Académique Documents
Professionnel Documents
Culture Documents
Experiment 2
Temperature Measurement
Group. 1/ Section. 51
Name ID
1
Table of Contents
Introduction ................................................................................................................................ 5
Theory ........................................................................................................................................ 5
Experimental Set-up................................................................................................................... 9
Procedure ................................................................................................................................. 14
Safety considerations............................................................................................................ 15
Data Collected .......................................................................................................................... 16
Results ...................................................................................................................................... 18
Sample Calculations................................................................................................................. 27
Discussion ................................................................................................................................ 29
Conclusion ............................................................................................................................... 31
References ................................................................................................................................ 34
2
Table of Figures
3
List of Tables
4
Introduction
Temperature is defined as the energy level of matter which can be evidenced by some change
in that matter. Temperature sensors come in a so many different variety but have one thing in
common - they all measure temperature by sensing some change in a physical characteristic
[1]. All temperature sensing devices have their own advantages and disadvantages. The most
important feature of the different types of temperature measurement devices is accuracy and
linearity of measurement.
In this experiment, seven different temperature measuring devices were used to compare and
understand accurate temperature measurements. The seven devices used in the experiment
are:
The main objectives of this experiment are to learn how to operate and make connections for
the platinum resistance thermometer (PRT), NTC Thermistor, thermocouples and
thermometers. The other objective is to prove the linearity of the platinum resistance
thermometer (PRT) and the non-linearity of the NTC Thermistor as well as compare the
linearity and output signal levels of J and K type thermocouples.
Theory
A platinum resistance thermometer (PRT) is a device which determines the temperature by
measuring the electrical resistance of a section of pure platinum wire. This platinum wire is
known as the temperature sensor. This device offers excellent combination of sensitivity,
range and reproducibility which makes it one of the best temperature measuring devices. The
reason why platinum wire is used is because it is a stable unreactive metal which can be
drawn down to fine wires but is not too soft. Using very pure wires, thermometers can be
made with closely similar resistance characteristics and achieve good reproducibility in use
[2]. An increase in temperature will increase the resistance of the sensor. This proves the
linearity of a PRT. The measurement is completed by applying a small measurement voltage
and utilizing a bridge type circuit. The maximum range recommended for an industrial PRT
is -200 to +650°C [3].
5
Figure 1 - PRT schematic diagram
NTC Thermistors are non-linear resistors, which alter their resistance characteristics with
temperature. The resistance of NTC will decrease as the temperature increases. One of the
main ways to measure temperature using an NTC Thermistor is with a Wheatstone
bridge. In addition, NTC Thermistors are frequently used to compensate for fluctuations
in temperature in coils and solenoids [4]. It uses an electrical resistor made of a
semiconductor material instead of a metal.
Thermocouple is a sensor used to measure temperature. They consist of two wire legs
made from different metals. The wires legs are welded together at one end, creating a
junction. This junction is where the temperature is measured. When the junction
experiences a change in temperature, a voltage is created. The voltage can then be
interpreted using thermocouple reference tables to calculate the temperature.
There are many types of thermocouples, each with its own unique characteristics in terms
of temperature range, durability, vibration resistance, chemical resistance, and application
compatibility. Type J, K, T, & E are “Base Metal” thermocouples, the most common
types of thermocouples. Type R, S, and B thermocouples are “Noble Metal”
thermocouples, which are used in high temperature applications.
The type J is very common and has a small temperature range and a shorter
lifespan at higher temperatures than the Type K. The temperature range for this
type is usually -210 to 760C. and standard accuracy is usually +/- 2.2C or +/- .75%
[5].
The type K made of nickel and chromium or nickel and aluminum is the most
common type of thermocouple. It’s inexpensive, accurate, reliable, and has a wide
temperature range. Its temperature range is usually –270 to 1260C and it has a
standard accuracy of +/- 2.2C or +/- .75% [5]
6
is to say the volume changes significantly with changes in temperature. Combined with
the narrowness of the tube, this means that the height of the column of liquid changes
significantly with small temperature variations [6]. In this experiment, three types of
thermometers were used to measure temperature:
Gas pressure thermometer: Measures the pressure exerted by a definite amount of
gas enclosed in a constant volume. The gas is usually hydrogen or helium is
enclosed in a glass bulb connected to a mercury manometer [7].
