Vous êtes sur la page 1sur 12

Kidney International, Vol. 57 (2000), pp.

1152–1163

Accuracy of hemodialysis modeling


M Z´ ,J A. P , and J G -C
ARIUSZ IOŁKO ACEK IETRZYK OANNA RABSKA HRZA˛STOWSKA

Department of Pediatric Nephrology, Polish-American Children’s Hospital; and Faculty of Electrical Engineering, Automatics,
Computer Science, and Electronics, University of Mining and Metallurgy, Krako´w, Poland

Accuracy of hemodialysis modeling. from patient weight are the simplest coefficients for mathemati-cal
Background. One- and two-compartmental models of he- models and have sufficient precision as well. The global value of
modialysis (HD) are well known. These models make it possi-ble to both compartments is slightly greater than the corre-sponding value
analyze the course of treatment and to predict the effect of dialysis for a one-compartmental model. The effective-ness of dialyzers is
procedures. Mathematical modeling helps physicians to match in practice lower than might be expected on the basis of the data
dialysis therapy to the individual needs of the patient; however, the provided by their manufacturers. Urea cellular clearance is two
efficiency of the models depends on the accuracy of the coefficients. times greater than creatinine and uric acid cellular clearances. The
How to select coefficients in the case of one-compartmental models clearance differences are more prominent for the cellular membrane
is known for urea and creatinine. Less information is available for than for artificial semiper-meable membranes.
two-compartmental models. Results on modeling of uric acid
concentrations have not been published.
Methods. The identification of the mathematical model coef-
Mathematical modeling for hemodialysis aids the phy-
ficients was based on the concentration measurements of three
markers of uremic toxicity (urea, creatinine, and uric acid) in both sician in providing an accurate dialysis prescription as well
patients’ blood and dialysate. Blood samples were taken from the as predicting the course of the individual’s renal disease.
arterial line several times throughout the dialysis period. Although the two-compartmental models are more effective
Simultaneously, dialysate samples were taken from a test port in the than one-compartmental models in de-termining the urea,
dialyzer outflow line. The mathematical model parameters were
determined so as to minimize the deviations between the measured
creatinine and uric acid concentra-tions, they are much more
points and the calculated curves. In this way, distribution volumes, complicated to use in the dialysis setting.
cellular clearances, and dialyzer mass transfer coefficients were
estimated. Studies were carried out on six chronically dialyzed
Results. For a one-compartmental model, the median value of patients in the course of three regularly performed he-
distribution volume V _ 0.56 DW was obtained, where DW is the
patient’s dry weight. For a two-compartmental model, modialysis (HD) sessions for each patient. The dialyses were
intercompartment volume Vi _ 0.36 DW and extracompart-ment performed at Dialysis Unit, Department of Pediat-ric
volume Ve _ 0.21 DW. The following median values for cellular Nephrology, Jagiellonian University, Krakow, Po-land.
clearances were established: urea 415 (mL/min), creatinine 207 Braun Secura hemodialysis machines were used, and low-
(mL/min), and uric acid 257 (mL/min). flux polysulfone Fresenius F5 and F6 and Baxter CA-110
Conclusions. One- and two-compartmental models describe the
concentration of the urea, creatinine, and uric acid very effectively, dialyzers were employed. During each HD ses-sion,
in contrast with phosphorus, in which modeling results are not concentrations of four components (urea, creati-nine, uric
satisfactory. Although two-compartmental mod-els are more acid, and phosphorus) were determined on an average of 11
effective, they are much more complicated than one-compartmental times at 20-minute intervals by the use of the automated dry
models, which justifies using the one-com-partmental model for method (Kodak Ectachem 700 XR analyzer). About 1600
hemodialysis modeling. A two-compart-mental model must be used
in the case of rebound phenomenon modeling. The total body water determinations of toxin concen-trations in blood and
values we have obtained are similar to the anthropometrically based dialysate were done. All the pa-tients gave their informed
values for urea and creatinine and to a lesser degree for uric acid. consent for their participation in the studies carried out in the
Distribution volumes for one- and two-compartmental models course of regular HD sessions.
obtained
Computer software was prepared to calculate the opti-mal
Key words: compartment models, distribution volume, clearance, dial-ysis coefficients for the mathematical models. The values of the
modeling, urea clearance. distribution volume for three commonly measured uremic
Received for publication January 21, 1999 and in
toxins (urea, creatinine, and uric acid) were com-puted and
revised form August 26, 1999 Accepted for compared with volumes obtained via five an-
publication September 16, 1999 thropometrically based methods for total body water
2000 by the International Society of Nephrology assignment. The usefulness of these methods is discussed

1152
Zio´łko et al: Hemodialysis modeling 1153

and analyzed according to the results derived from the uted volume V to experimental data, correction for mean
experimental data. ultrafiltration was done automatically.
Experimental data were also used to establish cellular Equation 1 has a simple interpretation. The rate of toxin
clearances for three toxins: urea, creatinine, and uric acid. concentration (left hand side of Equation 1) is a decreasing
Additionally, actual clearances of the dialyzers em-ployed (minus sign in the right hand side of (Equa-tion 1) function
were studied. The results were compared with the data of time t and is proportional to the time variable toxin
supplied by the manufacturers. We found that actual concentration (Ce), proportional to constant clearance (Kd )
clearance values were lower than those reported by the and inversely proportional to constant distribution volume (V
manufacturers. ). The solution of Equa-tion 1 has the form:
The study was undertaken with the following aims: (1 ) to
estimate the mathematical model coefficients by fitting
Ce(t) _ C0e_tKd/V (Eq. 2)
theoretical curves to the measurements of urea, creatinine,
and uric acid concentrations in blood and dialysate; (2 ) to where e < 2.72 is the base of the natural logarithm. Equation
compare one-compartmental curve fits with two- 2 describes the decrease of toxin concentration during the
compartmental curve fits; (3 ) to verify the mathematical hemodialysis treatment. Coefficient C0 is an initial toxin
formulae describing dialyzers; and (4 ) to compare the concentration.
practical effectiveness of dialyzers for three toxins with the Volume distribution (V ) must be known in order to
effectiveness expected on the basis of the manufacturers’ calculate the allowable dialyzer clearance (Kd ) or the
data. optimal dialysis time. From Equation 2:

