Instruction Manual
WE
IGH
IN
UNIT G
* S
2nd
UNIT 01/0
S 1/0
0.0
3 0
0g 3:0
5:4
5: P
0.0
0g
0 g
M
610
0g
Declaration of Conformity
Ohaus Corporation
19A Chapin Road
PO Box 2033
Pine Brook, NJ 07058 USA
hereby declares that the following products are in conformity with the EEC directives listed below (including any and all modifications).
Balance models: EP12001, EP12001C, EP22001, EP22001C, EP32001, EP32001C
ISO 9001 Registration for Ohaus Corporation.. Ohaus Corporation, USA, was examined and evaluated in 1994 by the Bureau Veritas Quality Interna-
tional, BVQI, and was awarded ISO 9001 registration. This certifies that Ohaus Corporation, USA, has a quality system that conforms with the international
standards for quality management and quality assurance (ISO 9000 series). Repeat audits are carried out by BVQI at intervals to check that the quality
system is operated in the proper manner.
Ted Xia
President Urs Müller
Ohaus Corporation General Manager
Pine Brook, NJ USA Ohaus Europe
Date: July 29, 2003 Greifensee, Switzerland
Additional Standards
Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates,
uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the
interference at his own expense.
TABLE OF CONTENTS
1. INTRODUCTION ...................................................................................................................................... EN-3
1.1 Description ........................................................................................................................................... EN-3
1.2 Features ............................................................................................................................................... EN-3
1.3 Safety Precautions ................................................................................................................................. EN-3
INDEX
Explorer Pro
EN-3
1. INTRODUCTION 2. INSTALLATION
1.1 Description 2.1 Unpacking
Thank you for deciding to purchase a Explorer® Pro Balance Open the package and remove the instrument and the
from Ohaus. Explorer Pro’s software design allows direct accessories. Check the completeness of the delivery. The
access to 11 application modes and all menus. The backlit following accessories are part of the standard equipment of your
dot matrix display, is capable of displaying either limited or new Explorer Pro balance.
multiple data fields at each application. Explorer Pro’s • AC Power
standard AutoCal™ automatically calibrates the balance due • Instruction Manual
to temperature changes that may affect calibration. Panel • Warranty Card
controls and soft key functions on the display clearly indicate • Weigh Below Hook
functions and data. Operation is extremely simplified, you
enter a menu, select an item, modify the item according to Remove packing material from the instrument.
menu selections and exit. Check the instrument for transport damage. Immediately
inform your Ohaus dealer if you have complaints or parts
Capacities from 12,000 grams to 32,000 grams are available. are missing.
Store all parts of the packaging. This packaging guarantees
To ensure you make full use of the possibilities offered by the best possible protection for the transport of your
your Explorer Pro balance, we advise you to read through instrument.
these operating instructions.
1.2 Features
• Ready to weigh without complicated setup
• 3 soft keys are application mode specific.
• Operating languages include English, Spanish, French,
German and Italian.
• Simplified menu navigation and balance setup.
• Capacity Bar guide and Fill bar guide.
• Flexibility to display either simple weighing results or
more advanced results including piece count, average piece
weight, sample size, tare weight, weight, under and over.
• Dot Matrix display with backlight
• Weighing, Parts Counting, Animal Weighing, Percent
Weighing, Check Weighing, Gross/Net/Tare Weighing and
Filling.
• Advanced weighing applications: Statistics, SQC,
Formulation, Differential Weighing.
• Library storage.
• Auto Calibration
Correct Incorrect
WEI
GH
IN
UN G
ITS
*
2nd
UN 01/0
ITS 1/03
0g
0.0 03:0
5:45
: PM
0
0.00
g
g
Explorer Pro
EN-5
2.3 Connecting Power and Communications
2.3.1 AC Adapter Installation
Connect the AC Adapter supplied to the three pin connector located
at the rear of the balance.
AC Adapter Connection
1 N/C
The following sections describe the hardware and software
2 Data Out (TXD)
provided with the balance
3 Data In (RXD)
4 N/C
Hardware
5 Clear To Send (CTS)
On the rear of the balance, the right-hand, 9-pin male subminiature
6 Data Terminal Ready (DTR)
“D” connector is provided for interfacing to other devices. The
7 Ground
pinout and pin connections are shown in the adjacent illustration.
8 Request To Send (RTS)
Refer to paragraph 3.5.6 for setup.
9 N/C
Output Formats
Data output can be initiated in one of three ways: 1) By pressing
PRINT; 2) Using the Auto Print feature; 3) Sending a print
command (“P”) from a computer.
* 0.00 0.00g
g
0g 6100g
Press the Enter button with the CALIBRATE soft key highlighted.
Explorer Pro EN-7
3 OPERATION
3.1 Overview of Controls
1 2
3
WEIGHING 01/01/03 03:05:45: PM
UNITS 2nd UNITS CALIBRATE 4
5
12
11
* 0.00 g
6
0.00g 9
10
7
9 0g 6100g
10
2 button When pressed in Menu mode, single press moves menu selection bar in an up direction and
highlights the activated field or changes the settings of a selected field in increasing order.
3 button When pressed in Menu mode, single press moves the cursor within a alphanumeric field
to the left (example date 07/02/2003). Selection does not wrap at end.
In application mode, single press moves to the next soft key selection to the left.
4 button When pressed in Menu mode, single press moves the cursor within a alphanumeric field
to the right (example date 07/02/2003). Selection does not wrap at end.
In application mode, single press moves to the next soft key selection to the right.
EN-8 Explorer Pro
5 button When pressed In Menu mode, single press moves menu selection bar in a down
direction and highlights the selected field or changes the settings of a selected field in
decreasing order.
6 Enter button When pressed in application modes, functions as an “Enter” button to accept Soft-key selection.
When pressed in menu, functions as an “Accept” button for accepting and saving
parameter of selected items.
7&8 >O/T< buttons When pressed, performs tare function or zero function.
9 LCD Display Backlit LCD display provides all necessary indications for operation.
10 Mode button When pressed, causes the balance to cycle through all active application modes.
11 Print button When pressed in any application mode, causes a print function to occur.
12 Menu button When pressed in any application mode, will bring up the Menu screen. Pressing the
button within a Menu screen will return to the application mode.
Explorer Pro EN-9
1 2 3
13
12
11 -* 200.0000
>0<
Result Line Result Line
dwt
6
7
8
Result Line Result Line
Result Line Result Line
0g 6100g 9
10
4 Soft Keys Three Soft-keys are displayed in the upper display area. Depending on the application mode
selected, the functionality changes. A Soft-key can be selected by using the left and right
arrow buttons. Pressing the Enter button on a highlighted soft-key will activate the
function.
10 Bar Graph Shaded area indicates capacity used for, under, accept, over for check weighing application,
or target values for filling application.
13
* Stability indicator.
EN-10 Explorer Pro
3.3 Menu To select a highlighted menu, press the Enter button. Each menu
item contains a display that allows setting balance parameters
3.3.1 Menu Structure
to individual requirements. These are described in detail in section
The Explorer Pro balance utilizes a menu structure that permits
3.5 Balance Settings.
entering various menus by using a dedicated Menu button.
Pressing the Menu button allows access to additional sub
Section 3.5.1 describes Calibration procedures and section 3.4
menus. See illustration below.
describes the Applications.
MENU
Calibration
Balance Options 3.3.3 Turning on the Balance
Readout
Application Modes The Explorer Pro balance is ready to operate after the
Units
Interface installation procedures are performed. When the balance is
Print Options first turned on and it completes its checks, it can be used to
GLP Print Options
Library weigh or tare materials without setting the menus.
