Alarms are displayed in the Alarm Area of each screen.

5.1 ALARM LIST

Code Description What to do

1 Memory error in parameters area Contact Technical Assistance

160 Memory error in Internal Flash Contact Technical Assistance

161 Memory error in External Flash Contact Technical Assistance

162 Wrong exchange data between the 2 CPUs Contact Technical Assistance

8 Pressure Transducer - Left Piston Side - CHA - • Verify that the wiring is not open circuited.
Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

9 Pressure Transducer - Left Piston Side - CHA - • Verify that the wiring is not short circuited
High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

10 Pressure Transducer - Right Piston Side -CHA - • Verify that the wiring is not open circuited.
Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

11 Pressure Transducer - Right Piston Side - CHA - • Verify that the wiring is not short circuited
High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

000303-MMR17X.uso.EN.00-TRI I 31 I 5 | Alarms
User manual | MIDMAC RS17-XX English

Code Description What to do

18 Pressure Transducer - Left Rod Side – CHA - Low • Verify that the wiring is not open circuited.
Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

19 Pressure Transducer - Left Rod Side - CHA - High • Verify that the wiring is not short circuited
Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

20 Pressure Transducer - Right Rod Side - CHA - • Verify that the wiring is not open circuited.
Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

21 Pressure Transducer - Right Rod Side - CHA - • Verify that the wiring is not short circuited
High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

165 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Piston Side

166 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Rode Side

170 Transducer -Left Piston Side - CHB - Low Signal • Verify that the wiring is not open circuited.
Signal from Transducer pressure sensor lower than • Verify the pressure transducer integrity.
the minimum value If the alarm persists, contact Technical Assistance

171 Transducer -Left Piston Side - CHB - High Signal • Verify that the wiring is not short circuited
Signal from Transducer pressure sensor higher than • Verify the pressure transducer integrity
the maximum value If the alarm persists, contact Technical Assistance

172 Transducer -Right Piston Side -CHB - Low Signal • Verify that the wiring is not open circuited.
Signal from Transducer pressure sensor lower than • Verify the pressure transducer integrity.
the minimum value If the alarm persists, contact Technical Assistance

000303-MMR17X.uso.EN.00-TRI I 32 I 5 | Alarms
User manual | MIDMAC RS17-XX English

Code Description What to do

173 Transducer- Right Piston Side -CHB - High Signal • Verify that the wiring is not short circuited
Signal from Transducer pressure sensor higher than • Verify the pressure transducer integrity
the maximum value If the alarm persists, contact Technical Assistance

174 Transducer -Left Rod Side - CHB - Low Signal • Verify that the wiring is not open circuited.
Signal from Transducer pressure sensor lower than • Verify the pressure transducer integrity.
the minimum value If the alarm persists, contact Technical Assistance

175 Transducer -Left Rod Side - CHB - High Signal • Verify that the wiring is not short circuited
Signal from Transducer pressure sensor higher than • Verify the pressure transducer integrity
the maximum value If the alarm persists, contact Technical Assistance

176 Transducer - Right Rod Side - CHB - Low Signal • Verify that the wiring is not open circuited.
Signal from Transducer pressure sensor lower than • Verify the pressure transducer integrity.
the minimum value If the alarm persists, contact Technical Assistance

177 Transducer - Right Rod Side - CHB - High Signal • Verify that the wiring is not short circuited
Signal from Transducer pressure sensor higher than • Verify the pressure transducer integrity
the maximum value If the alarm persists, contact Technical Assistance

178 Redundance Error - Pressure transducer Right • Verify the relative pressure values
Piston Side

179 Redundance Error - Pressure transducer Right • Verify the relative pressure values
Rode Side

4 Angle sensor - CHA - MIN • Verify the angle sensor integrity


Angle sensor reading (channel A) lower than the •
minimum value If the alarm persists, contact Technical Assistance

5 Angle sensor - CHA - MAX • Verify the angle sensor integrity


Angle sensor reading (channel A) higher than the •
maximum value If the alarm persists, contact Technical Assistance

180 Angle sensor - CHB - MIN • Verify the angle sensor integrity
Angle sensor reading (channel B) lower than the •
minimum value If the alarm persists, contact Technical Assistance

000303-MMR17X.uso.EN.00-TRI I 33 I 5 | Alarms
User manual | MIDMAC RS17-XX English

Code Description What to do

181 Angle sensor - CHB - MAX • Verify the angle sensor integrity
Angle sensor raeding (channel B) higher than the •
maximum value If the alarm persists, contact Technical Assistance

