Vous êtes sur la page 1sur 4

PRESSURE TRANSDUCERS TECHNICAL NOTE 2

SPECIFYING A PRESSURE TRANSDUCER –


HOW TO SELECT PRESSURE RANGE AND DEVICE TYPE
Mitch Berkson, Original, Proof pressure is specified as a multiple of Suppose that there is a requirement to mea-
the upper limit of the device’s operating (e.g., sure pressures in the range 0-50 psia and a
Dave Field, Update, 2007
1.5X for a 0-100 psi device would mean a 0-100 psia sensor has been selected. There
When choosing a pressure transducer for a proof pressure of 150 psi). are two issues which affect the theoretical
particular application, usually the first accuracy.
The typical testing performed by Sensata to
question which arises is: “For what pressure
verify proof pressure is for the pressure trans- First, the device accuracy (static error and
range should the transducer be rated?” This
ducer to withstand 1000 cycles of exposure total error bands) is specified as a percentage
simple question begets a bevy of related
to the proof pressure, at room temperature, of full span (%FS) and, in this example,
ones, namely: “What is the pressure range
holding the pressure for 30 seconds each is ±0.75% FS. If the full span of the
in which the device typically operates? Does
cycle, without degradation of performance. application is only 0-50 psia the accuracy,
the device occasionally need to measure
as a percentage of the application’s full span
pressures outside this range? What pressure Burst Pressure would be (100psia / 50psia)* ±0.75% FS =
must the device withstand and still oper-
Burst pressure is the maximum pressure to ±1.5% FS. In general, the accuracy in an
ate within specification when returned to which a device may be exposed for two application is given by:
its normal range? What pressure must the minutes without rupture that results in
device withstand without failing even if it Aapp = Adev * (FSdev/FSapp)
component separation from the transducer
will function properly after returning to the or application fluid leakage. It is not where: Aapp = application accuracy
normal operating range?” guaranteed that the device will function Adev = device accuracy
Closely related to the pressure specification within specification when returned to its FSdev = device full span
is the type of device to choose - absolute, normal operating range after being exposed FSapp = application full span
gage or sealed gage. The following sections to a pressure above its proof pressure (even
if that pressure is below its burst pressure). The second aspect of using a pressure sensor
will first address pressure selection and then
Like proof pressure, burst pressure is speci- for a smaller span than its rating is that of
device type.
fied as a multiple of the upper limit of the analog to digital converter (ADC) resolu-
Pressure Specifications device’s operating range. tion. In many systems, an ADC is used to
condition the output of the pressure sensor.
Three pressure specifications typically appear An application might use a 10 bit ADC
on the data sheet. Choosing Device Pressure referenced to 0V and 5V. A 10 bit ADC has
Proof pressure and burst pressure should 1024 (210) possible outputs. Since the usual
Normal Operating Range be chosen based on potential over-pressure pressure sensor output is from 0.5V to 4.5V,
The device pressure range is the normal in the operating environment or regulatory only 80% (4V/5V) of the ADC range will
operating region for the device. This is the requirement. be used. This reduces the number of pos-
region in which the data sheet specifications The highest system accuracy is attained only sible ADC outputs to 819 (80*1024). Each
are valid. least significant bit change corresponds to
when the normal operating pressure of the
a voltage change of 4.9mV (4V span/818).
pressure transducer is matched to the
Proof Pressure Since the output of the ADC is discrete,
application. In order to understand the ef- it may be in error by ±1/2LSB, or in this
Proof pressure is the maximum pressure to fect of using a device to measure pressures example, ±2.4mV. The error due to AC
which a device may be exposed after which which are in a smaller range than the devices quantization is ±0.06% (2.4mV/4V). Note
it will return to its normal operating pres- full span range, it is useful to examine that no mention has been made yet of the
sure range and perform within specification. an example. input pressure range of the pressure sensor.
PRESSURE TRANSDUCERS TECHNICAL NOTE 2
SPECIFYING A PRESSURE TRANSDUCER –
HOW TO SELECT PRESSURE RANGE AND DEVICE TYPE

