained
values are within one percent of the baseline photon and electron beam profiles for each
energy.
M-2A: Light/Radiation Field Coincidence, Jaw Position
Indicators, MLC Settings vs. Radiation Field for Two
Patterns (non-IMRT), and Photon Beam Profile Constancy
Recommended Items
Sun Nuclear FS-QA Phantom
EPID image of the phantom or film images at symmetric and asymmetric field sizes
Note: If you use film for analysis, a separate test is required to satisfy the Jaw Position
Indicators procedure.
3. Verify that the jaw positions indicators are within tolerance by matching the resultant
light field with the phantom etchings.
4. Acquire an EPID image (as shown below in the DoseLab Pro Single Image Analysis
module).
3. Place sufficient buildup over the film to produce darkening of the film at depth once
irradiated.
5. Using a new sheet of film (or an unexposed region), repeat Steps 2-4 for an asymmetric
jaw setting.
6. Using a new sheet of film (or an unexposed region), repeat steps 2-4 for a 10 cm x 10 cm
MLC-defined field.
7. Using a new sheet of film (or an unexposed region), repeat steps 2-4 for a 15 cm x 15 cm
MLC-defined field.
2. Select Machine QA | Flatness and symmetry from the main menu or by selecting the
icon.
3. Enter the appropriate image information in the Analysis settings area. (Please refer to the
“Dose Lab Pro User’s Manual” for a full description of all user-defined inputs.) At a
minimum, enter the field size of the acquired image.
b. If the image is being used to satisfy the photon beam profile constancy, identify the
baseline profile for comparison against in the Analysis settings. First acquire water
scans to ensure that the beam profile is within tolerance. Then acquire an image of the
phantom using the procedure above and use it as your baseline for subsequent
analysis.
Note: The baseline EPID profile will probably not match your water scan profile.
Recommended Items
Front pointers
Tel-align phantom
Workflow
1. Verify the front pointers’ calibration distance by standing each up next to each other
verify that the distances match those that share the same values.
3. Place a front pointer in the accessory tray so that the tip of the front pointer is at a
distance of 100 cm SSD.
4. Place the Tel-align phantom on the table and raise it until it touches the front pointer.
5. Verify that the table vertical readout position is within the distance check device for
lasers compared with front pointer tolerance relative to the value used on a daily basis for
test D-1B.
6. Remove the front pointer and its accessory tray and align the phantom with the cross-
hairs and the light field.
7. Verify that the lasers fall within the localizing laser’s tolerance relative to the phantom
etchings and described in test D-1B.
2. Verify that the coronal lasers are within the localizing lasers tolerance distance from the
tip of the front pointer.
3. Verify that the axial and sagittal lasers split the rod down the middle or are within the
localizing lasers tolerance from the middle of the pointer.
Optional Workflow (Testing the ODI and Table Vertical Travel at Additional Depths)
While not specifically stated in TG-142, you should test the ODI and table vertical travel at
depths other than 100 cm SSD.
1. Place a front pointer in the accessory tray so that the tip of the front pointer is at a
distance of 100 cm SSD.
3. Remove the front pointer and accessory tray, and verify that the ODI reading is within the
distance check device for lasers tolerance.
4. Repeat Steps 1-3 for distances of 80 cm SSD and 120 cm SSD (or clinically relevant
extremes).
Tip: Solid water stacked on the table is needed to obtain shallower depths.
Recommended Item
Bubble or digital level
Workflow
1. Verify that the gantry position is within the Gantry / collimator angle indicators tolerance
at all four cardinal angles using the level:
a. Place the level on a flat surface of the gantry head adjacent to the range finder hole
and compare the level readout with the digital gantry readout.
b. Verify that the lasers and cross-hairs intersect while the gantry is in the 90° and 270°
gantry positions.
2. With the gantry at 90° or 270° (Varian IEC scale), rotate the collimator and verify that
the position is within the Gantry / collimator angle indicators tolerance at all four cardinal
angles using the level by placing the level on the Teflon edge where the blocks slide in
and compare the level readout to the digital readout.
Workflow:
1. Insert the graticule tray into the accessory mount.
2. Turn on the field light and verify that the cross-hairs and the graticule tray markings are
within the accessory trays tolerance of one another. (This verifies alignment between
treatment isocenter and the physical graticule.)
