Vous êtes sur la page 1sur 5

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 54, NO.

6, JUNE 2007 1167

Simultaneous Fetal Magnetocardiography and


Ultrasound/Doppler Imaging
Hui Zhao, Mingli Chen, Barry D. Van Veen, Fellow, IEEE, Janette F. Strasburger, and Ronald T. Wakai*

Abstract—The difficulty of utilizing multimodality diagnostic demonstrate the feasibility of performing fetal biomagnetic
imaging techniques for fetal surveillance remains one of the recording and ultrasound/Doppler imaging simultaneously.
greatest challenges in providing enhanced prenatal care. In
this Letter we demonstrate the feasibility of performing fetal II. METHODS
magnetocardiography (fMCG) and ultrasound/Doppler imaging
simultaneously, using a multichannel SQUID magnetometer and a A. Characterization of the Interference
portable ultrasound scanner. Despite large magnetic interference
from the scanner, the implementation of simple noise reduction The ultrasound scanner used in this study was a Sonosite
procedures and appropriate signal processing techniques yielded TITAN (Bothwell, Washington), equipped with the obstetrical
fMCG recordings of sufficient quality for assessment of fetal and cardiac packages, electrocardioraphy (ECG) input, and a
heart rate and rhythm. A variation of reference channel filtering, 60-mm broadband (2–5 MHz) curved array transducer. Al-
referred to here as synthetic reference channel filtering, was
used to reduce nonstationary low-frequency interference. The though it is a portable scanner, it can perform all the major
combination of fMCG and/or fMEG with ultrasound/Doppler imaging modes required for fetal cardiac and brain imaging, in-
offers new possibilities for assessment of fetal well-being and fetal cluding 2D, zoom, color power Doppler, pulsed wave Doppler,
cardiac function. continuous wave Doppler, M-mode, and tissue harmonic
Index Terms—Echocardiography, echo/Doppler, fetal magneto- imaging. The magnetic measurements were made using a
cardiography, fetal monitoring, multimodality imaging, reference 3-axis fluxgate magnetometer (FVM-400, MEDA, Dulles,
channel filtering. Virginia) and a 37-channel SQUID magnetometer (Magnes, 4D
Neuroimaging).
For the initial characterization of the interference, the scanner
I. INTRODUCTION
was situated as it would be for a typical examination, approxi-
mately 1.0 m above the floor and about 1.0 m from the center
N the last few years, fetal magnetocardiography (fMCG) of the patient table. The SQUID magnetometer was centered
I and fetal magnetoencephalography (fMEG) have attracted
considerable interest due to their ability to noninvasively assess
over the patient table at a height of 1.0 m. The transducer was
placed on the patient table, several cm below the SQUID mag-
fetal cardiac and brain electrophysiology, respectively [1]–[4]. netometer. The resulting interference was dominated by large
A drawback of these techniques, which utilize extremely sensi- periodic interference of fundamental frequency 22 Hz and am-
tive SQUID magnetometers, is their high susceptibility to inter- plitude on the order of . This interference originated
ference from electronic equipment, which generally precludes predominately from the scanner, as it was highly dependent on
their simultaneous use in conjunction with most other fetal mon- the position of the scanner and was weakly affected by the po-
itoring methods. In particular, it has not been possible to record sition of the transducer. The interference was strongest for 2-D
echo/Doppler along with fMCG to investigate electromechan- imaging, was smaller for M-mode and Doppler imaging, and
ical coupling in the fetal heart or to directly monitor fetal be- was affected by the backlighting of the display. The display was
havior with 2d ultrasound during fMCG or fMEG recording. always adjusted to maximum backlighting, which minimized
Recently, several manufacturers have developed portable, the interference.
battery-operated ultrasound scanners, which generate much When the transducer was moved, the resulting low-frequency
less interference than ac-powered scanners. In this letter, we interference was extremely large, on the order of .
Channels closest to the transducer were often saturated. This
interference was clearly due to a permanent magnetic moment,
Manuscript received July 18, 2006; revised October 15, 2006. This work was as it was present even when the scanner was turned off.
supported in part by the National Institute of Health (NIH) under Grant R01 Rather than extensively characterizing the interferences as
HL63174, Grant R01 NS37740, and Grant R21 HD049022. Asterisk indicates
corresponding author.
a function of position and orientation of the transducer and
H. Zhao and M. Chen are with the Department of Medical Physics, University scanner, we estimated their respective magnetic moments,
of Wisconsin-Madison, Madison, WI 53706 USA (e-mail: rtwakai@wisc.edu). as described in the following paragraph. This allows the in-
B. D. Van Veen is with the Department of Electrical and Computer Engi-
neering, University of Wisconsin-Madison, Madison, WI 53706 USA.
terference to be estimated for any given source position and
J. F. Strasburger is with the Department of Pediatrics, Division of Cardiology, orientation.
Children’s Hospital of Wisconsin, Milwaukee, WI 53226 USA. To determine the magnetic moment of the transducer, we
*R. T. Wakai are with the Department of Medical Physics, 1300 Univer- first aligned the moment along the -axis of the fluxgate mag-
sity Ave., University of Wisconsin-Madison, Madison, WI 53706 USA (e-mail:
rtwakai@wisc.edu). netometer by orienting the transducer such that the -magnetic
Digital Object Identifier 10.1109/TBME.2006.889198 field component was maximal and the - and -components
0018-9294/$25.00 © 2007 IEEE
1168 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 54, NO. 6, JUNE 2007

