Académique Documents
Professionnel Documents
Culture Documents
MICROS
P/N : RAA 006 A Ind.F
ABX HEMATOLOGIE
B.P. 7290
Rue du Caducée
Parc Euromédecine
34184 MONTPELLIER Cedex 04
Tel.: 67 14 15 16
FAX : 67 14 15 17
REVISIONS
P/N
Index revision REVISION Section Date
A RAA 006 A A Creation 30/09/94
B RAH 154 AA Update following the V1.2 version program. all 13/03/95
C RAH 229 AA New software version V2.03 RAS 091 A 15/11/95
RAS 121 A
D RAH 262 AA New procedure RAS 134 A and miscellaneous corrections 19.02.96
E RAH 326 AA Update following the V2.2 version program. all 04.03.97
F RAH 504 AA Update all sections MICROS / MICROS 60 all 09.03.98
SECTION 1
1. INTRODUCTION ............................................................................................................................2
1.1. Preliminaries ...................................................................................................................2
1.2. Description of the manual ...............................................................................................2
SECTION 2
1. PRELIMINARIES .................................................................................................................................... 2
2. DESCRIPTION OF THE MANUAL.......................................................................................................... 2
1. INTRODUCTION................................................................................................................................... 2
2. HYDRAULIC PARTS .............................................................................................................................. 2
3. PNEUMATIC DIAGRAMS..................................................................................................................... 2
1. INTRODUCTION................................................................................................................................... 2
2. ELECTRIC PRINCIPLES ......................................................................................................................... 2
3. POWER SUPPLY ................................................................................................................................... 2
4. ELECTRONIC PRINCIPLES.................................................................................................................... 2
5. INTERCONNECTION DIAGRAMS........................................................................................................ 2
1. CUSTOMER MAINTENANCE
1.1 Introduction
WARNING ! : Never use solvent or an abrasive cleaning material to clean the system.
Page 4/2
RAA 006 A Ind.F 10/03/98
4 - MAINTENANCE PROCEDURES
A - Hydraulic
B - Mechanical
Page 4/3
RAA 006 A Ind.F 10/03/98
4 - MAINTENANCE PROCEDURES
3. PROCEDURES
CONCERNED PARTS :
- Hydraulic maintenance and adjustments
- Pneumatic maintenance and adjustments
- Electrical maintenance and adjustments
- Power supply maintenance and adjustments
- Electronic maintenance and adjustments
- Printer maintenance and setup
CAUTION : Each procedure has to be read entirely before the intervention is begun.
Page 4/4
RAA 006 A Ind.F 10/03/98
4 - MAINTENANCE PROCEDURES
Fig.4.1
Page 4/5
RAA 006 A Ind.F 10/03/98
4 - MAINTENANCE PROCEDURES
Fig.4.2
Page 4/6
RAA 006 A Ind.F 10/03/98
4 - MAINTENANCE PROCEDURES
Page 4/7
RAA 006 A Ind.F 10/03/98
MICROS DILUTION BLOCK MAINTENANCE
* CONCERNS
- Lyse dispenser O ring replacement
- Diluent dispenser O ring replacement
- Sampling needle dispenser O ring replacement
* REQUIRED TOOLS
- Allen key : 2,5
2
- Dynamometric screw driver A302 : MAG 019 A
A301 : MAG 020 A
* REQUIRED PRODUCTS
- Silicone grease : LAM 004 A
- Soft paper
* INTERVENTION TIME
- 15min
* FREQUENCY
- On request or or on a leak case.
- O rings replacement : 1/year
Page 1/5
RAS 084 A Ind.C 06.03.98
MICROS DILUTION BLOCK MAINTENANCE
* PROCEDURE
- Pull the piston assy in the upper position and clamp the diluent and lyse
tubing as described on the diagram 1.
- Disconnect the diluent/lyse/sampling tubings from the 3 syringes and the
tube on the sampling syringe side.
Diag.1
Page 2/5
RAS 084 A Ind.C 06.03.98
MICROS DILUTION BLOCK MAINTENANCE
Diag.2
- Pull out the pistons , and from the body with their respective o
ring still around (Diagram 3).
- Replace the lyse and diluent O rings by new ones. Check the cleanliness of
the piston and of the syringe bodies. If necessary clean with a soft paper.
Page 3/5
RAS 084 A Ind.C 06.03.98
MICROS DILUTION BLOCK MAINTENANCE
Diag.3
- Spread a little amount of silicone grease between two fingers and apply a
very thin film of grease on the 2 new sampling syringe O rings. Replace the
old ones .
Page 4/5
RAS 084 A Ind.C 06.03.98
MICROS DILUTION BLOCK MAINTENANCE
* REMARKS
Page 5/5
RAS 084 A Ind.C 06.03.98
MICROS VACUUM/WASTE SYRINGE MAINTENANCE
* CONCERNS
- O ring replacement.
* REQUIRED TOOLS
- Allen key : 2,5
- Dynamometric screw driver A302 : MAG 019 A
* REQUIRED PRODUCTS
- Silicone grease : LAM 004 A
* INTERVENTION TIME
- 10min
* FREQUENCY
- O ring replacement : - Type 1 : 1/year
- Type 2 : 2/year
- Type 3 : 3/year
Page 1/4
RAS 085 A Ind.C 10/03/98
MICROS VACUUM/WASTE SYRINGE MAINTENANCE
* PROCEDURE
- Manually pull the syringe piston up in order to freed it from the syringe
body.
NOTE : The syringe has to be linked to the atmosphere, pressing the valve , to
pull the piston out from the syringe body.
Turn the cylindrical gearing by hand to help the raising of the piston
(See Diagram 1).
Diag.1
Page 2/4
RAS 085 A Ind.C 10/03/98
MICROS VACUUM/WASTE SYRINGE MAINTENANCE
- Spread a little amount of silicone grease between two fingers and apply a
very thin film of grease on a new O ring.
- Reinstall in the reverse order. Apply the following torque to the screws :
700 mN/m.
CAUTION : Push the piston back inside the syringe body before tightening the o ring
fixation screws . Use the dynamometric screw driver to tighten the
screws to 400 mN/m
Page 3/4
RAS 085 A Ind.C 10/03/98
MICROS VACUUM/WASTE SYRINGE MAINTENANCE
* REMARKS
Page 4/4
RAS 085 A Ind.C 10/03/98
MICROS CT SAMPLING NEEDLE MAINTENANCE
* CONCERNS
- Needle O ring replacement
- Sampling needle replacement
- Piercing needle replacement
* REQUIRED TOOLS
- Allen key : 2,5
- Dynamometric screw driver A302 : MAG 019 A
* REQUIRED PRODUCTS
- Silicone grease : LAM 004 A
* INTERVENTION TIME
- 15min
* FREQUENCY
- See frequency chart table for cleaning.
