Vous êtes sur la page 1sur 7

Cryogenics 80 (2016) 174–180

Contents lists available at ScienceDirect

Cryogenics
journal homepage: www.elsevier.com/locate/cryogenics

Research paper

Pyrolytic graphite film thermal straps: Characterization testing


Ian M. McKinley ⇑, Colin H. Smith, Perry G. Ramsey, Jose I. Rodriguez
Jet Propulsion Laboratory, California Institute of Technology, USA

a r t i c l e i n f o a b s t r a c t

Article history: This paper reports on the experimentally-measured conductance, stiffness, and particulate contamina-
Received 7 September 2016 tion of pyrolytic graphite film thermal straps. This work was aimed at assessing the feasibility of replac-
Received in revised form 8 October 2016 ing standard aluminum foil in thermal straps with graphite film, which is more conductive and lighter.
Accepted 10 October 2016
Four different U-shaped straps with similar cross-sections and terminals were tested in the study.
Available online 1 November 2016
Three of the straps had a three-inch long flexible section. One of these was made from aluminum
1100 foil, and two were made from Pyrovo pyrolytic graphite film (PGF). One of the PGF straps was fab-
Keywords:
ricated with an aluminized mylar blanket that was sealed at the terminals. The last strap was made from
Pyrolytic graphite film
Flexible thermal link
PGF, was blanketed, and was six inches long. The conductance of each strap was measured as a function
Annealed pyrolytic graphite of mean strap temperature ranging from 60 K to 300 K. The peak measured conductance of the three-inch
PGF and aluminum straps were 1.0 W/K at 162 K and 0.28 W/K at 64 K, respectively. The conductance of
all straps converged to around 0.3 W/K as the mean strap temperature approached 60 K. In addition, the
peak conductance of the six-inch PGF strap was 0.83 W/K at 150 K. The fact that its peak conductance was
near the conductance of the three-inch PGF strap indicated that the thermal resistance of the terminals in
the PGF straps was significant. For a given temperature, the conductance varied by as much as 15% for
two units of the same strap design. One of the straps was thermally cycled from 300 K to 60 K ten times.
Its conductance was unchanged by the thermal cycling. Furthermore, one of the six-inch long PGF straps
was subjected to random vibration. The random vibration spectrum was designed so that one terminal
achieved a maximum displacement of 0.25 in. from its neutral position in three orthogonal axes while
the other was held stationary. The conductance of this strap was unaffected by the random vibration test.
The straps were also tested for the level of contamination introduced to the environment. The bare
aluminum and bare PGF straps had equal particulate cleanliness levels while the encapsulated PGF strap
had a lower one. Finally, the dynamic stiffness of one of the six-inch strap was measured to be less than
0.5 lb/in. in all directions for temperatures between 200 and 300 K.
Ó 2016 Elsevier Ltd. All rights reserved.

1. Introduction and background cryocooler. Finally, the strap must not introduce other
contaminants onto the focal plane.
Thermal straps are intended to achieve a thermal coupling and Thermal straps consist of two rigid end-pieces (terminals) sep-
a structural decoupling between cryogenic components [1]. They arated by a flexible middle section composed of numerous thin lay-
often physically connect a mechanical cooling source, such as a ers of foil or film. The flexible elements and terminals are thermally
cryocooler, to infrared (IR) detectors and focal planes possessing linked by soldering, swaging, welding, or other proprietary meth-
highly critical alignment requirements. Mechanical flexibility of ods [1]. The thermal performance of a thermal strap is evaluated
the strap is needed to accommodate launch loads as well as isolate in terms of its thermal conductance, G (in W/K) [1]. Pure copper
the forces exported by the cryocooler to the IR detector or focal straps provide excellent thermal performance, however they suffer
plane. In addition, small mass is needed to minimize the launch from large mass and stiffness. The thermal conductivity of pure
loads on thin-walled cryocooler coldfingers. Large thermal aluminum is 40% smaller than that of pure copper at 300 K [2].
conductance of the strap is needed for efficient operation of the However, pure aluminum possesses density and stiffness that are
30% and 60% than that of pure copper, respectively [3]. Aluminum
thermal straps often meet the mass, conductivity, flexibility, and
reliability requirements needed [4]. However, Pyrovo pyrolytic
⇑ Corresponding author at: 4800 Oak Grove Drive M/S 157-316, Pasadena, CA
graphite film (PGF) has emerged as a new candidate for the flexible
91109-8099, USA.
material in thermal straps. Its thermal conductivity at 300 K is
E-mail address: ian.m.mckinley@jpl.nasa.gov (I.M. McKinley).

