Vous êtes sur la page 1sur 6

FUNDAMENTALS OF LIQUID MEASUREMENT III

CLASS 2180

Peter W Kosewicz
Hydrocarbon Measurement Consultants, Inc.
P.O. Box 590347
Houston, Texas 77259

INTRODUCTION CUSTODY TRANSFER

We've learned when measuring crude oil that liquids In simple terms, "custody transfer" is the transfer of
expand and contract with increases and decreases ownership or responsibility for a liquid hydrocarbon
in temperature. The liquid volume also decreases from one party to another. Since ownership is being
when pressure is applied. All these effects are part transferred, either immediately or eventually, it is
of the physical properties of liquid petroleum fluids. essential that accurate accounting be used so that
We learned in Fundamentals of Liquid Measurement all parties involved in the transaction receive their
I how these physical properties effect the fair measure. With the prices for hydrocarbons these
measurement of liquid hydrocarbons. The objective days it is obvious how important accurate
of either static measurement or dynamic accounting---hence, accurate measurement
measurements is to determine the quantity and becomes. The words "custody transfer" have
quality of hydrocarbons transferred. However these become synonymous with accurate measurement.
measurements are rarely performed at the standard
conditions discussed in Fundamentals I, therefore However the terms "measurement (volume)
not only must temperature be measured, but also accuracy" and "meter accuracy" are not the same.
density, sediment and water, vapor pressure, Measurement accuracy as applied to volumes is the
pressure and viscosity must be measured. With absolute accuracy of the volume measured,
these measurements correction factors such as whereas "meter accuracy" is the accuracy of the
Volume Correction Factors (VCF) can be determined meter relative to a reference standard, such as a
to allow volumes determined at operating conditions prover.
to be expressed at standard reference conditions.
The term's repeatability and linear'~ are commonly
The means of measuring hydrocarbon liquids fall used to define meter accuracy. Repeatability is the
into one of two methods: variation in the meter's performance under constant
operating conditions, i.e. constant flow rate,
Static measurement temperature, density, etc. Where-as linearity is the
Dynamic measurement variation in the meter's performance over a range of
flow, commonly referred to as turndown ratio.
Static measurement is performed when the
hydrocarbon liquid is at rest and contained within a Therefore measurement accuracy of volumes is
container such as a tank, hence it is commonly influenced by the following factors:
referred to as Tank Gauging. On the other hand
when hydrocarbon liquids are measured while in • Meter repeatability
motion, this is referred too as dynamic measurement • Liquid density corrections due to
or Metering. varying temperature and
pressure
Another why to think of the difference in • Meter calibration or proving,
measurement techniques is to think of static including procedures
measurement as measuring the volume in a • Variations in operating
container at a point in time and dynamic conditions an their effect on a
measurement as measuring the volume in a meters performance
container over time. • Calibration of prover, i.e.
waterdraw.
This paper will examine the various types of meters,
their accessories and the devices to verify the
meter's performance.