were negligible. Measurements of the -magnetic field com-


ponent were then taken at several positions along the -axis.
Plotting this data on a log-log graph showed the expected
falloff. The magnetic moment was computed by fitting the data
to the equation for the magnetic field of a magnetic dipole;
it gave a magnetic dipole moment of .
The permanent moment of the scanner was determined in the
same manner and was found to be 0.58 . This strong
moment arose mainly from magnets used to latch the display
cover, which were subsequently removed with the aid of the
manufacturer.
The periodic interference from the scanner had to be mea-
sured with the SQUID magnetometer due to its lower ampli-
tude. The magnetic signal showed a falloff, implying that
the source was effectively quadrupolar. Empirically, the field
strength is described by , where .

B. Procedures for Reducing the Interference


Having identified the sources of the interference, the fol-
lowing procedures were implemented to minimize interference
in the fMCG recordings. The scanner was placed on the floor
as far away as possible from the SQUID magnetometer, but
within the reach of the operator, limited by the 2-m transducer
cable length. This required the operator to sit on a low chair in
order to operate the scanner. If the scanner was placed at a more
convenient, higher elevation; the amplitude of the interference
increased. For example, raising the scanner from the floor to
a height of 1 m increased the interference from about 4 pT to
about 12 pT.
The low-frequency interference was reduced by degaussing
the transducer using a video tape demagnetizer. This decreased
the magnetic moment by about a factor of 20; repeated de-
gaussing was unsuccessful in further reducing the moment.

C. Acquisition of Simultaneous fMCG and Ultrasound/Doppler


Recordings
Simultaneous fMCG and ultrasound/Doppler recordings
were obtained from 4 women with uncomplicated pregnancies
at gestational ages 26–40 weeks. The recordings were made
in a magnetically-shielded room (Lindgren RF Enclosures,
Fig. 1. (a) Power spectrum of recording in (b), showing low-frequency interfer-
Glendale Heights, Illinois) using the 37-channel SQUID ence and large, sharp spectral peaks at 22 Hz and harmonics. (b) Multichannel
magnetometer. tracing of a simultaneous fMCG/ultrasound recording taken from a subject at 35
Typically, the transducer was located 3–12 cm from the weeks after 1–80 Hz bandpass filtering. For clarity, only 1/3 of the 37 channels
are shown. Channel 28 is closest to the transducer. The sensor layout can be
SQUID magnetometer during imaging. The duration of the seen in Fig. 3. Notice that the interference is small in channel 34. (c) Recording
recordings was 5–10 min. During the last few minutes, the in (b) after notch filtering and synthetic reference channel filtering. In addition,
scanner was turned off so that a period of artifact-free recording maternal interference was removed using a matched filter in order to better show
the resolution of the fetal signal. P-waves (arrows) are visible in the recording.
was available for comparison.