http://dx.doi.org/10.1016/j.cryogenics.2016.10.002
0011-2275/Ó 2016 Elsevier Ltd. All rights reserved.
I.M. McKinley et al. / Cryogenics 80 (2016) 174–180 175

approximately 3.5 and 6 times greater than pure copper and pure specifies the strap material, centerline length, blanketing, mass,
aluminum, respectively [2]. In addition, its thermal conductivity at and conductance measured by Thermotive at 304 K of each strap.
100 K is approximately 10 and 16.5 times greater than pure copper The blanketing consisted of a 4 lm thick sealed mylar blanket
and pure aluminum, respectively [2]. Furthermore, its density is equipped with a HEPA filter to allow venting and mitigate potential
much smaller than that of copper and aluminum. The specific ther- particle contamination. Version 1 was a standard aluminum strap
mal conductivity of a material, defined as its thermal conductivity against which various characteristics of the Pyrovo PGF straps
divided by its density, provides a figure of merit for thermal strap could be compared. This version was made from aluminum 1100
comparison. The specific thermal conductivity at 300 K of pyrolytic foils which feature a thermal conductivity of 340 W/m K at 60 K
graphite, pure aluminum, and pure copper are 0.63, 0.088, and which is 32% lower than that of 99.999% (5-nines) aluminum [1].
0.045 W m2/kg K, respectively [2]. In addition to being an excellent Each terminal had five 3.6 mm diameter mounting holes in a circu-
thermal conductor, PGF is highly flexible possessing a modulus of lar pattern with a diameter of 33 mm. Version 3 had the same
elasticity of 23 GPa at 300 K as measured by beam deflection [5]. geometry as version 1 and was approximately twice as expensive.
The modulus of elasticity of PGF is 5 and 3 times less than that Also, the electrical conductance of the PGF straps varied from 1.0 to
of pure copper and pure aluminum, respectively [3]. 134.2 ohms. Note that this was not a parameter that was con-
Although PGF meets the requirements of providing excellent trolled for this study. Finally, note that Thermotive Pyrovo PGF
thermal coupling per unit mass as well as providing sufficient straps typically do not have any exposed graphite. However, Ther-
mechanical decoupling, there are uncertainties related to (i) the motive agreed to fabricate a pair of PGF straps (version 2) without
contamination it introduces to its environment and (ii) its reliabil- the standard mylar blanketing for this study.
ity. One approach to eliminating contamination is to encapsulate
annealed pyrolytic graphite (APG). Thermacore Inc. has developed
and patented an APG material that can be encapsulated by any 2.2. Conductance test
number of metals, composites, plastics, or ceramics. However,
the encapsulating shell is thick and causes the resulting structure The conductance of straps 1-I, 3-I, 3-II, 4-I, and 4-II was tested
to be stiff and unsuitable as a thermal strap material. Alternatively, as a function of temperature between 60 K and 300 K in a vacuum
Thermotive LLC builds straps out of PGF and blankets them with a chamber held at a pressure less than 1  105 torr (1.33 mPa).
thin, sealed mylar blanket that does not increase the stiffness of Fig. 2 shows a schematic and Fig. 3 shows a picture of the experi-
the PGF. In addition, the sealed mylar blanket is equipped with a mental apparatus used to test conductance. The bottom of each
high-efficiency particle air (HEPA) filter to allow venting of the strap was mounted to a copper plate, which was attached to the
