NEWSLETTER NO. 1081


March 10, 2008
Dear Sir or Madam,

Introduction of ProSound α10 Ver. 7.0

We are pleased to introduce ProSound α10 Ver.7.0 into the market. The performance
of ProSound α10 has been improved including newly released functions like
Trapezoidal Scan with the linear probes, Spatial Compound Scan with the convex
probes, Image Optimizer function for B-mode image and so on. The essence of these
upgrades is mentioned below as well as short remarks on the clinical settings.

Serial numbers of software Ver.7.0: M01871 and higher


Main appeal points from this upgrade:

(Image Quality and Performance points)

z New Spatial Compound Scan image for general applications using


standard convex probes like UST-9130.

z Wider view field of image using Trapezoidal Scan with the linear
probes.

z Quick adjustment of the image uniformity on B-mode using Image


Optimizer function. It is also available with AIP (Adaptive Image
Processing), SCS (Spatial Compound Scan) and Trapezoidal Scan
technologies.

z Multi Slice Imaging display and Flow 3D function are available with
the 3D/4D probes.

(Data Analysis and Operation)

z Newly added “Auto Doppler Angle Correction” is a very helpful tool


for cardiovascular examination.

z Flow information (Color Doppler/Power Flow) is added on the EFV


image.
-2-

Other main improvements of Ver.7.0:

A. Image quality improvement using the following technologies


1. Compound Impulse Transmission technology
2. Broadband Harmonic Imaging technology
These technologies were introduced from Ver.5.0.
The following probes are additionally connectable on the Ver.7.0.
- UST-52101: Tissue Harmonic Echo (THE) image quality is improved
within 8 cm from the surface.
- UST-9118: ExPHD function is available on this probe.
- UST-675P: ExPHD function is available on this probe.

The imaging performance of Trapezoidal Scan and Spatial Compound


Scan is improved with the following probes: (the higher frame rate is
available)
Trapezoidal Scan: UST-5411, UST-5412, UST-567, UST-5712, UST-5713T
and UST-5543
Spatial Compound Scan: UST-9130, UST-9115, UST-9135P and
ASU-1010

B. Newly added features & functions


1. Image Optimizer function for B-mode image:
A quick adjustment is possible to make a uniformity image of the
intensity.
2. Automatic Doppler Angle Correction for spectrum Doppler:
Placing the cursor on the Color Doppler, and the direction of flow is
automatically estimated and the scale of spectrum Doppler is
adjusted.
3. Memorized STC function:
It is possible to keep the STC curve by pressing one switch.
4. AIP Level function:
6 kinds of imaging parameters for AIP function can be switched at
once by using the AIP Level switch.

C. Improvements
1. Image Adjustment
a. The AIP function to adjust the AIP LEVEL is added (there are 6
selections).
b. Default position of Auto1P (Focus) is possible to set on the Preset.
c. The velocity range of spectrum Doppler is expanded for “pulse by
pulse” setting on the Triplex display mode.

2. CHE (Contrast Harmonic Echo)


a. Addition of lower frequency setting (1.5MHz transmission on
UST-9133)
It will improve a penetration of CHE image especially in combination
with the SonoVue from Bracco.
b. Addition of Trans-gastro radial endoscope CHE with UST-OLY-R5R1 (it is
produced by OLYMPUS).
-3-

c. The direct select function is available for “DDM” (Dual Display Mode).
d. The synchronized search function is added on the DDM.
It is working at the preset of Monitor Cine Memory proportion “1:1”.
e. The data on the cine memory is not erased when the Manual Flash
function is activated.
f. To keep the selected MI value for transmitting power is possible even if
the position of focus is moved.
g. The value of each step of Monitor Level can be changed evenly by 5%.
h. The value of Monitor GAIN can be changed from +/-30 to +/-40.