T_ V ln C0 (Eq. 3)
METHODS Kd CT
Models of hemodialysis therapy where T is the HD duration; CT is the toxin concentration
Wolf and his collaborators were the first who described value that should be arrived at as a result of HD; and ln is the
dialysis kinetics and dialyzer clearance almost 50 years ago natural logarithm.
[1]. Renkin was also a pioneer in the mathematical Equation 3 is used in the calculation of dialysis dura-tion
description of dialysis [2], while Sargent and Gotch suc- (T ) necessary to decrease the toxin concentration level from
cessfully introduced the one-compartmental model to clinical the value C0 to CT if the toxin distribution area in the patient’s
practice in the late seventies [3, 4]. The benefits and risks of body is V, and the dialyzer clearance Kd. Equation 3 could
this model application have been described and discussed in be helpful in clinical practice if the data (V, Kd and C0) are
many clinical and theoretical articles [5–13]. not encumbered with significant errors. The dependence of
relative deviations _T/T of HD duration on the relative
deviations of distribution volume _V/V, dialyzer clearance
Figure 1A shows the flows of any uremic toxin during
dialysis treatment according to the one-compartmental _Kd/Kd and toxin con-centrations _C0/Co and _CT/CT has the
form:
model. All body fluids and plasma water are considered as
T V Kd C C C
one volume of distribution [3, 4]. The mathematical model _ _ _ _ _ _ 1_ 0 _ _ T2 ln 0
for this case has the form of an ordinary differen-tial T V Kd C0 CT CT
equation:
(Eq. 4)
dCe _ _ Kd Ce (Eq. 1) The minus indicates the descending character of the
dt V dependence (for example, between dialysis time and dia-
where Ce is the toxin concentration; t is the time; Kd is the lyzer clearance). In clinical practice ln(C0/CT) < 1, be-cause
dialyzer clearance, and V is the compartment volume. toxin concentration values decrease by approxi-mately 2.7
It was assumed that the change of volume (V ) during HD times as a result of hemodialysis. It follows from Equation 4
had little influence on the modeling efficiency and could be that the deviations of the four argu-ments on the right hand
neglected. Another assumption was that urea generation and side of Equation 3 have a similar influence over the value of
the residual urea removal by the kidneys were very low in deviation _T/T of the dialysis duration. The V volume can be
comparison with the dialyzer clearance (Kd). The established relatively easily and precisely enough on the
mathematical model given by (Eq. 1) took only the most basis of anthropomet-ric data, mainly the patient’s body
important phenomena into consideration. The less important mass. These methods will be presented in later in this article.
phenomena sometimes cancelled each other out, and The clearance Kd can be determined on the basis of the
sometimes their effects could be represented by the information provided by the manufacturer of the dialyzer.
coefficients that characterize domi-nant phenomena. For Usually this is the theoretical value resulting from in vitro
example, in adjusting the distrib- experi-
1154 Zio´łko et al: Hemodialysis modeling

Fig. 1. Dialysis toxin flow according to a (A) one-


compartmental model and (B) two-com-
partmental model.

ments that are overestimated in comparison to in vivo or


actually measured values. Our observations and com-ments dCi _ Kc(Ce _ Ci)
in this area are also presented below. The initial C0 toxin dt Vi
concentration can be determined from a sample of the where Ce is the toxin concentration in blood; Ci is the toxin
patient’s blood immediately before a dialysis pro-cedure. concentration in interior water; Ve is the extracom-partment
The measurement must be precise. Thus, one has to use a volume; Vi is the intracompartment volume, and Kc is the
high level analyzer and an independent analysis of two blood intercompartment mass transfer coefficient.
samples is recommended. The toxin transfer between both compartments is de-
The dialyzer clearance scribed by the product of a volumetric mass transfer
C coefficient Kc multiplied by the difference between toxin
Kd _ V ln 0 (Eq. 5) concentrations (Ci _ Ce). The rate of toxin concentra-tion in
T CT a particular compartment depends also on its volume Ve or
calculation will be appropriate in relation to the planned HD Vi. The rate of toxin removal by dialyzer is (similarly to the
procedure. In this case, the HD duration T has to be assumed. one-compartmental model) a linear function of concentration
In the case of a two-compartmental model, it is as-sumed Ce with the proportionality coefficient equal to the ratio
that body fluids are divided into two parts: those directly and Kd/Ve. The model defined by Equation 6 may be used for
quantitative modeling of toxin concentration changes if all
those indirectly accessible to the dialyzer (Fig. 1B). The drop
constants (Ve, Vi, Kc, and Kd) and variables (Ce and Ci) have
of toxin level in the blood, which is directly accessible, is
been precisely interpreted clinically. Therefore, Ce is
faster because of its direct contact with the dialyzer
regarded as the toxin concentration value in the bloodstream
membrane. The consequence of toxin removal from this pool
flowing into the dialyzer, and it can be measured. The
is a gradual solute transfer from the interior body water. The variable Ci is a toxin concentration value with no precise
toxin concentration in two-compartmental model can be localization within the patient’s body, and that is why it
described by a set of two differential equations: cannot be measured.