blanket during pressure changes without allowing particles to pass cold tip of an APD DE-102 Gifford-McMahon (GM) cryocooler
through it. This blanket serves to (i) contain graphite contamina- inside a vacuum chamber. The temperature of the copper plate
tion that could exist, (ii) protect the edges of the PGF from damage was controlled by a Lake Shore 340 Temperature Controller and
during handling, (iii) facilitate cleaning, and (iv) reduce the heated by Dale resistors. The top of the strap was heated by a Dale
exposed surface area by a factor of 100. The goal of this study resistor providing constant heat load through an interface plate.
was to assess the thermal performance and contamination charac- This resistor had two wires connected to each terminal. One set
teristics of Pyrovo PGF thermal straps in comparison to ones made of wires supplied the current and the other measured the voltage
from aluminum 1100. Moreover, the reliability of the PGF straps at the resistor. In addition, the wires were used to suspend the
was assessed by subjecting them to thermal cycling and random top terminal so that it did not make contact with the bottom ter-
vibration. minal. The top and bottom terminal of the straps each had two
Lake Shore DT670 diodes mounted on bolt holes. The temperature
of each terminal was defined as the mean reading of the two
2. Materials and methods
diodes. The strap was enclosed in a temperature-controlled alu-
minum radiation shield in order to minimize radiative heat trans-
2.1. Thermal strap design
fer with the walls of the vacuum chamber. The temperature at the
bottom of the shield was approximately the same as the bottom
The thermal strap design consisted of a U-shaped flexible sec-
strap terminal. The temperature of the top of the shield was con-
tion made of 180 foils that were 0.001 in. (0.0254 mm) thick and
trolled to within 5 K of the top terminal of the strap by a Lake Shore
0.75 in. (19.05 mm) wide. Four versions of the thermal strap were
340 using Minco 180 ohm Kapton film heaters. The temperature of
conceived and two of each version were purchased from Thermo-
the top of the shield was such that the mean of the top and bottom
tive LLC. Fig. 1 shows a photograph of versions 2, 3, and 4. Table 1
shield temperatures was the same as the mean strap temperature.
In addition, the inside of the shield was lined with multi-layer
insulation (MLI) to further reduce radiative heat transfer between
the shield and the strap. Moreover, the aluminum strap was spiral
wrapped with 3 layers of 20 lm thick aluminized mylar as it would
be in a flight configuration. The mylar wrap served as a radiative
heat transfer shield between the strap and its surroundings. The
Pyrovo PGF straps were not spiral wrapped because their mylar
shield blanketing served as a radiative heat transfer shield. Finally,
the outside of the aluminum radiation shield was covered with MLI
to reduce heat transfer with the vacuum chamber.
For each conductance test, a constant heat load of 4 W was
applied to one end of the strap as the copper plate temperature
varied between 50 K and 300 K, and then a 6 W heat load was
applied to each strap for at least five temperatures between 50 K
and 300 K. Furthermore, strap 3-I was thermally cycled between
300 K and 60 K ten times. The conductance was measured at
Fig. 1. Photograph of different versions of the straps. 60 K and 300 K during each cycle.
176 I.M. McKinley et al. / Cryogenics 80 (2016) 174–180