Liquid red spirit thermometer: The alcohol/ spirit thermometer is an alternative to
the mercury-in-glass thermometer but has similar functions. The contents of an
alcohol thermometer are less toxic and will evaporate away fairly quickly. The
ethanol version is the most widely used due to the low cost and relatively low
hazard posed by the liquid in case of breakage [8].
Bimetallic thermometer: Bimetallic thermometers are made up of two metallic
strips formed by joining two different metals having different thermal expansion
coefficients. Basically, bimetallic strip is a mechanical element which can sense
temperature and transform it into a mechanical displacement [9].
In addition to the mentioned above temperature devices, various other devices can be used
which work on different principles. A few of these devices are mentioned below:
7
Figure 4 - Pyrometer
For the completion of the experiment and to get accurate results, the following equations are
used for calculations:
V
Resistance is calculated using Ohm’s law: R = I
[Equation 1]
V = voltage (V)
I = current (A)
X1 Y1 − −
Y = +Y [Equation 2]
−
X2 Y2
X3 Y3
|D viati |
Percentage error = × [Equation 4]
Sta ar
8
Experimental Set-up
The overall appearance of the temperature measurement and calibration apparatus, TD400, is
its metal frame and contains an icebox, water heater tank, low voltage connections, and a
digital display. The low voltage electrical connections are made for the low voltage
measurement devices [15]. These devices include: the platinum resistance thermometer
(PRT), negative temperature coefficient (NTC) thermistor, J type and K type thermocouples.
Other devices that are used in this apparatus but are not connected to the low voltage
connections are: liquid filled thermometers, gas and bi-metallic thermometers. This apparatus
could be placed anywhere on a desk near a draining sink or a bench top, as it is an easy
equipment to place anywhere desired. It can be connected to a software program called the
TecQuipment’s Versatile Data Acquisition System (VDAS®) [14]. The overall equipment is
shown in Figure 6. Figure 5 shows the apparatus that was used during the experiment at Abu
Dhabi University Chemical Engineering Laboratory.
Figure 5 - TD400 measuring and calibration device Figure 6 – TD400 apparatus at ADU
The equipment has a built-in temperature and pressure sensor. The temperature sensor works
accurately to show accurate temperature reference. The pressure sensor shows the local
(barometric) pressure. The display also shows the local boiling point of water based on the
barometric pressure [14].
For safety, the water heater has a water level sensor/float switch and a cut-out switch to
switch off the heater when the water level is low [15]. The water heater tank has a lid with
holes to hold the measurement tools during the experiments. It has a drain tap (shown in
Figure 8) that can be connected to a container, sink, or any water drain. This helps in easily
change water during runs in an experiment, save time and be safe [14]. At the front of the
heater tank, there is a small temperature scale to show the temperature of the water as it is
increasing and/or decreasing and is just for reference [14]
9
Figure 7 - Water heater tank Figure 8 - Water draining tap/ valve
The icebox is also safe by being insulated thermally. The insulation is made to hold the ice
and slow the melting process during experiments. There is a container-like in the icebox that
can be removed to pour out the ice or the ice water during and after experiments. Like the
heater tank, it has a lid with holes to hold the devices during the experiments.
The digital displays include the Reference Display and the Millivoltmeter (shown in Figure 9)
each with their own characteristics displayed. The reference display shows, the accurate
reference temperature given by the reference sensor, the barometric pressure, the boiling
point temperature of water based on the barometric pressure. The millivoltmeter shows four
voltages from each of the four input and output sockets on the sides of the display [15].
The low voltage connections have two main sockets: The PRT and NTC Thermistor sockets
and the Thermocouple J and K type sockets.
The PRT and NTC thermistor sockets (shown in Figure 10) have a constant current source,
constant voltage source, fixed resistances with high accuracy in a circuit that stimulates PRT
at 0°C (100 Ohm) and its connection wires (R1, R2, R3, R4), and fixed resistances with high
accuracy in a Wheatstone Bridge Circuit of three 100-R-resistors to match the 100-Oh,
resistance of a PT100 at 0°C [15].
10
Figure 10 - PRT and NTC thermistor sockets
The J and K type thermocouple sockets (shown in Figure 11) have a voltage amplifier, which
increases the output voltage from the thermocouples that are relatively small to a suitable
level for the experiment. The sockets also have four spare junction sockets that are not
electrically connected to anything, which can be used during experiments to keep tidy wiring.