Lockout
Factory Reset
Exit It is recommended that you read this manual carefully and set
the balance to operate to your specific applications before using.
3.3.2 Navigation
When the balance is first turned on, the main weighing screen is Power On/Off
displayed. To enter the menu and to change parameters, the To turn the balance ON, press the On/Off button. To turn OFF,
following controls are used: press the button again.
-Scrolls down through the menus. Recommended warm up period is twenty (20) minutes. Analytical
Class I balances require at least 2 hours.
- Moves the cursor within the field to the left.
The internal circuits of the balance are powered whenever it is
plugged into a power source.
- Moves the cursor within the field to the right.
There are two ways to exit the Menu; one is to scroll to Exit and
press the Enter button, the other is to press the Menu button. All
changes will be saved.
Explorer Pro EN-11
Removing the container and material from the pan will cause the
By pressing the Enter button and using the arrow buttons, each
balance to display the container’s weight as a negative number.
item on the Application Modes screen can be turned on or off.
When finished, press the Menu button, the balance returns to The tared container weight will remain in memory until the balance
weighing mode. All settings are saved. is zeroed again or the balance is turned off.
To use any of the modes that have been turned on, press the
Mode button repeatedly to cycle through all of the application
modes.
EN-12 Explorer Pro
2ND UNITS Tare the balance if required. With the NEW COUNT soft key
When 2nd UNITS is highlighted, each press of the Enter button highlighted, press the Enter button. Place 10 pieces on the pan
cycles through the measuring units that are turned On in the as instructed by the blinking text on the display. Press the Enter
Units menu. The last option when cycling through the units is button.
Off.
The display now indicates the Average Piece Weight (APW) and
UNITS the sample size.
When UNITS is selected and highlighted, each press of the Enter PARTS COUNTING 01/01/03 03:05:45 PM
button cycles through the measuring units that are On and will SAMPLE SIZE SETUP NEW COUNT
appear adjacent to the large numerals.
*
APW: 2.00g
10 Pcs
20.00 g
3.4.2 Parts Counting Sample Size: 10
Tare: 0.00g
Three different Parts Counting modes can be selected, Count,
Check or Fill in the Parts Counting Setup. Each mode is described 0g 6100g
separately.
Place parts to be counted on the pan. Balance displays number
USE of pieces.
PARTS COUNTING
In the Parts Counting mode, the balance displays the quantity of ADJUSTMENTS
parts you place on the pan. Since the balance determines the The Parts Counting display contains three soft keys at the top of
quantity based on the average weight of a single part, all parts the screen SAMPLE SIZE, SETUP and NEW COUNT.
must be reasonably uniform in weight.
NEW COUNT
The balance has a default setting of 10 pieces. This permits When NEW COUNT soft key is selected, follow the screen
starting parts counting immediately without setting up the balance. instructions “Add 10 Pieces, Press Enter “. After pressing the
The following display indicates the first time operation of parts Enter button, the balance calculates the APW and the display
counting. indicates the number of pieces.
SETUP
Select the SETUP soft key and press the Enter button. The PARTS
COUNTING SETUP is shown.
Explorer Pro EN-13
PARTS COUNTING SETUP For higher quantities, the APW will not be recalculated. During
Library Name : PARTS COUNTING
the auto optimization process, the message “Auto Optimizing,
Mode : Count Please wait” will be displayed in the message bar on the screen.
Unit :g
Sample Size : 10
APW : Off
Tare : 0.00
Display Settings: Custom, Default (default is Default)
Auto Optimize : Off Selecting Display Settings to Custom will bring up the DISPLAY
Display Settings : Custom
Size/APW Key : Unlocked SETTINGS screen that will allow individual display items to be
New Count Key : Unlocked
Exit turned on or off. Selecting Exit will return balance to previous
screen.
Using the arrow keys and Enter button, each item on the display DISPLAY SETTINGS
can be entered and modified as required. The following APW : On
Sample Size : On
information describes each entry on the screen. Tare : On
Weight : On
Exit
Library Name:
Allows storage and recovery of parameters and results. When
a library is recovered, the library name will replace the
application name in the application mode screen. The library
name has maximum of 18 characters.
Size/APW Key: Unlocked, Locked (default is Unlocked)
Mode: Count, Check, Fill (default is Count) Allows locking and unlocking the soft key function in the Parts
Three different modes are available, Count, Check or Fill. The Counting screen. If locked, the message “Key Is Locked” will
applications for Check and Fill are described in detail after this be displayed in the message bar.
section.
New Count Key: Unlocked, Locked (default is Unlocked)
Unit: g, kg, etc... (default is g) Allows locking and unlocking the soft key function in the Parts
Allows the selection of all activated units. Counting screen. If locked, the message “Key Is Locked” will
be displayed in the message bar.
Sample Size : 0-99 (default is 10)
Defines the amount of pieces used to calculate APW (average Exit:
piece weight). Setting the sample size will force the APW field When selected, balance returns to the Parts Counting Mode.
to Off and change the APW soft key to SAMPLE SIZE.
SAMPLE SIZE/APW
APW: 0-999999999 (default is Off) With SAMPLE SIZE soft key highlighted,press the Enter button.
Defines the average piece weight. Setting the APW will force The PARTS COUNTING display is shown with the value of the
the Sample Size field to Off and change the SAMPLE SIZE soft sample size highlighted. Using the arrow buttons and the Enter
key to APW. button, the sample size value can be changed.
Tare: 0-99999999 (default is 0) After the Enter button is pressed, and a new sample size has
Defines the weight of the container being used. been entered, remove the sample and place parts to be counted
on the balance using the new sample size.
Auto Optimize: On,Off (default is Off)
Selecting Auto Optimize On will recalculate the APW value APW
automatically during the parts counting process up to double With the APW soft key highlighted, press the Enter button. The
the amount of parts. APW is highlighted at the lower left of the screen. Using the
arrow buttons and the Enter button, the APW value can be
changed.
EN-14 Explorer Pro
USE ADJUSTMENTS
PARTS COUNTING-CHECK The PC Check display contains three soft keys at the top of the
This feature permits establishing a set quantity of pieces as a screen SAMPLE SIZE, SETUP and NEW COUNT.
criteria for similar items that can be quickly checked against the
sample. In the Parts Counting Check mode, the displayed data NEW COUNT
includes, present count (Pcs), APW, Sample Size, Tare, Weight When NEW COUNT soft key is selected, follow the screen
and a bar graph indicating UNDER, ACCEPT (blinking) and instructions “Add 10 Pieces, Press Enter “. After pressing the
OVER. Enter button, the balance calculates the APW and the display
indicates the number of pieces.
Enter the PARTS COUNTING SETUP menu and change Mode to
Check. SETUP
Using the left arrow key, select SETUP soft key and press the
The balance is now in the PC CHECK mode. Enter button. The PARTS COUNTING SETUP is shown.
PARTS COUNTING 01/01/03 03:05:45 PM
SAMPLE SIZE SETUP NEW COUNT PARTS COUNTING SETUP
Add 10 Pieces, Press Enter Library : PARTS COUNTING
Mode : Check
Unit :g
Sample Size : 10
APW : Off
APW: 0.00g 0.00 g Tare : 0.00
Sample Size: 10 Under: 5 Auto Optimize : Off
Tare: 0.00g Over: 10 Over : 10
Under :5
Display Settings : Custom
Size/APW Key : Unlocked
New Count Key : Unlocked
20.00 g
Defines the under limit in pieces.
Three new entries Weight, Over and Under appear on this screen. PARTS COUNTING 01/01/03 03:05:45 PM
Selecting Exit will return balance to previous screen. SAMPLE SIZE SETUP NEW COUNT
Under: On, Off (default is On) NOTE: The values appearing on the screen initially are from
Can be turned on or off. previous entries in parts counting.