- O ring replacement : 1/year
- Needle replacement : On request only
Page 1/5
RAS 086 A Ind.C 09/03/98
MICROS CT SAMPLING NEEDLE MAINTENANCE
* PROCEDURE
A - Needle or O ring replacement
- Disconnect the tube from the top of the needle . Manually lift up
the sampling needle (Diagram 1).
- Unscrew the 2 screws in order to freed the needle rinsing block
from the carriage frame .
Diag.1
- Remove the clip and then the rinsing block/needle assy from the
carriage taking care not to bend the needle.
Page 2/5
RAS 086 A Ind.C 09/03/98
MICROS CT SAMPLING NEEDLE MAINTENANCE
- Lift up the O ring holder and replace the O rings by new ones
previously greased. Wipe all excess of grease away.
- If necessary clean the inner surface of the rinsing block with a little piece of
paper.
Diag.2
Page 3/5
RAS 086 A Ind.C 09/03/98
MICROS CT SAMPLING NEEDLE MAINTENANCE
Diag.3
- Lift the needle in the upper position.
- Loosen the 2 screws just enough to enable the rinsing block to be lifted
up of about 5 mm.
- Pull the piercing needle foreward and replace it by a new one.
- Reassemble in the reverse order. Apply the same torque : 700mN/m to
tighten the two screws .
Page 4/5
RAS 086 A Ind.C 09/03/98
MICROS CT SAMPLING NEEDLE MAINTENANCE
* REMARKS
Blockage problems may occur on some instruments during the needle or
carriage motions giving some motor error messages.
Remove the axle screw of the puley and clean its 2 parts and the
washer.Reinstall the puley assy, the rounded edge facing the puley. Tighten
the screw with a torque of 400mN/m.
Check that the puley turns freely after the tightening. Add a drop of oil (LAM
007 A) if necessary.
A service kit (XEA 343 AS) will be available from May 1995 and will include :
Page 5/5
RAS 086 A Ind.C 09/03/98
MICROS OT SAMPLING NEEDLE MAINTENANCE
* CONCERNS
- Needle replacement
- O ring replacement
* REQUIRED TOOLS
- Allen key : 2,5
- Dynamometric screw driver A300 : MAG 013 A
* REQUIRED PRODUCTS
- Silicone grease : LAM 004 A
* INTERVENTION TIME
- 15 min
* FREQUENCY
- Needle replacement : On request only
- O ring replacements : 1/year
Page 1/3
RAS 087 A Ind.C 09/03/98
MICROS OT SAMPLING NEEDLE MAINTENANCE
* PROCEDURE
- Disconnect the tube from the top of the needle . Manually lift up
the sampling needle (Diagram 1).
- Unscrew the 2 screws in order to freed the needle rinsing block
from the carriage frame. Remove the clip .
Diag.1
- Remove the rinsing block/needle assy from the carriage taking care not to
bend the needle.
- If necessary clean the inner surface of the rinsing block by means of a little
piece of soft paper. Spread a little amount of grease in between the rinsing
block and its support.
- Lift up the O ring holder and replace the O ring by a new one
previously greased. Wipe all excess of grease away.
Page 2/3
RAS 087 A Ind.C 09/03/98
MICROS OT SAMPLING NEEDLE MAINTENANCE
* REMARKS
Blockage problems may occur on some instruments during the needle or
carriage motions giving some motor error messages.
Remove the axle screw of the puley and clean its 2 parts and the
washer.Reinstall the puley assy, the rounded edge facing the puley. Tighten
the screw with a torque of 400mN/m.
Check that the puley turns freely after the tightening. Add a drop of oil (LAM
007 A) if necessary.
A service kit (XEA 343 AS) will be available from May 1995 and will include :
Page 3/3
RAS 087 A Ind.C 09/03/98
MICROS CHAMBER MAINTENANCE
* CONCERNS
- RBC, WBC/HGB, Mixing chamber cleaning
- Aperture O ring replacement
- Coaxial O ring replacement
* REQUIRED TOOLS
- Allen key : 2,5
- Soft paper
- Dynamometric screw driver : MAG 013 A
- Cutting pliers
* REQUIRED PRODUCTS
- Liquid soap
- Distilled water
- A scalpel
- A Micropipette tip
* INTERVENTION TIME
- 30 min
* FREQUENCY
- RBC, WBC, Mixing chamber cleaning : 2 (type 1 & 2) or 3/year (type 3).
- Aperture O ring replacement : 1/year
- Electrode O ring replacement : 1/year
Page 1/7
RAS 088 A Ind.C 09/03/98
MICROS CHAMBER MAINTENANCE
* PROCEDURE
1 - RBC chamber cleaning :
- Run a drain chamber cycle ("SERVICE" menu, "DRAIN CHAMBERS " Sub
menu).
- Record the tube positions before dismantling the chambers.
- Disconnect the chamber tubes.
- Unclip the RBC chamber. Dismantle the electrode loosening the 2 fixation
screws (Diagram 1).
Diag.1
Page 2/7
RAS 088 A Ind.C 09/03/98
MICROS CHAMBER MAINTENANCE
- Clean the groove and replace the O ring , using the clamp as shown on
Diag.2.
Diag.2
Page 3/7
RAS 088 A Ind.C 09/03/98
MICROS CHAMBER MAINTENANCE
- Remove the fixation screws from the counting head (See diag.3).
Diag.3
WARNING ! : Do not manipulate the aperture using hard instruments. Clean the aperture
with a piece of soft paper or preferably, in between 2 fingers.
- Clean the chamber and the counting head with liquid soap, do not
introduce any sharp instruments inside so as to avoid damaging the inside
of the chamber and the aperture.
- Rinse thoroughly with distilled water
- Dry the exterior of the chamber with a soft paper.
- Re-assemble the chamber. Replace the O ring (Diag.3).
WARNING ! : Do not apply too much pressure on the counting head or on the electrode
fixation screws, as it can break the aperture (tightening torque =
100mN/m).
Page 4/7
RAS 088 A Ind.C 09/03/98
MICROS CHAMBER MAINTENANCE
2 - WBC chamber
- Run a drain chamber cycle ("SERVICE" menu, "DRAIN CHAMBERS" Sub
menu).
- Loosen the cover screws of the WBC chamber and remove the cover.
- Record the tube positions before dismantling the chamber.
- Disconnect the chamber tubes and top.
- Unclip the WBC chamber. Unscrew the photometer fixation screws
(Diagram 4).
Remove, clean the WBC chamber and replace the O rings as described in
"1 - RBC chamber".
Diag.4
Page 5/7
RAS 088 A Ind.C 09/03/98
MICROS CHAMBER MAINTENANCE
3 - Mixing chamber
- Run a drain chamber cycle ("Service" menu, "Drain chambers" Sub menu).