D. Postsignal Processing Techniques


Time-domain high-pass filtering was used to remove the low-
A frequency-domain comb notch filter was used to remove frequency interference from the transducer; however, it was not
the periodic interference at harmonics of 22 Hz. Reference possible to remove the interference without attenuating such
channel noise cancellation was also tried because the reference low-frequency signals of interest as the T-wave. We, therefore,
channels, consisting of 3 orthogonal magnetometer channels investigated a phase sensitive filtering method, referred to here
located approximately 15.8 cm above the signal channels, were as synthetic reference channel filtering, in which reference chan-
dominated by the 22-Hz interference. The weights of the ref- nels were synthesized in software from the signal channels. This
erence channels were computed with a linear minimum mean filter is, thus, able to remove interference arising from sources
square error algorithm, using data in a 1-sec moving window. near the signal channels, as opposed to conventional reference
ZHAO et al.: SIMULTANEOUS FMCG AND ULTRASOUND/DOPPLER IMAGING 1169

Fig. 2. Removal of interference by signal processing from a channel dominated by periodic interference: (a) 1–80 Hz band-pass filtered signal; (b) reference
channel filter output of signal in (a); (c) comb-notch filter output of signal in (a). Removal of interference by signal processing from a channel containing very
large low-frequency interference: (d) 1–80 Hz band-pass- and notch-filtered signal; (e) signal in (d) after synthetic reference channel filtering using fixed weights;
(f) signal in (d) after synthetic reference channel filtering using recomputed weights.

channel filtering, which is intended to eliminate interference Some channels showed notably low interference even though
from distant sources. they were not farthest away. This reflects the vector nature of
The basic idea is to divide the rank-37 signal space into a the coupling between the magnetic interference and the detec-
low rank signal plus interference subspace and an interference tion coils, which vary from channel to channel in their orienta-
only subspace, using an appropriate spatial filter. Synthetic ref- tion as well as their position.
erence channels are formed by projecting the data onto the inter- Notch filtering was highly effective in removing the periodic
ference subspace, which is devoid of signal. The signal channels interference from the scanner Fig. 2(c). No distortion of the
are obtained by projection onto the signal plus interference sub- signal was evident. The results were fairly insensitive to the de-
space and generally still contain interference, although this can tails of the notch filter function. If the amplitude of the periodic
be greatly minimized by recording for a brief period with the interference was increased as a result of raising the scanner to a
scanner turned off. more convenient height, the notch filter was still able to effec-
An effective method of defining the signal plus interference tively remove the interference. Reference channel filtering was
and interference subspaces is to apply principle component also effective in removing the periodic interference; however,
analysis (PCA) to a brief time-series, , containing a an increase in noise was visually apparent Fig. 2(b).
representative fetal QRS complex [5]. The signal subspace High-pass filtering with a 1-Hz cutoff frequency removed
was defined as the space spanned by the first several principle much of the low-frequency interference due to movement of
eigenvectors of . To minimize signal distortion, the the transducer, but typically the residual interference was still
number of eigenvectors was chosen such that they accounted large Fig. 2(d). Further processing using the synthetic reference
for at least 95% of the signal variance; in the vast majority of channel filter provided a significant improvement [Fig. 2(e) and
cases 2 eigenvectors was sufficient. The interference subspace (f)]. The filter output was influenced by several factors. Using
was defined as the orthogonal complement. fixed weight factors derived from one segment of the recording
The synthetic reference channels were derived by performing generally resulted in smoother signals, while recomputing the
PCA on a 1-sec portion of the data, in which the residual inter- weight factors using a sliding window improved the suppres-
ference was strong in the signal channels. The reference chan- sion of large interference.
nels were chosen as the first few principle components, where In all subjects, the signal-to-noise ratio of the fetal QRS
the number of channels was chosen such that the corresponding complex was high enough in the processed recordings to allow
eigenvectors accounted for at least 95% of the interference vari- computation of fetal heart rate tracings Fig. 3(a). Waveforms
ance. The weight factors were computed from data in the same obtained by averaging 50 consecutive complexes from record-
1-sec window and were applied to the entire recording. We also ings taken during and after ultrasound imaging [Figs. 3(b) and
investigated the effect of allowing the weight factors to vary by (c), respectively] showed good consistency in all subjects.
recomputing them using a 1-sec sliding window. An example of a simultaneous echo/Doppler-fMCG tracing
is shown in Fig. 4. The traces were precisely time-aligned using
III. RESULTS a square wave timing signal connected to the ECG input of
The amplitude of both the low-frequency and periodic in- the ultrasound scanner and the fMCG data acquisition system.
terference varied markedly across the sensor array. Channels Using this procedure the tracings can be easily aligned to within
closest to the transducer showed the largest interference, while one pixel, which corresponds to approximately 2–10 msec, de-
channels farthest away showed much less interference (Fig. 1). pending on the sweep speed.
1170 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 54, NO. 6, JUNE 2007