SAMPLE SIZE/APW Placing an item on the pan at this time will only respond to the
Select SAMPLE SIZE soft key and enter the sample size desired. previous settings for sample size and APW.
APW To use PC FILLING, the sample size and other parameters must
With the APW soft key highlighted, press the Enter button. be set first. Refer to adjustments and set the balance parameters
The APW is highlighted at the lower left of the screen. Using to fit specific requirements.
the arrow buttons and the Enter button, the APW value can be
changed. Follow screen instructions and place sample on pan. ADJUSTMENTS
The PC FILLING display contains three soft keys at the top of the
Remove the sample and place items on pan. In the following screen SAMPLE SIZE, SETUP and NEW COUNT.
example, the sample size was 10, the APW was 5.00g, the
under value was set at 99 and the over value was set at 101. NEW COUNT
The acceptable value was 100 pieces. When NEW COUNT soft key is selected, follow the screen
instructions “Add 10 Pieces, Press Enter “is displayed After
PARTS COUNTING 01/01/03 03:05:45 PM pressing the Enter button, the balance calculates the APW and
SAMPLE SIZE SETUP NEW COUNT
the display indicates the number of pieces.
One additional entry has been added to the PARTS COUNTING 3.4.3 Animal Weighing
SETUP, that is Target. Refer to the Adjustments section of parts Animal weighing permits you to weigh small animals and filters
counting for a description of all other settings. out animal movements. Choices of manual, semi automatic
and automatic operation is possible.
Target: 0-9999 (default is 1000)
Defines the target pieces for the 100% limit. The display data includes the filtered weight of the animal in
selected unit of measurement, and unfiltered weight of the animal
Display Settings: Custom, Default (default is Default) in 2nd unit of measurement, filter level, mode of operation and
Allows Display Settings to be set to default or custom. If Display a bar graph indicating present used capacity of the balance.
Settings is set to custom, information on the result lines of the
Parts Counting mode can be individually turned on or off. If ANIMAL WEIGHING 01/01/03 03:05:45 PM
Display Setting is set to default, all the information is turned FILTER SETUP START
on. Place animal, Press Enter
Remove the sample and add items on pan until 100% is Filter: Low 2038.73 g
highlighted at the bottom of the screen as shown. In the following Mode: Manual
example, the sample size was 10, the APW was 2.00g, the Target
value was set at 100 pieces. 0g 6100g
*
APW: 2.00g
100 Pcs
200.00 g
highlighted, press the Enter button to clear the weight. The
balance is now ready for another measurement. To change
the mode of operation from manual to semi automatic or
Sample Size: 10 Target: 100 automatic and change filtering levels, refer to Adjustments.
Tare: 0.00g Diff: 0
50% 90% 100%
0% 110%
Explorer Pro EN-17
3.4.4 Percent Weighing The reference weight is removed and a second weight is placed
Percent weighing allows placing a reference load on the balance, on the pan. The example illustrates a 50 g weight.
then viewing other loads as a percentage of the reference. The
load placed on the pan as a reference is displayed as 100%. PERCENT WEIGHING 01/01/03 03:05:45 PM
Subsequent loads are displayed as a percentage of the reference. REFERENCE SETUP NEW REF
The maximum limit is the capacity of the balance. The minimum
limit is 100d. The display data includes reference weight in
unit of measurement, difference value in percent, difference value
*
Ref: 200.00g
25.00 %
50.00 g
in measuring unit and a bar graph indicating present used Diff: -75.00 %
capacity of the balance. Diff: -150.00g
0g 6100g
PERCENT WEIGHING 01/01/03 03:05:45 PM
REFERENCE SETUP NEW REF ADJUSTMENTS
Add Reference Weight, Press Enter
The PERCENT WEIGHING display contains three soft keys at the
* top of the screen REFERENCE, SETUP and NEW REF.
USE SETUP
PERCENT WEIGHING Select SETUP soft key and press the Enter button. The PERCENT
With the NEW REF soft key highlighted, press the Enter button. WEIGHING SETUP display is shown.
Place the reference weight on the pan and press the Enter
button again. The sample shown indicates a 200g weight PERCENT WEIGHING SETUP
was placed on the pan. Unit :g
Reference : 0.00
Display Settings : Custom
Reference Key : Unlocked
PERCENT WEIGHING 01/01/03 03:05:45 PM New Reference Key : Unlocked
REFERENCE SETUP NEW REF Exit
Add Reference Weight, Press Enter
*
Ref: 200.00g
100.00 200.00 g
%
Diff: 0.00 %
Diff: 0.00g
0g 6100g
Explorer Pro EN-19
Over: 15.00 g
Target: 10.00 g Diff: -10.00 g
Display Settings: Custom, Default (default is Default)
Under: 5.00 g Diff: -100.00 %
Allows Display Settings to be set to default or custom. If Display
Under Accept Over
Settings is set to custom, information on the result lines of the
Percent Weighing mode can be individually turned on or off. If
Display Setting is set to default, all the information is turned USE
on. CHECK WEIGHING
Before using the Check Weighing feature, the Over, Target and
Reference Key: Locked, Unlocked (default is Unlocked) Under limits must be set.
Allows locking or unlocking the REFERENCE soft key in the
percent Weighing screen. Message “Key is locked” when Locked ADJUSTMENTS
is selected. The CHECK WEIGHING display contains three soft keys at the top
of the screen UNITS, SETUP and CALIBRATE.
New Reference Key: Locked, Unlocked (default is Unlocked)
Allows locking or unlocking the NEW REF soft key in the Percent CALIBRATE
Weighing screen. Message “Key is locked” when Locked is When the CALIBRATE soft key is highlighted, and the Enter
selected. button is pressed, calibration can be performed.
Exit: SETUP
When selected, balance returns to the Percent Weighing Mode. Select the SETUP soft key and press the Enter button. The
CHECK WEIGHING SETUP display is shown.
REFERENCE
Select REFERENCE soft key and press the Enter button. The
display screen changes with Ref field highlighted. Reference CHECK WEIGHING SETUP
weight can be specified. Library Name : CHECK WEIGHING
Unit :g
PERCENT WEIGHING 01/01/03 03:05:45 PM Over : 15.00
Target : 10.00
REFERENCE SETUP NEW REF Under : 5.00
Display Settings : Custom
Exit
*
Ref: 0.00 0.00 g
Diff:
Diff:
0g 6100g
0.00 g
Target: 0-99999999 (default is 10). Diff: -50.00 g
Pct: 0%
Defines target value. 50% 90% 100%
0% 110%
Under: 0-99999999 (default is 5)
Defines under limit value. USE
FILLING
Display Settings: Custom, Default (default is Default) A sample (target weight) is placed on the pan and the Enter
Allows Display Settings to be set to default or custom. If Display button is pressed. The balance stores this weight as shown on
Settings is set to custom, information on the result lines of the the display as Target. The sample is removed from the pan and
Check Weighing mode can be individually turned on or off. If the material is added to the pan. The display provides Target
Display Setting is set to default, all the information is turned Weight, Difference weight, Percentage and the split bar at the
on. bottom of the screen displays the percentage of the sample
compared against the target weight. When a container is used,
Reading allows a choice of either weight readings to be zero the balance before placing the actual sample on the pan.
displayed in large numeral or message that displays OVER, See Adjustments to set other balance parameters.
ACCEPT and UNDER in large letters. Display below indicates
message format. ADJUSTMENTS
The FILLING display contains three soft keys at the top of the
CHECK WEIGHING 01/01/03 03:05:45 PM screen TARGET, SETUP and NEW TARGET.