Finally, these sockets have three fixed resistors, which simulate extra resistances in the
circuits used with measuring devices. [15].
1. Resistance thermometers
The types of resistance thermometers are sometimes called PRT’s and RTD’s. They refer
to Platinum Resistance Thermometer and Resistance Temperature Detectors, respectively.
They measure temperature through change in electrical resistance of a length of wire.
Platinum is used because it gives off linearity for resistance change versus temperature
and is stable. It has a very wide operating temperature range, between -200 to +650°C.
Increase in temperature will increase the resistance to the sensor. [16]. Figure 12 shows
how a PRT looks while Figure 1 gives the schematic makeup of a PRT.
11
Figure 12 - Platinum resistance thermometer (PRT)
Thermocouples of two types, two K type and one J type (shown in Figure 14), when
they’re put in an electrical or electronic circuit, they convert the small potential difference
into voltage or current of calibrated value [15]. The J type thermocouple has white and
black insulation, whereas, the K type thermocouple has green and white insulation.
(Figure 15). The J and K type thermometers use the same connections to use the
thermocouple it must be one at a time.
12
Figure 14 - J and K type thermocouples
The gas and Bi-metallic thermometers have a mechanical dial calibrated to show
temperature. The gas thermometer (shown in Figure 16) works when the gas expands due
to temperature increase, so the pressure increases and pushes against the mechanical
calibrated dial. The bi-metal thermometer (shown in Figure 17) has two metal strips held
together to make one thick strip, which is a composite of the two. As the temperature
rises, one metal expands and the other composite bends, which pushed the mechanical
calibrated, dial. [15].
13
5. Liquid filled glass thermometers
The difference between the two thermometers is that one used low toxicity liquid while
the other uses the red spirit. Opposite to the gas and bi-metal thermometers, the liquid
filled glass thermometers work on the process of liquid volume expansion, which is due
to rise in temperature. As liquid expands, it moves up or down a capillary tube behind a
scale. [15]. Figure 18 shows two liquid thermometers.
Procedure
Note: The experiment was made in two runs. Run 1 measured with the PRT, J-type
thermocouple, and the bi-metallic thermometer. In contrast, Run 2 was measured with NTC
thermistor, K-type thermocouple, gas thermometer and red spirit (liquid-filled) thermometer.
Experiment Start-up
1. Close the drain tap behind the heater tank and switch off electrical supply.
2. Remove the lid of the heater tank and add distilled water until its almost half full.
3. Cover the tank with the lid.
4. Remove the lid of the icebox, add ice and put back the lid.
5. Assure that heater switch if turned off and switch on the electrical supply.
Run 1
14
5. Place the reference sample along with the PRT, J-type thermocouple, and the
bimetallic thermometer in the heater tank, switch on the heater and record the
reference temperature.
6. With 10°C-intervals, record the readings of the millivoltmeter and the calibration
scale of the bimetallic thermometer.
Note: when reading the bi-metallic thermometer scale, make sure to tap on the calibration
scale to avoid extreme errors made by friction.
7. When temperature equals 100°C, stop the experiment and switch off the heater.
8. Remove the lid of the heater tank and wait for the water in the heater tank to cool to
70°C (add ice if necessary).
9. Drain the water. (Water needs to be changed between runs).
Run 2
1. Re-add distilled water in the water heater tank till it’s almost half full.
2. Connect the reference sensor to the socket it belongs to.
3. Connect the NTC thermistor to the millivoltmeter and the constant current source.
4. Connect the K-type thermocouple to the millivoltmeter and the amplifier.
5. Place the reference sample along with the NTC thermistor, K type thermocouple, gas
thermometer and red-spirit-glass-thermometer into the icebox, wait for the reference
temperature to stabilize at 0°C and record the other readings.
6. Place the reference sample along with the NTC thermistor, K type thermocouple, gas
thermometer and red-spirit-glass-thermometer in the heater tank, switch on the heater
and record the reference temperature.
7. With 10°C-intervals, record the readings of the millivoltmeter, the red spirit glass-
thermometer and the calibration scale of the gas thermometer.
Note: when reading the gas thermometer scale, make sure to tap on the calibration scale
to avoid extreme errors made by friction.