dCe _ _Kc(Ci _ Cd) _ KdCe When HD starts (t _ 0), toxin levels are equal within all the
(Eq. 6)
body compartments:
dt Ve Ve
Ce(0) _ Ci(0) _ C0 (Eq. 7)
Zio´łko et al: Hemodialysis modeling 1155

and can be measured in the patient’s blood. This assump-tion Assuming that KA is known, the estimation of toxin
is justified from a practical viewpoint and it consti-tutes the concentration in blood Ce can be calculated from the
initial conditions for Equation 6. For the Ce variable, the measured dialysate toxin concentration Cd:
solution of Equation 6 with the initial condi-tions given by
Cd(1 _ _Qd/Qb)
Equation 7 is a time dependent function: if Qb ? Qd
Ce(t) _ 0.5C0 e _ta
1 [(1 _ a3)e _ta
2
ta
_ (1 _ a3)e 2]
(Eq. 8)
Ce _ 5 1__
Cd(1 _ Qd/KA)if Qb _ Qd
(Eq. 11)
where
where KA(Qd_Qb)

_ _ e QbQd
a1 _ Kc _ Kd _ Kc
2Ve 2Vi
Identification of model parameters
KdKc
a2 _ !a 21 _ The method of model parameter identification pre-sented
ViVe in this article is based on toxin concentrations
˜

Ce(ti) measured in blood and Ce(ti) calculated for blood from


a3 _ ViKd _ Kc(Vi _ Ve) . the toxin concentration measured in dialysate. The
2a2ViVe identification problem consists in finding the coefficient
˜

V, Ve, Vi, and Kc if Ce(ti) and Ce(ti) are known, where i _ 1,


Dialyzer clearance identification
. . . ,I are numbers of samples. These coefficients of the
Assigned dialysis clearance (Kd) measurements are based mathematical models given by Equation 1 and Equation 6
on data provided by manufacturers. Information is usually were found by computer minimization of the quantity
given in the form of tables or figures, and is mainly limited criterion:
to “in vitro” Kd values for a given blood flow (that is, 200 or I ˜ 2
1 Ce(ti) _ Ce(ti)
300 mL/min). The rate of solute removal can also be
calculated by applying the formula: Q_ ! Io i 0
1 2 _ Ce(ti) 2.
eKA/Qb _ eKA/Qd (Eq. 12)
QbQd if Qb ? Qd
QdeKA/Qb _ QbeKA/Qd The quantity criterion (Eq. 12) is the mean value of the
Kd _ ˜

KAQb differences between the toxin concentrations Ce and Ce


5KA _ Qb if Qb _ Qd obtained from measuring results and the values for Ce
calculated from Equation 2 or Equation 8, respec-tively, at a
(Eq. 9)
discrete time denoted by ti.
where Qb is the blood flow rate (mL/min); Qd is the dialysate The volume distributions (V, Ve, Vi) are the key pa-
flow rate (mL/min); and KA is the overall mass transfer rameters in this case of mathematical modeling. In our case,
coefficient (mL/min). these parameters were determined by minimizing Equation
The coefficient KA is a measure of the dialyzer’s intrin-sic 12, but in clinical practice they ought to be estimated from
capacity to transfer toxin from the blood to the dialy-sate anthropometric data. One of the aims of this article was to
compartment. The value of coefficient KA is some-times compare these two methods. Five different formulae for the
given by a dialyzer’s manufacturer and it should be generally distribution volume are re-cently used. The popular Watson
constant for a given toxin and a given dia-lyzer. In clinical formulae using age, height and weight in men and height and
practice, for an individual patient-dia-lyzer system, the KA weight in women are as follows [14]:
coefficient can be identified by applying the formula:
V _ 2.477 _ 0.09516A _ 0.1074H _ 0.3362DW
˜
male
5
QbQd Qb(Ce _ Cd)
Vfemale _ _2.097 _ 0.1069H _ 0.2466DW
if Qd ? Qd
Qd _ Qb ln ˜ _ CdQd
CeQb
(Eq. 13)
KA _
QdCd where A is age (years); H is height (cm); and DW is the
˜ _ if Qb _ Qd
Ce Cd patient’s weight after HD (dry weight in kg). For practi-cal
(Eq. 10) purposes, a simplified formula considering only the
˜
where Ce toxin concentration measured in blood flowing body weight of patients is frequently employed [15] to
calculate volume according to the relationships:
into dialyzer, and Cd is the toxin concentration measured ˜

in dialyzer outflow line. Both concentrations, C and Cd, Vmale _ 0.58DW and Vfemale _ 0.55DW
must be measured at the same time. (Eq. 14)
1156 Zio´łko et al: Hemodialysis modeling

Table 1. Results of measurements and modeling of urea concentrations


Concentration
Mass Concentration in Average
measured in
transfer blood calculated concentration in Concentration
Time blood dialisate coefficient from the concen- blood (c.3 _ c.6)/2 from model
N min mmol/L mmol/L KA mL/min tration in dialysate mmol/L (Eq. 8)
1 2 3 4 5 6 7 8
1 0 34.9 —b —b —b 34.9a 34.8
2 10 30.3 9.8 325 30.5 30.4 30.8
3 22 28.2 9.3 339 29.0 28.6 28.0
4 30 26.4 8.6 330 26.8 26.6 26.7
5 45 24.6 7.9 320 24.6 24.6 24.7
6 60 22.5 7.3 327 22.7 22.6 23.1
7 90 20.9 —b —b —b 20.9a 20.4
8 120 18.1 5.6 297 17.8 18.0 18.0
9 150 16.6 5.0 282 15.6 16.1 15.9
10 180 14.5 4.3 274 13.4 14.0 14.1
11 210 13.0 3.7 254 11.5 12.3 12.4
Median value of mass transfer coefficient KA _ 320 is equivalent to clearance Kd _ 160.
a
Taken from the third column
b
Lack of data