UNITS SETUP CALIBRATE
NEW TARGET
* UNDER
Over: 15.00 g 0.00g
With the NEW TARGET soft key highlighted, press the Enter button.
Follow display instructions to “Add Target Weight, Press Enter”.
Target: 10.00 g Diff: -10.00 g This establishes a new target weight and is displayed on the
Under: 5.00 g Diff: -100.00 % screen.
SETUP
When SETUP soft Key is highlighted, press the Enter button. The
Exit:
FILLING SETUP display is shown.
When selected, returns to Check Weighing mode.
UNITS
When the UNITS soft key is highlighted, each press of the Enter
button cycles through the measuring units that are On.
Explorer Pro EN-21
0g 6100g
TARGET 0g 6100g
With the TARGET soft key highlighted, press the Enter button.
ADJUSTMENTS
The FILLING display is shown with the value of the Target The G/N/T WEIGHING display contains three soft keys at the top
highlighted. Using the arrow buttons and the Enter button, the of the screen UNITS, 2nd UNITS and TARE.
Target value can be changed.
FILLING 01/01/03 03:05:45 PM TARE
TARGET SETUP NEW TARGET When the TARE soft key is selected, pressing the Enter button
will store the current display reading as the tare value.
*
Target: 0.00 g
0.00 0.00 g
g
2nd UNITS
When 2nd UNITS soft key is highlighted, each press of the
Diff: -0.00 g
Enter button changes the measuring units that are On and will
50% 90% 100% appear under the large numerals. The last option when cycling
0% 110% through the units is Off.
ADJUSTMENTS
START/ACCEPT Tare Required: Off, Single, Dual (default is Single)
When the START key is selected, pressing the Enter button Allows selection of Tare Required. When Off is selected, the
will start differential weighing application. The screen displays balance will not prompt for container. When Single is selected,
“Place Initial Container X, Press Enter” message and the the balance will prompt for the initial container weight. When
Differential Weighing process has started. When a new Dual is selected, the balance will prompt for the initial
process starts all the data from previous Differential Weighing container weight and final container weight.
is erased. At this point, the soft key has changed into an
ACCEPT key. When the ACCEPT key is selected, the weight Auto Sample: Off, On (default is Off)
value has been accepted as valid data. Allows Auto Sample to be turned on or off . When Auto Sample
is turned on, the Differential Weighing process is automatic.
SETUP
When executed, the DIFFERENTIAL WEIGHING SETUP screen Clear All Data: No, Yes (default is No)
will be displayed. All the data from the Differential Weighing process will be
erased when Yes is selected.
EDIT
When executed, an edit screen will display that will allow an Display settings: Custom/Default (default is Default)
user to re-sample a sample weight, or to delete an undesirable Allows Display Settings to be set to default or custom. If Display
sample, or to continue. Settings is set to custom, information on the result lines of the
Differential Weighing mode can be individually turned on or
Library Name: off. If Display Setting is set to default, all the information is
Allows storage and recovery of parameters and results. turned on.
When a library is recovered, the library name will replace
the application name in the application mode screen. The View Results:
library name has maximum of 8 characters. Allows the calculated results of the Differential Weighing
process to be viewed.
Mode: Weight, Percent, % Retention (default is Weight)
Allows the selection of the various calculations in the final Exit:
results. In the sub menu, will return to the previous menu.
In the main men, will return to the last application mode.
Units: g, kg, etc...(default is g)
Allows the selection of all activated units.
3.4.9 Formulation Add the first component to the container. The dual bar graph
Formulations can be named and have from 2 to 10 components at the bottom of the screen indicates the percentage of the
specified. Names are limited to 18 characters. Once named, component added. The target weight, difference and percent
they may be recalled and used at any time. Each component are also displayed. Add each component until the specified
of a given formulation can be specified as to its weight or Target weight is reached (100%). When adding an individual
percentage. Each element of a formulation is shown on a dual component, make sure the balance has stabilized and indicates
bar graph as a percentage and displays the desired weight. the specified weight. Continue until all components have been
Thus, each component may be placed on the pan until 100% added.
is indicated.
The next screen automatically appears and indicates the Target,
USE Result and Difference weights of each component of the formula
FORMULATION 01/01/03 03:05:45 PM as well as the totals. Pressing the PRINT button will send the
RESULTS SETUP START screen results to an externally connected printer or computer.
* 0.00 g
Press the Enter button to exit and return to the starting screen.
Results of any formulation may be displayed by selecting the
Item: 1 RESULTS soft key at the top of the screen. The actual formulation
Target: 0.00 g
Diff: 00.00 g Pct:0.00 % is selected by entering the FORMULA SETUP screen and scrolling
50% 90% 100% through the Library Names.
0% 110%
Before using the Formulation feature, the Formulation Setup To start a new formulation, repeat the entire process.
should be reviewed. Highlight the SETUP soft key and press
the Enter button. Assign a Library Name. Review the settings ADJUSTMENTS
and change as necessary. Enter the Mode type, Weight or START/ACCEPT
Percent. Enter the number of Items. When the START key is selected, pressing the Enter button will
FORMULATION SETUP start formulation application. The screen displays “Add Item
Library Name : FORMULATION “name field”, Press Enter” message and the Formulation process
Mode : Weight
Unit :g has started. When a new process starts, all the data from
Number Of Items :2 previous Formulation is erased. At this point, the soft key has
Display Settings : Default
Items Setup changed into an ACCEPT key. When the ACCEPT key is selected,
View Results
Exit the weight value has been accepted as valid data.
SETUP
When selected, the FORMULA SETUP screen will be displayed.
Exit the Formulation Setup. The initial screen appears with the
RESULTS
name of the formulation appearing at the upper left-hand corner
When selected, a screen will display all calculated results.
of the screen.
Library Name:
Place a container on the pan and press the >O/T< button to
Allows storage and recovery of parameters and results.
tare the container weight.
When a library is recovered, the library name will replace the
application name in the application mode screen. The library
With the START key highlighted, press the Enter button. Follow
name has maximum of 18 characters.
the instructions on the screen. The START key changes to ACCEPT.
Explorer Pro EN-25
With Items Setup selected, specify the names for each component (Used for Weighing and Animal Weighing)
of the formula (limit 9 characters) and either the weights or
STATISTICS 01/01/03 03:05:45 PM
percentages.
RESULTS SETUP START
*
0.00
ITEMS SETUP
Item Name 1 : Item 1
Item Name 2 : Item 2 g
Item Weight 1 : 0.00 Max: 0.00 g Sum: 0.00 g
Item Weight 2 : 0.00 Min: 0.00 g Mean: 0.00 g
Exit Target: 10.00 g
50% 90% 100%
0% 110%
View Results:
Allows the calculated results of the Formulation process to be
viewed.
Exit:
In the sub menu, will return to the previous menu.
In the main men, will return to the last application mode.
EN-26 Explorer Pro
STATISTICS 01/01/03 03:05:45 PM Place a container on the pan and press the >O/T< button to
RESULTS SETUP START tare the container weight.
*
0.00 g
With the START key highlighted, press the Enter button. Follow
the instructions on the screen. The START key changes to
ACCEPT. Add the sample to the container. Repeat for all
Max: 0.00 g Sum: 0.00 g
Min: 0.00 g Mean: 0.00 g samples.
Under: 5.00 g Over: 15.00 g
Under Accept Over
Weighing/Animal Modes
(Used for Check Weighing) When Weighing Mode or Animal Weighing Mode is selected,
Before using the Statistics feature, the Statistics Setup should
The bar graph at the bottom of the screen indicates the weight
be reviewed. Highlight the SETUP soft key and press the Enter
of each sample when added. The Maximum, Minimum, Sum
button. Assign a Library Name.
and Mean weights are also displayed on the screen.
