Académique Documents
Professionnel Documents
Culture Documents
Abstract - Anisotropic magnetoresistive sensor can be used to electrical isolation and protection of the sensor and
measure DC and AC current by sensing the field near the surrounding electronics. The use of the sensors' set/reset
conductor in a contactless way. In order to improve circuitry allows for ultra low offset and ultra low offset drift of
measurement accuracy, an analysis of the influence of the current measurements.
displacement of the examined wire is made. According to the
In order to improve measurement accuracy, an analysis of
result of the analysis, restriction on the displacement of the
examined wire is given. Current measurement method based on the influence of the displacement of the examined wire is
current carrying wire position estimation is presented. made. According to the result of analysis, restriction on the
Experimental results show that current measurement method displacement of the examined wire is given. Current
based on current carrying wire position estimation is of higher measurement method based on current carrying wire position
accuracy than that of direct summation method. For example, for estimation is presented. Experimental results show that current
sensor circular of 25.4 mm radius around the measurement hole measurement method based on current carrying wire position
of 10 mm radius, the maximum relative error of wire position estimation is of higher accuracy than that of direct summation
estimation method due to non-concentric wire is about 0.2%. method.
Index Terms - Anisotropic magnetoresistive sensor. Magnetic II. THEORETICAL ANALYSES FOR CENTERED CURRENT
sensor application. Contactless current measuremen. Sensor array. CARRYING WIRE
Position estimation.
In the following, we assume that the sensors (from 0 to N-
I. INTRODUCTION 1) are strictly vectorial, reacting only to the magnetic
component parallel to their sensitive axes. In addition, the
There are numerous methods used to measure current. sensitivities (S) will be considered to be identical for all the
The most common ways include the use of a sampling sensors. The sensors are placed in an ideal symmetry and
resistor, a transformer, or a contactless magnetic sensor. The equidistance around the measurement hole so that their centers
voltage drop across the sampling resistor which is in series are on one concentric circle (name it as "sensor circle"). The
with the load is proportional to the current going through the sensitive axis of each sensor is tangent to the circle at this
resistor. This offers good accuracy and low offset, but does point.
not provide electrical isolation and has high thermal drift. This The output of sensor i (from 0 to N-1) is responses to the
allows transient spikes to ruin the sensor and potentially respective magnetic field by a voltage
overload the electronics. The current transformer is made up
of a primary and a secondary coil wrapped around a magnetic VOi = SHSA,i (1)
core. The primary coil carries the current to be sensed and where HSA,i is the magnetic field component parallel to
induces a magnetic field in the core. A current is generated in the sensitive axe of sensor i.
the secondary coil that is proportional to the primary current According to the Ampere's law, the magnetic field
scaled by the turns ratio. The current transformer offers intensity at the distance d from a straight current line in a
electrical isolation, but only works for AC applications. homogeneous space is
Current transformers can also be large and bulky.
Often used configuration of contactless current
H=_I_
measurement devices comprise magnetic flux concentrators in 2nd (2)
the form of a ring around the hole, which the examined wire is For the centered position of the wire, the summed output
led through. A Hall sensor with the sensitive axis voltage of the device is
perpendicular to the cross-section of the ring is placed into an NSf
air gap formed in the ring. This type of current measurement VOSUM = NVoi = NSHsA = - -
21lR (3)
devices offers electrical isolation, but can also be large and
bulky. Anisotropic Magneto-Resistive (AMR) sensors offer a where N is the number of the sensors in the sensor array,
high sensitivity, small size, solid state solution. Further, these R is the radius of the sensor circle.
devices may be used for contactless current sensing, meaning From (3), the current being measured is calculated as
they won't break the electrical circuit [1, 2]. This allows for
Authorized licensed use limited to: UNIVERSITY OF ALBERTA. Downloaded on August 16,2010 at 16:43:31 UTC from IEEE Xplore. Restrictions apply.
2nR I I
1M =--VOSU M = KVOS UM HA =- - =----;============
NS (4) 2m1EA 21C~(R-rcos(a)Y +(O-rsin(a)Y . (6a)
4086
Authorized licensed use limited to: UNIVERSITY OF ALBERTA. Downloaded on August 16,2010 at 16:43:31 UTC from IEEE Xplore. Restrictions apply.
When the current carrying wire is not centered, the sum of sensors with non-identical sensitivities is greater than that for
the outputs of all the sensors is not equal to that for centered sensors with identical sensitivities. In order to reduce current
wire carrying the same current. The current measurement error measurement error, the sensitivity of magnetic sensor should
is produced. The current measurement relative error is be calibrated.