185 Redundance Error - Angle value • Verify the relative angle values

190 Length sensor - CHA - Low Signal • Verify the length sensor integrity
Analogue signal lower than the minimum value •
admitted If the alarm persists, contact Technical Assistance

191 Length sensor - CHA - High Signal • Verify the length sensor integrity
Analogue signal higher than the maximum value •
admitted If the alarm persists, contact Technical Assistance

36 Length sensor – CHA – MIN • Verify the length sensor integrity


Length sensor reading (channel A) lower than the If the alarm persists, contact Technical Assistance
minimum

37 Length sensor – CHA – MAX • Verify the length sensor integrity


Length sensor reading (channel A) higher than the If the alarm persists, contact Technical Assistance
maximum

192 Length sensor - CHB - Low Signal • Verify the length sensor integrity
Analogue signal lower than the minimum value If the alarm persists, contact Technical Assistance
admitted

193 Length sensor - CHB - High Signal • Verify the length sensor integrity
Analogue signal higher than the maximum value If the alarm persists, contact Technical Assistance
admitted

194 Length sensor – CHB – MIN • Verify the length sensor integrity
Length sensor reading (channel B) lower than the If the alarm persists, contact Technical Assistance
minimum

000303-MMR17X.uso.EN.00-TRI I 34 I 5 | Alarms
User manual | MIDMAC RS17-XX English

Code Description What to do

195 Length sensor – CHB – MAX • Verify the length sensor integrity
Length sensor reading (channel B) higher than the If the alarm persists, contact Technical Assistance
maximum

196 Redundance Error - Length value • Verify the relative length values

24 Difference Press. Piston Side • Verify the connection of the two transducers
Too much difference of pressure between the two low pressure in the low chambers.
chambers • Verify the pressure transducer integrity
If the alarm persists, contact Technical Assistance

25 Difference Press. Rod Side • Verify the connection of the two transducers
Too much difference of pressure between the two pressure in the low chambers.
high chambers • Verify the pressure transducer integrity
If the alarm persists, contact Technical Assistance

31 Error configuration • Verify the machine setting


No machine selected • Save configuration and restart
If the alarm persists, contact Technical Assistance

30 Error configuration • Verify the machine setting


More than 1 machine selected • Save configuration and restart
If the alarm persists, contact Technical Assistance

62 Derrick Joystick fault • Verify joystick integrity


Value of the joystick out of range

64 Joystick CANBus Timeout • Verify wiring connection


Canbus timeout from joystick

63 Tele Joystick faul • Verify joystick integrity


Value of the joystick out of range

198 CANBus timeout from HCU controller • Verify wiring connection

000303-MMR17X.uso.EN.00-TRI I 35 I 5 | Alarms
User manual | MIDMAC RS17-XX English

Code Description What to do

120 ACQ CANBus Timeout - CHA


• Verify wiring connection
CANBus timeout from ACQ cable reel – channel A

121 ACQ Fault Error - CHA • Verify the cable reel integrity
Fault from ACQ cable reel – channel A • If the alarm persists, contact Technical
Assistance

122 ACQ CANBus Timeout - CHB • Verify wiring connection

123 ACQ Fault Error – CHB • Verify the cable reel integrity
Fault from ACQ cable reel – channel B • If the alarm persists, contact Technical
Assistance

150 RDN Boom UP


Redundance Error - Output of boom up not working • Verify wiring connection
correctly

151 RDN Boom DOWN


Redundance Error - Output of boom down not • Verify wiring connection
working correctly

152 RDN Tele OUT


Redundance Error - Output of tele out not working • Verify wiring connection
correctly

153 RDN Tele IN


Redundance Error - Output of tele in not working • Verify wiring connection
correctly

200 Pressure Transducer -Left Piston Side - CHA - • Verify that the wiring is not open circuited.
Low Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

000303-MMR17X.uso.EN.00-TRI I 36 I 5 | Alarms
User manual | MIDMAC RS17-XX English

Code Description What to do

201 Pressure Transducer -Left Piston Side - CHA - • Verify that the wiring is not short circuited
High Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

202 Pressure Transducer -Right Piston Side -CHA - • Verify that the wiring is not open circuited.
Low Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

203 Pressure Transducer- Right Piston Side -CHA - • Verify that the wiring is not short circuited
High Signa from CPU1l • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

204 Pressure Transducer -Left Rod Side - CHA - Low • Verify that the wiring is not open circuited.
Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

205 Pressure Transducer -Left Rod Side - CHA - High • Verify that the wiring is not short circuited
Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