According to Bock, the compartment volume V de-pends related to its determination; the intercompartmental mass
only upon height and weight: transfer coefficient has a contractual character. Such Kc
ought to be fitted so that the toxin concentration described by
Vmale _ _14.249 _ 0.19678H _ 0.29571DW
Equation 8 meets the measured toxin con-centration in the
and Vfemale _ _9.926 _ 0.17003H _ 0.21371DW (Eq. 15) best way. If the amount of measurement data is sufficient,
The approach of Hume-Weyers et al [14] is similar to then it allows for a precise Kc calcula-tion.
Bock’s, but the coefficients are different:
Vmale _ _14.249 _ 0.19678H _ 0.296785DW
RESULTS
and Vfemale _ _35.270121 _ 0.344547H _ 0.183809DW
An example of measurements and modeling
(Eq. 16)
The results of a dialysis session modeling for a 23-year-
The equation old male patient (R.S.), whose postdialysis body weight was
V _ _0.07493713 · A _ 1.01767992 · G 58 kg is presented. The patient was dialyzed thrice a week
employing a polysulfone F-6 dialyzer (Fre-senius), and each
_ 0.12703384 · H _ 0.04012056 · DW
session lasted 3 hours and 30 minutes. The blood flow rate
_ 0.57894981 · D _ 0.00067247 · DW 2 through the dialyzer was 250 mL/ min, and the dialyzer flow
rate 500 mL/min. Blood sam-ples were collected from an
_ 0.03486146 · A · G _ 0.11262857 · DW · G
arterial line before entering the dialyzer, simultaneously with
_ 0.00104135 · A · DW _ 0.00186104 · H · DW (Eq. dialysate samples that were collected from a test port in the
17) dialyzer outflow line. The samples were collected 11 times
presented by Chertow et al is the most complicated but it according to the schedule presented in column 2, Table 1.
was obtained from data recorded for dialysate patients Determina-tions of urea, creatinine, uric acid and phosphorus
[16]. G in Equation 17 is equal to 1 for male and to 0 for in blood and in dialysate were made. On the basis of Equa-
female; D is equal to 1 for diabetes and equal to 0 for tion 10, dialyzer mass permeability coefficients KA were
nondiabetes. calculated for urea, creatinine and uric acid. By minimiz-ing
Few data are available in the literature on the inter- the criterion (Eq. 12), the distribution volumes (V, Ve, Vi)
compartmental mass transfer coefficient Kc, although this were determined for particular toxins in one- and two-
parameter is an integral part of the two-compartmen-tal compartmental models, as well as cellular clearances Kc. The
model (Equation 6). Most likely there are two reasons results are presented in the Tables and Figures. Only one
responsible for this situation. The first results from a much graph is presented for phosphorus, since the modeling results
less frequent use of the two-compartmental model than the were unsatisfactory (Fig. 11). The spe-cific kinetics of
one-compartmental model. The second reason for the lack of phosphate also have been observed by Maasrami et al [11].
information about Kc lies in difficulties
Zio´łko et al: Hemodialysis modeling 1157

Fig. 6. Mass transfer coefficient (KA mL/min) calculated for creatinine.


Fig. 2. Blood urea concentration. Symbols are: (r) measured; (j) one-
compartmental model; (m) two-compartmental model.

Fig. 7. Dialysate creatinine concentration.


Fig. 3. Mass transfer coefficient (KA mL/min) calculated for urea.

Fig. 4. Dialysate urea concentration.


Fig. 8. Blood uric acid concentration. Symbols are: (r) measured; (j)
one-compartmental model; (m) two-compartmental model.

Fig. 5. Blood creatinine concentration. Symbols are: (r) measured; (j) one-
compartmental model; (m) two-compartmental model.
Fig. 9. Mass transfer coefficient (KA mL/min) calculated for uric acid.

The results of urea concentration modeling for patient


R.S. are: (2 ) intercompartment mass transfer coefficient Kc _
477 mL/min,
(1 ) one-compartmental distribution volume V _ (3 ) extracompartment volume Ve _ 10623 mL,
32941 mL, (4 ) intracompartment volume Vi _ 22645 mL,
1158 Zio´łko et al: Hemodialysis modeling

(2 ) intercompartment mass transfer coefficient Kc _ 113


mL/min,
(3 ) extracompartment volume Ve _ 20097 mL,
(4 ) intracompartment volume Vi _ 25282 mL,
(5 ) mean value deviation (Eq. 12) between measured
concentrations (column 7 in Table 2) and the val-ues
of the two-compartmental model is Q _ 12 _mol/L.

Fig. 10. Dialysate uric acid concentration.