- Record the tube positions before dismantling the chamber.
- Unclip the chamber and clean it with liquid soap.
- Rinse thoroughly with distilled water
- Dry the exterior of the chamber with a soft paper.
- Re install in the reverse order.
Page 6/7
RAS 088 A Ind.C 09/03/98
MICROS CHAMBER MAINTENANCE
* REMARKS
- Once the cleaning done on the chambers, start the instrument and check
for leaks.
- Check the calibration too.
Page 7/7
RAS 088 A Ind.C 09/03/98
MICROS LIQUID VALVE MAINTENANCE
* CONCERNS
- Liquid valve assy replacement
- Valve body replacement
* REQUIRED TOOLS
- Allen key
- Flat screw driver
* REQUIRED PRODUCTS
- Soft paper
* INTERVENTION TIME
- 15 min
* FREQUENCY
- On request only
Page 1/4
RAS 089 A Ind.C 09/03/98
MICROS LIQUID VALVE MAINTENANCE
* PROCEDURE
- Several reasons can unabled the correct operations of the valves :
Diag.1
Page 2/4
RAS 089 A Ind.C 09/03/98
MICROS LIQUID VALVE MAINTENANCE
Diag.2 & 3
NOTE : When leaks occur on valves it is recommended to replace the entire valve
assembly by a new one.
- If only one valve has been damaged it is possible to dismantle the valve
body on its own as shown on the diagram 4 : use a screw driver to
disconnect the valve holder clip and remove the body.
Diag.4
Page 3/4
RAS 089 A Ind.C 09/03/98
MICROS LIQUID VALVE MAINTENANCE
* REMARKS
When piercing several times the same tube cap, some pieces of cork may
be dragged along towards the mixing chamber, the WBC chamber and
then the liquid valve .
This may damaged the operation of the valve.
Page 4/4
RAS 089 A Ind.C 09/03/98
MICROS POWER SUPPLY CHECK AND REPLACEMENT
* CONCERNS
- Voltage supply check
- Power supply module replacement
- Fan operation check
* REQUIRED TOOLS
- Allen key : 2,5
2
- Flat screw driver
- Volmeter
* REQUIRED PRODUCTS
- None
* INTERVENTION TIME
- 15 min
* FREQUENCY
- 1/year
Page 1/5
RAS 090 A Ind.C 09/03/98
MICROS POWER SUPPLY CHECK AND REPLACEMENT
* PROCEDURE
1 - Supply voltage check
CAUTION : The supply voltage check has to be done with the power supply module
connected only.
- The supply voltage check is done on the mother board (XAA 355 A)
according to the following chart table :
* Ground on TP31, TP 30 or TP 29
WARNING ! : The two screws on the top of the coaxial cover holding the board on the
MICROS frame are the grounding connection of the mother board.
It is mandatory to chek the correct tightening of these screws to obtain
correct voltage values.
Page 2/5
RAS 090 A Ind.C 09/03/98
MICROS POWER SUPPLY CHECK AND REPLACEMENT
SUPPLIES
-12V +- 0.5
+0.6 TP20
24V -0 TP22
5V +0.3
-0 TP23
12V -+0.5TP21
SUPPLY
+5V
GND
J26
+24V
+12V
-12V
Diag.1
NOTE : If the voltages values are not correct or among the ranges no adjustment
can be carried out either on the board or on the power supply module.
Replace the power supply module as described below :
Page 3/5
RAS 090 A Ind.C 09/03/98
MICROS POWER SUPPLY CHECK AND REPLACEMENT
Diag.2
WARNING ! : The power supply module internal fuse is not to be replaced even when
this one has blown down.
- Replace the power supply module by a new one and reinstall in the
reverse order.
- Switch on the instrument and check the voltages on the mother board as
described in the previous paragraph.
Page 4/5
RAS 090 A Ind.C 09/03/98
MICROS POWER SUPPLY CHECK AND REPLACEMENT
* REMARKS
The four leds in front of the test points are lit to indicate a voltage presence
but whatever its value!!!
Check the operation of the fan as following : When the fan has stopped,
move the cursor to the 4 SERVICE menu and press ENTER : the fan should
start.
Page 5/5
RAS 090 A Ind.C 09/03/98
MICROS OT TECHNICIAN FUNCTIONS
* CONCERNS
* REQUIRED TOOLS
- Flat screw driver
- thermometer
- Barflex
- Allen key : 2,5
- Voltmeter
* REQUIRED PRODUCTS
- WBC latex : LAD 001 AS
- RBC and PLT latex : LAD 002 AS
- Soft paper
- Flat piece of stiff plastic
* INTERVENTION TIME
- 60 min
* FREQUENCY
- See maintenance chart table.
* PROCEDURE
The arborescence of the «technician functions» is the following :
1 - CAL. PHOTOMETER 1 - WBC ADJUST
1 - VERSION
2 - ADJUSTMENTS
2 - MCV < >
7 - VACUUM
1 - VACUUM < >
2 - CHECK
8 - TECHNICIAN
1 - UNCAL. TEMPERATURE
3 - TEMPERATURE 2 - CALIBRATE
1 - VISIBLE
3 - TEMPERATURE
2 - INVISIBLE
MICROS OT
1 - CONTROL
4 - RUN MODE
2 - USER
1 - YES
5 - REAGENT PACK
2 - NO
6 - SERIAL NUMBER
1 - BURNING #
8 - BURNING 2 - BURNING
3 - CONTROL
Once entering the «SERVICE» sub menu, move to «TECHNICIAN FUNCTIONS» and press .
A specific password (421) is required to enter the sub menus.
Move the cursor by means of and and choose the required menus pressing the
key.
II - Adjustments :
NOTE - If the WBC chamber has been dismantled previously make sure no liquid has flown
in between the spectrophotomer and the chamber.
Clean the inner surfaces of the spectrophotometer as well as the chamber.
Reassemble the assy and tighten the two screws to the following torque : 400mN/m
(see RAS 088 A : Chamber maintenance)
An HGB channel is displayed on the LCD screen representing a voltage value as shown below :
VALUE
237
The HGB photometer calibration must be done 20min at least after the
IMPORTANT instrument has been switched on.
- By means of R97 (See diagram 1) adjust the HGB channel according to the room temperature using
the chart table given on next page.
IMPORTANT This adjustment must be done with the WBC chamber cover installed!!!
HGB
R97 TP27
TP19
APERTURE R91
VOLTAGE
60V +0.1
-
Diag.1
2 - Aperture voltage
- Once entering the «APERTURE VOLTAGE» menu, connect a voltmeter between the ground (TP30)
and TP19.
3 - Pressure check
- If the results are not correct check the O ring and the
Diag.2 tubing watertightness.