Table 1
Strap properties. Conductance G reported by Thermotive, LLC.

Version Unit Strap material Length (in.) Blanketed Mass (g) G at 304 K (W/K)
1 I Aluminum 1100 3 No 79.32 0.22
II 81.08 0.24
2 I Pyrovo PGF 3 No – 1.08
II 62.58 1.09
3 I Pyrovo PGF 3 Yes 64.07 1.11
II 64.57 1.04
4 I Pyrovo PGF 6 Yes 78.19 0.68
II 76.75 0.67

Fig. 2. Schematic of the conductance test apparatus.

2.3. Contamination test

Fig. 4 shows the test apparatus used to measure the contamina-


Fig. 3. Photograph of the test apparatus used to measure the conductance of the
tion introduced to the environment by the thermal straps as a thermal straps. The strap in the photograph is version 4. The inner MLI that lined
result of simulated handling. One terminal of the strap was bolted the walls of the radiation shield is not shown.
to a G10 standoff and the other was bolted to an interface plate
attached to the shaft of a solenoid linear actuator. A hinge was used
to reduce the amount of stress induced on the strap when the shaft plate, replacing the aluminum cover, and operating the actuator
was actuated. The apparatus shown in Fig. 4 was placed in a lam- for two minutes. This process was repeated for straps 2-II and then
inar flow hood for the duration of the tests. First, a control test was 3-II. The witness plates were analyzed for particles using JPL Ana-
performed with no strap in place. The entire apparatus was cleaned lytical Chemistry Technical Procedure #136 (Standard Procedure
with isopropanol, a three-inch (76.2 mm) diameter silicon wafer for Determination of Cleanliness Levels and Particle Percent Cover-
witness plate was rested on the bottom of the aluminum box, age Using Silicon Wafer Witness Samples) in accordance with the
and an aluminum cover was placed over the box. Then, the sole- Institute of Environmental Sciences and Technology Contamina-
noid was actuated with stroke of approximately 4.4 mm for two tion Control Division Standard 1246E (IEST-STD-CC1246E) [6]. In
minutes at 15 Hz. Next, the control witness plate was removed addition, the conductance was measured for strap 3-II following
and strap 1-II was unsealed from its as-delivered packaging and the contamination test in order to compare with the values prior
placed in the test apparatus. The bottom of the aluminum box to being subjected to the actuation and also with the other unit
was cleaned with isopropanol prior to inserting a new witness of the same version.
I.M. McKinley et al. / Cryogenics 80 (2016) 174–180 177

Table 2
Random vibration spectrum.

Frequency (Hz) g2/Hz


5 0.01
8 0.01
9 0.1
20 0.1
2000 0.0001
Overall 2.2 grms

verify that the random vibration test did not alter the thermal
performance of the strap.

2.5. Stiffness test


Fig. 4. Photograph of the test apparatus used to measure the particle contamination
of the thermal straps. Cover and witness plate are not shown. The stiffness of strap 4-II was measured in two directions at
room temperature and at 200 K. This test was performed by Moog
CSA Engineering. Random excitation was applied to one end of the
strap while the other end was held fixed. The test setup was similar
2.4. Random vibration tests
to that of the random vibration tests except with different instru-
mentation. The two directions tested correspond to the left
Fig. 5 shows the test apparatus used to subject strap 4-I to ran-
(X-direction) and bottom right (Z-direction) photographs in
dom vibration in the vertical direction in three different orthogonal
Fig. 5. Measurements included the force through the strap and
mounting orientations. One terminal of the strap was mounted to
the displacement across the strap. The stiffness as a function of
the support frame that was fixed relative to the ground. The other
frequency was computed from these measurements.
terminal of the strap was mounted to an LDS V722 single axis sha-
ker that moved vertically relative to gravity. The fixed ends of the
support frame were bolted to the vibration shaker saddle. Table 2 3. Results and discussion
lists the specified output acceleration of the shaker as a function of
frequency. This spectrum was designed so that the moving strap 3.1. Conductance
terminal achieved a 0.25 in. (6.35 mm) maximum displacement
from its neutral position. In each orientation, the shaker was run Fig. 6 shows steady-state temperatures measured at different
for 20 s at 6 dB, 20 s at 3 dB, and then two minutes at 0 dB. locations in the test apparatus shown in Fig. 2 as a function of
The output acceleration of the shaker was continuously controlled mean strap temperature. The temperatures shown were measured
by a single axis control accelerometer. In addition, single axis for the aluminum strap 1-I with 4 W of input power on the strap
response accelerometers mounted on the support frame in the cen- heater. The mean strap temperature T strap;mean was defined as the
ter and at the joint of the vertical and horizontal beams continu- mean temperature of the four diodes mounted on the strap termi-
ously measured the vertical acceleration of the support frame. nals. The mean shield temperature T shield;mean was defined as the
High definition video was recorded during each test run to verify mean temperature of the two diodes on the shield, and T strap;C
that the stroke of the moving terminal did not exceed 0.5 in. and T strap;H were defined as the mean temperatures of the two
(12.7 mm). A witness plate was placed below the flexible section diodes on the cold and hot terminal, respectively. Fig. 6 demon-
of the strap during each test run in order to measure the contam- strates that the mean temperatures of the strap and the shield
ination production of the strap during the test. The shaker was also were nearly indistinguishable during conductance measurements
run prior to attaching the strap with a control witness plate in of the aluminum strap 1-I indicating that there was minimal radia-
place. The witness plates were analyzed in accordance with tive heat transfer between the shield and the strap. The mean
standard IEST-STD-CC1246E [6]. Finally, the conductance was shield and strap temperatures were also nearly indistinguishable
re-measured for strap 4-I following the random vibration test to during conductance measurements of the other thermal straps.