The results of uric acid concentration modeling for
patient R.S. are:

(1) one-compartmental distribution volume V _ 22546


mL,
(2) intercompartment mass transfer coefficient Kc _ 208
mL/min,
(3) extracompartment volume Ve _ 11671 mL
(4) intracompartment volume Vi 16278 mL,
(5) mean value deviation (Eq. 12) between measured
concentrations (column 7 in Table 3) and the val-ues
Fig. 11. Blood phosphorus concentration. Symbols are: (r) measured; (j) one- of the two-compartmental model is (Q ) _ 3.0 _mol/L.
compartmental model; (m) two-compartmental model.
Clearance identification
Table 4 presents both the clearance values calculated using
classic methods employing the data provided by the
manufacturer, and the values identified through toxin
concentration measurements. The determinations were done
both in blood flowing into the dialyzer and in the dialysate.
The measurements were performed 10 or 11 times in each
course of three regular HD sessions for six patients. In all the
cases non-reutilized dialyzers were employed. On the basis
of Equation 10, the overall mass transfer coefficients KA
were calculated, and subse-quently, on the basis of Equation
9, the clearance values Kd were computed. The identified
clearances (Table 4) were typically 19% to 33% lower than
published for in vitro values and 10% to 27% lower than
published for in vivo values.

Fig. 12. Comparison of average errors for Equations 13 through 17, and
both one- and two-compartmental models. Symbols are: (m) Wat-son; (j)
Compartment volume identification
Hume; (h) simplified; (X) Chertow, and (_) Bock. The above presented methods were employed to cal-culate
the compartment volumes V, Ve and Vi in six chronically
dialyzed patients. All patient data are listed in Table 5. The
(5 ) mean value deviation (Eq. 12) between measured volumes calculated according to anthro-pometric
concentrations (column 7 in Table 1) and the val-ues measurements are listed in Table 6. In each column the first
of the two-compartmental model is Q _ 0.33 mmol/L. value was calculated when the patient’s post-HD weight was
introduced to the equation, and the second for his/her pre-HD
The results of the creatinine concentration modeling for weight. The differences between the two values are small (up
patient R.S. are: to 6.3%), but they are noticeable in statistical analysis. If the
patient’s predi-alysis weight was introduced into
(1 ) one-compartmental distribution volume V _ 31385 anthropometric equa-tions, then the resulting distribution
mL, volume was always
Zio´łko et al: Hemodialysis modeling 1159

Table 2. Results of measurements and modeling of creatinine concentrations


Concentration
Mass Concentration in Average
measured in
transfer blood calculated concentration in Concentration
Time blood dialisate coefficient from the concen- blood (c.3 _ c.6)/2 from model
N min mmol/L mmol/L KA mL/min tration in dialysate lmol/L (Eq. 8)
1 2 3 4 5 6 7 8
1 0 1499 —b —b —b 1499a 1522
2 10 1423 337 186 1489 1456 1442
3 22 1270 326 212 1441 1356 1356
4 30 1277 302 185 1335 1306 1304
5 45 1237 280 173 1237 1237 1218
6 60 1102 261 186 1153 1128 1144
7 90 1035 —b —b —b 1035a 1024
8 120 935 210 171 928 932 930
9 150 869 193 168 853 861 855
10 180 801 173 161 764 783 793
11 210 778 158 147 698 738 740
Median value of mass transfer coefficient KA _ 173 is equivalent to clearance Kd _ 113.
a
Taken from the third column
b
Lack of data

Table 3. Results of measurements and modeling of uric acid concentrations


Concentration
Mass Concentration in Average
measured in
transfer blood calculated concentration in Concentration
Time blood dialisate coefficient from the concen- blood (c.3 _ c.6)/2 from model
N min lmol/L lmol/L KA mL/min tration in dialysate lmol/L (Eq. 8)
1 2 3 4 5 6 7 8
1 0 421 —b —b —b 421a 418
2 10 371 94 207 373 372 379
3 22 342 89 217 354 348 343
4 30 326 81 201 322 324 325
5 45 298 75 205 298 298 297
6 60 274 69 205 274 274 275
7 90 242 —b —b —b 242a 241
8 120 214 53 200 211 213 214
9 150 189 48 208 191 190 191
10 180 167 43 213 171 169 171
11 210 159 39 197 155 157 154
Median value of mass transfer coefficient KA _ 205 is equivalent to clearance Kd _ 126.
a
Taken from the third column
b
Lack of data

Table 4. F6 Fresenius dialyzer clearance (blood flow rate 250 mL/min and tions of each toxin concentrations were made. The medi-ans
dialysate flow rate 500 mL/min)
from the three assessed sessions for each patient and each
In vitro In vivo Identified compartment are separately listed in Table 7.
Urea 208 188 169 _ 19 The anthropometrically derived volume results (Table
Creatinine 184 169 124 _ 12 6) were compared with the results (Table 7) obtained from
Uric acid No manufacturer data available 132 _ 9
the measurements and computer minimization of criterion
(Eq. 12). For each method and each toxin, the relative error
was calculated as:
greater than the values arrived at through kinetic identi- 1 6 3 V(formulan) _ V(identificationn,i)
fication. E_ oo (n,i) (Eq. 18)
18 V
In Table 7 the results of one- and two-compartmental n_1 i_1 identification

model volume identification for four commonly mea-sured along with standard deviation
uremic toxins (urea, creatinine, uric acid and phos-phorus) 1 6 3 V(formulan) _ V(identificationn,i) 2

are presented. Blood samples and dialysate sam-ples were __ oo E_ (n,i)


! V
drawn in three consecutive dialysis sessions, 10 to 11 times 18 n_1 i_1 1 identification 2

per each treatment. About 400 determina- (Eq. 19)


1160 Zio´łko et al: Hemodialysis modeling

Table 5. Patient data Table 6. Compartment volumes calculated from Equations 13 to 17