STATISTICS SETUP
Maximum : On
Minimum : On
Difference : On
Sum : On
mean ; On
Std. Deviation : On
Rel. Deviation : On
Auto Sample : On
Auto Print : On
Display Settings : Default
View Results
Exit
Explorer Pro EN-27
The STATISTICAL RESULTS screen automatically appears and Unit: g, kg, etc... (default is g)
indicates the Number of samples, Unit, Mean, Maximum, Allows the selection of all activated units.
Minimum, Standard Deviation, Relative Deviation, Sum and
Difference weights. Pressing the PRINT button will send the Number of Samples: xx (default is 3)
screen results to an externally connected printer or computer. Allows manual entry for number of items. Only integer values
up to 100 are valid.
STATISTICAL RESULTS
Number Of Samples: 3
Unit: g
Mean: 0.00 Filter: Low, Medium, High (default is Low)-Animal mode only
Maximum: 0.00 Can be set to one of three filtering levels. A higher level will
Minimum: 0.00
Std. Deviation: 0.00 provide more accurate measurement.
Relative Deviation: 0.00
Sum: 0.00
Difference: 0.00 Over, Under: Check mode only
Allows manual entries for the Over limit and the Under limit
Press <Enter> To Exit
values. Eight decimal places are allowed for these values.
Press the Enter button to exit and return to the starting screen. Target: Check mode and Filling mode
Results may be displayed by selecting the RESULTS soft key at Allows manual entry for the Target Weight value. Eight
the top of the screen. decimal places are allowed for target Weight value.
To start a new test, repeat the entire process. Maximum: On, Off (default is On)
Allows Maximum to be turned On or Off. When Maximum is
ADJUSTMENTS turned on, the maximum stored sample is included into the
START/ACCEPT final results.
When the START key is selected, pressing the Enter button will
start statistics application. The screen displays “Add Sample Minimum: On, Off (default is On)
#1, Press Enter” message and the Statistics process has started. Allows Minimum to be turned On or Off. When Minimum is
When started, all the data from previous statistics is erased. At turned on, the minimum stored sample is included into the
this point, the soft key has changed into an ACCEPT key. When final results.
the ACCEPT key is selected, the weight value has been accepted
as valid data. Difference: On, Off (default is On)
Allows Difference to be turned On or Off. When Difference is
SETUP turned on, the difference calculation between the maximum
When selected, the STATISTICS SETUP screen will be displayed. stored sample and the minimum stored sample is included
into the final results.
RESULTS
When selected, a screen will display all calculated results. Sum: On, Off (default is On)
Allows Sum to be turned On or Off. When Sum is turned on,
Library Name: the sum calculation of the samples is included into the final
Allows storage and recovery of parameters and results. results.
When a library is recovered, the library name will replace
the application name in the application mode screen. The Mean: On, Off (default is On)
library name has maximum of 18 characters. Allows Mean to be turned On or Off. When Mean is turned on,
the mean calculation of the samples is included into the final.
Mode: Weighing, Animal, Check, Filling (default is Weighing)
Allows the selection of the various calculations in the final
results.
EN-28 Explorer Pro
Std. Deviation: On, Off (default is On) These weight limits are identified as +Tl, +T2, NOMINAL and -
Allows Std. Deviation to be turned On or Off. When Std. Deviation TI, -T2. As samples are weighed and stored in the balance, a
is turned on, the standard deviation calculation of the samples trend analysis is developed and displayed on the balance. Up
is included into the final results. to 25 samples in a batch with up to ten batches are visible on
a trend screen for quality control purposes. Each batch of
Rel. Deviation: On, Off (default is On) samples is shown on the display which indicates the maximum/
Allows Relative Deviation to be turned On or Off. When Relative minimum standard deviation and mean values for each batch.
Deviation is turned on, the relative deviation calculation of the An on going examination of the relative deviation of the samples
samples is included into the final results. along with other statistical data can be viewed and is stored.
By observing the results of the VIEW TRACE screen, you can
Auto Sample: Off, On (default is Off) effectively monitor the filling process operation. Setup parameters
Allows Auto Sample to be turned On or Off . When Auto Sample can be stored in the library and up to 5 products with statistical
is turned on, the Statistics process is automatic. history can be stored in memory. All SQC information can be
printed.
Auto Print: Off, On (default is Off) USE
Allows Auto Print to be turned On or Off . When Auto Print is
SQC 01/01/03 03:05:45 PM
turned on, the final calculated results are automatically printed
RESULTS SETUP START
out at the end of the Statistics.
*
Display settings: Custom/ Default (default is Default)
Allows Display Settings to be set to default or custom. If Display
Tare: 0.00 g
0.00 g
Settings is set to custom, information on the result lines of the Max: 0.00 g Sum: 0.00 g
Statistics mode can be individually turned on or off. If Display Min: 0.00 g Mean: 0.00 g
Setting is set to default, all the information is turned on. 0g 6100g
Exit: Before using the SQC feature, the SQC Setup should be reviewed.
In the sub menu, will return to the previous menu. Highlight the SETUP soft key and press the Enter button. Assign
a Library Name.
In the main men, will return to the last application mode.
NOTE: All SQC settings made with a Library Name will be directly
View Results: associated with the Library name. When a Library Name is
Allows the calculated results of the Statistics process to be called up, all settings will be retained with the Library Name.
viewed.
Review the settings and change as necessary. Enter the Unit,
Number of Samples, Tolerance (either weighing or percentage),
3.4.11 Statistical Quality Control
Tare, Auto Sample, Auto Print, Display Settings (Default or
The Statistical Quality Control (SQC) feature is extremely useful
Custom), and Batch Name.
during various types of process filling operations when it is
desired to monitor and control the process to eliminate under SQC SETUP
and over filling. Provisions are made in the balance to Library Name : SQC
Unit :g
accommodate the weight of various packaging methods. During Number Of Samples :3
Tolerance : Weighing
operation, parameters of the packaged product are set into the Tare Required : Off
balance such as packaging weight, acceptable weight limits Auto Sample : Off
Auto Print : Off
and non-acceptable weight limits of the product. Clear All Data : No
Display Settings : Default
Batch Name : Batch 1
View Trace
View History
Explorer Pro EN-29
Defining Tolerances After you have taken the required number of data samples, you
When Tolerance is selected a choice of either Weighing or can view the sample View Trace and/or the View History. The
Percent is displayed. Define either the weight or percentage for View Trace screen below indicates 5 batches were processed.
your product starting with +T2, +T1, Nominal, -T1 and -T2.
Batch 5 Total Batches: 5
01/01/03 03:05:45 :01/01/03 03:08:45
Weight Tolerance
+ Tolerance 2 : 0.00 60.00 +T2
+ Tolerance 1 : 0.00
Nominal : 0.00 55.00 +T1
- Tolerance 1 : 0.00
- Tolerance 2 : 0.00 50.00 N
Exit
45.00 -T1
40.00 -T2
NOTE: Once a batch with all samples has been completed, you
After all entries have been made, exit the SQC SETUP menu. cannot make changes to the batch. A new batch must be run.
Pressing the Enter button will start SQC application. Continue Return to the SETUP soft key and rerun the Batch.
to place samples on the pan until all samples have been entered.
The following sample screens are for 5 batches. A Nominal Definitions for the symbols are shown below.
weight of 50.00g with T1 set at 5 g above and below and T2
set at 10g above and below. A sample history of 5 batches is MAXIMUM LIMIT
displayed.