4 - WBC adjust
- Put the WBC latex to mix on a Vortex during 5min or shake thoroughly
- Wait for several results to be displayed and check the stability of both values.
After one minute the last volume values displayed on the screen are printed out as well as the WBC,
PLT and RBC histograms and the CBC “s results. Check that these printed values correspond to the
following target values :
- LYM = 57 + 1
- GRA = 180 + 2
- If not rerun a «WBC ADJUST» cycle after having previously mixed the latex vial again. By means
of R74 (see diagram 4) adjust the volumes to the target values during the measuring phasis.
THRESHOLD
THRESHOLD R75 350mV+5 APERTURE R91
400mV+5 TP13 - VOLTAGE
TP4 - R74 TP7
GAIN
R87 TP15
R81 GAIN 60V +0.1
-
J17 GAIN
TP11
GND TP29
NU
RBC PULSE PROCESS WBC PULSE PROCESS
Diag.4
- Put the RBC and PLT latex to mix on a Vortex during 5min or shake thoroughly
- Present the vial of Latex to the open probe as shown on diagram 3 and press the sampling bar
located behind the sampling needle : the needle directly delivers the latex sample in the RBC chamber
dilution (for a usual analysis cycle the sample is first delivers to the mixing chamber) and a measuring
phasis begins.
- During the cycle measuring phasis (around 1 minute) the Platelet and the Red Blood cell volumes
are displayed on the screen every 3 seconds as shown below :
- Wait for several results to be displayed and check the stability of both values.
After one minute the last volume values displayed on the screen are printed out as well as the PLT
and RBC histograms and the CBC “s results.
- Check that these printed values correspond to the following target values :
- RBC = 74 + 1
- PLT = 59 + 1
- If not rerun a «RBC PLT ADJUST» cycle after having previously mixed the latex vial again.
- Adjust the volumes to the target values during the measuring phasis by means of the following
potentiometers (see Diagram 4) :
6 - Adjust sensors
* Needle sensor
Diag.5
Diag.6
- For a current number of steps out of ranges, unloosen the 2 screws 2 (diagram 5) and gently move
the sensor - upward if the current value is too low
- downward if the current value is too high.
- Tighten the screws 2 and rerun a «NEEDLE SENSOR» cycle. Check that the current value is
correct.
This function allows the adjustment of the needle heigth in the RBC chamber. The current number of
steps is 630 and may be adjusted if reproducibilty on RBC/PLT is incorrect.
In this case the value has to be raised 10 steps by 10 steps. Reproducibilty on PLT has to be controled
between each value modification.
- From the SENSOR ADJUST menu, move the cursor to the function <1> "NEEDLE SENSOR" and
press ENTER.
- Push the sampling carriage above the RBC/PLT chamber. Lower the sampling needle until it comes
into contact with the chamber edge as shown on the Diag.7.
Diag.7
- Press ENTER and record the number of step displayed on the LCD. Add 284 to the number displayed and
record the calculated value.
- Move the cursor to the function <2> "NEEDLE" and press ENTER. Enter the calculated value and validate.
* Transfer sensor
Diag.8
- For a current number of steps out of ranges, unloosen the 2 screws 2 (see diagram 8) and gently
move the sensor
NOTE 10 steps correspond to around 1 mm. The target number of steps is 533 + 5
- Tighten the screws 2 (see diagram 8) and rerun a «CARRIAGE» cycle. Check that the current value
is correct.
7 - Vacuum
Move to «1 - VACUUM < >» and press ENTER. Raise the value if the drain has to be faster. Lower
the value if the drain has to be slower. Press ENTER to confirm. Rerun a «CHECK» cycle to control
the drain of the chamber.
Compare to this «CHECK» cycle , the chamber is totally drained during an analysis
NOTE
cycle.
When entering the «temperature adjustment» menu the following sub menus are displayed :
1 - Uncal. temperature
When pressing the key the sensor temperature value uncalibrated is displayed. This value
should be equivalent to the diluent temperature + 2°.
2 - Calibrate :
The temperature must be calibrated according to the diluent temperature :
- Plunge a thermometer directly into the diluent container and leave it for a while until stabilization.
- Run 2 diluent primes («SERVICE» menu, «PRIME» sub menu, «DILUENT» selection).
- Enter the «CALIBRATE» menu. Note the temperature of the diluent and type in the value (if it is
different from the previous on the instrument).
3 - Temperature :
1 - Visible : Press to validate the temperature display.
The value displayed is the calibrated temperature.
2 - Invisible : Press to cancel the temperature display.
(Temperature invisible by default)
IV - Run mode
The instrument must be configurated in the «USER» mode (configurated by default in the «USER «
mode), the «CONTROL» mode intends for a factory use.
V - Reagent pack
This function is used to update the instrument from a bottle mode to a pack mode.
VI - Serial #
VII - Cycle #
Displays the : Startup number since the first use of the instrument.
Stand by number since the first use of the instrument.
CBC number since the first use of the instrument.
The startup, stand by and CBC numbers are adjustable in this menu but
IMPORTANT it is mandatory to keep the initial values (useful for maintenance
schedules).
VIII - Burning
CAUTION : This function which allows the burning of the instrument is intended to a factory
use.
*REMARKS
* CONCERNS
- Sensor replacements
- Needle motion check
- Carriage motion check
- dilution block motion check
- Vacuum/waste syringe motion check
- Valve operation check
- LCD contrast adjustment
- Piercing mechanism check (MICROS CT)
* REQUIRED TOOLS
- Allen key : 2,5
- Felt-pen
* REQUIRED PRODUCTS
- None
* INTERVENTION TIME
- 30 min
* FREQUENCY
- See maintenance chart table.
Page 1/7
RAS 092 A Ind.C 10/03/98
MICROS MECHANIC FUNCTIONS
* PROCEDURE
Enter the "4 - SERVICE" menu and move to "6 - MECHANIC" sub menu.
The following operations are available :
Diag.1
Page 2/7
RAS 092 A Ind.C 10/03/98
MICROS MECHANIC FUNCTIONS
- Manually raise the needle support in the upper position. The "0" should
switch to "1". This indicates the correct operation of the needle sensor
(diagram 1).
If nothing happened try to move the needle up and down again to get the
commutation from "0" to "1".
If the test is still wrong, check the correct connection of the sensor on the
connector J7 of the mother board.
WARNING ! : The new sensor must be installed exactly on the same position.
Proceed as described in RAS 091 A (MICROS OT) or RAS 121 A (MICROS CT)
"Sensor adjustment" to check the correct position of the sensor.
Diag.2
Page 3/7
RAS 092 A Ind.C 10/03/98
MICROS MECHANIC FUNCTIONS
If needed to replace, note its position with a felt-pen, unscrew the 2 screws
and remove the cell holder .