Fig. 5. Photograph of the test apparatus used to subject strap 4-I to random vibration.
178 I.M. McKinley et al. / Cryogenics 80 (2016) 174–180

PGF strap 3-I at 60 K and 300 K did not change after the strap
was cooled to 60 K ten times.
Fig. 7 also shows that, for a given mean strap temperature, the
difference in conductance between two PGF straps of the same ver-
sion was as large as 15%. This variability was likely due to differ-
ences in the thermal resistance between the PGF strap and the
terminals because all other strap characteristics are well-
controlled. In fact, the bond between the strap and the terminal
was achieved using both chemical bonds and mechanical means.
On the contrary, aluminum thermal straps typically have a pres-
sure weld bond swaged strap to terminal interface that provides
a highly repeatable, small thermal resistance [1,7]. The fact that
the conductance of the 600 (152.4 mm) PGF straps was much greater
than half the conductance of the 300 PGF straps indicates that the
strap-terminal thermal resistance is significant. The trends of the
measured conductance of these straps as a function of temperature
closely follow those of thermal conductivity versus temperature
for 5-nines aluminum and PG [1]. However, for any given temper-
ature between 60 K and 300 K, the ratio of the conductance of the
Fig. 6. Temperatures measured in different locations as a function of mean strap 300 PGF straps to that of the 300 aluminum strap was not as large as
temperature with 4 W applied to the aluminum strap 1-I. The maximum temper- the ratio of the thermal conductivity of PG to 5-nines aluminum
ature differences T shield;C  T strap;C and T strap;H  T shield;H were approximately 4 K. [1]. This is further evidence that the strap-terminal interface on
the PG straps has a large associated thermal resistance. Further-
Furthermore, the maximum temperature difference between the more, the conductance of the PGF straps measured in this study
cold temperatures of the strap and shield ðT shield;C  T strap;C Þ as well around 300 K differed from that measured by Thermotive at
as the hot temperatures of the strap and shield ðT strap;H  T shield;H Þ 304 K. This difference was due to the fact that the strap tempera-
were approximately 4 K. tures were defined differently in the two tests. The conductance
Fig. 7 shows the measured conductance as a function of mean reported by Thermotive defined the strap temperature using a sin-
strap temperature for all of the thermal straps tested. The conduc- gle temperature sensor on each terminal located in its center, and
tance was measured over at least ten minutes, and the mean is the conductance reported by Thermotive was verified in this study
shown. The conductance of the 3 in. (76.2 mm) PGF straps was by placing temperature diodes at these locations. Moreover, the
approximately 3.5 and 5 times greater than that of the 3 in. alu- conductance of the aluminum strap 1-I was the same regardless
minum strap at 300 K and 160 K, respectively. The conductance of how the strap temperature was defined. In other words, the
of the PGF straps peaked between 150 and 160 K while that of temperature gradient in the terminal of the aluminum straps was
the aluminum strap was largest at 60 K. In addition, the conduc- negligible while that in the PGF straps was not.
tance of all the straps measured was approximately 0.3 W/K at Fig. 8 shows the specific conductance as a function of mean
60 K. The conductance was not measured at lower temperatures strap temperature for the 300 straps 1-I, 3-I, and 3-II. The specific
because they converged at 60 K. In fact, the thermal conductivity conductance was defined as the conductance shown in Fig. 7
of aluminum 1100 increases with decreasing temperature from divided by the strap masses in Table 1. The mass of the aluminum
60 K to 40 K while that of pyrolytic graphite decreases [1]. Further- strap 1-I reported in Table 1 included the three-layer spiral
more, the conductance measured with 4 W and 6 W applied to the wrapped mylar that was applied during the conductance test. In
strap heater was indistinguishable; in fact, Fig. 7 contains conduc- addition, the mass of the aluminum strap 1-II reported in Table 1
tances measured at both heat loads. Finally, the conductance of did not include a mylar wrap. Fig. 8 captures the superiority of