3. RT3D
a. Addition of MSI (Multi Slice Imaging) Mode: RT-3D option
(SOP-ALPHA10-4)
MSI Mode is useful to observe multi slice images together to
check the organ at once.
b. Addition of Flow 3D Mode: option (SOP-ALPHA10-35)
It is possible to observe flow information on the 3D image.
c. The data transfer on AVM (Auto Volume Measurement) is
improved.
It is possible to transfer the data result of AVM for other kind of
measurements.
d. The volume rate of 3D/4D image is improved up to 30 vols/sec.
e. The Vol Cutter function is added.
f. The selection of Depth Attenuation is modified.
g. The selection of Scan angle is added.
h. The 3D View Guide function is added.
i. The Image Resolution function is added.

4. eFLOW
a. Additional compatible probes (UST-5296 & ASU-1012)

5. ET & FMD
a. Data analysis program will be directly booted up from thumbnail
display screen.
b. The display of Vessel diameter is possible on the data acquisition
screen at ET and WI mode.
c. Acquire time for ET and WI is expanded up to 30 seconds or 20
beats.
d. The zoom function for wave form of vessel diameter change is
expanded on the data acquisition screen.
e. The memory function for measurement condition is added at the
DICOM data store.
f. The Manual Measurement function is added on the wave form of
vessel diameter change at ET function.
g. The polarity reverse function is added to detect the R-wave on
ECG. (ET and FMD)
h. The function to expand the wave form of vessel diameter change
is added. (ET and FMD)
i. The function to cancel the R-wave on ECG is added. (ET and FMD)
j. The setup method for the manual AI value is changed. (ET, FMD
and WI)
-4-

k. The function to display the wave form of primary derivation of


vessel diameter change is added.
l. The value of DATmax is displayed. (ET)
m. The shape of waveform and the result of measurement data are
combined. (1:1 for ET and WI)
n. Raw data output to the text file is possible. (ET and FMD)
o. The WI report is available.

6. EFV
a. The Flow, Power Flow and eFlow mode are added on the EFV
image. (Neither directional Power Flow nor directional eFlow mode
will be available on this EFV function.)

7. Physiological signal display


a. RESP Bar (Respiration bar) display will be automatically changed to
the sweep display after freezing the image.
b. Setup position to display the physiological signal on the B-mode
and Sweep mode becomes independent.
c. The function which is monitoring the stability of heart rate interval
on ECG is added.

8. File Archive
a. AVI codex: “Motion-JPEG AVI” is added.
b. The graphics on the line data viewer are added.
c. To transfer the line data to the server is possible.
d. The seeking bar to playback the data acquired by DV-800 is
displayed.
e. The image format is memorized to store the data for export
function.
f. It is possible to set the time limit to detect the R-wave for
capturing the motion image at ECG synchronized acquisition
mode.
g. It is possible to clear the space of cine memory at the data
acquisition by the manual store switch, and then stopped the
acquisition of data automatically when the cine memory becomes
full.
h. The “Time Counter” display is available in playing back the line
data. (Except for cine playback)
i. To capture and store the single frame image from the “Video Clip”
images is possible.
j. To adjust the imaging parameter is possible in playing back the
line data.
k. The automatic data erasing function for stored data in the HDD is
available at scheduled time.
l. The display format corresponding to Teaching File (excluding
patient data) is added.
m. It is possible to support to make a Beep sound manually at the
time to finish Video Clip.
-5-

9. Display
a. The display icon to show the active condition for the following
function: “BbH” for Broad band harmonics, “SCI” for Spatial
compound imaging and “AIP” for Adaptive imaging processing is
available on the monitor.
b. It is possible to select the frequency display for Transmitting and
Receiving at tissue harmonic echo (THE) and contrast harmonic
echo (CHE).
c. The velocity scale to find turning velocity is displayed on the color
bar at the directional Power Flow and the directional eFlow.
d. The L & R indication on the Body mark display is possible.
e. The number of registration for the Body mark display is increased
from 15 to 45.

