Alarms
162 Wrong exchange data between the 2 CPUs Contact Technical Assistance
8 Pressure Transducer - Left Piston Side - CHA - • Verify that the wiring is not open circuited.
Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
9 Pressure Transducer - Left Piston Side - CHA - • Verify that the wiring is not short circuited
High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
10 Pressure Transducer - Right Piston Side -CHA - • Verify that the wiring is not open circuited.
Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
11 Pressure Transducer - Right Piston Side - CHA - • Verify that the wiring is not short circuited
High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
000303-MMR17X.uso.EN.00-TRI I 31 I 5 | Alarms
User manual | MIDMAC RS17-XX English
18 Pressure Transducer - Left Rod Side – CHA - Low • Verify that the wiring is not open circuited.
Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
19 Pressure Transducer - Left Rod Side - CHA - High • Verify that the wiring is not short circuited
Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
20 Pressure Transducer - Right Rod Side - CHA - • Verify that the wiring is not open circuited.
Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
21 Pressure Transducer - Right Rod Side - CHA - • Verify that the wiring is not short circuited
High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
165 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Piston Side
166 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Rode Side
170 Transducer -Left Piston Side - CHB - Low Signal • Verify that the wiring is not open circuited.
Signal from Transducer pressure sensor lower than • Verify the pressure transducer integrity.
the minimum value If the alarm persists, contact Technical Assistance
171 Transducer -Left Piston Side - CHB - High Signal • Verify that the wiring is not short circuited
Signal from Transducer pressure sensor higher than • Verify the pressure transducer integrity
the maximum value If the alarm persists, contact Technical Assistance
172 Transducer -Right Piston Side -CHB - Low Signal • Verify that the wiring is not open circuited.
Signal from Transducer pressure sensor lower than • Verify the pressure transducer integrity.
the minimum value If the alarm persists, contact Technical Assistance
000303-MMR17X.uso.EN.00-TRI I 32 I 5 | Alarms
User manual | MIDMAC RS17-XX English
173 Transducer- Right Piston Side -CHB - High Signal • Verify that the wiring is not short circuited
Signal from Transducer pressure sensor higher than • Verify the pressure transducer integrity
the maximum value If the alarm persists, contact Technical Assistance
174 Transducer -Left Rod Side - CHB - Low Signal • Verify that the wiring is not open circuited.
Signal from Transducer pressure sensor lower than • Verify the pressure transducer integrity.
the minimum value If the alarm persists, contact Technical Assistance
175 Transducer -Left Rod Side - CHB - High Signal • Verify that the wiring is not short circuited
Signal from Transducer pressure sensor higher than • Verify the pressure transducer integrity
the maximum value If the alarm persists, contact Technical Assistance
176 Transducer - Right Rod Side - CHB - Low Signal • Verify that the wiring is not open circuited.
Signal from Transducer pressure sensor lower than • Verify the pressure transducer integrity.
the minimum value If the alarm persists, contact Technical Assistance
177 Transducer - Right Rod Side - CHB - High Signal • Verify that the wiring is not short circuited
Signal from Transducer pressure sensor higher than • Verify the pressure transducer integrity
the maximum value If the alarm persists, contact Technical Assistance
178 Redundance Error - Pressure transducer Right • Verify the relative pressure values
Piston Side
179 Redundance Error - Pressure transducer Right • Verify the relative pressure values
Rode Side
180 Angle sensor - CHB - MIN • Verify the angle sensor integrity
Angle sensor reading (channel B) lower than the •
minimum value If the alarm persists, contact Technical Assistance
000303-MMR17X.uso.EN.00-TRI I 33 I 5 | Alarms
User manual | MIDMAC RS17-XX English
181 Angle sensor - CHB - MAX • Verify the angle sensor integrity
Angle sensor raeding (channel B) higher than the •
maximum value If the alarm persists, contact Technical Assistance
185 Redundance Error - Angle value • Verify the relative angle values
190 Length sensor - CHA - Low Signal • Verify the length sensor integrity
Analogue signal lower than the minimum value •
admitted If the alarm persists, contact Technical Assistance
191 Length sensor - CHA - High Signal • Verify the length sensor integrity
Analogue signal higher than the maximum value •
admitted If the alarm persists, contact Technical Assistance
192 Length sensor - CHB - Low Signal • Verify the length sensor integrity
Analogue signal lower than the minimum value If the alarm persists, contact Technical Assistance
admitted
193 Length sensor - CHB - High Signal • Verify the length sensor integrity
Analogue signal higher than the maximum value If the alarm persists, contact Technical Assistance
admitted
194 Length sensor – CHB – MIN • Verify the length sensor integrity
Length sensor reading (channel B) lower than the If the alarm persists, contact Technical Assistance
minimum
000303-MMR17X.uso.EN.00-TRI I 34 I 5 | Alarms
User manual | MIDMAC RS17-XX English
195 Length sensor – CHB – MAX • Verify the length sensor integrity
Length sensor reading (channel B) higher than the If the alarm persists, contact Technical Assistance
maximum
196 Redundance Error - Length value • Verify the relative length values
24 Difference Press. Piston Side • Verify the connection of the two transducers
Too much difference of pressure between the two low pressure in the low chambers.
