Académique Documents
Professionnel Documents
Culture Documents
CONTENTS
3- NGL Measurement
3.1Pressure Measurements
- Fluid Properties Relating to Pressure Measurement
and Calibration Principles.
- Primary Transducers and Pressure gauges
- Secondary Transducers and Transmitters
- Smart Transmitters
2
Chapter 1
Orifice Meter Station Fundamentals
The metering system shall measure flow rate and total quantity
and have access to gas quality parameters.
Filter/Separator
Filter to remove foreign particles and the separator to remove the
water.
Orifice meter
An orifice meter is a restriction element to create a pressure drop.
Flow Transmitter
The flow transmitter is an electronic device which converts the
variation in flow into a standard analogue signal 4 – 20 mA.
Temperature Transmitter
4
The temperature transmitter is an electronic device which convert
the variation in temperature into a standard analogue signal
4 –20 mA.
Pressure Transmitter
The pressure transmitter is an electronic device which converts the
variation in pressure into a standard analogue signal 4 – 20 mA.
Calorific Meter
Gas chromatography is normally employed for the calculation of
the heating value.
Flow Computer
The flow computer is an electronic device which receives the flow,
pressure and temperature analogue signals in addition to the data
of calorific meter and manipulate these variables to calculate:
- Gross volume
- Net volume
- Mass flow
5
Vena Contracta
The cross-sectional area of a fluid decreases as it is forced
through a restriction as shown in figure 1.2.
Figure 1.2
Vena-Contracta
6
As the fluid approaches the orifice the pressure increases slightly
and then drops suddenly as the orifice is passed. It continues to
drop until the “vena contracta” is reached and then gradually
increases until at approximately 5 to 8 diameters downstream a
maximum pressure point is reached that will be lower than the
pressure upstream of the orifice. The decrease in pressure as the
fluid passes thru the orifice is a result of the increased velocity of
the gas passing thru the reduced area of the orifice. When the
velocity decreases as the fluid leaves the orifice the pressure
increases and tends to return to its original level. All of the
pressure loss is not recovered because of friction and turbulence
losses in the stream. The pressure drop across the orifice (Fig.
1.3) increases when the rate of flow increases. When there is no
flow there is no differential. The differential pressure is proportional
to the square of the velocity, it therefore follows that if all other
factors remain constant, then the differential is proportional to the
square of the rate of flow.
Figure 1.3
Typical Orifice Flow Pattern Flange Taps Shown
7
a. Concentric Orifice Plate
The concentric orifice plate has a round opening (bore) in the
center of the plate (Fig. 1.4).
The plate should not depart from flatness along any diameter by
more than 0.01 inch per inch of the dam height (D-d)/2. To conform
with recommended practices, the orifice-to-pipe diameter ration
d/D (called Beta ratio), must not exceed recommended limits.
Figure 1.4
Concentric orifice plate
8
Figure 1.5
Eccentric orifice plate
9
Figure 1.6
Quadrant edge orifice plate
a. Flange Taps
These taps are located one inch from the upstream face of the
orifice plate and one inch from the downstream face with a + 1/64
to +1/32 tolerance.
b. Pipe Taps
These taps are located 2½ pipe diameters upstream and 8 pipe
diameters downstream (point of maximum pressure recovery).
d. Corner Taps
These taps are located immediately adjacent to the plate faces,
upstream and downstream. Corner taps are most widely used in
Europe, in line sizes less than 2 inches they are used with special
honed flow meter tubes for low flow rates.
10
Orifice meter types:
11
Figure 1.8 Senior Orifice Type
The Junior Fitting is much like the Senior, except the Junior does
not have a slide valve and a top chamber. The Junior fitting
requires only the following steps to remove an orifice plate from
the line:
1. Shut in meter tube.
2. Depressure tube.
3. Loosen set screws, remove top clamping bar, sealing bar, and
gasket.
4. Turn shaft, elevating orifice plate out of the fitting.
12
The procedure is reversed to install an orifice plate.
Meter Tubes
These lengths of pipe are known as meter tubes, meter runs, flow
sections, meter sections, etc.; however, the most generally
accepted terminology is meter tubes.
It has been clearly demonstrated by tests that the length and the
condition of the pipe used in meter tubes has a very important
bearing on the overall accuracy of the measurement.
13
14
15
Table (1)
3) Always slope the manifold lines gently from the orifice fitting to
the meter to eliminate any high or low points in the manifold lines.
PRESSURIZE
A) Be sure (1) & (2) are closed (if not, follow depressure
procedure).
B) Open (3), (4), (5), & (6) and close (7).
C) Slowly open (1) & (2).
16
D) Close (5) & (6) and open (7).
DEPRESSURIZE
A) Close (7).
B) Open (5) & (6).
C) Close (3)& (4) (or (1) & (2) if checking piping for leakage).
D) Slowly open (7) (this will evenly vent hi & low sides).
Pressure correction
Absolute pressure = gage pressure + 14.73 = psia
17
CHAPTER 2
Orifice Meter Station Operation / Maintenance
Velocity is the speed at which the fluid moves. This by itself does
not give any information about the quantity of fluid.
Q = v. A
Mass flow rate can only be calculated from the velocity or the
volumetric flow rates if the density is constant. If the density is not
constant, then mass flow measuring equipment is required for
mass flow rate.
W = Q. ρ
18
measurement is referred to as the inferred method of measuring
mass flow.
Base Conditions:
The conditions of temperature and pressure to which measured
volumes are to be corrected (Same as reference or standard
conditions).
For both liquid and gas applications, these base conditions can
change from one country to the next, from one state to the next, or
from one industry to the next. Therefore, it is necessary that the
base conditions be identified for standard volumetric flow
measurement.
Qv = 18.527*Cd*Ev*Y1*(d²)*[Tb/Pb]*[(Pf1*Zb*hw)/(Gr*Zf1*Tf)] 0.5
Where
Cd = Orifice plate coefficient of discharge
d = Orifice plate bore diameter calculated at flowing temperature
(Tf) - in.
Gr = Real gas relative density (specify gravity)
hw = Orifice differential pressure in inches of water at 60 degF
Ev = Velocity of approach factor
Pb = Base pressure - psia
Pf1 = Flowing pressure (upstream tap - psia
19
Qv = Standard volume flow rate - SCF/hr.
Tb = Base temperature - degR
Tf = Flowing temperature - degR
Y1 = Expansion factor (downstream tap)
Zb = Compressibility at base conditions (Pb,Tb)
Zf1 = Compressibility (upstream flowing conditions - Pf1, Tf)
20
Velocity of Approach Factor - Ev
21
A higher temperature means a less dense gas and higher flows,
but when this higher flow is corrected to base temperature, the
base flow is less.
Expansion Factor - Y1 or 2
The expansion factor corrects for the density change between the
measured tap density and the density at the plane of the orifice
face. Since the common static pressure tap used in natural gas
measurement is the downstream factor Y2; this factor is smaller
than the Y1 correction.
The real gases compress more than the ideal gas law predicts and
this must be corrected for when gas is measured at high pressure
and temperatures other than 60°F mathematically reduced to base
conditions. This correction, when applied outside of the square root
radical is called supercompressibility.
22
Critical Flow
The above square root flow formula applies to subsonic flow only.
Sonic or critical flow occurs when the velocity of the gas or vapor
reaches the speed of sound (approx. 700 miles per hour in air). A
gas cannot be made to travel any faster and remain in the same
state.
3- With the exception of the orifice meter, almost all flow meters
require a fluid flow calibration at flow and temperature conditions
closely approximating service operation in order to establish
accuracy.
BETA RATIO is the ratio of orifice plate bore divided by pipe I.D. is
referred to as the Beta Ratio or d/D where d is the plate bore and
D is the pipe I.D.
23
between 0.2 and 0.6 beta ratios. Below 1,000,000 Reynolds
number there will be some small increase in uncertainty with the
minimum Reynolds number of 4,000 being the limit of the
standard.
Rangeability
Repeatability
The ability of a flow meter to indicate the same readings each time
the same flow conditions exist. These readings may or may not be
accurate, but will repeat. This capability is important when a flow
meter is used for flow control.
Figure 2.2 Modern natural gas measuring systems may be quite complex,
and all elements must be considered in determining metering accuracy.
• The opening and closing of the metering system with gas flowing
through, shall always be done in manual mode (from the computer
part).
• The metering system shall measure the gas flow for different
operating conditions.
25
Figure 2.3 If a meter’s rangeability is not sufficient to cover the flow range
being measured, multiple transducers can be used or a different type meter
evaluated.
By far the most critical part of the meter is the orifice plate,
particularly the widely used square-edged concentric plate whose
construction requirements are well documented in standards such
as AGA-3 and ISO 5167-1. These standards define the plate's
edge, flatness, thickness with bevel details, if required and bore
limitations.
26
These simplify plate installation/removal for changing flow ranges
and for easy inspection.
In every case, the orifice must be installed concentric with the pipe
within limits stated by the standard.
Since this is not economical, almost all orifice systems are built to
meet the standard(s). This allows calculations to be made with
specified tolerances. Control over orifice metering accuracy
derives directly from data in the standard, which must be followed
without exception.
• Flow Computer.
• Flow proportional sampling system and/or On-line gas
chromatograph.
27
• Hourly automatic sampling of recorded and calculated
parameters of gas and its consumption; database maintenance.