5. Verify that the electronic graticule and the graticule tray markings are within the
accessory trays tolerance of one another. (This verifies alignment between the physical
graticule tray and the imaging isocenter.)
Workflow
1. Place the graph paper directly on the couch.
2. Set the gantry, collimator, and couch to 0° and the SSD to 100 cm.
4. Vary the symmetric jaw settings through clinically relevant field sizes (smallest size used
to the maximum size graph paper can accommodate) and verify that the light field and
the graph paper are within the jaw position indicators tolerance.
Tip: Laminate the graph paper used and mark the standard field sizes on the graph paper
to quickly assess congruency.
5. Repeat Step 4 with the asymmetric jaws at predetermined settings, including settings of 0
cm and 10 cm at minimum.
6. Rotate the collimator through full range of motion and verify that during the rotation the
cross-hairs of the graph paper and the collimator do not deviate beyond the cross-hair
centering tolerance for your machine.
1. Place the treatment couch at isocenter with the indexing bar latched at the 0 position on
the couch.
3. Verify that the couch digital readout is 0 cm (lat), 0 cm (vert), and 140 cm (long) or
within the treatment couch position indicators tolerance of these baseline values.
4. Move the couch laterally to the preset demarcation and verify that the couch readout is
within the treatment couch position indicators tolerance relative to the baseline value.
Tip: Test four lateral offsets total: two in each direction from isocenter at maximal offset
(23 cm for Varian IGRT couches) and an intermediate position.
6. Move the couch longitudinally until the indexing bar and the cross-hairs align.
7. Verify that the longitudinal offset is within the treatment couch position indicators
tolerance (The Varian IGRT couch top has 14 cm longitudinal offsets between each
indexing position.)
1. Place the graph paper on the table and align it with the cross-hairs at an SSD of 100 cm.
2. Move the couch laterally to various offsets and verify that the digital couch offset reading
is within the treatment couch position indicators tolerance relative to the graph paper.
3. Place the non-calibrated indexing bar (one used for patient treatment) at one of the
indexing positions on the couch.
4. Move the couch so you can see the cross-hairs from the collimator on the indexing bar.
7. Move the treatment couch so that the collimator cross-hairs and the cross-hairs drawn on
the indexing bar are aligned.
8. Verify that the couch readout is within the treatment couch position indicators tolerance
from the baseline offset value (14 cm for Varian IGRT couches).
Workflow
1. Set the gantry and collimator to 0°.
2. Insert each wedge into the accessory slot and turn on the field light.
3. Measure the distance from isocenter to the edge of the shadow and verify that the
distance is within the wedge placement accuracy tolerances relative to the baseline
values.
Workflow
1. Visually inspect each wedge and tray for loose screws.
2. Insert each into the accessory tray and verify that each insert and wedge latch correctly.
Recommended Items
• DoseLab Pro software
Workflow:
1. Set the MV-QA phantom up at isocenter using the phantom stand. Place the phantom in
the stand so that the two sagittal alignment markers are visible.
2. Align the phantom so that the light field cross-hairs (or room lasers) match the cross-hairs
on the face of the phantom and the sagittal alignment markers on the top of the phantom.
4. Repeat Steps 1-3 with the kV-QA phantom and the OBI system.
5. Import the images into DoseLab Pro’s Single Image Analysis module.
7. Select the appropriate phantom from the drop-down menu and verify the imported
DICOM information.
8. Select the Process button for automated analysis of the image. (Refer to the “DoseLab
Pro User’s Manual” for detailed information on the image analysis and preferences.)
Note: The example shown is an automated analysis of the Sun Nuclear kV-QA phantom
image.
9. Record the results, and export them to PDF, or store them in the DoseLab Pro database as
desired.
Workflow
1. Set up the Catphan phantom per the manufacturer’s recommendations.
4. Navigate to the stored location of the acquired CBCT and click OK.
5. Select the correct phantom and verify the imported DICOM information in the Image
information section of the CT Imaging QA window. (Please refer to the “DoseLab Pro
User’s Manual” for a complete description of the analysis performed and the
customizable user settings and tolerances).
6. Record the results, and export them to PDF, or store them in the DoseLab Pro database as
desired.
Workflow
1. Set up the ion chamber and solid water at reference conditions (field size, depth, and
SSD) for which the wedge factor is defined.
