and maintenance
Application Note
Introduction
Loop-powered isolator
Process instrumentation
requires periodic calibration
and maintenance to ensure that
Output
it is operating correctly. This from
application note contains 2-wire
information to help guide tech- transmitter Loop
nicians through some of the power
supply
more common loop calibration
tasks using Fluke tools.
Fluke loop calibrators are
ideal for a wide variety of cali- Two-wire transmitter
bration applications. They
include: Input types; typical
• Fluke 707 Loop Calibrator Thermocouple
• Fluke 705 Loop Calibrator Frequency
RTD
• Fluke 715 Volt/mA Calibrator DC volts
AC volts
• Fluke 787 and 789 DC current
ProcessMeter™ tools AC current Loop power
A summary of the features of Strain gauge supply
Resistance 24 V–48 V
Fluke’s loop calibrators and a
glossary of terms are at the end
of this application note. Figure 1
Loop V
V
mV
mA
A
mA
isolator
loop-powered isolator, the two-
OUTPUT
supply OFF
A
OUTPUT 0-24mA
SOURCE
mA
SIMULATE
COM
mA
mA
250
HART
LOOP POWER
V
AutoHOLD
HOLD
FAST MN MX
MIN MAX
CANCEL
Hz % ms
REL
RANGE
F
mV
ac+dc
C
A
V
ac+dc
mA
mV A
dB
OFF A
Simulate
ac+dc
TEMPERATURE
Positioner
Flow
Figure 3
Step 1: Basic set-up. Setting Step 3: Span (full open) Step 5: Linearity check
the Fluke 707 loop calibra- position check For valves with linear action,
tor current output Using the 25 % button, step linearity can be checked by
Place the calibrator in the 4-20 the calibrator output value to setting the 707 to 4 mA and
mA output current mode. 20 mA and allow the valve to stepping current to 12 mA
Connect the 707 to the input stabilize. Step the input to (50 %) while observing valve
terminals of the valve posi- 20.1 mA using the vernier knob travel. If your valve is of a non-
tioner (see Figure 3). in the depressed position turn- linear type, refer to the valve
Step 2: Zero adjustment ing clockwise while watching manual for proper operational
Set the 707 to an output of 4 or feeling for movement of the checks.
mA and allow some time for valve stem. Minimize this move- Step 6: Stroking the valve
valve stem movement to stabi- ment using the span adjust- Checking for smooth valve
lize. Quickly decrease the ment on the valve positioner. operation is easy to accomplish
current from 4 mA to 3.9 mA Using the vernier knob in using the slow ramp function of
by depressing and turning the the depressed position, adjust the 707.
current up and down between
vernier knob in a counter-
20.10 mA and 19.9 mA. There • Set the calibrator to mA
clockwise direction. You can source mode and select
operate the 707 with one hand should be no movement of the the slow ramp function ( )
∨
while feeling the valve stem valve stem above 20 mA and a by depressing the 25 %
with your free hand to check slight movement below 20 mA. and 0-100 % buttons
for any sign of movement. Step 4: Check zero and simultaneously.
Adjust for zero movement span again • Allow the calibrator to ramp
between these two current set- Many positioners have interac- through several cycles while
tings by using the zero tive zero and span controls. watching or feeling for any
adjustment on the positioner. This step will help ensure proper abnormal operation of the
Increase and decrease cur- valve position adjustment. valve, such as sticking in
rent from 4 mA to 4.1 mA using Repeat Step 2 and Step 3. one position momentarily or
the vernier knob in the erratic movement.
depressed position. Ensure that
the valve stem just begins
movement above the 4.1 mA
setting and is fully closed at
4 mA.
Figure 4
1 kΩ resistor
Figure 5
Figure 5.