• Warning signaling:
- High flow
- Low flow
- High pressure
- Low pressure
28
29
Figure 2.5 Remotely Mounted Differential-Pressure Flow meters for Gas
Service
The first step in testing any meter is a visual inspection for any
signs of improper operation, such as leakage and unstable flow.
The best throughput test is one that can be run directly in series
with a prover. The prover can come in many forms, but essentially
involves a basic volume that has been certified by the government
or industry group.
31
Gas provers are usually master meters with computer controls so
that testing requires little or no calculation or critical flow nozzles
(where accurate thermodynamic properties of the gas are
available). Critical nozzles require permanent pressure drops of
some 15 to 20% of the upstream static pressure and cannot be run
at static pressure below approximately 30 pounds per square inch
absolute.
Figure 2.6 A major problem with tests using charts—such as this one
comparing chart integration versus a flow computer—is that even though input
data are correct, data extraction by integration may introduce very large
errors.
32
a favorite target has been the maintenance costs along gas
pipelines.
Maintenance for all meters, particularly orifice and turbine for gas
measurement, has suffered. In order to avoid having to change
orifice plates to accommodate changing flow ranges, the largest
orifice bore acceptable to both parties is often installed along with
smart differential-pressure transmitters.
33
Figure 2.7 The orifice plates shown here range from “clean” (upper left) to so
dirty as to make measurement almost meaningless (lower right).
Figure 2.8 A ball hone such as this one may be required to remove certain
deposits in a meter tube.
36
This requires maintenance, since all pipelines contain liquids and
solids that will dirty meter tubes and necessitate cleaning. Meter
tubes for other types of meters should also be similarly inspected
and cleaned.
The first step is to inspect the meter tubes internal condition with
the fiber optic device. The meter tube must be blown down and
removed from service. Insertion of the inspection device requires a
tap 1/2 inch or larger in diameter. This may be a blow-off valve, a
tap hole, or a fitting slot.
The most critical parts of the meter tube to inspect are the 5 to 7
diameters immediately upstream of the meter and downstream of
straightening vanes or flow-conditioner plates (if used). Examine
for any liquid or solid deposits. Contrary to common belief, any
amount of deposit may cause measurements to be in error, usually
low. Evaluation of the necessity for cleaning can be made after
37
visual inspection, but cleaning is almost always the best choice. It
is simply better to err on cleaning than to take a chance on
measurement several tenths of a percent low.
The cleaning cycle should be repeated until the drain water runs
clear.
38
Figure 2.9 The orifice plate shown here in a gas measurement system clearly
shows that a “river of liquid” was flowing in the line. Needless to say, gas
measurement accuracy was far from accurate.
Plate Removal
Procedure:
1. To remove the orifice plate the operator must first balance the
pressure between the body chamber and the top chamber.
39
3. Wait several seconds while the top chamber pressure equalizes
with the actual line pressure.
4. Rotate the slide valve shaft (5) until it stops and OPEN is
indicated on the indicator plate.
5. Rotate the plate carrier shaft (6) located in the body chamber (4)
in a direction that will move the plate carrier out of the body
chamber (4) and into the upper gear shaft (7) located in the top
chamber.
6. Rotate the plate carrier shaft (7) located in the top chamber in a
direction that will lift the plate carrier completely out of the body
chamber and fully into the top chamber. Rotate shaft until the plate
carrier stops against the seal bar.
7. Once the plate carrier is completely out of the body chamber (4),
rotate the slide valve shaft (5) to the CLOSED position. This
rotation closes the slide valve thus separating and sealing the
body chamber (4) from the top chamber.
11. Once the fasteners (11) are loose, the operator should then
turn the plate carrier shaft (7) located in the top chamber in the
direction to raise the plate carrier and to tap the inside surface of
the sealing bar (9) to free both it and the gasket.
12. Slide the clamping bar (12) containing the fasteners (11),
sealing bar (9) and gasket (9A) from the top chamber.
13. Rotate the top chamber plate carrier shaft (7) to lift the plate
carrier out of the Senior Orifice Fitting.
40
CHAPTER 3
NGL Measurement
41
Figure 3.1 NGL fractionation train.
3.1Pressure Measurements
The Bourdon tube works on a simple principle that a bent tube will
change its shape when exposed to variations of internal and
external pressure. As pressure is applied internally, the tube
42
straightens and returns to its original form when the pressure is
released.
The tip of the tube moves with the internal pressure change and is
easily converted with a pointer onto a scale. A connector link is
used to transfer the tip movement to the geared movement sector.
The pointer is rotated through a toothed pinion by the geared
sector.
The main advantage with the Bourdon tube is that it has a wide
operating (depending on the tube material). This type of pressure
measurement can be used for positive or negative pressure
ranges, although the accuracy is impaired when in a vacuum.
Helix and spiral tubes are fabricated from tubing into shapes as
per their naming. With one end sealed, the pressure exerted on
the tube causes the tube to straighten out. The amount of
straightening or uncoiling is determined by the pressure applied.
43
These two approaches use the Bourdon principle. The uncoiling
part of the tube is mechanically linked to a pointer which indicates
the applied pressure on a scale.
This has the added advantage over the C-Bourdon tube as there
are no movement losses due to links and levers.
Bellows type sensors are also available which have the sensing
pressure on the outside and the atmospheric conditions within.
44
Figure 3.4 Basic mechanical structures
45
Figure 3.5 Spring and bellows gauge
3.1.2.5 Diaphragm
Many pressure sensors depend on the deflection of a diaphragm
for measurement.
Figure 3.6
Wheatstone circuit for strain gauges
Strain gauges are mainly used due to their small size and fast
response to load changes.
47
3.1.3.2 Piezoelectric
When pressure is applied to crystals, they are elastically deformed.
Piezoelectric pressure sensing involves the measurement of such
deformation. When a crystal is deformed, an electric charge is
generated for only a few seconds. The electrical signal is
proportional to the applied force.
Because these sensors can only measure for a short period, they
are not suitable for static pressure measurement.
More suitable measurements are made of dynamic pressures
caused from:
- shock
- vibration
- explosions
- pulsations
- engines
- compressors
This type of pressure sensing does not measure static pressure,
and as such requires some means of identifying the pressure
measured. As it measures dynamic pressure, the measurement
needs to be referenced to the initial conditions before the impact of
the pressure disturbance. The pressure can be expressed in
relative pressure units, Pascal RELATIVE.
Quartz is commonly used as the sensing crystal as it is
inexpensive, stable and insensitive to temperature variations.
Tourmaline is an alternative which gives faster response speeds,
typically in the order of microseconds.
3.1.3.3 Capacitance
Capacitive pressure measurement involves sensing the change in
capacitance that results from the movement of a diaphragm. The
sensor is energised electrically with a high frequency oscillator. As
the diaphragm is deflected due to pressure changes, the relative
capacitance is measured by a bridge circuit.
Two designs are quite common. The first is the two-plate design
and is configured to operate in the balanced or unbalanced mode.
The other is a single capacitor design.
The balanced mode is where the reference capacitor is varied to
give zero voltage on the output. The unbalanced mode requires
measuring the ratio of output to excitation voltage to determine
pressure.
This type of pressure measurement is quite accurate and has a
wide operating range.
Capacitive pressure measurement is also quite common for
determining the level in a tank or vessel.
48
Figure 3.7
Cross section of the Rosemount S-Cell™ Sensor
Figure 3.8
Capacitance pressure detector
Isolation Valves
49
Many pressure devices require tapping points into the process.
Isolation valves should be considered between the process fluid
and the measuring equipment if the device is required to be taken
out of service for replacement or calibration.
Sensor Construction
Depending on the environment in which the instrument is to be
used, selection of the correct sensor should also involve physical
conditions. The sensor may need to be isolated mechanically,
electronically and thermally from the process medium and the
external environment.
50
The smart transmitter utilizes capacitance sensor technology for
DP and GP measurements. Piezoresistive sensor technology is
utilized in the AP measurements.
The sensor electronics are installed within the sensor module and
include a temperature sensor (RTD), a memory module, and the
capacitance to digital signal converter (C/D converter). The
electrical signals from the sensor module are transmitted to the
output electronics in the electronics housing.
51
Figure 3.9 smart transmitter
52
53
3.2 Temperature Measurements
3.2.1 Temperature Scales
Temperature is a measure of the thermal energy in a body, which
is the relative hotness or coldness of a medium and is normally
measured in degrees using one of the following scales; Fahrenheit
(F), Celsius or Centigrade (C), Rankine (R), or Kelvin (K).
Kelvin scale (K) named after Lord Kelvin was proposed in the late
1800s. It is referenced to absolute zero but based on the Celsius
scale, i.e., a change of 1°C = a change of 1 K. The freezing and
boiling point of pure water are 273.15 K and 373.15 K,
respectively, at 1 atm, see Fig. 3.10. The degree symbol can be
dropped when using the Kelvin scale.
54
Figure 3.10 Comparison of temperature scales.
To convert °F to °C
°C = (°F − 32)5/9
To convert °C to °F
°F = (°C × 9/5) + 32
To convert °F to °R
°R = °F + 459.6
To convert °C to K
K = °C + 273.15
To convert K to °R
°R = 1.8 × K
To convert °R to K
K = 0.555 × °R
3.2.2 Thermowells
Temperature measurement transducers, in particular
thermocouples or RTD's; need different housings and mountings
depending on the application requirements.
Sensing devices are usually mounted in a sealed tube, more
commonly known as a thermowell; this has the added advantages
of allowing the removal or replacement of the sensing device
without opening up the process tank or piping. Thermowells need
to be considered when installing temperature-sensing equipment.