Note: Some systems (e.g., Pinnacle) use a reference field of 10 cm x 10 cm at a depth of
10 cm for all measurements, while some use a reference field size the same as that used
during the wedge measurement (i.e., a 15 cm x 15 cm wedge measurement is normalized
to a 15 cm x 15 cm open field and not a 10 cm x 10 cm open field measurement).
4. Divide the wedge reading by the reference reading and verify that the measured output
factor is within the Wedge Factor for all Energies tolerance relative to the baseline value
used for dose calculation.
Recommended Items
None
Workflow
Run a sample arc treatment and verify that the Clinac DynaLogs are within the SRS Arc
Rotation Mode and Arc Mode tolerances for your machine.
Workflow:
Please refer to AAPM’s TG-51 report for a complete description of how to perform output
calibration.
Workflow:
1. Set up the water tank and ion chamber at reference conditions (field size, depth, SSD).
Note: Most treatment planning systems normalize output measurements to an open 10 cm
x 10 cm field size at a depth of 10 cm, but some facilities may choose a depth of dmax
instead. You must verify the reference depth.
2. Acquire a set of reading at the reference conditions and average the results.
3. Acquire a set of readings at various field sizes (or applicator sizes in the case of
electrons). Readings should include at least two field sizes, preferably small (< 4x4 cm2)
and large (>4x4 cm2).
4. Divide the acquired reading by the reference readings and verify that the measured field
size-dependent output factors are within the Spot Check of Field Size Dependent Output
Factors for X-ray tolerance relative to the baseline values used for dose calculations.
6. Repeat Steps 2-4 for each electron energy, testing at least one applicator for each energy.
Note: Reference conditions for electron measurements are typically different than for
electron measurements (e.g., Pinnacle recommends dmax rather than a depth of 10 cm).
You must verify the reference conditions when measuring cone factors.
7. Acquire a set of readings at the reference conditions used to measure the wedge
transmission factors that were input into the treatment planning system. A single field
size is sufficient. For example, Pinnacle normalizes all wedge measurements to an open
10 cm2 field size at a depth of 10 cm, so a series of measurements under these conditions
would suffice for reference measurements.
Jimmy Jones, MS / Mobius Medical Systems, LP 34
Version 1.0
A Practical Guide to TG-142
A-1: Annual Dosimetry QA
8. Acquire a set of readings for each physical wedge with the physical wedge in both
possible orientations and average the results.
9. Divide the acquired wedge reading by the open reference reading and verify that the
measured output factor is within the Physical Wedge Transmission Factor Constancy
tolerance relative the value used for dose calculations.
Workflow
Most water tank manufacturers have profile acquisition software that allows you to set up
acquisition queues, shift acquired data where needed, and convert from ionization to dose
where needed. Use these tools and follow the manufacturer’s directions on how do so. Be
sure to complete full scans rather than spot checks of PDD10 and R50, since the scans can be
analyzed for profile constancy at the same time and fulfill other annual dosimetry
requirements set forth by TG-142.
Workflow
1. Set up the solid water (or water tank) and ion chamber at reference conditions, usually
100 cm SSD, 10 x 10 cm2 field size, and a depth of 10 cm.
2. Deliver 400 MU with a clinically relevant dose rate and record the measurement.
3. Using the same setup conditions, deliver 100 MU with the same dose rate and record the
measurement.
4. Verify that the 100 MU reading is1/4th of the 400 MU to within the X-Ray Monitor Unit
Linearity tolerance.
5. Deliver 2 MU with the same dose rate and verify that the reading is 1/50th of the 100 MU
reading to within the X-Ray Monitor Unit Linearity tolerance.
6. Deliver 100 MU with other clinically relevant dose rates, and verify that each100 MU
delivered readings are equal to one another within the X-Ray Output Constancy vs. Dose
Rate tolerance.
8. Shift the depth of the ion chamber to a reasonable depth for electrons (near dmax).
9. Deliver 400 MU with a clinically relevant dose rate and record the measurement.
10. Deliver 100 MU with the same dose rate and the same setup.
11. Verify that the 100 MU reading is 1/4th of the 400 MU reading to within the Electron
Monitor Unit Linearity tolerance.
Workflow
1. Insert the IMF device into the gantry accessory mount per vendor recommendations.
2. Insert the Daily QA3 device into the IMF device per vendor recommendations.
3. Deliver 100 MU with low energy photons to the device at a gantry angle of 0 (gantry
facing down) and the maximum field size permitted by the device.
