Daisuke Fujisawa / Regional Market Manager / Vaisala / Tokyo, Japan

Humidity
measurement
in cleanrooms
Choosing the right sites, such as expansion and contrac- In food processing plants, it is
tion, and hardening and softening necessary to keep the manufacturing
type of measurement of material, change in viscosity of
liquid, growth of microbes, increase
site at or below certain humidity. For
example, 40% or below seems to be
instrumentation is in static electricity, and corrosion
and rust, are largely affected by
a commonly used value. This helps
in restricting the growth of germs
humidity. and bacteria that can cause food
important in order poisoning.

to reach the best Dewpoint In semiconductor and electronics


product plants, the generation of
Dewpoint (Td) is a temperature at products changes more and more
humidity measurement which dew, or condensation, forms rapidly. As a result, the control
on cooling a gas. Dewpoint is a of humidity and dewpoint in the
results. Calibration parameter suitable for expressing manufacturing process has become
very small water content in a gas like stricter. In the manufacturing mini-
should also be carried air. In the micromachining of semi- environments, very high level control
conductors the conditions are very with an accuracy of +/-1%RH is often
out regularly, and to dry as water molecules are regarded required.
as contaminants. In this condition Humidity control is also impor-
traceable standards. relative humidity is practically stag- tant at liquid crystal display plants
nated at 0 %RH but dewpoint scale and paint plants. In this case, the
Products manufactured in clean- is still sensitive for water content durability and accuracy of the
rooms cover a wide range, including changes in the measured gas. humidity sensor is very important.
pharmaceuticals and semicon- These plants generate various gases,
ductors. Humidity, temperature, Different applications, which can affect sensor elements.
particles and pressure are often different needs
controlled, as these parameters Humidity and dewpoint
can have serious effects on product A pharmaceutical manufacturing sensor technologies
quality and production efficiency. plant often has a large number
of cleanrooms. The control and Humidity sensors, which measure
Relative humidity recording of temperature and water content in the air, are
humidity is strictly designated by broadly divided into two types.
Relative humidity (%RH) describes GMP (Good Manufacturing Practice). One measures humidity and the
the amount of water vapor that exists The most important feature required other dewpoint. In an atmosphere
in a gaseous mixture of air and water. from humidity sensors is small devia- where the humidity level is at least
It is a ratio of the amount of water tion. It is important to be able to 10%RH, humidity measurement is
vapor present compared to how perform precise calibration to check often used, while in low humidity,
much could be present at a given that the sensor does not drift over dewpoint measurement is preferred.
temperature. Issues at production the long-term. In some cases it is convenient to use

12 180/2009
dewpoint measurement even in high trodes. The reel is coated with lithium
humidity conditions. chloride. An alternating current is
passed through the winding and the
Humidity and dewpoint lithium chloride solution, causing
sensors include: resistive heating. As the reel heats,
water evaporates from the lithium
1. Psychrometer chloride solution at a rate which is
2. Mechanical hygrometer controlled by the vapor pressure of
3. Lithium chloride dewpoint indi- water in the surrounding air. When
cator the reel begins to dry, the resistance
4. Resistance type hygrometer of the lithium chloride solution
5. Capacitance type hygrometer increases, and less current flows
(dew indicator) through the winding. This allows the
6. Mirror dewpoint indicator reel to cool. This heating and cooling
of the reel reaches an equilibrium
Sensors 1-6 can measure general point where it neither takes on nor
humidity levels. Sensors 5 and 6 gives off water, and the equilibrium
are also used for low dewpoint temperature is directly proportional
measurement. The principle of each to the dewpoint of the surrounding air.
technology is described briefly in the 4. A resistance type hygrometer
following. utilizes the principle that electrical
1. A psychrometer is a simple form of resistance varies in a material that
a hygrometer, which consists of two absorbs moisture. Special sensors
thermometers. One has a dry bulb are used to measure the resistance Vaisala’s own cleanroom produces sensors
and the other a bulb that is kept wet to a current passing between wires. for radiosondes as well as different humid-
ity, barometric pressure and carbon dioxide
to measure wet-bulb temperature. This type of sensor is suitable for measurement products.
The wet bulb cools by evaporation of mass production and seems to be
the water. The amount of evapora- most used for home appliances and Sensors mostly used in clean-
tion, as well as cooling of the ther- consumer products. However, it may rooms include the resistance type
mometer, depends on the humidity of not measure accurately in very low hygrometer, capacitance type
the atmosphere. This data, together or very high humidity environments. hygrometer (dew indicator) and
with humidity tables or calcula- 5. A capacitance type hygrometer mirror dewpoint indicator. When
tions, is used to determine the vapor measures humidity by detecting selecting a suitable instrument, it is
pressure of water in the surrounding the change in capacitance of a thin important not only to pay attention
air, and relative humidity. This is a polymer film. This type of sensor can to the price and product specifi-
method often used in laboratories easily achieve sufficient accuracy, cations, but also to consider the
and humidity and temperature test and is mostly used in industry. measurement accuracy, manufac-
chambers. The patented HUMICAP® humidity turer’s application knowledge and
2. A mechanical hygrometer sensors manufactured by Vaisala use services available. All these factors
measures and records humidity using this technology. contribute to the actual user-experi-
an instrument that expands and 6. A mirror dewpoint indicator ence and operational success.
contracts with humidity changes, utilizes the occurrence of dew at
such as human hair. This type of dewpoint temperature when air Regular traceable
measurement has been used for containing water vapor is cooled. A calibration is important
a long time. The accuracy of the mirror is cooled until it reaches the
method is not very good. dewpoint of the gas in question. As One should always make sure that
3. A lithium chloride dewpoint dew condensation forms, it changes the data produced by the measure-
indicator is a measurement principle the light reflected from the mirror. ment equipment is reliable and
based on the hygroscopic character- When the mirror surface reaches an accurate. Periodic calibration is
istic (ability of a substance to attract equilibrium state whereby evapora- absolutely essential. Typical cali-
water molecules) of lithium chloride. tion and condensation are occurring bration intervals can be viewed in
The sensor consists of a reel covered at the same rate, the temperature of table 1.
with an absorbent fabric and a bifilar the mirror is equal to the dewpoint Table 2. presents an example
winding (two insulated wires, with temperature of the tested gas. of a traceability chain for installed
currents traveling through them in This type of sensor is often used in humidity and temperature units.
opposite directions) of inert elec- research institutes. From a global perspective, all