Fig. 7. Measured conductance as a function of mean strap temperature. Fig. 8. Measured specific conductance as a function of mean strap temperature.
I.M. McKinley et al. / Cryogenics 80 (2016) 174–180 179

the PGF straps over aluminum for temperatures above 60 K. In fact,


the specific conductance of the PGF straps was as large as six times
greater than that of the aluminum strap. Moreover, a 300 copper
strap tested by Sparr et al. [8] weighing 200 g had a conductance
of 0.029 W/K at 80 K corresponding to a specific conductance of
0.15 W/kg K. The larger specific conductance of the PGF straps
was due to the fact that they were not only more conductive, but
also lighter than the aluminum strap occupying the same volume.
Note that the specific conductance of the 600 PGF straps was not
included in Fig. 8 because these straps did not occupy the same
volume as the 300 straps.

3.2. Contamination

Table 3 shows the contamination results. The percent area cov-


erage (PAC) was defined as the fraction of a surface that is covered
by particles [6]. It was reported as the total particle projected area
per 0.1 m2. The particulate cleanliness level (PCL) was defined
according to Table 1 in Ref. [6]. Note that the control test results
were not included in Table 3 because the PCL of this test was
Fig. 9. Random vibration spectrum specified in Table 2 and measured by the
greater than that of the tests with the straps in place. The larger
control accelerometer as well as the measured response of the fixed support
PCL of the control test was likely due to a larger solenoid stroke structure.
combined with greater hinge movement. Since the results of the
control test were non-conclusive, the cleanliness levels of the
straps in Table 3 are relative, not absolute. The PCL of the bare alu- that the structure was sufficiently stiff to not exceed the desired
minum and bare PGF strap was the same indicating that the PGF stroke of 0.500 . In fact, review of the HD video of the tests confirmed
strap did not contaminate more than the standard aluminum strap. that the maximum stroke was 0.500 for all three test orientations.
In addition, the blanketed PGF strap had a smaller particulate Furthermore, analysis of the witness plates revealed that the PCL
cleanliness level than the bare straps. This indicates that the was less than or equal to 300 for all three test orientations. Note
vented mylar shield blanketing effectively contained any particles that the background witness plate had a PCL of 500. Similar to
that may have been discharged from the PGF strap during simu- the contamination test, the PCL of the background plate being lar-
lated handling. Furthermore, Fig. 7 shows that, for a given mean ger than that of the plates with the straps in place inhibited the
strap temperature, the conductance of the blanketed PGF strap 3- absolute particulate cleanliness level of the straps from being
II that was subjected to the contamination test was greater than determined. In this case, the larger background PCL may have been
or equal to that of the PGF strap 3-I that was not subjected to sim- due to the fixture settling during the first operation of the shaker.
ulated handling. This indicates that the simulated handling con- Finally, the conductance of strap 4-I measured before and after the
tamination test did not affect the thermal performance of the random vibration test was identical.
PGF strap.
3.4. Stiffness
3.3. Random vibration
The strap was extremely compliant in both directions at both
The strap 4-I was subjected to random vibration in order to temperatures tested. In fact, the stiffness and loss factor of the
assess how exposure to a spaceflight launch environment affected strap could only be accurately measured between 5 and 10 Hz. In
the thermal performance, cleanliness, and vibration isolation/ the Z-direction shown in the bottom right photograph of Fig. 5,
attenuation of the strap. Fig. 9 shows the output spectrum of the the stiffness remained below 0.5 lb/in. (1 lb/in. = 175.3 N/m)
shaker as measured and as specified in Table 2. It also shows the between 5 and 10 Hz at both room temperature and 200 K. In
measured responses of the support frame in the vertical direction the X-direction shown in the left photograph of Fig. 5, the stiffness
at the center and corner of the cross beam. The measured shaker remained below 0.25 lb/in. at both temperatures. In both direc-
spectrum closely matched that specified in Table 2. Fig. 9 illus- tions at both temperatures, the damping was too low to measure.
trates that the vibration responses in the support frame were lar- Finally, in a given direction, the measured stiffness did not change
gely attenuated demonstrating that the strap acted as a vibration appreciably with temperature indicating that the stiffness can be
isolator between the shaker and the support frame. The measured considered independent of temperature between 200 and 300 K.
large amplification in the support frame near 300 Hz was due to
the resonant frequency of the cross bar being excited as was pre-
dicted by a finite element model (FEM) of the thermal strap and 4. Conclusion
support frame test fixture. In addition, the support structure
responses were extremely low in the range of 1–50 Hz indicating This work determined that it is feasible to replace the standard
aluminum foil in thermal straps with lighter and more conductive
graphite film. The peak measured conductance of the three-inch
Table 3
Contamination test results including percent area coverage (PAC) and particulate PGF and aluminum straps were 1.0 W/K at 162 K and 0.28 at
cleanliness level (PCL). 64 K, respectively. The conductance of all straps converged to
around 0.3 W/K as the mean strap temperature approached 60 K.
Strap PAC (%) PCL Fibers
In addition, the peak conductance of the six-inch PGF strap was
1-II: bare aluminum 1100 0.0004 300 None
0.83 W/K at 150 K. The fact that its peak conductance was near
2-II: bare Pyrovo PGF 0.0003 300 None
3-II: encapsulated Pyrovo PGF 0.00003 100 None the conductance of the three-inch PGF strap indicated that the
thermal resistance between the film and the terminals in the PGF
180 I.M. McKinley et al. / Cryogenics 80 (2016) 174–180