4. Record the output readings at central axis and at specified points off-axis.
6. Verify the output at each point is equal for each gantry angle to within the X-Ray Output
Constancy vs. Gantry Angle (central axis) and Electron and X-Ray Off-Axis Factor
Constancy vs. Gantry Angle (off-axis diodes) tolerance.
2. Set the jaws of the collimator so that a thin field is produced (i.e., 0.4 cm x 20 cm –
symmetric about the central cross-hair).
3. Irradiate the film at various collimator angles. (An equidistant 30° spacing would result
in six spokes.)
Example: This collimator starshot film was produced with gafchromic film. The film
was placed at isocenter and buildup placed on top of the film to permit dose deposition
and film darkening. The film was irradiated with six collimator angles.
4. Repeat Steps 1-3 with a new piece of film and by varying the couch angle.
5. Repeat Steps 1-3 with a new piece of film and by varying the gantry angle.
Tip: For the gantry isocenter test, stand up the solid water vertically so that it is
perpendicular to the axis of rotation of the gantry. Position the film so that the center of
the film is still at 100 cm SSD and centered relative to the central axis laterally.
Note: Winston-Lutz style testing, described in greater detail in the section “Coincidence
of Radiation and Mechanical Isocenter,” is an alternate to verify rotation isocenter.
2. Import the scan into DoseLab Pro using the Single Image Analysis module.
4. Adjust the analysis settings as needed. (For a detailed description of the user-defined
settings, preferences, and analysis, please refer to the “DoseLab Pro User’s Manual.”)
5. Select Process.
7. Select a circle diameter that passes through all the spokes of the image and click.
8. Verify that the diameter of the circle touching all spokes is within the Collimator, Gantry,
and Couch Rotation Isocenter tolerances.
9. Record the results, print them to PDF, or store them in the DoseLab Pro database.
Workflow
1. Verify that the machine interlocks when you touch the applicator’s touch pad.
2. Verify that the machine interlocks when the applicator is not fully inserted into the
accessory slot.
Workflow
1. Set up the WL-QA phantom on the treatment couch at isocenter so that the cross-hairs
from the light field align with the central cross-hairs etched into the phantom. Repeat this
with the lateral faces of the phantom by rotating the gantry 90° in both directions.
2. Acquire an EPID image of the cube at all four cardinal gantry angles.
Tip: It is recommended that the field size for the EPID image be defined by the MLCs
and with a field size smaller than that of the cube itself, but large enough to contain the
tungsten sphere at the center of the cube. This aids in threshold analysis of the image.
4. Navigate to the storage location of the acquired images and select all four images. (You
can select up to eight images if you tested different combinations of
gantry/couch/collimator angles.)
5. Verify the DPI and SID image information were imported from the DICOM file.
6. Adjust the threshold settings as needed to delineate the edge of the field and the sphere,
using the “DoseLab Pro User’s Manual” if necessary.
7. Verify that the total displacement for each image is within the Coincidence of Radiation
and Mechanical Isocenter tolerances.
8. Record, export to PDF, or save within the DoseLab Pro database as desired.
Alternate Workflow
You can modify this test so that various collimator angles and couch angles are used during
the EPID image acquisition to satisfy the Collimator Rotation Isocenter and Couch Rotation
isocenter respectively. The workflow would be the same as for the gantry, but for each
respective test, the other two variables remain at 0° (i.e., gantry and collimator at 0° and
rotate couch through cardinal angles to acquire EPID images).
Workflow
1. Extend the treatment couch to its maximum longitudinal position towards the gantry with
the surface of the couch at isocenter.
3. Verify that the displacement of the table is less than the Table Top Sag tolerance.
Workflow
Rotate the couch through its full range and compare the digital readouts to the mechanical
readouts at various angles. Verify that their agreement is within the Table Angle tolerance.
Alternate Workflow
To verify couch centering, place a piece of graph paper on the couch and align it with the
light field cross-hairs. Rotate the couch through its full range and verify that the center of the
light field and graph paper cross-hairs stay within agreement through the full range of
rotation.
Note: Many of the weekly and monthly workflows satisfy the annual MLC procedures. The
reference to those tests is provided in the test description column of the TG-142 spreadsheet.