STANDARD MEAN VALUE OF SQC SET
Samples: 5 Batch Name: DEVIATION
Unit: g
Nominal Weight: 50.00 MINIMUM LIMIT
Mean: 49.9975
Maximum: 49.9996
Minimum: 49.9936
Std. Deviation: 0.0025
Relative Deviation: 0.0001
>+T2: 0.00 %, 0
>+T1: 0.00 %, 0
+T1 > N > -T1: 100.00 %, 5
<-T1: 0.00 %, 0 ADJUSTMENTS
<-T2: 0.00 %, 0
START/ACCEPT
When the START key is selected, pressing the Enter button will
start SQC application. The screen displays “Add Sample #1,
Pressing the Enter button will return the balance to the Start
Press Enter” message and the SQC process has started. At this
Screen. Highlight the SETUP soft key and press the Enter button.
point, the soft key has changed into an ACCEPT key. When the
Assign a new Batch Name. Continue until all batches have
ACCEPT key is selected, the weight value has been accepted as
been processed.
valid data.
Return to the SETUP soft key and press the Enter button. Scroll
SETUP
down to View Trace and or View History and press the Enter
When selected, the SQC SETUP screen will be displayed.
button.
RESULTS
When selected, a screen will display all calculated results of
the SQC batch.
EN-30 Explorer Pro
Mean Tare: Off, 00.000 (default is Off) A Density Determination Kit Part Number 470007-010 is designed
When Tare Required is set to Mean Tare or Add w/ Mean Tare, to be used with Ohaus Explorer Pro balances. Illustrations
this menu selection will allow manual entry of the Mean Tare in this procedure refer to the density kit, however, you may
value. use whatever lab apparatus that will suit the requirements
for density measurements. A built in reference density table
Auto Sample: Off, On (default = Off) for water at temperatures between 10ºC and 30ºC is included
Allows Auto Sample to be turned On or Off . When Auto Sample in the balance software.
is turned on, the Statistics process is automatic.
When making density measurements, the material should weigh
Auto Print: Off, On (default is Off) at least 10.0 mg on an analytical balance and 100 mg on a
Allows Auto Print to be turned On or Off . When Auto Print is precision balance.
turned on, the final calculated results are automatically printed
out at the end of the SQC process.
Explorer Pro EN-31
USE
Balance Preparation for Density Measurement
The Explorer Pro Balance is equipped with a weigh below
hook which can be attached to the bottom of the balance.
When making density measurements, it is necessary to sup-
port the balance on both sides allowing enough clearance
to for a beaker to fit under the balance, a platform or labo-
ratory jacks can be used. Remove the protective adhesive
cap from the bottom of the balance as shown in the figure
below. Screw the weigh below hook into the bottom of the
balance.
PROTECTIVE
BOTTOM OF BALANCE COVER
WEIGH BELOW
HOOK INSTALLATION
LOCATION
PLATE
BALANCE SUPPORT
EN-32 Explorer Pro
Solid Density Determinations for items More Dense Than The density Q is the quotient of the mass m and the volume V.
Water
DENSITY 01/01/03 03:05:45 PM
Q= m
TEMP SETUP START
V
*
Temp: 25.0
0.00
Air: 0.00 g
g
Density determinations are performed by using Archimedes’
principle.. This principle states that every solid body immersed
Aux D: 0.9971 Liquid: 0.00 g
in a fluid loses weight by an amount equal to that of the fluid it
Volume: 0.00 g displaces. The density table for water is included in the
0g 6100g Explorer balance software.
Before using the Density feature, the Density Setup should be The density of a solid is determined with the aid of a liquid
reviewed. Highlight the SETUP soft key and press the Enter whose density, Qo, is known (water is used as an auxiliary
button. Assign a Library Name. liquid). The solid is weighed in air (A) and then in the auxiliary
liquid (B). The density Q can be calculated from the two
NOTE: All Density settings made with a Library Name will be weighings as follows:
directly associated with the Library name. When a Library Name
is called up, all settings will be retained with the Library Name.
A
Q= •Q0
A -B
Review the settings and change as necessary. The balance allows direct determination of the buoyancy
P (P =A - B ) and consequently the above formula can be
DENSITY SETUP simplified:
Library Name : DENSITY A
Mode : Solid Q= •Q0
Auxiliary Liquid : H2O P
Auxiliary Density : 0.9971
Temperature (C)
Porous Material
: 25.00
: Off
Q = Density of the solid
Auto Sample : Off A = Weight of the solid in air
Auto Print : Off
Display Settings : Default B = Weight of the solid in the auxiliary liquid
Exit Q0 = Density of the auxiliary liquid at a given tempera-
ture (this value depends on the temperature). The
density table for water is included in Explorer balances.
P = Buoyancy of the solid in the auxiliary liquid (corre-
sponds to A -B).
Place a beaker underneath the balance (not supplied) A typical final display is shown which indicates all of the
and suspend a precision calibrated thermometer 0 °C to
parameters and values.
30 °C on the edge of the beaker.
• Record and store the temperature in the Setup Menu. DENSITY 01/01/03 03:05:45 PM
• Tare the balance. TEMP SETUP START
• Press Enter button to start the Density application.
• Follow directions of the prompts
• “Place weight in Air, Press Enter”. Place the solid sample
on the pan and press the Enter button.
• “Place weight in Liquid, Press Enter”. Suspend the solid
*
Temp: 25.0
19.90 Air: 100.19 g
g/cc
sample from the Weigh Below Hook into the beaker. Use Aux D: 0.9971 Liquid: 95.17 g
a fine wire or thread. Fill the beaker with auxiliary liquid Volume: 5.03 g
(liquid of known density Q 0' usually distilled water or 0g 6100g
ethanol). Insure that the liquid will cover the sample by
at least 1 cm after immersion.Ensure that no air bubbles
adhere to the immersed part of the supporting wire. Re-
move air bubbles by moving the wire or by means of a
fine brush.
Weigh the solid (buoyancy P ) by pressing the Enter
button.
• The display indicates the density in grams/ cc.
EN-34 Explorer Pro
Improving the Accuracy of the Result of Solid Density Liquid Density Determinations
The following tips should help you improve the accuracy of The density of a liquid can be made using a sinker of known
the results in the density determination of solids. volume. The sinker is weighed in air and then in the liquid
whose density is to be determined, The density, Q, can be
Temperature determined from the two weighings as follows:
Solids are generally so insensitive to temperature fluctuations Q = Density of the liquid
that the corresponding density changes are of no consequence. A = Weight of the sinker in air
However, as work is performed with an auxiliary liquid in the Q= A-B B = Weight of the sinker in liquid
V
density determination of solids, their temperature must be taken V = Volume of the sinker
into account as the temperature has a greater effect with liquids P = Buoyancy of the sinker in the liquid
and causes density changes in the order of magnitude 0.I to ( P = A-B)
1% per °C. This effect is already apparent in the third decimal
place of the result. In DENSITY SETUP, set Mode to Liquid Density and enter sinker
volume in cc’s.
To obtain accurate results, we recommend that you always
take the temperature of the auxiliary liquid into account on all After weighing the sinker in air and then weighing the sinker
density determinations. immersed in liquid, the balance calculates the density of the
liquid and is displayed in grams/cc.
Air Buoyancy
1 CM3 of air weighs approximately 1.2 mg (depending on the
physical condition). As a consequence, in the weighing in air,
each solid experiences a buoyancy of this magnitude (the so-
called “air buoyancy”) per cm3 of its volume.
Temperature: 00.00
When Auxiliary Liquid is set to water, the temperature of the
water can be manually entered.
EN-36 Explorer Pro
Exit:
In the sub menu, will return to the previous menu.
In the main menu, will return to the last application mode.