8. When temperature equals 100°C, stop the experiment and switch off the heater.
Experiment Shut-down
1. Remove the lid of the water heater tank, and wait till temperature decreases to 70°C.
To fasten the process, add ice to the tank, if necessary.
2. Open the drain valve and empty the tank.
3. Turn off the main power supply.
Safety considerations
During practical work proper supervision is required. All personnel should be aware of the
dangers that might occur when dealing with lab materials, high pressure and high temperature
conditions. People performing this experiment should wear appropriate PPEs including: lab
coats, safety goggles, hard covered shoes, and high-impact gloves. In addition, Students can
15
perform the experiment alone if they have competent supervision by an instructor/engineer.
Some physical, chemical, and mechanical hazards that may potentially arise while
performing the experiment are mentioned below:
Mechanical: be aware of hot materials (heater tank) as it can burn the skin.
Electrical: be gentle when using/pouring/moving water when the electrical supply is
on as electrical disasters can occur if water falls on any electric supply.
Physical: safely and careful deal with the hot tank and the ice as both can cause
physical damage to the skin.
Data Collected
To perform this experiment, seven devices were used in two different runs at different
temperatures. The below tables show the collected data of both the runs.
(oC)
16
Table 2 : Run II Data
87 16.8 51.2 88
17
Results
For each of the seven devices, the deviation and error were calculated from the standard
values provided in Appendix A.
A plot was then plotted for each of the device to show the deviation and comparison with
reference temperature
18
Resistance Vs. Reference Temperature Plot for PRT
150
140
Resistance Ω
19
Resistance Vs Reference Temperature
300
250
Resistance Ω
200
150
100
50
0
0 20 40 60 80 100 120
Reference Temperature (°C)
20
Voltage Vs. Reference Temperature from J type Thermocouple
6000
5000
4000
3000
Achsentitel
0
0 20 40 60 80 100 120
-1000
-2000
Achsentitel
21
Standard Voltage VS Measured Voltage
4095
3895
3695
3495
3295
3095
2895
2695
2495
2295
Measured Voltage(𝜇V)
2095
1895
1695
1495
1295
1095
895
695
495
295
95
-105
-100 100 300 500 700 900 1100 1300 1500 1700 1900 2100 2300 2500 2700 2900 3100 3300 3500 3700 3900 4100
-305
-505
-705
-905
Standard Voltage (𝜇V)
4000
2000
0
-2000 0 20 40 60 80 100 120
Reference Temperature(°C )
22
5. Results and calculations of gas pressure thermometer are presented in Table 7
0 0.38 0.38 -
100
80
60
40
20
0
0 20 40 60 80 100 120
Indicated Temperature (°C)
23
Temperature Vs. Number of Trials
120
100
Temperature (°C)
80
60
40
20
0
0 1 2 3 4 5 6 7 8 9 10
Number of Trials
0 -0.1 -0.1
24
Indicated Vs Reference Temperature plot for Bi-Metallic
Indicated Temperature (°C)
Thermometer
140
120
100
80
60
40
20
0
-20 0 22,4 32 42 52 62 72 82 92 98
Reference Temperature (°C)
100
Temperature (°C)
80
60
40
20
0
1 2 3 4 5 6 7 8 9 10
-20
Number of trials
25
7. Results and calculations of liquid red spirit thermometer are presented in Table 9
0 0.05 -0.05 0
37 37 0 0
47 47 0 0
77 79 -2 2.59
87 90 -3 3.44
97 101 -4 4.12
100
80
60
40
20
0
0 20 40 60 80 100 120
Reference Temperature (°C )
26
Reference Temperature&Red Spirit Indicated
Temperature
120
Temperature (°C )
100
80
60
40
20
0
0 1 2 3 4 5 6 7 8 9 10
Number of Trials
Sample Calculations
This section shows detailed mathematical calculations for finding the data required for
different measurement devices at reference temperature of 87 °C.
For PRT and NTC Thermistor data, a sample calculation of NTC thermistor only is shown
since NTC thermistor works in a similar way and same data are required to be calculated.
The data required are: experimental resistance, standard resistance, and resistance percentage
error.
Calculated Resistance
In order to find calculate resistance at 87 °C, the collected voltage values and constant
supplied current (1 mA) can be applied in Ohm’s law.