chambers • Verify the pressure transducer integrity
If the alarm persists, contact Technical Assistance
25 Difference Press. Rod Side • Verify the connection of the two transducers
Too much difference of pressure between the two pressure in the low chambers.
high chambers • Verify the pressure transducer integrity
If the alarm persists, contact Technical Assistance
000303-MMR17X.uso.EN.00-TRI I 35 I 5 | Alarms
User manual | MIDMAC RS17-XX English
121 ACQ Fault Error - CHA • Verify the cable reel integrity
Fault from ACQ cable reel – channel A • If the alarm persists, contact Technical
Assistance
123 ACQ Fault Error – CHB • Verify the cable reel integrity
Fault from ACQ cable reel – channel B • If the alarm persists, contact Technical
Assistance
200 Pressure Transducer -Left Piston Side - CHA - • Verify that the wiring is not open circuited.
Low Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
000303-MMR17X.uso.EN.00-TRI I 36 I 5 | Alarms
User manual | MIDMAC RS17-XX English
201 Pressure Transducer -Left Piston Side - CHA - • Verify that the wiring is not short circuited
High Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
202 Pressure Transducer -Right Piston Side -CHA - • Verify that the wiring is not open circuited.
Low Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
203 Pressure Transducer- Right Piston Side -CHA - • Verify that the wiring is not short circuited
High Signa from CPU1l • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
204 Pressure Transducer -Left Rod Side - CHA - Low • Verify that the wiring is not open circuited.
Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
205 Pressure Transducer -Left Rod Side - CHA - High • Verify that the wiring is not short circuited
Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
206 Pressure Transducer - Right Rod Side - CHA - • Verify that the wiring is not open circuited.
Low Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
207 Pressure Transducer - Right Rod Side - CHA - • Verify that the wiring is not short circuited
High Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
208 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Piston Side from CPU1
000303-MMR17X.uso.EN.00-TRI I 37 I 5 | Alarms
User manual | MIDMAC RS17-XX English
209 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Rode Side from CPU1
210 Pressure Transducer -Left Piston Side - CHB - • Verify that the wiring is not open circuited.
Low Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
211 Pressure Transducer -Left Piston Side - CHB - • Verify that the wiring is not short circuited
High Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
212 Transducer -Right Piston Side -CHB - Low Signal • Verify that the wiring is not open circuited.
from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
213 Pressure Transducer- Right Piston Side -CHB - • Verify that the wiring is not short circuited
High Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
214 Pressure Transducer -Left Rod Side - CHB - Low • Verify that the wiring is not open circuited.
Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
215 Pressure Transducer -Left Rod Side - CHB - High • Verify that the wiring is not short circuited
Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
216 Pressure Transducer - Right Rod Side - CHB - • Verify that the wiring is not open circuited.
Low Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
000303-MMR17X.uso.EN.00-TRI I 38 I 5 | Alarms
User manual | MIDMAC RS17-XX English
217 Pressure Transducer - Right Rod Side - CHB - • Verify that the wiring is not short circuited
High Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
218 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Piston Side from CPU1
219 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Rode Side from CPU1
220 Angle sensor - CHA - MIN from CPU1 • Verify the angle sensor integrity
Signal from Angle sensor lower than the minimum If the alarm persists, contact Technical Assistance
value
221 Angle sensor - CHA - MAX from CPU1 • Verify the angle sensor integrity
Signal from Angle sensor higher than the maximum If the alarm persists, contact Technical Assistance
value
222 Angle sensor - CHB - MIN from CPU1 • Verify the angle sensor integrity
Signal from Angle sensor lower than the minimum If the alarm persists, contact Technical Assistance
value
223 Angle sensor - CHB - MAX from CPU1 • Verify the angle sensor integrity
Signal from Angle sensor higher than the maximum If the alarm persists, contact Technical Assistance
value
224 Redundance Error - Angle value from CPU1 • Check the wiring connection
If the alarm persists, contact Technical Assistance
230 Length sensor - CHA - Low Signal from CPU1 • Verify the length sensor integrity
• If the alarm persists, contact Technical
Assistance
231 Length sensor - CHA - High Signal from CPU1 • Verify the length sensor integrity
If the alarm persists, contact Technical Assistance
000303-MMR17X.uso.EN.00-TRI I 39 I 5 | Alarms
User manual | MIDMAC RS17-XX English
232 Length sensor – CHA – MIN from CPU1 • Verify the length sensor integrity
Signal from Length sensor lower than the minimum If the alarm persists, contact Technical Assistance
value
233 Length sensor – CHA – MAX from CPU1 • Verify the length sensor integrity
Signal from Length sensor higher than the maximum If the alarm persists, contact Technical Assistance
value
235 Length sensor - CHB - High Signal from CPU1 • Verify the length sensor integrity
If the alarm persists, contact Technical Assistance
236 Length sensor – CHB – MIN from CPU1 • Verify the length sensor integrity
Signal from Length sensor lower than the minimum If the alarm persists, contact Technical Assistance
value
237 Length sensor – CHB – MAX from CPU1 • Verify the length sensor integrity
Signal from Length sensor higher than the maximum If the alarm persists, contact Technical Assistance
value
238 Redundance Error - Length value from CPU1 • Verify the relative length values
240 ACQ Fault Error – CHA from CPU1 • Verify the cable reel integrity
Fault from ACQ cable reel – channel A If the alarm persists, contact Technical Assistance
241 ACQ Fault Error – CHB from CPU1 • Verify the cable reel integrity
Fault from ACQ cable reel – channel B If the alarm persists, contact Technical Assistance