10. DICOM
a. Structured Reporting
Structured Reporting is added for Abdominal application.
b. The format Monocrom 2 is added. (only on the preset)
c. The format RGB Pixel is added. (only on the preset)

11.Measurements
(General items)
a. SR (Structure Report) for Abdominal measurement is added.
b.CSV formatted file output function for Abdominal and Urology
calculation results is added.
c. Font size to display the measurement result is selectable.
d. The digit number to display the result of measurement is expanded.
(only for Distance measurement)
e. The function of data shift for the measurement is added.
f. The function for Report printout is improved.
① The Black and White digital printer is connectable.
② “None” display of the Signature is selectable.

(For Basic measurements)


p. New mark for “M. Time” measurement is added.
q. The automatic transcribing function of HR for SV/CO measurement
data is added.

(For special measurements)


r. It is possible to transcribe Volume data on the RT3D automatic
volume measurement (AVM) function for other measurement.
s. Corresponding to Wave Intensity measurement report function is
added.

(For Abdominal measurements)


t. The SOL measurement function is improved.
① To calculate the volume data is possible.
② The Graph display function is added.
-6-

(For Cardiology measurements)


u. The number of data storage is increased.
v. BSA, SVI and COI at the LVOT flow and RVOT flow measurement
are added.

(For Vascular measurement)


w. The flip function of the drawing for Plaque Score Block indication is
added.

D. Upgrade Kits

There are 4 upgrade paths as follows:


z Free software upgrade kit (Service Pack)
z Software upgrade kit for sale
z Hardware upgrade kit for sale
z Optional items

1. Free software upgrade kit (Service Pack): SP-A10-S702.


z Only the software upgrade is possible
z Compatibility with newly added optional probes is not available. Please
use the software upgrade kit for sale for addition of optional probes.
z User’s manual for Ver.7.0 is not attached.
(The user’s manual is sold separately.)

2. Software upgrade kit for sale (CD+FD version)


a. PM-A10-S702EXP: for general export units
b. PM-A10-S702USA: for United States units
c. PM-A10-S702CAN: for Canadian units

3. Software upgrade kit for sale (HDD version)

The printer driver and new released probe parameters are included.

(Compatible with EP49600 for CPU and S/N M00101 to M01352)


a. PM-A10-S702EXP-H: for general export units
b. PM-A10-S702USA-H: for United States units
c. PM-A10-S702CAN-H: for Canadian units
(Compatible with EP535300 for CPU or S/N M01353 and higher)
a. PM-A10-S702EXP-H10 for general export units
b. PM-A10-S702USA-H10 for United States units
c. PM-A10-S702CAN-H10 for Canadian units
-7-

4. Optional items
Following Optional items are newly added.
z Flow 3D, SOP-ALPHA10-35 (it is necessary to work with RT-3D
function to activate this function. Therefore, it is required the
“EU-9102” and “SOP-ALPHA10-4” for Light version. Additionally,
PM-A10-H013 is also required to upgrade from S/N: M00101 to
M01870.)
Flow 3D function does not work without this kit.
In case main unit lower than S/N: M01352 has CPU EP496000, it is
necessary to change the PC board with PM-A10-H012 option, and
then upgrade to the ver.7.0.2 using M-A10-S702EXP/USA/CAN
-H10. to make the high performance of 3D processing.

E. Notes:
Additional Upgrade information for the unit which is already shipped.
z For Flow 3D:
PM-A10-H013 is additionally required: For M00101 to M01870.

(Information for the bug)


On the RT-3D function:
The error may sometimes occur when the 3D loop playback is used
at the certain condition as Render mode: Lighted and 3D Smooth: On.
Therefore, please use the “3D Smooth” function at default preset
condition. The default preset of “3D Smooth” is “OFF” condition.
“Smoothing(3D)” function has no problem. This problem starts from
Ver.5.0 of ProSound alpha10.
(Details)
Setup condition; “Render mode”: Lighted, “3D Smooth”: On
Process; Change the Loop playback of Volume Cine Memory from “On”
to “Off”.
Result; The program for showing the 3D images is stopped at 50% of
possibility. And then it is not possible to show the 3D image.
The symptom is not recovered till the Power switch is turned off and on
again.