Explorer Pro EN-37
Each of the submenus contain settings that will affect the operation AutoCal Delta: -100 - +100 (default is 0)
of the balance. Please review all submenus settings to obtain AutoCal Delta allows the internal calibration mass value to be
the best performance from the balance. Make the necessary adjusted. This permits calibrating the balance using an external
settings to suit specific needs. mass which is traceable to a certified standard.
3.5.1 Calibration Perform the internal calibration procedure. After the calibration is
Press the Menu button and select CALIBRATION. Press the Enter completed, zero the balance.
button, the CALIBRATION screen is displayed. Explorer Pro Place a certified mass equal to the span calibration value
balances offers a choice of five calibration methods: Internal
of the balance.
Calibration, Autocal Calibration, Span Calibration, Linearity
Calibration and User Calibration,
Compare the reading on the balance to the expected weight
being used. If the reading is above the expected weight, the
Linearity, Span and User calibration are disabled for Type
delta entered in digits is negative. If the reading is below the
Approved/LFT balances Class II and Class III.
expected weight, the delta entered in digits is positive.
CALIBRATION See following example:
Internal
AutoCal : Off Actual Weight Reading: 200.0014
AutoCal Delta (d): 0
Span Expected Weight Reading: 200.0000
Linearity
User
Delta Weight (d): 0.0014
User Cal Weight (g): 500.00 Delta weight in Digits: -14
Cal Test
Exit
Recalibrate using internal calibration. After calibration, place
the certified mass on the pan and see if the mass agrees with
the displayed value. If not, repeat procedure until internal
IMPORTANT !
calibration reading agrees with the certified mass.
DO NOT DISTURB THE BALANCE
DURING CALIBRATION.
Span Calibration:
Internal Calibration Span calibration utilizes two calibration points, one at zero
Calibration is accomplished using the internal calibration mass. load and the other at specified full span. See the following
Internal calibration can be performed at any time providing the table.
balance has warmed up to operating temperature and is level.
EN-38 Explorer Pro
CALIBRATION MASSES The display indicates the difference in weight between calibration
LINEARITY SPAN ONLY mass placed on the pan and the previous weight value which
CAPACITY MASSES MASSES was stored in the balance. After Cal Test, the display returns to
12000 g 5000g/10000 g 10000 g to the last application mode.
22000 g 10000g/20000 g 20000 g
32000 g 15000g/30000 g 30000 g 3.5.2 Balance Options
Press the Menu button, and select Balance Options.
It is recommended that masses must meet or exceed
ASTM Class 1 Tolerance. Calibration masses are Press the Enter button. The BALANCE OPTIONS screen is
available as accessories. displayed.
BALANCE OPTIONS
Language : English
Display Date/Time : Date/Time
Date Format : DD/MM/YY
Date : 00/00/00
Time Format : 12 Hour
Time : 00:00:00
Contrast :
Brightness :
Audible Signal : Off
Exit
Additional calibration values to be used are shown on the display
screen. The best accuracy is achieved using the mass closest Languages: English, Spanish, etc...(default is English)
to the full span value. After calibration, the display returns to the Allows the selection of English, Spanish, French, German or
last application mode. Italian as the operating language to be displayed.
Filter Level: Low, Medium, High (default is Medium) All application modes are displayed on this screen. Each mode
Sets the balance reading averaging level to a value of Low, may be set on or off. These modes are cycled through when the
Medium or High. A higher filter level will provide a more repeatable Mode button is pressed.
measurement.
3.5.5 Units
Stability Level: 0.5, 1.0, etc... (default is 1.0) Press the Menu button, and select Units. The UNITS screen
Sets the balance stability level for the stability indicator to either displays a list of the available weighing units. Depending on
0.5, 1.0, 2.0 or 5.0. A 0.5 setting is equivalent to end result of the model, some units are not available.
.5 display digit. A 5.0 setting is equivalent to end result of 5
display digits. UNITS
Milligram (mg) : Off
Gram : On
Auto Zero: Off, 0.5, etc... (default is 0.5) Kilogram : Off
Oz avdp (oz) : Off
Sets the balance auto zero level to either Off, 0.5, 1.0, 2.0 or Pound avdp (lb) : Off
Carat (ct) : Off
5.0. Auto Zero tracking compensates for the drift at zero load. Penny weight (dwt) : Off
Troy OZ (ozt) : Off
0.5d means 0.5 display digit drift compensation per second. Grain (GN) : Off
HK Tael (Hkt) : Off
5.0d means 5 display digit drift compensation per second. SG Tael (SGt) : Off
ROC Tael (ROt) : Off
Singapore Tael (SGt): On, Off (default is Off) 123.4 = .1234 x 1000 =.1234 x 103
Unit (SGt) = g x 0.02645547, displayed readability by 5. 12.34 = .1234 x 100 = .1234 x 102
1.234 = .1234 x 10 = .1234 x 101
ROC Tael (ROt): On, Off (default is Off) .1234 = .1234 x 1 = .1234 x 100
.01234 = .1234 x .1 = .1234 x 10-1
Unit (ROt) = g x 0.02666667, displayed readability by 5.
.001234 = .1234 x .01 = .1234 x 10-2
.000123 = .123 x .001 = .123 x 10-3
Explorer Pro EN-41
EXPONENTS
3.5.7 Print Options
Press the Menu button, and select Print Options.
E-3 Moves decimal point 3 places to the left.
E-2 Moves decimal point 2 places to the left.
Press the Enter button PRINT OPTIONS screen is displayed.
E-1 Moves decimal point 1 place to the left.
E0 Leaves decimal point in normal PRINT OPTIONS
position.
Auto Print : Off
E1 Moves decimal point 1 place to the right. Print Interval : Off
E2 Moves decimal point 2 places to the right. Stable Data : Off
Numeric Data : Off
E3 Moves decimal point 3 places to the right. Exit
LSD’s
LSD .5 Adds one decimal place display counts by 5’s.
LSD 1 Display counts by 1’s.
LSD 2 Display counts by 2’s. Auto Print: Off, interval, etc... (default is Off)
LSD 5 Display counts by 5’s.
When enabled, the Auto Print feature causes the balance to
LSD 10 Display counts by 10’s.
automatically output display data in one of three ways:
LSD 100 Display counts by 100’s.
continuously, at user specified time intervals, or upon stability.
OFF Turns off the auto print feature.
Interval Provides a user specified printing interval
3.5.6 Interface
Stable Provides printed data each time a stable
Press the Menu button, and select Interface.
reading is achieved.
Continuous Outputs data continuously.
Press the Enter button, The INTERFACE screen is displayed.
Exit: Press the Enter button GLP PRINT OPTIONS screen is displayed.
Will return to previous display.
EN-42 Explorer Pro
---------------SPAN CAL--------------
Cal: 400.000 g
Old:399.445 g Select either RUN, DELETE, or DELETE ALL. When RUN is selected,
Diff:-0.555 g a screen showing all stored libraries is displayed. Any stored
Wt. ref______________________ library can be recalled, enabled and run. All parameters from
Name_______________________ a previously run library are retained.
The default setting for all items on the LOCKOUT screen are Application Modes:
Unlocked.
Weighing: On
All others Off
3.5.11 Factory Reset
The FACTORY RESET allows parameters to be set to the default
Units:
values. Each menu item can be reset by setting to Yes. After
Grams On, all others Off.
accepting the changes, the balance will reset the selected menus
items. Global Reset will change all menus to a factory setting in
Interface:
one step.
Baud Rate: 2400
Parity: None
Press the Menu button, and select Factory Reset.
Data Bits: 7
Stop Bits: 2
Press the Enter button FACTORY RESET screen is displayed.