180/2009 13
measurements are based on the Table 1. Typical calibration intervals for measurement equipment.
globally agreed International Month
Measurement equipment
System of Units (SI). This ensures 6 9 12 24 36 60
that we use the same quantities, Mechanical pressure meters
and that measurements performed Precision barometers
with various types of equipment in Barometers
various locations are comparable. Liquid-in-glass thermometers
National laboratories are respon- Resistive temperature sensors and thermoele-
sible for maintaining and developing ments/thermometers

traceability and for providing the Dewpoint meters

highest accuracy calibrations. The Humidity meters

calibration services of the National Active electrical meters

Measurement Standard Laboratories Passive electrical meters


Lenght measurement equipment
may be limited to calibration of the
Lenght measurement equipment with electrical
highest grade primary standards. display
Commercial calibration services
suitable calibration interval
provide calibration services for
lower level standards and measure- Table 2. Example of a traceability chain for installed humidity and temperature measurement units.
ment equipment. These may be
International SI–Units SI–Units
manufacturer services providing Pressure Temperature
level
calibration services for their own
National Measurement
products, or laboratories providing Standards Laboratory National Standard National Standard
calibration services for any measure-
ment equipment. Non-accredited In-house Laboratory
Primary Standard Primary Standard
calibration services are the majority
service providers, including most of
the measurement equipment manu- Primary Humidity generator

facturers’ calibration services and a


considerable amount of commercial Working Standard Working Standard
calibration services. Without accredi-
tation the competence of these Customer
Calibration for humidity and temperature instruments
services is not proven. Before use,
the competence should be confirmed
by auditing the service. organization should be founded. The temperature, dewpoint and humidity
Each calibration service provider organization may contain just one instruments. Services are available
must maintain an effective trace- person or a whole department with through regional service centers, and
ability chain. At the very least, the management and calibration staff. available for both already installed
primary standard must be cali- Laboratory calibration is units and together with the delivery
brated at an outside laboratory and preferred to field calibration. In a of new units.
then used for calibrations. Some laboratory, the effects caused by the You can order your own Vaisala
commercial calibration services do environment can be minimized, and Calibration Book free of charge at
not include uncertainty estimations the number of factors influencing the www.vaisala.com/calibrationbook .
in their calibration certificates if not calibration are reduced significantly. The book contains useful informa-
ordered separately. Some calibration Field calibration is a quick and tion on everything you need to know
services are not able to calculate easy way of checking measurement about calibration.
uncertainty at all. One should always equipment without having to remove
consider the competence of these it from the process or process area. Further information:
services. Field calibration requires a working www.vaisala.com/humidity
Sometimes it is practical to standard as a reference. This working www.vaisala.com/dewpoint
maintain an in-house calibration standard can be hand-held or some
system. This may be the case if the other equipment used for calibrating References:
measurement equipment is difficult the instrument installed in the Arun S. Mujumdar; Handbook of
to transfer (calibration on-site) or process. Working standards are cali- Industrial Drying (2006)
when the amount of calibrated equip- brated at a higher level laboratory. Vaisala Calibration Book (2007)
ment is high. To set up an in-house Vaisala has accredited calibra-
calibration system, a suitable tion services for Vaisala pressure,