The length of the thermowell needs to be sized for the temperature
probe.
55
Consideration of the thermal response needs to be taken into
account. If a fast response is required, and the sensor probe
already has adequate protection, then a thermowell may impede
system performance and response time. Note that when a
thermowell is used, the response time is typically doubled.
Thermowells can provide added protection to the sensing
equipment, and can also assist in maintenance and period
calibration of equipment.
56
connection to the measuring device a second junction is formed,
called the reference or cold junction, this completes the
circuit.
Figure 3.12
Typical thermocouple and extension leads
Construction of RTD’s
RTD’s consist of a Platinum or Nickel wire element encased in a
protective housing having, in the case of the Platinum version a
base resistance of 100 Ohms at 0OC and the Nickel type a
resistance of 1000 ohms, again at 0OC.
They come packaged in either 2,3 or 4 wire versions, the 3 and 4
wire being the most common. 2 wire versions can be very
inaccurate as the lead resistance is in series with the measuring
circuit, and the measuring element relies on resistance change to
indicate the temperature change.
Thermistors
These elements are the most sensitive and fastest temperature
measuring devices in common use, unfortunately the price paid for
this is terrible nonlinearity (see Figure 2.6), and a very small
temperature range.
Thermistors are manufactured from metallic oxides, and have a
negative temperature coefficient, that is their resistance drops with
temperature rise.
They are also manufactured in almost any shape and size from a
pin head to discs up to 25 mm diameter x 5 mm thick.
I.C Sensors
Integrated Circuitry Sensors have only recently began to make
their presence felt in the process control world. As such they are
still limited in the variability of shape, size and packaging that is
advisable.
Their main advantages are their low cost (below $10.00) along
with their linear and high output signals.
57
Radiation Pyrometers
At the other end of the scale is the requirement to measure high
temperatures up to 4000OC or more. Total radiation pyrometers
operate by measuring the total amount of energy radiated by a hot
body. Their temperature range is 0OC to 3890OC.
The infrared (IR) pyrometer is rapidly replacing this older type of
measurement, and these work by measuring the dominant
wavelength radiated by a hot body. The basis of this is in the fact
that as temperature increases the dominant wavelength of hot
body radiation gets shorter.
SPDT switch
DUAL SWITCH(DPDT)
LCD Display
60
3.2.6 Thermocouple MV. vs. Temperature Table
50 0.002 0.003 0.003 0.003 0.004 0.004 0.004 0.005 0.005 0.006 0.006
60 0.006 0.007 0.007 0.008 0.008 0.009 0.009 0.010 0.010 0.011 0.011
70 0.011 0.012 0.012 0.013 0.014 0.014 0.015 0.015 0.016 0.017 0.017
80 0.017 0.018 0.019 0.020 0.020 0.021 0.022 0.022 0.023 0.024 0.025
90 0.025 0.026 0.026 0.027 0.028 0.029 0.030 0.031 0.031 0.032 0.033
100 0.033 0.034 0.035 0.036 0.037 0.038 0.039 0.040 0.041 0.042 0.043
110 0.043 0.044 0.045 0.046 0.047 0.048 0.049 0.050 0.051 0.052 0.053
120 0.053 0.055 0.056 0.057 0.058 0.059 0.060 0.062 0.063 0.064 0.065
130 0.065 0.066 0.068 0.069 0.070 0.072 0.073 0.074 0.075 0.077 0.078
140 0.078 0.079 0.081 0.082 0.084 0.085 0.086 0.088 0.089 0.091 0.092
150 0.092 0.094 0.095 0.096 0.098 0.099 0.101 0.102 0.104 0.106 0.107
160 0.107 0.109 0.110 0.112 0.113 0.115 0.117 0.118 0.120 0.122 0.123
170 0.123 0.125 0.127 0.128 0.130 0.132 0.134 0.135 0.137 0.139 0.141
180 0.141 0.142 0.144 0.146 0.148 0.150 0.151 0.153 0.155 0.157 0.159
190 0.159 0.161 0.163 0.165 0.166 0.168 0.170 0.172 0.174 0.176 0.178
200 0.178 0.180 0.182 0.184 0.186 0.188 0.190 0.192 0.195 0.197 0.199
210 0.199 0.201 0.203 0.205 0.207 0.209 0.212 0.214 0.216 0.218 0.220
220 0.220 0.222 0.225 0.227 0.229 0.231 0.234 0.236 0.238 0.241 0.243
230 0.243 0.245 0.248 0.250 0.252 0.255 0.257 0.259 0.262 0.264 0.267
240 0.267 0.269 0.271 0.274 0.276 0.279 0.281 0.284 0.286 0.289 0.291
250 0.291 0.294 0.296 0.299 0.301 0.304 0.307 0.309 0.312 0.314 0.317
260 0.317 0.320 0.322 0.325 0.328 0.330 0.333 0.336 0.338 0.341 0.344
270 0.344 0.347 0.349 0.352 0.355 0.358 0.360 0.363 0.366 0.369 0.372
280 0.372 0.375 0.377 0.380 0.383 0.386 0.389 0.392 0.395 0.398 0.401
290 0.401 0.404 0.407 0.410 0.413 0.416 0.419 0.422 0.425 0.428 0.431
300 0.431 0.434 0.437 0.440 0.443 0.446 0.449 0.452 0.455 0.458 0.462
310 0.462 0.465 0.468 0.471 0.474 0.478 0.481 0.484 0.487 0.490 0.494
320 0.494 0.497 0.500 0.503 0.507 0.510 0.513 0.517 0.520 0.523 0.527
330 0.527 0.530 0.533 0.537 0.540 0.544 0.547 0.550 0.554 0.557 0.561
340 0.561 0.564 0.568 0.571 0.575 0.578 0.582 0.585 0.589 0.592 0.596
350 0.596 0.599 0.603 0.607 0.610 0.614 0.617 0.621 0.625 0.628 0.632
360 0.632 0.636 0.639 0.643 0.647 0.650 0.654 0.658 0.662 0.665 0.669
370 0.669 0.673 0.677 0.680 0.684 0.688 0.692 0.696 0.700 0.703 0.707
380 0.707 0.711 0.715 0.719 0.723 0.727 0.731 0.735 0.738 0.742 0.746
390 0.746 0.750 0.754 0.758 0.762 0.766 0.770 0.774 0.778 0.782 0.787
400 0.787 0.791 0.795 0.799 0.803 0.807 0.811 0.815 0.819 0.824 0.828
410 0.828 0.832 0.836 0.840 0.844 0.849 0.853 0.857 0.861 0.866 0.870
420 0.870 0.874 0.878 0.883 0.887 0.891 0.896 0.900 0.904 0.909 0.913
430 0.913 0.917 0.922 0.926 0.930 0.935 0.939 0.944 0.948 0.953 0.957
440 0.957 0.961 0.966 0.970 0.975 0.979 0.984 0.988 0.993 0.997 1.002
450 1.002 1.007 1.011 1.016 1.020 1.025 1.030 1.034 1.039 1.043 1.048
460 1.048 1.053 1.057 1.062 1.067 1.071 1.076 1.081 1.086 1.090 1.095
470 1.095 1.100 1.105 1.109 1.114 1.119 1.124 1.129 1.133 1.138 1.143
480 1.143 1.148 1.153 1.158 1.163 1.167 1.172 1.177 1.182 1.187 1.192
490 1.192 1.197 1.202 1.207 1.212 1.217 1.222 1.227 1.232 1.237 1.242
°C 0 1 2 3 4 5 6 7 8 9 10
61
3.2.7 RTD Temperature vs. Resistance Tables
62
3.3 Level Measurements
3.3.1 Level Measurement Methods
Continuous Measurement
The units of level are generally meters (m). However, there are
numerous ways to measure level that require different
technologies and various units of measurement.
Such means may be:
- Ultrasonic, transit time
- Pulse echo
- Pulse radar
- Pressure, hydrostatic
- Weight, strain gauge
- Conductivity
- Capacitive
For continuous measurement, the level is detected and converted
into a signal that is proportional to the level. Microprocessor based
devices can indicate level or volume.
Point Detection
Point detection can also be provided for all liquids and solids.
Some of the more common types are:
- Capacitive
- Microwave
- Radioactive
- Vibration
- Conductive
63
A level measuring system often consists of the sensor and a
separate signal conditioning instrument. This combination is often
chosen when multiple outputs (continuous and switched) are
required and parameters may need to be altered.
Figure 3.16
Level by Visual Inspection
64
Figure 3.16 Gauging rod
65
Tensioning of the tape storage reel is sufficient to ensure correct
measurement, while not affecting the position of the float.
The shaft on the counter drive rotates as the float moves the
perforated tape up and down. The rotary motion of the shaft is
used to give a metric readout.
In atmospheric conditions, a seal is used to protect the sensing
head from the process fluid. However in pressurised applications,
it is better to fill the head with the sensing fluid, particularly if the
fluid is clean and lubricating.
Figure 3.17
Wire-guided float detectors and details of head
Hydrostatic Pressure
Some of the different types of level measurement with pressure
are:
- Static pressure
- Differential pressure
- Bubble tube method
- Diaphragm Box
- Weighing
P = h.ρ.g
66
where: P = pressure
h = height
ρ = relative density of fluid
g = acceleration due to gravity
For constant density, the only variable that changes is the height.