straps was significant. For a given temperature, the conductance Charles Phillips, Andres Andrade, Paul MacNeal, and Jim Post for
varied by as much as 15% for two units of the same strap design. their contributions to this work.
Furthermore, the conductance of the six-inch long PGF strap 4-I
was unchanged after the strap was subjected to random vibration References
that displaced one terminal a maximum of 0.25 in. from its neu-
tral position in three orthogonal axes while the other was held sta- [1] Bugby D, Marland B. Flexible conductive links. In: Donabedian M, editor.
Spacecraft thermal control handbook volume II: cryogenics. The Aerospace
tionary. The conductance was also not affected by thermally Press; 2003. p. 327–46.
cycling the strap from 300 K to 60 K ten times. The bare aluminum [2] Incropera FP, DeWitt DP, Bergman T, Lavine A. Fundamentals of heat and mass
and bare PGF straps had equal particulate cleanliness levels while transfer. 6th ed. New York, NY: John Wiley and Sons, Ltd.; 2006.
[3] Beer Jr FP, Johnston ER, DeWolf JT. Mechanics of materials. 4th ed. New York,
the blanketed PGF strap had a lower one. Finally, the dynamic stiff- NY: McGraw-Hill; 2006.
ness of the six-inch PGF strap was less than 0.5 lb/in. in all mea- [4] Urquiza E, Vasquez C, Rodriguez JI, Van Gorp B. Development and testing of an
sured directions for temperatures between 200 and 300 K. innovative two-arm focal-plane thermal strap (TAFTS). Cryogenics
2012;52:306–9.
[5] Pappis J, Blum SL. Properties of pyrolytic graphite. J Am Ceram Soc
Acknowledgements 1961;44:592–7.
[6] Institute of Environmental Science and Technology. Product cleanliness levels –
applications, requirements, and determination; 2013.
The work described in this paper was carried out at the Jet
[7] Williams B, Jensen S, Batty J. An advanced solderless flexible thermal link. In:
Propulsion Laboratory, California Institute of Technology under a Ross Jrl R, editor. Cryocoolers 9. Plenum Press; 1997. p. 807–12.
contract with the National Aeronautics and Space Administration. [8] Sparr L, Boyle R, Nguyen L, Frisch H, Banks S, James E, et al. Design and test of
potential cryocooler cold finger interfaces. In: Kittell P, editor. Advances in
In addition, the authors would like to thank Jerami Mennella, Euge-
cryogenic engineering. Plenum Press; 1994. p. 1253–62.
nio Urquiza, Ramesham Rajeshuni, Tim McCann, Howard Tseng,

Vous aimerez peut-être aussi