50 Pressure Transducer - Right Piston Side • Verify that the wiring is not open circuited.
Stabilizer - CHA - Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
000303-MMR17X.uso.EN.00-TRI I 40 I 5 | Alarms
User manual | MIDMAC RS17-XX English
51 Pressure Transducer - Right Piston Side • Verify that the wiring is not short circuited
Stabilizer - CHA - High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
52 Pressure Transducer - Left Piston Side Stabilizer • Verify that the wiring is not open circuited.
- CHA - Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
53 Pressure Transducer - Left Piston Side Stabilizer • Verify that the wiring is not short circuited
- CHA - High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
54 Pressure Transducer - Rod Side Stabilizer - CHA - • Verify that the wiring is not open circuited.
Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
55 Pressure Transducer - Rod Side Stabilizer - CHA - • Verify that the wiring is not short circuited
High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
320 Pressure Transducer - Right Piston Side • Verify that the wiring is not open circuited.
Stabilizer - CHB - Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
321 Pressure Transducer - Right Piston Side • Verify that the wiring is not short circuited
Stabilizer - CHB - High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
000303-MMR17X.uso.EN.00-TRI I 41 I 5 | Alarms
User manual | MIDMAC RS17-XX English
322 Pressure Transducer - Left Piston Side Stabilizer • Verify that the wiring is not open circuited.
– CHB - Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
323 Pressure Transducer - Left Piston Side Stabilizer • Verify that the wiring is not short circuited
- CHB - High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
324 Pressure Transducer - Rod Side Stabilizer - CHB - • Verify that the wiring is not open circuited.
Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value
325 Pressure Transducer - Rod Side Stabilizer - CHB - • Verify that the wiring is not short circuited
High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value
56 Pressure Transducer - Steering - Low Signal • Verify that the wiring is not open circuited.
Signal from Transducer pressure sensor lower than • Verify the pressure transducer integrity.
the minimum value If the alarm persists, contact Technical Assistance
57 Pressure Transducer - Steering - High Signal • Verify that the wiring is not short circuited
Signal from Transducer pressure sensor higher than • Verify the pressure transducer integrity
the maximum value If the alarm persists, contact Technical Assistance
330 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Piston Side Stabilizer
331 Redundance Error - Pressure transducer Right • Verify the relative pressure values
Piston Side Stabilizer
333 Redundance Error - Pressure transducer Rod • Verify the relative pressure values
Side Stabilizer
000303-MMR17X.uso.EN.00-TRI I 42 I 5 | Alarms
User manual | MIDMAC RS17-XX English
334 Difference pressure between right piston side • Verify the relative pressure values
and left piston side stabilizer
335 Difference pressure between right piston side • Verify the relative pressure values
and rod side stabilizer
336 Difference pressure between left piston side and • Verify the relative pressure values
rod side stabilizer
310 CanBus Timeout from slave (stabilizer) MIDAC+ • Verify wiring connection
351 RDN error activate stabilizer system [valve 1+7] Verify wiring connection
Redundance Error - Output of output MIDAC SLAVE
PIN 4
000303-MMR17X.uso.EN.00-TRI I 43 I 5 | Alarms
User manual | MIDMAC RS17-XX English
354 RDN error Fix stabilizer position [valve 8] Verify wiring connection
Redundance Error - Output of output MIDAC SLAVE
PIN 8
355 RDN error Operator presence signal to APC 200 Verify wiring connection
Redundance Error - Output of output MIDAC SLAVE
PIN 9
356 RDN error Set pressure [valve 11] Redundance Verify wiring connection
Error - Output of output MIDAC SLAVE PIN 5
000303-MMR17X.uso.EN.00-TRI I 44 I 5 | Alarms