In fact, any instrument that can measure pressure can be
calibrated to read height of a given liquid, and can be used to
measure liquid level in vessels under atmospheric conditions.
Most pressure sensors compensate for atmospheric conditions, so
the pressure on the surface of liquids open to the atmosphere will
be zero. The measuring units are generally in Pascals, but note
that 1 Pa is equivalent to 1 m head of water.
Hydrostatic pressure transducers always consist of a membrane
which is connected either mechanically or hydraulically to a
transducer element. The transducer element can be based on
such technologies as inductance, capacitance, strain gauge or
even semiconductor.
Figure 3.18
A pressure gauge used to measure the height of a liquid in an open tank
Differential
When the surface pressure on the liquid is greater (as may be the
case of a pressurised tank) or different to the atmospheric
pressure, then a differential pressure sensor is required. This is
because the total pressure will be greater than the head of liquid
67
pressure. With the differential pressure sensor, the pressure on the
surface of the liquid will be subtracted from the total pressure,
resulting in a measurement of the pressure due to the height of the
liquid.
Figure 3.19
For a pressurised tank, the level is measured using differential pressure
methods
Ultrasonic Measurement
Ultrasonic level sensors work by sending sound waves in the
direction of the level and measuring the time taken for the sound
wave to be returned. As the speed of sound is known, the transit
time is measured and the distance can be calculated.
68
Figure 3.20
Ultrasonic measurement
Radar Measurement
Radar gauges differ from ultrasonic in that they use microwaves
instead of sound waves. Like ultrasonic devices they measure
from the top of the vessel to determine the product level.
69
Figure 3.20
Radar differential pressure hybrid system
70
Measurement Principle:
Figure 3.21
Switch position on rising level
71
A non-magnetic barrier tube (5) provides a static pressure
boundary between the switch mechanism and the displacer
assembly. As the liquid level falls, the displacer lowers, causing the
spring to extend, and moving the attraction sleeve out of the
magnetic field of the switch mechanism.
2.2 Mounting
Adjust the displacers on the displacer cable for the desired switch
actuating levels (instruction tag is attached to cable). Screw
displacer cable fitting to threade connection link protruding from
the underside of control.
2.3 Wiring
3.1 What to do
3.1.1 Keep control clean
Be sure the switch housing cover is always in place on the control.
This cover is designed to keep dust and dirt from interfering with
switch mechanism operation. It protects against damaging
moisture and acts as a safety feature by keeping bare wires and
terminals from being exposed.
73
3.1.2 Inspect switch mechanisms, terminals, and connections
monthly
1. Dry contact switches should be inspected for excessive wear on
actuating lever or misalignment of adjustment screw at point of
contact between screw and lever. Such wear can cause false
switch actuating levels. See switch mechanism bulletin supplied
with control should switch adjustment or replacement be
necessary.
NOTE: Check wiring carefully and replace at the first sign of brittle
insulation.
74
only tend to attract dust and dirt which can interfere with
mechanism operation.
75
CHAPTER 4
Electronic Flow Measurements
Table 4.1
Main types of liquid flow
Figure 4.1
Flow profile
76
smaller currents flow in all directions - these are also known as
eddies. This type of flow has a flatter flow profile, such that the
velocity of forward flow in the centre of the pipe is nearly the same
as that near the walls of the pipe.
Swirl
Swirl occurs with laminar flows as fluid passes through elbows or
some other form of pipeline geometry. In a similar fashion to
turbulent flows, they affect the measurements of many instruments
and precautions should be taken to mount measuring devices well
downstream from the swirling fluid. The effects of swirls can
be minimised by the use of a flow conditioner, or straightener in the
line upstream.
Reynolds number
A Reynolds number defines the flow conditions at a particular
point. It is a way of representing fluidity and is a useful indicator of
laminar and turbulent flow.
Laminar flow exists if the Reynolds number is less than 2000, and
turbulence when the number is above 4000. There is not a clear
transition between laminar and turbulent flows, which does
complicate flow measurement in this range of operation.
RD = (K x ρ x v x D) / µcp
77
Figure 4-2 Cylinder and piston show principle of displacement metering
Figure 4.3
Rotating vane meter
Typical Applications
This type of meter is used extensively in the petroleum industry for
liquids.
Advantages
- Reasonable accuracy of 0.1%.
- Suitable for high temperature service, up to 180 oC.
- Pressures up to 7Mpa.
Disadvantages
- Suitable for clean liquids only.
78
Lobed impeller
This type of meter uses two lobed impellers, which are geared and
meshed to rotate at opposite directions within the enclosure. A
known volume of fluid is transferred for each revolution.
Figure 4.4
Rotating lobe meter
Typical Applications
The lobed impeller meter is often used with gases.
Advantages
- High operating pressures, up to 8Mpa.
- High temperatures, up to 200oC.
Disadvantages
- Poor accuracy at low flow rates.
- Bulky and heavy.
- Expensive.
Figure 4.5
Positive displacement meters measure volumetric flow rate directly by dividing
a stream into distinct segments of known volume, counting segments and
multiplying by the volume of each segment.
79
The viscosity of the fluid can affect the leakage, or slip flow. If the
meter is calibrated on a particular fluid, it will read marginally
higher should the viscosity rise.
Advantages
- High accuracy of 0.25%
- High operating pressures, up to 10MPa
- High temperatures, up to 300oC
- Wide range of materials of construction
Disadvantages
- Pulsations caused by alternate drive action
Application Limitations
Over speeding can damage positive displacement meters. When
the outlet pressure needs to be maintained due to an unacceptable
pressure drop across the meter, it is quite common for the inlet
pressure to be increased.
Filters may be required to filter debris and clean the fluid before
the meter. Regular maintenance is an obvious addition in this
case. The added pressure drop may also need to be considered,
especially if regular maintenance is not carried out.
80
Positive displacement meters become limited when high volume
measurement is required. They are primarily used for low volume
applications.
Advantages
- Can measure non-conductive fluids.
- Very high accuracy.
- Unaffected by viscosity.
- High Rangeability of up to 10:1.
Disadvantages
- Clean fluids only, limited life due to wear.
- Some unrecoverable pressure loss.
- Requires viscous fluid, not suitable for gas.
- Limited operating range.
- Mechanical failure likely to cause blockage in pipe.
- Cost.
81
The velocity of flow is related to the square root of the differential
pressure.
Formulae
The relationship between the flow rate and the change in pressure
can be shown as:
Primary Elements
4.2.1 Orifice Plate
For technical details, refer to section 1
82
Due to the relatively high cost of the Venturi Tube, applications are
generally limited to high flow rate fluids, such as main steam lines.
Figure 4.6
Flow measurement with a restriction
Advantages
- Less significant pressure drop across restriction.
- Less unrecoverable pressure loss.
- Requires less straight pipe up and downstream.
Disadvantages
- More expensive.
- Bulky - requires large section for installation.
83
These devices are more cost effective, but as such they provide
less accuracy than venturis, and have a higher unrecoverable
pressure loss.
The advantages and disadvantages highlight any differences from
the venturi tube and are over and above those listed for that
device.
Figure 4.7
The flow nozzle is preferable for high velocity applications
Advantages
- High velocity applications.
- Operate in higher turbulence.
- Used with fluids containing suspended solids.
- More cost effective than venturis.
- Physically smaller than the venturi.
Disadvantages
- More expensive than orifice plates.
- Higher unrecoverable pressure loss.
Advantages
- Shorter lay length.
- Lower unrecoverable pressure loss.
Disadvantages
- More complex to manufacture.
- Sensitive to turbulence.
- Accuracy based on flow data.
84
Figure 4.8
The Dall tube low-loss meter
Application Limitations
It is recommended that when using this type of tube, it should be
calibrated with the piping section in which it is to be used and over
the full range of flows in which it will operate.
These types of flow devices do not necessarily conform to the wide
range of conditions and extensive test data for orifice meters.
Figure 4.9
Principle of operation of the Pitot tube
Figure 4.10
Basic form of pitot-static tube
Advantages
- Low cost.
- Low permanent pressure loss.
- Ease of installation into existing systems.
Disadvantages
- Low accuracy.
- Low Rangeability.
- Requires clean liquid, gas or vapour as holes are easily
clogged.
86
4.2.6 Multiport Pitot Averaging
An annubar tube is a multi-impact opening type and improves the
accuracy of this type of measurement. In most industrial
applications this type of multi-impact or averaging is used to
compensate for changes in the velocity profile.
Annubar sensor are inserted perpendicular to the flow stream and
extend the full diameter of the pipe. There is a very low obstruction
to the flow, which causes minimal pressure loss.
Sensing ports are located on both upstream and downstream
sides of the Annubar.
These ports are connected to dual averaging plenums. The
number of ports is proportional to the diameter of the pipe. The
upstream ports produce an average impact (total) pressure and
the downstream ports produce the average reference (static)
pressure. The difference is an accurate and stable dynamic
pressure that is easily converted into flow rate.
Annubars also provide good measurement when located in difficult
piping. They can be located as close as two pipe diameters
downstream of an elbow and still give accurate and repeatable
measurements.
Figure 4.11
Multiport ‘Annubar’ Pitot averaging system
4.2.7 Elbow
A pipe elbow can be used as a primary device. Elbow taps have an
advantage in that most piping systems have elbows that can be
used. In applications where cost is a factor and additional pressure
loss from an orifice plate is not permitted, the elbow meter is a
viable differential pressure device.