Disconnect the wire from the connector J8 (See diagram 3)
Replace the previous sensor by a new one and reassemble in the reverse
order.
WARNING ! : The new sensor must be installed exactly on the same position.
Proceed as described in RAS 091 A (MICROS OT) or RAS 121 A (MICROS CT)
"Sensor adjustment" to check the correct position of the sensor.
Diag.3
- The motor ------> Check the connection on the board (connector J3)
- The syringes ------> Check the motion up/down
- The sensor ------> Check the connection on the board or replace it.
Page 4/7
RAS 092 A Ind.C 10/03/98
MICROS MECHANIC FUNCTIONS
Diag.4
Page 5/7
RAS 092 A Ind.C 10/03/98
MICROS MECHANIC FUNCTIONS
- The motor ------> Check the connection on the board (connector J4)
- The syringe ------> Check the motion up/down
- The sensor ------> Check the connection on the board or replace it.
Diag.5
Run a "6 - VALVES" cycle. The number of the valves from 1 to 13 is displayed
and each valve is activated once.
Check their correct operations. If a valve is suspected proceed as
described in the procedure : RAS 089 A.
Page 6/7
RAS 092 A Ind.C 10/03/98
MICROS MECHANIC FUNCTIONS
Enter the sub menu "7 - CHG. CONTRAST". Use the arrows to modify the
Enter the sub menu "8 - PIERCING". Follow the instructions : "PLEASE CLOSE
TUBE HOLDER DOOR" : a piercing operation is simulated.
This function gives the sample tube holder position and the heigth of the
needle in its lower position. See RAS 091 A (MICROS OT) or RAS 121 A
(MICROS CT)
NOTE : If the sample tube holder has been removed the following message is
displayed : "ERROR : NO SAMPLE TUBE HOLDER".
If the sample tube holder has been turned in between two piercing
positions (it means that the tube is not in front of the piercing needle) the
following message is displayed : "ERROR : TUBE HOLDER POSITION'".
* REMARKS
Page 7/7
RAS 092 A Ind.C 10/03/98
MICROS DRAIN DETECTION ADJUSTMENT
* CONCERNS
Drain detection sensor adjustment/replacement
* REQUIRED TOOLS
- Voltmeter
- Flat screw driver
- 5ml syringe
* REQUIRED PRODUCTS
- None
* INTERVENTION TIME
- 15min
* FREQUENCY
- On request or 1/year
Page 1/5
RAS 093 A Ind.C 09/03/98
MICROS DRAIN DETECTION ADJUSTMENT
* PROCEDURE
The Vacuum/Waste syringe and the chambers drains are controled by an
infrared sensor located below the syringe.
When a drain problem occurs during a cycle two error messages are
possible :
Diag.1
Page 2/5
RAS 093 A Ind.C 09/03/98
MICROS DRAIN DETECTION ADJUSTMENT
- Connect a voltmeter between the ground TP30 and TP6 (See Diag.2) and
adjust the voltage to 5,5V + 0,5V by means of R6.
Diag.2
Page 3/5
RAS 093 A Ind.C 09/03/98
MICROS DRAIN DETECTION ADJUSTMENT
- Fill the syringe up with diluent and connect it on the syringe waste tubing
instead of the syringe nipple (see Diag.3).
Diag.3
- Push diluent through the sensor and check that the voltage falls down
below 1Volt on the voltmeter.
NOTE : Check the commutation from 5,5Volts to ~ 0Volt once again pushing and
drawning alternately liquid and air through the sensor by means of the
syringe.
* REMARKS
However it is posible to control the correct operation of the sensor by means
of the LED located next to R6 (Upper one).
Indeed the LED should be lit when liquid circulates through the sensor and
switched off with air.
Page 5/5
RAS 093 A Ind.C 09/03/98
MICROS PCB VOLTAGE CHECKS/ADJUSTMENTS
* CONCERNS
* REQUIRED TOOLS
- Voltmeter
- Flat screw driver
* REQUIRED PRODUCTS
- None
* INTERVENTION TIME
- 20min
* FREQUENCY
- None
Page 1/4
RAS 094 A Ind.D 09/03/98
MICROS PCB VOLTAGE CHECKS/ADJUSTMENTS
* PROCEDURE
1 - RBC, PLT, WBC threshold checks/adjustments
- Ground on TP30.
- Adjust the thresholds according to the below chart table :
(See Diag.1)
- Ground on TP31.
- Adjust the motor voltages according to the below chart table :
(See Diag.1)
Page 2/4
RAS 094 A Ind.D 09/03/98
MICROS PCB VOLTAGE CHECKS/ADJUSTMENTS
Page 3/4
RAS 094 A Ind.D 09/03/98
MICROS PCB VOLTAGE CHECKS/ADJUSTMENTS
* REMARKS
The mother board XAA 355 A has been updated to index C according to the
following chart table :
The mother board XAA 355 has been updated to the major index B with the stepper
motor voltage adjustments.
Page 4/4
RAS 094 A Ind.D 09/03/98
MICROS CT TECHNICIAN FUNCTIONS
* CONCERNS
* REQUIRED TOOLS
- Flat screw driver
- thermometer
- Barflex
- Allen key : 2,5
- Voltmeter
* REQUIRED PRODUCTS
- WBC latex : LAD 001 AS
- RBC and PLT latex : LAD 002 AS
- Soft paper
- Flat piece of stiff plastic
* INTERVENTION TIME
- 60 min
* FREQUENCY
- See maintenance chart table.
* PROCEDURE
The arborescence of the «technician functions» is the following :
1 - CAL. PHOTOMETER
1 - WBC ADJUST
1 - VERSION
2 - APERTURE VOLTAGE
2 - # LYM < >
3 - CHECK PRESSURE
3 - # GRA < >
4 - WBC ADJUST
1 - RBC PLT ADJUST
2 - ADJUSTMENTS
1 - NEEDLE SENSOR 3 - MPV < >
6 - SENSOR ADJUST
2 - NEEDLE < >
2 - CHECK
4 - TRANSFER
8 - TECHNICIAN
8 - SAMPLING POSITION
1 - NEEDLE ADJUSTMENT
2 - NO 8 - PIERCING
6 - SERIAL #
1 - VISIBLE
1 - STARTUP # < >
2 - INVISIBLE
2 - STANDBY # < >
7 - CYCLE #
1 - BURNING #
8 - BURNING 2 - BURNING
MICROS CT
3 - CONTROL
Once entering the «SERVICE» sub menu, move to «TECHNICIAN FUNCTIONS» and press .
A specific password (421) is required to enter the sub menus.
Move the cursor by means of and and choose the required menus pressing the
key.
II - Adjustments :
NOTE - If the WBC chamber has been dismantled previously make sure no liquid has flown
in between the spectrophotomer and the chamber.
