M.A. van Straten and H.A. Claessens Organic acids are an important class of com-
Eindhoven University of Technology pounds that must be analyzed in different samples,
Dept. Chemical Engineering and Chemistry for example, fruit drinks, wines, and many other
Laboratory of Polymer Chemistry aqueous samples. Under most reversed-phase
P.O. Box 513 liquid chromatography (RPLC) conditions, organic
5600 MB, Eindhoven acids are difficult to retain and hence to analyze.
The Netherlands
For this type of analysis, a number of major issues
A. Dams must be considered: Under most usually applied
Agilent Technologies B.V. reversed-phase separation conditions these acids
P.O. Box 667 are ionized and show little or no retention. The
1180 AR, Amstelveen application of ion suppression chromatography
The Netherlands (ISC) is a convenient and straightforward
approach to increase retention and selectivity. In
ISC the pH of the eluent must be about 2 pH units
Abstract lower than the pKa-values of the analyte acids.
Under such conditions, acids are in their molecu-
The ZORBAX SB Aq column was successfully used to sep- lar form and can be retained and separated.
arate organic acids by Ion Suppression Chromatography
at low pH using Reversed Phase Liquid Chromatography In addition, under ISC-conditions the eluent is
columns and Diode Array Detection. Two standard mix- usually nearly completely aqueous, which may
tures of organic acids were separated and a number of result in collapsing many standard bonded station-
softdrinks and wines were analyzed. ary phases. Specific RPLC-phases, for example, the
ZORBAX SB Aq column can easily be applied
under completely aqueous conditions. Further-
more, under acidic eluent conditions, many RPLC-
columns are not very stable over an acceptable
period of time.
It was found, however, that the use of bulky side Chromatographic Conditions
chains significantly improves column/bonded
phase longevity under acidic conditions [1]. This The different organic acids were analyzed in the
concept of stable bond technology has resulted in a ISC mode. The eluent pH was adjusted at least
number of StableBond (SB) reversed phase 2 pH units below the pKa-value of the acids. Detec-
columns. This group of columns has already tion was performed using Diode Array Detection
proved their substantial column longevity under (DAD).
acidic eluent conditions [2]. The ZORBAX SB Aq
column was specifically designed to operate under Conditions
completely aqueous conditions without phase col-
lapsing. At the same time the column longevity Column: 4.6 mm × 150 mm × 5 µm
under acidic conditions is excellent. ZORBAX SB-Aq
Mobile phase: 20 mM aqueous
phosphate buffer pH
Experimental 2.0/acetonitrile=99/1(v/v)
Flow rate: 1.0 mL/min
The organic acids selected for study are listed in Injection volume: 1 µL
Table 1. Column temperature: 25 °C
Detection: UV-DAD detection wave-
Table 1. Organic Acids Selected for Study
length/window 210/8 nm,
Acid MW Formula Std # RT (min) reference wavelength
Tartaric 150.09 C4H6O6 1, 2 1.86 360/80 nm
Malic 134.09 C4H6O5 1, 2 2.15 Sample preparation: Filtration over 0.45-µm
Lactic 90.08 C3H6O3 1 2.23 filter
Ascorbic 176.12 C6H8O6 2 2.28
Acetic 60.05 C2H4O2 1, 2 2.42 Equipment
Citric 192.12 C6H8O7 1, 2 2.94 • Agilent 1100 series HPLC
Succinic 118.09 C6H4O6 1 3.19
Maleic 116.07 C4H4O4 2 3.47 • Vacuum degasser
Fumaric 116.07 C4H4O4 1, 2 4.50 • Quaternary pump
• Autosampler
Samples and Sample Preparation
• Thermostatted column compartment
Two standard solutions were prepared due to
co-elution of lactic acid and ascorbic acids: • DAD
2
Table 2. Retention Times (min)
Linearity
3
Area
Area = 3006.06548*Amt _11.226386
6000
7
5000
4000
6
3000
2000
5
1000
4
3
0
1, 2 Correlation: 0.99994
0 1 2
Amount (mg/mL)
Chromatographic Examples
mAU
4.502 - Fumaric acid
3.189 - Succinic acid
1.857 - Tartaric acid
60
50
40
30
20
10
7.140
7.513
1.660
1.512
2 4 6 8 min
4
0
20
40
60
80
100
120
mAU
0
20
40
60
80
100
120
140
mAU
Figure 4.
Figure 3.
1
1
1.861 - Tartaric acid
Orange juice.
2
2
2.156 - Malic acid
Malic acid 2.282 - Ascorbic acid
Ascorbic acid
2.417 - Acetic acid
Standard mixture 2.
2.935 - Citric acid
3
Citric acid
Succinic acid 3.473 - Maleic acid
4
Fumaric acid 4.504 - Fumaric acid
5
5
6
6
7
7
8
8
9
9
min
min
5
6
0
mAU
0
20
40
60
80
100
mAU
10
20
30
40
50
60
70
80
Figure 6.
Figure 5.
1
Apple juice.
White wine.
2
2
Tartaric acid
Malic acid
Lactic acid Malic acid
Acetic acid
3
Succinic acid Succinic acid
4
Fumaric acid Fumaric acid
5
6
6
7
8
8
9
min
min
Conclusions
• Ion suppression can be used to retain and
analyze for organic acids.
• The ZORBAX SB-Aq column separates organic
acids with good peak shape and retention at
low pH (ion suppression) and under high
aqueous conditions.
• This column has an alkyl bonded phase struc-
ture and is also designed to prevent phase col-
lapse. At the same time, very hydrophilic
compounds are retained under completely
aqueous conditions.
• The ZORBAX SB-Aq column has a unique selec-
tivity and resolution for very polar compounds.
References
1. J.J. Kirkland, J.H. Glajch, and R.D. Farlee,
(1989) Anal. Chem., 61, (1): 2-11.
2. H.A. Claessens, (2001) Trends in Analytical
Chemistry, 20, no. 10.
7
www.agilent.com/chem
Agilent shall not be liable for errors contained herein or for incidental or consequential
damages in connection with the furnishing, performance, or use of this material.