87
If an existing elbow is used then no additional pressure drop
occurs and the expense involved is minimal. They can also be
produced in-situ from an existing bend, and are typically formed by
two tappings drilled at an angle of 45o through the bend.
These tappings provide the high and low pressure tapping points
respectively.
Tappings at 22.5o have shown to provide more stable and reliable
readings and are less affected by upstream piping. However 45o
tappings are more suited to bidirectional flow measurement.
Figure 4.12
Elbow meter geometry
88
be performed for more accurate results, with the added advantage
that repeatability is good for this type of measurement.
Figure 4.13
Flow in a pipe band
Advantages
- Simplified installation.
- Inexpensive.
Disadvantages
- Low accuracy.
Advantages
89
- No moving parts.
- Large range of sizes and opening ratios.
- Suitable for most gases and liquids.
- Well understood and proven.
- Price does not increase dramatically with size.
Disadvantages
- Accuracy is affected by density, pressure and viscosity
fluctuations.
- Erosion and physical damage to the restriction affects
measurement accuracy.
- Cause some unrecoverable pressure loss.
- Viscosity limits measuring range.
- Require straight pipe runs for expected accuracies.
Application Limitations
- Limited accuracy in measurement.
- Low Rangeability of 4:1.
Pressure loss:
When obstructing the flow using the differential pressure method of
sensing, note that they do cause some unrecoverable pressure
loss in the line.
Disturbances:
Differential type instruments that cause a restriction are easily
affected by disturbances to inflow and outflow. Apart from bends in
the pipe, such fittings as T pieces and valves can cause
disturbances to the flow measurement and as such should be
separated from the instrument by a significantly long straight pipe
extension. The length of clean pipe should be greater than 5 x
diameter on the upstream side and no less than 4 x diameter on
the downstream side.
4.2.10 Troubleshooting
One of the most common inaccuracies induced in differential
pressure flowmeters is not allowing enough straight pipe. When
the flow material approaches and passes some change in the pipe
small eddies are formed in the flow stream. These eddies are
90
localised regions of high velocity and low pressure and can start to
form upstream of the change and dissipate further downstream.
Flowmeter sensors detect these changes in pressure and
consequently produce erratic or inaccurate readings for flow rate.
4.3 Velocity Flow Meters
4.3.1 Turbine
Basis of Operation
Figure 4.14
Turbine flow meter
They are not recommended for use with high viscosity fluids due to
the high friction of the fluid which causes excessive losses as the
92
turbine becomes too much of an obstruction. The viscosity of the
liquid must be known for use of this type of meter.
They are also subject to erosion and damage. Each meter must be
calibrated for its application.
K-Factor
93
Figure 4.15
Typical calibration curve for a turbine meter
Advantages
- High accuracy, repeatability and Rangeability for a defined
viscosity andmeasuring range.
- Temperature range of fluid measurement: -220oC to
+350oC.
- Very high-pressure capability: 9300psi.
- Measurement of non-conductive liquids.
- Capability of heating measuring device..
- Suitable for very low flow rates.
Disadvantages
- Not suitable for high viscous fluids.
- Viscosity must be known.
- 10 diameter upstream and 5 diameter downstream of
straight pipe is required.
- Not effective with swirling fluids.
- Only suitable for clean liquids and gases.
- Pipe system must not vibrate.
- Specifications critical for measuring range and viscosity.
Application Limitations
As turbine meters rely on the flow, they do absorb some pressure
from the flow to propel the turbine. The pressure drop is typically
around 20 to 30 kPa at the maximum flow rate and does vary
depending on flow rate.
Summary
Turbine meters provide excellent accuracy, repeatability and
rangeability for a defined viscosity and measuring range, and are
commonly used for custody transfer applications of clean liquids
and gases.
Vortex flowmeters can measure liquid, gas or steam and use the
principle of vortex shedding. Vortex shedding occurs when an
obstruction is placed in the flowing stream. The obstruction is
referred to as a bluff body and causes the formation of swirls,
called vortices, downstream from the body.
Figure 4.15
Principle of operation of vortex meter
95
flow, a series of vortices generates electrical pulses with a
frequency that is also proportional to the flow velocity. The velocity
can then be converted to volumetric flow rate.
Figure 4.16
96
Relationship between K-Factor and Reynolds number
4.3.3 Swirl
Figure 4.17
Construction of a typical vortex precession (swirl) meter
4.3.4 Electromagnetic
97
Faradays law states that moving a conductive material at right
angles through a magnetic field induces a voltage proportional to
the velocity of the conductive material. The conductive material in
the case of a magmeter is the conductive fluid.
The fluid therefore must be electrically conductive, but not
magnetic.
98
Figure 4.17
Principle of the electromagnetic meter
In general the pipeline must be full, although with the later models,
level sensing takes this factor into account when calculating a flow
rate.
This takes into account both the meter itself and the secondary
instrument. Because of its linearity, the accuracy at low flow rates
exceeds that of such devices as the Venturi tube. The magnetic
flowmeter can be calibrated to an accuracy of 0.5% of full scale
and is linear throughout.
The lining of the meter piping must also be lined with an insulating
material to prevent short-circuiting of the electric field.
100
For correct operation of the magmeter, the pipeline must be full.
This is generally done by maintaining sufficient backpressure from
downstream piping and equipment. Meters are available that make
allowances for this problem, but are more expensive and are
specialised. This is mainly a problem in gravity feed systems.
Magmeters are not greatly affected by the profile of the flow, and
are not affected by viscosity or the consistency of the liquid. It is
however recommended that the meter be installed with 5
diameters of straight pipe upstream and 3 diameters of straight
pipe downstream from the meter.
5.6.8 Advantages
- No restrictions to flow.
- No pressure loss.
- No moving parts.
- Good resistance to erosion.
- Independent of viscosity, density, pressure and turbulence.
- Good accuracy.
- Bi-directional.
- Large range of flow rates and diameters.
5.6.9 Disadvantages
- Expensive.
101
- Most require a full pipeline.
- Limited to conductive liquids.
102
Figure 4.18
Transit time measurement
103
FIGURE 4.19
Minimum straight run requirements for a 1% accuracy of a single-path transit
time flowmeter
104
5.8.4 Typical Applications
Transit-time ultrasonic flow measurement is suited for clean fluids.
Some of the more common process fluids consist of water,
liquefied gases and natural gas.
Figure 4.20
Doppler effect
105
erosion or viscosity. Most ultrasonic flowmeters are bi-directional,
and sense flow in either direction.
5.8.8 Advantages
- Suitable for large diameter pipes.
- No obstructions, no pressure loss.
- No moving parts, long operating life.
- Fast response.
- Installed on existing installations.
- Not affected by fluid properties.
5.8.9 Disadvantages
- Accuracy is dependent on flow profile.
- Fluid must be acoustically transparent.
- Errors cause by build up in pipe.
- Only possible in limited applications.
- Expensive.
- Pipeline must be full.
5.8.11 Summary
Doppler flowmeters are not high accuracy or high performance
devices, but do offer an inexpensive form of flow monitoring. Their
intended operation is for dirty fluids and find applications in
sewage, sludge and waste water processes.
106
4.4 Mass Flow Meters
Although most meters can infer mass flow rate from volumetric
flow measurements, there are a number of ways to measure mass
flow directly:
- The Coriolis meter
- Thermal mass flowmeter
Figure 4.21
Principle of Coriolis effect
107
The Coriolis effect causes a retarding force on a rotating section of
pipe when flow is moving outward, conversely producing an
advance on the section of pipe for flow moving towards the axis of
rotation.
Figure 4.22
Application of Coriolis force to a meter
Coriolis Meter
The force is applied to oscillate the flow pipes and the Coriolis
effect is the principle used to determine the acceleration due to the
torque (the amount of twisting).
Sensors are used to measure the amount of twist in the flow tubes
within the meter as a result of the flow tube vibration and deflection
due to the mass flow. The amount of twist measured is proportional
to the mass flow rate and is measured by magnetic
sensors mounted on the tubes.
108
Figure 4.23
Coriolis meter construction
Advantages
- Direct, in-line mass flow measurement.
- Independent of temperature, pressure, density, conductivity
and viscosity.
- Sensor capable of transmitting mass flow, density and
temperature information.
- High density capability.
- Conductivity independent.
- Suitable for hydrocarbon measurements.
- Suitable for density measurement.
Disadvantages
- Cost.
- Affected by vibration.
- Installation costs.
- Adjustment of zero point.
109
Application Limitations
- High temperature.
- Vibration.
- Amount of gas in fluid.
- Restricted to low flow rates.
- Limited to pipe sizes of up to 150mm.
Summary
Coriolis meters provide direct, in-line and accurate mass flow
measurements that are independent of temperature, pressure,
viscosity and density. Mass flow, density and temperature can be
accessed from the one sensor. They can also be used for almost
any application when calibrated.
The two main types of thermal mass flow measuring devices are:
- Thermal Anemometer
- Temperature rise flowmeter
Thermal anemometer
110
Figure 4.24
Thermal mass flowmeters
These devices are primarily used for gases and are dependent on
flow profile. They are therefore limited if the flow profile changes as
they only measure at one point in the flow stream. Their limitations
are similar to pitot tubes, which are also an insertion device into a
flow stream.
The temperature probe must protrude into the flow stream, and
therefore may be easily damaged by corrosion and erosion. In
addition the robustness of the system is compromised by the
protrusions into the fluid stream, increasing the chances of
leakage.
111
Advantages
- Fast response times, < 0.5milliseconds.