ducers to an application, there will be an up-


If a 0-50 psia application uses a 0-100 psia
sensor, the output range will be between per limit where the resolution of the sensor
Device Type – Gage or Sealed
0.5V and 2.5V for a range of 2V. In this starts to affect the accuracy. This is the case The pressure sensing element in Sensata
case quantization error is 2.4VmV/2V = with a large transducer pressure range and the pressure transducers is a pair of parallel
±0.12% FS. In general, the quantization small application range. This large mismatch plates which form a capacitor. One plate is
error of an application is given by: will result in a slight change of application pres- fixed to a ceramic diaphragm which flexes in
sure causing an out of specification output. response to pressure changes. The other
1
Eq(±%FS) = x100 It is also important to note that if other plate is attached, with a rigid glass seal, to a
FSapp
mechanical aspects of a higher full scale pres- ceramic substrate which is insensitive to
2 x (2n – 1) x
(ADCHI – ADCLO) sure transducer are desirable (such as proof pressure changes. As the pressure varies, the
pressure) then a device can be configured diaphragm flexes and the distance between
Where: Eq = quantization error with this sense element and calibrated down the capacitor plates changes.
n = ADC bits to include a higher degree of accuracy with a
In typical transducers, only one side of the
lower pressure range.
FSAPP = application full span diaphragm is accessible to the application
ADCHI = ADC high reference
voltage Measuring Pressures and the other inaccessible side, is perma-
nently referenced to some other pressure.
ADCLO = ADC low reference Outside Normal Device Range The pressure on the inaccessible side can fall
voltage
In some applications, there is a requirement into one of two categories: gage (or vented)
The quantization error is a function solely to detect pressures above or below the nor- and sealed. In a gage sensor, the inaccessible
of the ADC resolution and reference mal operating range of the pressure sensor. side of the diaphragm is vented to the atmo-
voltages, and the full span voltage range of One example is if the system requires alarms sphere. In a sealed sensor, there is no way for
the device as used in the application. By (“diagnostic bands”)at points outside the the atmosphere to move in or out of contact
using as much of the pressure sensor’s input normal operating range. In this situation, with the inaccessible side of the diaphragm.
pressure range as possible, quantization it may be undesirable to use a sensor with a
error is minimized. range encompassing the alarm points since It is possible to manufacture a sealed sensor
this will reduce the resolution in the normal with different pressures sealed in. Typically
These two aspects of theoretical accuracy for
operating range in order to accommodate the sealed pressure is atmospheric. Depend-
a pressure transducer should be kept in mind.
infrequent incursions to the alarm pressures. ing on the associated electronics, a transduc-
It is important to note that this discussion is
more about resolution than actual transducer Application note AN3 - Useful Pressure er can be built to operate within a specified
accuracy when in application. Sensata trans- Transducer Performance Outside Normal accuracy regardless of the sealed pressure.
ducers are always calibrated to have as little Operating range describes how far outside The two transducer types are illustrated
as possible error and 100% checked for the normal operating range the pressure schematically in Figure 1.
accuracy tolerances through out the entire transducer will function and its performance
operating pressure range. In applying trans- in these regions.

Application Pressure Application Pressure

Diaphragm Diaphragm

Vent to
Atmosphere
Sealed
Area

Substrate Substrate

Figure 1. Different types of pressure transducers


Figure 2. Gage vs. sealed device for measuring open or closed pressure vessels
PRESSURE TRANSDUCERS TECHNICAL NOTE 2
SPECIFYING A PRESSURE TRANSDUCER –
HOW TO SELECT PRESSURE RANGE AND DEVICE TYPE

Pressure Scales surface exposed to the atmosphere. When Depending on the pressure of the applica-
the atmospheric pressure increases, the gage tion, the error contributed by changes in the
Since the measurement of pressure is always ambient atmospheric pressure may be negli-
device will indicate a lower measured pres-
differential (i.e., the pressure being mea- gible. In the example above, if the full span
sure since it is measuring with respect to
sured is always compared to some other were 1,000 psia, the error contributed by
atmospheric pressure.
pressure), the nomenclature used to describe the altitude change would be 2.5 psi/1,000
a pressure measurement includes informa- How do these different operational character-
psi * 100% = 0.25% FS.
tion about the pressure reference. istics influence the choice of device type? The
only way to be unaffected by altitude or baro- In some systems (e.g., refrigeration), the
The following abbreviations are commonly metric pressure changes with a single sensor is fluid being measured is shielded from the
to use a gage device. Conversely, if the applica- effect of atmospheric pressure by the system
used:
tion requires measurement of the environmen- construction If in the previous example the
psi = pounds per square inch
container were closed so that changes in
psia = psi absolute tal pressure, a gage device cannot be used.
(i.e., relative to vacuum) atmospheric pressure had no effect on the
psig = psi gage Altitude Example fluid, then there would be no error due to
(i.e., relative to atmospheric the atmospheric pressure changes even if a
Ambient atmospheric pressure decreases
pressure) sealed device were used. In this situation,
with increasing altitude by approximately
using a gage device would cause an error
0.5 psi for every 1,000 foot increase.
Psi (without the “a” or “g”) should generally since atmospheric pressure changes experi-
At the typical mountain pass in Colorado
be avoided as it is unclear to what pressure enced by the vented, inaccessible side of the
of 10,000 feet, the ambient pressure is 14.7
the measured pressure is referenced. An diaphragm would not be offset by pressure
psia - 5 psia = 9.7 psia. changes on the application side. See illustra-
exception is when a pressure increment is
When a sealed sensor (either absolute or tion in Figure 2.
being discussed - since a 10 psia pressure
sealed gage) with its pressure port open
change is the same as a 10 psig pressure
change, the term psi could be used unam-
to the atmosphere is brought to higher Vent Path
elevations, the output will decrease because All gage devices must have a vent path to
biguously in this context.
the pressure applied to its diaphragm is atmospheric pressure. In competitive de-
Depending on the pressure of the applica- being reduced. The error due to altitude is vices, the vent path is wide open to dirt and
tion, the distinction between psia and psig calculated by finding the pressure change moisture and is a frequent cause of field
may be negligible. This is generally true in for the altitude range and dividing by the failures. In Sensata gage devices, the internal
high pressure applications. The difference full scale range of the sensor to get error in workings of the sensor are protected from
in measurement between an absolute and a % FS. moisture and dirt by the connector as well
gage device depends on atmospheric pres- as some patented features in the sensing ele-
Example (note: psi is used in some
sure. (At sea level, is approximately ment. The vented air travels from the sensing
places in this example when pressure
14.7 psi.) For a 1000 psi full span application element, through the standard electrical
changes or differences are being considered):
this is a 1.5% FS difference. wire and exits at the end of the wire. The
A 0-50 psia sealed sensor is being used to
wire end must remain in a clean environ-
measure the height of an open container
Performance of the of liquid by measuring the pressure at the
ment to avoid plugging of the vent (i.e.,
not plugged with solder or water or oil).
Transducer Types bottom of the tank. The system is moved The size and length of the leads for venting
In order to understand how the two device from sea level to Denver, Colorado at 5,000 affect the response time of the sensor. For a
types differ in performance, it is necessary feet. So 5,000’ * (0.5 psi/1,000’) = 2.5 psi 10 psi atmospheric pressure drop, a device
to change their environment while keeping change. 2.5 psi/50 psi * 100% = 5% FS with standard 18 AWG wire leads will have a
the applied pressure constant. altitude error. This error is due to the response time of less than one minute. This
decreased pressure of the atmosphere on the is significantly faster than the atmospheric
What is the effect of increasing atmospheric
liquid. If a gage part were used, the decrease pressure changes unless the sensor is rapidly
pressure on each of the device types? The
in pressure on the liquid would be compen- changing altitudes. This response time does
gage device is the only one which changes in
sated by a decrease in pressure on the side of not affect the sensor’s output response time
response to an atmospheric pressure change
the diaphragm exposed to the atmosphere to changes in working fluid pressure.
since it is the only one with a diaphragm
and the error would be zero.
© Copyright Sensata Technologies 2007
PRESSURE TRANSDUCERS TECHNICAL NOTE 2
SPECIFYING A PRESSURE TRANSDUCER –
HOW TO SELECT PRESSURE RANGE AND DEVICE TYPE

