REVISION CONTROL
Review/Approval History
Table of Contents
Page
1. Categories _______________________________________________________________5
2. Boot up__________________________________________________________________6
2.1 Symptoms_____________________________________________________________________ 6
2.2 Check route ___________________________________________________________________ 6
2.3 Reason of defect________________________________________________________________ 6
2.4 Defective parts statistics _________________________________________________________ 6
3. Black screen _____________________________________________________________7
3.1 Symptoms_____________________________________________________________________ 7
3.2 Check route ___________________________________________________________________ 7
3.3 Reason of defect________________________________________________________________ 7
3.4 Defective parts statistics _________________________________________________________ 7
4. Keyboard and trackball related _______________________________________________8
4.1 Symptoms_____________________________________________________________________ 8
4.2 Solution (keyboard)__________________________________________________________ 8
4.3 Reason of defect________________________________________________________________ 8
4.4 Defective parts statistics _________________________________________________________ 9
5. Lock up __________________________________________________________________10
5.1 Symptoms____________________________________________________________________ 10
5.2 Solution _____________________________________________________________________ 10
5.3 Reason of defect_______________________________________________________________ 10
6. Image related _____________________________________________________________11
6.1 Symptoms____________________________________________________________________ 11
6.2 Check route __________________________________________________________________ 11
6.3 Reason of defect_______________________________________________________________ 12
6.4 Defective parts statistics ________________________________________________________ 13
7. Intermittent problems _______________________________________________________14
7.1 Symptoms____________________________________________________________________ 14
7.2 Solution _____________________________________________________________________ 14
8. Screen flicker _____________________________________________________________15
8.1 Symptoms____________________________________________________________________ 15
8.2 Solution _____________________________________________________________________ 15
1. Categories
This document offers several methods to isolate defects or problems according to the symptoms observed.
We can classify the symptoms to several categories. The categories are very helpful for you to find
problems in different systems and it’s a general tool, not only used in the system described in this
document.
For different categories, we have different methods to solve problems. Most of the methods are
commonly used in all Mindray ultrasound systems but there might be slight differences when you use
them on different systems.
2. Boot up
2.1 Symptoms
Stops at Mindray LOGO, and nothing else happens
2.2 Check route
power supply
45.50% board
54.50% main board
3. Black screen
3.1 Symptoms
When power on the machine, screen is black.
3.2 Check route
power supply
45.00%
board
monitor
55.00%
Note:
The flow chart above only tell how to deal with the problem that keyboard can’t function,
When meeting trackball out of control and nothing else abnormal, please check the connection between
Trackball and keyboard or connection between trackball (keyboard) and DSC board.
Change the keyboard PCB, or trackball
clean or adjust
silicon rubber
pad position
2% 22.00% silicon rubber
28% pad
keyboard PCB
48.00%
power supply
board
5. Lock up
5.1 Symptoms
When you press some keys the system locks up, and nothing else abnormal. Or can not wake up from
system dormant
5.2 Solution
Change the main board
5.3 Reason of defect
Main board failure
6. Image related
6.1 Symptoms
Machine is working normally, but image is abnormal
6.2 Check route
Note: Some customers complain that the image quality is not good; generally, these problems can be
solved by adjusting the machine (such as monitor brightness, image parameters …….)
Sometimes the defect of monitor itself can also cause the image display not good.
6.3 Reason of defect
Some electronic or magnetic interference, probe failure, channel board failure, and main board failure.
machine is not
adjusted well
probe
17.00% 22.00%
monitor
22.00% 11.00%
main board
28.00%
power supply
board
7. Intermittent problems
7.1 Symptoms
Image with some ripples
7.2 Solution
Check the surrounding, including AC power input, is there any machines which will interfere with the
system. Do remember keep a good environment.
8. Screen flicker
8.1 Symptoms
Screen glinting or trembling
8.2 Solution
Exchange the CRT. Or check the surrounding, including AC power input, is there any machines which
will interfere with the system. Do remember keep a good environment.