Disadvantages
- Require 10 diameters of straight pipe upstream.
- Have similar limitations to pitot tubes.
Figure 4.25
Principle of ‘Temperature Rise’ method
As with the hot wire probe, the temperature sensors and the heater
must protrude into the flow stream, and therefore may be easily
damaged by corrosion and erosion. Also the robustness of the
system is compromised by the protrusions into the fluid stream,
increasing the chances of leakage.
Disadvantages
- Suitable for low gas flows only.
- Subject to erosion and corrosion.
- More tapping points, increased chances of leakage.
112
Developments of the insertion type of sensing have moved the
heating and sensing elements to outside the pipe to overcome the
problems with tapping points. By limiting the tapping points, the
chance of leakage (and associated maintenance) is greatly
reduced, if not eliminated.
Figure 4.26
Thermal flowmeter with external elements and heater
Advantages
- Non contact, non intrusive sensing
- No obstruction to flow
- Reduced maintenance
Disadvantages
- Suitable for low gas flows only
- Subject to erosion and corrosion
Summary
Thermal mass flowmeters are mainly used to measure the flow of
clean gases with known heat capacities. They are commonly used
in the refining and chemical industries.
Description
113
The uses a thermal differential technique to sense changes in
the heat transfer characteristics of a media. Figure 1 shows the
outline of the flow switch.
Flow Switching
When the sensor is inserted into a liquid or gas the heated RTD is
strongly affected by the velocity of the medium. Flow past the
114
heated RTD changes the heat transferred from the surface of the
sensor. This cooling effect reduces the temperature of the sensor.
The flow switch compares this change to a preset flow trip point to
switch the output. Figure 2 shows the flow switch signal change vs.
flow rate for air, light hydrocarbon liquids, and water.
The signal change vs velocity has the same general shape for all
three media but the change is larger for air and the sensitive range
is different for each.
For air and most gaseous media the range is 0.1 to 500 feet per
second (FPS).
For most liquid media the flow switch range is 0.01 to 5 FPS.
115
MAXIMUM
VELOCITY (FT/SEC) RANGE
116
Figure 4.30 Differential Pressure Transmitter
Operation
117
CHAPTER 5
Orientation of Instrumentation and Controls
ELECTRONICS
Several stages of development have taken place in the move to
electronics.
Figure 5-1 Typical temperature transducer. Devices such as this, along with
computers and other equipment, are vital parts of a metering system.
118
Simple computers may calculate a flow rate and totalize flow for a
meter. Or computers can be operating centers for measurement,
control, and communications in complex multimeter systems.
Computers can develop the complete volume calculation and print
appropriate hard copy or feed a central control or computer center
with the complete accounting procedures.
They can provide real-time operation and control information for
metering systems.
RELATED DEVICES
Most related devices have an accuracy in the general range of
±0.5 to ±1% of full scale. This makes it important to choose
transducers with the right ranges; measurements should be in the
upper two-thirds of device ranges. The higher the differential the
more accurate the meter reading, provided flow does not exceed
the differential device's range. Many users, not recognizing the
effects auxiliary instrumentation can have on the accuracy
obtained by a given meter, compromise flow accuracy accordingly.
Overall accuracy obtained includes the inaccuracies of each of the
auxiliary devices and how they are used, as well as the primary
meter accuracy. Sometimes the auxiliary devices may control the
119
accuracy of the results more than the basic meter. For example,
with gases (particularly near their critical points), a change of one
pound can represent several percent in flow. This says two things:
1) it would be better to measure at some other location where
conditions are farther away from critical points; or 2) the accuracy
required for pressure measurement must be increased several fold
to maintain the same limits for the corrected flow measurement.
122
devices must be calibrated against manometers or dead weights.
They are usually tested and calibrated at atmospheric pressure
and then rezeroed at line pressures. Some test devices that
operate at line pressures are available, but their use is often
restricted to laboratory work rather than field calibration.
For natural gas, the definitions in AGA-3 are the weight per unit
volume of gas compared to the weight per unit volume of air at the
same conditions of pressure and temperature. This definition of
123
.real specific gravity. ignores the corrections for compressibility
when these weighings are made at atmospheric pressures, since
such corrections are relatively small. However, this yields a
specific gravity different by a small amount from the ratio of
molecular weights (which is equal to the ideal specific gravity).
δf= (SG)(Ww,)
Figure 5-3 Recording gas gravitometer using the indirect weighing method.
124
can happen when variable components make up a sample with the
same specific gravity.
125
Figure 5-4 Typical gas sampling system.
126
points; (4) effects of sample-system material that comes into
contact with the sample; (5) effects of the type of gas to be
sampled on the choice of the process; (6) cleaning required; and
(7) care of the sample after it is obtained.
When this work is completed the committee will write a much more
complete standard on the problems of getting a good sample so
that the flow measurement will be improved. This will cover flows
as simple to sample as methane-rich streams that have been
dehydrated and cleaned to such complex samples as multiphase
flows. The value of this work is the empirical data that will be taken
to prove the decisions made some of which may require totally
new and different approaches.
Some progress reports show that root causes of gas-sample
distortion were found to include poor sampling systems and
procedures, surface effects, thermal dynamic problems, purge
problems, and leaks.
127
down because of shrinkage (such as mixtures of light hydrocarbon
liquids or two-phase gas flows).
CALORIMETRY
Where heating value is needed, a calorimeter can be installed to
continuously monitor a stream. Or samples can be taken at meter
locations and individual samples tested at a centrally located
calorimeter. After all inputs have entered a pipeline, a single unit is
often used to determine the heating value at all downstream
locations. The choice between chromatograph and densitometer
depends on product value (quantity and cost) and the contractual
requirement for corrections (i.e., most require correction for heat
value, while some only require that a minimum heat value be
maintained).
128
The Flow Meters for Liquids & Gases measure flow based on a
pressure differential created across a built-in calibrated nozzle.
The meter is self-contained and complete.
5.2.1 FLOW
Common devices for flow measurement fall into the following
categories:
a. Differential-head meters measure flow inferentially from the
differential pressure caused by flow through a primary element.
Flow is proportional to the square root of the differential pressure
produced. This differential is sensed by diaphragms, bellows, or
manometers.
ACCESSIBILITY
All locally mounted flow instruments should be readily accessible
from grade, platforms, fixed walkways, or fixed ladders. A rolling
platform may be used where free access is available to the space
below the instruments.
LOCAL INDICATION
Where local indication is desired and nonindicating transmitters
are used, output indicators should be provided. In applications
where flow can be manually controlled at a control-valve station,
flow indication should be clearly visible and readable from the
valve location to permit manual control when necessary. This flow
gauge should not be used to calibrate the transmitter.
VIBRATION
Most instruments are susceptible to damage, abnormal
wear, or malfunction if mounted in a location where they are
subject to vibration. If any part of the flow system or equipment
130
is subject to vibration, the affected instruments should
be provided with vibration-free supports.
PULSATION
Measurement of pulsating flow is difficult and should be avoided.
Head-type flowmeters and instruments with mechanical
movements, such as positive-displacement meters and turbines,
should not be used in pulsating-flow applications.
The measurement is not dependable, and the pulsing may
contribute to premature wear of the mechanical components.
PIPING
Process connections to the instruments should be furnished and
installed in accordance with applicable piping and material
specifications. All pipe should be deburred after cutting and blown
clean of cuttings and other foreign material before assembly. As an
alternative to pipe, tubing of suitable material may be used.
131
installation, ability to be bidirectional, very low pressure loss,
minimum requirement for upstream piping, and use of one type of
transmitter regardless of pipe size.
132
or on platforms with adequate support.
Installation
Installation of differential-pressure flow devices is generally
requires consideration of the factors described in 5.2.1.3.1 and
5.2.1.3.2.
5.2.1.4.1 Installation
A variable-area meter should be installed at a location that
is free from vibration and has sufficient clearance for occasional
float removal for service or inspection. The meter
should be readable and readily accessible for operation and
maintenance. In general, when a meter is to be used in regulating
service, it should be placed as close as possible upstream of the
throttling point. Variable-area meters must always be mounted
vertically, with the outlet connection at the top of the meter and the
inlet connection at the bottom.
133
avoid any strain on the meter body.
134
and other conductive liquids. A magnetic flowmeter consists of two
parts-a primary element, installed directly in the process line, and a
secondary element, the electronic transmitter. The meter
generates a signal proportional to the rate of flow.
5.2.1.5.1 Installation
Considerable care must be exercised when the magnetic
flowmeter's primary element is installed in the pipeline. Special
care must be taken to prevent damage to the liner and to ensure
that grounding requirements are met. The manufacturer's
installation recommendations should be followed, including
consideration of upstream and downstream piping requirements.
The transmitter is built on a rugged piece of pipe, but it should be
handled as a precision instrument.
The transmitter should be accessible from grade or from a platform
with enough space around it to permit removal of at least the top
housing if necessary. Sufficient access should be available for
removal of any inspection plates.
The magnetic flow transmitter tube may be installed in any position
(vertical, horizontal, or at an angle), but it must run full of liquid to
ensure accurate measurement. If the tube is mounted vertically,
flow should be from bottom to top to ensure that the pipe is full. If
the tube is mounted horizontally, the electrode's axis should not be
in a vertical plane. A small chain of bubbles moving along the top
of the flow line can prevent the top electrode from contacting the
liquid. Vertical mounting with a straight run on the inlet side and
upward flow is recommended if an abrasive slurry is being
measured. This arrangement distributes wear more evenly.
