Required material
Terminal/Controller MIQ/TC 2020 XT
Power supply module MIQ/PS
MIQ/CR3 combination output module
Sensor FDO 700 IQ
Sensor SensoLyt 700 IQ
Content
1 Mounting system ......................................................................................................... 5
1.1 Stacked mounting of MIQ modules ............................................................................... 5
1.2 Switching on a terminator ............................................................................................ 7
1.3 Installing a sensor ....................................................................................................... 9
1.4 Installing a terminal ................................................................................................... 11
3 Calibration ................................................................................................................ 33
3.1 FDO 700 IQ Check mode on air ................................................................................. 33
3.2 Change FDO cap....................................................................................................... 35
3.3 Maintenance condition ............................................................................................... 37
3.4 Calibrating a SensoLyt 700 IQ ................................................................................... 39
3.5 Calibration history of a SensoLyt 700 IQ .................................................................... 41
1 Mounting system
Solution:
The following example shows the stack expansion forwards. For details refer
to the system operating manual, section Installation.
1. Remove the covers:
Solution:
Detailed description: System operating manual, chapter Installation
SN terminator switch
Solution:
10
11
Solution:
12
2 Basic settings
13
Solution:
Steps Detailed description
14
Set the system date and system time to the current time.
15
Solution:
Steps Detailed description
16
17
Solution:
Either the altitude of the system site or the absolute air pressure
must be known for this setting. Both values are linked through the
barometric altitude formula.
18
Assign the name "O2 basin 1" to the FDO 700 IQ sensor
Assign the name "Inlet" to the SensoLyt 700 IQ sensor
19
Solution:
Steps Detailed description
20
R1
R2 pH LL
R3 pH UL
C1 O2
C2
C3 Temp. pH
21
Solution:
Steps Detailed description
22
23
Solution:
The measuring mode is indicated in the measured value display
(example: mg/l O2 = D. O. concentration). Alternatively, you can
switch to the setting menu of the sensor and view the setting as
follows:
24
Link all available relay outputs and current outputs with sensors
25
Solution:
Steps Detailed description
26
27
Solution:
Steps Detailed description
Special features
The entry of the end value is not accepted. The difference between the
start value and end value is too small (minimum difference: 5 % of the
measuring range)
28
29
Solution:
Steps Detailed description
30
With the aid of a SensoLyt 700 IQ, the pH value should be regulated
with two relays (dosing of HCl and
NaOH)
The regulation is carried out via frequency controllers that are
directly connected to the feed pumps.
NaOH should be dosed in the range pH 2 to 4. The relevant pump
should receive 0...120 pulses/minute in this range.
HCI should be dosed in the range pH 6 to 8. The relevant pump
should receive 10...100 pulses/minute in this range.
31
Solution:
Parameters to be set
Relay 1 Relay 2
Relay function Frequency output Frequency output
Measured parameter Main measured parameter Main measured parameter
Start value pH 4 pH 6
End value pH 2 pH 8
Frequency (f) min. 0 1/min 10 1/min
Frequency (f) max. 120 1/min 100 1/min
Steps Detailed description
Open the Settings menu:
Any measured value display: press “S”
1. Set the measured parameter,
measuring range, etc. of the sensor:
Settings of sensors and diff. sensors menu
Sensor operating manual
Column, Measuring range
Select the sensor General operating principles:
Make the required adjustments / settings System operating manual or terminal
according to the general operating operating manual
principles
Note: This step can be skipped if Exercise
2.6 was already done.
1. Link the sensor with the relay output:
Return to the Einstellungen/Settings menu
with “ESC” Operating manual of the output module
Menu, Settings of outputs and links General operating principles:
Column & Select the relay output (designation System operating manual or terminal
Rx) and press “OK” operating manual
Select the sensor from the list and press “OK”
Note: This step can be skipped if Exercise
2.7 was already done.
1. Link the sensor with a second relay
output as described in step 3
1. Set both relay outputs separately: Operating manual of the output module,
Press “ESC” to return to the chapter Basic information on relay
Settings of outputs and links menu functions
Column, Feature Operating manual of the output module,
Select the linked relay output chapter Settings of relay outputs
Select the Relay function line: Frequency General operating principles:
controller System operating manual or terminal
Make further adjustments / settings according operating manual
to the general operating principles
32
3 Calibration
3.1 FDO 700 IQ Check mode on air
33
Solution:
check on water saturated air
34
35
Solution:
Steps Detailed description
Hint:
It is not needed to disconnect the FDO 700 IQ or to switch off the
system before you exchange a FDO cap.
36
37
Solution:
Steps Detailed description
38
39
Solution:
Before calibrating ensure that the required calibration procedure is
set (2-point calibration procedure CAL TEC AUTO or CAL CON 2P
). If necessary, correct the setting in the setting table of the sensor.
40
41
Solution:
Steps Detailed description
42