8J = 1M - I X 100% Table II
I (9) THE MAXIMUM MEASUREMENT RELATIVE ERROR OF-
CURRENT FOR SENSORS OF NON-IDENTICAL SENSITIVITIES
For four-sensor configuration with identical sensitivity, (FOUR-SENSOR CONFIGURATION, SB=SC=SD=S, SA=l.l S,)
the maximum measurement relative error of current is
illustrated in Table I. As r/R increases, 18/lmax increases ~ 0.01 0.02 0.03 0.04 0.05
Table I
7R 0.06 0.07 0.08 0.09 0.10
iR 0.01 0.02 0.03 0.04 0.05 7R 0.11 0.12 0.12 0.14 0.15
1&lmax (%) 1.0E-6 1.6E-5 8.1E-5 2.6E-4 6.2E-4 1&lmax (%) 0.32 0.35 0.39 0.43 0.48
7R 0.06 0.07 0.08 0.09 0.10 7R 0.16 0.17 0.18 0.19 0.20
1&lmax (%) 1.3E-3 2.4E-3 4.1E-3 6.6E-3 0.010 1&lmax (%) 0.53 0.58 0.64 0.70 0.77
7R 0.11 0.12 0.12 0.14 0.15 7R 0.21 0.22 0.23 0.24 0.25
1&lmax (%) 0.015 0.021 0.029 0.038 0.051 1&lmax (%) 0.84 0.92 1.00 1.09 1.20
7R 0.16 0.17 0.18 0.19 0.20 7R 0.26 0.27 0.28 0.29 0.30
18Jlmax (%) 0.066 0.084 0.11 0.13 0.16 18Jlmax (%) 1.30 1.42 1.55 1.69 1.84
7R 0.21 0.22 0.23 0.24 0.25 7R 0.31 0.32 0.33 0.34 0.35
18Jlmax (%) 0.19 0.23 0.28 0.33 0.39 1&lmax (%) 2.00 2.18 2.37 2.58 2.80
7R 0.26 0.27 0.28 0.29 0.30 7R 0.36 0.37 0.38 0.39 0.40
1&lmax (%) 0.46 0.53 0.62 0.71 0.82 1&lmax (%) 3.04 3.30 3.57 3.87 4.19
7R 0.31 0.32 0.33 0.34 0.35 7R 0.41 0.42 0.43 0.44 0.45
1&lmax (%) 0.93 1.06 1.20 1.35 1.52 1&lmax (%) 4.53 4.90 5.30 5.72 6.17
7R 0.36 0.37 0.38 0.39 0.40 7R 0.46 0.47 0.48 0.49 0.50
1&lmax (%) 1.71 1.91 2.13 2.37 2.63 1&lmax (%) 6.65 7.17 7.72 8.31 8.94
7R 0.41 0.42 0.43 0.44 0.45 B Current measurement based on current carrying wire
position estimation
1&lmax (%) 2.91 3.21 3.54 3.89 4.28 According to the output signals of magnetic sensor, the
position of current carrying wire can be estimated.
1R 0.46 0.47 0.48 0.49 0.50 The direction of HA , HB , He, HD can be represented by
their deviation angle
1&lmax (%) 4.69 5.13 5.61 6.12 6.67
qJA = atan2(HAT , HAN)
In order to evaluate the influence of non-identical (lOa)
sensitivities, we assume that SB=Se=SD=S, SA=l.lS. The qJB = atan2(H BT' H BN) (lOb)
maximum measurement relative error of current is illustrated
in Table II. Comparing the data of table II with that of table I, qJc = atan2(HCT' H CN) (IOc)
it can be seen that the maximum measurement error for
4087
Authorized licensed use limited to: UNIVERSITY OF ALBERTA. Downloaded on August 16,2010 at 16:43:31 UTC from IEEE Xplore. Restrictions apply.
({JD = atan2(HDT' H DN) (IOd) «« =~(O-xY +(R- yY (16b)
From the direction of H A and H B , we can calculate the
position of current carrying wire (x(A, B), y(A, B)). de =~(-R-X)2 +(0- yY (16c)
~COs(9'A +~J . do =~(0-X)2 +(-R- yY (16d)
x(A, B) = -R ( ) sInqJB
cos qJA - qJB The sensitivity variation of magnetic sensor at point A, B,
(lla)
C and D due to current carrying wire position deviation can be
~COS(9'B -~) . compensated. The sum of output of magnetic sensors should
be
yeA, B) = R ( ) sIn({JA
cos qJA - ({JB
(lIb) VOS U M = -«; ~ VOAT 2 + VOAN 2 + -dB ~ VOBT 2 + VOBN 2
From the direction of H B and He, we can calculate the R R
position of current carrying wire (x(B, C), y(B, C)).
+ -de ~ VOCT 2 + VOCN 2 + -d D ~ VODT 2 + VODN 2
~cos (9'c - ~) . R R
Then the current can be calculated using (4).
.(17)
x(B, C) = -R ( ) SIn qJB Theoretically.the measurement error can be reduced to
COS ({JB - ({Jc
(12a) O.
y(B,C)=-R
hCOs(9'B +~)
(
.