Age Height Weight Volumes L
Patient Sex years cm kg Diagnosis
Patient Watson Simple Bock Hume Chertov
M.P. F 11 129.5 22.7–23.7 Chronic pyelonephritis
M.P. 17.3–17.6 12.5–13.0 16.9–17.2 13.5–13.7 20.1–20.3
M.D. F 24 161.5 48.7–51.9 Lupus nephritis
Ma.K. M 17 154 69.4–69.8 Chronic pyelonephritis M.D. 27.2–28.0 26.8–28.5 27.9–28.6 29.3–29.9 31.0–31.7
M.S. F 35 175 55.1–58.0 Renal cirrhosis Ma.K. 40.7–40.9 40.3–40.5 36.6–36.7 36.6–36.7 39.6–39.7
Mi.K. M 14 137 36.5–38.8 Chronic pyelonephritis M.S. 30.2–30.9 30.3–31.9 31.6–32.2 35.2–35.7 35.3–36.0
Chronic rejection after Mi.K. 28.1–28.9 21.2–22.5 23.5–24.2 23.5–24.2 26.4–27.1
kidney transplantation R.S. 35.2–36.2 31.9–33.6 32.1–33.0 32.1–33.0 34.8–35.7
R.S. M 23 153 55.0–58.0 Bladder extrophy with
secondary urolithiasis
and chronic renal failure

basis of the data provided by their manufacturers, that is,


19% for urea and 33% for creatinine. Besarab et al observed
that dialyzer urea clearances were typically 15% to 20%
where: V(formulan,i) is the compartment volume calculated lower than the published manufacturer’s values [17].
from anthropometrically based (Eq. 13 through Eq. 17)
for the n-th patient and i-th HD session; and Videntification(n,i) The above results are based on a relatively large num-ber
is the compartment volume identified for the n-th patient and of measurements and on precise calculation methods. It could
i-th HD session (a sum of two compartments in the case of therefore be postulated that the results are an accurate
the two-compartmental model). representation of the studied phenomena. How-ever, one
should still remember that our studies were carried out in
Determination of cellular clearance in human subjects
young patients (Table 5), whose mean age was 23 years.
The calculated cellular clearance Kc values are pre-sented
in Table 10. In Fig. 14 shows the medians, mean values and The sensitivity of mathematical models Equations 1 and 6,
standard deviations calculated from these data. The median as far as estimating coefficients is concerned, is relatively
value of 415 mL/min for urea is consis-tent with the mean low. This makes it difficult to precisely estimate model
value of 540 mL/min published by Schneditz et al [13] and parameters on the basis of experimental data. It is particularly
552 mL/min used in [11] by Maasrani et al. difficult to estimate those parameters that represent
phenomena of secondary importance. This low sensitivity
also has a beneficial aspect. The effectiveness of modeling
DISCUSSION can be sufficiently good even if the model coefficients have
been selected in a way that is somewhat lacking in precision.
Equation 11 makes it possible to determine a toxin In clinical circumstances, anthropo-metrically derived data
concentration level in the patient’s blood on the basis of are the most useful way to obtain coefficients for
measurements made in the dialysate. Thus, blood col-lection mathematical models. Their application in dialysis may be
from the patient is not necessary, although the amount limited to the Bock, Hume et al and “simplified” methods.
collected each time (0.1 ml drawn 11 times in the course of a The effectiveness of these three methods is similar, and thus
dialysis session) is small and does not constitute any threat each can be successfully employed in clinical practice.
even to anemic patients. Sometimes the legal aspect is more Equation 14 has one great advantage: it is sufficient to know
important. The collection of dialysate samples does not only the patient’s weight. It seems that for this reason, it is
require the patient’s consent, as happens in the case of blood the best among all the equations published in the literature for
samples, since dialysate is treated as a waste product. One calculat-ing the anthropometric distribution volume in a one-
should also bear in mind the fact that the accuracy of compartmental model. Equations 14 to 16 yield errors less
determining toxic levels in dialysate is greater than in blood than 8% magnitude in urea modeling (Table 8). The rates of
samples. Possi-bly, future generation dialyzing units will be errors slightly increased in creatinine modeling. The global
equipped with an analyzer of selected toxin levels in value of both compartments in two-compart-mental models
dialysate to facilitate the evaluation of on-line dialysis is slightly higher than the corresponding value for a one-
effects. Equation 11 will then be required for converting compartmental model. The identified compartment volumes
dialy-sate concentration values to blood toxin levels. In order for uric acid are distinctly lower (Table 7) than for urea and
for a calculation based on Equation 11 to be accurate, the creatinine. Thus, the values arrived at on the basis of
proper values of KA must be known. Unfortunately, they are Equations 14 to 16 are too high, and deviations can reach
not always available. As follows from the mea-surements 31%. The Chertow et al Equation 17 yields values that are too
presented earlier, the effectiveness of dialyz-ers is in practice high (the devia-tions range from 3% for a two-compartmental
lower than might be expected on the creatinine
Zio´łko et al: Hemodialysis modeling 1161

Table 7. Median values of compartment volumes V and Ve _ Vi identified from one- and two-compartmental models
Volume L
Urea Creatinine Uric acid Phosphorus
Patient one-comp. two-comp. one-comp. two-comp. one-comp. two-comp. one-comp. two-comp.
M.P. 14.4 5.0 _ 9.9 13.6 6.4 _ 10.8 14.4 4.8 _ 10.8 3.2 2.5 _ 30.7
M.D. 29.0 11.5 _ 17.7 27.3 14.0 _ 17.7 19.9 6.6 _ 14.6 33.6 11.6 _ 66.4
Ma.K. 36.0 3.6 _ 29.1 30.1 10.8 _ 23.9 27.0 7.0 _ 20.3 75.8 4.0 _ 78.3
M.S. 32.7 18.1 _ 21.1 33.8 25.8 _ 19.9 25.2 11.2 _ 15.7 19.5 8.6 _ 114.5
Mi.K. 16.9 5.2 _ 11.6 19.5 15.7 _ 3.9 17.2 5.0 _ 11.6 16.2 1.9 _ 24.2
R.S. 32.9 10.6 _ 22.6 30.4 22.7 _ 52.3 25.5 11.7 _ 16.3 20.1 7.5 _ 39.7