200
METER SELECTION describe the most common types of measuring
elements are:
In general two types of meters are used in custody
transfer measurement of liquid hydrocarbons. They • Nutating disc
are positive displacement (PD) meters and turbine • Reciprocating piston
meters. The selection of a specific meter type • Sliding-vane type rotary
depends on its application. In some applications a • Bucket-type rotary
specific type is preferred, while in others either meter • Lobed rotary
type would perform satisfactorily. • Helical rotary
• Certain combinations of the above
All meter installations must meet certain
fundamental requirements. These include accurate
proving facilities; adequate protective devices, such Positive displacement meters are commonly used
as strainers, relief valves, and air or vapor for all types of liquid measurement, especially for
eliminators; and dependable pressure and flow viscous liquids such as crude oil. However, their use
controls. Additionally accurate instrumentation for in light hydrocarbon services such as propane,
measuring the physical properties of the flowing butane, etc., may result in high maintenance due to
fluid, such as temperature, density, etc. A further the lack of lubrication. They are frequently used in
fundamental is that physical condition's during product services such as loading rack applications
normal metering operations and proving of the meter even though a turbine meter may be more desirable
must be identical. The following should be for the liquid being measured. The reasons being
considered when selecting a meter and its auxiliary that loading racks are intermittent service...a service
equipment. better suited to PD meters.
• Flow range and whether flow is Positive displacement meters are generally
intermittent or continuous considered to have a 511 turndown, i.e., maximum
• Pressuro maximum operating pressure flow to minimum flow. They are typically selected to
and maximum permissible differential limit their operating range to between 40 percent
pressure and 80 percent of their stated flow range due to
• Type of liquid and its characteristics inaccuracies at the low end and rapid meter wear at
• Temperature range and accuracy of the high end. However, if proved when operating
temperature compensation above 80 percent and below 40 percent and that
• Type of volume registration device proving meets the repeatability requirements, then
• Accuracy required they can be used successfully.
• Type and method of proving required
• Applicability of auxiliary meter Positive displacement meters have an advantage
registration equipment over turbine meters in that they are not affected by
• Maintenance requirements viscosity and they require no flow conditioning, i.e.,
• Foreign matter in fluid streams straightening vanes upstream of the meter. They are
• Installation space available subject to wear which means that erosive fluid
applications require special care in filtering and
frequent meter proving to maintain their accuracy.
TYPES OF METERS Positive displacement meters are used
predominantly on LACT (Lease Automatic Custody
Positive Displacement (PD) Meter--a positive Transfer) units and on crude oil ACT (Automatic
displacement meter is a device installed in a piping Custody Transfer) units.
system in which flowing liquid is constantly and
mechanically isolated into segments of known
volumes. These segments of liquid are counted as Turbine Meter--the liquid turbine meter is an
they are displaced, and their accumulated total inferential type of volumetric measurement device
continuously and instantaneously indicated in units installed directly in the pipe of a flowing system. The
of liquid quantity by a meter register, pulse generator turbine meter housing has an internal rotor that
or flow computer. The type of mechanism employed rotates with respect to the linear velocity of the fluid
to isolate the liquid segments, i.e. by the nature of passing through the cross-sectional area of the
their measuring element, generally differentiates meter housing. As the fluid passes through the
positive displacement meters. The terms used to meter housing, the angular velocity (rpm) imparted
to the turbine rotor is proportional to the linear

201
velocity of the flowing fluid. The volumetric flow rate Turbine meters register volume utilizing a pulse
can then be determined by the rotor spccd (rpm). generator called a pick-off coil and a pre-amp. As
The volumetric flow rate (Q) is assumed to be the blade tip or paramagnetic button passes through
proportional to this measured flow velocity (V) by the magnetic field generated by the pick-off coil a
assuming a constant flow area (A). pulse is generated. This pulse is collected or
counted by an electronic register, such as a flow
Q=(V)(A) computer. Typically each turbine meter has a K-
(ft%ec) = (fUsec) (ft2) factor assigned, which is a nominal number of
pulses per unit volume. Volume is determined by
dividing the accumulated pulses by the K-factor as
The accuracy of the flow measurement can be shown:
affected if the assumption that the flow area remains
constant does not. Factors contributing to the Volume = (Pulses) + (K-factor)
changing flow area include:

• Deposits (such as paraffin) Turbine meters being velocity or inference type


• Boundary layer thickness devices require flow stream conditioning for their
• Cavitation accurate performance. The detailed requirements for
• Debris flow conditioning can be found in API MPMS
q, Operating conditions (i.e., temperature, Chapter 5, section 3--:'Measurement of Liquid
pressure, density, etc) Hydrocarbons by Turbine Meters". Typical flow
conditioning consists of upstream and downstream
Additionally the assumption that the rotor velocity is straightening sections. The upstream section usually
directly proportional to the axial velocity through the contains a tube bundle, which allows the upstream
turbine meter can be affected by the following: section to be reduced in length. This tube bundle
serves to eliminate any "swirl" in the flow stream
• Bearing friction before reaching the meter presenting a Symmetrical
• Viscous •ction velocity profile to the turbine rotor.
• Rotor blade configuration
• Flow conditioning METER ACCESSORIES
Turbine meters are used primarily for refined product
Custody transfer meter installations require proper
and light hydrocarbon measurement. Although they
accessories irrespective of meter type. Common
reportedly have a 1011 turndown range, care should
accessories include strainers or filters, pressure
be taken to select a meter size to operate within 50
gages, thermometers and thermowells, registering
percent to 100 percent of its stated flow range.
devices, temperature and pressure transmitters,
Large inaccuracies may result when operating below
pulse generating devices, pulsation dampening
the 50 percent flow level. equipment, prover connections, air eliminators,
pressure and flow controllers, and straightening
Turbine meters are generally used in trash-free
vanes. A discussion of some of these devices is as
flowing applications since any possibility of build-up
follows:
on the rotor will impede its rpm---hence measuring
ability. Early installations of turbine meters suffered 1. Strainersmboth PD and Turbine meters
high failures due to frequent rotor bearing failures.
require this equipment upstream of the
The use of tungsten carbide bearings has greatly
meter to protect the internals. Foreign
reduced the failures and inaccuracies due to bearing
objects can cause meter damage,
friction.
excessive wear, or inaccurate
measurement. Housing need to be
When using turbine meters in light hydrocarbon
compatible with applicable piping
liquid mixtures such as ethane-propane, NGL's care
specifications and pressure ratings.
should be exercised, in that the linear operating 2. Pressure gages or transmitters--an
range is raised, and the meter should be selected to
accurate pressure determination is
operate over the 80 percent to 125 percent range of
required since the measured liquid is
the stated flow range. The lighter fluid impinging on always metered at a pressure higher
the rotor blades has less force at lower capacities
than atmospheric and is in a
causing erratic rotation and accompanying
compressed form. A compressibility
inaccuracies.