(Additional Notes)
The new function makes the data on the local HD automatically
eliminated after the certain period.
It is working without any notice to the operator for the system.
Therefore please be careful to setup this function to “ON”.
The factory default is “OFF” condition.
An error is caused when 3D volume data set is reconstructed on the
ProSound alpha 10 Ver.6.*.* and before with the data acquired by
SSD3500/SSD4000/ProSound alpha5.
However, it is fixed after the Ver.7.0.
-8-

Documents planned: Brochure: End of March


Specifications: PDF file: Available now

Special Note:
ProSound α10 of which serial number is M01706 or higher (Front-End PC board,
EP495000HH is incorporated) requires to connect optional Mechanical Scanning Probe
Connection Unit, EU-9124 instead of former EU-9109. This is mentioned in the
specifications of ProSound α10 Ver.7.0.

Yours very truly,

ALOKA CO., LTD.

Y. Kiyomura
General Manager
International Marketing and Sales Dept.

YK/Kk/ims
-9-

Notes for V7 enhancements:


A. Addition of ExPHD(THE) function for UST-9118 and UST-675
2D Image resolution is finely displayed at ExPHD(THE) condition compared to
fundamental image. Therefore, the inside of vessels will be presented more clearly
than before. However, the penetration to observe the uterus and prostate will not be
efficient.
There is THE setting on ProSound a7, but it is not available on ProSound a10.

Fundamental ExPHD Fundamental ExPHD

B. Addition of the frequency for Flow/Power Flow/eFlow at the UST-5296


This is only for tender specifications. Therefore, the reality of image will not be a
big improvement.

5.0 MHz 3.75 MHz 5.0 MHz 3.75 MHz

C. Modified the setting of THE High on the UST-52101

The resolution for near field of UST-52101 is improved on the THE setting “High”.
However, the global image quality on this probe is not widely changed. Other than
“High” condition, there is no merit to change the parameters.

D. Addition of eFlow on the ASU-1012


The penetration of ASU-1012 is better than UST-9118. Therefore, it has not any
problem on this probe. The transmitting frequency will be selected from 6.0, 5.0
and 3.75 MHz.
- 10 -

E. Auto Angle Correction for spectrum Doppler using Color Doppler display
This is the function to adjust the Doppler angle for spectrum display by using the
Color Doppler image. It is using the Color Doppler information, therefore the
information will not be correctly displayed in the following cases.

z No flow display on the image. It is not possible to estimate the direction of the
flow when the Color Doppler information is not displayed on the image even
though the switch is activated.

z No uniform color display like aliasing condition.

z Using the short-axis view for the vessel.

F. Eliminate the vertical line noise


The new scan algorithm is added to eliminate vertical line noise in the Color
Doppler. It independently builds up the frame data for B&W image and Color
Doppler image. The demerit of this function will be making a conflict between B&W
image and Color Doppler image at the low frame rate condition.
This function will be working for the following three probes; UST-5411, 9130 and
9135P.

Segment Scan : Ver.6.0 Frame by Frame : Ver.7.0


- 11 -

G. Image Optimizer
This is the function to adjust B-mode gain automatically by pushing one switch.

The flow of process is shown below;

Press the “Image Optimizer”

1) Acquired the histogram of intensity.


Adjust the intensity for whole image. 2) Cut off the brightest part for the
reference.
3) Refer to the following points;
- Refer to the brightness of vessels
Adjust the STC balance for whole gain adjustment.
- Not refer to the brightness of
vessels for STC and Angle Gain
Adjust the intensity for whole image. adjustment.

Adjust the Angle GAIN ILOW IHIGH


Cut off : lower thanILOW & higher thanIHIGH

Adjust the intensity for whole image.