Fig. 13. Comparison of standard deviation values for Equations 13


through 17, and both one-and two-compartmental models. Symbols are:
(m) Watson; (j) Hume; (h) simplified; (X) Chertow, and (_) Bock. Fig. 14. Comparison of cellular clearances for the selected toxins. Sym-
bols are: (m) median; (h) mean value; (whisker) standard deviation.

model to 46% for a one-compartmental uric acid model. The


where DW is the postdialysis weight of the patient (dry
average errors and the standard deviations for an-
weight). Equation 20 represents the median values (Ta-ble
thropometrically based methods are presented in Table 8 and
11) for the three toxins. In the case of the one-compartmental
Table 9, respectively.
model, the toxin distribution volume de-viations from 56%
When the postdialysis weight of the patient is intro-duced
of the patient’s weight do not exceed 10% for all the three
into Equations 14 to 16, satisfactory distribution volumes are
toxins. The global values (Ve _ Vi) for the two-
obtained for urea and creatinine, while the values for uric
compartmental model show relatively small deviations from
acid are too high (compare Table 6 and Table 7; see Fig. 12).
the value of 0.57DW arrived at using Equation 20. Clearly,
All anthropometrically based meth-ods, except Equation 17,
larger deviations are ob-served for particular compartments
were established in healthy individuals. Thus, the post-
in a two-compart-mental model. This means that the value of
hemodialysis weight of the patient, or the so-called dry
the sum of both compartments is of great importance in
weight, should be introduced into these equations.
modeling accuracy, but its distribution to particular
compartments is of secondary significance. According to
It seems that adequately precise volumes for Ve and Vi can
Yasamura’s and Friis-Hansen’s data in Maasrani et al’s study
be achieved from the patient’s body mass. Table 11 presents
[11], Ve _ V/3 and Vi _ 2V/3, which makes Ve _ Vi _ V. From
the distribution volume median values for the three uremic
our observations it follows that Ve _ 0.375V and Vi _ 0.643V,
toxins studied. As follows from these data, the following
which gives Ve _ Vi _ 1.018V. The aggre-gate volume of the
equations can be employed to calcu-late toxin distribution
two compartments in the bicompart-ment model is slightly
volumes:
greater than the volume in the single-compartment model.
V _ 0.56DW The results (Table 7) for the single-compartment model
differ from the values derived from Equation 20 by
Ve _ 0.21DW (Eq. 20)
approximately 7%, 2%, 4%, 2%,
Vi _ 0.36DW
1162 Zio´łko et al: Hemodialysis modeling

Table 8. Average errors (Eq. 18) for anthropometrically based methods of volume determination
Urea Creatinine Uric acid
one-comp. two-comp. one-comp. two-comp. one-comp. two-comp.
Watson 0.15 0.12 0.17 _0.02 0.39 0.32
Equation 14 0.01 0.02 0.03 _0.14 0.24 0.18
Bock 0.08 0.05 0.10 _0.08 0.31 0.25
Hume 0.07 0.04 0.08 _0.09 0.30 0.24
Chertow 0.21 0.17 0.23 0.03 0.46 0.39

Table 9. Standard deviations (Eq. 19) for anthropometrically based methods of volume determination
Urea Creatinine Uric acid
one-comp. two-comp. one-comp. two-comp. one-comp. two-comp.
Watson 0.26 0.27 0.20 0.29 0.23 0.24
Equation 14 0.16 0.19 0.16 0.23 0.24 0.25
Bock 0.16 0.18 0.12 0.23 0.15 0.16
Hume 0.17 0.19 0.11 0.24 0.21 0.22
Chertow 0.20 0.21 0.15 0.27 0.16 0.17

Table 10. Identified cellular clearances Table 11. Estimates of cellular clearances and distribution volumes

Kc mL/min Kc median V/DW Ve/DW Vi/DW


Patient Urea Creatinine Uric acid Urea 415 0.60 0.21 0.40
Creatinine 207 0.56 0.35 0.36
M.P. 54, 394, 490 130, 170, 63 232, 314, 208
Uric acid 257 0.46 0.17 0.30
M.D. 640, 288, 372 408, 11774, 533 349, 154, 334
Ma.K. 310, 447, 203 8395, 12048, 9755 165, 116, 257
M.S. 174, 290, 415 207, 197, 311 245, 121, 473
Mi.K. 768, 1108, 633 332, 165, 444 474, 480, 290
R.S. 477, 749, 332 113, 44, 63 208, 386, 159
and creatinine is mentioned by Maasrani et al [11]. They
obtained the mean value of Kc _ 552 for urea and Kc _ 274
for creatinine (our estimations are Kc _ 415 and 207,
11% and 2%. The sum of both compartments differs from respectively). These relationships were similar to the
that derived from Equation 20 by 6%, 0%, 11%, 11%, 13% relationships between dialyzer clearances for the ex-amined
and 1%. Thus, it seems that these deviations can be toxins (Table 4). The clearance differences are more
considered acceptable. Slightly greater deviation values are prominent for the cellular membrane than for arti-ficial
obtained for extracellular compartments. The sum of both semipermeable membranes. The median values of urea
compartments can be estimated in a more precise way than cellular clearances were almost three times greater than the
its particular components. values of urea dialyzer clearance. In the case of uric acid and
In modeling phosphorus kinetics, other mathematical creatinine, the cellular clearances are almost two times
models should be employed, as the presently obtained results greater than dialyzer clearances. Sig-nificant differences in
are unsatisfactory. The phosphorus concentration level at the cellular clearance were found for creatinine. For two patients
end of a dialysis session is stable, and some-times even (M.D. and Ma.K. in Table 10), much higher values were
increases, which cannot be satisfactorily mod-eled obtained than for other pa-tients. This means that creatinine
employing Equations 1 and 6. The compartment volumes may be wholly available directly through blood. In such
(Table 7) calculated on the basis of the phospho-rus cases, a one-compartmen-tal model is more adequate than a
concentration measured in blood have considerable errors. two-compartmental one. The single-compartmental
Similar effects were observed by Maasrani et al, and some concentration kinetics (a very high value of Kc), which is
explanations are given in [11]. seen sometimes for creatinine, is more rare for urea, and
In a two-compartmental model, the cellular clearances never occurs for uric acid, clearly results from our
were found to show individual differences for given tox-ins. measurements. In Schneditz et al’s study it is suggested that
It is well known that Kc urea is greater than Kc creatinine. A the high Kc values reported in the literature (up to 2000
similar effect was found in our studies: Kc urea was two mL/min) may include a fair amount of access recirculation
times greater (Table 11) than Kc creatinine and Kc uric acid. and cardiopulmonary recirculation [13].
Exactly this same proportion of urea
Zio´łko et al: Hemodialysis modeling 1163