206 Pressure Transducer - Right Rod Side - CHA - • Verify that the wiring is not open circuited.
Low Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

207 Pressure Transducer - Right Rod Side - CHA - • Verify that the wiring is not short circuited
High Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

208 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Piston Side from CPU1

000303-MMR17X.uso.EN.00-TRI I 37 I 5 | Alarms
User manual | MIDMAC RS17-XX English

Code Description What to do

209 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Rode Side from CPU1

210 Pressure Transducer -Left Piston Side - CHB - • Verify that the wiring is not open circuited.
Low Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

211 Pressure Transducer -Left Piston Side - CHB - • Verify that the wiring is not short circuited
High Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

212 Transducer -Right Piston Side -CHB - Low Signal • Verify that the wiring is not open circuited.
from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

213 Pressure Transducer- Right Piston Side -CHB - • Verify that the wiring is not short circuited
High Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

214 Pressure Transducer -Left Rod Side - CHB - Low • Verify that the wiring is not open circuited.
Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

215 Pressure Transducer -Left Rod Side - CHB - High • Verify that the wiring is not short circuited
Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

216 Pressure Transducer - Right Rod Side - CHB - • Verify that the wiring is not open circuited.
Low Signal from CPU1 • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

000303-MMR17X.uso.EN.00-TRI I 38 I 5 | Alarms
User manual | MIDMAC RS17-XX English

Code Description What to do

217 Pressure Transducer - Right Rod Side - CHB - • Verify that the wiring is not short circuited
High Signal from CPU1 • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

218 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Piston Side from CPU1

219 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Rode Side from CPU1

220 Angle sensor - CHA - MIN from CPU1 • Verify the angle sensor integrity
Signal from Angle sensor lower than the minimum If the alarm persists, contact Technical Assistance
value

221 Angle sensor - CHA - MAX from CPU1 • Verify the angle sensor integrity
Signal from Angle sensor higher than the maximum If the alarm persists, contact Technical Assistance
value

222 Angle sensor - CHB - MIN from CPU1 • Verify the angle sensor integrity
Signal from Angle sensor lower than the minimum If the alarm persists, contact Technical Assistance
value

223 Angle sensor - CHB - MAX from CPU1 • Verify the angle sensor integrity
Signal from Angle sensor higher than the maximum If the alarm persists, contact Technical Assistance
value

224 Redundance Error - Angle value from CPU1 • Check the wiring connection
If the alarm persists, contact Technical Assistance

230 Length sensor - CHA - Low Signal from CPU1 • Verify the length sensor integrity
• If the alarm persists, contact Technical
Assistance

231 Length sensor - CHA - High Signal from CPU1 • Verify the length sensor integrity
If the alarm persists, contact Technical Assistance

000303-MMR17X.uso.EN.00-TRI I 39 I 5 | Alarms
User manual | MIDMAC RS17-XX English

Code Description What to do

232 Length sensor – CHA – MIN from CPU1 • Verify the length sensor integrity
Signal from Length sensor lower than the minimum If the alarm persists, contact Technical Assistance
value

233 Length sensor – CHA – MAX from CPU1 • Verify the length sensor integrity
Signal from Length sensor higher than the maximum If the alarm persists, contact Technical Assistance
value

234 Length sensor - CHB - Low Signal from CPU1


• Verify the length sensor integrity
If the alarm persists, contact Technical Assistance

235 Length sensor - CHB - High Signal from CPU1 • Verify the length sensor integrity
If the alarm persists, contact Technical Assistance

236 Length sensor – CHB – MIN from CPU1 • Verify the length sensor integrity
Signal from Length sensor lower than the minimum If the alarm persists, contact Technical Assistance
value

237 Length sensor – CHB – MAX from CPU1 • Verify the length sensor integrity
Signal from Length sensor higher than the maximum If the alarm persists, contact Technical Assistance
value

238 Redundance Error - Length value from CPU1 • Verify the relative length values

240 ACQ Fault Error – CHA from CPU1 • Verify the cable reel integrity
Fault from ACQ cable reel – channel A If the alarm persists, contact Technical Assistance

241 ACQ Fault Error – CHB from CPU1 • Verify the cable reel integrity
Fault from ACQ cable reel – channel B If the alarm persists, contact Technical Assistance

50 Pressure Transducer - Right Piston Side • Verify that the wiring is not open circuited.
Stabilizer - CHA - Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

000303-MMR17X.uso.EN.00-TRI I 40 I 5 | Alarms
User manual | MIDMAC RS17-XX English