Finish the process


Setup the items related to the Image Calculated the intensity of echo level

Optimizer: 4) Process the STC and Angle Gain


Brightness Level: Setup the target intensity
(Selection from 40 to 80)
Brightness Level: AUTO
The system estimates the intensity level
automatically. 5) Adjust the STC

It is working with STC, Angle Gain, AIP,


Spatial Compound and Trapezoidal scan
function without any problem.
There is some time gap to show the adjusted
image after this function is selected.
During the procedure, image without
frame correlation is displayed.
- 12 -

Image Optimizer
ON

G. Trapezoidal Scan
It is shown an angle of sector part and shift the size of linear display on the table.

Model number Change the size of


Angle of Trapezoid
of probe linear part
UST-5412 20° Yes
UST-541 20° Yes
UST-5543 15° Yes
UST-567 15° No
UST-5712 10° No
UST-5713T 10° No

H. Color Doppler on EFV

EFV: OFF EFV: ON


The presentation of Color Doppler on the EFV image is not the same as standard
Color Doppler because of the color process. It is difficult to use with general flow
mode in case of pulsatile flow. Therefore, if you want to show the structure of
vessel in the organ by color, we will recommend to use the Power flow and eFlow
- 13 -

for this purpose.

Flow Power Flow

The EFV function makes a focus and collects the information in the middle of the
image, therefore if there is a small Flow area and the position is out of middle
area, the width of image is automatically expanded till 60% when the EFV switch
is “ON”.

(Flow area: Not on the center) (Expand the area when EFV is ON)
- 14 -

i. Compound display function at Convex probes

The spatial compound function is working for Convex array probes also.
At this moment, the following four probes are supported: UST-9130, 9135P, 9115
and ASU-1010.
STD Compound

STD Compound

The angles to show the compound image are presented on the table below.
However, the default preset for the angle is 10 degrees.
Model number of
Angle for compound image
probe
UST-9130 5°to 20°
UST-9135P 5°to 20°
UST-9115 5°to 15°
ASU-1010 5°to 15°
UST-5411
5°to 20°
UST-5412
UST-533
UST-534
UST-536
UST-547 5°to 15°
UST-567
UST-5543
UST-5548
UST-5712
5°to 10°
UST-5713T
- 15 -

A vertical line or gap appears on the image when the side of probe is not well
touched to the body like the data below.

j. Flow 3D
It is possible to show the Flow 3D image at one volume data set when the 3D scan
function is used in Flow, Power Flow or eFlow mode. The 3D scan function is the
one volume data acquisition mode with slow scan.
The RT-3D option is required to integrate system with SOP-ALPHA10-35.
ProSound alpha10: SOP-ALPHA10-4 + SOP-ALPHA10-35
ProSound alpha10 Lite: SOP-ALPHA10-4 + EU-9102 + SOP-ALPHA10-35
If the high performance of 3D processing is required, it is necessary to change
the PC board with PM-A10-H012 option, and then upgrade to the ver.7.0.1
using PM-A10-S702EXP/USA/CAN-H10.

The items to be adjusted on the Flow 3D image are as follows;


z Color Map (3D): Selection from “A” to “H”
Setup the color of Color Doppler part on the 3D image. Setup the color of B&W
image part by the “Display Color (3D)”.
z Color Map (Ref A, B, C): Selection from “A” to “H”
Setup the color of Color Doppler part on the Ref A, Ref B and Ref C image.
- 16 -

z 3D Display Data: Selection “BW+Flow”, “Flow” or “BW”


Setup the combination of display between B&W image and Color Doppler on
the Flow 3D display.
z Flow Opacity Control: Selection from “1” to “16”
Setup the opacity of color. Setup the combination of Low Level, High Level and
Power on the Flow Opacity Control at preset menu. It is possible to set 16
steps. Then read the condition from the menu.
z Clip only BW: Selection “ON” or “OFF”
It is clipper function to cut the B&W image on the Flow 3D image.
z Rendering mode: Selection “Shaded”, “GVR” and “Brightest”
The “GVR” mode will be presented the edge of vessels clearly compared with
“Shaded” mode. It is a better effect to suppress the noise signal. However the
thin vessels are also eliminated. It is the default setting for Abdominal studies.
The “Brightest” mode will enhance the low speed flow vessels compared with
“Shaded” mode because of display the whole blood flow front of B&W image.
However the noise signal is also increased.