Step 1: Setting the Fluke 707
to source 4-20 mA 707 with Set DMM to
dc current
shunt assembly
1. With the shunt assembly in installed (connections
vary)
the “source” jacks, power
the 707 on. Figure 6
2. Check the display. If the dis-
play does not read 4 mA,
turn the 707 off and on Step 3: Checking linearity Making fine adjustments
again while holding the Once zero and span controls Many signal conditioners with
mode button for two sec- have been properly set, signal 0-20 mA and 4-20 mA outputs
onds. The display should conditioner linearity may be are notorious for zero and span
now read 4.000 mA. verified using the following control interaction. If, when
Step 2: Calibrating procedure. This procedure will checking linearity in the cali-
1. Place a precision multimeter, check zero, 25 %, 50 %, 75 % bration section, your output
such as the Fluke 187 DMM, and span settings for signal meter displayed a value higher
set to dc current mode, in conditioner linearity. or lower than 4 mA, perform
series with the output of the 1. With the precision resistor the following steps to affect the
signal conditioner as shown assembly in place, adjust the required 4 and 20 mA display
in Figure 6. source current of the Fluke on the output meter.
2. Connect the test leads from loop calibrator to 4 mA (1.0 1. Note the value above or
the precision shunt assembly V) using the 25 % button. below 4 mA that was dis-
to the signal conditioner The DMM displays 4 mA. played on the output meter
input terminals, observing 2. Using the 25 % button, step when you returned to a
proper polarity. the source current to 25 %, source value of 4 mA on the
50 %, 75 % and 100 % and 707. Adjust the zero control
3. With the 707 set in the mA on the signal conditioner so
source mode, the display note the corresponding val-
ues. Table 2 shows the the value of the output
should read 4 mA (1.00 V meter shows one-half the
across the shunt). correct values of output for a
linear signal conditioner. (If difference of the remaining
4. Adjust the zero control on values differ from that mA value to 4 mA.
the signal conditioner for an shown in Table 2 by more Example: If your display
indication of 4 mA on the than the linearity specifica- reading at zero input in Step
DMM. tion of the signal conditioner, 7 was 3.50 mA adjust the
5. Press the 0-100 % button on contact the signal condi- output (with 4 mA source
the 707 until the display tioner manufacturer or current) to display a reading
reads 20 mA. Then adjust replace the device.) of 3.75 mA, which is one-
the span adjustment on the half the delta toward the
signal conditioner until the desired value of 4 mA; e.g.,
display on the DMM reads 4.00 - 3.50 = 0.50; 0.05 /
20 mA. 2 + 3.50 = 3.75, or one-half
the difference between the
6. Push the 0-100 % button reading and the desired
and the 707 should read 4. value.
Verify the meter connected
to the output signal condi-
tioner reads 4 mA.1
Fluke display Input Output
7. Calibration is now complete. reading % Input voltage current
Note: Many manufacturers specify 4.000 0 1.000 V 4.000 mA
a warm-up time prior to beginning
calibration. 8.000 25 2.000 V 8.000 mA
12.000 50 3.000 V 12.000 mA
1
If your output meter displayed a value higher 16.000 75 4.000 V 16.000 mA
or lower than 4 mA, consult the “Making fine
adjustments” section on this page. 20.000 100 5.000 V 20.000 mA
Table 2. Correct value of output for a linear signal conditioner
10
0%
0%
input
ing. Adjust the span control
of the signal conditioner –
one-half the delta from
20 mA.
Example: If the output dis-
play reads 21 mA, adjust the
span control to 20.5 mA
(one-half the delta to the
required value of 20 mA).
3. Repeat this “one-half step”
process until the required
output is obtained. There are
signal conditioners that have
non-interactive controls that
do not require this procedure.
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
Canada (800)-36-FLUKE or
Fax (905) 890-6866
Other countries (425) 446-5500 or
Fax (425) 446-5116
Web access: http://www.fluke.com
©2003 Fluke Corporation. All rights reserved.
Trademarks are the property of their respective owners.
8 Fluke Corporation Loop calibration and maintenance Printed in the Netherlands 4/2003 Pub ID 10437-eng