V . 𝑉 . 𝑉
Ohm's Law → R = I = = = . Ω
𝐴 𝐴
Standard Resistance
Table.1A in Appendix A shows the standard resistance of NTC thermistor at different
temperatures. Since Tref = 87 °C is located between 85 °C and 90 °C, interpolation is
preferable to find the accurate standard resistance.
− − . − . −
−
= −
− .
= −
R= 15.874 Ω
Percentage Error
𝐷 𝑖𝑎 𝑖
Percentage Error % = | |× %
𝑎 𝑎 𝑖 𝑎
27
= 15.874 Ω - . Ω = -0.926 Ω
− . Ω
Percentage Error % =| .
|× %= . %
For K & J type thermocouples, mathematical calculations for the required data is shown on K
type thermocouple as both work on the same principle of potential difference across two
different metals subject to a heat gradient. Hence, same data are required to be calculated
which are: actual measured voltage, standard voltage, and voltage percentage error.
Standard voltage
From Table.2A in Appendix A, which shows the standard voltages at different temperatures,
the standard voltage at 87 °C is 3557 𝜇V
Percentage Error
𝐷 𝑖𝑎 𝑖
Percentage Error % = | 𝑎 𝑎 𝑉 𝑎
|× %
Percentage Error % = | |× %= . %
Thermometers are manufactured in a way temperature can be measured directly from a scale
on the equipment. A sample of calculation for red spirit thermometer data is shown below.
These data are: indicated temperature and temperature percentage error.
Indicated temperature
At reference temperature of 87 °C, the liquid in red spirit thermometer rises to reach a
temperature of on the calibrated scale 90 °C
Percentage Error
𝐷 𝑖𝑎 𝑖
Percentage Error % = | |× %
𝑎
28
Deviation °C = Reference Temperature − Measured Temperature
= 87 °C-90 °C= -3 °C
−
Percentage Error % = | |× %= . %
Discussion
Through the temperature measuring experiment, knowledge of using different types of
temperature measuring devices is gained. From the found results, the accuracy and the
linearity of the measurements, which are the most important features, are compared. The
instruments worked with are: platinum resistance thermometer (PRT), negative temperature
coefficient (NTC) thermistor, K and J types thermocouples, gas and bi-metallic
thermometers, and liquid filled glass thermometer. The results found, the comparison
considered and the errors are discussed through the report.
The first device used was the PRT, with a voltage reading in millivoltage. The PRT was
equipped in the first run. Given that the constant current was set as 1 milliampere, the
resistance was calculated using Ohm’s law in order to understand the relationship between
the temperature and the resistance. The plotted graphs seen in Figure 19 shows the increase in
the resistance due to the increase in temperature. As the temperature rises, the resistance of
the wire increases increasing the voltage drop and eventually the resistance. The prediction of
𝑉
the relationship was reached from Ohm’s law ( 𝑅 = ) where it can be seen that R and V are
𝐼
directly proportional as I is constant. Through graphing the calculated resistance and the
standard resistance versus the temperature, the deviation of the graphs clearly represents the
linearity of the two and the error present between them. The range of percentage error of the
data was 0.94% to 1.65%, which is acceptable. This might have occurred due to several
points to be discussed later in this report.
NTC thermistor takes its name from the negative temperature coefficient of resistance. NTC
thermistor was used in the second run. Similar to PRT, constant current was used as 1
milliampere and the resistance was calculated using the Ohm’s law. However, since NTC
thermistor comprises of a semiconductor material, the resistance of the material decreases
with the increase of temperature. The found data supports the prediction as can be seen in
Figure 20. The graph also shows the non-linearity property of NTC thermistor. Through
graphing both calculated and standard values, the error present in the experiment can be seen
clearly. The error range in this case is 1.03% to 6.85%.
Also, K and J type thermocouples were used through the experiment. Since both
thermocouples are connected through the same port, the first run included the J type and the
second one included the K type. The data was collected for each and calculations done. At
first, the voltage was converted to the actual value in microvolts. Then, the standard voltages
at the reference temperature values were found. Through these data, the graph of the voltage
versus reference temperature was created. From the graphs in Figure 23 & Figure 21, the
29
linearity of the relationship can be clearly seen. Both K and J types have similar graphs.
However, an error can be seen in the both types thermocouple, which reached a range of
23.34% to 81.32% for the K type and 23.45% to 92.99% for the J type thermocouple. The
deviation is clearly present in the graphs and may be due to touching the tank walls, or wrong
wires connections missed.