(Remarks)
The directional Power Flow and directional eFlow function are not supported.
The color map between standard B-mode with color Doppler display and Flow 3Dis
not synchronized.

K. Multi Slice Imaging


This function is added to the RT-3D function.
When “Multi Slice Imaging” is selected
on the group menu, the function of Multi
Slice Imaging (MSI) will be activated, and
then it is able to display the multi slice
images.
The items to be setup on the MSI function
are as follows;
- 17 -

z MSI Layout: Selection “2x2” or “3x3”


Setup the layout of the display.
z MSI Slice Number: Selection from “8” to “19”
Setup the number of sliced images.
z MSI Slice Distance: Selection from “0.5mm” to “10.0mm” (0.5mm steps)
Setup the interval of sliced images.
z MSI Slice Direction: Selection “Ref A”, “Ref B” or “Ref C”
Setup the direction of sliced images.
z MSI Pre/Next: No item selection
Make a shift of displayed images.

The whole lines are shown the slice


planes. Solid indications are shown
the currently displayed images. The
position of solid lines are shifted by
“MSI Prev/Next” function.

It is the image indicated


by green line.

l. Modify the factory default preset to fit the LCD monitor

Preset items Modified reason


Depth
Depth(CV/LN/PA) Changed the depth range from 10cm to 12cm because of
size of LCD monitor too large (only for OBST-TV)
ImageB, M1
Frame Rate, Frame Rate Decreased the frame rate to reduce the gap of lines (For
(ExPHD) Convex: Abdomen, AbdomenD.P, OBST, OBST-TV, GYN,
GYN-TV, Prostate, Kidney, Brachytherapy, Operation)
Frame Smoothing Changed the frame smoothing from “OFF” to “ON” to
reduce the gap of lines (LoExtArtCV)
IP Select(B), (THE),
(ExPHD)
- 18 -

Contrast Decreased the contrast setting from “66” to “74” because


of related to “Image Enhance” and “View Gamma”
function. (ExPHD of Small Part and Breast)
AGC Increased the AGC setting to enhance the edge.
“Off” to “1” (ExPHD of Abdomen, AbdomenH.F,
AbdomenD.P and Kidney): “2” to “3” ( ExPHD of
Intestine): “Off” to “2” (ExPHD of Small Part): “Off” to
“1” (ExPHD of Breast)
Relief Changed the Relief setting from “Low” to “Off” to reduce
the blocky image. (B/THE/ExPHD of Intestine, Placenta,
Prostate, Brachytheraphy and Operation; B/THE of
OBST-TV and GYN-TV; ExPHD of OBST、Breast)
Smoothing Changed the smoothing for good connection: “1” to “3”
(ExPHD of Abdomen, AbdomenH.F, AbdomenD.P and
Kidney): “5” to “8” ( Intestine ) : “5” to “6” ( ExPHD of
OBST-TV): “5” to “6” (ExPHD of GNY-TV): “10” to “7”
(ExPHD of Breast): “1” to “6” (Operation)
Image Enhance Changed the image enhance to “1” related to “View
Gamma” setting
“10” to “1” (ExPHD of Small Part): “12” to “1” (ExPHD of
Breast)
View Gamma Changed the setting of View Gamma to make a clear
image
“G” to “C” (ExPHD of Abdomen, AbdomenH.F,
AbdomenD.P, OBST, GYN and Kidney and also Carotid):
“G” to “B” (Cardiac B): “Linear” to “C” (ExPHD of Small
Part and Breast): “Linear” to “B” (Operation)
IP Select (D)

Contrast Drop down the setting of contrast from “11” to “9”


(CardiacB) for avoiding the saturation of Brightness.