) slnqJc
III. EXPERIMENTAL RESULTS
Eight AMR sensors (HMCI02IS) were placed on a
COS qJB - qJc printed circuit board, realizing a circular arrary of 25.4 mm
(12b)
From the direction of Heand H D , we can calculate the radius around the measurement hole of 10 mm radius. Four
position of current carrying wire (x(C, D), y(C, D)). symmetrically distributed sensors on the upside of the board
are of their sensitive axises directed to tangential direction and
~cos ( 9'c + ~) . four symmetrically distributed sensors on the downside are of
their sensitive axises directed to normal direction.
x(C, D) = -R ( ) sInqJD
cos qJc - ({JD
(13a) The sensors are flipped periodically at 400 Hz. The output
of magnetic sensor is a square wave signal with amplitude
~COS(9'D -~) . proportional to the magnetic field intensity. The output signal
of each sensor is amplified by precise instrumentation
y(C, D) =R ( ) sln({Jc
amplifier AD623 and then demodulated [5]. The demodulated
cos ({Jc - ({JD
(13b)
signals are led through two multicore cables to the signal
From the direction of HD and HA , we can calculate the processing board, where they are converted into digital
position of current carrying wire (x(D, A), y(D, A)). quantities. On the signal processing board, microcontroller
~cos ( 9' ~) A - •
C805lF061 which has 16-bit 1 Msps ADC is used. The signal
processing board is connected to computer through RS-232.
xeD, A) = R ( ) sm e., The influence of the wire position is illustrated in Fig. 2.
cos ({JD - ({J A
(14a) The measurement was performed with a constant DC current
y(D,A)=-R
~COS(9'D +~)
(
.
) Sln({JA
of 3A and the current carrying wire moved around the edge of
the measurement hole. In order to eliminate the influence of
geomagnetic field, geomagnetic field intensity is measured
COS ({JD - ({JA
(14b) when current being measured is zero. From Fig.2, it can be
The position of current carrying wire can be estimated seen that the maximum relative error of wire position
using (x(A, B), y(A, B)), (x(B, C), y(B, C)), (x(C, D), y(C, D)) estimation method due to non-concentric wire is about 0.2%
and (x(D, A), y(D, A)). (absolute value), which is better than that of direct summation
1 method (2.6%). The residual error is caused mainly by cross-
x=-(x(A, B)+x(B, C)+x(C, D)+x(D, A)) axis effect of AMR sensors and asymmetries in the probe
4 .(15a)
layout.
1
y = -(yeA, B)+ y(B, C)+ y(C, D)+ y(D, A))
4 .(15b)
The distance between current carrying wire and point A,
B, C, D can be calculated.
4088
Authorized licensed use limited to: UNIVERSITY OF ALBERTA. Downloaded on August 16,2010 at 16:43:31 UTC from IEEE Xplore. Restrictions apply.
0.2 , - - - - - - - - - - - - - - - - - - - - - - - - , has been presented. It is based on current carrying wire
position estimation. Experimental results show that the
maximum relative error of wire position estimation method
O. 1 f----- ~ 1!----- ~ -----__:_:-----I~
due to non-concentric wire is much lower than that of direct
summation method.
e
~
~
R EFERENC ES
'"
'"
>
[I] ] 1. 1. Yao, J. Wang , 1. H. Wang , Y. S. Geng , " Investigation on the theory
~ -0 .1
of interference rejection for current measurement with magnetic sensor
'"'" arrays " , Proceedings ofthe CSEE, vol. 24, no. 12, pp. 160-164 ,2004.
- 0. 2 f--- - - - - - - - - - - - - - - - -----'- --j [2] 1. Wang , Y. S. Geng, 1. 1. Yao, 1. H. Wang , " Investigation on the signal
processing algorithms applied in the multi-sensor system measuring
electric current", Proceedings of the CSEE, vol. 25, no. 22, pp. 130-135,
- 0.3 L- -----' 2005.
Angl e [deg] f3] P. Mlejnek, M. Vopalensky, P. Ripka, "AMR current measurement
devi ce", Sensors and Actuators A, vo1.141, pp. 649-653, 2008.
Fig. 2. Relativ e error of current measurement, f4] P. Ripka, P. Kaspar, K. Draxler, " Precise DC current sensors " , IEEE
when the wire rotates around the measurement hole . Radius of the Instrumentation and Measurement Corfe rrence, Brussles, Belgium, June
measurement hole 10 mm, radius of the sensor circle 25.4 mm. 4-6 ,1996, pp. 1479-1483.
[5] X. S. Li, J. X. Wang, C. Tang, L. J. Wang , S. C. Wang , "Development of
IV. CONCL USIONS high accuracy magnetic electronic compass", Chinese journal of sensors
and actuators, vol. 19, no. 6, pp. 2441-2444, 2006.
An innovative contactless current measurement method
4089
Authorized licensed use limited to: UNIVERSITY OF ALBERTA. Downloaded on August 16,2010 at 16:43:31 UTC from IEEE Xplore. Restrictions apply.