5.2.1.6.1 Installation
Turbine meters are installed directly in the process line. The line
135
should be relatively free from vibration. Meters with integrally
mounted, direct-reading registers should be positioned so that they
can be easily read and maintained.
Turbine meters are normally installed in horizontal lines but may be
installed in vertical up flow lines. It is necessary to specify the
position in which the meter is to be calibrated. Calibration for the
installed position is required.
5.2.1.6.2 Piping
The accuracy and repeatability of measurements from turbine
meters depend on the upstream and downstream piping. In
addition to sufficiently long straight runs upstream and
downstream, straightening vanes are required for high accuracy.
5.2.1.6.3 Strainers
All turbine meter installations should have strainers to prevent
damage to the meter rotor. The strainer must be capable of
removing particles of a size that might damage the rotor and
bearings. The strainer should be located upstream of the required
meter run.
5.2.1.7.1 Installation
Positive-displacement meters are installed directly in the process
piping and can be a source of vibration. Adequate foundations
should be provided (refer to the manufacturer’s recommendations).
Positive-displacement meters are normally installed in horizontal
lines. Certain types are specifically designed for vertical lines.
Meters should be installed so that the meter case or body is not
subject to piping strain. The piping should be arranged so that the
meter is always full of liquid.
Adequate back pressure may be required to eliminate the
possibility of vapor release.
For continuous process services, a bypass around a positive-
displacement meter is recommended. For custody transfer,
bypasses are not permitted. Positive-displacement meters should
136
always be installed with an adequate strainer to prevent foreign
matter from damaging the meter or causing excessive wear; the
manufacturer’s recommendation on mesh size should be
observed. Where excessive amounts of debris are entrained in the
fluid, strainer pressure drop should be monitored.
The installation of a positive-displacement meter should be
designed to avoid air or vapor in the piping. Where the design
does not allow for this, air eliminators should be considered.
Air eliminators can leak or have inadequate capacity to protect the
meter from slugs of air or vapor; such eliminators should be
removed and replaced.
5.2.1.8.1 Installation
Vortex meters are installed directly in the process piping
and are normally supported by the piping. They may be installed in
any orientation. A vortex meter should be in stalled so that the
meter body is not subjected to piping strain. In liquid applications,
the piping should be arranged so that the meter is kept full.
Block and bypass valves should be provided when operating
conditions do not permit shutdown.
137
5.2.2 LEVEL
The general application of the commonly used level instruments
are devices for indicating, recording, and controlling the liquid
levels.
A wide variety of level instrumentation is currently available.
Selection and proper installation depends on a number of
variables, such as (a) the type of vessel, fluid, or material involved
(namely, solids, granules, liquids, or a liquid-liquid or liquid-foam
interface), (b) process conditions (namely, pressure, temperature,
specific gravity, boiling point, viscosity, and pour point), (c) what
the instrument is to accomplish (monitoring, on-off or modulating
control, or alarm), and (d) whether the signal is to be electronic or
pneumatic.
INTRODUCTION
Certain general procedures, practices, and precautions apply
to practically all of the instruments discussed in this section.
Where applicable, the material discussed in 3.2.2 through 3.2.9
should be considered a part of each of the subsequent
discussions.
ACCESSIBILITY
All locally mounted liquid level instruments, including gauge
glasses, should be readily accessible from grade, platforms, fixed
walkways, or fixed ladders. For maintenance purposes, rolling
platforms are frequently used when free access is available in the
area below the instruments.
READABILITY
In all applications in which a liquid level is regulated bay control
valve, some indication of the level a gauge glass, receiver
138
pressure gauges, or another indicator-should be clearly readable
from operating location.
VIBRATION
Some level instruments are susceptible to damage or malfunction
if they are subjected to vibration. To minimize vibration effects,
such instruments should be mounted on a rigid support adjacent
but not connected to the source of vibration.
Such an arrangement requires flexible tubing or conduit
connections between the source of vibration and the instrument.
Additionally, shockproof mounts may be considered.
Instruments should be carefully selected, since some instruments
are less susceptible to vibration effects.
139
Locally mounted indicating devices include armored gauge
glasses, magnetic gauges, and differential-pressure level
indicators.
141
Figure 5.8 External-Cage Displacement Instrument
142
Figure 5.9 Typical Stilling Well
Installation
Ultrasonic units should not be installed in areas with strong
electrical fields (motors, relays, electric generators, and so forth).
143
5.2.2.7 CAPACITANC LEVEL TRANSMlTTER
Operation
A capacitor consists of two conductive plates separated by an
insulator. Its capacitance is a function of the area of the plates, the
spacing between them, and the dielectric constant of the insulator.
Installation
The sensing element must be vertical and must not be in contact
with the vessel wall or internals. Applications in which both the
container walls and the medium are nonconductive may require a
counter electrode (ground reference) made from a conductive
material. The need for and type of ground reference should be
reviewed with the manufacturer.
The user should review the design of the sensing element's seal
for fire safety and should review the design of electrical and
electronic circuitry to ensure that it meets explosion proof or
intrinsic safety requirements or both.
145
Figure 5.11 Vessel's liquid level
GENERAL
The basic considerations for instruments used to initiate high- or
low-level alarm signals.
146
Figure 5.12 Arrangement of High- and Low-Level Alarm Switches With
Parallel Gauge Glass
147
5.2.3 PRESSURE
The instruments covered are pressure gauges and switches,
pressure transmitters, and locally mounted controllers and
recorders.
APPLICATION PRACTICE
Hydrocarbons or other process fluids that may be hazardous
or otherwise undesirable in the control room in the event of
leakage should not be piped to any instruments located in a central
control room. It is industry practice to transmit the pressure of such
fluids either electrically or pneumatically to receiving instruments.
ACCESSIBILITY
All locally mounted pressure instruments should be readily
accessible from grade, platforms, fixed walkways, or fixed ladders.
LOCAL INDICATION
148
Where local indication is desired and nonindicating (electronic
and pneumatic) transmitters, switches, and locally mounted
pressure controls are used, these instruments should be
supplemented with directly connected process pressure gauges
(see Figure 19), output indicators, or both.
VIBRATION
If any part of the pressure system or equipment is subject to
vibration, the instrument should be mounted on a vibration-free
remote support.
Coiled tubing, armored hose, or a capillary system should be
provided between the pressure source and the instrument.
PULSATION
Instruments that measure the pulsating pressures of reciprocating
pumps and compressors should be equipped with pulsation
dampeners to prevent premature failure of the movements or the
pressure elements.
Figure 20 indicating pressure gauges with liquid-filled cases should
also be considered for pulsating-service applications .
149
5.2.3.2 Pressure Transmitters
CONNECTIONS
The process connection for pressure transmitter is generally
3/4 inch in size, with the first block valve conforming to applicable
process piping specifications. The most common size for
instrument connections is 1/2 inch.
INSTALLATION CONSIDERATIONS
The installation of a pressure transmitter requires careful weighing
of a variety of factors. It is important to know the physical
characteristics and operating conditions of the process fluid.
Following are some guidelines (see Figure 24):
a. Impulse piping should be as short as possible.
b. Larger, heavier transmitters should be supported by means
other than the process connection.
c. Placement of taps on the bottom of the line should be avoided
because of the possible presence of sediment or scale.
d. Transmitters in liquid or condensable-vapor service such as
steam should be self-venting (that is, mounted below the process
connection, with all lines sloping toward the instrument) to prevent
gas from being trapped in the instrument.
e. Transmitters in gas service should be self-draining (that is,
mounted above the process connection, with all lines sloping
toward the process connection) to prevent liquid from being
trapped in the instrument.
f. The installation must protect the transmitter from both ambient
and process temperatures. If the process temperature is outside
the transmitter's limits, the following measures can be used to
ensure that the temperature at the transmitter is within the
manufacturer's specifications:
l. Providing a sufficient length of uninsulated piping to lower or
raise the temperature of the process fluid at the transmitter.
2. Purging the transmitter. When purging, piping of sufficient
diameter to minimize friction effects should be used.
3. Using a diaphragm seal and capillary to transmit pressure
to the transmitter.
150
Figure 5.15 Typical Installation of Pressure Transmitters for Gas, Liquid, and
Steam Service
5.2.4 TEMPERATURE
The installation of devices which measure and display temperature
in refinery process services and to do the following:
a. Display the temperature at the point of measurement.
b. Use the temperature for local control of the process
variable.
c. Transmit the temperature to a remote location for
indication, recording, alarm, and/or control at that point.
5.2.4.1 Thermowells
Direct exposure of temperature-sensing devices to process fluids
is usually impractical. Thermowells (see Figures 5.16) are
employed in temperature measurement to protect thermal
elements.
INSERTION LENGTH
The insertion length, U (see Figure 5.16), is the distance from the
free end of the temperature-sensing element or well, up to but not
including the external threads or other means of attachment to a
vessel or pipe.
151
IMMERSION LENGTH
The immersion length is the distance from the free end of the
temperature-sensing element or well to the point of immersion in
the medium whose temperature is being measured.
MATERIALS
The materials selected for thermowells must be suitable for the
temperature and corrosion environment encountered.
For general services in which carbon steel piping is normally used,
the minimum quality material usually specified is Type 304 or Type
3 16 stainless steel.
CONSTRUCTION
Typical thermowell construction and installation details are shown
in Figures 5.16A, B & C. Thermowells may be screw mounted, as
shown in Figure 5.16.