FLOW
Color Map. Setting Changed the setting of Abdom from “D” to “A” to adjust
the color between CRT and LCD. (It is including the next
preset: Small Part, Breast, Prostate, Kidney,
Brachytherapy, Operation, Orthopedic and Hip Joint)
Edge Smooth Changed the setting from “Off” to “On” to reduce the
rough image on the edge of vessel. (Small Part, Breast,
Orthopedic and Hip Joint)
IP Select (Flow)
Smoothing Increased the setting to reduce the rough image. (Small
Part, Breast, Orthopedic and Hip Joint)

IP Select (Power)
Frame Corre, Smoothing Increased the setting to reduce the rough image. (Small
Part, Breast, Orthopedic and Hip Joint)

RT3D
- 19 -

3D Zoom, Ref A, B & C Changed the magnification of the zoom because of the
size of LCD viewing monitor.
3D Zoom: 120 to 80% (for Abdomen, AbdomenH.F,
AbdomenD.P, Intestine, GYN, GYN-TV & Kidney)
120 to 100% (for OBST, Fetal Heart, Fetal Brain,
Placenta, OBST 3D & Small Part)
Ref A, B &C: 120 to 100% ( for Abdomen, AbdomenH.F,
AbdomenD.P, Intestine, OBST, Fetal Heart, Fetal Brain,
Placenta, OBST 3D, GYN, Kidney & Small Part)

M. Modified points of Preset Default on ProSound alpha 10 Ver.7.0 (for LCD)

There are described the modification points of Preset Default between CRT monitor
and LCD monitor on the ProSound alpha 10 Ver.7.0 below;
(It is only shown the different points from Ver.6.0)

1. Depth (CV/LN/PA)
Changed the depth of B-mode at the OBST-TV application, because the size of
image displayed on the LCD screen using the transvaginal probe is too large
compared to the previous system.

Object Preset: OBST-TV


Title Sub-title Default of Ver.6 Default of Ver.7
Depth Depth 10cm 12cm
(CV,LN,PA)

2. Flow: Color Map, Setting


Changed the Color Map selection from “Abdom D” to “Abdom B” due to adjust the
color between CRT and LCD.

Object Preset: Abdomen, Abdomen H.F, Abdomen D.P, Intestine, OBST, OBST-TV,
Fetal Brain, Placenta, GYN, GYN-TV, Small Part, Breast, Prostate, Kidney,
Brachytherapy, Operation, Orthopedic and Hip Joint
Title Sub-title Default of Ver.6 Default of Ver.7
Flow Color Map, Setting Abdom D Abdom B

3. RT3D: 3D Zoom, Ref A,B,C


Changed the ratio of 3D Zoom considering the size of LCD monitor.

Object Preset: Abdomen, Abdomen H.F, Abdomen D.P, Intestine, GYN, GYN-TV
Title Sub-title Default of Ver.6 Default of Ver.7
RT3D1 3D Zoom 120% 80%
Ref A,B,C 120% 100%

Object Preset: OBST, Fetal Heart, Fetal Brain, Placenta, OBST 3D, Small Part,
Breast, Prostate, Kidney, Brachytherapy, Operation, Orthopedic and Hip Joint
Title Sub-title Default of Ver.6 Default of Ver.7
RT3D1 3D Zoom 120% 100%
Ref A,B,C 120% 100%
- 20 -

4. Image setting:
Described the changed points for LCD monitor.

Object Preset: Abdomen, Abdomen D.P


Title Sub-title Default of Ver.6 Default of Ver.7
Image-B, M1 Frame Rate High Med
Frame Rate(Ex) High Med
IP Select(Ex) AGC Off 1
Smoothing 1 3
View Gamma G C

Object Preset: Abdomen H.F


Title Sub-title Default of Ver.6 Default of Ver.7
IP Select(Ex) AGC Off 1
Smoothing 1 3
View Gamma G C

Object Preset: Intestine


Title Sub-title Default of Ver.6 Default of Ver.7
IP Select(B) Relief Low Off
IP Select(THE) Relief Low Off
IP Select(Ex) AGC Off 3
Relief Low Off
Smoothing 5 8

Object Preset: OBST


Title Sub-title Default of Ver.6 Default of Ver.7
Image-B, M1 Frame Rate High Med
Frame Rate(Ex) High Med
IP Select(Ex) Relief Low Off
View Gamma G C