Print Options:
FACTORY RESET
Calibration : No Auto Print: Off
Balance Options : No Print Interval: Off
Readout : No
Application Modes : No Stable Data: Off
Units : No
Interface : No Numeric Data: Off
Print Options : No
GLP Print Options : No
Lockout : No GLP Print Options:
Global Reset : No
Accept Changes : No Balance ID=OHAUS
User Name = OHAUS
All others Off
Lockout:
All Unlocked
EN-44 Explorer Pro
When the balance is first turned ON and LFT has been previously
set ON, the initial display will indicate that LFT is ON.
The last digit on the weighing display will be highlighted in white Example of Sealing Method
and is used to indicate the auxiliary digit.
4.2 Troubleshooting
SYMPTOM PROBABLE CAUSE(S) REMEDY
Unit will not turn on. Power cord not plugged in or properly Check power cord connections.
connected to balance.
Incorrect weight reading. Balance was not tared before weighing. Press >O/T< with no weight on the
pan, then weigh item.
Balance out of level. Adjust leveling feet.
Cannot display weight in desired unit. Desired unit not enabled. Enable units in Units menu.
Cannot access desired application. Desired application mode is not enabled. Enable desired application in
Application Modes menu.
Unable to change menu settings. Menu locked. Verify that Lockswitch is in the Off
position, unlock menu in the
Lockout menu.
RS232 interface not working. Interface parameters not properly set up. Verify interface settings in RS232
menu correspond to those of
peripheral device.
Tare Errors
2.0 Balance is unable to stabilize within time limit after taring. Environment is too hostile or balance needs
recalibration.
Calibration Errors
3.0 Incorrect or no calibration mass used for calibration. Recalibrate with correct masses.
RS232 Errors
4.4 RS232 buffer is full.
User Errors
7.0 User entry out of bounds.
7.2 Number outside of display capacity.
CheckSum Errors
9.1 Bad factory checksum. If error persists, have the balance serviced.
9.2 Bad factory checksum. If error persists, have the balance serviced.
9.3 Bad factory checksum. If error persists, have the balance serviced.
9.4 Auto Cal data failed checksum. This failure will disable access to the Auto Cal feature (if installed).
9.5 Factory calibration data failed checksum.
9.6 Bad program checksum.
9.7 Bad CMOS checksum.
9.8 User calibration data failed checksum.
9.9 Factory temperature compensation data failed checksum.
Explorer Pro EN-47
4.6 Accessories
Description
5. TECHNICAL DATA
5.1 RS232 Commands
Command
Character Description
PM Application mode.
PV Version: print name, software revision and LFT ON (if LFT is set ON)
xAL Set Auto-Zero level to x. x = 0 for OFF, x = 1 for 0.5d, x=2 for 1.0d, x=3 for 2.0d, x=4 for
5.0d.
On Turns balance on
xM Set current Application mode to x. x = 1 for Weighing, x = 2 for Parts Counting, x = 3 for
Animal Weighing, x = 4 for Percent Weighing, x = 5 for Check Weighing, x = 6 for G/N/T,
x = 7 for Filling, x = 8 for Differential, x = 9 for Formulation, x = 10 for Statistical, x = 11
for SQC, 12 = for Density
xAW Set Animal Weigh Level to x. x = 1 for LOW, x = 2 for MEDIUM, x = 3 for HIGH.
AC Abort calibration
5.2 Specifications
Admissible Ambient Conditions
Use only in closed rooms
Ambient Temperature range: 5 ºC to 40 ºC
Atmospheric humidity: 80% rh @ to 30 ºC
Voltage fluctuations: –15% +10%
Installation category: II
Pollution degree: 2
Power supply voltage: 12 VAC, 50/60 Hz or 12 VDC, 1A
Weighing units *** g, kg, lbs, oz, oz t, ct, dwt, taels (3), mommes, gn, ti, N, custom unit
Application modes Basic Weighing, Parts Counting, Differential Weighing, Statistics, Formulation,
Filling, Animal Weighing, Percent Weighing, Check Weighing, Density
Determination, Gross-Net-Tare Weighing, & SQC.
Features RS232 Port, GLP Protocol, Selectable Language, Display text, Selectable
Displayed Information Settings, Selectable Environment Settings, Selectable
Auto-Print Settings, Integral Weigh Below Hook, Contrast and Brightness
Control, Protective In-Use Cover
Repeatability (Std. dev.) (g) 0.1
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the
warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be
defective at no charge, provided that the product is returned, freight prepaid, to Ohaus.
This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive
materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than
Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the
authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for
any consequential damages.
As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for
further details.
Explorer Pro EN-51
A Index
AC Adapter Installation 5 Features 3
Accessories 47 FILLING 20
Admissible Ambient Conditions 49 Filter Level 39
Ambient Temperature rang 49 Formulation 24
Animal Weighing 16
Application Modes 39
G
Applications 11 G/N/T WEIGHING 21
Atmospheric humidity 49 Global Part No. 47
Audible Signal 39 GLP Print Options 41, 42
Auto Print 41
Auto Tare 39 H
Auto Zero 39
AutoCal 37 Hardware 5
AutoCal Delta 37 Hardware Lockswitch 44
Average Piece Weight 12 I
B In-Use Display Cover Kit 47
Balance ID 42 Initial Calibration 6
Balance Location 4 INSTALLATION 3
Balance Options 38 Installation category 49
Balance Preparation for Density Measurements 31 Interface 41
Balance Settings 37 Internal Calibration 37
Baud Rate 41 L
Brightness 39
Languages 38
C LCD Display 9
Cal Test 38 Legal for Trade 39
CALIBRATE 6, 37 Leveling the Balance 4
CALIBRATION MASSES 38 LFT DEFAULT TABLE 44
Capacity 49 Library 42
CARE AND MAINTENANCE 45 LIMITED WARRANTY 50
CHECK WEIGHING 19 Linearity Calibration 38
Cleaning 45 Lockout 43
Connecting Power and Communications 5 M
Contrast 38
Cord Set 47 Menu 10
Custom Unit (Cst) 40
N
D
Data Bits 41 O
Date 38 Output Formats 5
Date Format 38 Overview of Controls 7
Density 30 Overview of Display Indicator 9
Differential Weighing 22
Display Date/Time 38 P
E Parity Bit 41
PARTS COUNTING 12
Error Codes List 46 PARTS COUNTING-CHECK 14
EXPONENTS 41 PARTS COUNTING-FILL 15
F PERCENT WEIGHING 18
Pollution degree 49
Factory Reset 43 Power On/Off 10
EN-52 Explorer Pro
Index
Power Pack 47
Power supply voltage 49
Print Interval 41
Print Options 41
Printer 47
Printer Cable 47
Printing Data 44
Project Name 42
R
Readout 39
Replacement Parts 47
RS232 Commands 48
RS232 Interface 5
RS232 Interface Cable 47
S
Safety Precautions 3
Selecting the Location 4
Service Information 47
Span Calibration 37
Specifications 49
Stability Level 39
Stabilization 10
Stable Data 41
Statistical Quality Control 28
Statistics 25
Stop Bits 41
Switch Plate 44
T
TECHNICAL DATA 48
Time 38
Time Format 38
Troubleshooting 45
U
Units 39
Unpacking 3
User Calibration 38
User Name 42
V
Voltage fluctuations 49
W
Weighing 11
Z
Explorer Pro EN-53
Index
EN-54 Explorer Pro
Explorer Pro EN-55
EN-56 Explorer Pro
Ohaus Corporation
19A Chapin Road,
P.O. Box 2033
Pine Brook, NJ 07058, USA
Tel: (973) 377-9000,
Fax: (973) 593-0359
*80250963*
PN 80250963A © Ohaus Corporation 2003 all rights reserved.