Fig. 3. (a) Fetal heart rate tracing and (b and c) sensor layout display of averaged fMCG waveforms obtained from a subject at 35 week’s gestation. Two-dimen-
sional ultrasound imaging was performed from 0–305 sec. The signals were saturated by large artifact from 305–325 s after the scanner was turned off and moved
away from the magnetometer. The remainder of the recording was artifact-free. The waveforms in (b) were obtained by averaging 50 consecutive complexes in the
first part of the recording during ultrasound scanning and show good consistency with the averaged waveforms in (c), which were derived from the artifact-free
part of the recording.

IV. DISCUSSION
To our knowledge, this is the first study to demonstrate
the feasibility of performing fMCG and ultrasound/Doppler
imaging simultaneously. With appropriate signal processing,
demagnetization of the tranducer, and other simple measures,
it was possible to derive fMCG recordings of sufficient quality
for assessment of fetal heart rate and rhythm.
In this study we used an off-the-shelf scanner; however,
modest manufacturing modifications could significantly re-
duce the level of magnetic interference. The main desired
improvement is the elimination of all magnetic components
from the transducer and the scanner. Reduction of high-fre-
quency interference from the scanner would further reduce the
signal processing requirements and might allow fMCG to be
visualized in real-time.
The ability to combine fMCG and/or fMEG with ultra-
Fig. 4. Echo/Doppler recording of right ventricular inflow and outflow showing sound/Doppler offers new possibilities for fetal investigation.
simultaneous fMCG in the bottom tracing. The negative Doppler velocities rep- Simultaneous echo/Doppler and fMCG may allow assess-
resent inflow during diastole across the tricuspid valve; the positive velocities
represent systolic outflow into the pulmonary artery. Above the Doppler tracing
ment of cardiac function and electromechanical coupling
is a timing signal recorded through the scanner’s ECG input (grey) and through in utero with efficacy similar to that of echo/Doppler and ECG
the fMCG data acquisition system (white and opposite polarity), which is used postnatally. fMCG complements ultrasound/Doppler with its
to align the Doppler and the fMCG. The timing signal is square, but is dis-
torted by high-pass filtering. The amplitude of the QRS complex in the fMCG ability to assess beat-to-beat fetal heart rate variability and to
is approximately 3 pT. provide waveform information; fMEG can detect fetal brain
ZHAO et al.: SIMULTANEOUS FMCG AND ULTRASOUND/DOPPLER IMAGING 1171

activity directly. Currently, the large low-frequency noise is [2] H. Eswaran, J. Wilson, H. Preissl, S. Robinson, J. Vrba, P. Murphy,
problematic for fMEG, but eventually this may be overcome D. Rose, and C. Lowery, “Magnetoencephalographic recordings of vi-
sual evoked brain activity in the human fetus,” Lancet, vol. 360, pp.
through the use of nonmagnetic transducers and/or transducer 779–780, 2002.
stabilization. [3] P. van Leeuwen, B. Hailer, W. Bader, J. Geissler, E. Trowitzsch, and
D. H. Gronemeyer, “Magnetocardiography in the diagnosis of fetal ar-
rhythmia,” Br. J. Obstet. Gynaecol., vol. 106, pp. 1200–1208, 1999.
[4] J. M. Lengle, M. Chen, and R. T. Wakai, “Improved neuromagnetic
REFERENCES detection of fetal and neonatal auditory evoked responses,” Clin. Neu-
rophysiol., vol. 112, pp. 785–792, 2001.
[1] R. T. Wakai, “Assessment of fetal neurodevelopment via fetal mag- [5] M. Chen, R. T. Wakai, and B. Van Veen, “Eigenvector based spatial
netocardiography,” Exp. Neurol., vol. 190, no. Suppl 1, pp. S65–S71, filtering of fetal biomagnetic signals,” J. Perinatal Med., vol. 29, pp.
2004. 486–496, 2001.

Vous aimerez peut-être aussi