Object Preset: OBST-TV


Title Sub-title Default of Ver.6 Default of Ver.7
Image-B, M1 Frame Rate Med Low
IP Select(B) AGC Off 1
Relief Low Off
IP Select(THE) AGC Off 1
Relief Low Off
IP Select(Ex) Smoothing 5 6

Object Preset: Placenta


Title Sub-title Default of Ver.6 Default of Ver.7
IP Select(B) Relief Low Off
IP Select(THE) Relief Low Off
IP Select(Ex) Relief Low Off
- 21 -

Object Preset: GYN


Title Sub-title Default of Ver.6 Default of Ver.7
Image-B, M1 Frame Rate High Med
Frame Rate(Ex) High Med
IP Select(Ex) AGC Off 1
Smoothing 1 3
View Gamma G C

Object Preset: GYN-TV


Title Sub-title Default of Ver.6 Default of Ver.7
Image-B, M1 Frame Rate Med Low
IP Select(B) AGC Off 1
Relief Low Off
IP Select(THE) AGC Off 1
Relief Low Off
IP Select(Ex) Smoothing 5 6

Object Preset: Cardiac B


Title Sub-title Default of Ver.6 Default of Ver.7
IP Select(B) View Gamma G B
IP Select(THE) View Gamma G B
IP Select(Ex) View Gamma G B
IP Select(D) Contrast 11 9

Object Preset: Carotid, Vein, LoExtArt LN, UpExtArtery


Title Sub-title Default of Ver.6 Default of Ver.7
IP Select(Ex) View Gamma G C

Object Preset: LoExtArt CV


Title Sub-title Default of Ver.6 Default of Ver.7
Image-B, M1 Frame Smoothing Off On

Object Preset: Small Part


Title Sub-title Default of Ver.6 Default of Ver.7
IP Select(Ex) Contrast 66 10(74)
AGC Off 2
Image Enhance 10 1
View Gamma Linear C
Flow Edge Smooth Off On
IP Smoothing 1 5
Select(Flow)
IP Smoothing Off 4
Select(Power)
- 22 -

Object Preset: Breast


Title Sub-title Default of Ver.6 Default of Ver.7
IP Select(Ex) Contrast 72 10(74)
AGC Off 1
Relief Low Off
Smoothing 10 7
Image Enhance 12 1
View Gamma Linear C
Flow Edge Smooth Off On
IP Smoothing 1 5
Select(Flow)
IP Frame Corre 5 7
Select(Power) Smoothing 1 4

Object Preset: Prostate, Brachytherapy


Title Sub-title Default of Ver.6 Default of Ver.7
Image-B, M1 Frame Rate(Ex) Med Low
IP Select(B) Relief Low Off
IP Select(THE) Relief Low Off
IP Select(Ex) Relief Low Off

Object Preset: Kidney


Title Sub-title Default of Ver.6 Default of Ver.7
Image-B, M1 Frame Rate High Med
Frame Rate(Ex) High Med
IP Select(Ex) AGC Off 1
Smoothing 1 3
View Gamma G C

Object Preset: Operation


Title Sub-title Default of Ver.6 Default of Ver.7
Image-B, M1 Frame Rate(Ex) Med Low
IP Select(B) Relief Low Off
Smoothing 1 6
View Gamma Linear B
IP Select(THE) Relief Low Off
Smoothing 1 6
View Gamma Linear B
IP Select(Ex) Relief Low Off
Smoothing 1 6
View Gamma Linear B
IP Frame Corre 13 7
Select(Power)
- 23 -

Object Preset: Orthopedic, Hip Joint


Title Sub-title Default of Ver.6 Default of Ver.7
IP Select(B) View Gamma Linear C
IP Select(THE) View Gamma Linear C
Flow Edge Smooth Off On
IP Smoothing 3 5
Select(Flow)
IP Frame Corre 11 7
Select(Power) Smoothing 3 4

End of documents