Vous êtes sur la page 1sur 4

April 23, 1996 SERVICE NOTE Erratum Page 1 of 1

IMAGEUR 16 (6”) FOR STENOSCOP 2 SN 4011A


Rev. 1 2122943–100

IMPORTANT
READ THIS FIRST

ERRATUM FOR SN 4011A


NEW 16 (6”) II TUBE FOR STENOSCOP 2
Before using the enclosed SN, note the following changes since the previous version:

REV DATE REASON FOR CHANGE

0 Dec 30, 1994 Initial release.

1 Apr 23, 1996 The original Service Note, SN 4011 did not cover all existing models of the
Imageur 16. SN 4011A gives the procedure to follow in the event of II tube
replacement according to Imageur model.

LIST OF EFFECTIVE PAGES NUMBER SIZE REVISION


Page count Page count 2122943–100TPH A4 1
Erratum 1
English text 3
Copyright 1996 by General Electric Co.
April 23, 1996 SERVICE NOTE Page 1 of 3

IMAGEUR 16 (6”) FOR STENOSCOP 2 SN 4011A


Rev. 1 2122943–100

NEW 16 (6”) II TUBE FOR STENOSCOP 2

APPLIES TO:
All Hyperlux 16 (16-cm/6-inch) and Imageur 16 (16-cm/6-inch) models.

PURPOSE:
The II tube TH9448 F H506, Part No. 45066699 has been replaced by the II tube TH9449 HP H506,
Part No. 2120442.
Special features of II tube TH9449 HP H506 are:
 All-metal tube.
 No G3 electrode (no field diameter adjustment).
 The G2 electrode voltage range is greater (1.8 kV to 2.3 kV, instead of 0.8 kV to 1.2 kV for the
TH9448 F H506).

PROBLEM:
The voltage range of the G2 electrode is no longer compatible with the voltage range provided by the
TH7191 high voltage power supply (HVPS) (Part No. 58079229) or with the Brandenburg HVPS (Part
No. 45066585).

SOLUTION:
This Service Note describes the procedure to follow according to the model of Hyperlux 16 or
Imageur 16 considered.

High Volt Power Supply II Tube


Image r
Imageur
Old New Old New
Hyperlux 16 Brandenburg 821 Brandenburg 821–1 TH 9448 F H506 TH 9449 HP H506

Imageur 16 PUT Brandenburg 833 Thomson TH7191 TH 9448 F H506 TH 9449 HP H506
Brandenburg 833–2
Thomson TH7191

Imageur 16 CCD Thomson TH7191 Thomson TH7191 TH 9449 HP H506 TH 9449 HP H506

Copyright 1996 by General Electric Co.


April 23, 1996 SERVICE NOTE Page 2 of 3

IMAGEUR 16 (6”) FOR STENOSCOP 2 SN 4011A


Rev. 1 2122943–100

A. Hyperlux 16 for Stenoscop with Monicon 1 or Monicon 2


These devices use HVPS type Brandenburg, model 821 (C381367), 38 V DC supplied. In the event of
an II tube replacement, this HVPS cannot drive the new II tube TH 9449 (2120442) because the range
of the G3 electrode is not wide enough.
A new HVPS has been redesigned by Brandenburg as Model 821–1 (Part No. 2131405). This HVPS
is also 38 (dc) supplied, and replaces the Model 821 (C381367). It is compatible with both old and new
II tube models.
Before the HVPS replacement, follow the instructions delivered with the new HVPS in accordance with
the II tube model used (TH 9448 or TH 9449).

B. Imageur 16 for Stenoscop with Pick-Up Tube


This Imageur exists with HVPS from Brandenburg, Model 833 (45066585), Brandenburg Model 833–2
(45067332) or Thomson Model TH7191 (58079229). These HVPS are all 24 V (dc) supplied.
In the event of II tube replacement, use the Thomson TH7191 (58079229, replace the HVPS if necessary)
in accordance with the instructions given in the Paragraph D.

C. Imageur 16 for Stenoscop with CCD


These Imageurs use II tubes TH 9449 (2120442) and HVPS Thomson TH7191 (58079229) exclusively.
In case of the II tube replacement, it is not necessary to change the HVPS, the same HVPS can be used.
Follow the instructions given in Paragraph D.

D. HVPS replacement ( on Imageurs )


1. Remove the connectors of the old II tube and solder them to the electrodes of the new II tube.

CAUTION Solder the old G3 connector to the G2 electrode of the new II tube TH9449HP.

2. Connect the G2 electrode of the new II tube to the G3 connector of the HVPS
(see Illustration 1).
3. Connect the other electrodes to the corresponding connectors of the HVPS. The G2 output from
the HVPS is now unconnected.
4. Install the remaining old G2 connector to the HVPS at the G2 output.
5. Adjust the vg2 (LV) Control Voltage of the HVPS to 0 V, using the following potentiometers:
S Imageur 16 for Stenoscop pick-up tube: R5 on Power Supply and Interco Board.
S Imageur 16 for Stenoscop CCD: R4 on CCD Interface Board.

6. Set the vg3 (LV) voltage to obtain the VG2 (HV) voltage corresponding to that shown in the
Inspection Acceptance Report for the new II tube, using the following potentiometers:
S Imageur 16 Stenoscop pick-up tube: R2, on Power Supply and Interco Board.
S Imageur 16 Stenoscop CCD: R7, on CCD Interface Board.
Use the following equation: vg2  1  VG2  1300 vg2 = low voltage
1300
VG2 = high voltage

CopyrightE 1996 by General Electric Co.


April 23, 1996 SERVICE NOTE Page 3 of 3

IMAGEUR 16 (6”) FOR STENOSCOP 2 SN 4011A


Rev. 1 2122943–100

ILLUSTRATION 1
II TUBE TH 9449 HP H506 CONNECTION

HV
SUPPLY II TUBE

7. Set the vg1(LV) voltage to obtain the VG1 (HV) voltage corresponding to that shown in the
Inspection Acceptance Report for the new II tube, using the following potentiometers:
S Imageur 16 for Stenoscop pick-up tube: R8, on Power Supply and Interco Board.
S Imageur 16 for Stenoscop CCD: R1, on CCD Interface Board.
Use the following equation: vg1  1  VG1  40 vg1 = low voltage
34.5
VG1 = high voltage

8. Insert a copy of this Service Note in the Stenoscop Service Manual.


9. Mark “Modified per SN 4011A”:
S On the Revision History page,
S On the cover of the Directions.

Are parts in stock? Yes n No Parts not needed


(If No, Service Note is not to be distributed)

Support: OLC ESC ASC


1–800–321–7937 (USA & Canada) +33 1 39 20 00 07 81–426–56–0033
1 (414) 524–5300 (Elsewhere)
Author: Didier Hervot
X-Ray Core Quality Team IB
Tel: +33 1 30 70 48 08

CopyrightE 1996 by General Electric Co.

Vous aimerez peut-être aussi