202
factor must be established and applied meter at maximum flowing conditions
to the metered volume. plus 1.25 times the equilibrium vapor
3. Temperature transmitters and pressure at flowing temperature.
devices--liquid is generally measured . Straightening vanes--turbine meters
at some temperature other than base preferably should have straightening
60°F, in order to correct the volume to a vanes upstream of the meter. The vanes
net 60°F volume, a volume correction condition the liquid flow profile for the
must be made. This can be done turbine meter. If straightening vanes are
manually, in a computer (flow) or directly not u s e d , sufficient straight run
by the meter. An additional thermowell upstream and downstream piping must
is normally installed adjacent to the be provided to condition the flowing
transmitter to allow checks to be made stream.
periodically.
4. Pulse generating device c= pulsing
device connected directly to the meter is METER PERFORMANCE
required for transmitting the metered
volume and for proving the accuracy of The meter's operating performance should be
the meter. The pulse rate varies with verified by proving it periodically. The frequency of
meter size and type and typically may proving is guided by the type of operations, i.e.
be 1000 or 8400 pulses/Bbl for PD batched systems, regulatory requirements,
meters and 500, 1000 pulses/Bbl for agreements between parties and operational
turbine meters. experience. Therefore the prover is an integral part
5. Registering devices--a local register of any metering system. For the meter to register
should be provided to show gross accurately its registration needs to be compared to a
readings. This device is typically either a device with a known volume. All provers' function in
mechanical or electronic totalizer. the same manner, they measure the volume of fluid
Increasingly Measurement Row passing through the meter. The two measurements,
Computers are utilized to accumulate the prover and meter, are compared in order to
not only the pulses but also the determine a meter factor as follows:
temperature and pressure variables and
in some systems the flowing real time Meter Factor = Prover Volume / Meter Volume
density.
6. Pulsation dampers---meters located in This meter factor is applied to the meter's
line with positive displacement pumps registration to correct it.
should use a dampening device to
eliminate the pulsating effects. The The types of provers include the following:
typical device usually consists of an
appendage to the pipeline, which has an Calibrated Prover Tank
inflatable bag inside and is precharged Pipe Prover
with nitrogen. • Bi-directional
7. Air eliminators--air or vapor • Uni-directional
eliminators should be installed • Reduced volume
preceding the meter since large • Small volume
compressed volumes will tend to
expand through the meter causing Master meters are utilized from time to time in
overspeed and serious damage. An air specific applications.
eliminator is usually an enlarged section
of pipe or vessel that is vented and Meters that are used to measure liquids of different
provided with a level control device. densities, viscosity's or other characteristics, which
8. Back pressure control--some may affect meter slippage or volume corrections,
installations require a method of should have meter factors developed for each type
controlling backpressure on the meter of fluid. Additionally, meters subjected to varying
sufficiently high to prevent vaporization rates of flow should be proved at a sufficient number
across the meter. API MPMS Chapter 5, of points to allow preparation of a performance curve
section 3 recommends that the and an appropriate meter factor selected from this
backpressure be equal to or greater curve. For all practical purposes, a change of 15
than twice the pressure drop across the