Code Description What to do

51 Pressure Transducer - Right Piston Side • Verify that the wiring is not short circuited
Stabilizer - CHA - High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

52 Pressure Transducer - Left Piston Side Stabilizer • Verify that the wiring is not open circuited.
- CHA - Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

53 Pressure Transducer - Left Piston Side Stabilizer • Verify that the wiring is not short circuited
- CHA - High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

54 Pressure Transducer - Rod Side Stabilizer - CHA - • Verify that the wiring is not open circuited.
Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

55 Pressure Transducer - Rod Side Stabilizer - CHA - • Verify that the wiring is not short circuited
High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

320 Pressure Transducer - Right Piston Side • Verify that the wiring is not open circuited.
Stabilizer - CHB - Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

321 Pressure Transducer - Right Piston Side • Verify that the wiring is not short circuited
Stabilizer - CHB - High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

000303-MMR17X.uso.EN.00-TRI I 41 I 5 | Alarms
User manual | MIDMAC RS17-XX English

Code Description What to do

322 Pressure Transducer - Left Piston Side Stabilizer • Verify that the wiring is not open circuited.
– CHB - Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

323 Pressure Transducer - Left Piston Side Stabilizer • Verify that the wiring is not short circuited
- CHB - High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

324 Pressure Transducer - Rod Side Stabilizer - CHB - • Verify that the wiring is not open circuited.
Low Signal • Verify the pressure transducer integrity.
Signal from Transducer pressure sensor lower than If the alarm persists, contact Technical Assistance
the minimum value

325 Pressure Transducer - Rod Side Stabilizer - CHB - • Verify that the wiring is not short circuited
High Signal • Verify the pressure transducer integrity
Signal from Transducer pressure sensor higher than If the alarm persists, contact Technical Assistance
the maximum value

56 Pressure Transducer - Steering - Low Signal • Verify that the wiring is not open circuited.
Signal from Transducer pressure sensor lower than • Verify the pressure transducer integrity.
the minimum value If the alarm persists, contact Technical Assistance

57 Pressure Transducer - Steering - High Signal • Verify that the wiring is not short circuited
Signal from Transducer pressure sensor higher than • Verify the pressure transducer integrity
the maximum value If the alarm persists, contact Technical Assistance

330 Redundance Error - Pressure transducer Left • Verify the relative pressure values
Piston Side Stabilizer

331 Redundance Error - Pressure transducer Right • Verify the relative pressure values
Piston Side Stabilizer

333 Redundance Error - Pressure transducer Rod • Verify the relative pressure values
Side Stabilizer

000303-MMR17X.uso.EN.00-TRI I 42 I 5 | Alarms
User manual | MIDMAC RS17-XX English

Code Description What to do

334 Difference pressure between right piston side • Verify the relative pressure values
and left piston side stabilizer

335 Difference pressure between right piston side • Verify the relative pressure values
and rod side stabilizer

336 Difference pressure between left piston side and • Verify the relative pressure values
rod side stabilizer

310 CanBus Timeout from slave (stabilizer) MIDAC+ • Verify wiring connection

59 Error configuration in stabilizer condition – park • Verify the relative condition


brake

60 Error configuration in stabilizer condition – • Verify the relative condition


steering pressure

340 Error configuration in stabilizer condition – • Verify the relative condition


cylinder leaking

350 RDN error RELEASE FLOW Verify wiring connection


Redundance Error - Output of output MIDAC
MASTER PIN 9

351 RDN error activate stabilizer system [valve 1+7] Verify wiring connection
Redundance Error - Output of output MIDAC SLAVE
PIN 4

352 RDN error stabilizer down [valve 6] Verify wiring connection


Redundance Error - Output of output MIDAC SLAVE
PIN 6

353 RDN error stabilizer up [valve 5 ] Verify wiring connection


Redundance Error - Output of output MIDAC SLAVE
PIN 7

000303-MMR17X.uso.EN.00-TRI I 43 I 5 | Alarms
User manual | MIDMAC RS17-XX English

Code Description What to do

354 RDN error Fix stabilizer position [valve 8] Verify wiring connection
Redundance Error - Output of output MIDAC SLAVE
PIN 8

355 RDN error Operator presence signal to APC 200 Verify wiring connection
Redundance Error - Output of output MIDAC SLAVE
PIN 9

356 RDN error Set pressure [valve 11] Redundance Verify wiring connection
Error - Output of output MIDAC SLAVE PIN 5

000303-MMR17X.uso.EN.00-TRI I 44 I 5 | Alarms

Vous aimerez peut-être aussi