152
Figure 5.16 thermowell installation
153
Table 1
Installation
Thermocouples are generally installed in thermowells to minimize
temperature lag (response time), the thermocouple must be in
contact with the bottom of the well.
Thermocouple Transmitters
Thermocouple transmitters can be mounted in the thermocouple
head or in the control room. Two types of field mounted
transmitters are available. One is mounted directly on the
thermowell, and one is mounted remotely from the thermowell.
APPLICATIONS
The use of filled-system devices is limited by the capillary tubing
that maybe employed and the maximum temperature to which the
bulb may be exposed. Systems with compensation are built to self-
adjust for changes in temperature either of the case or of the
capillary and case. This self-adjustment assures accurate
measurement of the temperature where the bulb is located.
Dimensional, functional, and physical characteristics vary
depending on the manufacturer.
INSTALLATION GUIDELINES
In all installations of filled-system temperature instruments, it is
necessary to protect the bulb and capillary tubing from mechanical
damage. It is usually desirable to use armored capillary tubing and
to support the tubing run between the bulb and the controller or
transmitter to protect it from accidental damage. The capillary
tubing should not be cut, opened, or pinched in any manner. For
safety purposes, the vent hole on the top of the bulb packing gland
should be free from obstructions.
155
normally, it means that the process variable is within the high and
low operating limits.
156
As shown in Figure 5.18, a typical alarm system informs the
operator in the control room of abnormal process conditions in the
field. In this case, we see that the level in the process vessel has
As you can see, the alarm system performs a very simple and
important function. It is basically an ON-OFF control circuit that
uses a limit-sensing device connected to a warning device.
Depending on the process variable, the limit sensing device will
be:
A pressure switch
A float-operated switch
A temperature switch
A flow-actuated switch
157
and off when the process goes off-limit. By reading the location for
the off-limit condition, an operator will know what corrective action
to take.
Figure 5.19 Visible alarms
158
In process operations, the alarm system is often called an
annunciation. The simplest annunciation sequence is shown in
Figure 1-4. In this sequence, the abnormal process condition trips
the limit switch. This in turn activates the annunciation system with
its audible and visible alarms. In some cases the alarms will stay
on until the process condition returns to normal. In other cases, the
alarms can be
turned off by the operator and reset for the next off-limit condition.
160
ACTIVATE
SHUTDOWN
PROCEDURES
Troubleshooting Aids
We must ask our self this question “Why is the discharge pressure
not normal on the chemical pump discharge” to prompt a
discussion.
163
CHAPTER 6
Instruments and Controls
Automatic Control
To relieve our operator from the tedious task of manual control,
we automate the controls - i.e. we install a PID controller (Figure
6.1b). The controller has a Set Point (SP) that the operator can
adjust to the desired temperature.
164
So, when everything is up and running, our PID controller
compares the process variable to its set point and then calculates
the difference between the two signals, also called the Error (E).
Then, based on the error, a few adjustable settings and its internal
structure (described below), the controller calculates an output that
positions the control valve. If the actual temperature is above its
set point, the controller will reduce the valve position and vice
versa.
1. Sensing element
2. Measuring element
3. Controller element
4. Final control element
165
Figure 6.1c Control loop
The block diagram in Figure 6.1c shows the basic instruments in a typical
process control loop. The sensor responds to changes in the process
variable and sends this information to a transmitter. The transmitter
changes the process information to a standard pneumatic or electronic
signal and sends it to a controller. The controller uses the signal from the
transmitter to compute the corrective action to be taken by the final
control device. The indicator provides the process operators with a
continuous indication of the process conditions.
166
• Describe the differences between a reverse and a direct acting
controller;
• Indicate what deadtime is and how it impacts on a process.
167
Use of Open Loop Control
This control is often based on measured disturbances to the inputs
to the system. The most common type of open loop control is
feedforward control.
168
Figure 6.4 Concept of Feedback Control
169
Figure 6.5 Direct and Reverse Acting Controllers
170
Proportional Control
This is the main and principal method of control. It calculates a
control action proportional to the ERROR (ERR). Proportional
control cannot eliminate the ERROR completely.
PB = 100%
Kc
As long as the flow out of the tank remains constant, the level
(which is our process variable in this case) will remain at its set
point.
172
But, if the operator should increase the flow out of the tank, the
tank level will begin to decrease due to the imbalance between
inflow and outflow. While the tank level decreases the error
increases and our proportional controller increases the controller
output proportional to this error (Figure 6.9). Consequently, the
valve controlling the flow into the tank opens wider and more water
flows into the tank.
With our P-controller the offset will remain until the operator
manually applies a bias to the controller's output to remove the
offset. It is said that the operator has to manually "Reset" the
controller. Or we can add Integral action to our controller.
173
Figure 6.10 Integral Control Action
Most controllers use integral time in minutes as the unit for integral
control, but some others use integral time in seconds, Integral
Gain in Repeats / Minute or Repeats / Second. Table 2 compares
the different integral units.
177
Cascade Control
of Heat Exchanger
F
T Prim
I
I
S ary
C RS P
C
P
F T
T T
Heat
Exchan
ger Retur
Proc n
ess
178
Figure 6.15 Pneumatic Actuator: Air-to-Close/Spring-to-Open
179
anywhere between fully open and fully closed in response to
changes in supply air pressure.
A positioner is a device that regulates the supply air pressure to a
pneumatic actuator. It does this by comparing the actuator’s
demanded position with the control valve’s actual position. The
demanded position is transmitted by a pneumatic or electrical
control signal from a controller to the positioner. The pneumatic
actuator in Figure 6.15 is shown in Figure 6.16 with a controller
and positioner added.
180
Figure 6.16 Pneumatic Actuator with Controller and Positioner
181
the actuator in Figure 36 will fail open. This type of arrangement is
referred to as "air-to-close, spring-to-open" or simply "fail-open."
Some valves fail in the closed position. This type of
actuator is referred to as "air-to-open, spring-to-close" or "fail-
closed." This "fail-safe" concept is an important consideration in
nuclear facility design.
183
The coil is connected to an external current supply. The spring
rests on the armature to force it downward. The armature moves
vertically inside the coil and transmits its motion through the stem
to the valve.
When current flows through the coil, a magnetic field forms around
the coil. The magnetic field attracts the armature toward the center
of the coil. As the armature moves upward, the spring collapses
and the valve opens. When the circuit is opened and current stops
flowing to the coil, the magnetic field collapses. This allows the
spring to expand and shut the valve.
A major advantage of solenoid actuators is their quick operation.
Also, they are much easier to install than pneumatic or hydraulic
actuators. However, solenoid actuators have two disadvantages.
First, they have only two positions: fully open and fully closed.
Second, they don’t produce much force, so they usually only
operate relatively small valves.
The motor moves the stem through the gear assembly. The motor
reverses its rotation to either open or close the valve. The clutch
and clutch lever disconnects the electric motor from the gear
assembly and allows the valve to be operated manually with the
handwheel.
184
Most electric motor actuators are equipped with limit switches,
torque limiters, or both. Limit switches de-energize the electric
motor when the valve has reached a specific position. Torque
limiters de-energize the electric motor when the amount of turning
force has reached a specified value. The turning force normally is
greatest when the valve reaches the fully open or fully closed
position. This feature can also prevent damage to the actuator or
valve if the valve binds in an intermediate position.
185
To step through the list headers, press the Page button .
You will eventually return to the initial Instrument list.
To step through the parameters within a list, press the Scroll
Button . When you reach the end of the list you will return to the
list header.
From within a list, you can return to the list header at any time by
pressing the Page button .
Parameter names
For an explanation of each parameter, see the configuration
parameter tables.
186
CONFIGURATION PARAMETER TABLES
187
188
189
*In place of the dashes, the last three characters indicate the alarm
type. If an alarm is not configured the displayed name will differ:
e.g. for the first alarm AL 1’ will be shown.
190
*In place of the dashes, the last three characters indicate the alarm
type. If an alarm is not configured the displayed name will differ:
e.g. for the first alarm ‘AL 1’ will be shown
191
6.5 Periodic Calibration of the Controllers
Equipment needed:
1. Calibrated millivolt source & Ice bath or a calibrated
internally compensated thermocouple reference
2. Calibrated gage
3. Calibration record card
4. Calibration sticker
Temperature Verification:
1. Place controller and test instrument on a flat surface.
2. Connect units to appropriate power source.
3. Turn on power to each unit and allow 30 minutes for
warming.
4. Connect the test instrument to the controller.
5. If using a millivolt source, input appropriate mV levels from
table 1, otherwise input the temperatures from table 1 one at
a time. A minimum of three temperature points should be
tested.
6. Record and compare the temperature values.
7. Establish stable level.
8. Compare level with calibrated standard.
9. Repeat at 3 point’s minimum across the scale of the gage.
10. Record and compare the levels.
192
Controller Evaluation
11. Determine if the controller meets required accuracy (see
notes 1&2).
12. If controller does not meet accuracy requirements,
proceed to Calibration Procedure below.
13. If controller does meet accuracy requirements, complete
and attach a calibration sticker.
14. File record of calibration.
15. Return controller to service.
Table 1
Millivolt Temp.
0 32°F
1.942 100°F
4.906 200°F
7.947 300°F
11.023 400°F
14.108 500°F
17.186 600°F
20.253 700°F
23.317 800°F
26.396 900°F
29.484 999°F
193