Clean the inner surfaces of the spectrophotometer as well as the chamber.
Reassemble the assy and tighten the two screws to the following torque : 400mN/m
(see RAS 088 A : Chamber maintenance)
An HGB channel is displayed on the LCD screen representing a voltage value as shown below :
VALUE
237
The HGB photometer calibration must be done 20min at least after the
IMPORTANT instrument has been switched on.
- By means of R97 (See diagram 1) adjust the HGB channel according to the room temperature using
the chart table given on next page.
IMPORTANT This adjustment must be done with the WBC chamber cover installed!!!
HGB
R97 TP27
TP19
APERTURE R91
VOLTAGE
60V +0.1
-
Diag.1
2 - Aperture voltage
- Once entering the «APERTURE VOLTAGE» menu, connect a voltmeter between the ground (TP30)
and TP19.
3 - Pressure check
- If the results are not correct check the O ring and the
Diag.2 tubing watertightness.
4 - WBC adjust
- Put the WBC latex to mix on a Vortex during 5min or shake thoroughly
As the WBC gain is a factory adjustment it is mandatory not to readjust it
IMPORTANT without taking the following precautions :
Carry out previously an autoconcentrated cleaning to make sure of the
cleanliness of the WBC counting circuit.
If necessary clean the WBC chamber aperture as described in RAS 088 A.
Make sure the Latex has been thoroughly mixed before.
- Run a blank cycle to check the cleanliness of the instrument. Install the vial in the sampling position
required for the blood control and turn the tube holder in the compartment inside position.
CAUTION : Make sure the cap of the vial has been removed.
Diag.3
- Wait for several results to be displayed and check the stability of both values.
After one minute the last volume values displayed on the screen are printed out as well as the WBC,
PLT and RBC histograms and the CBC “s results. Check that these printed values correspond to the
following target values :
- LYM = 57 + 1
- GRA = 180 + 2
- If not rerun a «WBC ADJUST» cycle after having previously mixed the latex vial again. By means
of R74 (see diagram 4) adjust the volumes to the target values during the measuring phasis.
THRESHOLD
THRESHOLD R75 350mV+5 APERTURE R91
400mV+5 TP13 - VOLTAGE
TP4 - R74 TP7
GAIN
R87 TP15
R81 GAIN 60V +0.1
-
J17 GAIN
TP11
GND TP29
NU
RBC PULSE PROCESS WBC PULSE PROCESS
Diag.4
- Put the RBC and PLT latex to mix on a Vortex during 5min or shake thoroughly
CAUTION : Make sure the cap of the vial has been removed.
Enter the «RBC PLT ADJUST» sub menu and close the door of the piercing mechanism (Diag.3).
The needle directly delivers the latex sample in the RBC chamber dilution (for a usual analysis cycle
the sample is first delivers to the mixing chamber) and a measuring phasis begins.
- During the cycle measuring phasis (around 1 minute) the Platelet and the Red Blood cell volumes
are displayed on the screen every 3 seconds as shown below :
- Wait for several results to be displayed and check the stability of both values.
After one minute the last volume values displayed on the screen are printed out as well as the PLT
and RBC histograms and the CBC “s results.
- Check that these printed values correspond to the following target values :
- RBC = 74 + 1
- PLT = 59 + 1
- If not rerun a «RBC PLT ADJUST» cycle after having previously mixed the latex vial again.
- Adjust the volumes to the target values during the measuring phasis by means of the following
potentiometers (see Diagram 4) :
6 - Adjust sensors
* Needle sensor
- For a current number of steps out of ranges, unloosen the 2 screws 5 (diagram 5) and gently move
the sensor - upward if the current value is too low
- downward if the current value is too high.
- Tighten the screws 5 and rerun a «NEEDLE SENSOR» cycle. Check that the current value is
correct.
This function allows the adjustment of the needle heigth in the RBC chamber. The current number of
steps is 630 and may be adjusted if reproducibilty on RBC/PLT is incorrect.
In this case the value has to be raised 10 steps by 10 steps. Reproducibilty on PLT has to be controled
between each value modification.
- From the SENSOR ADJUST menu, move the cursor to the function <1> "NEEDLE SENSOR" and
press ENTER.
- Push the sampling carriage above the RBC/PLT chamber. Lower the sampling needle until it comes
into contact with the chamber edge as shown on the Diag.6.
Diag.6
- Press ENTER and record the number of step displayed on the LCD. Add 204 to the number displayed and
record the calculated value.
- Move the cursor to the function <2> "NEEDLE" and press ENTER. Enter the calculated value and validate.
* Transfer sensor
Diag.7
NOTE 10 steps correspond to around 1 mm. The target number of steps is 350 + 5
- Tighten the screws 2 (see diagram 7) and rerun a «CARRIAGE» cycle. Check that the current value
is correct.
* Transfer
Once entering the menu «ADJUST SENSORS» (Selection 6), enter the sub menu «TRANSFER».
Then move the cursor to «2 - CHECK» and press ENTER. The needle moves above the RBC chamber.
Check the centering of the needle over the chamber.
If the position of the needle has to be adjust, move to «1 - TRANSFER < >». The value shown on
the screen is the number of steps from its initial position to the RBC chamber position. The current
value is 585.
Press ENTER and raise the value to move the needle towards the left or lower the value to move the
needle towards the right. 10 steps correspond to around 1mm. Rerun a «2 - CHECK» to control the
centering of the needle.
7 - Vacuum
Move to «1 - VACUUM < >» and press ENTER. Raise the value if the drain has to be faster. Lower
the value if the drain has to be slower. Press ENTER to confirm. Rerun a «CHECK» cycle to control
the drain of the chamber.
Compare to this «CHECK» cycle , the chamber is totally drained during an analysis
NOTE
cycle.
8 - Sampling position
Two states 0/1 are possible for the 3 switches (see diag
11) :
- Switched OFF : 1
- Switched ON : 0
The code «0 0 0» means that the tube holder has been turned in between two sampling
NOTE positions. The code «1 1 1» means that the tube holder has been removed.
Four of the positions are used on the actual tube holder : (See Diagram 11)
Diag.12
The number of step value which is recorded is the maximum value found to the bottom
NOTE
of the position less 3 mm (Diag.12).
It is possible to enter directly the required number of steps for each sampling position. Proceed as
following :
Enter the sub menu that corresponds to the number of the needle : the current number of steps is
displayed. Enter the new value.
Increase the number of steps to move the needle deeper or decrease the value to raise the needle.
Confirm the new value with . The minimum and maximum step values are as follow :
NUMBER OF STEPS
NEEDLE
MINI. DEFAULT MAXI.
1 1 762 782
2 1 630 650
3 1 860 880
4 1 925 945
5 1 1005 1025
Table 3 6 1 580 600
* Piercing
The piercing function allows to check the piercing operation on each sampling position. From the
SAMPLING POSITION menu, move the cursor to the function <8> Piercing and press ENTER.