The next device is the gas thermometer which was used in the second run. The gas
thermometer showed in Table 7 shows great results compared to the reference temperature
with a range of error of 0.13% to 1.89%. From the graph in Figure 24, the straight line shows
the accuracy of the values from the gas thermometer. It shows the reference and indicated
temperatures over the different trials and as clear, both line overlap one another showing the
perfect results found. A very slight percentage error can also be seen and is a result of several
mistakes. These errors will be discussed by the end of the discussion.
Similar to the gas thermometer, the bi-metallic thermometer was used to measure the
temperature in the first run. The temperature values found compared to the reference
temperature were almost accurate but with a higher percentage error of that of the gas
thermometer. The range of percentage error was found to be 0.20% to 19.37%. The errors
occurring while measuring with the bi-metallic can be similar to those of the gas thermometer
and will be discussed.
Figure 26 & 27 show the graphs and how the values are close to those of the reference. The
slight error can also be seen in the figures as the line deviate from the standard at the
beginning. The mistakes predicted will be discussed later as they may be shared with the
other devices.
The last device, to measure the temperature, was the liquid filled glass thermometer used in
the second run. It showed a high accuracy from the results found with a range of percentage
error of 0% to 4.12%. Figure 28 & 29 shows how the values are compared to the reference
each in a different way. From the figures, the accuracy of the device shows as the numbers
are close to the reference and the curves overlap one another in Figure 29. The errors still
occur of course and are explained below in the discussion.
Comparing the instruments results with one another, the accuracy can be compared through
the graphs found in the results. The least percentage error comes from PRT with least of 0%
While the highest percentage error is found from K type thermometer with upper range of
92.99%. As mentioned before, the results include errors; therefore, the conclusion about
which is most accurate is not final. The experiment also shows the operating difficulty of
each instrument. The liquid glass filled thermometer is considered easiest to operate and the
difficulty increase. Moreover, the similarities of the devices can be seen as well while
comparing them including how they operate, the main concept behind the measurement, and
the relations between results and actual values.
Away from the perfection of the theory, the actual work includes a various amount of errors
in any experiment. Temperature measurement is no exception. Each of the data collected
from the different devices included errors that are similar or specific. All these devices share
30
the errors occurring due to getting the values at the exact reference temperature, and the not
letting the device touch the sides of bottom of the tank. In addition, devices that required
visual observation such as the gas, bi-metallic, and liquid filled glass thermometers, include
the error of parallax. Moreover, the friction in the gas or bi-metallic thermometer which is
removed by tapping the glass is also a source of error. In addition, wires used for connections
can be the cause of errors. In the end, all the used instruments have a percentage error
depending on the instrumentation finite precision.
Conclusion
In conclusion, the objectives were reached through the experiment and the different
temperature measuring instruments were compared. Measuring tools used included platinum
resistance thermometer (PRT), negative temperature coefficient (NTC) thermistor, K and J
types thermocouples, gas and bi-metallic thermometers, and liquid filled glass thermometer.
Through the report, a full understanding of how to operate these instruments and how to get
the desired data was achieved. The results were interpreted, understood, and compared
through the discussion. The report reached results which include the linearity of all
instruments except for NTC, and the accuracy of PRT above all the other instruments.
Furthermore, the success of the experiment relies on the elimination of the errors explained
above. In addition, one of the ways to improve the results would be to use very cold water in
the heater before starting. Another way is to take accurate readings from the temperature
devices at the exact reference temperature.
31
Appendix
Table.1A: Standards for PRT– Resistance for Temperature
32
Table.3A: Standards for J Type thermocouple
33
References
[5]"Thermocouple-Thermocouples-What is a thermocouple-Types of
thermocouples", Thermocoupleinfo.com, 2017. [Online]. Available:
http://www.thermocoupleinfo.com/. [Accessed: 28- Sep- 2017].
[10]T. Agarwal, "6 Different types of Temperature Sensors with their Specifications", Buy
Electronics & Electrical Projects in India, 2017. [Online]. Available:
https://www.edgefx.in/6-different-types-of-temperature-sensors-with-their-specifications/.
[Accessed: 29- Sep- 2017].
34
[13]2017. [Online]. Available:
https://web.mst.edu/~cottrell/ME240/Resources/Temperature/Temperature.pdf. [Accessed:
01- Oct- 2017].
35