Sealed Gage Device A sealed gage part behaves the same as an Rule 2: If the pressure being measured
absolute; its output is affected by changes in changes when atmospheric pressure changes
Some users of sealed sensors find it more atmospheric pressure. They differ in output and that should result in a change in device
convenient for the transducer to output a by a 14.7 psi offset as shown in Figure 3. output, use a sealed (absolute or sealed gage)
value of 0 psi at atmospheric pressure (14.7 device.
psia) rather than at 0 psia. For these cases an
absolute transducer is calibrated to output
Selection Summary Rule 3: If the pressure being measured does
The following three rules can be used to de- not change when atmospheric pressure
0 psi at a standard sea level pressure of 14.7
cide whether a sealed or gage device should changes, use a sealed device (absolute or
psia. To make it clear that an absolute refer-
be used: sealed gage).
ence pressure is being used but with a gage
The flow chart in Figure 4 can be used to
scale calibration, the industry uses the term Rule 1: If the pressure being measured
decide whether a gage, sealed gage or
“sealed gage” which is denoted by psis. changes when atmospheric pressure changes
absolute part should be used in an spplication.
AND that is considered an error AND that
error is too high, use a gage device
Absolute (psia) scale
0 14.7 114.7 Is the effect of
atmospheric pressure on No Gage
sensor output within Sensor
application spec?

0 100 Yes
Gage (psig) and sealed gage (psis) scale

Want low end of No Sealed sensor (low


sensor output level sensor output at
range at 0 psia? 0 psis = 14.7 psia)
Figure 3. Absolute vs. Gage and Sealed
Gage Pressure Scales Sensata Pressure Sensors
Yes

Company Description Absolute sensor


(low level sensor
output at 0 psia)
Sensata Technologies is one of the world’s leading suppliers of sensing, electrical protection,
control and power management solutions across a broad array of industries and markets.
Figure 4. Transducer Type Selection Flow Chart
Our products improve safety, efficiency and comfort for millions of people every day in
automotive, appliance, aircraft, industrial, military, heavy vehicle, heating, air conditioning,
data, telecommunications, recreational vehicle and marine applications.

For more information, please visit our web site at: www.sensata.com/pressure or call toll free at 1-888-438-2214
or email us at sensors@sensata.com.

Important Notice: Sensata Technologies (Sensata) reserves the right to make changes to or discontinue any product or service identified in this publication without notice.
Sensata advises its customers to obtain the latest version of the relevant information to verify, before placing any orders, that the information being relied upon is current.
Sensata assumes no responsibility for infringement of patents or rights of others based on Sensata applications assistance or product specifications since Sensata does not
possess full access concerning the use or application of customers’ products. Sensata also assumes no responsibility for customers’ product designs.

Printed in U.S.A., Reprinted February, 2008

Vous aimerez peut-être aussi