Select the required position on the tube holder and close the sample door. The piercing cycle is
carried out and the number of steps for this position is displayed :
NEEDLE 2
CURRENT : 635 STANDARD : 630
When entering the «temperature adjustment» menu the following sub menus are displayed :
1 - Uncal. temperature
When pressing the key the sensor temperature value uncalibrated is displayed. This value
should be equivalent to the diluent temperature + 2°.
2 - Calibrate :
The temperature must be calibrated according to the diluent temperature :
- Plunge a thermometer directly into the diluent container and leave it for a while until stabilization.
- Run 2 diluent primes («SERVICE» menu, «PRIME» sub menu, «DILUENT» selection).
- Enter the «CALIBRATE» menu. Note the temperature of the diluent and type in the value (if it is
different from the previous on the instrument).
3 - Temperature :
1 - Visible : Press to validate the temperature display.
The value displayed is the calibrated temperature.
2 - Invisible : Press to cancel the temperature display.
(Temperature invisible by default)
IV - Run mode
The instrument must be configurated in the «USER» mode (configurated by default in the «USER «
mode), the «CONTROL» mode intends for a factory use.
V - Reagent pack
This function is used to update the instrument from a bottle mode to a pack mode.
VI - Serial #
VII - Cycle #
Displays the : Startup number since the first use of the instrument.
Stand by number since the first use of the instrument.
CBC number since the first use of the instrument.
The startup, stand by and CBC numbers are adjustable in this menu but
IMPORTANT it is mandatory to keep the initial values (useful for maintenance
schedules).
VIII - Burning
CAUTION : This function which allows the burning of the instrument is intended to a factory
use.
*REMARKS
* CONCERNS
- Changing the instrument language from the V1.2 version.
* REQUIRED TOOLS
- Pair of pliers.
* REQUIRED PRODUCTS
- None
* INTERVENTION TIME
- 5 min
* FREQUENCY
- On request only.
Page 1/4
RAS 122 A Ind.B 09/03/98
MICROS CHANGING THE INSTRUMENT LANGUAGE
* PROCEDURE
- Switch off the instrument.
- Remove both jumpers from E1 (on mother board, see diagram 1).
E1
Diag.1
Page 2/4
RAS 122 A Ind.B 09/03/98
MICROS CHANGING THE INSTRUMENT LANGUAGE
JUMPER CONFIGURATION ON E1
ENGLISH
FRENCH
SPANISH
GERMAN
Page 3/4
RAS 122 A Ind.B 09/03/98
MICROS CHANGING THE INSTRUMENT LANGUAGE
* REMARKS
Page 4/4
RAS 122 A Ind.B 09/03/98
MICROS REAGENT PACK
* CONCERNS
- Replacement of the waste connector O ring
- Replacement of the diluent/clean/lyse O rings
* REQUIRED TOOLS
- Pair of pliers
* REQUIRED PRODUCTS
- None
* INTERVENTION TIME
- 15 min
* FREQUENCY
- O ring replacements : 1/year
Page 1/3
RAS 123 A Ind.B 09/03/98
MICROS REAGENT PACK
* PROCEDURE
- Disconnect the pack if this one is still connected.
Diag.1
Page 2/3
RAS 123 A Ind.B 09/03/98
MICROS REAGENT PACK
* REMARKS
Page 3/3
RAS 123 A Ind.B 09/03/98
TRANSFORMATION OF A MICROS
MICROS "BOTTLE" INTO MICROS "PACK"
* CONCERNS
This procedure allows the technician to transform a MICROS OT or CT "bottle" version into a MICROS
OT or CT "pack" version.
1 - This procedure is available only for MICROS having a serial number above
WARNING !
600 as some frame parts are not compatible below this number.
2 - If the MICROS is operating with a program version below V1.3, a specific
password is required to carry out the instrument setup at the end of the
procedure. If this password is unknown, the procedure has to be carried out
by authorized technician only.
* REQUIRED TOOLS
- Hexagonal key 2.5.
- Large and flat screwdriver.
- Cutters.
- Pliers.
- Ruler.
* REQUIRED PRODUCTS
- MICROS reagent packs.
* INTERVENTION TIME
1hour 30minutes
* FREQUENCY
On request
* PROCEDURE
Diag.1
Diag.2
Diag.3
Diag.4
- Cut the remaining part of the lyse tube to a length of 480mm. This tube will be used for the diluent
output of the reagent pack.
- Disconnect the blue tube from the detergent straw to the valve <4> input #2 and remove the tube.
- Connect a tube 2.29x1100 to the valve <5> input #1. Route the tube from the valve to the top of the
reagent compartment using the previous route of the reagent tubes.
Diag.6
Diag.8
IMPORTANT Cut the ends of the used tubes to have a better tigthening on the connectors.
- Install a MICROS reagent pack in its location according to the instructions given on the pack
package. Check for leaks.
- From the main menu, move the cursor to the function <4> SERVICE and press ENTER. Move the
cursor to the function <8> TECHNICIAN and press ENTER. Enter the technician password (421) and
press ENTER.
- For MICROS instrument operating with a program version below V1.3, a specific password is required.
Move the cursor in front of the hidden function and press ENTER. Enter this password and press
ENTER. Move the cursor next to the function <2> PACK and press ENTER. Move the cursor next to
the function <2> YES and press ENTER. Return to the SERVICE menu with the <ESC> key.
- For MICROS instrument operating with a program version V1.3 or above, move the cursor next to
the function <5> PACK and press ENTER. Move the cursor next to the function <2> YES and press
ENTER. Return to the SERVICE menu with the <ESC> key.
- From the SERVICE menu, move the cursor next to the function <3> PRIME and press ENTER.
Move the cursor next to the function <1> CHANGE PACK and press ENTER.
The CHANGE PACK function allows a complete priming of the reagent lines and
NOTE allows the initialization of the cycle counter. The PRIME function is recommended
only to prime the reagent lines.
- Follow the displayed instructions and check for leaks. Run an autocleaning cycle from the SERVICE
menu, then several blank cycles.
* REMARKS
* ATTACHED DOCUMENTS
* CONCERNS
- This procedure allows the technician to check and to lubricate the air syringe and the liquid syringe
of the ABX MICROS.
* REQUIRED TOOLS
- Hexagonal keys.
* REQUIRED PRODUCTS
* INTERVENTION TIME
- 30 minutes.
* FREQUENCY
- Once a year.
- Absorbant paper.
* PROCEDURE
Diag.2
Diag.3
Place some absorbant paper at the instrument rear connections (diluent and
NOTE
waste) as some liquids may come out when the syringe is pushed.