Workflow
1. Set up the ion chamber in solid water or a water tank so that the sensitive volume
intersects the cross-hairs of the light field.
2. Acquire a reading with the jaws open: 200 MU with a 10 x 10 cm2 jaw defined field at
100 cm SSD and gantry at 0° (Varian IEC, gantry pointed down) is recommended if there
are no previous measurements. If previous baselines were established, mimic the same
reference field setup from those measurements.
3. Deliver the same number of MUs as in Step 2 and record the output.
4. Divide the offset closed field output by the open field output and verify that the interleaf
leakage is within the MLC Transmission tolerance relative to the baseline.
6. Acquire a film of the closed MLCs to visually verify there are no failing regions, since
the above workflow tests only the leakage in one region.
Workflow
1. Set the graph paper on the treatment couch at isocenter with the gantry, table rotation,
and collimator set to 0°.
3. Generate an MLC-defined field and note the shadow of the leaves position on the graph
paper.
Tip: Use a sheet of graph paper exclusively for this test so that the MLC pattern can be
drawn onto the graph paper for easier analysis.
5. Verify that the leaves return to their previous positions to within the Leaf Position
Repeatability tolerance.
Alternate Workflow
Leaf position repeatability is also satisfied with DoseLab Pro’s FractionCHECK module,
which is described earlier in this document.
Workflow:
1. Set the film at isocenter with the gantry, table rotation, and collimator all set to 0°.
2. Place enough buildup on the film to permit dose deposition and film darkening.
3. Generate a thin MLC-defined field symmetric about the central axis (similar to in section
“Collimator Rotation Isocenter, Gantry Rotation Isocenter, Couch Rotation Isocenter
QA,” but with the MLC field rather than the jaws).
4. Irradiate the film at 45° collimator intervals with the same MLC-generated field. Include
an additional irradiated field to produce a minimum of 5 “spokes” on the image.
8. Adjust the analysis settings as needed, referring to the “DoseLab Pro User’s Manual” as
necessary.
9. Select Process.
11. Select a circle diameter that passes through all the “spokes” of the image and click.
12. Verify that the diameter of the circle touching all “spokes” is within the MLC Spoke Shot
tolerance.
13. Record, export to PDF, or save within the DoseLab Pro database as desired.
Recommended Item
Tape measure
Workflow:
1. With gantry head up, position the MV detector at a preset distance.
Note: Varian customer acceptance uses a value of 0 / 0 / 0 so that the detector itself is at
100 cm away from the source. The imaging layer (without the covers on) is 1.2 cm below
the surface for E-Arm models. It is recommended at the time of commissioning to
determine the distance of the imaging layer with the covers on. This can be done by first
verifying the position with the covers off, then placing the covers on without moving the
detector, and reading out the new distance with the ODI.
2. Measure the distance using either the front pointer or ODI and verify that the distance is
within the Full Range of Travel SDD tolerance.
3. Position the arm at its most distal position and drive the longitudinal and lateral axes to
both limits.
4. Record the total range of travel for each axis and verify that the range is within the Full
Range of Travel SDD tolerance relative to the baseline values determined at customer
acceptance.
Tip: Place a piece of tape on the imager panel and the cross-hairs drawn on it to aid in
displacement measurements.
Solid water
Workflow
1. Set up the Profiler 2 device per the manufacturer’s recommendations.
2. Add solid water on top of the device so that the diode plane is at a depth of 10 cm.
4. Verify that the off-axis ratios at 80% field size are within 2% of the baseline values.
Workflow
DoseLab Pro’s VMAT analysis is based on the tests performed in Ling et al. (2008) and
therefore it is necessary to be able to duplicate these tests. You can manually set these tests or
obtain the plans on Varian’s website. Plans include a Picket Fence test performed during an
arc, as well as arcs in which the dose rate, MLC leaf speed, and gantry speed are modulated.
1. Deliver each plan and for each acquire either an EPID image or film image.
2. For the two MLC tests in which the gantry speed, MLC speed, and/or dose rate are
modulated, select the VMAT/Dynamic MLC button in DoseLab Pro.
3. Load the open field image and the MLC field image. DoseLab Pro performs an
automated analysis of the images and provides a detailed summary.
4. Load the picket fence images into DoseLab Pro’s Single Image Analysis module and
perform the Picket Fence analysis as described in section “MLCA Qualitative Test.”