The parameters Ve, Vi and Kc identified on the basis of Replacement of Renal Function by Dialysis, edited by MAHER JF, New
York, Kluwer Academic Publishers, 1989, pp 87–143
experimental data have considerable standard devia-tions as 5. BURGELMAN M, VANHOLDER R, FOSTIER H, BEIGAIR S: Estimation
a result of the relatively small sensitivity of the quantity of parameters in a two-pool urea kinetic model for hemodialysis. Med
criterion (Eq. 12) near the optimal solution. The two- Eng Phys 19:69–76, 1997
6. DEPNER TA: Prescribing Hemodialysis. New York, Kluwer Aca-demic
compartmental model is mathematically more complicated Publishers, 1991
than the corresponding one-compartmental model. The 7. GOTCH FA: Kinetic modeling in hemodialysis, in Clinical Dialysis,
edited by NISSENSON AR, FINE RN, GENTILE D, Norwalk, Appleton and
effectiveness in hemodialysis modeling is similar for both Lange, 1990, pp 118–146
models, thus justifying the use of a one-compartmental 8. HAKIM RM, DEPNER TA, PARKER TF: Adequacy of hemodialysis. Am J
model. The two-compartmental descrip-tion of toxin kinetics Kidney Dis 20:108–122, 1992
9. KOPPLE JD, JONES MR, KESHAVIAH PR, BERGSTROM J, LINDSAY RM,
ought to be used when explaining a steep increase of toxin MORAN J, NOLPH KD: A proposed glossary for dialysis kinetics. Am J
concentration immediately after the end of the dialysis (the Kidney Dis 26:963–981, 1995
so-called rebound phenom-enon). 10. LOPOT F (editor): Urea kinetic modelling, in EDTA/ERCA Series
(vol 4), Belgium, D. Verlinde, 1990
11. MAASRANI M, JAFFRIN MY, FISCHBACH M, BOUDAILLIEZ B: Urea,
creatinine and phosphate kinetic modeling during dialysis: Applica-tion
ACKNOWLEDGMENTS to pediatric hemodialysis. Artif Kidney Dial 18:122–129, 1995
12. PASTAN S, COLTON C: Transcellular urea gradients cause minimal
This research was supported in part by Scientific Research Commit-tee depletion of extracellular, during hemodialysis. ASAIO Trans 35:247–
grant KBN 0553/P4/93/05. 250, 1989
13. SCHNEDITZ D, FARIYIKE B, OSHEROFF R, LEVIN MW: Is intercompar-
Reprint requests to Dr. Jacek Pietrzyk, Department of Pediatric Ne- tmental urea clearance during hemodialysis a perfusion term? A
phrology, Polish-American Children’s Hospital, ul.Wielicka 265, 30–663 comparison of two pool urea kinetic models. J Am Soc Nephrol 6:1360–
Krako´w, Poland. 1370, 1995
14. WATSON PE, WATSON ID, BATT RD: Total body water volumes for adult
males and females estimated from simple anthropometric
REFERENCES measurements. Am J Clin Nutr 33:27–39, 1980
15. HUME R, WEYERS E: Relationship between total body water and surface
1. WOLF AV, REMP DG, KILLEY JE, CURRIE GD: Artificial kidney area in normal and obese subjects. J Clin Pathol 24:234–238, 1971
function: Kinetics of HD. J Clin Invest 30:1062–1070, 1951
2. RENKIN EW: The relationship between dialysance, membrane area, 16. CHERTOW GM, LAZARUS JM, LEW NL, MA L, LOWRIE EG: Develop-
permeability, blood flow in the artificial kidney. ASAIO Trans 2:102– ment of a population-specific regression equation to estimate total body
105, 1956 water in hemodialysis patients. Kidney Int 51:1578–1582, 1997
3. SARGENT JA, GOTCH FA: Mathematical modeling of dialysis ther-apy. 17. BESARAB A, ROSS R, FRINAK S, ZASUWA G: Increased urea kinetic
Kidney Int 18(Suppl 10):2–10, 1980 modeling, possible mechanisms and its significance. ASAIO J 43:303–
4. SARGENT JA, GOTCH FA: Dialysis principles and biophysics, in 310, 1997

Vous aimerez peut-être aussi