Diag.4
Place some absorbant paper at the instrument rear connections (diluent and
NOTE
waste) as some liquids may come out when the syringe is pushed.
- Re-install the instrument cover, reconnect the waste and diluent tubes, reconnect the power cable.
*REMARKS
* CONCERNS
This procedure allows the technician to set up the LX300 printer according to the user's needs.
* REQUIRED TOOLS
- none.
* REQUIRED PRODUCTS
- none
* INTERVENTION TIME
- 15 minutes
* FREQUENCY
- on request
- none
* PROCEDURE
NOTE On the MICROS, it is necessary to select the printer format RESERVED 1 of the
"PRINTER" menu (function 4) accessible through the "OPTIONS" menu (function 5 of
the main menu) then "RESULTS" (function 1).
Diag.1
1 - Printer configuration :
The printer configuration is printed out when pressing the key <FONT> when the printer is switched
ON. The configuration used for the MICROS is the factory configuration :
Diag.2
Each parameter can be modify by the corresponding parameter chart. Each chart is accessible using
the keys <PAUSE>, <FONT> and <LF/FF> according to the control LED combinaisons :
Diag.3
Diag.4
The control pannel keys allow the user to set up the main functions of the printer : paper advance,
paper ejection, and font selection. Control LEDS indicate the printer status.
LED PAUSE : The orange LED PAUSE lights when the printer stops printing. During each power ON,
this LED blinks for few seconds and 4 audible beeps occur. When the printer runs out of paper, the
LED blinks and 3 audible beeps occur. This LED lights also when the paper is in its tear off position.
When a problem occurs, this LED lights ON and 5 audible beeps occur.
LEDS FONT 1 and FONT 2 : These 2 green LEDS indicate the selected font. Refer to the printer
user's manual to select the font.
Key FONT : During normal operation, the FONT key allows the font selection. For each pressure on
this key, the selection is modified. Refer to the printer user's manual to select the font. When this key
is pressed during the printer power ON, the printer setup menu is entered.
Key LF/FF : During normal operation, a quick pressure on this key allows a ligne feed of the paper.
Keep the pressure on this key to feed a whole page. This key can be used to load or eject the paper.
When this key is pressed during the printer power ON, the printing test starts.
Key PAUSE : When this key is pressed during the printing, the printout stops. Press again on this key
to restart the printout.
PAPER PARK : If Z folded paper is used, the paper can be driven to its parking position when
pressing simultaneously on the keys LF/FF and FONT.
MICRO ADJUST : This function allows to adjust the loading paper position. See the user's manual
for details.
4 - Printer description :
Diag.5
Diag.6
*REMARKS
* CONCERNS
This procedure allows the technician to transform a MICROS OT or CT "pack" version into a MICROS
OT or CT "bottle" version.
* REQUIRED TOOLS
- Hexagonal key 2.5 or torx keys when applicable.
- Cutters.
- Pliers.
- Ruler.
* REQUIRED PRODUCTS
- MICROS reagent bottles.
* INTERVENTION TIME
1hour 30minutes
* FREQUENCY
On request
* ATTACHED DOCUMENTS
* PROCEDURE
Diag.1
Diag.2
Diag.5
- Remove the tube 2.29x510 from the detergent output of the reagent pack to the input #2 of the
valve <4>. Replace it by a blue tube 2.29x1100 from the output #2 of the valve <4> to the detergent
straw.
- Route the tubes from the valves to the top of the reagent compartment using the previous route of
thereagent tubes.
IMPORTANT Cut the ends of the used tubes to have a better tigthening on the connectors.
Diag.7
- Install the reagent straws at the end of each tube, waste tube included.
- Install the label on the reagent tube next to the straws : MINICLEAN on the blue tube.
- Place the diluent container at the same level than the instrument, install the MINILYSE and
MINICLEAN into the reagent compartment. Install the bottle stoppers and place the reagent straws.
- Place the waste container below the level of the instrument. Place the rubber stopper and the waste
straw into the container.
Diag.8
Diag.9
- From the main menu, move the cursor to the function <4> SERVICE and press ENTER. Move the
cursor to the function <8> TECHNICIAN and press ENTER. Enter the technician password (421) and
press ENTER.
- For MICROS instrument operating with a program version below V1.3, a specific password is required.
Move the cursor in front of the hidden function and press ENTER. Enter this password and press
ENTER. Move the cursor next to the function <2> PACK and press ENTER. Move the cursor next to
the function <1> NO and press ENTER. Return to the SERVICE menu with the <ESC> key.
- For MICROS instrument operating with a program version V1.3 or above, move the cursor next to
the function <5> PACK and press ENTER. Move the cursor next to the function <1> NO and press
ENTER. Return to the SERVICE menu with the <ESC> key.
- From the SERVICE menu, move the cursor next to the function <3> PRIME REAGENTS and press
ENTER. Move the cursor next to the function <1> ALL REAGENTS, press ENTER and check for
leaks. Run an autocleaning cycle from the SERVICE menu, then several blank cycles..
NOTE The ALL REAGENTS function allows a complete priming of the reagent lines.
* REMARKS
* CONCERNS
This procedure allows the technician to set up the barcode reader using the
barcode programmation labels for the PANASONIC reader.
* REQUIRED TOOLS
None.
* REQUIRED PRODUCTS
None.
* INTERVENTION TIME
10 minutes.
* FREQUENCY
On request.
Page 1/4
RAS 163 A Ind.A 06/03/98
MICROS BARCODE READER SETUP
* PROCEDURE
- Make sure that the BARCODE reader is properly fitted in the DIN plug of the RS
adaptor wiring.
- Make sure that the MICROS mother board XAA 355 B has an index E or above.
- Switch on the instrument and within the 15 first seconds, read from the top to the
bottom the 3 barcode labels located on the top left of the next page of this
procedure. The audible beep occurs after each reading.
- After the 3 labels, the audible signal beeps 5 times, read all the labels from top to
bottom and from left to right.
- When the last label is read, the signal beeps 5 times in order to indicate the end
of the setup. Check on the test labels located on the last page that the reading is
correct.
* NOTE : In case of difficulties to read 2 consecutive 0, move away the barcode reader
from the page after the first reading and then read the second 0.
Page 2/4
RAS 163 A Ind.A 06/03/98
MICROS BARCODE READER SETUP
No STF
(5 beeps)
1 bit stop Ø
Code identifier : no Ø
5 Beeps
Page 3/4
RAS 163 A Ind.A 06/03/98
MICROS BARCODE READER SETUP
READ TEST AFTER SETUP
EAN 8 C39
1234567 TEST
EAN 8 2/5
9076543210
1234567
EAN 13 CODABAR
543210
123456789012 543210
EAN 13 C 128
98765
98765
123456789012
Page 4/4
RAS 163 A Ind.A 06/03/98