203
percent from the base rate is considered to be a sampling system to capture the sample for
sufficient change to warrant multiple proving's. analysis.
Although there are a number of prover types and
proving techniques vary in complexity they all Regardless of which test method is used for analysis
function in the same basic manner. The basic one must first start with a sample of the hydrocarbon
concept of all pipe provers is the same; accumulate liquid to be analyzed. Samples can be obtained
the registration of the meter during a time when the through two principle means:
displacer through a pipe of known volume. This
known volume is called the calibrated section and is
precisely defined by the detector switches. As the • Manual Sampling
displacer passes the detector switch, it activates the
prover counter to commence collecting the pulses • Automatic Sampling
generated by the meter and when the displacer
activates the second detector, the counter stops its
accumulation of pulses. Adjusting the total pulses by Manual sampling procedures and equipment are
the meters K-factor yields the meter's indicated addressed in ASTM D 4057 (API Ch 8.1) =Manual
volume, which can then be compared to the provers Sampling of Petroleum and Petroleum Products".
known volume. This ratio yields the meter factor for Automatic sampling is covered in ASTM D 4177 (AP!
the given meter. Ch 8.2) "Automatic Sampling of Petroleum and
Petroleum Products". The successful and accurate
Prover Calibration---the known volume of the analysis of any sample depends on the appropriate
provers calibrated section must be precisely handling and mixing of that sample from point of
determined. This calibration process for pipe provers extraction to its placement into the analytical
i s known as a waterdraw calibration. Tank provers apparatus. These procedures are covered in ASTM
are calibrated utilizing a similar process known as D 5854 (API Ch 8.3) "Practice for Mixing and
waterfill method. Using precisely calibrated Test Handling of Liquid Samples of Petroleum and
Measures the prover is calibrated in a manner Petroleum Products".
similar to proving a meter. As water is circulated
through the prover, it moves the displacer. When the While both sampling methods can be used with
displacer activates the first detector, a solenoid metering systems, depending on the quality
valve is typically activated and the fluid is displaced parameter, most metering systems use the
from the prover into the Test Measure, stopping Automatic Sampling method. From the sample
when the second detector is activated. The process receiver multiple samples can be extracted for the
is repeated until a repeatability criterion is met. The vadous analytical tests.
volume recorded from these calibration runs is
corrected to base conditions and then related to the
volume of the calibrated section. This volume at
base conditions is called the Prover Base Volume, SUMMARY
and becomes the basis for the meter proving
process. The process of calibration, the '~Vaterdraw" A dynamic metering system allows a degree of
is covered in the API MPMS standard Chapter 4, flexibility, that static measurement does not, in
section 7 "Field Standard Test Measures" and a new operations, while meeting the exacting requirements
standard API MPMS Chapter 4, section 9 "Prover for Custody Transfer Measurement. A typical system
Calibration" is composed of multiple individual components that
interact to provide the information, conditioning or
control, which results in accurate measurement.
SAMPLING
The quality of those individual components, their
While some of the physical properties needed to installation, maintenance and calibration has a direct
complete the measurement transaction, such as bearing on accurate measurement. As we move
temperature and pressure are measured using from the mechanical systems to the electronic
electronic transmitters. A means needs to be utilized environment and increased automation, these
to determine the density of the transaction. This can requirements will not decrease, but substantially
be accomplished on-line using a vibrating element increase. The increasing use of Vibrating Element
densitometer, as is done typically in refined product, densitometer's for real-time density determination,
light hydrocarbon liquids and NGL's or off-line using especially for hydrocarbon stream with varying
compositions, requires increased verification and

204
proving of the primary device. The miniaturization of
electronics has resulted in Measurement Flow
Computers with the robustness and processing
capabilities, which used to reside on a mainframe
computer. However, this requires that the flow
computer be "calibrated" through the use of audit
trails for both input data and calculation processes.

As we utilize new metering technology such as


helical turbine meters and integrate it into existing
systems we must search for measures for
verification of its performance. Similar
verification/proving concerns exist as the industry
explores the use of other metering technology such
as Coriolis Meters and Liquid Ultrasonic Meters.
Both of these meters are inferential meters similar to
turbine meters. However, they require signal
processing integrated over time in order to generate
a pulse output signal from their electronics
packages. The Coriolis meter measures the
deflection force generated by the fluid moving
through the sensors. The Ultrasonic meter measures
the time it takes for a pulse to travel from one sensor
to another. The signals of both are processed and a
volume is imputed from the signal.

The key to accurate measurement in a dynamic


mode is the requirement that all measurements must
be verified against a "master" device, which is
traceable to NIST (National Institute for Standards
and Technology). This verification must be
performed on an ongoing basis.

205

Vous aimerez peut-être aussi